Endogenous glucocorticoid analysis by liquid chromatography–tandem mass spectrometry in routine clinical laboratories

Endogenous glucocorticoid analysis by liquid chromatography–tandem mass spectrometry in routine clinical laboratories

Accepted Manuscript Title: Endogenous glucocorticoid analysis by liquid chromatography-tandem mass spectrometry in routine clinical laboratories Autho...

1MB Sizes 0 Downloads 41 Views

Accepted Manuscript Title: Endogenous glucocorticoid analysis by liquid chromatography-tandem mass spectrometry in routine clinical laboratories Author: James M. Hawley Brian G. Keevil PII: DOI: Reference:

S0960-0760(16)30144-3 http://dx.doi.org/doi:10.1016/j.jsbmb.2016.05.014 SBMB 4729

To appear in:

Journal of Steroid Biochemistry & Molecular Biology

Received date: Revised date: Accepted date:

2-9-2015 11-5-2016 12-5-2016

Please cite this article as: James M.Hawley, Brian G.Keevil, Endogenous glucocorticoid analysis by liquid chromatography-tandem mass spectrometry in routine clinical laboratories, Journal of Steroid Biochemistry and Molecular Biology http://dx.doi.org/10.1016/j.jsbmb.2016.05.014 This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Endogenous glucocorticoid analysis by liquid chromatography-tandem mass spectrometry in routine clinical laboratories.

James M Hawley 1 and Brian G Keevil 1,2

1. University Hospital South Manchester, Manchester, UK. 2. Manchester Healthcare Academy, Manchester, UK

Corresponding author: James M Hawley Address: Clinical Biochemistry Department, Clinical Sciences Building, University Hospital South Manchester, Southmoor Road, Manchester, M23 9LT. Telephone: 0044 161 291 2136 Fax: 0044 161 291 2927 Email: [email protected]

Page 1 of 66

Highlights



A review of the clinical applications of LC-MS/MS glucocorticoid analysis.



The major endogenous glucocorticoids are discussed.



Serum, urine and salivary cortisol are comprehensively covered.

Page 2 of 66

Non-standard abbreviations

3-HSD = 3-hydroxysteroid dehydrogenase 5THF = Allo-tetrahydrocortisol 5THF = Tetrahydrocortisol 11HSD1 = 11 beta hydroxysteroid dehydrogenase type I 11HSD2 = 11 beta hydroxysteroid dehydrogenase type II 11-DOC = 11-Deoxycortisol ACTH = Adrenocorticotropin hormone AME = Apparent mineralocorticoid excess APCI = Atmospheric pressure chemical ionisation APPI – Atmospheric pressure photo ionisation CBG = Cortisol binding globulin CRH = Corticotropin releasing hormone CRM = Certified Reference Material EQA = External quality assessment ESI = Electrospray Ionisation GC-MS = Gas chromatography-mass spectrometry GC-MS/MS = Gas chromatography-tandem mass spectrometry HPA = Hypothalamic – pituitary – adrenal HPLC = High performance liquid chromatography LC-MS/MS = Liquid chromatography-tandem mass spectrometry LC = Liquid chromatography ONDST = Overnight dexamethasone suppression test

Page 3 of 66

LLE = Liquid-liquid extraction m/z = mass to charge ratio NEQAS = National External Quality Assurance Scheme PFP = Pentafluorophenyl PPT = Protein precipitation RMP = Reference measurement procedure SLE = Supported liquid extraction SPE = Solid phase extraction THE = Tetrahydrocortisone UFC = Urine free cortisol UPLC = Ultra high performance liquid chromatography

Page 4 of 66

Abstract Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is a powerful analytical technique that offers exceptional selectivity and sensitivity. Used optimally, LC-MS/MS provides accurate and precise results for a wide range of analytes at concentrations that are difficult to quantitate with other methodologies. Its implementation into routine clinical biochemistry laboratories has revolutionised our ability to analyse small molecules such as glucocorticoids. Whereas immunoassays can suffer from matrix effects and cross-reactivity due to interactions with structural analogues, the selectivity offered by LC-MS/MS has largely overcome these limitations. As many clinical guidelines are now beginning to acknowledge the importance of the methodology used to provide results, the advantages associated with LCMS/MS are gaining wider recognition. With their integral role in both the diagnosis and management of hypo- and hyperadrenal disorders, coupled with their widespread pharmacological use, the accurate measurement of glucocorticoids is fundamental to effective patient care. Here, we provide an up-to-date review of the LC-MS/MS techniques used to successfully measure endogenous glucocorticoids, particular reference is made to serum, urine and salivary cortisol.

Page 5 of 66

Introduction Glucocorticoids are steroid hormones that exert their physiological effects through an interaction with complementary glucocorticoid receptors. These receptors are expressed within almost all human cells. Their stimulation initiates a signalling cascade that terminates in the activation of genes that help regulate normal development and control day-to-day homeostasis. The synthesis, metabolism and the pleiotropic actions of glucocorticoids have been extensively reviewed [1, 2, 3]. Similarly, the clinical indications for their measurement have been outlined in international guidelines [4, 5] and recommendations [6].

Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is a powerful technique that has transcended research and analytical chemistry laboratories into routine clinical diagnostics [7]. The exquisite sensitivity and specificity LCMS/MS offers is however reliant on the optimisation of several processes.

This review aims to provide the reader with an up-to-date account of LCMS/MS endogenous glucocorticoid measurement. The optimisation of all relevant LC-MS/MS processes including: sample preparation, mobile phase composition, chromatographic separation and MS/MS detection are considered in this review, with particular reference made to serum, urine and salivary cortisol quantification.

Page 6 of 66

Endogenous glucocorticoids In humans, the major circulating glucocorticoid is cortisol. This can be measured in many biological specimens including: whole blood, plasma, serum, urine, saliva and hair. This variety of sample types serves many different purposes from clinical diagnosis, research studies and athletic drug screening, with each matrix offering its own analytical advantages and challenges. However, in routine clinical biochemistry, only serum, urine and saliva have found prominence with their use is endorsed for the assessment of the hypothalamic-pituitary-adrenal (HPA) axis [4,5,6].

Serum/plasma cortisol Cortisol is one of the best characterised analytes routinely measured in clinical laboratories. Since 1973, the National Institute for Standards and Technology have made available cortisol standard reference material to promote assay standardisation. Subsequently, several reference measurement procedures (RMPs) have been described that utilise this reference material for calibration. These methods have been used to produce higher-order certified reference materials (CRMs) that are metrologically traceable to SI units [8].

The majority of these RMPs have been validated using isotope dilution (ID) gas chromatography-mass spectrometry (GC-MS). However, advances in sample preparation, liquid chromatography and mass spectrometry have

Page 7 of 66

heralded a second generation of LC-MS/MS reference systems. The first of these was described in 2004 and utilises sequential off-line solid phase extraction (SPE) and liquid-liquid extraction (LLE) for sample preparation prior to LC-MS/MS analysis [9]. More recently, an on-line SPE method has been described [10] and our own group have developed a candidate RMP using supported liquid extraction (SLE) technology for sample clean-up [11].

Although RMPs offer accurate and precise results, they do not represent a timely and cost-effective option for routine analysis. Instead, they are used to assign materials with a target value so they can serve as certified reference materials (CRMs) to calibrate commercial immunoassay kits [12]. However, despite the availability of serum cortisol RMPs and CRMs, there remains a problem with the standardisation of routine commercial immunoassays. External quality assessment (EQA) schemes provide laboratories with a means to review the analytical performance of their method relative to that of other users. Whereas the intra-method variation for a given sample may be relatively small, it can be significantly increased when all methods are accounted for. The report below provides an example of this variation (Figure 1).

[Insert Figure 1 here.]

This longstanding poor performance has been attributed to matrix effects and poor specificity [13, 14]. Consequently, these assays present a potential risk to accurate patient diagnosis and management. This has provided a

Page 8 of 66

requirement for robust and accurate cortisol methods in routine clinical laboratories. These requirements have been fulfilled by both LC-MS/MS and gas chromatography coupled to either a single quadrupole mass spectrometer (GC-MS) or a triple quadrupole mass spectrometer (GC-MS/MS). However, whereas serum/plasma cortisol analysis by GC-MS and GC-MS/MS has predominantly been limited to RMPs, LC-MS/MS has been favoured for routine clinical analysis. This may be explained by the general requirement to derivatise non-volatile compounds such as cortisol to make them amenable to GC analysis [8]. Despite improving sensitivity, derivatisation can extend sample preparation time and increase total analytical error. Moreover, it has been suggested that the conditions used for derivatisation can hydrolyse conjugates and metabolites to produce the analyte of interest, thereby reducing analytical specificity [15]. Nevertheless, GC-MS and GC-MS/MS still hold an advantage over LC-MS/MS owing to their higher chromatographic resolution, they therefore remain powerful techniques for specialised applications such as urine steroid profiling [16]. Conversely, LC-MS/MS assays do not generally require derivatisation, are typically higher-throughput and now have comparable sensitivity and specificity to GC-MS or GC-MS/MS. Consequently, their application has generally been favoured for routine clinical diagnostics.

Several LC-MS/MS methods have been used to successfully quantify serum cortisol (Table 1). Of those listed, sample volume requirements range from 25 to 1000 µL, with larger volumes generally necessary to improve assay

Page 9 of 66

sensitivity. Accordingly, many of these assays are designed to quantify a panel of endogenous and exogenous glucocorticoids, many of which are present in significantly lower concentrations than that of cortisol. However, in routine clinical practice, it is notable that serum and plasma are not always of sufficient volume for the multitude of tests that are typically required. This can be especially true of paediatric samples. Hence, for serum/plasma cortisol analysis, small sample volumes are generally preferable.

Sample pre-treatment and extraction is variable across methods (Table 1). The choice of sample preparation may be influenced by a number of factors including: the relative abundance and physiochemical properties of the analyte of interest, sample matrix, potential interferences, the presence of ion suppression or enhancement, the sensitivity required, the time in which the result needs to be issued, operator proficiency and financial considerations. For serum cortisol, successful sample clean-up has been achieved in a number of ways including: protein precipitation (PPT) [17, 18, 19], LLE [20, 21, 22, 23, 24], SPE [25, 26, 27] and also using a combination of these procedures [28, 29, 30].

Collectively, these approaches aim to remove protein and other potential interferences. Deproteinisation reduces the risk of blocking frits, injectors, columns or narrow bore HPLC tubing and helps reduce matrix effects which can have detrimental effects on analysis [31, 32]. In addition to protein, salts and phospholipids can co-elute with the analyte of interest thereby altering the ionisation efficiency. This can lead to in-source competition for ionisation

Page 10 of 66

resulting in ion suppression or enhancement [31]. This in turn may compromise the accuracy of the final result. Sample deproteinisation is particularly relevant to serum/plasma cortisol analysis. At present, all current clinical guidelines advocate total cortisol measurement [4,5]. Since cortisol circulates as both free (5%) and protein bound (95%) fractions, it is only through liberating cortisol from its binding proteins that allows its total concentration to be reliably quantified.

PPT methods rely on the addition of a precipitant to liberate cortisol from its binding proteins. Several options are available with metal ions [17] and organic solvents [18,19] popular choices. The relative merits and limitations of different forms of PPT have been extensively tested [33]. PPT methods are rapid, amenable to automation, and relatively inexpensive. However, as a consequence of the required sample dilution, assay sensitivity is reduced. Nevertheless, for cortisol alone, this does not appear to jeopardise the utility of the assay since LOQs below current clinical requirements are still achievable, even with sample volumes as small as 25 µL [17].

Conversely, as both LLE and SPE remove more matrix effects and concentrate the sample, both sensitivity and specificity are improved. These extraction techniques have therefore been favoured for multiplexed assays where additional sensitivity and specificity may be required [20, 21, 22, 23, 25, 26, 27, 28, 29, 30]. However, relative to protein precipitation, LLE and in particular SPE, are technically more demanding, require a longer preparation

Page 11 of 66

time and are more expensive. For these reasons they are not always practical for routine clinical work.

Page 12 of 66

Table 1

Analyte(s)

Sample volume (L)

Preparation

Column

Mobile phase (organic/buffer/modifier)

Run time (mins)

Ionisation

LOQ (nmol/L)

Reference

Cortisol

25

PPT (Zinc sulphate / Methanol)

C8

Methanol / 2 mmol/L ammonium acetate / 0.1% formic acid

2.6

ESI +

12.5

[17]

Cortisol

250

LLE (Ethyle acetate)

C18

Methanol / 1 mmol/L ammonium chloride / NA

8

ESI -

1.38

[24]

Cortisol, Cortisone

100

LLE (MTBE)

Phenyl

Methanol / NA / NA

3

APPI +

13.76 (Cortisol) 13.87 (Cortisone)

[20]

Cortisol, Cortisone

100

LLE (Ethyle acetate)

C18

Methanol / 5 mmol/L ammonium acetate / NA

Not Stated

ESI +

2.75 (Cortisol), 6.94 (Cortisone)

[23]

Cortisol, Dexamethasone

1000

PPT (Acetonitrile) / SPE (C18)

C18

Methanol / NA / 0.2% formic acid

Not Stated

ESI +

< 20.0 (Cortisol)

[29]

Cortisol, Cortisone, Dexamethasone

200

SPE (Strata X)

Aromax

Acetonitrile / NA / 0.05% formic acid

Not stated

APCI +

0.8 (Cortisol) 0.8 (Cortisone)

[25]

Cortisol, Dexamethasone, Prednisone, Prednisolone

500

SPE (HLB)

C18

Methanol / 5 mmol/L ammonium acetate / NA

6

ESI -

14.85 (Cortisol)

[26]

Cortisol, Cortisone, Prednisolone, Prednisone

50

PPT (Acetonitrile), evaporation, reconstitution

Phenyl

Acetonitrile / NA / 0.1% formic acid

8

ESI +

2.76 (Cortisol), 1.4 (Cortisone)

[18]

Cortisol, Cortisone, 17-OHP, 21-DOC, 11-DOC, Deoxycorticosterone

500

LLE (Methylene chloride)

C18

Methanol / NA / NA

10

ESI +

2.75 (Cortisol),* 13.87 (Cortisone)* 14.4 (11-DOC) *

[21]

Cortisol, Cortisone, Prednisolone, Dexamethasone, 11-DOC,

250

SPE (HLB)

C18

Methanol / 2 mmol/L ammonium acetate / 0.1% formic acid

3

ESI +

3.75 (Cortisol), 3.75 (Cortisone), 5.00 (11-DOC)

[27]

Cortisol, Cortisone, Dexamethasone, Prednisolone, Prednisone, Methylprednisolone

500

PPT (Acetonitrile) / LLE (Dichloromethane)

C18

Methanol / 2mmol/L ammonium acetate / 0.1% formic acid

12

ESI +

8.3 (Cortisol), 8.3 (Cortisone)

[28]

Cortisol, Cortisone, Testosterone, Dihydrotestosterone, Progesterone

100

PPT (Methanol)

C18

Methanol / NA / 0.1% formic acid

12

ESI +

5.24 (Cortisol), 0.83 (Cortisone)

[19]

85

LLE (Ethyl acetate:hexane, 80:20 (v/v))

C18

Acetonitrile / NA / 0.1% formic acid

6.1

ESI -

1.95 (Cortisol), 1.58 (Cortisone), 0.098 (11-DOC)

[22]

500

PPT (HCl) / SPE (HLB)

C18

Methanol / 5 mmol/L ammonium acetate / NA

Not stated

ESI -

9.93 (Cortisol)

[30]

Cortisol, Cortisone, 11-DOC, Dexamethasone, Prednisolone, Prednisone, Testosterone, Androstendione, Progesterone Cortisol, Dexamethasone, Methylprednisolone, Prednisone, Prednisolone, Mycophenolic acid, Mycophenolic acid glucuronide

Page 13 of 66

There is marked heterogeneity in mobile phase composition across the methods listed in Table 1. Whereas water is typically used for the aqueous phase, the choice of organic solvent varies between methanol and acetonitrile. As a protic solvent, methanol can promote [M+H]+ ion formation. However, when mixed with water, its viscosity can increase resulting in a high column backpressure. This is generally undesirable and can result in column deterioration or system leaks. Since the advent of UPLC, there has been a trend towards using sub 2 micron particle columns as they can enhance peak resolution. Our experience has found these columns to be especially useful for steroid analysis as they can help separate isobaric interferences [34]. However, columns with < 2 m particles can be susceptible to rapid degradation when subjected to very high backpressures. This can result in deterioration of chromatographic separation and often necessitates that low flow rates are used which consequently increases injection-to-injection time. Acetonitrile offers a useful alternative to methanol as it has a lower viscosity when mixed with water and therefore produces lower backpressures. However, as an aprotic solvent, acetonitrile does not enhance [M+H]+ ion formation, is generally more expensive and, at times, of limited availability.

While several methods do not make use of a buffer [18, 19, 20, 21, 22, 25, 29], others have used ammonium acetate [17, 28, 26, 27, 30, 23] or ammonium chloride [24]. The presence of a buffer can improve the ionisation of cortisol, this may be through adduct formation (e.g. ammonium chloride) or through enhancing in-source ionisation (e.g. ammonium acetate). As a neutral molecule, control of pH has minimal effect on cortisol analysis however,

Page 14 of 66

formic acid has been used to enhance ionisation both in the positive and negative ionisation modes [17, 18, 19, 22, 25, 28, 29, 27]. In the positive ionisation mode, formic acid can provide a rich supply of protons which favour the formation of [M+H]+ ions. Conversely, in the negative ionisation mode, cortisol can form a specific adduct with formic acid as discussed below. In our experience, it is good practice to experiment with different mobile phase compositions to optimise ionisation. In all cases, we have found that LC-MS grade solvents, buffers and modifiers produce the best results. These are typically purer than their HPLC- or reagent-grade equivalents.

By virtue of its hydrophobicity (Log P = 1.43 [35]) and steric properties, cortisol can be chromatographically separated from structurally analogous compounds by traditional alkyl bonded reverse phases. Several C18 columns are now available with differences in end-capping, carbon load and ligand surface area coverage just some of the factors that contribute to the varying performance between columns and manufacturers. Other than C18 columns, a C8 phase has successfully been used [17]. C8 columns rely on similar hydrophobic interactions however, owing to their shorter carbon chain length, analytes are less avidly retained thus permitting shorter run-times. The use of phenyl columns have also been described [18, 20, 25]. This mode of separation relies on pi-pi interactions between aromatic compounds and the stationary phase. Phenyl ligands are therefore capable of exploiting the aromatic ring system present in cortisol to chromatographically separate it from structurally similar compounds. This has reportedly helped separate cortisol from the prednisolone metabolite tetrahydroprednisolone, a potential

Page 15 of 66

isomeric interference [20]. However, in two of the assays developed, it is unlikely that full advantage of the phenyl phase is taking place [18, 25]. In these methods, acetonitrile has been used in the mobile phase. Acetonitrile is known to delocalise the electron clouds that help form pi-pi interactions, therefore suggesting the separations observed involve other retention mechanisms.

Although all assays listed achieve an LOQ below any current clinical requirements, the method described by Ray et al is notably lower than the others (0.80 nmol/L) [25]. This method uses SPE for sample purification and atmospheric pressure chemical ionisation (APCI) to convert cortisol into an ion prior to MS/MS detection. APCI relies on the vaporisation of the column eluent by a heated nebuliser, the remaining molecules (including the analyte of interest) are subsequently ionised by a corona discharge needle. This produces stable ions which then enter the mass spectrometer. APCI is particularly efficient at enhancing the ionisation of compounds that ionise less effectively in solution such as cortisol [36]. This efficiency has contributed to APCI generally being considered a harder form of ionisation than alternatives. Indeed, it has been suggested that APCI can cause an exchange between the deuterated atoms of labelled internal standards with the hydrogen atoms of water therefore compromising accurate quantification [37]. This observation has also been refuted, suggesting instrument-specific differences may effect ionisation [36]. In a novel approach, one cortisol method has used atmospheric pressure photoionisation (APPI) [20]. APPI uses ultraviolet light to irradiate the column eluent which initiates a cascade of reactions resulting

Page 16 of 66

in the formation of stable ions which are channelled into the mass spectrometer. Other than APCI and APPI, the majority of methods have used ESI to ionise cortisol. In ESI, the column eluent enters a stainless steel capillary to which a voltage is applied. A heated gas, usually nitrogen, is passed along the capillary resulting in the nebulisation of the eluent and production of charged droplets. These droplets undergo coulombic fission and are converted into stable ions through desorption and desolvation before entering the mass spectrometer. Relative to APCI and APPI, ESI in the positive ionisation mode does not remove as many matrix effects [37, 38]. However, if careful attention is paid to sample preparation and chromatography, this is not problematic for serum/plasma cortisol analysis since all ESI methods listed in Table 1 achieve sufficient performance characteristics to be clinically useful.

Despite the variation in preparation, separation and ionisation, the transitions used to detect cortisol are similar across methods. To complement this section of the review please refer to Figure 2. This shows a precursor ion scan in the positive ionisation mode of a 1 mg/L methanolic solution of cortisol infused into a Waters Acquity TQD™ LC-MS/MS system under the conditions described by Owen et al [17] (Fig. 2A). Here, the m/z 363 peak corresponding to the [M+H]+ cortisol ion is clearly visible, it is this ion that is scanned for in the first quadrupole of all methods using the positive ionisation mode. In figure 2B we present the product ion scan acquired from applying a collision energy of 28 eV to the precursor ion in the collision cell. In addition,

Page 17 of 66

we have included the corresponding theoretical fragments for the common product ions (Fig. 2B).

The majority of assays listed in Table 1 have favoured the quantification of cortisol using the m/z 363>121 transition [20, 23, 25, 27, 28, 29]. This corresponds to detection of the A-ring fragment following cleavage of the adjoining B-ring bonds [39]. When optimised, the m/z 363>121 is generally the most sensitive transition however, it is not always specific with interference reported from prednisolone and its metabolites [11, 17, 18] (Figure 3). Conversely, the m/z 363>327 [18, 19] and m/z 363>97 [17] transitions have been found to be more selective yet not as sensitive. Whereas the m/z 363>327 transition corresponds to cleavage of two hydroxyl groups, the m/z 363>97 transition represents cleavage and subsequent rearrangement on the A-ring fragment [40, 41]. In addition, Ray et al have also used m/z 363>267 as a secondary transition.

[Insert Figure 2 here]

[Insert figure 3 here]

Whereas ESI in the positive ionisation mode is favoured for cortisol analysis, four of the methods have used ESI in the negative ionisation mode [22, 24, 26, 30]. Of these, two have exploited of the ability of cortisol to form an adduct

Page 18 of 66

in this mode. Huang et al have used ammonium chloride to form a [M+35]- precursor ion and subsequently scanned for the m/z 397>35 product ion following fragmentation of cortisol and the chloride ion in the collision cell [24]. This allows for a very sensitive method however; it is notable that scanning for product ions in the lower range of the mass scale is generally not recommended as specificity can be compromised. The method described by Methlie et al has instead monitored the formic acid adduct [M+46] - and the corresponding m/z 407>297 transition following collision induced dissociation [22]. Conversely, two methods do not rely on adduct formation and instead scan for the native [M-H]- ion which fragments to [M-H-CH2O]-, this corresponds to m/z 361>331 [26, 30]. Relative to ESI in the positive ionisation mode, detection of cortisol in the negative ionisation mode has been reported to improve signal-to-noise [26, 30].

Interestingly, four of the methods summarised have used structural analogues as an internal standard including: flumethasone [26, 30], prednisolone-d6 [28] and 6-methylprednsiolone [24]. To ensure the internal standard provides true compensation for variability in ionisation, its retention time should be approximately equal to that of the analyte of interest. Therefore, given their widespread availability, it is highly recommended that isotopically labelled internal standards are used in preference to structural analogues for assay validation.

In general, short run times are desirable for a prompt result turnaround. The assay developed by Owen et al has been specifically designed for

Page 19 of 66

implementation into a routine clinical laboratory [17]. The other assays listed in Table 1 are mainly multiplexed and designed to ensure the adequate separation of all compounds scanned for. As such, they require longer run times.

One application where LC-MS/MS serum cortisol quantification has been found to be increasingly important is in patients prescribed metyrapone. This 11-hydroxlase inhibitor is used to block cortisol synthesis in patients with Cushing’s syndrome. In addition, it is used in some countries to help diagnose adrenal insufficiency. Serum cortisol measurement by routine immunoassays has been found to be prone to interferences in samples collected from patients prescribed metyrapone [42, 43]. Inhibition of 11- -hydroxlase can cause gross elevations in cortisol precursors (e.g. 11–deoxycortisol) that can interfere with immunoassay cortisol quantification therefore potentially masking hypoadrenalism [43, 44]. As metyrapone dosing regimens may be titrated against cortisol concentrations this has important ramifications on patient care. This fact has been recognised by the Endocrine Society in their recent clinical practice guideline concerning Cushing’s Syndrome treatment [44]. In this document, LC-MS/MS is recommended for serum and urine cortisol quantification in patients prescribed metyrapone as it circumvents the problems associated with immunoassays. McWhinney et al also suggest that when metyrapone is administered as part of a dynamic function test for the diagnosis of adrenal insufficiency, concomitant measurement of 11deoxycortisol along with cortisol can aid result interpretation [27].

Page 20 of 66

Urine free cortisol and its metabolites Urine free cortisol (UFC) is advocated for the investigation of hypercortisolism because it represents the non-protein bound cortisol fraction that is freely filtered at the glomerulus [4]. Whereas immunoassays suffer from interference due to cross reactivity with cortisol A-ring metabolites and other steroids, the selectivity and sensitivity of LC-MS/MS overcomes this to provide accurate results [45]. Consequently, many urinary free cortisol LC-MS/MS methods have been published (Table 2).

Urine cortisol measurements are only recommended following a 24h collection period [4], hence obtaining sufficient sample volume is rarely problematic. This is reflected by the relatively larger volumes that are used in urine assays compared to that of serum (Table 2).

Two of the methods listed make use of the simplest form of sample preparation namely ‘dilute-and-shoot’ [46, 47]. This can significantly reduce both preparation time and inter-analyst variation. However, this is at the expense of introducing relatively dirty samples into the LC-MS/MS system. This can be detrimental as the risk of blocking tubing or frits increases, column lifetime can reduce, and the mass spectrometer itself may require additional services as salts can deposit within the system. Although more comprehensive sample clean-up is recommended, analysts wishing to explore dilute-and-shoot should at least centrifuge the specimen prior to sampling to ensure particulates do not enter the LC-MS/MS system. In an evolution of dilute-and-shoot, Kushnir et al al have used a 2D chromatography system

Page 21 of 66

with an online C18 guard cartridge to trap cortisol [36]. In this method, the sample is first injected onto a guard column where it is trapped and subjected to an organic wash. This wash is sufficiently organic to remove hydrophilic interferences to waste without disrupting the interaction between the analyte of interest and the guard column phase. After the sample is suitably purified, the flow path is switched and the guard column eluent is diverted to the analytical column. The organic composition of the mobile phase is then increased to liberate the analyte of interest from the guard column onto the analytical column. This allows further sample clean-up in the form of traditional reverse-phase chromatography prior to MS/MS detection.

Page 22 of 66

Table 2 Analyte(s)

Sample volume (L)

Preparation

Column

Mobile phase (organic/buffer/modifier)

Run time (mins)

Ionisation

LOQ (nmol/L)

Reference

Cortisol

500

dilution (1:1 water)

C18

Methanol / NA / 0.003% trifluoroacetic acid

3

APCI +

13.8

[46]

Cortisol

500

Online C18 cartridge

C18

Methanol / 5 mmol/L ammonium formate / NA

8

APCI +

5.5

[36]

5

ESI +

21.5

[48]

4.5

ESI +

2

[38]

8

ESI -

1.38 6. 0 (Cortisol), 6.0 (Cortisone)

[24]

Cortisol

100

PPT (trichloroacetic acid)

C18

Cortisol

1000

SPE (HLB)

C18

Cortisol

250

LLE (ethyl acetate)

C18

Methanol / 2 mmol/L ammonium acetate / 0.1% formic acid Methanol / 2mmol/L ammonium acetate / 0.1% formic acid Methanol / 1 mmol/L ammonium chloride / NA

Cortisol, Cortisone

500

LLE (methylene chloride)

C18

Methanol / NA / NA

3

ESI +

Cortisol, Cortisone, 6-sulfatoxymelatonin

250 (minimum)

dilution (1:1 water)

C18

Acetonitrile / NA / 0.1 % formic acid

6

ESI +

3.8 (Cortisol) * 4.7 (Cortisone) *

[47]

Cortisol, Cortisone

100

Turboflow

C18

Methanol / NA / NA

9.5

APCI +

5.5 (Cortisol) 2.8 (Cortisone)

[54]

Cortisol

300

SPE (C8 Column)

C18 monolithic

Methanol / 2 mmol/L ammonium acetate / 0.1% formic acid

4.5

ESI +

5

[50]

Cortisol, Cortisone

1000

SPE (MAX)

C8

Methanol / 20 mmol/L ammonium formate / NA

10

ESI +

0.4 *

[51]

Cortisol, Cortisone, Tetrahydrocortisol, Allotetrahydrocortisol, Tetrahydrocrotisone Cortisol, Cortisone, Tetrahydrocortisol, Allotetrahydrocortisol, Tetrahydrocrotisone Cortisol, Cortisone, Tetrahydrocortisol, Allotetrahydrocortisol, Tetrahydrocrotisone

1000

SPE (HLB)

C18

Methanol / NA / 0.1% formic acid

18

ESI +

2000

SPE (HLB) / dilution (1:1 water)

Biphenyl

Methanol / NA / 0.1% formic acid

19.5

ESI -

480

PPT (acetonitrile)

C18

25

ESI -

Acetonitrile / 5 mmol/L ammonium acetate / 0.1% formic acid

Page 23 of 66

0.3 (Cortisol) 0.3 (Cortisone) 2.7 (Tetrahydrocortisol) 2.7 (Allo-tetrahydrocortisol) 2.3 (Tetrahydrocortisone) 8.3 (Cortisol) 1.7 (Cortisone) 0.8 (Tetrahydrocortisol) 0.8 (Allo-tetrahydrocortisol) 1.6 (Tetrahydrocortisone) 0.14 (Cortisol) 0.14 (Cortisone) (Tetrahydrocortisol) (Allo-tetrahydrocortisol) (Tetrahydrocortisone)

[37]

[52]

[53]

[49]

Alternative approaches to sample preparation have involved protein precipitation with trichloroacetic acid [48] or acetonitrile [49], LLE using ethyl acetate [24] or methylene chloride [37] and SPE [45, 50, 51, 52, 53]. With its enhanced purification providing improved sensitivity and specificity, SPE is the favoured approach. The advantages SPE confers are highlighted by the lower limits of quantification achieved using this technique relative to others. Although few authors disclose any preliminary work that influenced their choice of SPE phase, Fong et al evaluated their method using both a mixed mode anion exchange (MAX) phase and a mixed mode cation exchange (MCX) phase, superior recovery and imprecision were observed using the MAX cartridges [51]. In a novel approach, Sánchez-Guijo et al have introduced samples directly into a Turboflow™ chromatography system [54]. This uses turbulent solvent flow in a short chromatography column packed with porous particles to separate smaller molecules from larger ones (e.g. proteins) in complex biological matrices. Whereas the small molecules interact with particle pores, larger ones are size-excluded and discarded. Once separated, an analytical column can be used for improving separation of the smaller molecules prior to MS/MS analysis. The use of the Turboflow™ dramatically reduces bench-time preparation and inter-operator variability.

The majority of methods listed in Table 2 achieve chromatographic separation of cortisol using the hydrophobic interaction of C8 and C18 columns. However, in two original approaches, one method makes use of a monolithic phase [50] whereas another utilises a biphenyl column [53]. Monolithic phases rely on a silica based porous rod system to achieve chromatographic

Page 24 of 66

separation. The stability of these porous rods allow higher flow rates (up to 1 mL/min) without the consequential rise in pressure that is observed with particle based chromatography. This permits the development of efficient and rapid methods. Conversely, the biphenyl phase, like its phenyl counterpart, relies on the pi-pi interaction between hydrophobic aromatic compounds and the biphenyl ligand, this can produce exceptional separation of structurally similar compounds such as the stereoisomers tetra-hydrocortisol and allotetra-hydrocortisol [53].

Although most assays have used an isotopically labelled internal standard, one method has used 6--methylprednisolone to compensate for variation in ionisation [47]. Interestingly, another method evaluated methylprednisolone as an alternative for cortisol-d4 [53]. The results demonstrated that the structural analogue was 10-15% less accurate than the isotopically labelled internal standard for the quantification of cortisol and its metabolites.

Whereas ESI in the positive ionisation mode is the most common approach, three methods have used ESI in the negative ionisation mode [24, 49, 53]. Under these conditions, Cuzzola et al specify that the formic acid adduct [M+HCOO]- consistently fragments to [M-H-CH2O]- for all analytes providing high sensitivity and specificity [53]. In contrast, three methods have utilised APCI in the positive ionisation mode, this reportedly increased signal-to-noise by 2-3 fold compared to ESI [36, 46, 54]. Despite the differences used to achieve ionisation, all methods have determined an LOQ within the desirable clinical range for routine UFC measurements. The differences observed for

Page 25 of 66

ionisation may be directly related to the buffers and modifiers used, or the insource ionisation which can vary between manufacturers

Aside from hypercortisolism, quantification of urine cortisol and its metabolites can also be effective in the diagnosis of other pathologies related to disordered steroid production or metabolism. Several enzymes are involved in modulating the availability, metabolism and clearance of cortisol (Figure 4) [55]. Hepatic and adipose tissue both express the enzyme 11-hydroxysteroid dehydrogenase type I (11HSD1) which converts the inactive cortisone to the active cortisol. This reaction can also be reversed in the presence of 11hydroxysteroid dehydrogenase type II (11HSD2) which is expressed in renal and colon cells. Furthermore, the action of the hepatic A-ring reductases (5 and 5reductase), in conjunction with 3-hydroxysteroid dehydrogenase (3-HSD), inactivate cortisol to the tetrahydrometabolites: allotetrahydrocortisol (5THF), tetrahydrocortisol (5THF), and tetrahydrocortisone (THE). Thus, as conjugation is predominantly a hepatic process, the ratio of cortisol and cortisone tetrahydrometabolites (i.e. 5THF + 5THF / THE) provides an index of hepatic 11HSD1 activity. Similarly, the free cortisol:cortisone ratio provides information on the activity of 11HSD2. In addition, the ratio of 5THF / 5THF reflects the relative activities of global 5 and 5reductase activity [56, 57].

[Insert Figure 4 here]

Page 26 of 66

Much of the early work measuring these metabolites was undertaken using GC-MS with the syndrome of apparent mineralocorticoid excess (AME) one of the first conditions identified [58]. AME is characterised by a congenital defect in 11HSD2 which prevents the oxidation of cortisol to cortisone. Without this protective mechanism, cortisol can accumulate and stimulate the mineralocorticoid receptor (MR). Interestingly, AME can also be acquired through inhibition of 11HSD2 secondary to carbenoxolone therapy or excessive consumption of glycyrrhizinic acid (commonly found in liquorice). Unregulated MR stimulation results in sodium and water retention consequently leading to hypertension. Two congenital forms of AME have been described, both of which can be determined by virtue of an elevated urine free cortisol to cortisone ratio [59]. However, Type II is distinguishable from Type I as A-ring reduction is the major deficiency [60, 61].

In addition to AME, GC-MS has helped to identify glucocorticoid metabolic imblances in patients with glucocorticoid suppressible aldosteronism and adrenal adenomas [62, 63], females with polycystic ovarian syndrome [64], obesity [65], and congenital adrenal hyperplasia secondary to 21-hydroxylase deficiency [66]. Furthermore, a debate has recently emerged in the literature suggesting that the cross-reactivity observed with cortisol metabolites in immunoassays may actually increase their diagnostic sensitivity for Cushing’s Syndrome [67]. Therefore, to further increase the diagnostic sensitivity of LCMS/MS it may be necessary to measure cortisol and its metabolites to provide a complete assessment of hypercortisolism. Collectively, this evidence has

Page 27 of 66

provided an impetus for the development of LC-MS/MS assays capable of measuring cortisol and its metabolites.

Consequently, several of the assays listed in Table 2 have been developed to measure cortisone concomitantly with cortisol [37, 47, 49, 51, 52, 53, 54]. In comparison to cortisol, cortisone differs by the presence of a C=O group in position 11. Taylor et al suggest that the presence of this bond may be responsible for the different ionisation pattern observed with cleavage of the C-ring (as opposed to the B-ring with cortisol) producing the most abundant fragment m/z 361 > 163 when analysed in the positive ionisation mode [37]. Interestingly, as prednisolone and cortisone share the same molecular weight (360.44 Da) crosstalk has also been reported in their analysis [22, 28]. This can be prevented by ensuring adequate chromatographic separation of the two compounds.

In normal adults, approximately twice as much cortisone is excreted compared to cortisol, as such sensitivity has not proved problematic. Hence, it has been suggested that the the cortisol:cortisone ratio can add value to result interpretation and the evaluation of 11HSD2 activity [68].

More recently, several LC-MS/MS assays have been designed to provide a more complete picture of cortisol metabolism through measurement of cortisol, cortisone and their tetrahydrometabolites: 5THF, 5THF and THE. One of the earliest methods published quantified only the tetrahydrometabolites and required a derivatisation procedure that increased

Page 28 of 66

sensitivity 15-80-fold [69]. Campino et al have since published a method using an F5 column that did not allow for chromatographic separation of the stereoisomers 5THF and 5THF and therefore provided limited information regarding the 5 and 5reductase activity [70]. F5 columns act as a lewis acid due to the presence of 5 electronegative fluorine groups. This ensures the phenyl ring is electron deficient which can help separate closely related ring systems such as those present in steroids. However, care is still required to ensure the gradient encourages separation of closely related compounds such as stereoisomers. More recently a series of methods have been published that achieve separation of all compounds including the 5THF and 5THF [49, 52, 53, 71]. Interestingly, two of the methods have favoured ESI in the negative ionisation mode owing to its enhanced signal-to-noise relative to ESI in the positive ionisation mode [53] However, this observation may be instrument-specific and it is always worth experimenting in both modes. Importantly, the cycle times injection-to-injection for these three assays range from 18 – 25 minutes [49, 52, 53, 71], run times this long are not always conducive with routine, high-throughput laboratories. Another potential problem with these assays is they are unable to truly compensate for variation in ionisation due to a lack of availabile isotopically labelled internal standards. Taking the method of Zhai et al as an example, here the structural analogue methylprednisolone elutes after approximately 6 minutes, the analytes of interest subsequently elute between 8 and 18 minutes [49]. Thus, it is apparent that there is a significant period where no allowance is made for variation in ionisation efficiency. In an original approach, Marcos et al have

Page 29 of 66

used a C18 column to achieve excellent separation of a panel of 67 corticosteroids including many additional glucocorticoid metabolites [71].

Salivary cortisol By virtue of its lipophilic nature, serum free cortisol can passively diffuse into saliva via the acinar cells in a process that is independent of salivary flow rate [72]. Therefore, like UFC, salivary cortisol quantification provides an index of serum free cortisol concentration [73]. The use of saliva for cortisol measurement has several advantages over both urine and serum: as a matrix itself, saliva contains relatively fewer interfering compounds and in that regard presents less of an analytical challenge. In addition, its non-invasive, costeffective collection is appealing to both patients and clinicians. Consequently, advances in sampling techniques coupled with a strong evidence base have propelled salivary cortisol from a novel assay confined to endocrine and psychological research into routine clinical practice. Indeed, it is now recommended as a first-line diagnostic test in the investigation of Cushing’s Syndrome [4]. Its transition into clinical laboratories has in part been driven by LC-MS/MS with accurate quantification of low salivary cortisol concentrations now possible. Originally measured by RIA and immunoassay, salivary cortisol has been reported to suffer from non-specific cross-reactivity with endogenous and exogenous glucocorticoids. In addition, poor standardisation between assays has contributed to assay-specific variation meaning results are not directly comparable [74, 75]. This problem is compounded by the expression of 11HSD2 in the parotid glands which ensures the cortisone:cortisol ratio is significantly higher than in serum.

Page 30 of 66

The structural similarity of cortisone to cortisol has proved difficult for immunoassays to differentiate between the two. These limitations have largely been circumvented by the application of LC-MS/MS. Table 3 provides a summary of recently published LC-MS/MS salivary cortisol assays.

Page 31 of 66

Table 3

Analyte(s)

Sample Volume (L)

Preparation

Column

Mobile Phase (organic/buffer/modifier)

Run time (mins)

Ionisation

LOQ (nmol/L)

Reference

Cortisol

100

LLE (dichloromethane)

C18

Methanol / NA / NA

10

ESI -

0.07

[78]

Cortisol

100 / 200

PPT (heat denaturation) /online SPME (Supel Q PLOT capillary)

C8

Methanol / NA / 1% acetic acid

5

ESI +

0.01*

[80]

Cortisol

250

PPT (acetonitrile / acetic acid)

C8

Methanol / NA / 0.5% acetic acid,

5

ESI +

1.4

[79]

Cortisol, cortisone

50

online SPE (C18)

C18 monolithic

Methanol, 0.1% formic acid

4

ESI +

0.75 (Cortisol) 0.75 (Cortisone)

[83]

Cortisol, cortisone

500

PPT / SPE (strata X) / derivitisation

C18 monolithic

Methanol, 500 mmol/L ammonium formate / NA

Not stated

ESI +

0.66 (Cortisol) 2.77 (Cortisone)

[76]

Cortisol, Cortisone

100

online SPE (C8)

C18

Methanol / 2 mmol/L ammonium acetate /0.1% formic acid

5

ESI +

0.39 (Cortisol) 0.78 (Cortisone)

[81]

Cortisol

200 / 100

online SPE (C8)

C18

Methanol / 2 mmol/L ammonium acetate / 0.1% formic acid

4

ESI +

2

[82]

Cortisol, Cortisone, Melantonin

50

Turboflow

C18

Methanol, 1.5 mmol/L ammonium formate / 0.1% formic Acid

13

ESI +

0.55 (Cortisol), 5.5 (Cortisone)

[84]

Page 32 of 66

These methods share a common requirement for small sample volumes ranging from 50-250 L however, one method requires 500 µL [76]. Generally, saliva is not a difficult specimen to obtain, with the availability of dedicated collection vessels promoting standardised collection procedures. Nevertheless, despite following these procedures, salivary yields can be poor in paediatric populations and acutely unwell patients. For this reason, small sample volumes remain desirable for routine analysis.

Salivary cortisol quantification is used to assess the nadir of an individuals diurnal rhythm. Consequently, the cortisol concentration in normal subjects at this time can challenge the functional sensitivity of immunoassays. LC-MS/MS analysis is better suited to this task because it is 10-100 fold more sensitive than RIA and confers greater specificity [77]. To achieve this, LC-MS/MS salivary cortisol methods generally benefit from a sample preparation that purifies and concentrates cortisol prior to analysis. This has been accomplished by various techniques including LLE using dichloromethane [78] and protein precipitation with evaporation of the supernatant prior to reconstitution [79]. Heat denaturation has also been combined with an online solid phase micro extraction (SPME) technique that consists of a 2D chromatography system that loads the sample onto a fused silica capillary [80]. In a similar approach the use of off-line [76] or on-line [81, 82, 83] SPE has been used for sample purification. When performed off-line, SPE can take a relatively long time to prepare with the process heavily reliant on operator proficiency. Conversely, online SPE reduces inter-operator variability and ensures maximum recovery of the analyte as loss is minimal compared to the

Page 33 of 66

offline evaporation of an eluate. Online SPE using cartridges has advantages over the use of a guard column as the washes can be performed separately to the main flow path hence solvents other than those used in the mobile phase can be utilised for sample clean up [83]. Another approach to online sample clean-up has used Turboflow™ technology to purify the sample prior to introduction to the LC-MS/MS, this requires little sample pre-treatment ensuring minimal inter-operator variation [84].

Two assays have been validated using a monolithic C18 phase [76, 83]. As discussed above, the application of a monolithic phase as opposed to a traditional silica particle based column can facilitate the development of rapid and efficient methods such as that described by Jones et al which achieves separation of cortisol and cortisone within 4 minutes (Figure 5) [83].

[Insert Figure 5 here]

Sensitivity can be crucial in salivary cortisol methods, as such the majority of assays listed in Table 3 use ESI in the positive ionisation mode to monitor the m/z 363>121 transition [81, 82, 83, 84]. Conversely, Jonsson et al have reported observing interference in the m/z 363>121 transition and have instead opted to use m/z 363>309 [79]. Vieira et al use hydroxylamine for derivatisation and scan for the corresponding m/z 393>136. Another method has analysed cortisol in the negative ionisation mode using the m/z 361>331 transition [78]. Finally, one method uses just a single quadrupole to scan for the [M+H]+ ion corresponding to m/z 363 [80]. The use of single quadrupole

Page 34 of 66

mass spectrometers is generally not recommended, specificity is considerably reduced and therefore the chance of detecting an interfering compound is increased.

Salivary cortisone and cortisol can be measured concurrently [76, 81, 83, 84]. It has been suggested that this can facilitate result interpretation [81]. The non-invasive nature of salivary collection lends itself to the assessment of hydrocortisone replacement therapy whereby day-curves can be produced to ensure patients are provided with optimal steroid replacement. Measuring cortisone and cortisol concomitantly can demonstrate contamination from oral hydrocortisone shown by an unexpectedly high cortisol:cortisone ratio [81]. Perogamvros et al have proposed that the salivary free cortisone concentration may more accurately reflect the concentration of free serum cortisol [85].

Although much work has focused on the application of salivary cortisol in the diagnosis of Cushing’s syndrome, there is evidence to suggest it can also be used in the diagnosis of adrenal insufficiency [81, 86, 87]

Future directions Over the past decade there has been an exponential rise in the number of LCMS/MS glucocorticoid publications. This trend does not appear to be declining with advances in sample preparation, chromatography, and mass

Page 35 of 66

spectrometry providing scientists with the instrumentation required to improve existing techniques. This, coupled with increasing clinical knowledge of glucocorticoid metabolism, is providing a solid platform to continue advancing the field.

For many scientists and clinicians with an interest in glucocorticoid analysis, the development of robust serum/plasma free cortisol assays remains an important goal. Free hormone quantitation is analytically challenging as analysis itself can disturb the equilibrium between the free and protein bound fractions bringing into question the integrity of the final result. Although free serum cortisol assays have been described, they typically require extensive sample preparation such as ultrafiltration [24, 88] or equilibrium dialysis [89]. These techniques are not conducive to high-throughput laboratories and it is possible that advances in sample preparation or LC-MS/MS methodology could advance free hormone analysis into a routine application. This would require extensive clinical studies to support its utility.

The case for LC/MS/MS measurement of urine cortisol, cortisone and their metabolites has been outlined above. Although initial assays have been described, their cycle times are relatively long when compared to most routine LC-MS/MS methods [49, 52, 53]. Future work is required to develop these assays for routine use. In addition, there is a general requirement for manufactures to provide isotopically labelled internal standards for the glucocorticoid metabolites.

Page 36 of 66

Another potentially exciting development in the field of LC-MS/MS glucocorticoid analysis is the quantification of hair total cortisol. As cortisol has been found to deposit in the hair shaft, it is postulated that hair cortisol analysis may allow clinicians to assess chronic exposure to cortisol to aid in the diagnosis of Cushing’s Syndrome, monitor hydrocortisone replacement therapy and to assess pathologies associated with chronic stress [90]. Despite the growing evidence base [91, 92, 93, 94], the utility of hair cortisol measurements has yet to be recognised in any clinical guidelines. One possible reason for this is the pre-analytical factors involved with hair collection [95]. Further work is required to fully assess the clinical utility of glucocorticoid hair measurements.

It is also important to appreciate that unlike cortisol, no RMPs currently exist for any other endogenous glucocorticoids or their metabolites. For this reason, metrological traceability cannot be guaranteed and assay standardisation may in future be identified as a problem. Thus, there remains a requirement for RMPs to be developed that are sufficiently accurate and precise they can be used to produce CRMs that will promote assay standardisation.

One of the major advantages LC-MS/MS holds over immunoassay is the ability for laboratories to develop novel assays relatively quickly. With the introduction of new equipment and the recognition of new endogenous metabolites opportunities will continue to be presented for novel assay development.

Page 37 of 66

Historically, one factor that has limited the wider uptake of LC-MS/MS is the relative expense of the equipment compared to other analysers. As the number of manufacturers has increased and technology has advanced, entrylevel instruments have gradually become more affordable. This has helped a greater number of laboratories adopt LC-MS/MS technology. However, their cost and the expertise required to successfully operate them remain an issue, especially in the developing world. To circumvent this, there is a general requirement for manufacturers to develop equipment that can integrate into routine, high-throughput, automated chemistry analysers that will require minimal operator input.

Summary The field of glucocorticoid quantitation is being propelled by LC-MS/MS. With its adoption by routine clinical chemistry laboratories, there has been increasing awareness of the advantages this powerful technology offers over existing methodologies. This recognition has now moved beyond the laboratory with many clinicians and research scientists cognisant of the advantages LC-MS/MS offers for patient diagnosis and management as well as for high impact clinical research.

Despite its growing reputation, there have been general concerns raised regarding LC-MS/MS analysis. Many of these centre on the variability observed across methods that are designed to measure the same analyte [96, 97, 98, 99]. This concern has been justified in this review. Although many methods rely on established validation criteria (e.g. FDA IVD guidelines [100])

Page 38 of 66

there is widespread variation to the approaches used with many of these, especially the older ones, reliant on inadequate validation.

Several of the methods reviewed have used structural analogues as opposed to a stable, isotopically labelled internal standard. This can be problematic if the internal standard and analyte of interest do not co-elute. To compensate for variation in ionisation efficiency, it is essential that the physiochemical properties of the internal standard resemble those of the analyte of interest as closely as possible. Although isotopically labelled internal standards are recommended over structural analogues, problems can still be encountered. Duxbury et al have shown that when using cortisol-d2 as an internal standard to measure cortisol, the naturally occurring isotope cortisol-365 can falsely increase the internal standard concentration therefore reducing the relative response for cortisol [101]. This can largely be avoided by selecting internal standards with more than 2 substituted deuteriums. Davison et al have also reported that the location of the deuterium labels can be crucial as some sites will back-exchange with hydrogen atoms thereby causing a reduction in internal standard area and an increase the relative response of the analyte [102]. Many of the concerns surrounding internal standards have been addressed with the introduction of superior C13 based compounds [38].

When clinical samples are to be measured, the assay should be validated using samples taken using blood collection tubes from several manufacturers and using different anticoagulants. These could all effect how well the analyte of interest ionises because of differing ionisation efficiency profiles. Matrix

Page 39 of 66

effects can be evaluated using infusion experiments to qualitatively test for ion suppression [31] and also quantitatively by comparing the recovery of spiked analyte from aqueous and plasma samples [103]. It is also possible to monitor interferences directly using targeted analysis of individual phospholipids [104].

Another common problem is the investigation of analytical interference. Method validation typically includes analysis of a panel of endogenous and exogenous compounds structurally analogous to the analyte of interest. The potential interferent is normally added in excess to a solution and injected into the LC-MS/MS system, any interference is deemed significant if a signal greater than the LOQ of the assay is detected in any of the m/z transitions scanned for. Although sufficient for endogenous compounds, this approach is not comprehensive enough for exogenous compounds since it is often parent compound metabolites that manifest as an interferent. For example, although prednisolone itself can be separated from cortisol, its metabolites have been reported to interfere in LC-MS/MS analysis [11,17, 18]. As it is not always possible to investigate drug metabolites, it is best practice to develop a method capable of identifying potential interference. To achieve this, analysts can scan for quantifying and qualifying ion transitions for the analyte of interest and calculate their relative ratio. This uses the principle that the analyte fragments in a similar way for patient samples as it does for interference-free samples (e.g. calibrators). Thus, comparison of the ion ratio of a patient sample to that of an interference-free sample can help to identify interference. This feature is now integrated into most software packages.

Page 40 of 66

Documentation detailing the specific steps for development and validation of quantitative LC-MS/MS assays has been reviewed [105]. Moreover, recent guidance specifically covering LC-MS/MS validation is now available and should help promote the standardisation of method development [106].

With improvements in technology allowing faster scan speeds, analysis of multiple analytes in one assay is now routinely undertaken. However, although multiplexed assays can be advantageous, their application in routine clinical work can prove difficult. Producing fresh calibrators for a singleanalyte assay can itself prove challenging since it is always necessary to ensure patient results do not significantly differ from those previously issued. For multiplexed assays, the likelihood of a problem being encountered during the re-calibration process is significantly increased. Some manufacturers are now beginning to produce metrologically traceable LC-MS/MS kits that include all reagents and consumables required for a given assay. This can help promote standardisation and minimise result variability between laboratories.

In conclusion, the implementation of LC-MS/MS in routine clinical laboratories provides scientists and clinicians with state-of-the-art technology that can produce accurate and precise results. Its application to glucocorticoid analysis is particularly relevant where the non-specific cross-reactivity of commercial immunoassays can potentially compromise patient diagnosis and management. Furthermore, the ability of LC-MS/MS to simultaneously measure several analytes potentiates its clinical utility. With careful attention

Page 41 of 66

to detail during assay validation, many of the pitfalls associated with this technology can be avoided and robust assays developed.

Acknowledgements The authors would like to thank Dr. Finlay Mackenzie of UK NEQAS for granting permission to use the information provided in Figure 1.

Page 42 of 66

Figure 1: Method variation for cortisol analysis. This is a recent report from the UK NEQAS serum cortisol EQA scheme. The wide dispersion of results is clearly visible along with a bimodal distribution which is attributed to matrix effects. (The authors wish to thank Dr Finlay Mackenzie of UK NEQAS for granting permission to use this report).

Figure 2: Collision induced dissociation of cortisol. Precursor (A) and product (B) ion scans for a 1 mg/L solution of cortisol infused into a Waters Acquity TQD™ LC-MS/MS system. The m/z 363 peak corresponding to the [M+H]+ cortisol ion is clearly visible in panel A. In panel B the precursor ion has been subjected to collision induced dissociation and the product ion scan has detected the major fragments. The structures are the theoretical fragments that correspond to the accurate mass of the product ions.

Figure 3: A sample serum cortisol chromatogram. Chromatograms taken from a routine cortisol analysis as described by Owen et al [17]. The cortisol concentration was quantified as 15.6 nmol/L using the m/z 363>97 transition (A). The presence of prednisolone in this sample has been detected (C), the reported isobaric interference is apparent in the m/z 363>121 transition (B). Panel D represents the internal standard, cortisol-d4.

Figure 4: An overview of endogenous cortisol metabolism. The metabolites shown can be used to provide a comprehensive overview of glucocorticoid metabolism.

Page 43 of 66

Figure 5: A sample salivary cortisol chromatogram. A chromatogram taken from the routine analysis of a sample for salivary cortisol and cortisone using the method described by Jones et al [83]. The cortisol concentration was quantified as 1.5 nmol/L (A) whereas cortisone was 8.3 nmol/L (C). Panels B and D corresponds to cortisol-d4 and cortisone-d7 respectively.

Table 1: A summary of the methodologies used to measure serum/plasma endogenous glucocorticoids (* = Limit of Detection).

Table 2: An overview of the methodologies used to measure urine endogenous glucocorticoids (* = Limit of Detection).

Table 3: A summary of the methodologies used to measure salivary endogenous glucocorticoids (* = Limit of Detection).

Page 44 of 66

References

1. Arlt W, Stewart PM. Adrenal corticosteroid biosynthesis, metabolism, and action. Endocrinol Metab Clin North Am. 2005; 34:293-313.

2. Cain DW, Cidlowski JA. Specificity and sensitivity of glucocorticoid signalling in health and disease. Best Pract Res Clin Endocrinol Metab. 2015;29:545-56.

3. Spiga F, Lightman SL. Dynamics of adrenal glucocorticoid steroidogenesis in health and disease. Mol Cell Endocrinol. 2015;15:227-34.

4. Nieman LK, Biller BM, Findling JW, Newell-Price J, Savage MO, Stewart PM, Montori VM. The diagnosis of Cushing’s syndrome: an Endocrine Society Practice Guideline. J Clin Endocrinol Metab. 2008;93:1526-40. doi: 10.1210/jc.2008-0125. Epub 2008 Mar 11

5. Bornstein SR, Allolio B, Arlt W, Barhel A, Don-Wauchope A, Hammer GD, Husebye ES, Merke DP, Murad MH, Stratakis CA, Torpy DJ. Diagnosis and treatment of primary adrenal insufficiency: An Endocrine Society Clinical Practice Guideline. J Clin Endocrinol Metab. 2016;101:363-89.

Page 45 of 66

6. Bancos I, Hahner S, Tomlinson J, Arlt W. Diagnosis and management of adrenal insufficiency. Lancet Diabetes Endocrinol. 2015;3;216-26.

7. Adaway JE, Keevil BG, Owen LJ. Liquid chromatography tandem mass spectrometry in the clinical laboratory. Ann Clin Biochem. 2015;52:1838.

8. Thienpont L, Siekmann L, Lawson A, Colinet E, De Leenheer A. Development, validation, and certification by isotope dilution gas chromatography-tandem mass spectrometry of lyophilised human serum reference materials for cortisol (CRM 192 and 193) and progesterone (CRM 347 and 348). Clin Chem. 1991; 37:540-6.

9. Tai SS, Welch MJ. Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry. Anal Chem. 2004;76:1008-14.

10. Kawaguchi M, Eyama S, Takatsu A. Automated isotope dilution liquid chromatography-tandem mass spectrometry with on-line dilution and solid phase extraction for the measurement of cortisol in human serum sample. J Pharm Biomed Anal. 2014;96:220-3.

Page 46 of 66

11. Hawley JM, Owen LJ, MacKenzie F, Mussel C, Cowen S, Keevil BG. Candidate reference measurement procedure for the quantification of total serum cortisol with LC-MS/MS. Clin Chem. 2016;62:262-9).

12. Vesper HW, Thienpont LM. Traceability in laboratory medicine. Clin Chem. 2009;55:1067-75.

13. Dodd AJ, Ducroq DH, Neale SM, Wise MP, Mitchem KL, Armston A, Barth JH, El-Farhan N, Rees DA, Evans C. The effect of serum matrix and gender on cortisol measurement by commonly used immunoassays. Ann Clin Biochem. 2013;51:379-385. doi: 10.1177/0004563213514567

14. Manguso F, Bennato R, Lombardi G, Viola A, Riccio E, Cipolletta L. Electrochemiluminescence immunoassay method underestimates cortisol suppression in ulcerative colitis patients treated with oral prednisone. World J Gastroenterol. 2014;20:10895-9.

15. Soldin SJ, Soldin OP. Steroid hormone analysis by tandem mass spectrometry. Clin Chem 2009;55:1061-6.

Page 47 of 66

16. Krone N, Hughes BA, Lavery GG, Stewart PM, Arlt W, Shackleton CHL. Gas chromatography/mass spectrometry (GC/MS) remains a preeminent discovery tool in clinical steroid investigations even in the era of fast liquid chromatography tandem mass spectrometry (LC/MS/MS). J Steroid Biochem Mol Biol. 2010;121:496-504

17. Owen LJ, Adaway JE, avies S, Neale S, El-Farhan N, Ducroq D, Evans C, Rees DA, Mackenzie F, Keevil BG. Development of a rapid assay for the analysis of serum cortisol and its implementation into a routine service laboratory. Ann Clin Biochem. 2013;50:345-52.

18. Ionita IA, Fast DM, Akhlaghi F. Development of a sensitive and selective method for the quantitiatve analysis of cortisol, cortisone, prednisolone, and prednisone in human plasma. J Chromatogr B Analyt Technol Biomed Life Sci. 2009;877:765-72.

19. Broccardo CJ, Schauer KL, Kohrt WM, Schwartz RS, Murphy JP, Prenni JE. Multiplexed analysis of steroid hormones in human serum using novel microflow tile technology and LC-MS/MS. J Chromatogr B Analyt Technol Biomed Life Sci. 2013;934;16-21. Doi: 10.1016/j.jchromb.2013.06.031.

Page 48 of 66

20. Kushnir MM, Neilson R, Roberts WL, Rockwood AL. Cortisol and cortisone analysis in serum and plasma by atmospheric pressure photoionization tandem mass spectrometry. Clin Biochem. 2004;37:357-62.

21. Kao PC, Machacek DA, Magera MJ, Lacey JM, Rinaldo P. Diagnosis of adrenal cortical dysfunction by liquid chromatography-tandem mass spectrometry. Ann Clin Lab Sci. 2001;31:199-204.

22. Methlie P, Hustad SS, Kellman R, Almås B, Erichsen MM, Husebye E, Løvås K. Multisteroid LC-MS/MS assay for glucocorticoids and androgens, and its application in Addison’s disease. Endocr Connect. 2013;2:125-136.

23. Lee S, Lim HS, Shin HJ, Kim SA, Park J, Kim HC, Kim H, Kim HJ, Kim YT, Lee KR, Kim YJ. Simultaneous determination of cortisol and cortisone from human serum by liquid chromatography-tandem mass spectrometry. J Anal Methods Chem. 2014; doi:10.1155/2014/787483 (Epub).

24. Huang W, Kalhorn T, Baillie M, Shen D, Thummel K. Determination of free and total cortisol in plasma and urine by liquid chromatographytandem mass spectrometry. Ther Drug Monit. 2007;29:215-224.

Page 49 of 66

25. Ray JA, Kushnir MM, Rockwood AL, Wayne Meikle A. Analysis of cortisol, cortisone and dexamethasone in human serum using liquid chromatography tandem mass spectrometry and assessment of cortisol: Cortisone ratios in patients with impaired kidney function. Clin Chim Acta. 2011;412:1221-1228. doi:10.1016/j.cca.2011.03.016

26. Frerichs VA, Tornatore KM. Determination of the glucocorticoids prednisone, prednisolone, dexamethasone, and cortisol in human serum using liquid chromatography coupled to tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2004;802:329-38.

27. McWhinney BC, Briscoe SE, Ungerer JPJ, Pretorius CJ. Measurement of cortisol, cortisone, prednisolone, dexamethasone and 11deoxycortisol with ultra high performance liquid chromatographytandem mass spectrometry: Application for plasma, plasma ultrafiltrate, urine and saliva in a routine laboratory. J Chromatogr B Analyt Technol Biomed Life Sci. 2010; 878: 2863-9.

28. Sæves I, Vethe NT, Bergan S. Quantification of 6 glucocorticoids in human plasma by liquid chromatography tandem mass spectrometry: method development, validation, and assessment of matrix effects. Ther Drug Monit. 2011;33:402-10.

Page 50 of 66

29. Hempen C, Elfering S, Mulder AH, van den Bergh FA, Maatman RG. Dexamethasone suppression test: development of a method for simultaneous determination of cortisol and dexamethasone in human plasma by liquid chromatography/tandem mass spectrometry. Ann Clin Biochem. 2012;49:170-6.

30. DiFrancesco R, Frerichs V, Donnelly J, Hagler C, Hochreiter J, Tornatore KM. Simultaneous determination of cortisol, dexamethasone, methylprednisolone, prednisone, prednisolone, mycophenolic acid and mycophenolic acid glucuronide un human plasma utilising liquid chromatography-tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2007;859:42-51.

31. Annesley TM. Ion suppression in mass spectrometry. Clin Chem. 2003;49:1041-4.

32. Taylor PJ. Matrix effects: the Achilles heel of quantitative highperformance liquid chromatography-electrospray-tandem mass spectrometry. Clin Biochem. 2005;38:328-34.

33. Polson C, Sarkar P, Incledon B, Raguvaran V, Grant R. Optimisation of protein precipitation based upon effectiveness of protein removal and ionisation effect in liquid chromatography-tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2003;785:263-75.

Page 51 of 66

34. Keevil BG. Novel liquid chromatography tandem mass spectrometry (LC-MS/MS) methods for measuring steroids. Best Pract Res Clin Endocrinol Metab. 2013;27:663-74.

35. Chemspider. http://www.chemspider.com (accessed February 2015).

36. Kushnir MM, Rockwood AL, Nelson GJ, Terry AH, Meikle AW. Liquid chromatography-tandem mass spectrometry analysis of urine free cortisol. Clin Chem. 2003;49:965-967. doi: 10.1373/49.6.965

37. Taylor RL, Machacek D, Singh RJ. Validation of a high-throughput liquid chromatography-tandem mass spectrometry method for urinary cortisol and cortisone. Clin Chem. 2002; 48:1511-9.

38. Vogeser M, Seger C. Pitfalls associated with the use of liquid chromatography-tandem mass spectrometry in the clinical laboratory. Clin Chem. 2010;56:1234-44.

39. Antignac J-P, Le Bizec B, Monteau F, Poulain F, André F. Collisioninduced dissociation of corticosteroids in electrospray tandem mass spectrometry and development of a screening method by high performance liquid chromatography/tandem mass spectrometry. Rapid Commun Mass Spectrom. 2000;14:33-39.

Page 52 of 66

40. Thevis M, Beuck S, Höppner S, Thomas A, Held J, Schäfer M, Oomens J, Schänzer W. Structure elucidation of the diagnostic product ion at m/z 97 derived from Androst-4-en-3-one-based steroids by ESICID and IRMPD spectroscopy. J Am Soc Mass Spectrom. 2012;23:537-46.

41. Pozo OJ, Eenoo PV, Deventer K, Grimalt S, Sancho JV, Hernández F, Delbeke FT. Collision-induced dissociation of 3-keto anabolic steroids and related compounds after electrospray ionisation. Considerations for structural elucidation. Rapid Commun Mass Spectrom. 2008;22:4009-24.

42. Owen LJ, Halsall DJ, Keevil BG. Cortisol measurement in patients receiving metyrapone therapy. Ann Clin Biochem. 2010;47:573-5.

43. Monaghan PJ, Owen LJ, Trainer PJ, Brabant G, Keevil BG, Darby D. Comparison of serum cortisol measurement by immunoassay and liquid chromatography-tandem mass spectrometry in patients receiving -hydroxlase inhibitor metyrapone. Ann Clin Biochem. 2011;48:441-6.

44. Nieman LK, Biller BM, Findling JW, Murad MH, Newell-Price J, Savage MO, Tabarin A. Treatment of Cushing;s Syndrome: An Endocrine Society Clinical Practice Guideline. J Clin Endocrinol Metab. 2015;100:2807-31.

Page 53 of 66

45. Wood L, Ducroq DH, Fraser HL, Gillingwater S, Evans C, Pickett AJ, Rees DW, John R, Turkes A. Measurement of urinary free cortisol by tandem mass spectrometry and comparison with results obtained by gas chromatography-mass spectrometry and two commercial immunoassays. Ann Clin Biochem. 2008;45:380-88.

46. Nasser AE, Varshney N, Getek T, Cheng L. Quantitative analysis of hydrocortisone in human urine using a high-performance liquid chromatographic-tandem mass spectrometric-atmospheric-pressure chemical ionisation method. J Chromatogr Sci. 2001;39:59-64.

47. Sniecinska-Cooper AM, Shah AJ, Dimitriou D, Iles RK, Butler SA, Bayford R. Determination of urinary cortisol, cortisone and 6sulfatoxymelatonin using dilute and shoot ultra-high pressure liquid chromatography mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2015;978-979:18-23.

48. Wear JE, Owen LJ, Duxbury K, Keevil BG. A simplified method for the measurement of urinary free cortisol using LC-MS/MS. J Chromatogr B Analyt Technol Life Med Sci. 2007;858:27-31.

49. Zhai X, Chen F, Zhu C, Lu Y. A simple LC-MS/MS method for the determination of cortisol, cortisone and tetrahydro-metabolites in human urine: Assay development, validation and application in depression patients. J Pharm Biomed Anal. 2015;107:450-5.

Page 54 of 66

50. McCann SJ, Gillingwater S, Keevil BG. Measurement of urinary free cortisol using liquid chromatography-tandem mass spectrometry: comparison with urine adapted ACS:180 serum cortisol chemiluminescent immunoassay and development of a new reference range. Ann Clin Biochem. 2005;42:112-8.

51. Fong BM, Tam S, Leung KS. Improved liquid chromatography-tandem mass spectrometry method in clinical utility for the diagnosis of Cushing’s syndrome. Anal Bioanal Chem 2010.396;783-90.

52. Allende F, Solari S, Campino C, Carvajal CA, Lagos CF, Vecchiola A, Valdivia C, Baudrand R, Owen GI, Fardella CE. LC-MS/MS method for the simultaneous determination of free urinary steroids. Chromatographia. 2014;637-42.

53. Cuzzola A, Mazzinin F, Petri A. A comprehensive study for the validation of a LC-MS/MS method for the determination of free and total forms of urinary cortisol and its metabolites. J Pharm Biomed Anal. 2014;94:203-9.

Page 55 of 66

54. Sánchez-Guijo A, Hartmann MF, Shi L, Remer T, Wudy SA. Determination of free cortisol and free cortisone in human urine by online turbulent flow chromatography coupled to fused-core chromatography-tandem mass spectrometry (TFC-HPLC-MS/MS). Anal Bioanal Chem. 2014;406:793-801.

55. Remer T, Maser-Gluth C, Wudy SA. Glucocorticoid measurements in health and disease-metabolic implications and the potential of 24-hour urine analyses. Mini Red Med Chem. 2008;8:153-70.

56. Andrew R, Phillips DI, Walker BR. Obesity and gender influence cortisol secretion and metabolism in man. J Clin Endocrinol Metab. 1998;83:1806-9.

57. Tomlinson JW, Finney J, Hughes BA, Hughes SV, Stewart PM. Reduced glucocorticoid production rate, decreased 5 alpha-reductase activity, and adipose tissue insulin sensation after weight loss. Diabetes.2008;57:1536-43.

58. Ulick S, Levine LS, Gunczler P, Zanconato G, Ramirez LC, Rauh W, Rösler A, Bradlow HL, New MI. A syndrome of apparent mineralocorticoid excess associated with defects in the peripheral metabolism of cortisol. J Clin Endocrinol Metab. 1979;49:757-64.

Page 56 of 66

59. Mantero F, Palermo M, Petrelli MD, Tedde R, Stewart PM, Shackelton CHL. Apparent mineralocotioid excess: Type I and Type II. Steroids. 1996;61:193-96.

60. Ulick S, Tedde R, Mantero F. Pathogenesis of the type 2 variant of the syndrome of apparent mineralocorticoid excess. J Clin Endocrinol Metab. 1990;70:200-6.

61. Shackleton CHL. Mass spectrometry in the diagnosis of steroid-related disorders and in hypertension research. J Steroid Biochem Mol Biol. 1993;45:127-40.

62. Chu MD and Ulick S. Isolation and identification of 18-hydroxycortisol from the urine of patients with primary aldosteronism. J Biol Chem. 1982;257:2218-24.

63. Gomez-Sanchez CE, Gill JR Jr, Ganguly A, Gordon RD. Glucocorticoid-suppressible aldosteronism: a disorder of the adrenal transitional zone. J Clin Endocrinol Metab;1988:67:444-8.

64. Vassiliadi DA, Barber TM, Hughes BA, McCarthy MI, Wass JA, Franks S, Nightingale P, Tomlinson JW, Arlt W, Stewart PM. Increased 5 alpha-reductase activity and adrenocortical drive in women with polycystic ovary syndrome. J Clin Endocrinol Metab. 2009;94:3558-66.

Page 57 of 66

65. Stewart PM, Boulton A, Kumar S, Clark PM, Shackleton CHL. Cortisol metabolism in human obesity: impaired cortisonecortisol conversion in subjects with central adiposity. J Clin Endocrinol Metab;1999:1022-7.

66. Kamrath C, Hartmann MF, Wudy SA. The balance of cortisol-cortisone interconversion is shifted towards cortisol in neonates with congenital adrenal hyperplasia due to 21-hydroxylase deficiency. J Steroid Biochem Mol Biol. 2014;143:386-91.

67. Raff H, Auchus RJ, Findling JW, Nieman LK. Urine free cortisol in the diagnosis of Cushing’s syndrome: is it worth doing and, if so, how? J Clin Endocrinol Metab. 2015;100:395-7.

68. Palermo M, Shackleton CHL, Mantero F, Stewart PM. Urinary free cortisone and the assessment of 11-hydroxysteroid dehydrogenase activity in man. Clin Endocrinol (Oxf). 1996;45:605-11.

69. Yamashita K, Nakagawa R, Okuyama M, Honma S, Takahashi M, Numazawa M. Simultaneous determination of tetrahydrocortisol, allotetrahydocortisol and tetrahydrocortisone in human urine by liquid chromatography-electrospray ionisation tandem mass spectrometry. Steroids. 2008:73;727-37.

Page 58 of 66

70. Campino C, Carvajal CA, Cornejo J, San Martin B, Olivieri O, Guidi G, Faccini G, Pasini F, Sateler J, Baudrand R, Mosso L, Owen GI, Kalergis AM, Padilla O, Fardella CE. 11-hydroxysteroid dehydrogenase type-2 and type-1 (11- HSD2 and 11–HSD1) and 5-reductase activities in the pathogenesis of essential hypertension. Endocrine. 2010;37:106-14.

71. Marcos J, Renau N, Casals G, Segura J, Ventura R, Pozo OJ. Investigation of endogenous corticosteroids profiles in human urine based on liquid chromatography tandem mass spectrometry. Anal Chim Acta. 2014;812:92-104.

72. Vining RF, McGinley RA, Symons RG. Hormones in saliva: mode of entry and consequent implications for clinical interpretation. Clin Chem. 1983;29:1752-6.

73. Vining RF, McGinley RA, Maksvytis JJ, Ho KY. Salivary cortisol: a better measure of adrenal cortical function than serum cortisol. Ann Clin Biochem. 1983;20:329-35.

74. Raff H, Homar PJ, Burns EA. Comparison of two methods for measuring salivary cortisol. Clin Chem. 2002;48:207-208.

Page 59 of 66

75. Miller R, Plessow F, Rauh M, Gröschl M, Kirschbaum C. Comparison of salivary cortisol as measured by different immunoassays and tandem mass spectrometry. Psychoneuroendocrinology. 2013;38:50-7. doi: 10.1016/j.psyneuen.2012.04.019. Epub 2012 May 27.

76. Vieira JGH, Nakamura OH, Carvalho VM. Determination of cortisol and cortisone in human saliva by a liquid chromatography-tandem mass spectrometry method. Arq Bras Endocrinol Metab. 2014;58:844-50.

77. Matsui F, Koh E, Yamamoto K, Sugimoto K, Sin H-S, Maeda Y, Honma S, Namiki M. Liquid chromatography-tandem mass spectrometry (LCMS/MS) assay for simultaneous measurement of salivary testosterone and cortisol in healthy men for utilisation in the diagnosis of late-onset hypogonadism in males. Endocr J. 2009;56:1083-93.

78. Turpeinen U, Välimäki MJ, Hämäläinen E. Determination of salivary cortisol by liquid chromatography-tandem mass spectrometry. Scand J Clin Lab Invest. 2009;69:592-7. doi: 10.1080/00365510902890331.

79. Jönsson BAG, Malmberg B, Amilon Å, Garde AH, Ørbæk P. Determination of cortisol in human saliva using liquid chromatographytandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2003;784:63-8.

Page 60 of 66

80. Kataoka H, Matsuura E, Mitani K. Determination of cortisol in human saliva by automated in-tube solid-phase microextraction coupled with liquid chromatography-mass spectrometry. J Pharm Biomed Anal. 2007;44:160-5

81. Perogamvros I, Owen LJ, Newell-Price J, Ray DW, Trainer PJ, Keevil BG. Simultaneous measurement of cortisol and cortisone in human saliva using liquid chromatography-tandem mass spectrometry: Application in basal and stimulated conditions. J Chromatogr B Analyt Technol Biomed Life Sci. 2009;877:3771-5. doi: 10.1016/j.jchromb.2009.09.014

82. Owen LJ, Haslam S, Adaway JE, Wood P, Glenn C, Keevil BG. A simplified liquid chromatography tandem mass spectrometry assay, using on-line solid phase extraction, for the quantitation of cortisol in saliva and comparison with a routine DELFIA method. Clin Ann Biochem. 2010;47:131-36. doi: 10.1258/acb.2009.009053

83. Jones RL, Owen LJ, Adaway JE, Keevil BG. Simultaneous analysis of cortisol and cortisone in saliva using XLC-MS/MS for fully automated online solid phase extraction. J Chromatogr B Analyt Technol Biomed Life Sci. 2012;881-882:42-8.

Page 61 of 66

84. Fustinoni S, Polledri E, Mercadente R. High-throughout determination of cortisol, cortisone and melantonin in oral fluid by on-line turbulent flow liquid chromatography interfaced with liquid chromatography/tandem mass spectrometry. Rapid Commun Mass Spectrom. 2013;27:1450-1460.

85. Perogamvros I, Keevil BG, Ray DW, Trainer PJ. Salivary cortisone is a potential biomarker for serum free cortisol. J Clin Endocrinol Metab; 2010;95:4951-8.

86. Aardal-Eriksson E, Karlberg BE, Holm AC. Salivary cortisol – an alternative to serum cortisol determinations in dynamic function tests. Clin Chem Lab Med. 1998;36:215-22.

87. Cornes MP, Ashby HL, Khalid Y, Buch HN, Ford C, Gama R. Salivary cortisol and cortisone responses to tetracosactrin (synacthen). Ann Clin Biochem. 2015;52:606-10. doi: 10.1177/0004563215577838.

88. Montskó G, Tarjá nyi Z, Mez ősi E, Kovács GL. A validated method for measurement of serum total, serum free, and salivary cortisol, using high-performance liquid chromatography coupled with high-resolution ESI-TOF mass spectrometry. Anal Bioanal Chem. 2014;406:2333-41.

Page 62 of 66

89. Kirchhoff F, Briegel J, Vogeser M. Quantification of free serum cortisol based on equilibrium dialysis and isotope dilution-liquid chromatography-tandem mass spectrometry. Clin Biochem. 2011;44:894-9.

90. Wester VL, van Rossum EF. Clinical applications of cortisol measurement in hair. Eur J Endocrinol. 2015;173:M1-10.

91. Staufenbiel SM, Pennix BW, de Rijke YB, van den Akker EL, van Rossum EF. Determinants of hair cortisol and hair cortisone concentrations in adults. Psychoneuroendocrinology. 2015;60:182-94.

92. Stalder T, Kirschbaum C, Alexander N, Bornstein SR, Gao W, Miller R, Stark S, Bosch JA, Fischer JE. Cortisol in hair and the metabolic syndrome. J Clin Endocrinol Metab. 2013;98:2573-80.

93. Noppe G, de Rijke YB, Dorst K, van den Akker EL, van Rossum EF. LC-MS/MS-based method for long-term steroid profiling in human scalp hair. Clin Endocrinol (Oxf). 2015;83:162-6.

94. Russell E, Koren G, Rieder M, Van Uum SH. The detection of cortisol in human sweat: implications for measurement of cortisol in hair. Ther Drug Monit. 2014;36:30-4.

Page 63 of 66

95. Hoffman MC, Karban LV, Benitez P, Goodteacher A, Laudenslager ML. Chemical processing and shampooing impact cortisol measured in human hair. Clin Invest Med. 2014;37:E252-7.

96. Vesper HW, Bhasin S, Wang C, Tai SS, Dodge LA, Singh RJ, Nelson J, Ohorodnik S, Clarke NJ, Salameh WA, Parker CR Jr, Razdan R, Monsell EA, Myers GL. Interlaboraotry comparison study of serum total testosterone [corrected] measurements performed by mass spectrometry methods. Steroids. 2009;74:498-503.

97. Grebe SK, Singh RJ. LC-MS/MS in the clinical laboratory – where to from here? Clin Biochemist Rev. 2011;32:5-31.

98. Büttler RM, Martens F, Fanelli F, Pham HT, Kushnir MM, Janssen MJW, Owen LJ, Taylor AE, Soeborg T, Blankenstein MA, Heijboer AC. Comparison of 7 published LC-MS/MS methods for the simultaneous measurement of testosterone, androstendione and dehydroepiandrosterone in serum. Clin Chem. 2015;61:1475-1483.

99. Owen LJ, Keevil BG. Testosterone measurement by liquid chromatography tandem mass spectrometry: the importance of internal standard choice. Ann Clin Biochem. 2012;49:600-2.

Page 64 of 66

100.

Guidance for industry. Bioanalytical method validation. US

Department of Health and Human Services. Food and Drug Administration, Center for Drug Evaluation and Research (CDER) and Centre for Vetinary Medicine (CVM). May 2001.

101.

Duxbury K, Owen L, Gillingwater S, Keevil BG. Naturally

occurring isotopes of an analyte can interfere with doubly deuterated internal standard measurement. Ann Clin Biochem. 2008;45:210-2.

102.

Davison AS, Milan AM, Dutton JJ. Potential problems with using

deuterated internal standards for liquid chromatography-tandem mass spectrometry. Ann Clin Biochem. 2013;50:274.

103.

Van Eeckhaut A, Lanckmans K, Sarre S, Smolders I, Michott Y.

Validation of bioanalytical LC–MS/MS assays: evaluation of matrix effects. J Chromatogr B Analyt Technol Biomed Life Sci;2009:877:2198-207.

104.

Xia YG, Jemal M. Phospholipids in liquid chromatography/mass

spectrometry bioanalysis: comparison of three tandem mass spectrometric techniques for monitoring plasma phospholipids, the effect of mobile phase composition on phospholipids elution and the association of phospholipids with matrix effects. Rapid Commun Mass Spectrom. 2009;23:2125-38.

Page 65 of 66

105.

Honour JW. Development and validation of a quantitative assay

based on tandem mass spectrometry. Ann Clin Biochem. 2011;48:97111. ` 106.

CLSI. C-62-A. Liquid chromatography-mass spectrometry

methods; approved guideline. Wayne (PA): CLSI; 2014.

Page 66 of 66

Figure

Figure 2

A

B

Figure 3

363 > 97 (cortisol quantifier)

A

363 > 121 (cortisol qualifier)

B

363 > 121 (interference)

361 > 147 (prednisolone)

C

367 > 97 (cortisol-d4)

D

Time (mins)

Figure 4

11bHSD2

11bHSD1

Cortisol

Cortisone 5b-reductase 3a-HSD

Allo-tetrahydrocortisol (5aTHF)

Tetrahydrocortisol (5bTHF)

Tetrahydrocortisone (THE)

Figure 5

363 > 121 (cortisol quantifier)

A

367 > 121 (cortisol-d4)

B

361 > 163 (cortisone quantifier)

C

369 > 169 (cortisone-d7)

D Time (mins)