Accepted Manuscript Application of high-resolution continuum source flame atomic absorption spectrometry to reveal, evaluate and overcome certain spectral effects in Pb determination of unleaded gasoline
Zofia Kowalewska, Hanna Laskowska, Michał Gzylewski PII: DOI: Reference:
S0584-8547(16)30210-5 doi: 10.1016/j.sab.2017.03.020 SAB 5233
To appear in:
Spectrochimica Acta Part B: Atomic Spectroscopy
Received date: Revised date: Accepted date:
14 September 2016 19 March 2017 20 March 2017
Please cite this article as: Zofia Kowalewska, Hanna Laskowska, Michał Gzylewski , Application of high-resolution continuum source flame atomic absorption spectrometry to reveal, evaluate and overcome certain spectral effects in Pb determination of unleaded gasoline. The address for the corresponding author was captured as affiliation for all authors. Please check if appropriate. Sab(2017), doi: 10.1016/j.sab.2017.03.020
This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
ACCEPTED MANUSCRIPT Application of high-resolution continuum source flame atomic absorption spectrometry to reveal, evaluate and overcome certain spectral effects in Pb determination of unleaded
RI
PT
gasoline
a
NU
SC
Zofia Kowalewskaa *, Hanna Laskowskab, Michał Gzylewskia
Faculty of Civil Engineering, Mechanics and Petrochemistry, Warsaw University of Technology,
MA
Łukasiewicza 17, 09-400 Płock, Poland
OBR JSC (Warter Group), Chemików 5, 09-411 Płock, Poland
D
b
PT E
Abstract
High-resolution continuum source and line source flame atomic absorption spectrometry (HR-
CE
CS FAAS and LS FAAS, respectively) were applied for Pb determination in unleaded aviation or automotive gasoline that was dissolved in methyl-isobutyl ketone. When using
AC
HR-CS FAAS, a structured background (BG) was registered in the vicinity of both the 217 nm and 283.306 nm Pb lines. In the first case, the BG, which could be attributed to absorption by the OH molecule, directly overlaps with the 217 nm line, but it is of relatively low intensity. For the 283 nm line, the structured BG occurs due to uncompensated absorption by OH molecules present in the flame. BG lines of relatively high intensity are situated at a large distance from the 283 nm line, which enables accurate analysis, not only when using simple variants of HR-CS FAAS but also for LS FAAS with a bandpass of 0.1 nm. The lines of the 1
ACCEPTED MANUSCRIPT structured spectrum at 283 nm can have “absorption” (maxima) or “emission” (minima) character. The intensity of the OH spectra can significantly depend on the flame character and composition of the investigated organic solution. The best detection limit for the analytical procedure, which was 0.01 mg L-1 for Pb in the investigated solution, could be achieved using HR-CS FAAS with the 283 nm Pb line, 5 pixels for the analyte line measurement and
PT
iterative background correction (IBC). In this case, least squares background correction
RI
(LSBC) is not recommended. However, LSBC (available as the “permanent structures”
SC
option) would be recommended when using the 217 nm Pb line. In LS FAAS, an additional phenomenon related to the nature of the organic matrix (for example, isooctane or toluene)
NU
can play an important role. The effect is of continuous character and probably due to the simultaneous efficient correction of the continuous background (IBC) it is not observed in
MA
HR-CS FAAS. The fact that the effect does not depend on the flame character indicates that it is not radiation scattering. For LS FAAS, the determination of Pb using the 283 nm line, a 0.1
D
nm bandpass and a fuel lean flame is strongly recommended. The analysis of certified
PT E
reference materials, recovery studies and the analysis of real samples with low Pb content supported the satisfactory accuracy of Pb determination in automotive or aviation gasoline
CE
when the recommended analytical variants are applied. The studies in this work shed new light on spectral phenomena in air-acetylene flames. The
AC
structured background due to absorption by the OH molecules must be taken into account during Pb determination in other materials as well as in some other elemental determinations, especially at low absorbance levels. The usefulness of HR-CS FAAS for revealing and investigating a structured background was demonstrated. HR-CS FAAS does not reveal fully corrected spectral effects with a continuous character, which can be found in LS FAAS.
2
ACCEPTED MANUSCRIPT Keywords: lead determination, high-resolution continuum source atomic absorption spectrometry, structured background, LSBC, gasoline *Corresponding author. Tel:+48 604569041; fax:+48 24 3653307 E-mail address:
[email protected] (Z. Kowalewska)
RI
PT
1. Introduction
SC
Lead is a very important element in the petroleum industry due to the application of tetraalkylleads (TALs) and especially the most widely used compound TEL (tetraethyllead) as
NU
an antiknock agent [1-3]. TEL, which was commercially introduced in 1923, turned out to be an efficient and useful additive to gasoline. It was estimated that the application of this
MA
compound significantly contributed to the development of transport and the global economy [2]. However, TEL has also been called “the mistake of the 20th century” [2,3]. Its poisoning
D
effect on human health, first of all, its neurotoxicity, has been known since the beginning of
PT E
TEL production, but the facts were not properly taken into account for many years. The broad application of TALs contributed to significant environmental pollution all over the world,
CE
including in Greenland snow and high alpine sites [4,5]. After the peak years of maximum leaded gasoline usage in the 1960s-70s, the process of phasing out TALs began. At present,
AC
only a few countries in the world allow the use of leaded automotive gasoline [2,3]. The aviation industry is more conservative than the automotive industry, and various grades of leaded aviation gasoline are still produced, although usually with diminished Pb content. The most popular aviation gasoline, Avgas 100LL, can contain up to 0.56 g L-1 Pb [6]. The complete phasing out of leaded gasoline is planned [3], and unleaded aviation gasoline [7,8] technologies are being developed.
3
ACCEPTED MANUSCRIPT Automotive gasoline sold in European Union can contain up to 5 mg L-1 Pb [9]. The admissible Pb content in unleaded aviation gasoline is at a similar level, for example, 13 mg L-1 in 91 UL gasoline [7] or 5 mg L-1 in 85 UL gasoline [8]. Because the natural concentration of Pb (coming from the original crude oil) in middle distillates is usually 0.001-0.200 mg L-1 [10-22], a higher content of Pb in commercial unleaded gasoline indicates contamination.
PT
Lead is not only harmful to human health, but it also damages the catalytic convertors
RI
used in modern cars. For that reason, associations of cars producers have the aim of
SC
diminishing the threshold concentration of Pb (as other elements) to 1 mg L-1 for all grades of automotive gasoline [23].
NU
Various analytical techniques were used to develop methods of Pb determination in gasoline, including graphite furnace atomic absorption spectrometry [10-16,24], inductively
MA
coupled plasma mass spectrometry [17-20], inductively coupled plasma optical emission spectrometry [21,25], microwave plasma optical emission spectrometry [26] and various
D
variants of voltammetry [27-28]. Some methods include Pb preconcentration [20-22].
PT E
The officially approved methods for Pb determination in unleaded gasoline are the standard methods noted in gasoline specifications [7-9]. In the case of ASTM 7547
CE
specification [7], these are the ASTM 5059C standard method with the application of wavelength dispersive X-ray spectrometry for Pb determination in the range 2.6-132 mg kg-1
AC
[29] and the ASTM 3237 standard method with the application of flame atomic absorption spectrometry (FAAS) [30]. In compliance with European Union norm [9], the only officially approved method is the EN 237 standard method with the application of FAAS [31]. Two versions of this method are available, that is, the recommended one and the alternative one. The ASTM 5059C method is outside the scope of this work due to its relatively poor sensitivity and detectability. The primary features of the FAAS methods have been compared in Table 1. 4
ACCEPTED MANUSCRIPT FAAS has been used for Pb determination since the 1960s and many studies that were performed in the field (primarily in the 60s and 70s for leaded gasoline) led to the development of an effective samples and standards preparation procedure, which enabled an equalizing signal for various Pb species [32]. This procedure is still used in the FAAS standard methods [30-31] (Section 2.2).
PT
The direct causes for undertaking the topic of Pb determination in unleaded gasoline in
RI
this work were sporadic disagreements in the results of analyses that were performed
concerning the quality of the investigated fuel.
SC
according to official methods (Table 1 [30,31]), which led to contradictory conclusions
NU
The aim of this work was to find a source of error and to explain the inaccuracy of Pb determination in unleaded gasoline using FAAS. At present, apart from the classical version
MA
of FAAS with a Pb hollow cathode lamp as a line source of radiation (LS FAAS), a new version known as high resolution-continuum source flame atomic absorption spectrometry
D
(HR-CS FAAS) [33-35] can be used. It was expected that HR-CS FAAS, which makes it
PT E
possible to reveal the spectral neighbourhood of the investigated analyte line, would be helpful for performing the tasks in this study. It was also expected that the results of these
CE
investigations would make it possible to set recommendations for routine analysis using both
AC
HR-CS FAAS and LS FAAS.
2. Experimental
2.1. Instrumentation and its operation
The ContrAA 700 (Analytik Jena, Jena, Germany), a high-resolution continuum source absorption spectrometer, was equipped with a high-intensity xenon short-arc lamp as a 5
ACCEPTED MANUSCRIPT radiation source, a high-resolution double monochromator consisting of a prism and an echelle grating and a linear charge-coupled device array detector (200 pixels were available for analytical purposes). A spectral interval of approximately 0.245 nm (at 217.001 nm) or 0.314 nm (at 283.306 nm) was recorded simultaneously with a high dispersion of Δλ≈λ/175 000 per pixel. The resolution corresponded to 1.22 pm (at 217.001 nm) or 1.57 pm (at
PT
283.306 nm) per pixel. If it is not indicated otherwise, the experiments with the ContrAA 700
RI
were performed using 217.001 nm or 283.306 nm Pb lines (the abbreviations 217 nm and 283
SC
nm will be used further in this study), three pixels for Pb atomic absorption evaluation (CP±1; CP=central pixel) and iterative background correction (IBC) (more about the background
NU
correction can be found in references [33,36,37]).
The SOLAAR “M” (Thermo Electron, Waltham, USA), a line source atomic
MA
absorption spectrometer, was equipped with a hollow cathode lamp as a source of specific Pb radiation (the current of the lamp was 8 mA) and a deuterium lamp of the Quadline type was
D
used for background correction. The spectrometer was operated at the 217 nm or 283 nm Pb
PT E
line using a 0.2 nm or 0.1 nm bandpass.
In both HR-CS and LS FAAS, the sample introduction-waste removal systems were
CE
resistant to organic solvents. An air-acetylene flame was generated using burners with a 100 mm slit and an optimized burner height of 7 mm. In most of the investigated analytical
AC
variants, an oxidizing flame with a blue colour was used (a flame lean in acetylene, with the abbreviation “L”). In some experiments, a reducing flame with a yellow colour (which is rich in acetylene, with the abbreviation “R”) was generated. In LS FAAS, the maximum flow of additional air was always applied and the acetylene flow rate was 0.8 L min-1 (L flame) or 1.1 L min-1 (R flame). In HR-CS FAAS, an acetylene flow rate of 0.83 L min-1 and an additional air flow rate of 1.25 min-1 were applied to obtain an oxidizing flame while the fuel rich flame was obtained using an acetylene flow rate of 0.92 L min-1, but without additional oxidant. 6
ACCEPTED MANUSCRIPT The flow rate of the aspirated solutions was approximately 5 mL min-1. It was ensured that changes in the flow rate were lower than ± 6% when the flame character was changed as well as when solvent or various investigated solutions were aspirated. It concern not only typical calibration or sample solutions containing 90% solvent (1:10 dilution, v:v) but also certified reference material (CRM) solutions diluted in the solvent in a ratio of 1:5 or 1:2.5
PT
v:v.
RI
The list of analysis modes is presented in Table 2. In each analysis mode, the reference
SC
measurement [33,36] was the measurement of a blank solution.
NU
2.2. Standards, reagents, and samples
MA
The Pb standard, lead chloride (II) 99.999%, was from Sigma Aldrich (Saint Louis, USA). Iodine-sublimated and methyltrioctylammonium chloride (MTOACl) were obtained
D
from Merck (Darmstadt, Germany). Methyl-isobutyl ketone (MIBK) 99% (POCH, Gliwice,
PT E
Poland) was used as a solvent thoroughout the work. Toluene, isooctane and ethanol were purchased from Eurochem BGD (Tarnów, Poland). An isooctane:toluene mixture (1:1 v:v)
CE
was labelled as IT, while a 10% v:v solution of ethanol in IT was labelled as E10. The reagents used in this work were at least of analytical grade.
AC
Lead chloride was dissolved into a 10% v:v solution of MTOACl in MIBK and further stepwise diluted [30,31]. The calibration solutions used throughout the work contained 0, 0.25, 0.5, 1.0, 1.5, and 2.0 mg L-1 Pb and 10% v:v the IT mixture, which was added for matrix matching. The first of the solutions was calibration blank as well as the procedure blank. All the measured solutions (that is the calibration solutions and the samples solutions) were pre-treated with iodine, which was used in a 3% m:v solution in toluene. After reacting
7
ACCEPTED MANUSCRIPT with iodine (for a few minutes), MTOACl (as a 1% v:v solution in MIBK) was added, and the final measured solutions contained 0.006% iodine and 0.1% MTOACl [30,31]. Certified reference materials called “Lead in Gasoline Level 4”, containing 2.64 ± 0.13 mg L-1 Pb (AccuStandards, Inc., New Haven, USA) and “XRF petroleum CRM”
work, and they were labelled as CRM1 and CRM2, respectively.
PT
containing 13.21 ± 0.26 mg L-1 Pb in isooctane (VHG, Manchester, USA) were applied in this
RI
Isooctane, toluene, IT, ethanol and E10 (a 10% v:v solution of ethanol in IT) were pre-
SC
treated as samples and used as model samples. The real samples consisted of automotive gasoline EN 95 [9], aviation gasoline UL 85 [8], aviation gasoline UL 91 [7] and process
NU
solvents with the boiling range of gasoline.
MA
The investigated samples were diluted in a 1:10 v:v ratio, unless indicated otherwise.
D
3. Results and discussion
PT E
3.1. Investigation of the model solutions and CRMs using HR-CS FAAS
CE
Preliminary experiments that were performed using HR-CS FAAS showed significant differences in the spectra obtained for various samples and analysis conditions, which can
AC
indicate spectral phenomena as a source of inaccuracy in Pb determination. Therefore, systematic studies of spectral interferences were undertaken. For these studies, isooctane, toluene, IT mixture, ethanol, E10 and both CRMs were selected. The substances/materials represent groups of components/matrix of gasoline. They were prepared as samples using a 1:10 dilution factor [30,31]. In accounting for the alternative version of PN 237 [31] (Table 1), both CRMs were also diluted in a 1:5 ratio. Taking into account the low level of Pb in
8
ACCEPTED MANUSCRIPT CRM1, and to better reveal the possible matrix effects, the CRM1 was also diluted in a 1:2.5 v:v ratio. The recommendation of both standard procedures [30,31] is the application of an oxidative flame with a light blue colour. However, the colour and its intensity are not an accurate specification. There are also limitations in the flame adjustment. On one hand, the
PT
amount of available air could be too small (a hardware limitation), and the flame would be of
RI
green or even yellowish colour due to the presence of non-burned incandescent carbon. On
SC
the other hand, in the case of a very lean flame, flashback is possible, which would result not only in an extinguished flame but also in a ruptured protective membrane. The determination
NU
of the influence of the flame character on the spectra/analysis results could be important in the research undertaken in this work.
MA
It was found that when a fuel lean flame was adjusted (blue colour), there were no perceptible changes of the flame appearance during aspiration of the calibration solutions and
D
most of the sample solutions. Only for CRM solutions diluted in a 1:2.5 v:v ratio a yellowish
PT E
layer of a height of approximately 12 mm was observed over the burner slit. For a fuel rich flame, the flame was always of an intense yellow colour. The height of the yellow part was
CE
30-75 mm and depended on the aspirated solution (which will be discussed later). Some of the obtained spectra are presented in Figures S1-S4 (Appendix). The Figures
AC
refer to measurements at the 217 nm or at the 283 nm Pb lines, performed using fuel lean or fuel rich flame. The Figures refer to the analysis of isooctane, toluene and ethanol as samples or to analysis of various dilutions of CRM 1. Because a structured background was registered in some experiments, the example spectra were more precisely characterised and specified in Table S1 (Appendix) and in Figures 1. Figs. 1a-1b were obtained for MIBK while Fig. 1c for CRM2.
9
ACCEPTED MANUSCRIPT Generally, the spectra obtained for CRM1 and CRM2 were similar to one another, and the spectra obtained for ethanol (which was prepared as a sample in a 1:10 v:v dilution) were similar to the spectra obtained for MIBK (a direct solvent measurement). In considering Figures S1-S2 (Appendix), refereeing to the 217 nm Pb line, it can be stated that when a fuel lean flame is used , the baseline noise is very small. The application of
PT
a fuel rich flame causes a significant increase in the baseline noise, which is the highest for
RI
the lowest CRM1 dilution. The increase of noise can be attributed to the more intense
SC
radiation scattering due to the presence of large amount of incandescent carbon in the fuel rich flame.
NU
Additionally, a structured background can be clearly distinguished from the baseline noise in measurements in the vicinity of the 217 nm Pb line using the fuel rich flame (Figures
MA
S1d-S1f, S2d-S2f, Appendix). A similar background can be distinguished in the spectrum taken for MIBK (solvent) with the indicated characteristic lines (Fig. 1a).
D
To identify the origin of the obtained structured spectra, spectra published in the
PT E
literature were inspected. Because the pattern of the structured background depends on the molecule and does not depend on the analytical technique, both flame and graphite furnace
CE
atomization data could be taken into account. Generally, it is much more difficult to identify the molecules causing the structured spectra than the interfering atoms with single line spectra
AC
because tabulated spectroscopic data are available only for the interfering atoms [33,36,38]. For the structured spectra in the vicinity of the 217 nm line, most of the published data concern samples of complicated composition that contain huge concentrations of elements such as S, P, Si, N and some metals (e.g., Ca), which all can be blamed for the spectral interferences [33-35,39-45]. The samples investigated here do not contain these elements at a significant level, and none of the published spectra resemble the recorded spectra. An overview spectrum of the “pure” (without solution aspiration) air-acetylene flame should also 10
ACCEPTED MANUSCRIPT be considered [33,34]. In the 200-240 nm region, an absorption occurs in the flame, which can be attributed to absorption by an unknown molecule or by an NO molecule [33,34]. Careful inspection of the overview spectrum of the NO band system shows that the 217 nm Pb line is situated between two groups of rotational lines in the area of a very small NO absorption at the noise level [33]. Thus, the detected structured spectra are not caused by the
PT
NO molecule. Finally, only a suggestion offered by the ContrAA software [36] seems to shed
RI
light on the origin of the signals recorded in this work. It was found that the considered
SC
spectra are similar to the OH molecule spectrum [36]. Although the OH spectrum [36] has a less subtle structure, or is less disturbed by noise, it contains the characteristic lines, or groups
NU
of lines, described in Table S1 (Appendix) and depicted in Fig. 1a.. To the best of the authors’ knowledge, this study is the first in which the appearance of a structured spectrum in the
MA
vicinity of the 217 nm Pb line, attributed to the OH molecule, is reported to a wider audience. It is important to note that the direct overlap of the 217 nm Pb line with a less intense
D
rotational line in the structured spectrum occurs (there is a group of rotational lines in the
PT E
range of 216.978 to 217.005 nm, which correspond to the pixel number range of 99 up to 105). This result indicates the possibility of spectral interferences when using both LS FAAS
CE
and simple variants of HR-CS FAAS.
In all the measurements performed in the vicinity of the 283 nm Pb line, a structured
AC
background was registered, with primary peaks numbered in Figs. 1b - 1c and characterized in Table S1 (Appendix); considering the suggested spectra from ContrAA [36] and some spectra published in the literature [33,34,46], there is no doubt that the structured spectra registered in this work in the vicinity of the 283 nm Pb line are the OH molecule spectra. The spectra are a part of a band, which belongs to the Δ=+1 vibrational sequence in the electronic transition between the X2 and A2∑+ state [33].
11
ACCEPTED MANUSCRIPT The fine-structured band registered in the vicinity of the 217 nm line could be attributed to a vibrational sequence of a higher change in the vibrational excitation in the same electronic transition. The spectra registered in the vicinity of the 283 nm line, using the fuel lean flame (Figs. S3a-S3c and S4a-S4c, Appendix), contain peaks of greater amplitude (deep minima,
PT
high maxima) than all the spectra registered in the vicinity of the 217 nm Pb line. Among the
RI
spectra registered in the vicinity of the 283 Pb line, some spectra contain peaks of relatively low amplitude (with a background “absorbance” range from approximately -0.020 up to
SC
approximately 0.020), and their peaks can have a “positive-negative character” (Figs. S3a-S3b
NU
and S4a, Appendix). Some OH molecule spectra contain only high maxima of absorbance up to approximately 0.040; for example, the spectra for ethanol (Fig. S3c, Appendix) and for
MA
MIBK (Fig. 1b). However, the OH molecule spectra contain only deep minima in the case of lower dilutions of CRMs (with a minimum “absorbance” of approximately -0.070 up to
D
approximately -0.030, Figs. S4b and S4c, Appendix, as well as Fig. 1c). It is important to
PT E
underline that the intensity of the OH spectrum registered in the vicinity of the 283 nm Pb line using the fuel lean flame significantly depends on the sample and its concentration.
CE
The application of a fuel rich flame instead of a fuel lean flame causes a significant decrease of the structured background that is registered in the vicinity of 283 nm Pb line,
AC
which is now in the range of absorbance of approximately -0.002 up to approximately 0.010. This finding would suggest that in the fuel rich flame, the population of OH molecules is decreased. This finding seems not agree with the observations for the 217 nm line and would dispute the origin of the structured BG at 217 nm as coming from an absorption by OH molecules. However, in fact, the appearance and intensity of a structured BG due to OH molecule depends mainly on the compensation capabilities for a given analysis, i.e. on a difference between a result of a measurement for sample solution and for a blank solution, not 12
ACCEPTED MANUSCRIPT on the total amount of OH molecules. The compensation capabilities may be different for various wavelengths as in different wavelength range various additional phenomena can occur and influence BG correction efficiency. For example, when using 217 nm Pb line and fuel rich flame a very intense radiation scattering by incandescent carbon occurs as scattering effect is irreversibly proportional to the fourth power of wavelength [47]. It is sure, that
PT
processes occurring within a flame are very complex.
RI
It is important to note that the rotational lines of the OH molecule spectrum do not
SC
directly overlap with the 283 nm Pb line and are mostly situated outside the 0.1 nm wavelength range around the 283.306 nm Pb line (Figs. 1b and 1c). Therefore, their effect is
NU
not expected to be observed in HR-CS FAAS and is not expected or is at least significantly diminished in LS FAAS when a narrow bandpass of 0.1 nm or less is used.
MA
The influence of the registered molecular spectra on the results of Pb determination can be verified by an evaluation of the results of the analysis of CRMs. The results of CRM1
D
and CRM2 analyses, as presented in Figs. 2a and 2b, respectively, were obtained for various
PT E
sample dilution and analysis modes. Another important evaluation was performed on the basis of the model sample analysis, the results of which are presented in Fig. 3 (Figs. 2 and 3 also
CE
contain data for LS FAAS, which will be interpreted later). The model samples, isooctane, toluene, ethanol, and E10, as well as the solvent, MIBK, did not contain an amount of Pb
AC
detectable by FAAS, and the results of their analysis, as expected, are close to zero. The results for the CRMs are mostly inside the confidence interval (all the results obtained with the lean flame are inside the interval). The results obtained using both the Pb lines in different ranges of the electromagnetic spectrum are mutually consistent as well, which additionally confirms the lack of spectral interferences in HR-CS FAAS. Spectral interferences are not present for the 283 nm Pb line because the lines of the OH molecule spectrum are distant from the Pb line. For the 217 nm line, the structured 13
ACCEPTED MANUSCRIPT background overlaps with the Pb line, but it does not have a practical effect when a lean flame is used. The higher intensity of the structured background in the vicinity of the 217 nm Pb line when using the fuel rich flame could be responsible for some of the decrease in the Pb determination results (Fig. 2a, M7 Procedure). In HR-CS FAAS, a tool to eliminate the effects of the direct overlapping of an
PT
interferent line and an analyte line is the least squares background correction (LSBC).
RI
Generally, to obtain an analyte spectrum that is free from overlap in LSBC, an interferent
SC
spectrum is generated and mathematically subtracted from the sample spectrum. In the only publication to address the OH structured absorption at the 283 nm Pb line [33], the efficiency
NU
of LSBC in eliminating the background absorbance in aqueous solution analysis (with an absorbance from -0.002 to 0.004) was demonstrated. At present, a convenient method, known
MA
as the permanent structures (PS) option, is available [36]. This option is a special version of LSBC that uses the interferent spectrum that is stored and ready to be used in ContrAA
PT E
discussed in Section 3.3.
D
software. The application of LSBC/PS for Pb determination in unleaded gasoline will be
CE
3.2. Investigation of model solutions and CRMs using LS FAAS
AC
In LS AAS, BG correction by the application of a deuterium lamp is not feasible to correct the structured BG due to the absorption by diatomic molecules [47]. However, the effect of the OH molecule in a flame is specific because it is usually a stable (permanent) effect, which should be eliminated by the blank correction, even in LS FAAS [33,36]. The effect of OH molecule absorption is not expected in graphite furnace AAS. The absorption by an OH molecule in an air-acetylene flame was investigated for the first time by H. Massmann [47,48] (for the 306.8 nm Bi line). His important finding, which 14
ACCEPTED MANUSCRIPT corresponds to the data from this work, is the dependence of the OH molecule background compensation efficiency on the flame character [48]. Massmann also found that the OH molecule underwent Zeeman effect (a Zeeman system would not correct the BG caused by the OH molecule). Generally, literature on the OH molecule absorption in FAAS is very scarce. This is
PT
probably due to the equalizing action of the flame, which eliminates the OH molecule
RI
absorption effect in most FAAS analyses.
SC
In this study, the results of Pb determination in gasoline using HR-CS FAAS showed differences in the OH spectra, depending on the measured solution. Similar experiments, to
NU
those using HR-CS FAAS were performed using LS FAAS (Table 2, Figs. 2 and 3). To obtain additional information on the BG in LS FAAS, signals for the total absorbance (T, the
MA
measurement of the Pb lamp intensity), the background absorbance (BG, the measurement of the deuterium lamp intensity) and the atomic absorbance (AA, obtained by subtraction of BG
D
from T) were inspected for a few seconds of measurement time. The BG absorbance values
PT E
read from these plots are presented in Table 3, and some examples of the evolution of the AA, BG and T signals, as obtained for model solutions using 217 or 283 nm Pb lines, are depicted
CE
in Figs. S5 and S6, respectively (Appendix). In these figures, in some situations the wrong BG correction is indicated by negative BG and/or T and/or AA signals. A negative AA
AC
signal/result is evidence of an overcorrection, for an understated result (the average BG measured using the deuterium lamp is higher than the actual BG at the analyte line). An undercorrection presents the opposite situation. Taking into account the data from Table 3, Figs. 2 and 3 as well as Figs. S5 and S6 (Appendix), it can be stated that the BG registered in LS FAAS has a wide range (from a highly negative “absorbance” of approximately -0.22 up to a highly positive absorbance of approximately 0.5-0.6) and that an incorrect BG correction can occur even though the BG is 15
ACCEPTED MANUSCRIPT relatively small. The BG absorption is usually much smaller when using the 283 nm line than when using the 217 nm line. A very important finding is the different behaviours of the various substances tested as model samples. The most spectacular difference occurs when a lean flame, 217 nm line and 0.2 nm bandpass are applied (M3 procedure): an undercorrection occur in the isooctane
PT
analysis and an overcorrection, in the toluene analysis (Fig. 3 as well as Figs. S5a and S5b,
RI
Appendix), when performed against standards prepared using an IT mixture (1:1, v:v) for
SC
matrix matching [31]. These differences occur even though isooctane, toluene and the IT mixture are diluted in a 1:10 v:v ratio.
NU
Clearly, gasoline can have a highly variable hydrocarbon composition. For example, automotive gasoline can contain up to 35% v:v of aromatic hydrocarbons and up to 18% v:v
MA
of olefins [9], while 85 UL gasoline can contain up to 35% aromatic hydrocarbons and only 4% olefins [8]. The application of an IT 1:1 v:v mixture for matrix matching is a
D
simplification. A better solution seems to be the recommendation to apply lead-free gasoline
PT E
for matrix matching/calibration curve preparation [30], but it should be a gasoline of appropriate composition (known and similar to that of the samples).
CE
The spectral effect for a sample can significantly depend on the analysis condition. For example, for isooctane, in always using the 217 nm Pb line, an undercorrection (M3
AC
Procedure with a 0.2 nm bandpass and fuel lean flame), a right correction (M4 Procedure with 0.1 nm bandpass and fuel lean flame) or an overcorrection (M9 Procedure with 0.2 nm bandpass and fuel rich flame) can occur, which can be observed in Figs. S5a, S5d and S5g (Appendix), respectively, as well as in Fig. 3. As shown in Fig. 3, the positive and negative errors for the model samples are much lower when using the narrower bandpass of 0.1 nm, which concerns both the 217 and the 283
16
ACCEPTED MANUSCRIPT nm Pb lines. The better accuracy when using the 0.1 nm bandpass and the 283 nm Pb line can be explained by the cutting off of the OH molecule line regions. The results for E10, the 10% v:v solution of ethanol in IT, were correct (Fig. 3) in the investigated cases upon application of the lean flame (M3-M6 Procedures). This finding indicates that a 10% v:v addition of ethanol to automotive fuel [9] does not influence the
PT
accuracy of Pb determination.
RI
The investigations using LS FAAS show severe difficulties in BG correction when
SC
using a rich flame. The difficulties occur at both Pb lines, although they are much greater at the 217 nm line. The signals can be unstable when using rich flame (e.g., Fig. S5g,
NU
Appendix). However, the primary problems are from severe BG overcorrection. Among the various compounds, the highest overcorrection was observed for ethanol as a sample or for
MA
solvent (MIBK) measurement. The BG absorbance readings were -0.15 or -0.22, respectively, and an absurd reading of a Pb concentration of “-5 mg L-1” in the recalculation for the initial
D
sample can occur. The recommendation of both standard procedures [30,31] to maintain the
PT E
fuel lean flame is strongly confirmed here. The BG registered for both CRMs is similar and decreases similarly with sample
CE
dilution. The BG registered for both CRMs at the 217 nm Pb line does not depend significantly on the bandpass (0.2 nm or 0.1 nm) and on the flame character (Table 3).
AC
Erroneously high results are obtained for low dilutions of CRMs using the 217 nm Pb line (Figs. 2a), which can be correlated with high BG as well as with CRM hydrocarbon composition. CRM1 contains approximately 85% v:v isooctane, according to a chromatographic analysis in OBR JSC, while CRM2 contains approximately 100%, according to the certificate. Thus, for CRMs, a behaviour similar to the behaviour of isooctane can be expected, and the behaviour was actually observed. The effect is lower when using the narrower bandpass of 0.1 nm, similarly to the case of isooctane. The effect that occurs at the 17
ACCEPTED MANUSCRIPT 217 nm Pb line due to the BG, which is not fully compensated for, is clearly observed in the isooctane analysis as overstated results. This effect was not found in HR-CS FAAS. HR-CS FAAS has the basic option of continuous event correction, which is performed using correction pixels in WRC mode or absorption minima in IBC mode [33,36,37]. The IBC option used here probably corrected for some effects that occurred during various model
PT
samples analyses. Therefore, BG absorbance of the 0.6 level and undercorrection were not
RI
observed in HR-CS FAAS. It is important to add that analysis by HR-CS FAAS (IBC and
SC
WRC options) is simultaneous, while LS FAAS measurements of the BG and T are performed sequentially. For quickly changing BGs, a time shift can lead to errors.
NU
It is interesting to notice that differences in the flame appearance for various model samples can be distinguished (especially for the fuel rich flame, Table 3). As the flow rate of
MA
the aspiration of various solutions was always similar (due to similarity of their properties as viscosity and surface tension, among others), the relationship between the flame appearance
D
and the combustion differences of the investigated substances could be considered. For
PT E
isooctane, toluene and ethanol, the lower heating value (LHV) differs significantly as it is equal to 44.3, 40.6 and 28.9 MJ kg-1 (25°C), respectively. Thus, the higher the heating value
CE
of a sample, the more heat is released when the sample solution is combusted, the temperature of the flame is higher, and the zone of the yellow luminous flame is larger. This agrees well
AC
with the data from Table 3 in terms of both, flame appearance and BG absorbance in LS FAAS at the 217 nm line. For example, for isooctane, the LHV is higher than that for IT (blank); thus, the yellow luminous zone is larger and the BG absorbance is also higher. On the other hand, for toluene and ethanol, the LHV is lower than that of IT (blank), the yellow zone is smaller, and the BG absorbance is lower (for ethanol, it is even “negative”). The phenomena described above cannot be simply/only correlated with higher or lower scattering of radiation. Undoubtedly, the radiation scattering is higher in the fuel rich 18
ACCEPTED MANUSCRIPT flame due to the greater amount of incandescent carbon, but this cannot be simply correlated with the BG observed at the 217 nm line. For example, a similar BG was registered for CRM1, diluted 1:2.5 v:v, independently of the flame character (217 nm line, Procedures M3 and M9, Table 3). All the above data demonstrate the complexity of the processes occurring when
PT
organic solutions are aspirated into an air-acetylene flame. It is not possible to fully explain
RI
all the findings in this work. It appears that apart from radiation scattering and the structured
SC
background observed at both wavelengths and interpreted as absorption by the OH molecule (283 nm) or probable absorption by the OH molecule (217 nm), other phenomena can occur
NU
due to the variety of organic constituents present in the samples.
MA
3.3. Figures of merit, analysis of real samples and recommendations
D
The parameters that characterize the sensitivity (characteristic concentration), linearity
PT E
of calibration (correlation coefficient) and detectability (detection and determination limits) of the Pb determination were determined for HR-CS FAAS (Table 4) and LS FAAS (Table S2,
CE
Appendix). The detection and determination limits were calculated as three or seven standard deviation values for ten blank solutions prepared in parallel.
AC
In all the investigated options, the linearity of the calibration curve is good, and the sensitivity is satisfactory compared to expectations [47,50-54]. Usually, in HR-CS FAAS the best detection limit for narrow lines is achieved using 3 pixels, while for broad lines, the application of 5 pixels is more advantageous [49]. The experimental data (Table 4) show that the detectability is similar when using 3 or 5 pixels for analyte signal evaluation. This similarity occurs at both Pb lines. This finding can partially
19
ACCEPTED MANUSCRIPT explain differences in the choices undertaken in various works concerning Pb determination using HR-CS FAAS, e.g., 3 pixels [50,51] or even 7 pixels [52-54]. Another factor that can influence the detection is the applied background correction mode, i.e., the basic mode of background correction (WRC, an approximation of the baseline using correction pixels, and IBC, an approximation of the baseline using the absorption
PT
minima), as well as LSBC (the PS option) [36,37,49]. The data in Table 4 are supported by
RI
Figs. S7-S10 (Appendix), obtained for ethanol and CRM1, using the 217 nm or 283 nm Pb
SC
lines, with the application of IBC or WRC, and with or without the additional option of PS. It can be seen that the detectability depends significantly on the BG correction mode, that is, the
NU
baseline fitting mode (it is always better to use the IBC version than the WRC version) and the application of the PS option (the PS option is recommended when using the 217 nm line,
line, especially in the IBC version).
MA
especially in the WRC version, and it is strongly not recommended when using the 283 nm Pb
D
The data correspond very well to the findings concerning molecular absorption at both
PT E
wavelengths. PS eliminates the effect of the structured background directly overlapping with the 217 nm line; therefore, its application is advantageous at this line. The application of PS
CE
worsens the determination at the 283 nm wavelength because the Pb line is not overlapped at all by the OH molecule spectrum, and treatment of the high-intensity rotational lines of the
AC
OH molecule in the PS option increases the noise level and influences the baseline (Figs. S9c and S10c, Appendix).
The detectability of the optimized versions of LS FAAS and HR-CS FAAS makes it possible to detect 0.010-0.020 mg L-1 Pb in a measured solution. Taking into account the 1:10 v:v dilution ratio, the detection limits in the original gasoline sample are 0.1-0.2 mg L-1. This is satisfactory for the control of Pb level in automotive or aviation gasoline. There is also a
20
ACCEPTED MANUSCRIPT room to decrease the lower range of the standard methods of Pb determination [30-31]. In this way, the expectations of car producers would be met [23]. According to the findings from the previous sections, the results of CRMs analysis would not be enough to provide good accuracy because both investigated CRMs as well as the CRMs available on the market contain primarily isooctane, and their analysis would not
PT
reveal some difficulties associated with the variable composition of gasoline samples.
RI
Therefore, analyses of real samples were performed. Careful inspection of the area in the
SC
vicinity of the 217 and 283 nm Pb lines did not reveal any unexpected effects/lines, apart from those reported in the previous sections. Because the Pb content in all the investigated
NU
aviation and automotive gasoline samples was below the best detection limit (0.010 mg L-1), spiking tests were performed. The recovery that was achieved in the spiking tests for the
MA
gasoline samples, as well as for the model samples, was satisfactory (Table S3, Appendix). However, in the spiking test measurements, the same matrix is investigated and background
D
compensation is relatively easier.
PT E
Some real samples (process solvents) contaminated with a low level of Pb were analysed as well (Table 5). Sample 2 was analysed twice, with the second analysis occurring
CE
within a month of the first one. The sample was analysed at various dilutions, and each time, the dilutions were prepared on the day of analysis. As shown in Table 5, all the results
LS FAAS.
AC
obtained using HR-CS FAAS are mostly consistent, but there are discrepancies when using
In view of all the findings in this work, the recommendations for Pb determination using LS AAS are as follows: the 283 nm line, a 0.1 nm bandpass, a fuel lean air-acetylene flame and a 1:10 v:v sample dilution. For HR-CS FAAS, the best option is as follows: application of the 283 nm line, IBC baseline fitting without LSBC, 5 pixels for signal evaluation and a fuel lean air-acetylene flame. In HR-CS FAAS, lower sample dilutions (1:5, 21
ACCEPTED MANUSCRIPT 1:2.5 v:v) are possible, which would enable the method to reach a 0.04 mg L-1 detection limit in the original sample.
4. Conclusions
PT
The application of HR-CS FAAS for Pb measurement in solutions of gasoline in
RI
methyl-isobutyl ketone enabled us to reveal that the absorption of the investigated radiation
SC
by OH molecules (which are always present in air-acetylene flames) can significantly differ when the blank/standards and the samples are aspirated into the flame. Therefore, the
NU
absorption can remain uncompensated for, despite blank subtraction, and can be a source of error.
MA
OH molecule absorption occurs as a structured BG at the 283 nm Pb line (previous literature on the topic was very scarce). The structured background at the 217 nm Pb line,
D
which can also be attributed to OH molecule absorption, was announced for the first time in
PT E
this work. Both effects were carefully investigated here, in terms of matrix variability as well as analysis conditions, and this research led to proposed conditions in which spectral
CE
interferences due to absorption by OH molecules do not exist or are at least significantly reduced in both HR-CS FAAS and LS FAAS.
AC
In LS FAAS, apart from the OH molecule effect, additional phenomena could be distinguished by the differences in the results obtained for model samples, for example, isooctane and toluene, representing different groups of hydrocarbons. In some analyses, a higher concentration of isooctane leads to a high background (with absorbance up to 0.6) and the overstating the Pb determination results. The background, unlike OH molecule absorption as well as radiation scattering, does not significantly depend on the flame conditions. This absorbance is not observed in HR-CS FAAS. 22
ACCEPTED MANUSCRIPT The studies in this work are on a very specific topic but have more general meaning and shed new light on spectral phenomena that occur in air-acetylene flames. The structured background due to absorption by OH molecules should be taken into account during the determination of other analytes in other materials, especially at low absorbance levels. Some correlations between lower heating values of model samples, flame appearance
PT
and BG intensity or BG character have been found. The topic will be developed in future
SC
RI
works.
NU
Acknowledgements
The experimental part of this work was performed in OBR JSC (Warter Group), and
MA
the authors are grateful for the opportunity to perform the research. The authors also thank Bożena Słupska from OBR JSC for providing technical assistance.
PT E
D
This work was in part financed by the Warsaw University of Technology (statutory funds no. 504/02351/7191/40.000101), and its support is gratefully acknowledged.
AC
CE
Appendix A. Supplementary data
Supplementary data to this article can be found online at http://dx.doi.org/...
References:
23
ACCEPTED MANUSCRIPT 1. M.G.A. Korn, D.S.S.Santos, B. Welz, M.G.Vale, A.P. Teixeira, D.C.Lima, S.L.C. Ferreira, Atomic spectrometric methods for the determination of metals and metalloids in automotive fuels – A review, Talanta 73 (2007) 1-11. 2. G. Oudijk, The rise and fall of organometallic additives in automotive gasoline, Environ. Forensics 11 (2010) 17-49.
PT
3. E. O’Brien, Z. Gethin-Damon, Chronology of leaded gasoline / leaded petrol history, (A.
RI
Roberts, Ed.). http://www.lead.org.au/Chronology-Making_Leaded_Petrol_History/, 2011 (accessed 19.03.17).
SC
4. R. Łobiński, C.F. Boutron, J.P Candelone, S. Hong, Present century snow core record of
NU
organolead pollution in Greenland, Environ. Sci. Technol. 28 (1994) 1467-1476. 5. S.J. Hill, Speciation of lead, in R. Cornelis, H. Crews, J. Caruso and K.G. Heumann (Eds),
MA
Handbook of Elemental Speciation II: Species in the Environment, Food and Medicine & Occupational Health, John Wiley & Sons Ltd, Chichester, England, 2005.
D
6. ASTM D910:2016, Standard Specification for Leaded Aviation Gasolines.
PT E
7. ASTM D7547:2015, Standard Specification for Hydrocarbon Unleaded Aviation Gasoline 8. WT-10/OBR PR/PD/59 Technical Specification for Aviation Gasoline OBR 85UL, VIIIth
CE
Edition, Płock, Poland.
9. EN 228:2012, Automotive fuels - Unleaded petrol - Requirements and test methods.
AC
10. Z. Sikora, Z. Kowalewska, H. Laskowska, T. Wojnar, Lead distribution during crude oil distillation proces, Petroleum & Coal 38 (1996) 38-39. 11. Z. Kowalewska, E. Bulska, A. Hulanicki, Organic palladium and palladium-magnesium chemical modifiers in direct determination of lead in fractions from distillation of crude oil by electrothermal atomic absorption analysis, Spectrochim. Acta Part B 54 (1999) 835-843.
24
ACCEPTED MANUSCRIPT 12. M. Luz, A.N. Nascimento, P.V. Oliveira, Fast emulsion-based method for simultaneous determination of Co, Cu, Pb and Se in crude oil, gasoline and diesel by furnace atomic absorption spectrometry, Talanta 115 (2013) 409-413. 13. A. Anselmi, P. Tittarelli, D. A. Katskov, Determination of trace elements in automotive fuels by filter furnace atomic absorption spectrometry, Spectrochim. Acta Part B 57 (2002)
PT
403-411.
RI
14. R.C. Campos, H.R. Santos, P. Grindberg, Determination of cooper, iron, lead and nickel in
SC
gasoline by electrothermal atomic absorption spectrometry using three-component solutions, Spectrochim. Acta Part B 57 (2002) 15-28.
NU
15. M.N.M. Reyes, R.C. Campos, Graphite furnace atomic absorption spectrometric determination of Ni and Pb in diesel and gasoline samples stabilized as microemulsion using
MA
conventional and permanent modifiers, Spectrochim. Acta Part B 60 (2005) 615-624. 16. R.J. Cassela, D.M. Brum, C.F. Lima, T.C.O. Fonseca, Stabilization of aviation gasoline as
D
detergent emulsion for lead determination by electrothermal atomic absorption spectrometry,
PT E
Fuel Process. Technol. 92 (2011) 933-938. 17. T.D. Saint’Pierre, L.F. Dias, S.M. Maia, A.J. Curtius, Determination of Cd, Cu, Fe, Pb and
CE
Tl in gasoline as emulsion by electrothermal vaporization inductively coupled plasma mass spectrometry with analyte addition and isotope dilution calibration techniques, Spectrochim.
AC
Acta Part B 59 (2004) 551-558. 18. F. Chen, S. Jiang, Determination of Hg and Pb in fuels by inductively coupled plasma mass spectrometry using flow injection chemical vapor generation, Anal. Sci. 25 (2009) 1471-1476. 19. G. Caumette, Ch.Ph. Lienemann, I. Merdrignac, H. Paucot, B. Bouyssiere, R. Łobiński, Sensitivity improvement in ICP MS analysis of fuels and light petroleum matrices using a
25
ACCEPTED MANUSCRIPT microflow nebulizer and heated spray chamber sample introduction, Talanta 80 (2009) 10391043. 20. P.N.Nomngongo, J.C. Ngila, Determination of trace Cd, Cu, Fe, Pb and Zn in diesel and gasoline by inductively coupled plasma mass spectrometry after sample clean up with hollow fiber solid phase microextraction system, Spectrochim. Acta Part B 98 (2014) 54-59.
PT
21. P.N.Nomngongo, J.C. Ngila, S.M.Musyoka, T.A.M. Msagati, A solid phase extraction
RI
procedure based on electrospun cellulose-g-oxolane-2,5-dopne nanofibers for trace
SC
determination of Cd, Cu, Fe, Pb and Zn in gasoline samples by ICP-OES, Anal. Methods 5 (2013) 3000-3008.
NU
22. D.S.S. Santos, M.G.A. Korn, M.A.B. Guida, G.L.dos Sanots, V.A. Lemos, L.S.G. Teixeira, Determination of copper, iron, lead and zinc in gasoline by sequential multielement
MA
flame atomic absorption spectrometry after solid phase extraction, J. Braz. Chem. Soc. 22 (2011) 552-557.
D
23. The 5th World Fuel Charter, Recommended Automotive Fuel Specifications according to:
PT E
The Alliance of Automobile Manufacturers, along with the Truck and Engine Manufacturers Association, the European and Japanese Auto Manufacturers Associations.
CE
https://www.acea.be/uploads/publications/Worldwide_Fuel_Charter_5ed_2013/, 2013 (assessed 19.03.2017).
AC
24. M. Ozcan, S. Akman, Determination of Cu, Co and Pb in gasoline by electrothermal atomic absorption spectrometry using aqueous standard addition in gasoline-ethanol-water three-component systems, Spectrochim. Acta Part B 60 (2005) 399-402. 25. I.B. Brenner, A. Zander, S. Kim, Direct determination of lead in gasoline using emulsification and argon and argon-oxygen inductively coupled plasma atomic emission spectrometry, J. Anal. At. Spectrom. 11 (1996) 91-97.
26
ACCEPTED MANUSCRIPT 26. G.L.Donati, R.S. Amais, D. Schiavo, J.A. Nobrega, Determination of Cr, Ni, Pb and V in gasoline and ethanol fuel by microwave plasma optical emission spectrometry, J. Anal. At. Spectrom. 28 (2013) 755-759. 27. C.E. Cardoso, W.F. Pacheco, R. Sarubi, M.L.N. Ribeiro, P.A.M. Farias, R.Q. Aucelio, Voltammetric determination of copper and lead in gasoline using sample preparation as
PT
microemulsions, Anal. Sci. 23 (2007) 1065-1069.
RI
28. J.M. Trindade, L.C. Martiniano, V.R.A. Concalves, A.G. Souza, A.L.B. Marques, G.L. Baugis, T.C.O. Fonseca, Ch.Song, J. Zhang, E.P. Marques, Anodic stripping voltammetry
SC
coupled with design experiments for simultaneous determination of Zn2+, Cu2+, Pb2+ and Cd2+
NU
in gasoline, Fuel 91 (2012) 26-32.
29. ASTM D 5059:2014 (method C), Standard Test Methods for Lead in Gasoline by X-Ray
MA
Spectroscopy.
30. ASTM 3237:2012, Standard test method for lead determination by atomic absorption
D
spectroscopy
PT E
31. EN237:2004, Liquid petroleum products. Petrol. Determination of low lead concentrations by atomic absorption spectrometry
CE
32. S.T. Holding, J.M. Palmer, Determination of lead in gasoline by atomic absorption spectrophotometry – evaluation of standard methods, Analyst 109 (1984) 507-510.
AC
33. B. Welz, H. Becker-Ross, S. Florek, U. Heitmann, High-Resolution Continuum Source AAS, Wiley-VCH, Weinheim, Germany, 2005. 34. B. Welz, M. G. R. Vale, E.R.Pereira, I.N.B. Castilho, M.B. Dessuy, Continuum source atomic absorption spectrometry: past, present and future aspects – a critical review, J. Braz. Chem. Soc. 25 (2014) 799-821.
27
ACCEPTED MANUSCRIPT 35. B. Welz, S. Mores, E. Carasek, M.G.Vale, M. Okruss, H. Becker-Ross, High-resolution continuum source atomic and molecular spectrometry – a review, Appl. Spectrosc. Rev. 45 (2010) 327-354. 36. Aspect CS 2.1.1.0, software to ContrAA 700, Analytic Jena, Jena, Germany. 37. Z. Kowalewska, J. Pilarczyk, Ł. Gościniak, Spectral aspects of the determination of Si in
PT
organic and aqueous solutions using high-resolution continuum source or line source flame
RI
atomic absorption spectrometry, Spectrochim. Acta Part B 120 (2016) 45-56.
SC
38. PLASUS Ingenieurburo, Plasus Specline, Version 2.13, Koninsbrun, Germany, 2005, in: the Reference 44 of this work.
NU
39. D.L.G. Borges, A.F Silva, A.J. Curtius, B. Welz, U.Heitmann, Determination of lead in biological samples by high-resolution continuum source graphite furnace atomic absorption
MA
spectrometry with direct solid sampling, J. Anal. At. Spectrom. 21 (2006) 763-769. 40. D.L.G. Borges, A.F. Silva, A.J. Curtius, B. Welz, U.Heitmann, Determination of lead in
D
coal using direct solid sampling and high-resolution continuum source graphite atomic
PT E
absorption spectrometry, Microchim. Acta 154 (2006) 101-107. 41. A.R. Borges, E.M. Becker, L.L. Francois, A. Jesus, M.G.R. Vale, B. Welz, M.B. Dessuy,
CE
J.B. de Andrade, Investigation of spectral interferences in the determination of lead in fertilizers and limestone samples using high-resolution continuum source graphite furnace
AC
atomic absorption spectrometry, Spectrochim. Acta Part B 101 (2014) 213-219. 42. A.I. Barros, T.V. Silva, E.C. Ferreira, J.A. Neto, Determination of lead in eye shadow and blush by high-resolution continuum source graphite furnace atomic absorption spectrometry employing direct solid sampling, J. Braz. Chem. Soc. 26 (2015) 140-146. 43. A.T. Duarte, A.R. Borges, A.V. Zmozinski, M.B. Dessuy, B. Welz, J.B. Andrade, M.G.R. Vale, Determination of lead in biomass and products of the pyrolysis process by direct solid or liquid sample analysis using HR-CS GF AAS, Talanta 146 (2016) 166-174. 28
ACCEPTED MANUSCRIPT 44. J.L. Raposo Jr., S.R. Oliveira, J.A. Nobrega, J.A.G. Neto, Internal standardization and least-squares background correction in high-resolution continuum source flame atomic absorption spectrometry to eliminate interferences on determination of Pb in phosphoric acid, Spectrochim. Acta Part B 63 (2008) 992-995 45. J.L. Raposo Jr, S.R. de Oliveira, J.A. Noberga, J.A.G. Neto, Mathematical equation
PT
correction to spectral and transport interferences in high-resolution continuum source flame
RI
atomic absorption spectrometry: determination of lead in phosphoric acid, Ecl. Quim. Sao
SC
Paulo, 35 (2010) 19-24.
46. M.A. Bechlin, F.M. Fortunato, E.C. Ferreira, J.A.G. Neto, J.A. Nobrega, G.L. Donati,
NU
B.T. Jones, Bismuth as a general internal standard for lead in atomic absorption spectrometry, Anal. Chim. Acta 831 (2014) 24-30.
MA
47. B. Welz, M. Sperling, Atomic Absorption Spectrometry, Wiley-VCH, Weinheim, Germany, 1999.
D
48. H. Massmann, The origin of systematic errors in background measurements in Zeeman
PT E
atomic-absorption spectrometry, Talanta 29 (1982) 1051-1055. 49. U. Heitmann, B. Welz, D.L.G. Borges, F.G. Lepri, Feasibility of peak volume, side pixel
CE
and multiple peak registration in high-resolution continuum source atomic absorption spectrometry, Spectrochim. Acta Part B 62 (2007) 1222-1230.
AC
50. B.G. Nieto, M.J. Gismera, M.T. Sevilla, J.R. Procopio, Fast sequential multi-element determination of major and minor elements in environmental and drinking waters by highresolution continuum source flame atomic absorption spectrometry, Anal. Chim. Acta 854 (2015) 13-19. 51. M.B. Dessuy, R.M. Jesus, G.C. Brandao, S.L.C. Ferreira, M.G.R. Vale, B. Welz, Fast sequential of antimony and lead in pewter alloys using high-resolution continuum source flame atomic absorption spectrometry, Food Addit.Contam. A 30 (2013) 202-207. 29
ACCEPTED MANUSCRIPT 52. V.R.A. Filho, J.A.G. Neto, Different lubricating oil treatment for the determination of Cu, Cr, Fe, Ni, Sb, Pb, and Zn by HR-CS FAAS, Anal. Let. 41 (2008) 1555-1570. 53. C.C. Leite, A. Jesus, M.L. Potes, M.A. Vieira, D. Samios, Direct determination of Cd, Co, Cu, Fe, Mn, Na, Ni, Pb, and Zn in ethanol fuel by high-resolution continuum source flame atomic absorption spectrometry, Energy Fuels 29 (2015) 7358-7363.
PT
54. V.R Amorim, V.P. Franzini, J.A.G.Neto, Exploring potentialities of the HR-CS FAAS
RI
technique in the determination of Ni and Pb in mineral waters, Ecl. Quim. Sao Paulo, 33
SC
(2008) 49-56.
NU
Figure captions
1. Characterisation of the OH molecule absorption spectra: the 217 nm Pb line and fuel rich
MA
flame for the measurement of MIBK (a), the 283 nm Pb line and fuel lean flame for the measurement of MIBK (b) and the 283 Pb line and fuel lean flame for the measurement of
D
CRM2 1:5 v:v (c).
PT E
2. Results of CRM1 (a) and CRM2 (b) analysis under different conditions. For a given procedure, the average result of measurements for two sample solutions (prepared in parallel
CE
at a given dilution ratio) is shown as a bar with marked uncertainty (± standard deviation, n=2). The central horizontal lines represent the certified values. For the CRM1, the distance
AC
between the central line and the border line represents the square root of the sum of the squares of the standard uncertainties of the certified values and the analysis values. For CRM2, the border lines are distant from the central line by the reproducibility of the EN 237 method [31]. 3. Readings of Pb concentrations for the solvent (MIBK) and solutions of organic compounds, prepared as samples (dilution ratio in MIBK 1:10 v:v). No Pb addition. For a given procedure and a given model sample two results for two solutions prepared in parallel are shown as bars 30
ACCEPTED MANUSCRIPT with marked uncertainty (± standard deviation, n=2). For E10, only one solution was prepared and measured. The lines parallel to X axis are distant from the X-axis by determination limit (assumed to represent confidence interval of results for analyses of samples which do not contain Pb). In the case of the main figure it is determination limit of the M2 Procedure (the lowest between M1-M8 Procedures). In the case of the sub-figure it is determination limit of
PT
the M9 Procedure (the lowest between M9-M10 Procedures). The readings of Pb content
AC
CE
PT E
D
MA
NU
SC
RI
outside horizontal lines are assumed to suffer from systematic errors.
31
MA
NU
SC
RI
PT
ACCEPTED MANUSCRIPT
AC
CE
PT E
D
Figure 1a
32
MA
NU
SC
RI
PT
ACCEPTED MANUSCRIPT
AC
CE
PT E
D
Figure 1b
33
MA
NU
SC
RI
PT
ACCEPTED MANUSCRIPT
AC
CE
PT E
D
Figure 1c
34
D
MA
NU
SC
RI
PT
ACCEPTED MANUSCRIPT
AC
CE
PT E
Figure 2a
35
MA
NU
SC
RI
PT
ACCEPTED MANUSCRIPT
AC
CE
PT E
D
Figure 2b
36
MA
NU
SC
RI
PT
ACCEPTED MANUSCRIPT
AC
CE
PT E
D
Figure 3
37
ACCEPTED MANUSCRIPT Table 1 Primary features of standard methods of Pb determination in unleaded gasoline using FAAS. ASTM 3237 [34]
EN 237 [ 35],
EN 237 [ 35],
recommended mode
alternative mode
2.5-25
2.5-10
3.0-10
Wavelength, nm
283.306
217.000
283.306
Sample used for matrix
Lead-free gasoline
Isooctane :toluene
Isooctane :toluene
mixture, 1:1 (v:v)
mixture, 1:1 (v:v)
PT
Range of the method, mg L-1
RI
matching in calibration curve
SC
preparation 1:10
1:10
1:5
Standard
PbCl2
PbCl2
PbCl2
Repeatability, mg L-1
1.3
0.12
0.08
Reproducibility, mg L-1
2.6
0.62
0.40
AC
CE
PT E
D
MA
NU
Sample dilution ratio, v:v
38
ACCEPTED MANUSCRIPT
Table 2 List of analysis modes. Analysis Apparatus mode
Wavelength , nm
Bandpass,
Flame character L - fuel lean
nm
HR-CS
217.001
HR-CS
283.306
LS FAAS LS FAAS LS FAAS LS FAAS HR-CS
217.0 217.0 283.3 283.3 217.001
L
0.2 0.1 0.2 0.1 -
L L L L R
283.306
-
R
217.0 283.0
0.2 0.2
R R
NU
SC
FAAS M3 M4 M5 M6 M7
HR-CS FAAS LS FAAS LS FAAS
AC
CE
PT E
D
M9 M10
MA
FAAS M8
L
-
FAAS M2
-
R - fuel rich flame
RI
M1
PT
flame
39
ACCEPTED MANUSCRIPT Table 3 Readings of background absorbance in measurements using LS FAAS and some observation of flame appearance. Sample
Readings of background absorbance
solution
dilution
M3
M4
M5
M6
M9
M10
the yellow
(217
(217
(283
(283
(217
(283
layer
nm,
nm,
nm,
nm,
nm,
nm,
of rich flame,
0.2 nm,
0.1 nm,
0.2 nm,
0.1 nm,
0.2 nm,
0.2 nm,
mm*
L)
L)
L)
R)
R)
RI
SC L)
NU
Model samples
The height of
PT
Sample/
1:10
0.07
0.08
0.007
0.008
0.07
-0.020
50
Toluene
1:10
0.006
-0.024
0.008
0.002
-0.010
0.020
40
Ethanol
1:10
-0.001
-0.05
0.007
-0.002
-0.15
-0.016
30
MIBK
-
-0.002
-0.06
0.002
-0.002
-0.22
-0.013
30
E10
1:10
0.020
0.015
0.005
0.005
0.040
0.002
45
D
0.007
0.006
0.05
-0.012
45
0.2
0.2
0.015
0.019
0.3
-0.024
60
0.6
0.6
0.020
0.037
0.5
-0.034
75
1:10
0.05
0.05
0.007
0.009
0.05
-0.015
45
1:5
0.2
0.2
0.017
0.021
0.3
-0.020
60
AC
1:2.5
0.05
CRMs
0.05
1:5
CRM2
PT E
1:10
CE
CRM1
MA
Isooctane
*for the calibration solutions (IT 1:1 v:v matrix) the height of the yellow luminous layer was approximately 45 mm; all measurements were performed at the height of 7 mm over the burner slit
40
ACCEPTED MANUSCRIPT
Table 4 Sensitivity and detectability data for the determination of Pb using HR-CS FAAS. 217 nm Pb line, background correction: IBC Number of pixels
1
3a
3
5
5
off
off
On
off
on
0.245
0.0887 0.0887 0.0649 0.0650 0.0591 0.0572 0.0569c
7
9
11
off
off
evaluation Permanent structure
SC
Characteristic concentration, mg
NU
L-1 Correlation
off
RI
correction
PT
taken for signal
0.9983 0.9990 0.9990 0.9993 0.9993 0.9995 0.9996 0.9996
coefficient, R 0.018
0.016
solution, mg L-1 Determination limit
0.016
0.014
0.016
0.017
0.023
0.033
1
PT E
D
for a solution, mg L-
0.016
MA
Detection limit for a
217 nm Pb line, background correction: WRC
taken for signal evaluation
1
AC
Permanent structure correction
3
3
5
5
7
9
11
off
off
On
off
on
off
off
off
0.242
0.0876 0.0880 0.0640 0.0643 0.0580 0.0559 0.0553
CE
Number of pixels
Characteristic
concentration, mg L-1 Correlation
0.9981 0.9986 0.9987 0.9990 0.9991 0.9989 0.9985 0.9981
coefficient, R Detection limit for a
0.041
0.041
0.027
0.046
0.030
0.055
0.064
0.075
41
ACCEPTED MANUSCRIPT solution, mg L-1 Determination limit
0.062
for a solution, mg L1
283 nm Pb line, background correction: IBC Number of pixels
1
3b
3
5
5
off
off
On
off
on
0.412
0.156
0.157
0.112
0.113
7
9
11
off
off
Permanent structure
Characteristic concentration, mg
NU
L-1 Correlation
0.101
0.0993 0.0992
SC
correction
off
RI
evaluation
PT
taken for signal
0.9998 0.9997 0.9998 0.9996 0.9997 0.9996 0.9996 0.9996
coefficient, R 0.021
0.012
solution, mg L-1 Determination limit
0.062
0.010
MA
Detection limit for a
0.078
0.013
0.016
0.021
0.022
1
PT E
D
for a solution, mg L-
283 nm Pb line, background correction: WRC
taken for signal evaluation
1
AC
Permanent structure correction
3
3
5
5
7
9
11
off
off
On
off
on
off
off
off
0.409
0.1542 0.1544 0.1108 0.1109 0.0995 0.0975 0.0971
CE
Number of pixels
Characteristic
concentration, mg L-1 Correlation
0.9998 0.9997 0.9997 0.9996 0.9996 0.9995 0.9995 0.9996
coefficient, R Detection limit for a
0.028
0.018
0.019
0.019
0.021
0.027
0.036
0.046 42
ACCEPTED MANUSCRIPT solution, mg L-1 Determination limit
0.042
for a solution, mg L1
the characteristic concentrations were 0.086 and 0.13 mg L-1 and the detection limits for a solution were 0.051 and 0.025 mg L-1 for the analysis when using a fuel rich flame; cthe best parameters are marked by bold letters.
AC
CE
PT E
D
MA
NU
SC
RI
PT
a,b
43
ACCEPTED MANUSCRIPT Table 5 Results of the analysis of contaminated process solvents, mg L-1 (average results of 2-3
Analysis M1
M2
M3
M4
M5
M6
dilution,
number
(CS,
(CS,
(LS,
(LS,
(LS,
(LS,
217 nm)
283 nm)
217 nm,
217 nm,
283 nm,
283 nm,
0.2 nm)
0.2 nm)
0.1 nm)
Sample 1
2.38±0.04 2.26±0.06 2.31±0.10 2.33±0.07 1.81±0.10 -
NU
1:10 *
0.1 nm)
SC
v:v
PT
Sample
RI
parallel analyses ± standard deviations). Fuel lean flame, PS option not applied.
Sample 2 analysis
1.55±0.08 1.44±0.07 1.14±0.03 1.57±0.09 1.67±0.09 1.58±0.08
MA
1:10
1
1.58±0.20 1.52±0.02 1.73±0.02 -
PT E
2 analysis
1.45±0.05 1.50±0.08 1.26±0.10 -
1.43±0.01 -
1
-
-
-
1.19±0.08 -
-
CE
1:5
1.59±0.33 -
D
analysis
analysis
1:2.5
AC
2
analysis
1.61±0.11 1.57±0.06 1.01±0.21 -
0.98±0.02 -
1
-
-
-
<0.4
-
-
analysis 2 Sample 3
44
ACCEPTED MANUSCRIPT 1:10
0.55±0.02 0.54±0.03 0.66±0.07 0.63±0.05 1.12±0.11 0.60±0.05
*for sample 1, the fuel rich flame was also used, and the results of the Pb determination were 1.91±0.20 and 0.33±0.21 mg L-1 for the 217 and 283 nm Pb lines, respectively (0.2 nm
AC
CE
PT E
D
MA
NU
SC
RI
PT
bandpass in both cases, procedures M9 and M10, respectively).
45
ACCEPTED MANUSCRIPT
HIGHLIGHTS
Inconsistencies in FAAS determination of Pb in unleaded gasoline are explained Structured background due to absorption by OH molecules disturbs Pb determination Absorption by OH molecules significantly depends on the gasoline components
PT
Absorption by OH molecules depends on the Pb line, bandpass and flame conditions.
AC
CE
PT E
D
MA
NU
SC
RI
LSBC is not recommended for the 283 nm line and is advantageous for the 217 nm line
46