Capnography during cardiopulmonary resuscitation

Capnography during cardiopulmonary resuscitation

Resuscitation 83 (2012) 789–790 Contents lists available at SciVerse ScienceDirect Resuscitation journal homepage: www.elsevier.com/locate/resuscita...

113KB Sizes 12 Downloads 185 Views

Resuscitation 83 (2012) 789–790

Contents lists available at SciVerse ScienceDirect

Resuscitation journal homepage: www.elsevier.com/locate/resuscitation

Editorial

Capnography during cardiopulmonary resuscitation

Use of waveform capnography during cardiac arrest was recommended in the 2010 International Consensus on Cardiopulmonary Resuscitation (CPR) Science with Treatment Recommendations.1 A primary benefit includes confirmation that a tracheal tube (or other airway device) has been placed correctly and is providing ventilation of the lungs.2 Capnography is considerably more reliable than either clinical assessment by auscultation or observation of chest wall movement. The recent 4th National Audit Project of The Royal College of Anaesthetists and Difficult Airway Society examined Major Complications of Airway Management in the UK. It did not focus on airway management during cardiac arrest, but it included 11 instances where failure to use or correctly interpret capnography led to unrecognised oesophageal intubations during cardiac arrest, most of which led to avoidable death or brain injury.3,4 We can assume that the incidence of unrecognised oesophageal intubation is higher when waveform capnography is not used during cardiac arrest. There is strong evidence to support the use of waveform capnography in this situation (CPR will generate an attenuated, but not absent, end-tidal CO2 trace),5 with data demonstrating 100% sensitivity and 100% specificity in identifying correct tracheal tube placement.6 In contrast to waveform capnography, studies of alternative devices to determine correct tube placement (such as colorimetric end-tidal CO2 detectors, syringe aspiration oesophageal detector, self-inflating bulb oesophageal detector and non-waveform end-tidal capnometers) have been shown to have accuracy that is not substantially better than clinical assessment.7–15 During cardiac arrest, waveform capnography may also be used to guide the effectiveness of chest compressions and to provide an early indication of return of spontaneous circulation (ROSC). However, there is currently insufficient evidence to recommend the use of end-tidal CO2 monitoring as a method of prognostication during cardiac arrest.1 In this issue of the journal, Heradstveit and colleagues from the Emergency Medical Service of Haukeland University Hospital in Bergen, Norway, present data that suggest capnography may become a more useful tool for optimising and individualising ALS in patients experiencing of out-of-hospital cardiac arrest (OHCA).16 A retrospective observational study was performed using data collected routinely from 918 OHCA patients. Of these, 194 were excluded and 149 did not have capnography data recorded. Data were analysed from the remaining 575 patients. Based on all available clinical evidence, the cause of cardiac arrest was considered to be cardiac in 58%, respiratory in 20%, pulmonary embolism (PE) in 2% and unknown in 19%. Patients who developed cardiac arrest from a respiratory (excluding PE) cause were observed to have

0300-9572/$ – see front matter © 2012 Elsevier Ireland Ltd. All rights reserved. http://dx.doi.org/10.1016/j.resuscitation.2012.04.002

significantly higher end-tidal CO2 values compared with those of other causes. The authors suggest that this is because these patients have a period of rising CO2 from reduced or absent ventilation prior to cardiac arrest, rather than having a better cardiac output during CPR. In contrast, the few patients developing cardiac arrest from pulmonary embolus had significantly lower end-tidal CO2 values, presumably because of reduced blood flow through the pulmonary circulation, despite effective CPR. End-tidal CO2 values also varied with the person performing chest compressions. The authors propose using end-tidal CO2 values to optimise the quality of CPR and indicate when poor technique or rescuer fatigue may be compromising effective compressions; they suggest that a fall in end-tidal CO2 may be an indication for a different person to take over. Interestingly, patients who gained ROSC had significantly higher end-tidal CO2 values compared to those who did not have ROSC, regardless of the initial cardiac rhythm or cause of the arrest. Importantly, the authors caution the use of specific cut-off values of end-tidal CO2 to prognosticate during cardiac arrest. They correctly state that there are several confounding factors that affect the end-tidal CO2 value. These are demonstrated in the study findings and include the cause of cardiac arrest, the initial rhythm, effectiveness of bystander and first responder CPR, and time from cardiac arrest to initiation of end-tidal CO2 monitoring. Heradstveit and colleagues also discuss the important limitations of their study. The average, highest and lowest end-tidal CO2 recordings were estimated by the attendant anaesthetist. This may have lead to recording errors and also introduces the possibility of (unintended) bias. During cardiac arrest, patients’ lungs were manually ventilated by the anaesthetist and while this may have led to over or under-ventilation the impact of this on end-tidal CO2 in a ‘low-flow’ state is surprisingly small. However, hyperventilation during cardiac arrest decreases coronary perfusion pressure and, based on animal data, decreases survival.17 Determination of the cause of cardiac arrest is difficult and incorrect categorisation of the type of cardiac arrest may influence the data. The authors acknowledge that the number of cardiac arrests of unknown cause is high and had no post-mortem results to confirm or refute any presumed pre-hospital diagnosis. In particular, only 12 patients had a presumed diagnosis of PE (based on clinical history and an initial rhythm of pulseless electrical activity or asystole), which was much lower than previously reported. This confirms that diagnosis of PE is difficult during cardiac arrest. The fact that significantly lower end-tidal CO2 values were reported in patients with PE compared with patients with other causes of cardiac arrest is an interesting finding but should be interpreted with caution.

790

Editorial / Resuscitation 83 (2012) 789–790

Waveform capnography during cardiac arrest is now well established in clinical practice. It has a recognized role in determining correct airway placement and pulmonary ventilation. Defibrillators are appearing that incorporate continuous capnography monitoring. It is useful for assessing effectiveness of chest compressions, and for early detection of ROSC. However, it must be emphasized that end-tidal CO2 is currently much more of use in determining confirmation of tracheal tube placement and ventilation, rather than the additional uses described in this paper. The study by Heradstveit and colleagues asks more questions than it answers. The next step is to undertake high quality prospective studies to determine whether robustly recorded end-tidal CO2 monitoring during cardiac arrest can provide the clinician with robust information about diagnosis of cause, appropriate interventions and prognosis. Conflict of interest statement The authors have no conflict of interests to declare. References 1. Deakin CD, Morrison LJ, Morley PT. 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations Part 8: Advanced life support. Resuscitation 2010;81S:e93–174. 2. Nolan JP, Hazinski MF, Billi JE, Boettiger BW, Bossaert L, de Caen AR, et al. Resuscitation 2010;81S:e1–25. 3. Major complications of airway management in the UK: results of the 4th National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society. Part 1 Anaesthesia. Cook TM, Woodall N, Frerk C. British Journal of Anaesthesia 2011; 106: 617-31. 4. Major complications of airway management in the UK: results of the 4th National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society. Part 2 Intensive Care and Emergency Department. Cook TM, Woodall N, Harper J, Benger J. British Journal of Anaesthesia 2011; 106: 632-42. 5. Falk JA, Rackow EC, Weil MH. End-Tidal Carbon Dioxide Concentration during Cardiopulmonary Resuscitation. N Engl J Med 1988;318:607–11. 6. Silvestri S, Ralls GA, Krauss B, et al. The effectiveness of out-of-hospital use of continuous end-tidal carbon dioxide monitoring on the rate of unrecognized misplaced intubation within a regional emergency medical services system. Ann Emerg Med 2005;45:497–503.

7. Ornato JP, Shipley JB, Racht EM, et al. Multicenter study of a portable, handsize, colorimetric end-tidal carbon dioxide detection device. Ann Emerg Med 1992;21:518–23. 8. Varon AJ, Morrina J, Civetta JM. Clinical utility of a colorimetric end-tidal CO2 detector in cardiopulmonary resuscitation and emergency intubation. J Clin Monit 1991;7:289–93. 9. Anton WR, Gordon RW, Jordan TM, Posner KL, Cheney FW. A disposable endtidal CO2 detector to verify endotracheal intubation. Ann Emerg Med 1991;20: 271–5. 10. MacLeod BA, Heller MB, Gerard J, Yealy DM, Menegazzi JJ. Verification of endotracheal tube placement with colorimetric end-tidal CO2 detection. Ann Emerg Med 1991;20:267–70. 11. Schaller RJ, Huff JS, Zahn A. Comparison of a colorimetric end-tidal CO2 detector and an esophageal aspiration device for verifying endotracheal tube placement in the prehospital setting: a six-month experience. Prehosp Disaster Med 1997;12:57–63. 12. Bhende MS, Thompson AE. Evaluation of an end-tidal CO2 detector during pediatric cardiopulmonary resuscitation. Pediatrics 1995;95:395– 9. 13. Li J. Capnography alone is imperfect for endotracheal tube placement confirmation during emergency intubation. J Emerg Med 2001;20:223– 9. 14. Pelucio M, Halligan L, Dhindsa H. Out-of-hospital experience with the syringe esophageal detector device. Acad Emerg Med 1997;4:563–8. 15. Bozeman WP, Hexter D, Liang HK, Kelen GD. Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation. Ann Emerg Med 1996;27:595–9. 16. Heradstveit B, Sunde K, Sunde GA, Wentzel-Larsen, Heltne JK. Factors complicating interpretation of capnography during advanced life support in cardiac arrest – a clinical retrospectove study in 575 patients. Resuscitation 2012;83:813–8.

Edward Scarth ∗ Tim Cook Royal United Hospital, Combe Park, Bath BA1 3NG, United Kingdom ∗ Corresponding

author. Tel.: +44 0 7930393497. E-mail address: [email protected] (E. Scarth) 3 April 2012 4 April 2012