Journalof
VirologicalMethods,
EIsevier/North-Holland
COMPARISON
‘Groupe
99
ELECTRON MICROSCOPY,
AND COUNTERIMMUNOELECTROPHORESIS
ROTAVIRUS
G. OBERTI, R. GLOECKLER’, Facultk
99-107
OF IMMUNOSORBENT
IMMUNOASSAY OF H~AN
3 (1981)
Biomedical Press
de Recherches de MPdecine:
ENZYME
FOR DETECTION
IN STOOLS
J. BURCKARD’ and M.H.V. VAN REGENMORTEL’
SW In Pathogknie and ‘lwtitut
des Infections
Virales U 74, Laboratoire
de Biofogie MoEculaire
et Ceilulaire,
de Virologie,
15 rue Descartes,
67000
Stra~bourg, France
(Accepted 23 March 1981)
The detection
of human rotaviruses by routine electron microscopy examination
of stool speci-
mens has been compared with the sensitivity of detection obt~nable by three different immunoassays. These assays are: 1) immunosorbent electron microscopy (ISEM), which consists of the ~~ologi~d trapping of viruses on electron microscope grids coated with protein A and specific viral antiserum; 2) an enzyme-linked immunosorbent assay (ELISA), in which the primary antibody is rabbit antirotavirus immunoglobulin, the secondary antibody is chicken anti-rotavirus immunoglobulin extracted from egg yolk of immunized hens, and the indicator antibody is alkaline phosphatase-conjugated rabbit anti-chicken immunoglobu~n~ 3) counterimmun~electrophoresis (CIE). A total of 63 stool specimens from infants with gastroenteritis were examined. Of these, 23 and 24 specimens were found to contain rotavirus by eIectron microscopy and CIE, respectively. When scored by ELISA and ISEM, 37 and 39 were found to be positive, respectively. Confirmatory inhibition assays were necessary to eliminate some false positive reactions in ELISA. Detection of human rotaviruses in stools by ISEM is as sensitive as by ELISA, but in weakly positive specimens, ISEM offers the additional advantage of a direct visual demonstration of the presence of the aetiological agent. INTRODUCTION
Rotavirus children
infections
(McNulty,
in stool specimens
represent
one of the major causes of gastroenteritis
1978; Holmes, by electron
1979). The causative agent can be detected
microscopy
ex~nation
(Kap~ian
in young directly
et al., 1974), but
in many cases the virions can be visualized only after they have been concentrated by ultracentrifugation. Several serological techniques such as counterimmunoelectrophoresis (Middleton et al., 1976; Grauballe et al., 1977), enzyme-linked immunosorbent assay (Ellens and De Leeuw, 1977; Scherrer and Bernard, 1977; Zissis and Lambert, 1980), radio~unoa~ay (Kalica et al., 1977; Middleton et al., 1977; Sarkkinen et al., 1980), enzyme-linked fluorescence assay (Yolken and Stopa, 1979) and immunosorbent electron microscopy (Nicolai’eff et al., 1980) have also been used for detecting rotavirus in faecal material. Most of these immunological assays are more sensitive than routine electron microscopy, and also allow a more rapid detection of rotavirus in stool specimens. Ot66-0934/81/0000--0000!$02.5#
Q !+cv
, ~(~rt~~-~~l~~ll~.~~d Biomedical
Press
100
In the present study, we compared
the sensitivity
of detection
of rotavirus in human
stools by electron microscopy with the sensitivity achieved by three different immunoassays: immunosorbent electron microscopy (ISEM), an enzyme-linked immunosorbent assay (ELISA),
and counterimmunoelectrophoresis
1973; Roberts
and Harrison,
1979; Nicolaieff
(CIE). The ISEM technique et al., 1980) consists
(Derrick,
in the serological
trapping of viruses on electron microscope grids coated with specific ~tise~m against the virus. The particular ELISA method that was used utilizes chicken anti-rotavirus immunoglobulin extracted from egg yolk of laying hens immunized with rotavirus (Polson et al., 1980; Van Regenmortel and Burckard, 1980) and an alkaline phosphatase rabbit anti-chicken globulin conjugate. MATERIALS
AND METHODS
Specimens A total of 63 stool specimens were obtained from 63 infants admitted to the hospital in Strasbourg with symptoms of acute gastroente~tis. Four volumes of d~t~led water were added to one volume of faecal material and the mixture was homogenized and clarified by low-speed centrifugation (5000 g for 20 min). Purification
ofrotavirus
The procedure described by Rodger et al. (1975) was applied to a stool which had been found by electron microscopy to contain a large number of rotavirus particles (Nicolai’eff et al., 1980). The concentration of virus was expressed by the protein content determined by the method of Lowry et al. (195 1).
Rabbits
and laying hens were immunized
by a series of intramuscular
injections
of
rotavirus in adjuvant as described previously (Nicolaieff et al., 1980). Rabbit immunoglobulins were prepared by precipitation from antiserum with an equal volume of 4 M ~o~urn sulphate. Chicken ~munoglobu~s were obtained from the egg yolk of immunized hens as described previously (Polson et al., 1980). Routine electron microscopy The clarified stool extract was applied to carbon-coated electron microscope grids (300 mesh). After staining with 2% phosphotungstic acid (pH 7.0) the grids were screened for 15 min for the presence of rotavirus particles in a Philips EM 300 electron microscope at a magnification of 39,000 X . If a single particle was found, the specimen was scored as positive.
101
Immunosorbent electron microscopy Electron
microscope
pl drops of protein
grids, coated with a ffirn of formvar-carbon,
were floated on 50
A solution (26 pg/ml) for 4 min and were then placed successively on
three drops of 0.1 M sodium phosphate buffer (pH 7.0) for 1 min. Thereafter, the grids were floated for 10 min on 50 p.l drops of rabbit anti-rotavirus serum diluted 1 : 500. After an intermediate rinse on a drop of phosphate buffer, the antibody-coated grids were left overnight
on drops of stool extracts
diluted
with an equal volume of 0.2 M
phosphate buffer. After the serological trapping reaction, the grids were rinsed by placing them on a series of drops of distilled water. Virions were visualized by staining with 1% uranyl acetate in 45% ethanol for 2 min. Grids were screened for 5 min at an instrumental magnification of 8 100 X . Four areas were photographed with 70 mm film. Particle counts were made with a binocular
magnifier
on four 68 X 73 mm micrographs
(N’icolaieff et
al., 1980). Counterimmunoelectrophoresis The method used was similar to that described by Middleton et al. (1976). One percent agarose (Indubiose, Pharmindustrie, Clichy, France) in 0.05 M barbital buffer (pH 8.6) was applied to microscope slides. Stool extracts were placed in the cathode wells and specific rabbit antiserum in the anode wells. After electrophoresis at 10 V/cm for 2 h, the slides were washed in saline and stained with 1% tannic acid. Enzyme-linked immunosorbent assay An indirect ELISA method utilizing chicken and rabbit rotavirus antibodies was used (Scherrer and Bernard, 1977; Van Regenmortel and Burckard, 1980). Polystyrene microtitre plates (Cooke M 129 B, Dynatech) were used. Wells were coated at 37°C by 1 h incubation with 250 pl of rabbit immunoglobulins (10 pg/ml) diluted in 0.05 M sodium carbonate, saline, pH albumin in cubated at
pH 9.6. Plates were rinsed three times with 300 pl of phosphate-buffered 7.4, containing 0.05% Tween-20 (PBS-T) and once with 1% bovine serum PBS-T. Dilutions of purified rotavirus or of stool extracts in PBS-T were in37°C for 2 h in the coated wells. After rinsing, chicken immunoglobulins
(5 Erg/ml) in PBS-T were allowed to react at 37°C for 2 h with the trapped virus. After further rinsing, the wells were incubated for 2 h at 37°C with an anti-chicken globulin conjugate diluted 1 : 800. This conjugate was prepared by coupling IgG obtained from a rabbit immunized with chicken immunoglobulins with alkaline phosphatase (Boehringer, Mannheim), as described previously (VanRegenmortel and Burckard, 1980). After further rinsing, the bound enzyme conjugate was detected by adding 250 pl of the substrate p-nitrophenyl phosphate at 1 mg/ml in 0.1 M diethanolamine buffer, pH 9.8. After 1 h hydrolysis at 37”C, the tests were read with a photometer (Vernon, Paris) by measuring the absorbance at a wavelength of 405 m-n through the bottom of the microtitre plate.
102
Conjirrnato y ELISA In order to test the specificity absorbance kinen
and confirm
the validity of ELISA results in which the
values were lower than 0.7, a blocking
et al. (1980)
recently
showed
inhibition
the importance
test was performed.
of such confirmatory
Sark-
inhibition
tests to prove the specificity of low positive reactions. The blocking test was done as follows: 250 /_dvolumes of the specimens were added to three wells coated with rabbit anti-rotavirus globulins. After a 2 h incubation at 37”C, a 2.50 /.d volume of rabbit antirotavirus serum (1 : 200 dilution) was added to the first well and the same volume of normal rabbit serum (1 : 200) to the second well; the third one was filed with 250 fi PBS-T. After an incubation of 2 h at 37”C, the subsequent steps of the ELISA were performed as described above. The test was considered positive if a 50% or greater decrease in absorbance was obtained with the specimen incubated with rotavirus antiserum as compared with normal serum or dilution
buffer (Sarkkinen
et al., 1980).
RESULTS
The sensitivity of rotavirus detection by ELISA and ISEM was determined with a series of dilutions of purified virus, calibrated according to protein content. As shown in Fig. 1, as little as 2 ng/ml of purified rotavirus antigen could be detected by ELISA. In these tests, twice the buffer blank absorbance value was taken as the cut-off line. When the same dilutions of purified virus were examined by ISEM, the sensitivity of detection was essentially the same as by ELISA (Table 1). When 63 stool specimens were screened for the presence of rotavirus by four different
I
’ 1.25
’
2.5
1 5
ROTAVIRUS
Fig. 1. Sensitivity the means readings
’ 10
ANTIGEN
of detection
of readings
and corresponds
’ 20
from
’ 40 hg/ml)
of purified three
rotavirus
wells. The dotted
antigen
by ELISA.
line represents
to twice the value of the buffer
blank.
The absorbance
the cut-off
point
values
are
for significant
103
TABLE
1
Sensitivity
of rotavirus
detection
by immunosorbent
Rotavirus
antigen
40
electron
concentration
microscopy
(ng/ml)
20
10
5
2.5
1 0
Experiment
1
32a
25
11
3
2
Experiment
2
51
17
7
5
4
1
Experiment
3
47
23
10
7
3
0
130b
65
28
15
9
1
Total a
The values
for the three
separate
experiments
mm micrographs
taken at an instrumental
b
counts
Total
particle
represent
the total particle
counts
on three 68
x
73
of 8100 X
magnification
on nine micrographs.
techniques, the results presented in Table 2 were obtained. The sensitivity of detection by electron microscopy and CIE was similar and led to 36-38% of the samples being scored as positive. For these two techniques, the stool extracts were used undiluted and at a dilution of 1 : 2 respectively. By ELISA and ISEM, 59-6175 of the specimens were scored as positive when the stool extracts were diluted 1 : 20 and 1 : 2 respectively. The superiority of ELISA and ISEM was further confirmed by examining higher dilutions of the specimens by these two techniques TABLE
(Table 3). In the case of samples that contained
2
Comparison
of four methods
for detecting Rotavirus
No. of samples
rotavirus
in 63 stool specimensa
detection
by
Electron
CIE
ELISA
+
+
ISEM
microscopy 22
+
1
+
+ t
2
+ +
10 4
tb
2 22 Total:
63
100% a
Stool
CIE; 1 b
24
37
39
36%
38%
59%
61%
were examined
at the following
dilutions:
1
: 1 for electron
microscopy;
1
: 2 for
other
tech-
: 20 for ELISA, and 1 : 2 for ISEM.
Initially,
niques. c
specimens
23c
four samples
After confirmatory
The figures
represent
were scored ELISA
as positive
by ELISA
on these four samples
the total number
of samples
and as negative
by the three
(see Table 4) only two remained
scored
as positive
by each technique.
positive.
104
TABLE
3
Representative
results
weakly
stool specimens
Specimen
positive No.
of rotavirus
Method
of
detection
1
ELISA
0.875
ISEM
39
ELISA
0.65
ISEM a
Absorbance
b
Total
particle
1
451
ELISA
4
: 20
1
2.25
ISEM 3
of specimen
323b
ELISA
detection
Dilution
2.4a
ISEM 2
antigen
22
by ELISA
: 200 1.8
44 2.05 53
1
and ISEM with two highly
: 2000
1
: 20,000
1
and two
: 200,000
1.05
0.35
0.20
8
0
0
1.25
0.375
0.175
0
0
12
0.40
0.200
0.200
NTC
4
0
0
NT
0.35
0.200
0.200
NT
1
0
0
NT
at 405 nm. counts
on three
68 X 73 mm micrographs
taken at an instrumental
magnification
of
8100x. c
Not tested.
large numbers of particles, detection was still achieved at a specimen dilution of 1 : 2000, whereas by CIE and electron microscopy, the limit of detection was reached at a dilution of 1 : 20. In the case of weakly positive specimens, the limit of detection by ELBA and ISEM was reached at a dilution of 1 : 200. The absorbance values in the ELISA were then very close to the cut-off value for a positive result (Fig. l), whereas in ISEM, the low particle counts still allowed an unambiguous scoring of the specimens as positive. As shown in Table 2, four specimens were positive by ISEM and negative by ELISA, a finding which is in agreement with the slightly superior sensitivity of ISEM demonstrated in Tables 1 and 3. It was therefore somewhat surprising that the reverse situation was also found to exist, and that four specimens were initially scored as positive by ELISA and as negative by ISEM. In view of the possibility of non-specific binding in solid-phase immunoassays (Sarkkinen et al., 1980) these samples, together with all specimens with absorbance values below 0.7, were examined by confirmatory ELISA. Table 4 illustrates representative results obtained in four such confirmatory inhibition assays. Two of these tests were scored as negative, since they could not be specifically inhibited by rotavirus antiserum. DISCUSSION
Middleton et al. (1976) found CIE to be about 4-8 times less sensitive than electron microscopy, whereas Grauballe et al. (1977) reported that it was about 2--4 times more sensitive. Our own results agree with those of Tufvesson and Johnsson (1976) and Spence
10.5
TABLE 4 Confiimatory
ELBA inhibition test for establishing the specificity of rotavirus detection Absorbance when specimen incubated with dilution buffer
Absorbance when specimen incubated with rotavirus antiserum
Absorbance when specimen incubated with normal
2
0.85 0.75
0.325 0.275
0.75 0.85
3 4
0.55 0.50
0.175 0.225
0.200 0.200
Control antigen GO n&n)
2.15
0.75
1.85
Specimen No.
1
Interpretation of the test
SeNrn
+ +
+
et al. (1977) who reported that both techniques are about equally sensitive for detecting ro taviruses. Compared to electron microscopy and CIE, solid-phase immunoassays such as ELISA and radio~uno~say are generally considered to allow the detection of lower concentrations of viral antigens (Daugbarty and Ziegler, 1977; Middleton et al., 1977; Van Regenmortel, 198 1). Our own results (Table 2) confirm the findings of previous workers who showed that ELBA is a highly sensitive method for detecting rotaviruses in faecal extracts (Yolken et al., 1977; Ellens and De Leeuw, 1977; Scherrer and Bernard, 1977). The main d~advantage of solid-phase ~~unoassays lies in the possible occurrence of false positive reactions caused by non-specific binding (Sarkkinen et al., 1980). Since we did not observe any false positive reactions with specimens that gave ELISA absorbance values higher than 0.7 (all these specimens were positive by ISEM), we found it possible to eliminate non-specific reactions by performing confirmatory inhibition assays on all specimens with absorbance values below 0.7. The ISEM technique was equally sensitive as ELISA, but in weakly positive specimens, the direct visualization
of particles provided
a distinct
advantage
since no confirmatory
tests were necessary. When examing stool specimens by ISEM, we found that one microscopist could handle approximately 80 samples a day. Since microscope grids coated with protein
A and antiserum
can be stored for several weeks before use, they can be used
immediately as stool specimens become available. ~t~e~rn-coated grids can also be dispatched to centres that have no microscope facilities and returned for examination after completion of the trapping reaction. The trapping of virus particles on antiserumcoated grids occurs very rapidly and, with most samples, particles are visible after 60 min
incubation.
trapping Nicola’ieff
reaction
However, proceed
for maximum overnight.
sensitivity
As discussed
et al., 1980; Van Regenmortel,
of detection, previously
(Mime
198 l), the 1SEM procedure
we routinely
let the
and Luisoni, is simple
1977;
and rapid
106
and leads to fairly clean preparations, even when the stool extracts are allowed to react with the grids for 18 h. In weakly positive specimens, virus particles could be visualized within 5 min. The layer of serum proteins present on the grids prevents the adsorption of contaminants,
and as a result, small numbers
of virus particles are more easily visual-
ized than by routine electron microscopy. The primary coating of protein A on the grids considerably increases the efficiency of trapping by the adsorbed antibodies, since the IgG molecules are attached to the grid by their Fc regions and have their binding sites preferentially exposed (Shukla and Cough, 1979; Lesemann and Paul, 1980). The ISEM procedure has been used extensively in plant virology (Derrick, 1978; Milne and Lesemann, 1978; Roberts and Harrison, 1979; Lesemann et al., 1980; Nicolalieff and Van Regenmortel, 1980) and is also likely to find many applications in the diagnosis of human and animal virus infections. REFERENCI:S
Daugharty,
H. and
(Academic
D.W. Ziegler,
Derrick,
K.S., 1973, Phytopathology
Derrick,
K.S., 1978, Science PC.,
Holmes, Kaliq
J. Genner, R.H.
1977, J. Clin. Microbial.
A. Meyling
Purcell,
1977, J. Immunol. Kapikian,
A.Z.,
R.G.
1974, Science
and C. Kurstak
35,203.
M.M. Sereno,
R.G. Wyatt,
H.W. Kim, R.M. Chanock
and A.Z. Kapikian,
118, 1275. Wyatt,
W.J.
Rodriguez,
S. Ross,
W.L. Clin c, R.H.
Parott
and
R.M. Chanock,
185, 1049.
D.E. and H.L. Paul, 1980, Acta Hortic. D.E., R.F. Bozarth
and R. Koenig,
O.H., N.J. Rosebrough, MS.,
1977, J. Gen. Viral.
25, 1.
Lesemann, Lowry,
eds. L:. Kurstak
6,530.
and A. Hornsleth,
Lesemann, McNulty,
Diagnosis,
63, 538.
I.H., 1979, Prog. Med Viral. A.R.,
in: Comparative
199,538.
Ellens, D.J. and P.W. De Leeuw, Grauballe,
1977,
Press, New York) Vol. 2, p. 459.
110, 119.
1980, J. Gen. Viral. 48, 257.
A.L. Parr and R.J. Randall,
1951, J. Biol. Chem.
193, 265.
1978, J. Gen. Viral. 40, 1.
Middleton,
P.J., M. Pet&,
Middleton,
P.J.,
C.M. Hewitt,
M.D. Holdaway,
M.T. Szymanski
M. Petric,
and J.S. Tam,
M.T. Szymanski
1976, J. Clin. Pathol.
and J.S. Tam,
1977, Infect.
29, 191.
Immun.
16,
439. Mime, R.G. and E. Luisoni, (Academic
1977,
in: Methods
Milne, R.G. and D.E. Lesemann,
1978, Virology
Nicolaieff,
A. and M.H.V. Van Regenmortel,
Nicolafeff,
A., G. Obert
Polson, Rodger,
S.M., R.D. Schnagl
Scherrer,
R. and S. Bernard,
Shukla,
D.D. and K.H. Gough,
Spence,
L., M. Fauvel,
and PI<. Halonen, 1977,
1975, J. Viral.
1980, Immunol. 16, 1229.
1980, J. Viral. Methods
Ann. Microbial.
(Inst. Pasteur)
I, 33 I.
128 A, 499.
1979, J. Gen. Virol. 45,533.
R. Petro and S. Bloch,
B. and T. Johnsson,
1980, J. Clin. Microbial.
1979, Ann. Appl. Biol. 93, 289.
and I.H. Holmes,
H .K., H. Tuokko
Tufvesson,
and H. Koprowski
90,299.
and M.H.V. Van Regenmortel,
I.M. and B.D. Harrison,
Sarkkinen,
eds. K. Maramorosch
1980, Ann. Viral. (Inst. Pasteur)
and M.H.V. Van Regenmortel,
A., M.B. Von Wechmar
Roberts,
in Virology,
Press, New York) Vol. 6, p. 265.
1977, J. Clin. Microbial.
1976, Acta Pathol.
Microbial.
&and.
5, 248. B 84. 225.
13 1 L, 95. 12, 101. Commun.
9,475.
107
Van
Regenmortel, Wagner
(Plenum
Van Regenmortel,
M.H.V.,
1981,
M.H.V.
R.H. and P.J. Stopa,
Yolken,
R.H.,
1980, Virology
1979, J. Clin. Microblol.
H.W. Kim, T. Clem,
Virology,
and
R.R.
R.G. Wyatt,
106, 327.
10, 317.
A.R. Kalica,
2, 263.
Zissis, G. and J.P. Lambert,
eds. H. Fraenkel-Conrat
Vol 17, p. 183.
and J. Burckard,
Yolken,
Lancet
in: Comprehensive
Press, New York)
1980, J. Clin. Microbial.
11, 1.
A.Z. Kapikian
and R.M. Chanock,
1977,