A Quantitative Method for Measuring labeled Compounds With Whole-Body Autoradiography in Tissue Sections
EEVA-LIISASAINIO AND PERT-~SAINIO
Whole-body autoradiography is an effective method for localizing labeled compounds in various organs. However, the technique is limited in its ability to quantify such material. Using tissue sections, this study investigated certain parameters involved in the quantitative estimation of labeled compounds by whole-body autoradiography. These included correlation between thickness of the section and radioactivity counted, the precision of such measurements, and the reproducibility of the autoradiographic films as tested by image analysis transmission. The precision of radioactivity measurements using tapes with a tissue section or a “punch biopsy” (punching off a piece of tissue from the section) was compared. The results revealed excellent linearity between the thickness of the section and the radioactivity counted (r = 0.97) when section thickness was IO-30 km. The measurement precision using tapes was better than with the “punch specimen” method. The reproducibility of photographic films was good when transmission was measured by image analysis. It was concluded that a thickness of 30 km is ideal for use in whole-body autoradiographic studies. It appeared that radioactivity measurement of tissue sections on tapes was superior to direct measurement from organs. Image analysis was employed and statistically evaluated for the first time in this study, and the promising findings suggest that it is likely to become the method of choice for future studies of this type. Key Words: Densitometric
Whole-body; Autoradiography; Quantitation; image analysis; Labeled compounds
Tissue thickness;
INTRODUCTION
Whole-body autoradiography (WBA), first described by Sven Ullberg in the early 195Os, has proved a very useful technique for localizing labeled chemical compounds in animal studies. WBA has been employed in pharmacology, toxicology, nutrition research, and occupational health, among other disciplines. The technique makes it easy to recognize the tissues with particular affinity for the compound and/ or its metabolites under study, so that further procedures like histopathological characterization or biochemical tests can be focused on relevant organs or tissues.
From the Departments of Pharmacology and Toxicology (E.L.S.) and Oral Pathology, Roentgenology and Forensic Odontology (P.S.), University of Kuopio, P.O. Box 6, SF-70211, Kuopio, Finland. Address reprint requests to: P. Sainio, Department of Oral Pathology, Roentgenology, and Forensic Odontology, University of Kuopio, P.O. Box 6, SF-70211 Kuopio, Finland. Received August 1990; revised and accepted December 1990. 53 journalof
Pharmacological
Methods
0 1991 Elsevier Science Publishing
26, 53-59 (1591) Co., Inc.. 655 Avenue of the Americas, New York, NY loo10
54
E.-L. Sainio and P. Sainio In some countries, WBA is now a mandatory technique in trials for drug registration. Despite its superiority in detecting labeled compounds in tissues, however, WBA is unable to quantify such compounds satisfactorily. Quantification of tissue loading has often been expressed in the form low (+), moderate (+ +) or high (+ + +), to indicate
the degree
of affinity
based on the use of standards
of a compound containing
al., 1974) or commercial methacrylate (Irons et al., 1981) have been reported.
for a tissue.
different
Quantitative
methods
levels of radioactivity
(Cross et
sources of increasing 14C-specific activity The radioactivity of tissue samples can also
be measured with a liquid scintillation counter following combustion of the tissue. These values are correlated to concentrations on the x-ray film (Busch, 1977). In the present study, a quantitative WBA method with particular attention focused on the following:
was developed
1. The degree
of varying
radioactivity, 2. Comparison
of linearity
between
of the coefficients
the tape or “punch 3. The reproducibility
tissue
of variation
biopsy” methods, of exposed photographic
tissue when transmission density metric image analysis of exposed first time. 4. Summarizing MATERIALS
and drawing
AND
sections
in measuring
and evaluated,
thickness
tissue
and their
radioactivity
films in serial sections
by
of the same
was determined by image analysis. Densitofilms was employed for this purpose for the
conclusions
on the use of this WBA method.
METHODS
Test Animals and Chemicals The experimental under
regularly
animals
alternating
were
male outbred
periods
Han:
NMRI
mice (30 g), maintained
of light (14 hr) and darkness
(IO hr) and with
free access to food (standard diet from Hankkija Ltd, Turku, Finland) and water. Carboxyl-14C nicotinic acid (specific activity 53 mCi/mmol, code CFA.197) was purchased from Amersham, U.K., carboxymethylcellulose Department of Pharmacy, Kuopio, Finland (Ph.Nord and ethylether from E. Merck, Darmstadt, Germany.
Experimental Carboxyl-14C vein into mice.
(CMC) from the University grade), and Hexan (n-hexane)
Procedures nicotinic acid (5 FCi) in 0.9% NaCl (100 ~1) was injected via a tail After 30 min, the animals were anesthetized with ethylether. They
were then laid right side down in a shallow container constructed frame containing 1.5% CMC (w/v). The container was immediately
of a floor and a lowered into a
chamber containing hexane cooled with dry ice. The temperature was -74°C and freezing was complete in about 30 min. The solidly frozen specimens were allowed to stabilize overnight at -20°C before sections were cut. Sagittal sections from the whole animal were taken at -20°C by microtome (PMV Cryo Microtome 450MP, Stockholm) of several tissue thicknesses, using adhesive tape (Scotch Brand Magic Transparent Tape, No. 810, 3M). The sections were al-
A Quantitative lowed
to dry for 24 hr at -20°C
(Hyperfilm TM-B 5-week exposure. Sample
they were
placed
max, code RPN. 9, Amersham, made The films were developed by normal
Preparation
A cylinder
before
Method for Whole-Body
for Radioactivity
with a round,
Determinations
sharp edge (diam.
55
Autoradiography
on X-ray sensitive
film
by CEA Ltd, Sweden) for a photographic processing. from Sections
on Tape
3.829 mm) was used to cut out circular
pieces of tissue section from the tape (Figure I). The cutting procedure was performed on a cork base plate. A magnification glass (4 x ) was used for small biopsy areas. The specimens were placed in a liquid scintillation bottle and digested with 0.5 mL Lumasolve
(Lumac,
Netherlands)
overnight
at +37”C.
The following
day,
scintillation liquid (4 mL) (Lipoluma, Lumac, Netherlands) was added to the bottles and the radioactivity was immediately counted with a LKB Wallac 1216 Rackbeta liquid
scintillation
Direct Small surgical
counter.
Determination
of Tissue Radioactivity
pieces of frozen knife into liquid
Processing
and counting
Densitometric
Image
animal tissue (about 5-10 mg) were punched off with a scintillation bottles for digestion with 0.5 mL Lumasolve. were
as described
above.
Analysis
The autoradiograms were digitalized through a CCD video camera (TK870E, JVC, Japan) fitted with a mace bus model H7-C611AF and joined to an IBAS image analyzer UBAS 1, 2, Kontron,
Munich,
Germany),
and were
displayed
on a monitor
resolution
of 512 x 512 pixels and 256 shades of grey. The IBAS analyzer
FIGURE 1. the tape.
Sample taking by the tape method.
with
a
evaluates
1 = cylinder used, 2 = piece of tissue on
56
E.-L. Sainio and P. Sainio the mean
value
of gray shades
in the measured
area.
ograms as a standard the IBAS analyzer was calibrated 100 (background). Other values were linearily scaled
Using
one of the autoradi-
to a scale of 0 (darkest) to between these values. The
autoradiograms were scanned on a light box and the focus and exposure kept standardized. The imbalance of the light source was corrected. The percentage transmission
was measured
using a standardized
circle on the screen
placed
within
the
area occupied by the tissue. The radius of the circle was 8 pixels with a mean transmission value of 105 pixels within the measuring frame. The measured tissue area was the same in all sections. Large blood vessels and capsular avoided. Four measurements were carried out for each point.
structures
were
Statistical Analysis The results were analyzed with Student’s independent two-tailed t-test. P < 0.05 was taken as significance point. The percentage coefficient of variation (CV%) was calculated
accordingto
determinations
the formula
CV%
= 100 x
SD/x,
where
n =
number
of
and X is the mean.
RESULTS There was a strong linear relationship
between
the thickness
of sections
and their
radioactivity count (r = 0.97). The precision of the measurements is indicated by the coefficients of variation (CV%) shown in Table 1. The CV varied from 5.7% (liver cranial, 30 pm thickness) to 11.2% (salivary gland, IO pm). In addition, results from two different liver locations (liver cranial and liver caudal) were compared statistically, revealing no significant (P < 0.05) variation in radioactivity between them according to measurements performed on 10 different WBA sections from each location (results not tabulated). The reproducibility of radioactive tested
by measuring
the transmission
tions of the same thickness
TABLE 1
exposures
on the autoradiographic
percentages
from the same organ.
from five successively The transmission
Radioactivity in Different Tissues and the Corresponding
THICKNESS or SECTION (I&
10 IJ. 10 CL 10 I* 20 CL 20 P 20 P 30 W 30 CL 30 CL
SAMPLE/LOCALIZATION IN MICE Liver cranial Liver caudal Salivary gland Liver cranial Liver caudal Salivary gland Liver cranial Liver caudal Salivary gland
films was taken
CV% Values
MEAN + S.D.’ (DPM) 241 2 245 i 45 2 425 ? 461 ? 68 f 595 ? 624 ir 9426
19 20 5 32 38 7 34 12
sec-
measurements
CV% 7,6 8,1 11,2 7,4 8,1 IO,2 5,7 890 6,s
’ The measurements were performed according to the method described for determination of the radioactivity on tape, under material and methods (n = IO).
A Quantitative
Method for Whole-Body
57
Autoradiography
100 -
._ ; I ,m c-
40 20 .-
03
2
1
3
4
5
Organs FIGURE 2. Reproducibility of exposed autoradiographic films. Transmission was measured by densitometric image analysis on films obtained from five successively cut sections that were placed on films and exposed for 6 weeks. The value in bars are mean f S.D. Organs: 1 = liver, CV% = 2.6; 2 = testis, CV% = 5.2; 3 = salivary glands, CV% = 7.8; 4 = wall of small intestine, CV% = 4.6; 5 = seminal vesicle, CV% = 11.0. were
performed
CV variation
by densitometric
from 2.6%
(liver)
image
analysis.
to 11 .O% (seminal
The
result
(Figure
2) showed
a
vesicle).
The coefficient of radioactivity variation in three different organs was determined by punching off minute amounts of the tissue into a liquid scintillation bottle and measuring the radioactivity. The findings (Table 2) were CVs of 9.6% 8.6% in the small intestine and 30% in the salivary gland.
in the liver,
DISCUSSION Statistical
comparison
of radioactivity
measurements
by the two different
methods
of tissue collection revealed that the use of tape was accompanied by smaller coefficients of variation (5.7% for liver-11.2% for salivary gland) than when direct punching off was used (9.6%-30%, respectively). The better result with tape is due to the fact that a constant amount of tissue was always being measured. Direct determination of radioactivity in tissue samples can often suffer from problems such as variable
chemiluminescence,
which
means
that different
sizes of tissue sample
TABLE 2. Radioactivity Measured From a Punched Off Piece of Tissue. Details Described Under Material and Methods (n = 10) ORGAN Liver Small
intestine
Salivary
glands
DPM/~Y+GTISSUE
CV%
1890 f 182 226 c_ 26
986 II,5
152 + 46
30,o
58
E.-L. Sainio and P. Sainio can give non-comparable results. The use of a cylindrical cutting form for tissue sample-taking provides the advantage of accurate comparison of radioactivity between samples from different organs. The tissue biopsy cylinder of 3.8 mm in diameter was specially designed for our study purpose. Commercially manufactured skin biopsy punches of 2, 3, 4, and 6 mm in diameter are also available (Stiefel Laboratories Ltd, UK). The linear
relationship
between
tissue thickness
and radioactivity
means that sec-
tions thicker or thinner than 20 pm can be used. A thickness of 30 km seems ideal on the basis of our experience and findings. The use of thicker sections than 20 Frn shortens the advantage
the exposure
time needed,
of decreased
gives a more distinct
background.
Previously,
20 km
picture
and also has
has been
considered
ideal for such measurements (Irons and Gross, 1981). WBA has often been criticized for not being able to distinguish the administered compound from its metabolites. It may well be possible to use thicker sections and to apply analytical methods to identify and measure metabolites further. The densitometric measurements in serial sections made from the photographic films showed that the standard deviations fell within better earlier study (Yonekura et al., 1983). Our results conformed ings that, provided adequately tested hardware is used and strument calibration and adjustment are made, the analytic sitometry different
limits compared
to an
with earlier study findsuitable checks on inaccuracy of videoden-
is high (McEachron et al., 1988). Our study showed that radioactivity at tissue levels on separate films is totally comparable in the same animal.
The present study also agrees with the finding that the lower the absorbance are, the higher the coefficients of variation (Schweitzer et al., 1987). Our mental compound was nicotinic would give similar results. The mathematical relationship
acid.
However,
between
it is likely
radioactivity
that other
in tissues
and
values experi-
chemicals autoradi-
ographic image density has been shown to be linear using polymethacrylate standards (Irons and Gross, 1981). Thus, in principle it should be possible to obtain exact quantitation of radioactivity in $i/wt of tissue by measuring radioactivity in organs with this tape section method, correlating these values with the corresponding transmission values and drawing a standard curve. However, this is troublesome and time consuming and earlier results with this method have not been satisfactory (Franklin, 1983). Cross et al. (1974) have introduced
the idea of constructing
a scale of different
concentrations of radioactivities in tissues. This has promoted development of commercially manufactured standards (Amersham International plc, England). According to our experience these are quite suitable and easy to use. A good reproducibility of autoradiograms is an absolute prerequisite for using the standards. Our study showed that these requirements are fulfilled and the standards can be used for obtaining the tissue concentrations of labeled compounds in FCi/wt. Validity of brain paste standards or standards based on radioactive blood homogenates have also been reported suitable for quantitative WBA (Ito and Brill, 1990; Schweitzer et al., 1987). Our
technique
has
other
advantages
when
compared
with
the
traditional
“punched
off”
method.
A Quantitative
Method for Whole-Body
Autoradiography
measurement
of tissue radioactivity
from tapes can
Direct
serve as a preliminary guide for assessing the approximate exposure time for films. It also makes it possible to provide an estimate of the radioactivity in different organs point. This method histological It seems
so results can be treated could come to augment
realistically from the prescreening
examination or micro-autoradiography. clear that WBA will develop into the preferred
the distribution WBA has been
needed needed
the statistical viewof study targets in
method
and kinetics of chemical compounds in tissues. inhibited by the difficulties with quantification.
for determining
So far, adoption of Further advances
also depend on progress in developing more sensitive films with lower background. However, the basic principles of the method are already established. This is the first study involving examination based weight The
to further authors
thank
for preparation
statistically-handled on image analysis.
progress Mr.
quantification The promising
of results from densitometric findings will hopefully lend
in this field.
E. Pesonen
and Mr. A. Dunberg
for technical
assistance
and Mrs.
S. Oksman
of the manuscript.
REFERENCES Busch
U
(1977)
(WBAR): its
Whole-body
autoradiography
use for pilot studies
of pharmacokinet-
in rats. Acta fharmacol
Cross
SAM,
titative
Groves
method
sues sectioned Int / Appl Franklin
films
RD, Gross
ibration
icol Appl Ito T,
Brill
autoradiography.
Factors
affecting
the sensitivity autoradiog-
EA (1981) Standardization autoradiography analysis
compounds
McEachron
DL,
(1988) The
autoradiog-
radionuclides.
Appl
Ra-
Gallistel
CR, Eilbert
JL, Tretiak
Oj
analytic and functional
video densitometry
Schweitzer
system.
accuracy of a
/ Neurosci
A, Fahr A, Niederberger
ple method
and cal-
333.
for rou-
Yonekura
of the distribution
in animal tissues.
Tox-
Pharmacol59:250-256. AB (1990) Validity
whole-body
short-lived
for
Methods
(1983)
paste stan-
AB,
Quantitative
opharmaceuticals, system
of tissue
Y, Brill
Appl
j Nucl
of ?-whole-
Radiat
Som P, Bennett auoradiography Part
1:
by videodensitometry:
munication.
W (1987) A sim-
the quantitation
body autoradiograms.
13:163-169.
of whole-body
quantitative
25~63-74.
used for whole-body
tine semiquantitative of “C-labeled
in tis-
/sot 25:381-386.
raphy. Xenobiotica Irons
T (1974) A quanradioactivity
for whole-body
Radiat
ER (1983)
of X-ray
AD, Hesselbo
for
raphy using
diat /sot 41:661-667.
Toxico/41:28-29.
for measuring
dards
Med
Digital
/sot 38:329-
GW, with
Fand I radi-
film-analysis Concise
24:231-237.
com-
59