An instrument for measuring the respiration of small animals

An instrument for measuring the respiration of small animals

Life Sciences No . 10, pp . 779-781, 1963 . Pergamon Press, United States . Inc . Printed in the AN INSTRUMENT FOR MEASURING THE RESPIRATION OF SMAL...

90KB Sizes 6 Downloads 183 Views

Life Sciences No . 10, pp . 779-781, 1963 . Pergamon Press, United States .

Inc . Printed in the

AN INSTRUMENT FOR MEASURING THE RESPIRATION OF SMALL ANIMALS V . Jaclonan, W . C . Barnes, E . G . Erdöl Mellon Institute, Pittsburgh 13, Pennsylvania (Received 24 September 1963) Studies in this laboratory required a sensitive method for assaying the respiration of small animals .

For this purpose, a satisfactory and prac-

tical instrument has been developed here which may be of general interest . The device measures pressure differences in a flow transducer, and converts them into an electrical signal by means of a diaphragm and a differential transformer .

The resulting AC signal is converted to DC signal and registered on a

suitable recorder . A polyethylene tube was inserted in the trachea of rats, guinea pigs and rabbits with the help of tracheotomy . dogs through the

mouth .

The tube was connected to a flow detector which was

bored from a polystyrene rod (Fig . 200 mesh copper screen .

The tracheal tube was introduced in

la) .

The flow of air was restricted with a

The pressure differences were transmitted to the trans-

ducer by a 1/8" polyethylene tube . A sensitive beryllium copper diaphragm is used to convert the pressure differentials of the flow meter into a linear movement (Fig . lb) .

The dia-

phragm is solidly attached to the core of a differential transformer (Schaewitz 033SL) .

The amplitude of the resulting electrical signal depends on the linear

displacement of the differential transformer core .

Standard sensitivity ratios

of the differential transformer are 1 mV output signal/ .001" core movement/1 V transformer excitation . The small counterweight serves to pre-stress the diaphragm and should not be used for zeroing purposes .

Zero adjustment is done by moving a cap

screw ~rttich in turn moves the differential transformer coil into zero position . 779

780

INSTRUMENT FOR MEASURING RESPIRATION

No . 10

v 0 u tr o E cdwo w

d n n 0 t U

u T ~U 00 ~0




0

F

0

No . 10

781

INSTRUN~TP FOR N~ASURII~G RESPIRATION

The electrical circuit is quite simple (Fig . lc) .

The excitation

voltage of the differential transformer is taken from a 6 .3 V filament transformer .

A 5000 ohm 5 watt variable resistor is placed in series with the

differential transformer primary for minor sensitivity adjustments . The low voltage signal received from the differential transformer output is a 60 cycle signal which must be rectified or demodulated to obtain a DC voltage which is suitable for recording in instruments such as the Grass poly graph .

The rectification is accomplished with a Leeds and Northrup No .

chopper (see Fig . lc) .

3338-1

This device is essentially a very fast relay which

switches back and forth at the frequency of the exciting voltage (60 cycles in this case) .

The use of the chopper provides a half wave signal of pulsating

DC which is suitable for a DC recorder input and needs no filtering because the recorder input filter usually is sufficient . The 0 .5 mfd capacitor in the chopper driving circuit serves a dual purpose .

First,

it sets the phase or the time of the chopper contacts properly,

then it drops the supply voltage from 110 V to the 27 V needed to drive the chopper . During testing, sensitivities of 1 mV per 0 .6" of water pressure were easily obtained with the device and signal frequencies of 100 cps were clearly distinguished with a Grass polygraph recorder . The routine operation of the instrument is relatively simple and trouble-free .