JACC: CLINICAL ELECTROPHYSIOLOGY
VOL. 2, NO. 6, 2016
ª 2016 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION
ISSN 2405-500X/$36.00
PUBLISHED BY ELSEVIER
http://dx.doi.org/10.1016/j.jacep.2016.09.008
EDITORIAL COMMENT
Contact Force–Guided Pulmonary Vein Isolation The Quest for Perfection Continues* Saurabh Kumar, BSC(MED)/MBBS, PHD,a Hugh Calkins, MD,b Gregory F. Michaud, MDa
T
he past few years have seen a burgeoning up-
feedback, catheter stability on fluoroscopy or map-
take in the use of contact force (CF)–sensing
ping
technology in catheter ablation of atrial
impedance (baseline or after ablation), had only
fibrillation (AF). On the basis of a number of pre-
modest predictive value for actual tissue contact
clinical studies, it is well accepted that CF is a key
(5,6). Even the most experienced operators quickly
determinant of lesion size, volume, and depth (1,2).
realized that recognition of both high and low tissue
There are a number of key drivers for the rapid accep-
contact in the absence of CF monitoring was impre-
tance of CF-sensing technology. The first is the
cise (4,6). A number of subsequent retrospective,
suboptimal procedural success rate, even for parox-
case control, and prospective studies demonstrated
systems,
electrogram
characteristics,
and
ysmal AF (w80% freedom from atrial arrhythmias
that CF sensing reduced procedural and fluoroscopy
at $3 years of follow-up after multiple procedures)
time, improved impedance falls, reduced radiofre-
despite a well-defined mechanism attributed to the
quency (RF) ablation time required to isolate PVs,
pulmonary veins (PVs) (3). Ablation failure is thought
and decreased the likelihood of residual gaps after
to be due to nontransmural lesions and gaps in the
completion of the anatomical circumferential abla-
ablation line; it is hoped that CF sensing might over-
tion ring and of acute and chronic PV reconnection
come the limitation of lesion nontransmurality by
when
avoiding low force applications (4). It was also hoped
(7,8). However, it also emerged that simply using
that monitoring CF would lower the risk of complica-
the CF-sensing catheter or obtaining a higher abso-
tions, particularly the risk of cardiac tamponade, by
lute CF did not necessarily translate to improved clin-
preventing excessive CF leading to atrial perforation
ical outcomes such as AF recurrence, suggesting the
and/or steam pop formation (4). Another driver for
presence of other confounding factors mediating the
the rapid acceptance of CF sensing is the realization
relationship
that surrogate markers for contact, such as tactile
Furthermore, there remains significant interest in
compared
with
between
non–CF-sensing
CF
and
catheters
outcomes
(9–11).
elucidating the optimal CF required to achieve maximal efficacy and minimal complications. *Editorials published in JACC: Clinical Electrophysiology reflect the views
SEE PAGE 691
of the authors and do not necessarily represent the views of JACC: Clinical Electrophysiology or the American College of Cardiology. From the
a
Cardiovascular Division, Brigham & Women’s Hospital,
The recent contribution of Reddy et al. (12), in this
Department of Medicine, Harvard Medical School, Boston, Massachu-
issue of JACC: Clinical Electrophysiology examines 2
setts; and the bDivision of Cardiology, Johns Hopkins University School of
critical
Medicine, Baltimore, Maryland. Dr. Kumar is a recipient of the Neil
and interlesion distance, and their relationship to
Hamilton Fairley Overseas Research scholarship, which is cofunded by the National Health and Medical Research Council and the National Heart
confounders,
namely
catheter
stability
AF recurrence at 12-month follow-up in a sub-
Foundation of Australia; and is the recipient of the Bushell Travelling
analysis
Fellowship funded by the Royal Australasian College of Physicians. Dr.
SMARTTOUCH Catheter for the Treatment of Symp-
Calkins is a consultant for Medtronic. Dr. Michaud has received consul-
tomatic Paroxysmal Atrial Fibrillation) multicenter
ting fees/honoraria from Boston Scientific, Medtronic, and St. Jude Medical; and research funding from Boston Scientific and Biosense Webster.
from
the
SMART-AF
(THERMOCOOL
study (9). Patients with drug-refractory paroxysmal AF underwent PV isolation with confirmation of entrance
Kumar et al.
JACC: CLINICAL ELECTROPHYSIOLOGY VOL. 2, NO. 6, 2016 NOVEMBER 2016:700–2
block using the SMART-TOUCH CF-sensing catheter and the CARTO-3 mapping system (Biosense Webster,
Contact Force–Guided Pulmonary Vein Isolation
F I G U R E 1 Factors Confounding the Relationship Among CF, Lesion Size, and
Ablation Outcomes
Diamond Bar, California). Of the 162 patients in the SMART-AF study, 40 patients in whom complete CF and ablation stability data from the VISITAG module (Biosense Webster) were available were the subjects of the present analysis. Of note, the VISTAG module was retrospectively analyzed for lesions that met post hoc specified, including stability of 2 mm giving consideration to cardiac and respiratory movement, application of RF energy for a minimum of 10 s at each ablation site, and a minimum CF of 5 g. The average CF during the procedure was classified into tertiles (<6.5 g, 6.5 to 10.3 g, and >10.3 g). The main finding was the presence of a U-shaped relationship between average CF and 12-month success: a CF range of 6.5 to 10.3 g was associated with a 3-fold increase in procedural success compared with <6.5 g, whereas CF ranges of <6.5 g and >10.3 g yielded a lower and similar procedural success. Moreover, procedural success increased when stable CF was applied $73% of the time within the investigator’s pre-selected CF range. Last, patients with lesions that did not meet pre-defined VISITAG criteria with an interlesion distance >10.6 mm had 3-fold worse procedural success compared with patients who did not (12).
A myriad of factors potentially explain the variable relationship among CF, lesion size, and outcomes in the published data. Superscript numbers are reference numbers. CF ¼ contact force; FTI ¼ force-time integral; PV ¼ pulmonary vein.
The results of these well-conducted analyses strengthen emerging data that the contact quality, not just the absolute quantity of CF is likely a significant
applications when strict criteria for catheter stability
contributor to procedural success. Indeed, Shah et al.
were applied. Higher CF did, however, increase the
(1) elegantly demonstrated 6 years ago that constant
extent of tissue edema as assessed by cardiac magnetic
compared with variable and intermittent contact
resonance, suggesting potential for future gap forma-
(where there is loss of CF in diastole) resulted in greater
tion as the lesion edge recovers from tissue edema (14).
lesion size (1). A lower average CF made it more likely
Continuity index (number of catheter movements
that contact was intermittent, with 51% of lesions
between subsequent RF applications) (15), catheter
exhibiting diastolic loss of CF (0 g) if the average
stability in addition to contact (16), effects of cardiac
CF was <4 g compared with 3% if average CF was
and respiratory motion (17), anatomic location of CF
>20 g (13). The importance of low CF lesions was
delivery (6,18), anatomical wall thickness (18), and
highlighted by Neuzil et al. (8), who showed that
lesion density and interlesion distance (19) have all
chronic PV reconnection had a stronger association
emerged as important factors potentially influencing
with minimum CF and force-time integral within a PV
ablation efficacy with the use of CF-sensing catheters
segment rather than the average CF or force-time in-
(Figure 1). Moreover, there are other intraprocedural
tegral in that segment, suggesting that the durability of
markers of lesion completeness, such as loss of pace
PVI is contingent upon the worse lesion delivered in
capture (20) and electrogram-based assessment of
any PV segment. Further work showed that procedural
lesion transmurality (21), and novel cardiac magnetic
success was markedly enhanced when the RF lesions
resonance applications that provide additional infor-
were delivered in the operator pre-defined CF working
mation beyond CF that may also play an important role
range $80% of the time (9) or when $90% lesions were
in improving procedural efficacy.
delivered with an average CF of $10 g (10). More
Although limitations of the present study include
recently in a porcine model, Williams et al. (14) showed
analysis of only one-fourth of the original SMART-AF
that chronic RF lesion size was not significantly
study population, its retrospective nature, and the
different between low CF (<10 g) and high CF (>20 g)
small sample size, it lays the foundation for further
701
702
Kumar et al.
JACC: CLINICAL ELECTROPHYSIOLOGY VOL. 2, NO. 6, 2016 NOVEMBER 2016:700–2
Contact Force–Guided Pulmonary Vein Isolation
prospective, randomized studies that should examine
that further studies are needed to elucidate optimal
the role of catheter stability in addition to CF in
CF accommodating aforementioned factors, such as
enhancing procedural efficacy. It is possible that the
catheter stability. Until then, the quest for achieving
optimal CF ranges derived may have differed with
optimal procedural success in AF ablation remains
different stability criteria. Nevertheless, an emerging
very much alive.
paradigm is that facilitating homogenous contact is likely a key determinant of procedural success. The
REPRINT REQUESTS AND CORRESPONDENCE: Dr.
discrepancy between the finding of an optimal force
Saurabh Kumar, Cardiovascular Division, Brigham &
range of 6.5 to 10.3 g in the present study (12),
Women’s Hospital, 75 Francis Street, Boston, Massa-
compared with >20 g in prior studies (8,13), means
chusetts 02445. E-mail:
[email protected].
REFERENCES 1. Shah DC, Lambert H, Nakagawa H, Langenkamp A, Aeby N, Leo G. Area under the real-time contact force curve (force-time integral) predicts radiofrequency lesion size in an in vitro contractile model. J Cardiovasc Electrophysiol 2010;21:1038–43. 2. Yokoyama K, Nakagawa H, Shah DC, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 2008;1:354–62. 3. Ganesan AN, Shipp NJ, Brooks AG, et al. Longterm outcomes of catheter ablation of atrial fibrillation: a systematic review and meta-analysis. J Am Heart Assoc 2013;2:e004549. 4. Kuck KH, Reddy VY, Schmidt B, et al. A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm 2012; 9:18–23. 5. Kumar S, Haqqani HM, Chan M, et al. Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans. Heart Rhythm 2013; 10:962–9. 6. Kumar S, Morton JB, Lee J, et al. Prospective characterization of catheter-tissue contact force at different anatomic sites during antral pulmonary vein isolation. Circ Arrhythm Electrophysiol 2012;
9. Natale A, Reddy VY, Monir G, et al. Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol 2014;64: 647–56. 10. Reddy VY, Dukkipati SR, Neuzil P, et al. Randomized, controlled trial of the safety and effectiveness of a contact force-sensing irrigated catheter for ablation of paroxysmal atrial fibrillation: results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) Study. Circulation 2015;132:907–15. 11. Ullah W, McLean A, Tayebjee MH, et al. Randomized trial comparing pulmonary vein isolation using the SmartTouch catheter with or without real-time contact force data. Heart Rhythm 2016; 13:1761–7. 12. Reddy VY, Pollak S, Lindsay BD, et al. Relationship between catheter stability and 12-month success after pulmonary vein isolation: a subanalysis of the SMART-AF trial. J Am Coll Cardiol EP 2016;2:691–9. 13. Reddy VY, Shah D, Kautzner J, et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm 2012;9:1789–95.
7. Shurrab M, Di Biase L, Briceno DF, et al. Impact of contact force technology on atrial fibrillation ablation: a meta-analysis. J Am Heart Assoc 2015; 4:e002476.
14. Williams SE, Harrison J, Chubb H, et al. The effect of contact force in atrial radiofrequency ablation: electroanatomical, cardiovascular magnetic resonance, and histological assessment in a chronic porcine model. J Am Coll Cardiol EP 2015; 1:421–31.
8. Neuzil P, Reddy VY, Kautzner J, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol 2013;6:327–33.
15. Kautzner J, Neuzil P, Lambert H, et al. EFFICAS II: optimization of catheter contact force improves outcome of pulmonary vein isolation for paroxysmal atrial fibrillation. Europace 2015;17: 1229–35.
5:1124–9.
16. Okumura Y, Watanabe I, Iso K, et al. Clinical utility of automated ablation lesion tagging based on catheter stability information (VisiTag Module of the CARTO 3 System) with contact force-time integral during pulmonary vein isolation for atrial fibrillation. J Interv Card Electrophysiol 2016;47: 245–52. 17. Kumar S, Morton JB, Halloran K, et al. Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmias. Heart Rhythm 2012; 9:1041–7.e1. 18. Chikata A, Kato T, Sakagami S, et al. Optimal force-time integral for pulmonary vein isolation according to anatomical wall thickness under the ablation line. J Am Heart Assoc 2016;5: e003155. 19. Park CI, Lehrmann H, Keyl C, et al. Mechanisms of pulmonary vein reconnection after radiofrequency ablation of atrial fibrillation: the deterministic role of contact force and interlesion distance. J Cardiovasc Electrophysiol 2014;25:701–8. 20. Steven D, Sultan A, Reddy V, et al. Benefit of pulmonary vein isolation guided by loss of pace capture on the ablation line: results from a prospective 2-center randomized trial. J Am Coll Cardiol 2013;62:44–50. 21. Bortone A, Brault-Noble G, Appetiti A, Marijon E. Elimination of the negative component of the unipolar atrial electrogram as an in vivo marker of transmural lesion creation: acute study in canines. Circ Arrhythm Electrophysiol 2015;8: 905–11.
KEY WORDS catheter stability parameters, contact force, paroxysmal atrial fibrillation, pulmonary vein isolation, radiofrequency catheter ablation