Defect Control in Semiconductors K. Sumino (ed.) © Elsevier Science Publishers B.V. (North-Holland), 1990
CONTROL OF SEMICONDUCTOR SURFACE, REFLECTION X-RAY FLUORESCENCE (TXRF)
1547
TRACE
METAL
CONTAMINATION
BY
TOTAL
R. S. HOCKETT Charles Evans & Associates, 301 Chesapeake Drive, Redwood City, CA USA
94063
The control of ultra-surface metallic impurities is necessary to control defects in semiconductor devices. One analytical method which can provide some spatial information of the contamination, distinguish between particulate versus plated contamination, and thereby allow the control of these impurities is Total reflection X-Ray Fluorescence (TXRF). Examples of real commercial processes are presented and include: cleaning, rinsing, dry etching, rapid thermal processing, and ion implantation.
1. INTRODUCTION The
rays from the top 3-4nm of surface atoms are
control
metallic
of
impurities
ultra-surface
is
necessary
to
(<3nm) control
then detected using a conventional Li-drifted silicon detector which functions as an energy
defects in semiconductor devices, since these
dispersive
impurities can both cause structural defects and
normally operated at 30 kV and 60 mA in air
make
them
active1.
electrically
These
impurities can be introduced at many different process
steps,
etching,
for
rapid
example:
thermal
cleaning,
ion
The
system
is
using full diameter wafers up to 150 mm.
The
analysis area is 8 mm in diameter.
dry
processing,
spectrometer3.
Quantitation is achieved using a silicon substrate
which
has
been
intentionally
implantation and ambient transfer from materials
contaminated with Ni of a known areal density
of construction.
using a diluted wet chemistry standard3.
Often these
impurities are
introduced heterogeneously
which may
particulate
or
contamination
a
indicate
heterogeneous
the
calibration
for
One analytical
using sensitivity
method
some
once
can
information of
provide
the heterogeniety,
spatial
distinguish
is
Once
completed,
the
calibration for the other metals is completed
source of plated contamination. which
Ni
when
the
Correlation
factors which were measured instrument
of
was
TXRF
manufactured.
with
Rutherford
between particulate versus plated contamination,
Backscattering
and
these
Phase
X-Ray
Absorption Spectroscopy (VPD/AAS) has validated
thereby
impurities
and
allow
is
the
Total
control reflection
of
2
Fluorescence (TXRF) .
the
In TXRF using a Perkin-Elmer XSA-8000 X-Ray
Surface
Analyzer,
x-rays
from
Spectrometry
Decomposition quantitation
(RBS)C
followed
and by
approach4.
Vapor Atomic
Analytical
statistical process control is completed using a
periodic
TXRF
measurements
on
two
samples.
conventional molybdenum x-ray anode impinge the
Precisions
surface of the silicon wafer at a glancing angle
depend upon the specific element and its amount,
(1.3 milliradians), below the critical angle for
but practical examples include <15% for zinc at
'total
external
reducing
the
reflection,
background
thereby
silicon
greatly
fluorescence
0.4xl012
(one
relative
atoms/cm2
and
standard
for
iron
deviation)
at
atoms/cm . Interference-free detection limits '
signal commonly experienced in conventional x-
are on the order of 1011 atoms/cm2.
ray fluorescence.
technique
The surface fluorescence x-
0.9xl012
using
a Mo
anode
is
The TXRF
sensitive
to
1548 surface trace metals with atomic numbers in the
four world-scale silicon manufacturers was made
ranges of 20-35 and 56-83.
using 2
It has been shown both experimentally and theoretically3
that
the
change
in
the
TXRF
TXRF6.
Japanese.
Two
of
the
manufacturers
are
One cassette per vendor was used for
twenty-five wafers per vendor.
The data are
signal as a function of the incident x-ray angle
summarized
results
is different for plated contamination (e.g., one
statistically valid differences between wafers
thousandths
taken from the four cassettes for Cu, Fe, Zn, Ca
of
a monolayer
of
Cu)
than
for
in
Table
1.
The
show
residue contamination (e.g., Ni atoms dispersed
and Br.
in a a 50 nm thick organic residue.)
has wafers with surface Cu out of statistical
Figure 1
presents experimental data which are consistent
Furthermore, one of the cassettes (B)
control (Cu varying over two orders of magnitude within the cassette of wafers) while the other
CO
surface impurities on these same wafers are in
to
statistical
C
control.
importance
of
This
using
indicates
a
the
multi-element
characterization technique for defect control.
H
TABLE 1. AVE RESULTS FOR FOUR SILICON VENDORS (units of 10 12 atoms/cm2)
0.5
1.7 X-ray incident angle (mrad)
FIGURE 1 Normalized TXRF signal versus incident x-ray angle for Ni dispersed in a thin film residue and for Cu plated in a monolayer.
Fe
Cu
Zn
Ca
Br
A
0.5
0.5
2.5
<2
0.3
B
0.6
60
0.5
<2
<0.2
C
0.4
0.8
1.5
3.5
0.2
D <0.3
0.5
0.5
<2
2
Because of the high Cu contamination found
with the theoretical curves for these two cases. For residue contamination the TXRF signal is approximately constant with increasing angle up to about 1.2-1.6 mrad where it begins to drop. For
plated
contamination
the
TXRF
signal
increases monotonically with increasing angle up to about 1.4-1.6 mrad where it also begins to drop.
This difference in signal versus angle
on wafers in cassette B and because of the high range
of
Cu
from
wafer-to-wafer
order to resolve
this
issue, the TXRF angle
several wafers.
particulate
contamination, so that in the same series of measurements
the
contamination
related
to
the
for
The Figure 1 curve for Cu are
the results for one of the wafers.
plated
In
variation of the Cu signal was measured
reveals
and
the
came from plating or from particulates.
can be used to qualitatively distinguish between contamination
within
cassette, there was some question whether the Cu
Cu
is plated
rather
The graph than
in a
particulate or residue form. Surface
maps
by
TXRF
revealed
particulates and to plating can be in principle
heterogeneous Ca and Fe on a 100 mm diameter
distinguished.
silicon
wafer7.
The
maps
reveal
distinct
patterns: a symmetric, circular ring of Ca and 2. APPLICATIONS
Fe; and a non-symmetric Fe spot.
These patterns
2.1 Cleaning
can help the cleaning process engineer identify
A comparison of the surface impurities on
and control defects.
100 mm diameter prime silicon wafers taken from
the
sources
of
these
cleaning
1549 The
plating
chemicals
used
efficiency in
two
of
metals
different
from
cleaning 2,5
chemistries has also been studied by TXRF
.
The data reveal that Cu is especially plated to silicon
in buffered
especially
plated
HF
chemistry
to
silicon
and in
Fe
of recleaning
ammonium
The data reveal that all three dry processes
leave
heavy
metal
contamination, but only two of the three can be
&
silicon
wafers from three different silicon vendors was also studied by TXRF8.
cleaning. etching
is
hydroxide: hydrogen peroxide chemistry. The effectiveness
oxides after dry etching photoresist and after
Many of the surface
H Ό
impurities were homogenized among the vendors by
e
the RCAII reclean, but Cu was not.
u
A TXRF multi-point spectra of an oxide
o 0.5
wafer which had received a rinse in a semitool revealed
a heterogeneous
contamination.
The
TXRF angle variation shown in Figure 2 in this case reveals the Fe and Zn contamination was particulate rather than a heterogeneous plating.
1.7 X-ray incident angle (mrad)
FIGURE 3 Normalized TXRF signal versus incident x-ray angle for Fe plated from an RCAI cleaning of a silicon wafer. cleaned
up
by
a
cleaning
process
after
dry
etching. The TXRF angle variation for a different dry etched and cleaned sample but of the same type
is
shown
in Figure
4.
The
mono tonic
increase in signal up to 1.7 mrad reveals the heavy
metal
contamination
particulate, nor plated, but oxide surface.
O
0.5
to
be
imbedded
neither in the
This finding is consistent with
the model proposed by others9 of a dry etcher
1.7
driving metals from the resist into the oxide
X-ray incident angle (mrad) FIGURE 2 Normalized TXRF signal versus incident x-ray angle for Fe and Zn in a particle on the surface of an oxide after semitool rinsing.
surface where the metals are difficult to remove by cleaning. 1
S>
In contrast, the TXRF angle variation shown in Figure 3 of a bare wafer cleaned by an RCAI type chemistry, which has been shown to plate out Fe5,
reveals
plated
Fe
contamination,
0)
as
expected. o
2.2 Dry Etching
55
The problem of heavy metal contamination from dry etching of photoresists on oxides has been identified as important using qualitative surface
9
SIMS .
Table
2 gives TXRF
data
on
0.5
1.7 X-ray incident angle (mrad)
FIGURE 4 Normalized TXRF signal versus incident x-ray angle for Fe and Ni imbedded in the surface of an oxide after dry etching photoresist and cleaning.
1550 TABLE 2.
TABLE 3. RESULTS FOR ION IMPLANTATION AND RTA
RESULTS FOR DRY ETCHING
(units of 1012 atoms/cm2)
(units of 1012 atoms/cm2; a blank means not detected) SAMPLE
Cr
Fe
Cu
Zn
Oxide Control pos 0 pos 1 pos 2
SAMPLE
Fe
Zn
Cr
Control
0.3
10
<0.4 <0.4
RTA
2
0.3
75
3
0.2
0.4
65
0.9
<0.2
<0.4
60
As I m p l a n t
As Implant + RTA
0.25
As
Cleaning Control pos 0
does the ion implant introduce surface Fe, but
0.3
pos 1
the rapid
pos 2
The combination of implant plus anneal results 1 1
The TXRF angle variation for a different
0.5
75
As-implanted sample with 0.75xl012
0.2
by similar
0.45
0.2
Fe atoms/cm2 reveals the Fe is distributed on
0.2
particulate or residue in nature.
pos 1 pos 2
(RTA) does also.
in the Fe diffusing to the implant defects.
AFTER DRY ETCHER A pos 0
thermal annealing
the top surface of the materials and is not
AFTER DRY ETCHER B pos 0
2
0.6
pos 1
2
0.4
Cross contamination of As at the 1012/cm2
pos 2
2
0.3
level has also been found using TXRF on the surface of material implanted with P and B at
AFTER DRY ETCHER C pos 0
6
1.5
0.9
0.55
pos 1
4
1
0.5
0.3
pos 2
7.5
1.5
1
0.7
AFTER DRY ETCHER A PLUS CLEAN
1015/cm2 doses10. 2.4 Rapid Thermal Processing Rapid thermal processing (RTP) equipment has been found to have cross contamination even
pos 0
0.9
0.55
0.2
when
pos 1
0.6
0.85
0.2
Tungsten
pos 2
0.7
0.3
equipment (from
is
quartz1 .
HF-cleaned
tungsten
suicide
annealing)
transferred onto the surface of silicon wafers
AFTER DRY ETCHER B PLUS CLEAN
from the contaminated quartz using oxygen in the ambient and with an activation energy of about
pos 0 pos 1
the
0.8 eV.
0.3
The TXRF angle variation for a sample
processed at 1000°C for 120 seconds in air is
pos 2 AFTER DRY ETCHER C PLUS CLEAN
shown
in
Figure
5.
The
data
reveal
the
contamination is not particulate, but is not
pos 0 pos 1
0.3
strictly plated either.
The increase in signal
pos 2
0.35
at
suggests
the
higher
angles
some
of
the
tungsten is incorporated in the surface oxide 2.3 Ion Implantation and Annealing Ion implantation can introduce heavy metal contamination
from materials
of construction.
Table 3 gives TXRF data on ion implanted silicon and annealing.
The data reveal that not only
grown during the RTP. 2.5 SIMOX The high implant dose of oxygen for SIMOX substrates contamination
can as
introduce shown by
TXRF
heavy
metal
in Table
4.
These metals normally leave the ultra-surface
1551 reduces the
S,
thus
intensity of the primary x-rays,
reducing
signals.
the
However,
Ga
and
this
As
also
fluorescence results
in
a
reduction of fluorescent signals from impurities a)
on the surface of the GaAs substrate.
The final
result is a loss in detection limits.
The TXRF
detection
a u o
limits
for
GaAs
substrates
for
equivalent integration times are about one order 0.5
of magnitude worse than for silicon substrates.
1.7 X-ray incident angle (mrad)
However,
these
detection
limits
are
still
capable of detecting Fe and Cu on the surface of FIGURE 5 Normalized TXRF signal versus incident x-ray angle for W cross contamination incorporated into a thin oxide grown during rapid thermal processing (1000°C-120sec-air).
commercially available GaAs substrates. SUMMARY In summary the TXRF with its angle variation capability has been shown to be very useful in
region
after
diffuse
to
insulator
annealing the
and
defects
layer.
are
in
The
the
TXRF
expected
to
silicon
on
angle
variation
reveals the Ti and Fe are particulate on the
detecting
metallic
qualitatively
contamination
distinguishing
and
in
between
plated,
particulate, or imbedded contamination.
Thus it
is expected to be used for defect control in
surface while the Cr, Mn and Ta are more like
semiconductor processing.
imbedded contamination.
has shown the breadth of application for this
be
imbedded,
so
The Fe also appears to
that
there
may
be
both
particulate Fe and imbedded Fe.
(Units of 10
12
1. Fumio Shimura, Semiconductor Silicon Crystal Technology, published by Academic Press (New York) 1989.
2
atoms/cm ) Fe
la
0.6
10
0.7
6
17
4
4
15
4
(mrd)
Ti
Cr
Mn
0.5
4
2
1.3
7
9
1.8
<2
11
2.6 Gallium Arsenide arsenide
application substrates
fluorescence arsenic.
of
TXRF
results
signals
from
the
to in
gallium intense
gallium
and
These intense matrix signals result in
extremely high deadtimes for the detector if the XSA-8000
instrument
is
capability. REFERENCES
Angle
The
A variety of examples
operated
similar to that for silicon.
in a manner
For this reason a
200 micron thick Mo filter is inserted between the x-ray source and the GaAs substrate.
This
2. Peter Eichinger, Heinz J. Rath and Heinrich Schwenke, "Application of Total Reflection X-Ray Fluorescence Analysis for Metallic Trace Impurities on Silicon Wafer Surfaces," Semiconductor Fabrication: Technology and Metrology. ASTM STP 990. edited by Dinesh C. Gupta, American Society for Testing and Materials, 1989. 3. W. Berne ike, J. Knoth, H. Schwenke, and U. Weisbrod, Fresenius Z. Anal. Chem. 333. 524 (1989). 4. R. S. Hockett, N. Fujino, and P. Eichinger, to be presented at the Fall Meeting of the Japan Applied Physics Society, Fukuoka, Japan, September 1989; at the Fall Meeting of the ECS, Hollywood, Florida, October 1989; and at the Meeting of the European Committee of the ECS, Berlin, September 1989. 5. V. Penka and W. Hub, Fresenius Z. Anal. Chem. 333, 586 (1989).
1552 6. R. S. Hockett and W. Katz, accepted publication in the J. ECS, (1989).
for
7. P. Eichinger, "Metallic Trace Contamination in IC Technology: Impact, Origin, and Analytics," STEP Europe Conference, October 2728, 1989, Brussels. 8. V. Penka and W. Hub, Spectrochimica Acta 44B. 483 (1989). 9. Shuzo Fujimura and Hiroshi Yano, J. ECS 135. 1195 (1988). 10. U. Biebl, "Dry Processing Equipment as a Source of Wafer Contamination," STEP Europe Conference, October 27-28, 1989, Brussels 11. R. S. Hockett, "The Correlation between Quantitative Surface Metallic Contamination and RTP-Induced Surface Defects," accepted for publication by the Materials Research Society in Proceedings of the Symposium on Rapid Thermal Annealing/Chemical Vapor Deposition, San Diego, April 24-29, 1989.