Estimating ultrasound propagation velocity in tissues from unwrapped phase spectra

Estimating ultrasound propagation velocity in tissues from unwrapped phase spectra

F?ASONIC IMAGING 2, 223-231 (1980) ESTIMATING ULTRASOUND PROPAGATION VELOCITY IN TISSUES V. N. Gupta, FROM UNWRAPPED PHASE SPECTRA P. K. Bhagat,...

497KB Sizes 0 Downloads 35 Views

F?ASONIC

IMAGING

2,

223-231

(1980)

ESTIMATING ULTRASOUND PROPAGATION VELOCITY IN TISSUES V. N. Gupta,

FROM UNWRAPPED PHASE SPECTRA P. K. Bhagat,

and A. M. Fried'

Wenner-Gren

Biomedical Research Laboratory University of Kentucky Lexington, Kentucky 40506

A digital computer-based technique for estimating the velocity of sound in tissue is proposed. The technique utilizes the unwrapped phase spectra of fullband reference and attenuated ultrasonic pulses normally used in In addition, it does not require explicit meaattenuation measurements. surement of pulse-transit-times. The accuracy of the technique was tested on several computer simulated models, on specimens of known velocities (aluminum and plexiglass), and on formalin-fixed rabbit kidneys. The results indicate that the technique has an accuracy comparable to the pulse-transit-time method. Key words:

1.

Phase

spectra;

ultrasound;

velocity.

INTRODUCTION

Systematic study of the interaction of biological systems with applied energy provides insight into the organization and structure of the tissue Ultrasound at low intensity levels has been used to visualize medium. the tissue interfaces. The attenuation coefficient and propagation velocity are the two acoustic parameters most commonly used in quantitative tissue characterization. The attenuation coefficient, for a given tissue specimen, is generally estimated by computation of the difference between the log-magnitude spectra of the reference and that of the tissue attenuated signals [l-4]. Velocity, on the other hand, can be computed using one of the several available techniques [5]. Notable among these are the interferometric, pulse-transit-time, and the sing-around techniques. The interferometric technique is limited mostly to liquids and quasi-liquids. It is based on the principle that when the distance between the transducer and the reflector, within the medium under investigation, is an integral multiple of half-wavelengths, the detecting system indicates a corresponding minimum in the driving voltage [6]. In the pulse-transit-time method, two distinct variations are available. In the first, the pulse is transmitted by one transducer and detected by another, the axes of the two transducers being coincident. In the second, only one transducer is used, alternately, in the transmitand receive-mode, with a reflector positioned normal to the transducer axis. The pulsetransit-times together with specimen thickness and velocity in the coupling medium is used to compute the velocity in tissue [7]. A.M. Fried is with Department of Diagnostic Radiology, Kentucky Medical Center, Lexington, KY 40536.

University

of

0161-7346/80/030223-09$02.00/O

223

Copyright :U 1980 b-vAcademic Press, Inc. All righrs of reproduction in any form reserved.

GUPTA, BHAGAT, AND FRIED

The sing-around technique is a variant of the pulse-transit-time Instead of measuring the pulse-transit-time across a known technique. length of the tissue specimen, the number of times the pulse travels back and forth through the tissue specimen is counted over a known period of time [8]. In all the techniques discussed above, one has to conduct separate measurements for velocity estimation. In the present study, we propose a digital computer-based technique for measurement of ultrasonic propagation velocity which does not require explicit measurement of transit times. Instead, propagation velocity is estimated from the phase spectra of reThe underlying hypothesis implicit ference and tissue-attenuated signals. in this technique is the premise that for nondispersive media the phase of a traveling wave is proportional to the frequency for a given propagation velocity. Measurements in our laboratory [9] and elsewhere [lo] indicate that biological media (soft tissue) exhibit negligible dispersion over the frequency range l-10 MHz.

2.

THEORETICAL

DETAILS

When an ultrasonic beam propagates through a weakly scattering, dissipative medium, it suffers a loss of energy due to attenuation. The term attenuation is defined here to include losses due to reflection, scattering, refraction and absorption. Consider the pulse-echo system of figure 1, where a tissue specimen positioned between the ultrasonic transducer and a plane plexiglass reflector is immersed in a dissipationless coupling medium. Let the transducer emit a short, bandlimited burst of ultrasonic energy p.'(t) and let pi(t) and p,(t) be the echoes reflected from the reflector wlthout and with the tissue specfmen in position. The relationships between pi'(t), pi(t) and p,(t) can be expressed as Pi(t)

= ' p Im Pi'(w) -co

Pr(t)

= ap

e-j2wao'vo

ejwt

dw,

and

jm p l(w) e-2a(~)ae-j2[(~o-~)/vot~/v] i -co

,jut

dw,

where

Pi’(w)

=k

joD pi'(t)

e-jut

dt,

-m = angular frequency in radian/s, II = distance between the transducer and the Lo = tissue-specimen thickness, = reflection coefficient for the plexiglass aP = sound velocity in the coupling medium, ;o = sound velocity in tissue, and a(w) = frequency dependent attenuation. w

receiver, reflector,

In deriving Eqs. (1) and (Z), we have implicitly assumed that the coupling medium and tissue specimen are non-dispersive, i.e., the group velocity is the same as the phase velocity. This is based on results published from this laboratory [g] and elsewhere [lo], where maximum dis-

224

UNWRAPPED

PHASE

SPECTRA

TISSUE

ULTRASONIC

Fig.

1.

A schematic

representation

of the

persion was observed to be less than 2 percent over the frequency range 1 to 10 MHZ. Further, the reflection losses at the assumed to be negligible since the acoustic water are approximately equal. Fourier

Equations (1) transforms Pi(w)

and (2) as

= ap Pi'(w) = 1 Pi(w)1

pulse-echo

for

several

tissue-water impedances

may be rewritten

in terms

system. tissue

specimens,

interfaces are of the tissue and of their

respective

e-j2wao'vo (4)

e-jeo(@)

and P,(u)

= aP Pi'(w) = 1 P,(w)1

e -2aa(w)

,-j2u[(k,-a)lv,+alv]

e-jel(,).

(5)

Defining sO

= 2i?o/vo

(61

= 2mo-a)/v,+a/v1,

(7)

and s1 it

can be shown that

a(w)

=

[ln 1 Pi(w)]

-

ln I Prb)l

1 l(2a)

(8)

and V

= [(so-S, f/T,,+11 vos

(9)

where T = 22/v is defined to be the duration of the backscattered signal The use d?iginating from the entire depth of the tissue specimen. of Eq. (8) for attenuation measurements has been reported earlier [l]. Equation (9) provides velocity estimates using the phase components of the Fourier transforms of pi(t) and p,(t), namely, e,(w) and e,(w).

225

GUPTA, BHAGAT, AND FRIED

3.

IMPLEMENTATION

OF VELOCITY ESTIMATIONALGORITHM

As defined in the previous section, the pulse propagation velocity can be estimated from the phase components of the measured reference and attenuated spectra. However, the computations of the propagation-lags s and s are compounded by the fact that:(i) the phase spectra e (w) and e:(w), ai estimated on a digital computer, are discontinuous funceions; (this is due to the fact that arc-tangent routines in a computer provide both the reonly the principal values lying between -IT and 71); and (ii) ference and the subsequently attenuated echo signals are bandlimited, being :0p~5et~~lo:~:o~~eo~~~~~e~~~ w 5 w*.and ;ero elsewhere. The presence of he band idth w2-w,) affects the computations of so and s,. Elimination of (i) is (ii) is known as conversion for which algorithms exist of a band-limited signal to frequency-scaling operation w -w1) i--w2-w,)'

w=lT as proposed

known as phase unwrapping and elimination of of a bandlimited signal to fullband signal in the literature [ll, 123. The conversion a fullband signal can be achieved through a

O
by Tribolet

(10)

[12].

Defining w2

=n-

'2 D2

(11)

and w, = (1 - ;)

w2,

(12)

where I I D and D are integers, Tribolet has proposed that Eq. (10) can be I $pl$ien!.ed usizg digital interpolation/decimation algorithms Pa. With this implementation, the digitized reference and attenuated pulses, p.(n) and p (n) (corresponding to p.(t) and p (t) respectively), can be canverted t)S fullband discrete signals pi(n) bnd p,(n) respectively. Note that s and s in Eq. (9) represent the propagation-lags corresponding to the yinear bhase components of the bandlimited pulses p.(n) and p (n) respectively. Conversion of these pulses to fullband in&oducesra constant K=(D D )/(I I ) in the estimation of s and s If we define n and n as the2inte4eF time-lags (as computed Prom thd'unwrapped phase eseimates 1 for the fullband pulses pi(n) and p,(n) respectively, then so and s, can be expressed as sO

= Kn,T and s,

= Kn,T

,

(13)

where T = sampling interval. The origin of K can be explained as follows. The implementation of Eq. (10) is achieved through four elementary mappings [12] two of which perform decimation by the ratio I /D and I /D and the other two modulation (frequency shifts by IT). The goniequenc & o# the two stages of decimation is that the propagation lags s 'and s are reduced by the reciprocal of which is tile cons i ant K of Eq. T,;Tctor of I, 12/D,D2,

226

UNWRAPPEDPHASE SPECTRA

By substituting as

puted

Eq.

v = [&p

SC 4.

MATERIALS

(13)

in Eq.

(9),

velocity

in tissue

can be com-

+ 11 vO*

(14)

AND METHODS

A detailed description of the experimental apparatus and procedure Briefly, the samples are mounted on a plexihas appeared elsewhere [l]. glass holder and placed in the far field of an ultrasonic transducer, resonant at 5 MHz, with 80 percent bandwidth (3-7 MHz). A plane plexiglass reflector is used to reflect the ultrasonic pulses. The transducer, specimen holder and plane reflector are housed in a plexiglass tank filled with Hank's Balanced Salt Solution which provides dissipationless coupling between transducer, specimen and the reflector. An MP 203 pulser (Metrotek, Inc.) and an UTA-3 (Aerotech Labs.) transducer analyzer were used to excite the transducer and gate and capture the The echoes were digitized at 50 MHz by reference and attenuated echoes. The onset of the digitization means of a Biomation 8100 transient recorder. process was controlled by the 'trigger-sync' signal from the UTA-3. The digitized sweeps of 2048 points were stored on a floppy diskette attached to the Altair 8800 microcomputer and subsequently transferred to a remote PDP 11/34 minicomputer over a serial RS 232 interface. The entire processing was accomplished on a PDP 11/34 minicomputer. chosen

The cut-off to be

frequencies

w1 and w2 of the

transducer

bandwidth

were

which, given a sampling interval of 20 ns, correspond to a passband from 2.85714 MHz to 7.14285 MHz. Thus the cut-off frequencies chosen are approximately equal to those prescribed by the manufacturer (3 and 7 MHz). The discrete bandlimited signals p.(n) and p (n) were converted to fullband signals using the interpolatioa/decimati& algorithms [13]. A 2048-point FFT routine in conjunction with the phase unwrapping algorithm [11] was used to compute the linear, integer phase shifts no and n . The of the backscattered signal was measured from an osci -1 loduration T scope dispfsy of the entire backscattered sweep. Finally, Eq. (14) was used to compute the velocity of sound in the specimen. The system calibration was performed by measuring the velocity of sound in plexiglass and aluminum and subsequently in fixed rabbit kidneys, Velocities were also computed from a knowledge of transit-times [7]. 5.

RESULTS AND DISCUSSION

The accuracy of the proposed technique was first tested on several computer simulated models. An experimentally recorded pulse (Fig. 2A) was used as the reference echo which was time-displaced by known amounts (Figs. 2B-E) to form the hypothetical attenuated echoes. The developed

227

GUPTA, BHAGAT, AND FRIED

E.

, 0

TIME

Fig.

2.

(nT,T=20ns)

Simulated

pulse

600 separations.

technique was tested using computer generated data to measure known separations. Table 1 shows simulated and computed separations. The results indicate that separations as low as 0.12 ~J.S (six samples at a sampling If the signals were fullband to begin interval of 20 ns) can be detected. with, then the technique could, theoretically, detect a separation of one Finite transducer bandwidth implied by the presence of the consample. stant K (discussed earlier) is the limiting factor in the measurement of pulse separations. The minimum separation (number of samples) detectable is the smallest integer equal to or greater than K. Thus, for the 5 MHz transducer (3-7 MHz bandwidth) used in the study and the sampling frequency of 50 MHz, K was determined to be 35/6. Therefore, the minimum detectable separation is six samples. This can be reduced by lowering the sampling frequency and thereby lowering the constant K. However, lowering the sampling frequency results in increased aliasing; therefore, greater precision in velocity can only be attained at the cost of increased aliasing. Measurements were next made using aluminum and plexiglass. For comparative purposes, velocity data were also obtained using the pulse-transit Since the velocities for HBSS, plexiglass and aluminum time technique. are appreciably different, one would expect the unwrapped phase estimates

228

UNWRAPPED

Table

I.

Time lag

PHASE

SPECTRA

measurements--simulated

Simulated Time Lag (PSI 1.

2. 3. 4.

and measured. Measured Time Lag (us)

.120 .500 2.100 6.840

.117 .467 : 2.100 6.767

for the reference and attenuated echoes from aluminum and have different slopes. That this is indeed the case, can figure 3 which shows the unwrapped phase estimates for the In addition, the linearity aluminum and plexiglass echoes. estimates also indicates the nondispersive nature of HBSS, plexiglass.

plexiglass to be seen from reference, of the phasealuminum and

Finally, the technique was used to measure the sound velocity in six Table II summarizes the results obtained formalin-fixed rabbit kidneys. in this study as well as the published velocity data [5, 141. The results indicate that the proposed technique is at least as accurate as the pulseThe minor differences in the reported and published transit-time method. results can be attributed to different experimental conditions that may have been used in making the reported measurements. The significance eliminates the need thereby eliminating duce human bias in technique utilizes estimates. Further,

of the proposed technique is twofold: first, it for separate transit-time or distance measurements, the need for setting trigger thresholds which introthe conventional velocity measurements; second, the the measurements normally conducted for attenuation the term (2~) in Eq. (8) can be replaced by v*T,, to

FREQUENCY

Fig.

3.

Unwrapped

229

(MHZ)

phase

estimates.

GUPTA, BHAGAT, AND FRIED

Table

Specimen

II.

Velocity

Pulse-Transit Method (m/s)

Aluminum[5] Plexiglass[5] Rabbit kidneys[l4] avelocity reported name for the grade is not known.

data.

Phase-Estimate Method (m/s)

Published Values (m/s)

6397 2746 1545

6395 2746 1543 in [5] of the

6400 2680a 1566

is for polymethylmethacrylate; plexiglass specimen used

in the

the chemical present study

enhance the resolution of attenuation estimates. An interesting application of the technique might be phase tomography along the lines of attenuation tomography as proposed by Kak and Dines [15]. We are currently exploring this possibility. 6.

CONCLUSIONS

In summary, a new technique for estimating velocity of sound in tissue has been proposed. The technique utilizes the phase spectra of the reference and attenuated pulses normally used in estimating attenuation. The technique is simple to use2 and is, at least, as accurate as the conventional pulse-transit-time method. ACKNOWLEDGMENTS The authors acknowledge the technical assistance provided by Mr. V. C. Wu and Mr. M. Shafer and also wish to thank Dr. C. F. Knapp for the use of his minicomputer facilities. This study was supported NASA Contract NAS9-15452.

in part

by an NIH-BRSG Grant

RR05374 and

REFERENCES Cl]

Kadaba, M.P., Bhagat, tering of ultrasound Biomed. Eng. BME-27,

P.K., and Wu, V.C., Attenuation in freshly excised animal tissues, 76-83 (1980).

and backscatIEEE Trans.

[21

Kak, A.C. and Dines, K.A., Signal processing of broadband ultrasound: measurement of attenuation of soft biological IEEE Trans. Biomed. Eng. BME-25, 321-344 (1978).

[3]

Mimbs, J.W., Yukas, D.E., Miller, Detection of myocardial infarction ation of ultrasound, Circulation

pulsed tissues,

J.G., Weiss, R.N., and Sobel, B.E., in vitro based on altered attenuResearch 5, 192-198 (1977).

2

The Fortran listings can be obtained from

of the computer the first author.

230

programs

for

a PDP 11/34

minicomputer

UNWRAPPEDPHASE SPECTRA

[4]

M., Estimating the acoustic KUC, R., and Schwartz, ficient slope for liver from reflected ultrasound Sonics Ultrasonics SU-26, 353-362 (1979).

[5]

W;,

[6]

Fry, Sot.

[71

Sollish, B. D., A Device for Measuring Ultrasonic in Tissue, in Ultrasonic Tissue Characterization National Bureau of Standards Soec. Publ. 525, Printing Office, Washington, DC, 1979).

[8]

Biomedical

P.N.T., W.J., Amer.

The double 11, 17-28

Ultrasonics,

crystal (1949).

acoustic

Ficken, G.W., Jr. and Hiedemann, around' method for determination (1956). --Amer. 28, 921-923 Bhagat, P.K., Kadaba, M.P., based system for ultrasonic mentation (In Press 1980).

[lo]

Wells, P.N.T., Review: absorption biological tissue, Ultrasound Med.

[ll]

Tribolet, Speech,

J.M.,

Press,

New York,

Gupta, tissue

NY,

J. Acoust.

interferometer,

Propagation Velocity M. Linzer, ed., 53-56, (U.S. Government

II, PP.

E.A., Simple form of the 'singof sound velocities, J. Acoust.

[9]

Signal

(Academic

attenuation coefsignals, IEEE Trans.

Sot.

V.N., and Wu, W.C., Microprocessorcharacterization, Medical Instruand dispersion of ultrasound Biol. 1, 369-376 (1975).

A new phase unwrapping algorithm, IEEE Trans. Processing, ASSP-25, 170-177 (1977). Seismic Applications of Homomorphic Inc., Englewood Cliffs, NJ, 1979).

Signal

in Acoust.

Processing,

[12]

Tribolet, J.M., (Prentice-Hall,

[13]

Crochiere, R.E., A General Program to Perform Sampling Rate Conversion of Data by Rational Ratios, in Programs for Digital Signal Processing, edited by Digital Signal Processing Committee IEEE Acoustics, Speech, and Signal Processing Society (IEEE Press, pp. 8.2-l-8.2-7, 1979).

[14]

Goss, S.A., Johnston, of empirical ultrasonic Sot. Amer. 64, 423-457

[15]

Dines, K.A., soft tissues,

R.L., and Dunn, F., Comprehensive properties of mammalian tissues, (1978).

and Kak, A.C., Ultrasonic attenuation Ultrasonic Imaginql, 16-33 (1979).

231

compilation

J..Acoust.

tomography

of