Chapter 13
Qualitative analysis INTRODUCTION The main purpose of qualitative analysis is t o establish the identity of an, as yet, uncharacterised sample. This requirement may occur for one of several reasons: (a) The sample may have resulted from a new synthesis programme, or the new synthesis, by a different route, of an established material. It is important to establish if the product is the same as the one being sought and if impurities, particulary undesirable impurities, may have been introduced. (b) Isolation of compounds from complex naturally occuring products, e.g., alkaloids from plant material, (c) Confirmation of the identity of potentially hazardous or restricted chemicals. For example, in the enforcement of legislative procedures concerned with pollution and forensic science. All who are experienced in analytical chemical methods will be all too familiar with the problems that can arise when more attention is given t o quantitative rather than t o qualitative assessment of a sample. Typical examples are remarks like: (a) “The sample definitely contained three impurities”. These were shown much later to be impurities in the solvent used t o dissolve the sample. ( b ) “It must be pure, since it chromatographed as a single spot on a TLC plate”. (c) “Examination by LC showed the sample to be a two-component mixture”. - The separation was monitored using a UV absorption detector operating at a wavelength of 365 nm. These examples pin-point the potential dangers of making assumptions, often completely unjustified, about the nature of the sample without having established the same. The first remark quoted was simply an expression of the lack of attention t o practical detail that was given t o the examination: one injection of the solvent used t o dissolve the sample would have shown up the fault. This is the type of error which most operators are likely t o make, at least once. The second remarks made above presuppose that the separation of possible impurities on the TLC plate would have occurred if they were present in the sample, a situation which needs much extra work to establish. The third remark equally presupposes some definite knowledge that any impurities will have some definite spectral characteristics. When using highly selective detectors to study an unfamiliar sample it is recommended that a non-specific detector should also be fitted, in series if possible, to ensure that no components are missed. These examples illustrate the serious mis-interpretation that can occur
288
QUALITATIVE ANALYSIS
through insufficient attention to the qualitative aspects of a chromatographic procedure. Apart from the situations described in this section, perhaps one of the greatest sources of mis-interpretation of a LC analysis is the failure to establish whether or not the entire sample has eluted from the column: although it is easy to recognise this possibility, the work necessary to confirm that all the sample has eluted is quite difficult to achieve. One approach could be to inject a known mass of sample, collect the total column effluent and remove the solvent; in principle, weighing the residue will indicate the recovery of the sample. For this result to be accurate the mass injected should be at least 1mg and “blanks” should be run on the system t o make sure that residues from the solvents or column packing do not invalidate the results. An alternative method is t o study the sample by a steric exclusion technique as, in most cases, all the sample will elute within the region h’ = 0 and 1. This approach unfortunately only confirms that the sample will elute from a steric exclusion column and indicates nothing about the degree of retention on, say, a bonded-phase packing. Independent studies by TLC can also be helpful provided the coating on the TLC plate and the column packing are strictly comparable and the mobile phases are identical. The recommended approach is to remove spots from the TLC plate, extract the component and run it in the liquid chromatograph to establish the retention behaviour of the observed spots. METHODS OF ESTABLISHING OR CONFIRMING THE IDENTITY OF AN ELUTING PEAK From the earlier remarks it will be apparent that careful attention to the qualitative aspects of an analysis is important in all but the most predictable chromatographic separations. There are a number of methods by which the identity of the eluting component may be checked. These rely either on comparison of the chromatographic characteristics of the components with reference materials of known identity or on the characteristics of the detection system. In this sense the “detection” may be considered as either in-line, i.e., asa flow through detector, or isolated, i.e., by collecting fractions for further study. Identification methods based on comparison of retention data In principle, it could be considered that since the retention characteristics of a sample component are dependent on its chemical and physical properties, i.e., molecular size, functional groups and solubilities, correlation of retention times (or volumes) with the type of sample should provide a means of tentatively identifying an eluting component. This rather oversimplified concept has some decree of truth but if it is to be applied to the identification
ESTABLISHING T H E IDENTITY OF A PEAK
289
of eluting components without additional data considerable caution must be exercised as to the resolving power of the chromatographic system used. Work is often reported where the resolving power of the column is not capable of separating compounds of similar structure, thus the results may be strictly invalid. Any chromatographic system must offer a high resolving power for the samples being studied. In Chapter 2 it was shown that the resolving power is directly dependent on the capacity, the selectivity and the square root of the efficiency of the column. The most frequently used method of establishing the identity of an eluting component is comparison of the retention data of an “unknown” peak with the retention of a similar injection made under identical operating conditions of a reference substance, which, based on other considerations, e.g., a known synthesis precursor, could possibly occur in the sample. For this method t o be successful a number of supplementary points must be considered, First, it is imperative that the precision of the measurement of retention time must be very good in relation t o the variation of retention time due t o the selectivity of the chromatographic system for compounds of similar structure. The precision of retention times clearly is very dependent on the repeatability of operations such as recorder chart speeds, injection technique and, particularly, the flow through the column during the analysis. Errors in the measurement of solute retention due t o poor injection technique are not significant in most instances except with inexperienced operators. Errors from recorder chart speed variations are infrequent with modern instrumentation: indeed, many chromatographers rely on a digital integrator to measure retention times. On the other hand, variations in mobile phase flow occur for a number of reasons. Significant changes in flow can result from a change in the temperature of the chromatographic system and column permeability. A change in inlet pressure can affect the flow in a fairly straightforward manner, but in very high-pressure systems flow changes can be complicated by an increase in viscosity with increasing pressure or compressability of the mobile phase. Secondly, retention volumes, if derived simply by the expression “retention time X flow-rate = retention volume” can suffer from the same limitations as were discussed for retention time measurements in the previous paragraph. However, in this instance alternative methods are available in that one may, with the appropriate equipment, measure retention directly in terms of the volume of mobile phase passing through the column from the moment of injection of a sample t o its detection. This may be achieved by using a siphon counter, as described in Chapter 12, t o aid accurate assignment of elution volumes during the characterisation of molecular weights of polymers. In any situation where retention characteristics are being compared, there are several practical ways which can be employed t o endorse the tentative identification of a component.
2 90
QUALITATIVE ANALYSIS
(1)When it is believed that the “unknown” peak has been identified, prepare a 50: 50 mixture of the unknown and the anticipated reference compound and analyse the mixture. Clearly, if correctly identified only one peak will be observed. (2) If the equipment used has recycling capability (as described in Chapter 7), an even more critical test is t o recycle the mixture, prepared for the first test described above, through the chromatographic column system for as many times as it is practicable t o establish whether the “unknown” and “reference” substances can be resolved. ( 3 ) Where possible, the exercise should be repeated with other chromatographic phase systems which exhibit different types of selectivity, i.e., a normal phase, a reversed-phase system and a liquidsolid (adsorption) system. Confidence in the identification of a component is considerably increased if three different systems fail t o separate the prepared mixture of “unknown” and “reference” compounds. In studies of this nature, perhaps a negative result is far more decisive. Thus, if the retention characteristics of a reference substance and an unknown are different, they are most definitely not the same substance. On the other hand, if the rentions are identical, they may be the same substance. When tabulating retention data, use of the capacity factor term ( k ’ )is t o be recommended. This term is a measure of the effective retention of the compound on a column and is not influenced by column geometry and mobile phase flow-rate; thus comparison of results is simplified. Table 13.1 is an example of recording the results from studies of a group of compounds on two different chromatographic systems; the order of elution and extent of retention of the compounds are clearly seen from such a format. The use of retention volumes or capacity factors as characteristics of a sample is perhaps more common in the field of GC. In a number of instances the relationship between the logarithm of the retention volume and the TABLE 13.1 COMPARISON BETWEEN THE VALUES OF k’ IN RPC AND NORMAL-PHASE CHROMATOGRAPHY (NPC) (Reproduced from ref. 1 with permission.) Solute
RPC*
NPC**
p-Cresol 2,6-Xylenol 2,4-Xylenol 2,3,4-Trimethylphenol 2,4,5-Trimethylphenol
1.64 2.68 2.95 4.50 4.80
11.13 4.81 1.88 7.81 7.06
*Column, pBondapak Cla ;solvent, methanol-water ( 1 : l ) . **Column, Partisil 5; solvent, n-hexane-ethyl acetate (95 : 5 ) .
ESTABLISHING THE IDENTITY OF A PEAK
291
carbon number has been shown to be linear in a homologous series in liquidliquid partition and reversed-phase systems, in a similar manner to that observed in GLC. A vast amount of information exists relating retention characteristics of samples in GC systems with chemical structure. The very considerable bulk of such data that have been published gives some indication of the large amount of work which should be performed to ensure that one is assigning the correct structure to an eluting component. Recent studies carried out on hydrocarbon samples using adsorption [ 21 and reversed-phase [ 31 chromatography indicate the potential of a comparable method in the liquid phase. Both of these papers, however, although quite detailed, are in reality only scratching the surface of the subject. Identification methods using in-line selective detectors Perhaps the most common application of this approach is the tentative identification of components which exhibit characteristic UV or visible absorption spectra. If a photometric detector with multi-wavelength capability is employed, the analysis of a sample may be repeated several times, each run being monitored at a different wavelength. Comparison of a number of chromatograms obtained in this manner will show the size of the peaks due t o the separated components varying in accordance with their spectral characteristics, which in many instances will be known or can readily be determined. A useful adaptation of this approach is to monitor the ratio of absorbances at two different wavelengths. This ratio is numerically constant throughout the elution of a pure compound from a chromatographic column. Any change in the absorbance ratio, say from monitoring the leading edge t o the tailing edge of the same peak, indicates that the observed single peak contains more than one component. Some spectrophotometric detectors, e.g., the diode array type, are able to monitor continuously the entire UV-visible absorption spectrum during the elution of solutes from a column (see p. 108). More conventional spectrophotometers do not normally scan at sufficiently fast rates to accomplish this task. However, a more faithful representation of the spectrum is obtained by stopping the liquid flow during the time that the spectrum is recorded [ 41 . Since diffusion in the liquid phase is very slow, this method is more attractive for LC procedures than for those in the gas phase. The use of more than one detector, linked in series or parallel after the chromatographic column, can provide comparative information which reduces the possibility of incorrect assignment of the identity of a component. A simple example is the use of an UV absorbance detector in line with a differential refractive index (RI) detector. The latter will respond t o most substances, whereas the first mentioned detector is quite selective in its response. A ratio of the peak heights will normally provide the easiest method of comparing the relative responses of a “reference” and an “unknown”
292
QUALITATIVE ANALYSIS
sample. When recording the data, the sensitivity or attenuation settings of the detectors should not be overlooked. Comparisons of this kind are best performed on the same instrument as post-column band broadening will reduce the peak height, especially in the second detector, if they are coupled in series. The relative response of, say RI/UV, should be a characteristic of a compound provided both detectors are operating within their linear range; injections of samples of different masses will check this point. Other types of detectors that provide good qualitative information include UV in combination with electrochemical or fluorescence detectors. Fig. 13.1 illustrates the simultaneous UV/fluorescence detection of LSD (lysergic acid diethylamide) in illicit tablets. The combination of detection methods in this manner reinforces the certainty of identification of the components being sought. Detector non-linearity can be somewhat more acute in fluorescence measurements as the absorbance of a compound will reduce the intensity of the excitation radiation, which in turn will lower the fluorescent emission. This phenomenon is known as the “inner filter effect” and can cause some non-linearity in the response of a fluorescence detector if the background
0
2 4 6 Time (minutes)
Fig. 13.1. Use of combined fluorescence ( A ) and absorbance (B) detection for increased confidence in the identification of LSD in illicit tablets. Operating conditions: column, 0.25m x 2.1 mm I.D.: packing, Zorbax SOIL; mobile phase, methanol-dichloromethane acetic acid (30: 70 :0.1); temperature 24 C;inlet pressure, 8 MPa (1200p.6.i.); flow-rate, 0.6 cm3/min; detection by UV absorbance at 334 nm (0.08 a.u.f.6.) and fluorescence (16nA full scale; excitation wavelength 334 nm; emission wavelength 408 nm and above). (Reproduced by courtesy of Du Pont and from ref. 11with permission.)
ESTABLISHING THE IDENTITY OF A PEAK
293
absorbance (at the wavelength of the excitation radiation) is in excess of approximately 0.05 absorbance units. Monitoring of column effluents by mass spectrometry The combination of GC-MS and computerised data handling systems has proved to be one of the most powerful analytical methods for identifying minute components which may be present in chemical samples. Perhaps the greatest successes have been in its application in the fields of forensic science, pollution and biochemistry. Following the impact of the GC-MS technique, it is logical t o consider a similar approach involving LC, particularly since the separation of a wide range of sample types may be studied. As outlined in Chapter 5, there are several fundamental differences associated with the concept of an interfaced LC-MS system. First, the sample will in general be non-volatile, polar and/or of high molecular weight: if it were volatile, then GC should have been the chromatographic method to employ. Secondly, the mobile phase is very dense, particularly when it is compared to helium, the most popular carrier gas in GC-MS. There is also the risk of corrosion within the spectrometer due to the solvent itself or dissolved solids, e.g., from buffer solutions from the effluent of an ion-exchange separation. On the other hand, quantitative collection of a sample from an effluent leaving a liquid chromatograph is very easily accomplished as the component is in solution. If the identity of a compound is being sought, a portion of such acollected fraction can be evaporated on to a direct insertion probe and introduced manually into the mass spectrometer, thus avoiding the inherent problems associated with an in-line system. Mass spectral measurements of compounds eluting from a liquid chromatograph can be made in several ways. The simplest is to collect the fraction of column effluent containing the sample and determine its spectral characteristics as an independent exercise. Alternatively, the column effluent may be “sampled” automatically, either in a discontinuous or in a continuous manner. Instrumental requirements for these methods vary considerably and are described in the following paragraphs. Manual procedure for evaporating the collected fractions and examining the residue This method may be very simply achieved if, as is usual, the vapour pressure of the mobile phase is considerably higher than that of the sample. Juhasz et al. [5] have described the application of a refinement of this method, whereby the column effluent is collected in a small sample tube containing 6 mg potassium bromide. A steady stream of dry nitrogen is fed into the small sample tube via a hypodermic needle t o assist the evaporation of the solvent. In this manner the residue of the collected fraction is coated
294
QUALITATIVE ANALYSIS
on to the potassium bromide, which is subsequently formed into a disc suitable for examination in an IR spectrophotometer fitted with beam condensing optics. The potassium bromide disc may be subsequently transferred to the direct insertion probe of a mass spectrometer. On heating the probe, the sample is vaporised into the spectrometer allowing a spectrum of good quality to be obtained. This approach possesses the distinct advantage of simplicity and does not require a complex interfacing system. The greatest limitation is the amount of operator handling of the sample which can increase the possibility of contamination of the fraction and also lead to loss of sample, particularly during the evaporation of the solvent.
Semiuutomated sample collection and insertion into a spectrometer Lovins et al. [6] have described a liquid chromatograph-to-mass spectrometer interface where a motor-driven insertion probe is employed. In this approach the fraction of column effluent which is to be studied is initially collected in a small reservoir. On opening a valve, the solution passes through narrow-bore capillary tubing to the tip of the probe, where the solvent is flash evaporated in the reduced atmosphere of the fore-chamber. When this stage is completed, the valve controlling the entry of solution is closed automatically, the fore-chamber is reduced to low pressure whereon a highvacuum valve isolating the mass spectrometer ion source from the forechamber is opened and the tip of the sample probe is advanced into the ion source. All operations are accomplished by motorised components giving the interface a semi-automatic capability. By this method the complete operation from sample collection to obtaining a mass spectrum is reported as taking 3--5min with minimal operator attention. In common with the first method described, some loss of sample has been observed when the solute has a fairly high vapour pressure at the temperature at which the mobile phase is being flash evaporated.
In-line, coupled liquid chromatograph-mass spectrometer systems Several approaches have been described, notably by Horning et al. [7], Arpino et al. [8] and Jones and Yang [9] whereby the effluent from a liquid chromatograph is introduced into a mass spectrometer in a similar manner to that employed in GC-MS. The interfaces described t o date are generally fairly simple in design when one considers that they have to overcome the very large difference in sample environment, i.e., from solution in a liquid at high pressure to a vapour in a high-vacuum system. Not surprisingly, a good deal of the emphasis in design has been given to providing a sufficiently high pumping capacity in the MS analyser to avoid the pressure within the system exceeding approximately
OTHER CONSIDERATIONS
295
Torr, at which point the spectrometer will no longer operate efficiently. With the exception of very corrosive mobile phases, e.g., acids and buffer solutions, it would appear that simultaneous introduction of solvent and sample molecules can simplify the mass spectra by providing an “atmosphere” within the ion source comparable to that employed in chemical ionisation mass spectrometry. This latter variation of the technique allows simple spectra to be obtained from labile substances, often showing molecular ions. When using electron impact for ionisation, similar molecules are more completely fragmented, leading t o a spectrum showing ions of much lower m/e value, which can be confused with ions produced from the molecules of mobile phase. Homing et al. [ 101 have illustrated the practical utility of this approach using a quadrupole mass spectrometer fitted with a 63Ni radioactive ionisation source operating at essentially atmospheric pressure. This ionisation source is situated at the end of a heated capillary from which the column effluent is vaporised into a nitrogen stream immediately adjacent to an aperture, 10-25pm in diameter, leading into the mass spectrometer. As indicated in Chapter 5, many approaches are currently being investigated to effect the optimum coupling of LC and MS. Packed microbore columns with internal diameter of 1mm or less play an important role in these studies (see p. 52). One of the most interesting uses of a combined LC-MS system comes from the inherent “tuning” characteristics of the spectrometer. As in GC-MS it is possible to focus the spectrometer t o any desired m / e value and record the variation in the concentration of that ion with time. In applications where specific chemical species are being sought, for example, drug metabolism studies and pesticide residue analysis, most of the co-extracted substances which could interfere with a conventional chromatographic analysis will be rejected by the m / e value set on the mass spectrometer. This arrangement leads to high sensitivity and very selective detection of the components of interest. It is quite probable that this last-mentioned approach could well play a major role in the future analytical chemical research studies in areas of toxicology, metabolism and pollution. The most serious drawback of the technique will no doubt be the high cost of the equipment required, particularly if, as is often the case, a computerised data handling system proves necessary. OTHER CONSIDERATIONS WHEN SEEKING TO IDENTIFY AN ELUTED COMPONENT While on the subject of the identification of components eluting from a liquid chromatograph it is important to bear in mind the purity of the solvents used in the separation process. Clearly, when seeking to collect a sample component for further study it is imperative to select solvents which are free
296
QUALITATIVE ANALYSIS
from any non-volatile impurities and fairly easily vaporised. Careful distillation of all solvents prior to use will normally prevent any difficulties in respect of the mobile phase. In this application the use of packing materials which have the stationary phase bonded chemically to the support is to be recommended as these will not normally bleed stationary phase. In critical investigations, especially where trace components are being concentrated for identification, the possible presence of small proportions of organosilane reagents or their decomposition products should not be ignored. In an analogous manner, the ability to select a phase system that will enable minor components which must be identified to elute before the major components of a sample will greatly facilitate the collection of pure materials. In this way the chromatographic system may be overloaded significantly with respect to the major component in order to collect a larger amount of the impurity. If the minor component elutes later in the chromatogram, contamination by residual amounts of the major component is frequently encountered.
REFERENCES 1 2 3 4 5 6
7 8 9 10
11
H. Colin and G. Guiochon, J. Chrornatogr., 158 (1978)183-205. M. Popl, V. Dolansky and J. Mostecky, J. Chromatogr., 91 (1974)649-658. R. B. Sleight, J. Chromatogr., 83 (1973)31-38. A. M. Krstulovic, R. A. Hartwick and P. R. Brown, J. Chromatogr., 163 (1979) 19-28. A. A. Juhasz, J. Omar Doali and J. J. Rocchio, Amer. Lab., 6, No. 2 (1974)pp. 23-24,26, 28-29. R. E. Livins, S. R. Ellis, G. D. Tolbert and C. R. McKinney, Anal. Chem., 45 (1973) 1553-1 556. E. C. Horning, D. I. Carroll, I. Dzidic, K. D. Haegele, M. G. Horning and R. N. Stillwell, J. Chromatogr., 99 (1974)13-21. P. J. Arpino, B. D. Darokins and F. W. McLafferty, J. Chromatogr. Sci., 12 (1974) 574-578. P. R.Jones and S. K. Yang, Anal. Chem., 47 (1975)1000-1003. E. C. Horning, D. I. Carroll, I. Dzidic, K. D. Haegele, M. G. Horning and R. N. Stillwell, J. Chromatogr. Sci.,12 (1974)725-730. D.R. Baker, R. C. Williams and J. C. Steichen, J. Chromatogr. Sci., 12 (1974)499.