A novel anchoring system for use in a nonfusion scoliosis correction device

A novel anchoring system for use in a nonfusion scoliosis correction device

Accepted Manuscript A novel anchoring system for use in a non-fusion scoliosis correction device M. Wessels, PhD J.J. Homminga, PhD E.E.G. Hekman, MSc...

2MB Sizes 3 Downloads 44 Views

Accepted Manuscript A novel anchoring system for use in a non-fusion scoliosis correction device M. Wessels, PhD J.J. Homminga, PhD E.E.G. Hekman, MSc G.J. Verkerke, PhD PII:

S1529-9430(14)00461-6

DOI:

10.1016/j.spinee.2014.04.028

Reference:

SPINEE 55879

To appear in:

The Spine Journal

Received Date: 5 March 2013 Revised Date:

19 March 2014

Accepted Date: 16 April 2014

Please cite this article as: Wessels M, Homminga J, Hekman E, Verkerke G, A novel anchoring system for use in a non-fusion scoliosis correction device, The Spine Journal (2014), doi: 10.1016/ j.spinee.2014.04.028. This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

ACCEPTED MANUSCRIPT

Title A novel anchoring system for use in a non-fusion scoliosis correction device

RI PT

Authors

Wessels M, PhD,* Homminga JJ, PhD,* Hekman EEG, MSc,* Verkerke GJ, PhD,*†

Academic affiliation

SC

* Laboratory of Biomechanical Engineering, Faculty of Engineering Technology, University of Twente,

M AN U

The Netherlands

† Department of Rehabilitation Medicine, University of Groningen, University Medical Center Groningen, The Netherlands

Corresponding author

HR W 207, PO box 217

TE D

Martijn Wessels

EP

7500 AE Enschede, The Netherlands

AC C

Email: [email protected]

Tel: 0031 53489 2351

ACCEPTED MANUSCRIPT

Abstract

2

Background context

RI PT

1

Insertion of a pedicle screw in the mid and high thoracic region has a serious risk of facet joint

4

damage. Since flexible implant systems require intact facet joints, we developed an enhanced fixation

5

that is less destructive to spinal structures. The XSFIX is a posterior fixation system that uses cables that

6

are attached to the transverse processes of a vertebra.

7

Purpose

9

M AN U

8

SC

3

Purpose of this study is to determine whether a fixation to the transverse process using the XSFIX is strong enough to withstand the loads applied by the XSLATOR (a novel, highly flexible non-fusion implant system) and thus whether it is a suitable alternative for pedicle screw fixation.

11

Study Design/Setting

12

TE D

10

The strength of a novel fixation system using transverse process cables was determined and compared to the strength of a similar fixation using polyaxial pedicle screws on different vertebral levels.

14

Methods

EP

13

Each of 58 vertebrae, isolated from four adult human cadavers was instrumented with either a

16

pedicle screw anchor (PSA) system or a prototype of the XSFIX. The PSA consisted of two poly-axial

17

pedicle screws and a 5 mm diameter rod. The XSFIX prototype consisted of two bodies that were fixed to

18

the transverse processes, interconnected with a similar rod. Each fixation system was subjected to a

19

lateral or an axial torque.

AC C

15

1

ACCEPTED MANUSCRIPT

1 2

Results The PSA demonstrated fixation strength in lateral loading and torsion higher than required for use in the XSLATOR. The XSFIX demonstrated high enough fixation strength (in both lateral loading and

4

torsion), only in the high and mid thoracic region (T10-T12).

5

Conclusions

6

RI PT

3

This experiment showed that the fixation strength of XSFIX is sufficient for use with the XSLATOR only in mid and high thoracic region. For the low thoracic and lumbar region, the PSA is a more rigid

8

fixation. Since the performance of the new fixation system appears to be favorable in the high and mid

9

thoracic region, a clinical study is the next challenge.

11

M AN U

10

SC

7

Introduction

Pedicle screws are commonly used for posterior anchoring of spinal implants.1-3 Although pedicle screws are now commonly used on the lumbar and thoracic spine,2 the shape of the pedicles in the

13

thoracic region is inconsistent, especially in spinal deformities such as scoliosis.4,5 As a result of this,

14

pedicle screw placement is very challenging in this region.6-9 Currently, pedicle screws are inserted by

15

using either the free hand technique (as in the Roy-Camille procedure), the open lamina technique

16

where a partial laminectomy is performed, or under application of an image-guidance system. In each of

17

these techniques, required accuracy is very high.10,11 The pedicle entry point for a high or mid thoracic

18

pedicle screw is situated at the base of the superior facet.12,13 Even a well-placed pedicle screw may still

19

violate the facet joint. Particularly in the thoracic region, the size and position of the screw head can

20

result in obstruction of movements of the facet joints. Facet joint violation in pedicle screw anchoring in

21

the thoracic and lumbar region is a common occurrence in spinal surgery.14,15 The effect of limited

AC C

EP

TE D

12

2

ACCEPTED MANUSCRIPT

motion of the joints, such as in spondylodesis is often desirable , but particularly troublesome in the

2

recent trend toward dynamic, non-fusion systems that aim to maintain the flexibility of the spine.

3

Consequently, a demand for an enhanced fixation, that is less destructive to spinal structures, has grown.

4

RI PT

1

For these reasons, a novel fixation system, referred to as XSFIX, was developed; a bridge

construction that consists of two anchors, with a bridge in between. Each of these anchors consists of a

6

connector that is attached to a transverse process by means of a cable. This new fixation system can be

7

applied for non-fusion spinal implant systems that will correct spinal deformities e.g. scoliosis. The

8

XSLATOR, a non-fusion scoliosis correction system which is currently being developed (Figure 1), is able

9

to lengthen (cranially-caudally) to allow for spinal growth and daily motion (including flexion/extension)

M AN U

SC

5

by making use of sliding bearings. The axial forces (caused by all spinal motion including growth) and

11

sagittal forces (caused by flexion/extension) remain extremely low, while lateral force and axial torque

12

are significant but low.16 To consider the XSFIX as a solid fixation, the anchor must be able to transfer the

13

forces and moments forces delivered by the XSLATOR, which are (including a design safety factor of 2) a

14

maximal lateral force of 200 N and a maximal axial torque (around the local spinal longitudinal axis) of 7

15

Nm. 16 These loads are smaller than the loads applied by fusion systems, because the non-fusion implant

16

systems use the visco-elastic properties of the spine to correct the deformity.

EP

TE D

10

Purpose of this study is to determine whether our new XSFIX fixation system is strong enough to

18

withstand the loads applied by the XSLATOR and thus whether it is a suitable alternative for a similar

19

bridge construction that uses pedicle screws.

20

Materials and Methods

AC C

17

3

ACCEPTED MANUSCRIPT

1

60 Vertebrae (42 thoracic and 18 lumbar; table 2) were isolated from four adult human cadavers (aged between 84 and 98 years) and cleaned of all soft tissues. All specimens were kept frozen until

3

testing. Prior to the experiment, all vertebrae were DXA-scanned to determine the bone mineral density

4

scores. All spines were identified as osteoporotic (Table 1); T-scores were well below -2.5.17

5

RI PT

2

Each vertebra was instrumented with either a pedicle screw anchor (PSA) system or a prototype of the XSFIX system (Table 2). The PSA consisted of two poly-axial pedicle screws (one in each pedicle) that

7

were rigidly interconnected with a 5 mm diameter rod. The XSFIX prototype consisted of two stainless

8

steel bodies that were each fixed to the left and right transverse processes of the vertebrae with black

9

Nylon 6-6 cable ties (5mm wide, tensile strength 250 N), and were rigidly interconnected with a 5 mm

M AN U

SC

6

10

diameter rod (Figure 2). Tension in the cable ties was consistently applied (at 80 N) using a cable tie

11

installation tool.

12

Before testing, each vertebra was embedded with PMMA bone cement in a metal housing that was rigidly clamped onto the testing setup. Each fixation system was subjected to one of two loading

14

conditions until failure.

TE D

13

A lateral force (in the local coordinate system) was applied to the bridge, with a loading rate of

16

1.0 N/s. An axial torque (in the local coordinate system) was applied to the bridge, with a loading rate of

17

0.05 Nm/s (Figure 3). In the test setup a narrow or a wide fork construction was used to generate torque,

18

depending on the available space (Figure 4).

20

AC C

19

EP

15

A hydraulic tensile/compression machine (MTS, Berlin, Germany) was used to generate the appropriate loading on the bridge. Time and loads were recorded digitally with a sample rate of 2 Hz. The

4

ACCEPTED MANUSCRIPT

1

vertebrae from the four spines were alternately assigned to a type of fixation and to a loading condition

2

(Table 2).

Failure was defined as the point at which the displacement suddenly increased (as shown by the

4

rotation-time graphs (Figure 5a and Figure 5b). The absolute maximal force for each construct was also

5

measured, which represented the point of complete separation of the fixation system from the vertebra.

6

However, as clinically relevant failure clearly started long before that point, the initiation of failure seems

7

a more relevant strength criterion.

SC

RI PT

3

For statistical analysis, no difference in vertebral strength between spines was assumed. Results

9

were analyzed based on four regions: high thoracic (HT): T1-T5; mid thoracic (MT): T6-T10; low thoracic

10

(LT): T11-T12 and lumbar (L): L1-L5. The anatomy of T11 and T12 was markedly different from the other

11

thoracic vertebrae by poorly articulated transverse processes (which are essential for attaching the XSFIX

12

system). Because of this, T11 and T12 represent a separate group. Fixation failure strengths for torque

13

and lateral force were compared between vertebral levels and between fixation techniques using

14

ANOVA (with significance set at p=0.05).

TE D

M AN U

8

The overall performance of XSFIX and PSA was assessed by assigning three grades to each spinal

16

region. The grades ‘poor’, ‘good’ and ‘very good’ were used to evaluate on fixation strength, facet joint

17

damage, and placement. Fixation strength was assessed using the mean (Mean) and standard error (SE)

18

values for each VB level.

AC C

EP

15

19

For each loading condition, the fixation strength was graded:

20



‘poor’ if Mean – SE < RV, 5

ACCEPTED MANUSCRIPT

1



‘good’ if 1 ·RV ≤ Mean - SE < 2·RV, and

2



‘very good’ if 2·RV ≤ Mean – SE,

where RV is the required value (RV = 200 N for lateral force and RV = 7 Nm for torsion).

RI PT

3

For all four VB levels, the lowest grades of PSA and XSFIX, which represent the limiting factor in

5

fixation strength, were selected. For example, if a grade for XSFIX at the LT level was ‘poor’ for torsion

6

and ‘sufficient’ for lateral force, then fixation strength was rated ‘poor’.

SC

4

Facet joint damage was graded:

8



‘poor’ if the facet joint was intrinsically damaged,

9



‘good’ if the facet joint remained unharmed but with minor risk of joint obstruction, and

10



‘very good’ if no damage or possible obstruction occurred.

11

Placement of the fixation was graded:

12



‘poor’ if fixation was not possible,

13



‘good’ if placement was possible and straightforward but with risk of structural damage, and

14



‘very good’ if placement was very easy with very low risk of structural damage.

16

TE D

EP

AC C

15

M AN U

7

Results

The PSA typically showed slow migration of the screws until the structures ruptured completely. The

17

XSFIX generally showed large deformation of the vertebral arc before actual rupture of the structures.

18

Both fixation types typically show a point of sudden increase in rotation (or displacement), see Figure 5a

6

ACCEPTED MANUSCRIPT

1

and Figure 5b, indicating an early collapse of bone structures, defining the clinically relevant point of

2

fixation failure.

No differences in fixation strength for either spinal region were found between spines loaded with

4

lateral force (p>0.445) or torsion (p>0.13). General type of failure for XSFIX in all spines was transverse

5

process failure in axial torsion and pedicle failure in lateral loading. General types of failure for PSA in all

6

spines were pedicle screw migration and vertebral body rupture (Figure 6). Anchoring the XSFIX system

7

to T11 and T12 was problematic and sometimes impossible as the cables slipped off the transverse

8

processes when tightened. Although mean strength difference between XSFIX and PSA in both lateral

9

loading and in axial torsion appeared large, differences showed no significance due to the small sample

11

SC

M AN U

10

RI PT

3

size in the LT region.

In lateral loading, both the PSA and the XSFIX demonstrated fixation strengths above 200 N that are typically expected for a non-fusion implant system such as the XSLATOR. The fixation strength of the

13

XSFIX is significantly lower than that of the PSA in both the high thoracic region (p=0.019) and the lumbar

14

region (p=0.045). In the mid and low thoracic levels the mean strength of XSFIX is not significantly

15

different from the strength of the PSA system (p=0.52, Table 3 and Figure 7). The small sample size of the

16

LT region level causes the mean difference to be not significant (Figure 7a).

EP

In all spinal regions, the PSA demonstrated fixation strengths in torsion above 7 Nm, which was

AC C

17

TE D

12

18

estimated to be maximally generated by the non-fusion implant system. The XSFIX demonstrated fixation

19

strengths in torsion above 7 Nm only in the high and mid thoracic regions. In the lumbar region, the

20

fixation strength of the XSFIX significantly lower than that of the PSA (p=0.015) and lower than 7 Nm. In

21

the high and mid thoracic levels the mean strength of the XSFIX in torsion is not significantly different 7

ACCEPTED MANUSCRIPT

from the strength of the PSA system (p=0.989). The fixation strength (torsion) of the XSFIX was

2

significantly lower in the lumbar region compared to both the high thoracic (p=0.024) and mid thoracic

3

region (p=0.041, Table 3 and Figure 7b). In the mid thoracic region, mean fixation strength difference (in

4

torsion) is large, however not significant caused by small sample size (Figure 7b).

5

In Table 4, a schematic overview is given of the assessment of the XSFIX and PSA in respect to the

6

different spinal regions. Figure 8 shows the cumulated scores of both fixation systems for each vertebral

7

level. Evidently, the scores of the PSA are higher than the XSFIX for the low thoracic and lumbar regions.

8

The XSFIX shows higher accumulated scores than the PSA for the high and mid thoracic regions.

9

Discussion

SC

M AN U

10

RI PT

1

Some limitations apply to this study. First, all tested vertebrae were osteoporotic; the results may differ for healthy vertebrae. However, both the PSA and the XSFIX obtain their strength from bone

12

structures of the vertebral arc, which are relatively unaffected by osteoporosis.18 Furthermore, this test

13

can be seen as a worst-case scenario. It is reasonable to assume that the relative performance of the

14

new fixation system in relation to a similar fixation using pedicle screws is comparable to the

15

performance in adolescent bone.

EP

Second, both the XSFIX and the PSA were placed without the physical limitations that are present

AC C

16

TE D

11

17

during in vivo surgery. Pedicle screws were placed using visual cues not available during surgery. The

18

XSFIX was also easier to attach, due to the removal of the ribs. If the clinical reality prevents proper

19

placement of these systems, failure strengths may be different.

8

ACCEPTED MANUSCRIPT

1

Third, in this experiment the pedicle screw fixation (PSA) consisted of a bridge construction. Clinically, pedicle screws are used as stand-alone systems. A single screw will transfer a torque directly to

3

the vertebra, resulting in high peak pressures and increased risk of early fixation failure. Our test results

4

for the PSA cannot be compared to single pedicle screw arrangements; though it can be stated our

5

results will give much higher strength values.

6

RI PT

2

We defined failure strength as the point at which a sudden increase in displacement (rotation) took place. Because the PSA showed migration of the screws and the XSFIX showed large unrecoverable

8

deformation of the vertebral arc before actual rupture of the structures, we attribute the sudden

9

increase in deformation to microscopic bone failure. Nonetheless, the failure strength criterion we used

11

M AN U

10

SC

7

seems clinically most relevant.

In some cases, premature XSFIX failure occurred in the small ratchet of a cable tie. An improved version of the ratchet that consists of high strength materials will prevent early fixation failure. In

13

addition, the material aspect needs further attention because the material used in our ties (Polyamide

14

6/6) is not suitable for implantation due to issues with creep, relaxation, and biocompatibility. Replacing

15

the material for a creep resistant and biocompatible material such as carbon fiber reinforced PEEK

16

(polyetheretherketone) may well be considered.19,20

EP

TE D

12

In the high and mid thoracic regions the torsion strength of the XSFIX is higher than the required

18

value of 7 Nm for use with the non-fusion implant system such as the XSLATOR. In the low thoracic and

19

lumbar regions, the torsion strength of the XSFIX is lower than the required 7 Nm, but in these regions,

20

the PSA is strong enough. Lateral fixation strength of the XSFIX meets the requirement of 200 N for use

21

with the non-fusion implant system for all spinal regions. However, the anchoring of the XSFIX to T11 and

AC C

17

9

ACCEPTED MANUSCRIPT

T12 is problematic, due to the poorly pronounced transverse processes of those vertebrae.

2

Consequently, the XSFIX is not very suitable for those vertebrae. Similarly, the lumbar vertebrae have

3

long transverse processes that are very slender and fragile, making the lumbar vertebrae not suited for

4

use of the XSFIX. In fact, some of the lumbar transverse processes broke during tensioning of the cable

5

ties. Fortunately, pedicle screws can be inserted easily into these vertebrae without obstructing or

6

damaging the facet joints.

RI PT

1

The highly flexible implant XSLATOR requires three vertebral levels for fixation.16 Either PSA or XSFIX

8

must be used for anchoring. The force requirements for the fixation are based on the maximal forces on

9

the center vertebra. The actual force on the other two vertebrae is only 50% of the referred maximal

10

force. The XSLATOR therefore cannot be compared to other (fusion or non-fusion) scoliosis correction

11

constructs.

M AN U

SC

7

The grading system used in table 4 is rather subjective. However, the grading system was used

13

merely to illustrate our conclusions. Despite the subjectivity of the grading, we think the outcome

14

regarding comparison between both systems will be similar. Nevertheless, Table 4 shows that the XSFIX

15

is an excellent fixation for the high and mid thoracic regions. Especially in this region, the application of

16

pedicle screws is risky, because of the small diameter of the pedicles. Therefore, the XSFIX could play an

17

important role in the fixation of devices to the spine. In contrast, grades for fixation strength and

18

placement of XSFIX in the low thoracic and lumbar region are valuated much lower, whereas PSA

19

produced decent grades for those regions.

20

With this experiment, we have a strong indication that the XSFIX is comparable to the pedicle screw

21

performance (strength, placement and joint damage) in the mid and high thoracic region and that for the

AC C

EP

TE D

12

10

ACCEPTED MANUSCRIPT

lumbar region the pedicle screw is the method of choice. With this outcome, animal experiments can be

2

started to gather statistical evidence that the XSFIX is indeed preferred for fixation in the mid and high

3

thoracic region. Therefore, the experiments described in this manuscript contributed to a restriction of

4

animal experiments by giving these experiments a specific focus.

5

RI PT

1

In summary, this experiment showed that the XSFIX is a solid fixation with strength that is

comparable to the PSA strength in torsion, when placed on a vertebra in the T2-T10 region. Fixation

7

strength of XSFIX is sufficient for use with a non-fusion implant system, such as the XSLATOR, in mid and

8

high thoracic regions. For the low thoracic and lumbar region, the PSA is a more rigid fixation. Since the

9

performance of the new fixation system appears to be favorable in the high and mid thoracic region, a

10

clinical study is the next challenge.

11

References

1. Steffee AD, Biscup, RS, Sitkowskj DJ. Segmental spine plates with pedicle screw fixation - a new

TE D

12

M AN U

SC

6

internal fixation device for disorders of the lumbar and thoracolumbar spine. Clin Orthop Relat Res

14

1986;203:45-53.

16

17 18

19 20

2. Gaines RW Jr. The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 2000;82:1458-76.

AC C

15

EP

13

3. Cotrel Y, Dubousset J, Guillaumat M. New universal instrumentation in spinal surgery. Clin Orthop Relat Res 1988;227:10-23.

4. Zeiller S, Lee J, Lim M, Vaccaro A. Posterior thoracic segmental pedicle screw instrumentation: Evolving methods of safe and effective placement. Neurology India 2005;53:458-65. 11

ACCEPTED MANUSCRIPT

3 4

5 6

7

instrumentation: critical appraisal of current "state-of-art". Neurology India 2009;57:715-21.

6. Vaccaro AR, Rizzolo SJ, Allardyce TJ, et al. Placement of pedicle screws in the thoracic spine. Part I:

RI PT

2

5. Mattei T, Meneses M, Milano J, et al. "Free-hand" technique for thoracolumbar pedicle screw

Morphometric analysis of the thoracic vertebrae. J Bone Joint Surg Am 1995;77:1193-9.

7. Vaccaro A.R., et al. Placement of pedicle screws in the thoracic spine. Part II: An anatomical and radiographic assessment. J Bone Joint Surg Am 1995;77:1200-6.

SC

1

8. Xu R, Ebraheim NA, Ou Y, et al. Anatomic considerations of pedicle screw placement in the thoracic spine. Roy-Camille technique versus open-lamina technique. Spine 1998;23:1065-8.

9

9. Lonstein JE, et al. Complications associated with pedicle screws. J Bone Joint Surg Am

13 14

15 16

17 18

TE D

12

10. Rampersaud YR, Simon DA, Foley KT. Accuracy requirements for image-guided spinal pedicle screw placement. Spine 2001;26:352-9.

11. Papadopoulos EC, Girardi FP, Sama A, et al. Accuracy of single-time, multilevel registration in image-guided spinal surgery. Spine J 2005;5:263-7.

EP

11

1999;81:1519-28.

12. Chung KJ, Suh SW, Desai S, Song HR. Ideal entry point for the thoracic pedicle screw during the

AC C

10

M AN U

8

free hand technique. Int Orthop 2008;32(5): 657-662.

13. Modi HN, Suh SW, Fernandez H, et al. Accuracy and safety of pedicle screw placement in neuromuscular scoliosis with free-hand technique. Eur Spine J 2008;17(12):1686-1696.

12

ACCEPTED MANUSCRIPT

1

14. Chung KJ, Suh SW, Swapnil K, et al. Facet joint violation during pedicle screw insertion: a cadaveric study of the adult lumbosacral spine comparing the two pedicle screw insertion techniques. Int

3

Orthop 2007;31:653-656.

4

RI PT

2

15. Chen Z, Zhao J, Xu H, et al. Technical factors related to the incidence of adjacent superior

segment facet joint violation after transpedicular instrumentation in the lumbar spine. Eur Spine J

6

2008;17(11):1476-1480.

8

9 10

16. Wessels M, Hekman EEG, Verkerke GJ. Mechanical behavior of a novel non-fusion scoliosis correction device. J Mech Behav Biomed 2013;27:107-114.

M AN U

7

SC

5

17. WHO. Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. Report of a WHO Study Group. World Health Organ Tech Rep Ser 1994;843:1-129.

18. Riggs BL. Overview of osteoporosis. West J Med 1991;154:63-77.

12

19. Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials

14

2007;8:4845-69.

20. Toth JM, et al. Polyetheretherketone as a biomaterial for spinal applications. Biomaterials 2006;27:324-34.

16

Figure Legends

17

AC C

15

EP

13

TE D

11

Figure 1: The XSLATOR consists of two implants that are posteriorly fixed to the vertebrae. The three

18

bridges can relatively slide axially keeping axial forces extremely low. The bridges are fixed to the spine

19

using transverse process cables (XSFIX; right, top) or 2 pedicle screws (right, bottom). 13

ACCEPTED MANUSCRIPT

2

3 4

5

Figure 2: The two different fixations were attached to the vertebrae and the vertebrae were fixed in cups with PMMA. (Left) Vertebra with the XSFIX. (Right) Vertebra with the PSA in setup.

Figure 3: Loading with a vertical (lateral) force on XSFIX (left), PSA (middle), and torque around a

RI PT

1

vertical axis on PSA (right) was performed using a hydraulic tensile/compression machine.

Figure 4: The cemented cups were clamped into a framework, which was clamped onto a table. Torque was applied either between the cables (left) using a narrow rotating fork or on the ends of the

7

rod (right) using a wide rotating fork.

Figure 5a: Typical rotation-time curve of a XSFIX fixation. The rotation-time curve shows a sudden

M AN U

8

SC

6

9

increase in rotation at time=385 s, indicating a collapse of structures. At time=488 s one of the

10

transverse process is separated from the vertebral arc resulting in a sharp increase in rotation.

11

Figure 5b: Typical rotation-time curve of a PSA fixation. Also here, at time=998 s, a sudden increase of rotation can be seen in the curve (defined as the failure point). At time=1851 s, the vertebral body

13

ruptured.

15

Figure 6: (Left) Transverse process rupture, indicated with a white arrow, in torsion test with XSFIX.

EP

14

TE D

12

(Right) Vertebral body rupture in torsion test with PSA (black arrow).

Figure 7: High-low plots with mean failure loads in lateral loading (a) and axial torsion (b) at the

17

different vertebral levels. The symbol * indicates significant difference. Dashed lines indicate lower limit

18

values required for use in the non-fusion implant system.

19 20

AC C

16

Figure 8: The accumulated scores (for individual scores, see table 4 ) of the XSFIX are higher for the high and mid thoracic level but lower for the low thoracic and lumbar level. 14

ACCEPTED MANUSCRIPT

Table 1: T-scores taken from DXA scans show osteoporosis in all spines. T-score

Gender

Age

80

T6-T12

L1-L5

1

-4.4

-3.6

Male

2

-3.3

-4.1

Male

3

-4.8

-4.3

Male

4

-6.4

-5.5

Female

RI PT

Spine

84

AC C

EP

TE D

M AN U

SC

98

92

ACCEPTED MANUSCRIPT

Table 2: Loading types (‘F’ for lateral force and ‘T’ for torque) and fixations (XS-FIX/PSA) were altered between different spines and spinal levels. Four different spinal levels were defined for comparison: high thoracic (HT), mid thoracic (MT), low thoracic (LT), and lumbar (L).

S1

F

S2

T

S3

T4

T6

T7

T8

T

T

T

F

F

T

F

T

S4

T5

F F

T9

T10

T

T

T11

T

F

F

F

S3

T

T

F T

F

F

T

T

T

F

EP

T

T

F

F

L3

L4

L5

F

F

T

T

T

TE D

S4

F

L2

T

T F

L1

T

F

F

S2

T12

T T

S1

AC C

PSA

T3

L

M AN U

XS-FIX

T2

LT

RI PT

T1

MT

SC

HT

T T

T F T

T

F

F F

T

F

T T F

ACCEPTED MANUSCRIPT

Table 3: Mean fixation strengths and standard errors of the two fixations at all different vertebral levels. HT

L

Mean

SE

Mean

SE

Mean

SE

Mean

SE

XS-FIX

341.3

61.3

402.2

97.8

317.5

112.5

451.0

79.1

PSA

702.7

92.1

506.7

60.7

594.6

28.8

631.8

14.1

XS-FIX

13.9

2.2

12.3

1.8

5.0

1.8

4.5

1.5

PSA

9.6

1.9

13.1

3.4

14.0

4.0

16.2

2.8

AC C

EP

TE D

M AN U

SC

Torque [Nm]

LT

RI PT

Force [N]

MT

ACCEPTED MANUSCRIPT

Table 4: Evaluation of fixations on basis of three criteria (strength, facet joint damage, and placement) shows that the XS-FIX scores well on all criteria in the high and mid thoracic regions. The PSA scores best in the low thoracic and lumbar regions. Grades used for assessment are ‘poor’ (− or -1), ‘good’ (+ or 1) and ‘very good’ (++ or 2).

Fixation strength

+(1)

+(1)

−(-1)

−(-1)

Facet joint damage

+(1)

+(1)

+(1)

+(1)

Placement

+(1)

+(1)

−(-1)

+(1)

Fixation strength

+(1)

+(1)

+(1)

+(1)

Facet joint damage

−(-1)

−(-1)

+(1)

+(1)

Placement

+(1)

+(1)

+(1)

++(2)

SC

M AN U

TE D

L

RI PT

LT

EP

PSA

MT

AC C

XS-FIX

HT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT