Attomolar detection of cytokines using a chemiluminescence immunoassay based on an antibody-arrayed CMOS image sensor

Attomolar detection of cytokines using a chemiluminescence immunoassay based on an antibody-arrayed CMOS image sensor

Accepted Manuscript Title: Attomolar detection of cytokines using a chemiluminescence immunoassay based on an antibody-arrayed CMOS image sensor Autho...

454KB Sizes 5 Downloads 104 Views

Accepted Manuscript Title: Attomolar detection of cytokines using a chemiluminescence immunoassay based on an antibody-arrayed CMOS image sensor Author: Donggu Hong Hyou-Arm Joung Do Young Lee Sanghyo Kim Min-Gon Kim PII: DOI: Reference:

S0925-4005(15)30098-8 http://dx.doi.org/doi:10.1016/j.snb.2015.07.042 SNB 18773

To appear in:

Sensors and Actuators B

Received date: Revised date: Accepted date:

3-4-2015 8-7-2015 12-7-2015

Please cite this article as: D. Hong, H.-A. Joung, D.Y. Lee, S. Kim, M.-G. Kim, Attomolar detection of cytokines using a chemiluminescence immunoassay based on an antibody-arrayed CMOS image sensor, Sensors and Actuators B: Chemical (2015), http://dx.doi.org/10.1016/j.snb.2015.07.042 This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Research Highlights

1

Highlights

2 3 4

 Attomolar detection for interleukins was achieved with directly performing sandwich immunoassay onto a surface of CMOS image sensor chip.  A CMOS image sensor was used as a detector for CL imaging and substrate for the sandwich immunoassay.

6

 The sensor specific data calibration method was used to obtain stable and reproducible results.

7

 This assay system was evaluated with detection of interleukin-5 with a limit of detection of 0.074 fg mL ,

cr

ed

M

an

us

 Multiple target detection was simultaneously performed.

ce pt

9

which corresponds to a 50 aM, and with dynamic range in 1 fg mL-1 to 20 ng mL-1.

-1

Ac

8

ip t

5

Page 1 of 24

*Manuscript

1

Attomolar detection of cytokines using a chemiluminescence immunoassay based

2

on an antibody-arrayed CMOS image sensor

3

ip t

Donggu Honga+, Hyou-Arm Jounga+, Do Young Leec, Sanghyo Kimd* and Min-Gon Kima,b*

4 5 6

a

7

gwagiro, Gwangju 500-712, Republic of Korea

8

b

9

Cheomdan-gwagiro, Gwangju 500-712, Republic of Korea

us

cr

School of Physics and chemistry, Gwangju Institute of Science & Technology, 261 Cheomdan-

an

Advanced Photonics Research Institute, Gwangju Institute of Science & Technology, 261

10

c

11

Republic of Korea

12

d

13

Seognam-si, Gyeonggi-do 461-701, Republic of Korea

OPTOLANE Inc., 633 Sampyeong-dong, Bundang-gu, Seongnam-si, Gyeonggi-do, Korea 463-400,

M

Department of bionanotechnology, Gachon university, San 65, Bokjeong-dong, Sujeong-gu,

14

d

15

te

16 *

Corresponding authors: [email protected], [email protected]

18

+

These authors are equally contributed to this paper

19

Ac ce p

17

20

Tel.: +82(31)750-8554

21

E-mail: [email protected]

22 23

Tel.: +82(62)715-3330

24

Fax: +82(62)715-3419

25

E-mail: [email protected]

26 27 28 29

Page 2 of 24

ABSTRACT

31

A chemiluminescence immunoassay based on a CMOS image sensor has been developed for simultaneous

32

attomolar detection of cytokines. Attomolar level detection sensitivity is achieved through direct immobilization

33

of antibodies on the surface of the CMOS image sensor and using a simple data accumulation process with noise

34

calibration after the sandwich immunoassay. Here, the proposed method was showed that the logarithmic linear

35

range for interleukin-5 (IL-5) detection is from 1 fg mL-1 to 20 ng ml-1 and the limit of detection (LOD) is 0.074

36

fg mL-1, corresponding to a concentration of 7 aM. The LOD represents ca. 1105-fold improvement compared

37

with use of conventional ELISA. Furthermore, the multiple targets including IL-2, IL-4, IL-5, and IL-6 can be

38

detected with simultaneous and low-cost manner on array type CMOS image sensor, and we hope that this

39

sensor system will be a powerful tool in biological studies.

an

us

cr

ip t

30

40

M

41

Keywords

43

CMOS image sensor, attomolar detection, cytokines, chemiluminescence, sandwich immunoassay

46 47 48 49 50 51

te

45

Ac ce p

44

d

42

52 53 54 55 56

Page 3 of 24

57

1. Introduction Antibody array technology enables simultaneous detection and analysis of protein antigens present in

59

samples. In a protein microarray format, the method can be used to track interactions and activities of proteins in

60

a large scale and high throughput manner [Templin et al., 2002]. Additionally, the technique requires only small

61

quantities of samples and, as a result, it can be employed to generate abundant data sets in a single and relatively

62

cost-effective experiment [Melton, 2004]. For this reason, antibody arrays have been applied in a wide range of

63

biomarker discovery and diagnostic studies [Devarajan et al., 2003; Srinivas et al., 2002; Fang et al., 2003].

64

Although many commercial products are available for qualitatively profiling protein expression and identifying

65

candidate biomarkers, almost all of these platforms have detection sensitivities that are less than that of the

66

enzyme-linked immunosorbent assay (ELISA) system, which has limit of detection (LOD) in the picomolar

67

region.

an

us

cr

ip t

58

New strategies are required in order to develop much more sensitive sensing systems that can be utilized for

69

simultaneous detection and analysis of unknown biomarkers. Studies aimed at this goal have led to the

70

development of sensors that have LODs at the attomolar level. For example, new techniques have been

71

developed that are based on surface plasmon resonance (SPR) or SPR imaging [Lee et al., 2006; Hu et al.,

72

2014], electrochemical immunosensors [Chikkaveeraiah et al., 2012; Tang et al., 2012], and photonic crystal

73

arrays [Deng et al., 2013]. In particular, various approaches such as carbon-nanotube-based sensors and

74

nanowire-based field effect transistor (FET) sensors with various signal amplification strategies were used for

75

highly sensitive biomarker measurements [Elnathan et al., 2012; Zheng et al., 2005; Yu et al., Byon at al.,

76

2006]. Additionally, several sensitive approaches, such as those that employ rolling circle amplification (RCA),

77

electrochemiluminescence (ECL), chemiluminescence (CL) and phtoluminescence have been devised and are

78

compatible with the antibody array format [Konry et al., 2009; Sardesai et al., 2013; Zong et al., 2013; Liu et

79

al., 2013]. Despite their strong multiplex detection ability and high sensitivity, attomolar sensitivity still remains

80

a challenge in simultaneous biomarker measurement. Accordingly, simple, low-cost methods are still required

81

for the detection of biomolecules at attomolar levels.

Ac ce p

te

d

M

68

82

Detection methods based on CL are among the most sensitive and they have wide dynamic ranges [Kricka,

83

1993]. As a result, these types of sensors have been widely used in the biological research [Brien et al., 1991;

84

Constantine et al., 1994; Nori et al., 2008]. Light emission in an enzyme-catalyzed CL system is a

85

consequence of a chemical oxidation reaction of a reagent catalyzed by peroxidase. For example, horseradish

86

peroxidase (HRP) catalyzes oxidation of luminol in the presence of hydrogen peroxide to produce CL. However,

Page 4 of 24

87

the catalytic ability of this peroxidase is sometimes insufficient for specific experimental requirements and, as a

88

result, supplements are added to enhance CL signals [Aslan and Geddes, 2009]. In addition, light emission

89

efficiencies can be maximized by carrying out the CL reaction on the surface of the detector in this system. Recent developments have been made in the design of highly sensitive light detectors. For example,

91

complementary metal oxide semiconductor (CMOS) image sensor has become good alternatives to those based

92

on charge coupled device (CCD). CMOS based sensors have many attractive features including minimal power

93

dissipation [Fossum, 1997], improved image quality [Takayanagi and Nakamura, 2013], and lower power

94

consumption [Gamal and Eltoukhy, 2005], operation voltage and cost. For this reason, several bio-sensing

95

applications of CMOS image sensor have been described. For example, biofluorescence imaging inside a brain

96

tissue phantom has been carried out using a dedicated CMOS image sensor along with CL detection of nucleic

97

acids hybridized onto a probe-immobilized CMOS photodetector array [Ng et al., 2006; Mallard et al., 2005].

98

Although utilized in in selected cases, CMOS image sensors have not been employed extensively in

99

biomolecular studies.

M

an

us

cr

ip t

90

In the investigation described below, we have successfully implemented a CL based system for the

101

simultaneous and attomolar level detection of cytokines on an antibody arrayed CMOS image sensor. The

102

CMOS image sensor in this system not only serves as a detector for capturing images but it also acts as a

103

substrate for the immunoreaction in order to reduce CL light loss entering to a CMOS image sensor. In this way,

104

CL light emitted near the surface of the sensor is directly transmitted to photodiodes which convert light into an

105

electric and then digital signal with minimal loss. A calibration method is used to analyze digital data obtained

106

from images in order to reduce variations between the default digital signals from chips and between different

107

chips, and it is also used for correcting background signals arising from electrothermal effects on the sensors.

108

Interleukin-5 (IL-5), one of the cytokines that is present at levels below several pg ml-1, is used as a model target

109

to evaluate the sensitivity of the new detector. Furthermore, simultaneous detection of several cytokines on one

110

chip is carried out using this system.

Ac ce p

te

d

100

111 112 113

2. Material and Methods

114

2.1. Reagents and apparatus

115

CMOS image sensor chips and a reader module were provided by Optolane Inc. (Korea). 4-

Page 5 of 24

Aminoantipyrine (AAP), luminol sodium salt, 4-iodophenol (4-IP), p-coumaric acid, N,N-dimethylformamide

117

(DMF), glutaraldehyde (GA) solution (Grade I, 50%), sodium bicarbonate (NaHCO3) and sodium carbonate

118

(Na2CO3), hydrogen peroxide (H2O2), Tween 20, (3-aminopropyl)trimethoxysilane (APTMS) were obtained

119

from Sigma Aldrich (Korea). Bovine serum albumin (BSA) was obtained from Fitzgerald. 1X Phosphate

120

buffered saline (PBS) (137 mM NaCl, 2.7 mM KCl, 4.3 mM Na2HPO4, 1.4 mM KH2PO4) and Tris-HCl (1.5 M,

121

pH 8.8) were obtained from BIOSESANG, Inc. The TMB substrate reagent set was obtained from BD

122

Biosciences (United State). Supersignal west femto maximum sensitivity substrate and supersignal west pico

123

chemiluminescent substrate were obtained from Thermo Acientific (United State). Cytokine ELISA kits for

124

interleukin-2, -4, -5 and -6 (IL-2, IL-4, IL-5, IL-6), which include antigen capture antibody, biotinylated

125

detection antibody and streptavidin-conjugated horseradish peroxidase (STA-HRP), were obtained from BD

126

biosciecnces (United State). CL intensities were integrated using a chemiluminescence imaging system

127

(ChemiDoc, BIO-RAD). The IR spectrums were obtained using FTIR spectrometer (FT/IT-4200, JASCO).

an

us

cr

ip t

116

129

M

128

2.2. Preparation of the capture antibody immobilized CMOS image sensor The chip was modified to introduce aldehyde groups needed for immobilization of the capture antibodies.

131

The aldehyde functionalized surface was prepared by using the following procedure. The 4 µL of 1% BSA/PBS

132

solution was applied on the chip, a cover glass was placed on the chip and the assembly was allowed to stand for

133

1 h at room temperature (RT). The chip was then washed with PBST (0.05% tween 20 in 1X PBS) and distilled

134

water (DW) twice. The aldehyde-functionalized chip surface was prepared by adding 4 µL of 2% (v/v)

135

glutaraldehyde in PBS solution onto BSA adsorbed chip surface, a cover glass was placed on the chip and the

136

assembly was allowed to stand for 4 h at RT, and the chip surface was washed with PBST and DW twice. The 4

137

µL of capture antibody (4 µg mL-1 in 0.1 M carbonate buffer (pH 9.5)) is manually applied to the surface of a

138

chip with a micro pipette, then chip is incubated at 4 °C overnight. Finally, after twice wishing, the 1% BSA in

139

PBS solution was introduced on CMOS image senor for blocking the residual GA. The volume (4 µL) was

140

enough to cover the surface of CMOS image sensor (3 mm  4 mm). The IR spectrums were obtained BSA,

141

GA immobilization steps respectably.

Ac ce p

te

d

130

142 143 144

2.3. Sandwich immunoassay for detection of interleukin-5 As depicted in Scheme 1A, the IL-5 sandwich immunoassay is performed on the capture antibodies-

Page 6 of 24

functionalized chip surface. Solutions of the IL-5 antigen and biotinylated detection antibody are prepared in 1%

146

BSA/PBS. The sandwich immunoassay based on the general ELISA procedure is then performed by adding a

147

solution (4 µL) of the IL-5 antigen, installing a cover glass for 2 h, and then adding 4 µL of 1 µg mL-1 of the

148

biotinylated detection antibody and 4 µg mL-1 of STA-HRP mixture, followed by installation of a cover glass for

149

1 h. Between each step in this procedure, the surface is washed with PBST twice. Finally, the chip is incubated

150

in a sealed petri dish in order to prevent sample evaporation. To measure IL-5 using the CMOS image sensor, an

151

optimized CL solution was used. After introducing the CL reaction solution to the CMOS image sensor at the

152

end of the immunoassay, 1-sec exposure images were obtained for 120 sec before and after the reactions. Finally,

153

the images were analyzed using the proposed calibration method. The detailed CL optimization procedures are

154

described in the following section, and the data processing and calibration methods are described in detail in

155

section 3.4.

an

us

cr

ip t

145

156

2.4. CL optimization

M

157

In order to establish stable conditions for carrying out the HRP catalyzed CL reaction in our system, an

159

optimum buffer pH and enhancer concentration must be identified and a method was needed for reducing

160

interference of incident photon rate by phenolic compounds produced in the CL reaction. For this purpose, the

161

CL characterization test was performed with fixed compositions of 2 mM luminol in 50 mM sodium hydroxide

162

(NaOH) and 2 mM H2O2. CL detection was performed after STA-HRP reacts with biotinylated antibody, and its

163

intensity is measured using the chemiluminescence imaging system. All composition of CL solution were

164

optimized with 96-wellplate in order to follow standard method and the optimized condition was applied to

165

CMOS image sensor.

167

te

Ac ce p

166

d

158

2.4.1. Optimization of CL buffer pH

168

Various pH buffer conditions were tested using a 96-well microtiter plate. STA-HRP (1 µg mL-1) was added

169

to each well followed by incubation for 30 min and washing with PBST and DW twice. A fixed composition CL

170

solution containing 0.5 mM 4-IP in DMF along with various buffers including pH 7.2 (1X PBS), 7.4 (0.1M

171

phosphate buffer), 8.3 (0.1M borate buffer) 8.8 (0.1 M Tris-HCl buffer) and 10.3 (0.1 M carbonate buffer), were

172

used. The CL intensity was measured for 7 min.

173 174

2.4.2. Optimization of 4-IP concentration

Page 7 of 24

4-IP concentration optimization was also carried out. This was performed in a manner that is similar to that

176

used for buffer pH optimization. Specifically, 10 ng mL-1 STA-HRP was added to each well of a 96-well

177

microtiter plate, followed by addition of a fixed composition of CL solution in 0.1 M Tris-HCl buffer (pH 8.8)

178

and various concentrations (1 nM–0.01 mM) of 4-IP. The CL intensity was measured for 7 min.

ip t

175

179 180

2.4.3. Characterization of AAP effect

An improvement of the CL intensity by AAP was observed in the following way. First, 100 ng mL-1 of STA-

182

HRP was adsorbed to each well of a 96-well microtiter plate, followed by addition of fixed compositions of CL

183

solution in 0.1 M Tris-HCl (pH 8.8) buffer with DMF and various concentrations (1 mM–100 nM) of AAP. The

184

CL intensity was measured for 7 min.

us

cr

181

186

an

185 2.4.4. Comparison of various CL solutions

In order to evaluate the optimized CL composition, comparison was made between the CL solution

188

containing AAP and commercial CL solution on the CMOS image sensor chip. For this comparison, CL

189

solutions, consisting of 50 mM sodium hydroxide (NaOH), 2 mM H2O2, and 0.5 mM enhancer (4-IP or p-

190

coumaric acid) with and without 100 nM AAP, were prepared. SuperSignal West Femto Maximum Sensitivity

191

Substrate and SuperSignal West Pico Chemiluminescent Substrate were prepared for comparison purposes. The

192

IL-5 immunoassay was carried out on six CMOS image sensors for each CL solution. The image sensors were

193

first modified with aldehyde functionalized surfaces. Then, IL-5 sandwich immunoassay was performed with an

194

antigen concentration of 1 ng mL-1. For comparison, six different CL solutions were applied to each chip at the

195

final step. The CL response was measured using a CMOS image sensor reader module and analyzed using the

196

noise calibration method.

d

te

Ac ce p

197

M

187

198

2.5. Selectivity test

199

The four cytokines IL-2, IL-4, IL-5, and IL-6 were used to evaluate the selectivity of the new assay system

200

(Supplementary Fig. S1). Capture antibodies were manually applied to the chip surface with a micro pipette, and

201

then blocking step was proceeded with 1% BSA solution. For this purpose, IL-2, IL-4, IL-5 and IL-6 antigens

202

serving as multiple detection targets were dissolved in 1% BSA/PBS. Solutions of the four biotinylated

203

detection antibodies and STA-HRP in 1% BSA/PBS were prepared in advance. The stepwise procedure was

204

initiated by first preparing the aldehyde functionalized surface, and then 1 µL of the four capture antibodies

Page 8 of 24

were manually spotted and immobilized on each designated region for 2 h. After the blocking step using 1%

206

BSA/PBS, 4 µ L of a mixture of the target antigens (10 pg mL-1) was applied to the surface of the chip and

207

covered with a cover glass. Then, 4 µL of the detection antibodies and STA-HRP mixture were applied with a

208

cover glass onto the chip surface for 1 h. During each of these steps, the surface was washed with PBST twice.

209

The subsequent CL reaction was measured by using the chip reader module for 2 min. The detection time was

210

enough to distinguish the signal difference among several signals (data not shown).

cr

211

ip t

205

3. Results and discussion

213

3.1. Conceptual description of immunoassay on CMOS image sensor

us

212

In this new sandwich immunoassay system, the entire process is carried out directly on the surface of a chip

215

coated by a silicon nitride layer as a top surface. Generally, the immobilization method of biomolecules on

216

surface plays a key role in various biosensors [Samanta and Sarkar, 2011]. Diverse methods for surface

217

modification are known to enhance biocompatibility [Samanta and Sarkar, 2011, Mallard et al., 2005; Wang

218

and Jin, 2004]. Currently, more efficient protein immobilization methods such as polymer coating and layer-by-

219

layer assembly are used for the improvement of stability and performance with high versatility and flexibility

220

manner [Wei et al., 2014; Ariga et al., 2014]. However, since the CMOS image sensor in the new assay system

221

is directly attached to the printed circuit board (PCB), these conventional surface modification methods, which

222

typically use strong reagents such as sulfuric acid in piranha cleaning, APTMS used in silanization, and plasma

223

modification by using strong impact gases such as O2 plasma, cannot be performed [Wang and Jin, 2004; Lan

224

et al., 2005]. Thus, a BSA protein-based aldehyde functionalization method, which causes less damage to the

225

chip surface, was used in the assay system. The target protein, IL-5, was detected using HRP labeled detection

226

antibody and a CL reagent. The CL reaction involving HRP promoted oxidation of luminol is initiated by

227

applying the optimized CL reagents. CL light generated on the chip surface then enters the detector and is

228

converted to a digital signal. The digital signal, generated before and after the CL reaction is then subjected to

229

the calibration method to produce the detection results in a highly sensitive manner.

Ac ce p

te

d

M

an

214

230 231

3.2. Optimization of capture antibody immobilization on CMOS image sensor

232

To immobilize the capture antibodies, a BSA protein-based aldehyde functionalization was conducted.

233

Initially, the BSA was adsorbed on the CMOS image sensor and GA was reacted to form the aldehyde-activated

Page 9 of 24

surface. All modification steps were analyzed by IR spectra. As shown in Fig. S2A, The FTIR reflectance peak

235

of GA on CMOS image sensor surface present at 2820 cm-1 (C-H stretch) and 1730 cm-1 (C=O stretch). By

236

confirmation of these two peaks on GA treated CMOS image sensor, it was found that the GA for immobilizing

237

capture antibody was successfully applied to the chip surface. The capture antibody was then reacted on the

238

aldehyde-functionalized surface. Finally, 1% BSA solution was used as a blocking agent instead of

239

ethanolamine or sodium borohydride in order to reduce potential damage. The effect of the aldehyde-modified

240

surface on the integrity of the immunoassay was estimated using IL-5 and the CL imaging system. Compared to

241

a normal, untreated surface, the aldehyde-functionalized surface provided higher sensitivity of the sandwich

242

immunoassay for concentrations ranging from 10 ng mL-1 to 100 pg mL-1. The sensitivity of surface

243

immobilized antibodies generally depends on the amount of immobilization and orientation [Samanta and

244

Sarkar, 2011].

an

us

cr

ip t

234

In addition, the proposed method showed similar intensities for up to 6 times surface washing steps (Fig.

246

S2B). However, the signal decreased with increasing the number of the regeneration steps, and the CMOS

247

image sensor was not functional after the fourth regeneration (Fig. S2C). Based on this result, this sensor is

248

suitable for disposable use.

251

d te

250

3.3. Optimization of chemiluminescence reaction

Ac ce p

249

M

245

252

Environmental conditions, such as pH, solution composition, and concentration, are very important factors

253

that affect the oxidizing conditions of luminol and hydrogen peroxide [Lee and Seliger, 1972; Lind et. al.,

254

1983; Faulkner et al., 1993]. In particular, the reaction pH and the selection and optimization of the enhancer

255

concentration are important factors in enzyme-catalyzed CL reactions [Iwata et al., 1995; Dotsikas et al., 2007].

256

The reaction pH, enhancer concentration, and 4-AAP concentration were optimized 96-well plates, and these

257

optimized conditions were applied to the CMOS image sensor. The 4-idophenol (4-IP), which is a sensitive CL

258

enhancer [Chen et al., 2012; Wu and Liu, 2009], was used, and the AAP concentration was optimized to

259

improve the CL reaction efficiency using the CMOS image sensor. Generally, several phenolic compounds

260

including 4-IP are converted to hydrophobic polyphenol in the presence of HRP and hydrogen peroxide as HRP

261

substrates. Conversely, soluble dimers are formed in the presence of 4-AAP. Therefore, 4-AAP is possible to

262

improve the light efficiency of the CMOS image sensor.

263

Initially, we tested the pH and 4-IP concentration. As a result, the optimum pH was obtained as 8.8 (0.1 M

Page 10 of 24

Tris-HCl buffer) (Fig. S3A) and the highest CL intensity is generated when the concentration of 4-IP was 0.5

265

mM (Fig. S3B). After that the effect of AAP on the CL reaction was evaluated on the optimized pH and IP

266

concentration. The results show that the CL intensity is highest when 100 nM of the AAP concentration was

267

employed (Fig. S3C). The observation that the decrease of CL intensity at high concentrations of AAP is likely

268

the reaction of this substance with hydrogen peroxide, which lowers the effective concentration of hydrogen

269

peroxide that is needed for the CL reaction. In order to confirm the optimized condition on CMOS image sensor,

270

we designed six different optimized conditions (see Fig. S4) to confirm the operation of optimized condition in

271

CMOS image sensor simply. A comparison of the results arising from experiments in which six different

272

chemiluminescence solutions are used shows that the highest CL intensity arises when 4-IP and AAP are

273

employed in combination (Fig. S4). According to the 4-AAP characterization test, the optimized conditions in

274

96-well plate were applied to the CMOS image sensor. As a result, we used these optimized condition in

275

following experiments.

an

us

cr

ip t

264

277

M

276

3.4. Characterization of the default digital value

In order for the assay system to have a high sensitivity, variations of the digital values arising from

279

measurements utilizing a single chip or different chips need to be taken into account. A calibration method was

280

developed for reducing the variations. Accordingly, digital intensity data collected before and after the CL

281

reaction for 120 s were analyzed using the calibration method as shown in Scheme 1B. The subtractive method

282

is embodies in the equation:

Ac ce p

te

d

278









    _  _   _ _ 

283

where ACL and CCL are respective average data of integrated digital intensity in reacted pixels and background in

284

non-reacted pixels after CL reaction during i sec in presence of chemiluminescent reagents. ABuffer and CBuffer are

285

also respective average data of integrated digital intensity in reacted pixels and background in non-reacted pixels

286

before CL reaction during i sec in presence of buffer solution (0.1 M Tris-HCl, pH 8.8) only.

287

In order to obtain sensitive result, the slight difference of default digital value in different regions even on a

288

single chip was considered for multiple detections of cytokines. The correction process, in which the average

289

pixel data obtained before reaction (ABR) is subtracted from the average pixel data obtained after CL reaction

290

(AAR), gave a highly invariant default digital value. As shown in Fig. 1A, the standard deviation (SD) of AAR

291

(black bar) is approximately 15-fold larger than that of AAR-ABR (grey bar), which shows that the deviation in

Page 11 of 24

a single chip is nearly offset by utilizing the AAR-ABR correction. Additionally, another factor that can lead to a

293

non-specific signal in absence of the CL reaction is associated with the electrothermal effect caused by operating

294

chip reader module during measurement. Inspection of the plot of AAR data shows that the intensity of the

295

digital signal increases substantially upon increasing the measurement time (Fig. 1B). However, the corrected

296

(AAR-ABR) data is close to zero, which means that the electrothermal effect of a chip is easily accounted for

297

using this correction process. Even though they are made by using the same mechanical process, chips could

298

have variable electrical properties that cause variation in their digital output intensities. This phenomenon is

299

exemplified by the results of experiments (Fig. 2A) which show that differences exist in the digital intensities

300

between IL-5 immunoassays performed on three different chips with the same IL-5 antigen concentration of 1

301

ng mL-1. Importantly, the data in Fig. 2B show that these differences nearly disappear when the background

302

correction method involving subtraction of background noise is applied.

an

us

cr

ip t

292

303

3.5. Attomolar detection of interleukin-4 and interleukin-5

M

304

In order to evaluate the new performance of this proposed system, comparison tests were made using a 96-

306

well plate, slide glass and the CMOS image sensor. The sandwich immunoassay was performed using the

307

standard ELISA procedure with CL, and the glass slide and CMOS image sensor were prepared using same

308

surface modification method. As the data in Fig. 3A show, IL-5 can be detected by employing this technique

309

with 96-well plate at a concentration as low as 1 pg mL-1 and the calculated LOD of 0.91 pg mL-1 with

310

correlation coefficient (r) of 0.9856 (n = 3). Additionally, the assay carried out utilizing a glass slide has a lower

311

10 pg mL-1 sensitivity for detection of IL-5 and the LOD of 41.7 pg mL-1 (r = 0.8702, n = 2). In order to further

312

explore the sensitivity of the new system, digital intensities (I) arising from analysis of various IL-5

313

concentrations were recorded as a function of time (s). The results (Fig. 3A) show that the digital intensity

314

increases with increasing IL-5 concentration from 0.001 to 20000 pg mL-1. IL-5 can be detected at a

315

concentration as low as 1 fg mL-1 with a calculated LOD of 0.074 fg mL-1 (r = 0.9891, n = 3), which

316

corresponds to 50 aM. This higher sensitivity of CMOS image sensor than which of glass slide and 96-well plate

317

is attributed to distance between detector and sample as well as the CMOS image sensor optimized calibration

318

method. This LOD of immunoassay on the CMOS image sensor is the most sensitive simultaneous

319

immunoassay result among the biosensors described in Table 1.

Ac ce p

te

d

305

320

Another target, IL-4, was assayed using the new system. As shown in Fig. S5, inspection of the data shows

321

that the LOD for detection of IL-4 is 1.74 fg mL-1, corresponding to 980 aM (r = 0.9051, n = 3). Although it

Page 12 of 24

appears that IL-4 detection using the CMOS image sensor indicated is less sensitive and reproducible than that

323

for IL-5 detection, the results show that the possibility exists for using this system to detect other targets with

324

attomolar level sensitivity. Noteworthy points are that this system requires only a small 4 µL volume of a

325

solution containing IL-5 antigen to cover the chip surface and the concentration data collected at early time

326

points (1 min) are highly linear. Most importantly, the new the CMOS image sensor-based system is

327

approximately 1105-fold more sensitive than the conventional ELISA method. Furthermore, we measured the

328

1 pg mL-1 IL-5 spiked human serum samples, and compared with above assay result of IL-5 antigen diluted in 1%

329

BSA/PBS. From this experiment, we obtained similar intensity results of serum sample with digital intensity

330

value compared with IL-5 immunoassay result (Fig. S7). Although, this result was not enough to evaluate the

331

real sample application, we thought that this result showed the possibility of real sample application.

us

cr

ip t

322

333

an

332

3.6. Selectivity and simultaneous detection of multiple cytokines

The selectivity and multiple detection ability of the new assay were evaluated with a mixture of 10 pg mL-1

335

of each of four different cytokines, IL-2, -4, -5 and -6, using a 9 well-arrayed CMOS image sensor in which the

336

surfaces of a chip contained one of the four different immobilized capture antibodies. As the data in in Fig. 3B

337

demonstrate, the digital intensities selectively increased in response to IL-2, -4, -5 and -6 in each detection

338

region. Thus, only the complimentary antigen in each case reacts with its specific capture antibody to generate a

339

detectable CL signal.

Ac ce p

te

d

M

334

340

Simultaneous multiple cytokines detection was also performed on a 9 well-arrayed chip surface. When no

341

antigen sample is introduced, the digital intensity is close to zero in all detection areas. However, a highly

342

intense signal arises when the sample containing 10 pg mL-1 of all antigens is tested. Because the four cytokine

343

samples were selected from eight types by using a cross-activity test, the ability to perform multiple detection

344

demonstrates that little non-specific reaction takes place during the sandwich immunoassay. The fact that this

345

test was also successfully conducted using the well-type CMOS image sensor suggests that multiple detection

346

for various proteins is possible by applying micro-dispensing technique on the CMOS image sensor surface.

347

Finally, the multiple cytokines spiked samples depend on various concentrations was measured, the LODs

348

were obtained as 0.72 fg mL-1, 0.29 fg mL-1, 0.17 fg mL-1, 0.55 fg mL-1 for IL-4, IL-2, IL-5 and IL-6 respectably,

349

and all antigens successfully measured from the 1000 pg mL-1 to 1 fg mL-1 (Fig. S6). Although, some results

350

showed the relative higher standard deviations, it is showed the possibility to use as an alternative ELISA

351

method with ultra-sensitive manner.

Page 13 of 24

352

4. Conclusion In the investigation described above, we have implemented a chemiluminescence-based simultaneous

354

attomolar level detection for cytokines that utilizes a CMOS image sensor. Critical factors associated with the

355

chemiluminescence process that generates the signal in this assay, such as buffer pH and enhancer concentration,

356

were optimized, and the ability of AAP to reduce surface interference arising from byproducts was confirmed.

357

Optimized amounts of the components for the CL process were applied to the surface on which the sandwich

358

immunoassay was performed. Most importantly, a background signal calibration method was devised to correct

359

for variations in data caused by electrothermal effect and mechanical production in a single chip and between

360

chips. Use of the data calibration method enables the cytokine to reach an attomolar level sensitivity and to have

361

a high specificity for protein detection. Evaluation of the assay system showed that it can be employed to detect

362

IL-5 with a limit of detection of 0.074 fg mL-1, which corresponds to a 50 aM, and with dynamic range in 1 fg

363

mL-1 to 20 ng mL-1. In addition, the system successfully used to detect four different cytokines in a simultaneous

364

manner, and simply confirmed the similar CL intensities between buffer and human serum spiked IL-5 (Figure

365

S7). Although, this result was not enough to evaluate the real sample measurement, this result showed the

366

possibility of real sample application. The attractive features of the strategy used to devise the new

367

immunoassay suggest that it can be utilized to design portable, cost-effective and simple point-of-care detection

368

devices and potentially be applied to creating more efficient multiple detection methods in microarray based

369

assays for unknown and important biomarkers.

371 372 373

cr

us

an

M

d

te

Ac ce p

370

ip t

353

374 375 376 377 378

Page 14 of 24

Acknowledgements

380

This work was financially supported by grants from the GRL Program (NRF-2013K1A1A2A02050616), the

381

NLRL Program (NRF-2011-0028915), and the R&D Joint Venture Program funded by the Ministry of Science,

382

ICT and Future Planning.

ip t

379

383

cr

384 385

us

386

an

387 388

M

389 390

394 395 396 397 398 399

te

393

Ac ce p

392

d

391

400 401 402 403

Page 15 of 24

References

405

Ariga K., Yamauchi Y., Rydzer G., Ji Q., Yonamine Y., Wu K., Hill J.P., 2014, Chem. Lett. 43, 36-68.

406

Aslan, K., Geddes, C.D., 2009, Chem. Soc. Rev. 38, 2556–2564.

407

Brien, J.F., McLaughlin, B.E., Nakatsu, K., Marks, G.S., 1991, J. Pharmacol. Methods 25, 19-27.

408

Byon H.R. and Choi H.C., 2006 J. Am. Chem. Soc. 128, 2188-2189.

409

Chen, F., Lin, Z., Zheng, Y., Zeng, H., Nakajima, H., Uchiyama, K., Lin, J.M., 2012, Anal. Chim. Acta. 739,

410

77-82.

411

Chikkaveeraiah B.V., Mani V., Patel V., Gutkind S., Fusling J.F., 2011 Biosens. Bioelectron. 25, 4477-4483.

412

Constantine, N.T., Bansal, J., Zhang, X., Hyams, K.C., Hayes, C., 1994, J. Virol. Methods 47, 153-164.

413

Deng G., Zu K., Sun Y., Chen Y., Zheng T., Li J., 2013 Anal. Chem. 85, 2833-2840.

414

Devarajan P, Mishra J, Supavekin S, Patterson L.T., Steven Potter S., 2003, Mol. Genet. Metab. 80, 365-376.

415

Dotsikas Y., Louks Y.L., 2007 Talanta 71, 906-910.

416

Elnathan R., Kwiat M., Pevzner A., Engel Y., Burstein L., Khatchtourints A., Lichtenstein A., Kantaev R.,

417

Patolsky F., 2012 Nano. Lett. 12, 5245-5254.

418

Fang, Y., Lahiri, J., Picard, L., 2003, Drug Discov. Today 8, 755-761.

419

Faulkner K. and Fridovich I., 1993 Free Radic. Biol. Med. 15, 447-451.

420

Fossum, E.R., 1997, IEEE Trans. Electron. Device 44, 1689-1698.

421

Gamal, A.E., Eltoukhy, H., 2005, IEEE Circuits and Devices Magazine 21, 6-20.

422

Hu W., He G., Zhang H., Wu X., Li J., Zhao Z., Qiao Y., Lu Z., Liu Y., Li C.M., 2014 Anal. Chem. 86, 4488-

423

4493.

424

Iwata R., Ito H., Hayashi T., Sekine Y., Koyama N., Yamaki M., 1995 Anal. Biochem. 231, 170-174.

425

Jang, H.R., Wark, A.W., Baek, S.H., Chung, B.H., Lee, H.J., 2014, Anal. Chem. 86, 814–819.

426

Konry, T., Hayman, R.B., Walt, D.R., 2009, Anal. Chem. 81, 5777-5782.

427

Kricka, L.J., 1993, Clin. Biochem. 26, 325-331.

428

Lan, S., Veiseh, M., Zhang, M., 2005, Biosens. Bioelectron. 20, 1697-1708.

429

Lee H.J., Nedelkov D., Corn R.M., 2006 Anal. Chem. 78, 6504-6510.

430

Lee, J., Seliger, H., H., 1972, Photochemistry and Photobiology 115, 227–237.

431

Lind J., Merenyi G., Eriksen T.E., 1983 J. Am. Chem. Soc. 105, 7655-7661.

432

Liu D., Wu F., Zhou C., Shen H., Yuan H., Du Z., Lan M., Li L.S., 2013 Sens. Actuators, B 186, 235-243.

433

Mallard, F., Marchand, G., Ginot, F., Campagnolo, R., 2005, Biosens. Bioelectron. 20, 1813-1820.

Ac ce p

te

d

M

an

us

cr

ip t

404

Page 16 of 24

Melton, L., 2004, Nature 429, 101-107.

435

Ng, D.C., Tokuda, T., Yamamoto, A., Matsuo, M., Nunoshita, M., Tamura, H., Ishikawa, Y., Shiosaka, S., Ohta,

436

J., 2006, Sens. Actuators B 119, 262-274.

437

Nori, V., Seitzman, J., 2008, 46th AIAA Aerospace Science Meeting and Exhibit, 1-14.

438

Samanta D. and Sarkar A., 2011 Chem. Soc. Rev., 40, 2567-2592.

439

Sardesai, N.P., Kadimisetty, K., Faria, R., Rusling, J.F., 2013, Anal. Bioanal. Chem. 405, 3831-3838.

440

Srinivas P.R., Verma M, Zhao Y, Srivastava S., 2002, Clin. Chem. 48, 1160-1169.

441

Takayanagi, I., Nakamura, J., 2013, Proceedings of the IEEE 101, 61-73.

442

Tang C.K., Vaze A., Rusling J.F., 2011 Lab. Chip 12, 281-286.

443

Templin, M.F., Stoll, D., Schrenk, M., Traub, P.C., Vöhringer, C.F., Joos, T.O., 2002, Drug Discov. Today 7,

444

815-822.

445

Wang, Z.H., Jin, G., 2004, Colloids Surf. B: Biointerfaces 34, 173-177.

446

Wei Q., Becherer T., Angioletti-Uberti S., Dzubiella J., Wischke C., Neffe A.T., Lendlein A., Ballauff M., Haag

447

R., 2014, Angew Chem Int Ed Engl. 53, 8004-8031.

448

Wu, Y., Liu, S., 2009, Analyst 134, 230-235.

449

Yu X., Munge B., Patel V., Jensen G., Bhirde A., Gong J.D., Kim S.N., Gillespie J., Gutkind J.S.,

450

Papadimitrakopoulos F., Rusling J.F., 2006 J. Am. Chem. Soc. 128, 11199-11205.

451

Zheng G., Patolsky F., Cui Y., Wang W.U., Lieber C.M., 2005 Nat. Biotechnol. 23, 1294-1301.

452

Zong, C., Wu, J., Xu, J., Ju, H., Yang, F., 2013, Biosens. Bioelectron. 43, 372-378.

454 455 456 457

cr

us

an

M

d

te

Ac ce p

453

ip t

434

458 459 460 461 462 463

Page 17 of 24

464

Figure captions

465 Scheme 1. Overall schematic representation of the interleukin-5 detection on a CMOS image sensor (not to scale).

467

Chemiluminescence emission light passes through the pixel after specific sandwich immunoassay is performed and is

468

converted to a digital value (from 0 to 255) depending on the intensity of the emission (A). Detail information for the data

469

acquisition and calibration steps. The correction process for stabilizing the default digital intensity variation, which subtracts

470

the average pixel data obtained before reaction (ABR) from average pixel data obtained after CL reaction (AAR). ABR data

471

are obtained using 0.1 M Tri-HCl buffer (a, step 1). AAR data are obtained using optimized chemiluminescence reagents.

472

120 images per each reaction (step 1 and 2) are acquired by the CMOS image sensor and processed utilizing this calibration

473

method (B).

us

cr

ip t

466

474

Fig. 1. (A) AAR-ABR data calibration method reduces the variation of the default digital values among pixels in a single

476

chip. The standard deviation (SD) was investigated in randomly selected five regions on a single chip. The correction

477

process, in which the average pixel data obtained before reaction (ABR) is subtracted from the average pixel data obtained

478

after CL reaction (AAR), gave a highly invariant default digital value. The big difference among several regions appears

479

when only AAR was calculated (black bar), and it nearly disappears when AAR-ABR correction was applied (grey bar). The

480

SD in AAR-ABR correction is approximately 15 times lower than AAR data. (B) Electrothermal effect during chip operation

481

causes a distortion of data analysis. Increase of default digital intensity in absence of CL reaction can interfere in the data

482

analysis after CL reaction, and the increase phenomenon can be removed through the correction using the AAR-ABR

483

method.

M

d

te

Ac ce p

484

an

475

485

Fig. 2. The data calibration method is necessary for reducing the variations among different chips. (A) The different intensity

486

result among three chips that exhibited the same concentration of IL-5 antigen was present when the digital data were

487

calculated with data after CL reaction only. (B) The differences nearly disappear by applying a data calibration method.

488 489 490

Fig. 3. Immunoassay on CMOS image sensor provides a good sensitivity and selectivity result. (A) Calibration curve for

491

quantitative detection of IL-5 on a CMOS image sensor (△ △, left ordinate), slide glass (○, left ordinate) and 96-well plate (□,

492

right ordinate). Data points and each error bar represent averages and standard deviations of the two (slide glass) and three

493

(CMOS image sensor and 96-well plate) independent measurements. (B) Selectivity and simultaneous multiple target

494

detection ability were evaluated with four different cytokines, IL-2, 4, 5 and 6. All data bars and each error bar in the

495

histogram represent the average and the standard deviations of the two independent measurements.

496

Page 18 of 24

497

Figures

498 499

Scheme 1.

ip t

500 501

cr

502

us

503 504

505 506

Ac ce p

B

te

d

M

an

A

507 508 509 510 511 512

Page 19 of 24

513

Fig. 1.

514 515

an

us

cr

ip t

A

M

516 517 518

520 B

Ac ce p

521

te

d

519

522 523 524

Page 20 of 24

525

Fig. 2.

526 527 528

ip t

529

us

cr

A

532 533 534 535 536 537 538 539 540

te

531

Ac ce p

530

d

M

an

B

541 542 543 544 545 546

Page 21 of 24

547

Fig. 3.

548 549

A

M

an

us

cr

551

ip t

550

B

Ac ce p

te

553

d

552

554 555 556 557 558 559

Page 22 of 24

Table 1. Biomarker detection limits using various methods for simultaneous detection described in this article

562

Ac ce p

561

te

d

M

an

us

cr

ip t

560

Page 23 of 24

*Author Biographies

Biographies Donggu Hong is a Ph.D. candidate at the School of Physics and chemistry, Gwangju Institute of Science & Technology Hyou-Arm Joung is a Ph.D. candidate at the School of Physics and chemistry, Gwangju Institute of Science & Technology