High resolution gamma-ray spectroscopy at high count rates with a prototype High Purity Germanium detector

High resolution gamma-ray spectroscopy at high count rates with a prototype High Purity Germanium detector

Accepted Manuscript High resolution gamma-ray spectroscopy at high count rates with a prototype High Purity Germanium detector R.J. Cooper, M. Amman, ...

1MB Sizes 0 Downloads 56 Views

Accepted Manuscript High resolution gamma-ray spectroscopy at high count rates with a prototype High Purity Germanium detector R.J. Cooper, M. Amman, K. Vetter

PII: DOI: Reference:

S0168-9002(17)31459-6 https://doi.org/10.1016/j.nima.2017.12.053 NIMA 60394

To appear in:

Nuclear Inst. and Methods in Physics Research, A

Received date : 3 April 2017 Revised date : 8 December 2017 Accepted date : 15 December 2017 Please cite this article as: R.J. Cooper, M. Amman, K. Vetter, High resolution gamma-ray spectroscopy at high count rates with a prototype High Purity Germanium detector, Nuclear Inst. and Methods in Physics Research, A (2017), https://doi.org/10.1016/j.nima.2017.12.053 This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

*Manuscript Click here to view linked References

1

High Resolution Gamma-Ray Spectroscopy at High Count Rates with a Prototype

2

High Purity Germanium Detector

3 R.J. Coopera*, M. Ammana, K. Vettera,b

4 5 6 7

a

Lawrence Berkeley National Laboratory, 1 Cyclotron Road, Berkeley, CA 94720, USA b

Department of Nuclear Engineering, University of California, Berkeley, CA 94720,

8

USA

9 10

Abstract

11

High-resolution gamma-ray spectrometers are required for applications in nuclear

12

safeguards, emergency response, and fundamental nuclear physics. To overcome one of

13

the shortcomings of conventional High Purity Germanium (HPGe) detectors, we have

14

developed a prototype device capable of achieving high event throughput and high

15

energy resolution at very high count rates. This device, the design of which we have

16

previously reported on, features a planar HPGe crystal with a reduced-capacitance strip

17

electrode geometry. This design is intended to provide good energy resolution at the

18

short shaping or digital filter times that are required for high rate operation and which are

19

enabled by the fast charge collection afforded by the planar geometry crystal. In this

20

work, we report on the initial performance of the system at count rates up to and

21

including two million counts per second.

22

23

Keywords: Gamma-ray detectors, gamma-ray spectroscopy, high-purity germanium, high

24

count rate.

25 26

Corresponding author: Tel.: +1 510 486 7296

27

Email address: [email protected] (R.J. Cooper)

28 29 30

1.0 Introduction

31

While High Purity Germanium (HPGe) detectors are typically limited to use in count rate

32

regimes up to around 10 kcps, a number of applications exist where high count rate

33

operation is desirable. These applications range from basic nuclear physics [1] to the

34

non-destructive assay of spent nuclear fuel [2]. For such applications, the ability to

35

maintain both fine energy resolution and high event throughput at count rates exceeding 1

36

mega counts per second (Mcps) would allow the analysis of relatively weak features in

37

complex gamma-ray spectra to be performed with increased sensitivity. Unfortunately,

38

however, conventional HPGe spectrometers typically sacrifice energy resolution for

39

higher throughput by employing short pulse shaping times in order to reduce deadtime.

40

This fundamental trade-off defines the challenge of achieving both high throughput and

41

fine energy resolution at high count rates.

42 43

In Ref. [3], results from the operation of a coaxial HPGe detector adapted for high rate

44

performance were reported. The system was able to achieve 8 keV FWHM at 662 keV

45

and 39% throughput when operating at an incident count rate of 1.03 Mcps. To our

46

knowledge, this represents the best reported result for high resolution gamma-ray

47

spectroscopy at high count rate. In Ref. [4], a coaxial HPGe detector was operated using

48

specially developed electronics and demonstrated less than 2.4 keV FWHM at 662 keV at

49

an incident rate 1 Mcps. However, the associated dead time at this count rate was 98%.

50 51

To overcome these resolution and throughput limitations, we have developed a novel

52

HPGe detector concept and have demonstrated its performance at millions of counts per

53

second. In Ref. [5], the design and basic electrical and spectroscopic performance of the

54

first prototype device were presented.

55

performance of a second, slightly modified prototype.

In this work, we report on the high-rate

56 57 58

2.0 Prototype Detector

59

The prototype detector discussed in Ref. [5] is based on a 100 mm diameter x 16 mm

60

thick planar, p-type HPGe crystal and features a single-sided strip electrode configuration

61

of ten 71 mm long, 2.85 mm wide charge collecting strips, each surrounded by 0.95 mm

62

wide steering grid strips. The collecting strips are separated by 4.75 mm and the steering

63

grid is employed in order to maintain good charge collection efficiency in the regions

64

between collecting strips. All contacts were produced using amorphous germanium [6].

65

The device we report on here features the same basic design but with the full-area contact

66

(i.e. the contact on the opposite face to the strips) produced using lithium diffusion. This

67

modification was made in order to improve the leakage current stability of the detector at

68

high count rates. This modification was made after a significant increase in leakage

69

current was observed when two prototype detectors that featured full-area contacts

70

fabricated using amorphous germanium were operated under high rate irradiation. A

71

series of small test devices were fabricated in order to investigate this effect but the

72

behavior was not consistently reproducible across all devices. The underlying cause of

73

the increased leakage current is not currently well understood. However, as it has not

74

been observed in any device in which the full-area contact is fabricated using lithium

75

diffusion, it is possible that it is related to barrier lowering at the amorphous germanium

76

contact as a result of charge build up during high-rate irradiation.

77 78

The detector is operated with a positive bias of 1000 V applied to the full-area contact,

79

the collecting strips connected to virtual ground through charge sensitive preamplifiers,

80

and a positive bias of 200 V applied to the steering grid. Figure 1 shows a photograph of

81

the detector, taken with the segmented electrode on display, along with a schematic

82

diagram of the planar geometry HPGe crystal and the electrode configuration.

83

Employing a planar geometry allows a strong, uniform electric field to be established

84

throughout the crystal and ensures fast charge collection. This, coupled to the short drift

85

distance between electrodes, results in signal rise times which are short and exhibit

86

minimal variation. Fig. 2 shows the distribution of signal risetimes recorded from a

87

typical strip while the detector was uniformly illuminated with a

88

risetime was calculated on a signal-by-signal basis as the time taken for the signal to rise

89

from 10% to 90% of its maximum value. The distribution in Fig. 2 was generated using

90

five thousand events from within the 662 keV photopeak and shows that the risetime

91

varies from a few tens of nanoseconds to just over 250 ns, consistent with the simulated

137

Cs source. The

92

values presented in Ref. [5]. When using digital trapezoidal filtering, this small variation

93

in rise time allows short gap times (e.g. 350 ns or shorter) to be employed in digital

94

trapezoidal filtering without degrading the energy resolution. This offers a significant

95

advantage over coaxial HPGe geometries for high count rate applications. However,

96

employing a planar geometry crystal does result in reduced detection efficiency relative

97

to larger volume coaxial geometries, particularly at high energy. Employing thinner

98

planar crystals would offer even smaller variations in the rise time and therefore the

99

potential for increased count rate performance. However, they would also result in even

100

greater reductions in efficiency. The relative efficiency [7] of the prototype detector has

101

been measured to be approximately 6% when operated in singles mode (i.e. when each

102

strip is read out individually and the resulting spectra summed). In any real application,

103

however, the detector would be operated in add-back mode [8] where the full energy

104

deposited by Compton scatted gamma rays is recovered by summing the energies

105

associated with simultaneous interactions in multiple strips. Monte Carlo simulations

106

suggest that the relative efficiency would be approximately 20% in this case. The add-

107

back technique may also be applied to multiple detectors. For the spectroscopy of high

108

energy gamma rays it is envisioned that multiple detectors could be employed in a stack

109

configuration to increase the total detection efficiency. This approach only meaningfully

110

increases the detection efficiency in systems where each individual detector is able to

111

maintain low dead time losses (i.e. high event throughput). Given the planar geometry, a

112

single device is optimally suited to the spectroscopy of lower energy gamma rays where

113

the detection efficiency is greater. For example, the intrinsic efficiency at 81 keV, as

114

measured using a 133Ba source, is 81%.

115 116 117

The specific strip electrode geometry was designed to reduce the capacitance associated

118

with each readout electrode relative to conventional HPGe strip detectors while

119

maintaining a relatively modest channel count. This is achieved by employing wide gaps

120

between the collecting strips. At 4.75 mm, the separation between collecting strips in the

121

device is approximately an order of magnitude larger than that of typical, conventional

122

strip detectors. By reducing the inter-strip capacitance, the series noise is reduced and

123

this in turn allows for improved energy resolution at short peaking times. By enabling

124

the use of both short peaking and gap times without significant degradation of the energy

125

resolution, the device is designed to allow high energy resolution and high throughput to

126

be maintained at count rates of hundreds of kcps per strip (i.e. millions of counts per

127

second across the entire detector).

128 129

At an intermediate count rate of approximately 2 kcps per strip, the average energy

130

resolution of the collecting strips is 2.4 keV FWHM at 662 keV, as measured using the

131

digital nanoMCA [9] with a peaking time of 1 s and a gap time of 350 ns applied in

132

online trapezoidal filtering. The average width of a pulser peak is 1.8 keV FWHM.

133

Figure 3 shows an example of a gamma-ray spectrum from a single, central strip (strip 6)

134

recorded in the presence of a

135

the 662 keV peak is 2.3 keV FWHM. In addition to the 662 keV 137Cs photopeak and the

136

peak associated with the pulser, the K and K X-rays associated with 137mBa can also be

137

observed just above the low energy threshold. As shown in the inset, these X-rays, which

137

Cs source and simultaneous pulser input. The width of

138

have characteristic energies of 32 keV and 36 keV, can easily be resolved. The energy

139

resolution of the 36 keV peak is 1.8 keV FWHM.

140 141

Fig. 4 shows the energy resolution of each of the ten collecting strips, as measured with

142

the nanoMCA with a peaking time of 1 s and a gap time of 350 ns. It can be seen that

143

the energy resolution of the edge strips (strips 1 and 10) at 662 keV is slightly degraded

144

relative to that of the others. As the width of the pulser peak remains generally constant

145

across all strips, this suggests that this is a result of reduced charge collection

146

performance rather than increased electronic noise.

147 148

3.0 Performance at High Count Rates

149

The performance of the prototype detector was assessed at a range of count rates using an

150

uncollimated 1.8 mCi

151

each source-to-detector distance, the ten collecting strips were read out using charge

152

sensitive preamplifiers and spectra were acquired by sequentially connecting individual

153

strips to the single-channel nanoMCA. The energy was calculated using the MCA’s

154

onboard trapezoidal filter algorithm with a peaking time of 500 ns and a gap time of 350

155

ns. In each case, the measured count rate was defined as the rate at which counts were

156

recorded to the energy spectrum. By performing fits to the full energy peaks, values of

157

energy resolution were extracted from each spectrum. For each measurement, the per-

158

strip event throughput was then calculated using the average dead time fraction reported

159

by the nanoMCA.

160

correcting the measured count rate according to this reported dead time, effectively

137

Cs source positioned at various distances from the device. At

The incident count rate was calculated on a per-strip basis by

161

dividing the number of recorded counts by the live time of the measurement. The

162

uncertainty associated with the incident count rate was calculated by assuming that the

163

live time is subject to uncertainties in both the determination of the dead fraction and the

164

real time. An uncertainty of 1% in the determination of the dead fraction was assumed

165

and a 10 ms uncertainty in the real time was assigned based on the specifications of the

166

nanoMCA. These uncertainties, along with the statistical uncertainty associated with the

167

number of counts recorded, were then propagated through the calculation of the incident

168

count rate. Similarly, the assumed uncertainty in the dead fraction was propagated in

169

order to assign uncertainties to the throughput. The incident count rates were verified

170

using the results of corresponding Monte Carlo simulations performed using MCNP6

171

[10]. At all source-to-detector distances, good agreement between the simulated and

172

measured count rates was observed. At the shortest source-to-detector distance, for

173

example, the discrepancy between the simulated count rate of strip 6 and that estimated

174

from the data was approximately 2% of the measured value.

175 176

Figure 5 shows the spectrum acquired from strip 6 during a measurement in which the

177

incident single-strip count rate was calculated to be 178.5(+/-1.0) kcps. The continuum

178

of counts above the 662 keV photopeak is a result of piled-up events. The inset image

179

shows the region of the spectrum below 100 keV, where the X-rays from 137mBa can once

180

again be observed. The additional peaks at 59 keV and 69 keV result from fluorescence

181

X-rays emitted from the Tungsten shielding surrounding the

182

amount of symmetric tailing is observed below the Full Width at Tenth Maximum

183

(FWTM) points on both the high and low energy sides of the 622 keV peak. This tailing

137

Cs source. A small

184

is not observed at low and intermediate count rates and the magnitude of the observed

185

tailing increases with count rate. This is not an unexpected result and may be caused by

186

uncertainty in determining the signal baseline when the count rate is large.

187 188

Figure 6 shows how the energy resolution of the 662 keV photopeak varies as a function

189

of the incident count rate for two central strips (strip 5 and 6) as well as the two edge

190

strips (strips 1 and 10). Figure 7 shows how the event throughout varies as a function of

191

incident count rate for the same strips. A linear fit was applied to the energy resolution

192

data and a double exponential fit to the throughput data. The results of these fits are

193

plotted as dashed lines on each figure panel and serve to allow interpolation between the

194

measured data points. The double exponential model was used to fit the throughput as a

195

function of incident count rate as the data is not well described by a single exponential

196

function. This implies that the behavior of the detector and signal processing chain is not

197

well described by a simple, paralyzable dead-time model which assumes a Poisson

198

distribution of pulse arrival times. This may be due, in part, to the contribution from

199

image charge signals and the effect of signals arising from charge sharing interactions but

200

is likely also a result of the triggering architecture employed by the nanoMCA system.

201 202

The results presented in Figs. 6 and 7 show that both the representative central strips and

203

the edge strips perform well across a range of count rates. The fitted curves suggest that

204

strip 5, the performance of which is generally representative of all non-edge strips,

205

maintains an energy resolution of 3.3 keV FWHM and a throughput of 78% when

206

operating at an incident count rate of 100 kcps. At 100 kcps, the throughput of the edge

207

strips is similar to that of the central strips but the energy resolution is slightly worse at

208

3.5 keV FHWM.

209 210

At the maximum count rate studied, the incident count rates associated with the central

211

strips were 211.5(+/-1.3) kcps (strip 5) and 212.5(+/-1.4) kcps (strip 6). At these count

212

rates, the throughput values associated with strips 5 and 6 were 61.4(+/-0.4)% and

213

61.2(+/-0.4)%, respectively. The energy resolution of each strip was 4.1 keV FWHM at

214

662 keV. Due to the small distance between the source and the detector during these

215

measurements, the incident count rate associated with the edge strips was slightly lower

216

than that of the central strips at 170.2(+/-0.8) kcps for strip 1 and 173.0(+/-0.9) kcps for

217

strip 10. The edge strips exhibited throughput values of 67.2(+/-0.3)% (strip 1) and

218

67.0(+/-0.3)% (strip 10). These values are consistent with the throughput of the central

219

strips at the same incident count rates. The energy resolution of each edge strip was 4.0

220

keV FWHM at 662 keV. During these measurements, the sum of the incident count rates

221

of all ten collecting strips was 2.0 Mcps. Figure 8 shows the incident count rate (top),

222

event throughput (middle), and energy resolution (bottom) associated with all collecting

223

strips.

224

throughput is 62.8(+/-0.8)%.

The performance of all ten collecting strips is similar and the total event

225 226 227

4.0 Summary

228

We have demonstrated the performance of a novel, prototype HPGe detector for high-

229

rate, high-resolution gamma-ray spectroscopy. The detector design features a reduced-

230

capacitance, single sided strip electrode configuration with ten collecting strips and a grid

231

of narrow steering electrodes to maintain charge collection performance. The ability of

232

all ten strips to maintain high event throughput and fine energy resolution has been

233

demonstrated during measurements in which the incident count rate across the entire

234

detector was 2.0 Mcps. During these measurements, a single centrally located strip

235

operating at 212.5(+/-1.4) kcps, exhibited event throughput of 61.2(+/-0.4)% and energy

236

resolution of 4.1 keV FWHM at 662 keV.

237 238

This performance represents a significant improvement over previously demonstrated

239

systems for high-resolution gamma-ray spectroscopy at high count rate. The use of a

240

segmented electrode geometry allows the incident count rate to be distributed over

241

multiple channels, the short charge collection times offered by the planar geometry

242

allows the use of short gap times without loss of energy resolution, and the reduced

243

capacitance strip-electrode geometry reduces the degradation of energy resolution at short

244

peaking times. Due to the reduced capacitance of the electrode design, the device also

245

exhibits low electronic noise and good energy resolution at low energy. This may be

246

particularly advantageous for applications such as the non-destructive assay of spent

247

nuclear fuel where spectroscopy of the low energy features associated with Plutonium

248

and Uranium isotopes would allow direct measurements of fissile content to be made.

249 250

At a count rate of 1 Mcps, the small coaxial HPGe system presented in Ref. [3] was able

251

to achieve 8 keV FWHM at 662 keV and 39% throughput. The results presented here

252

represent improvement by a factor of greater than two in both energy resolution and

253

throughput at the same incident count rate. Extrapolations based on the performance of

254

this detector (Figs. 6 and 7), suggest that in its current implementation, an incident count

255

rate of 500 kcps per strip (equivalent to 5 Mcps total) could be reached before the

256

throughput dropped to 39%. At this count rate, the corresponding energy resolution

257

would be approximately 6.2 keV FWHM at 662 keV.

258 259

With the use of front-end electronics and a digital data acquisition system specifically

260

optimized for high rate operation, increases in both throughput and energy resolution may

261

be achieved. For example, reducing the capacitance of the front end electronics could

262

further improve the energy resolution at short peaking times while the application of

263

digital pile-up recovery algorithms could be used to improve the throughput.

264 265 266

Acknowledgements

267

This work was performed under the auspices of the US Department of Energy by

268

Lawrence Berkeley National Laboratory under Contract DE-AC02-05CH11231. The

269

project was funded by the US Department of Energy, National Nuclear Security

270

Administration, Office of Defense Nuclear Nonproliferation Research and Development

271

(DNN R&D).

272 273 274

References

275

[1] B. Loher et al., Nuclear Instruments and Methods in Physics Research A 686 (2012)

276

1-6.

277

[2] A.J. Gilbert et al., Proceedings of IEEE Nuclear Science Symposium and Medical

278

Imaging Conference, 2015.

279

[3] B.A. VanDevender et al., IEEE Transactions on Nuclear Science, 61 5 (2014) 2619-

280

2627.

281

[4] J. Plagnard et al., Applied Radiation and Isotopes 60 (2004) 179-183.

282

[5] R.J. Cooper et al., Nuclear Instruments and Methods in Physics Research A 795

283

(2015) 167-173.

284

[6] M. Amman and P.N. Luke, Nuclear Instruments and Methods in Physics Research A

285

452 (2000) 155.

286

[7] IEEE Std. 325 1996

287

[8] M. Schumaker and C.E. Svensson, Nuclear Instruments and Methods in Physics

288

Research A 575 (2007) 421-432.

289

[9] http://www.labzy.com/index.html

290 291

[10] T. Goorley et al. Annals of Nuclear Energy 87 (2016): 772-783.

292 293 294 295 296 297

298 299

Figure 1.

300

segmented electrode on display, along with a schematic diagram of the planar geometry

301

HPGe crystal (top) and electrode configuration.

302 303 304 305

A photograph of the prototype HPGe detector (bottom), taken with the

700 600

Counts

500 400 300 200 100 0 306

0

100

200

300

Signal Risetime (ns)

307

Figure 2. Distribution of signal risetimes in the prototype detector. The risetime is

308

defined as the time taken for the signal to rise from 10% to 90% of its maximum. The

309

distribution was generated using the signals associated with five thousand events from

310

within the 662 keV photpeak of 137Cs.

311 312 313 314 315 316 317 318 319

320 137

321

Figure 3. A gamma-ray spectrum from a single strip recorded in the presence of a

322

source and simultaneous pulser input at a count rate of around 2 kcps. The width of the

323

137

324

shows the low energy portion of the spectrum where the 32 keV and 36 keV X-rays

325

associated with 137mBa can also be observed.

326 327 328 329 330 331 332 333 334

Cs

Cs and pulser peaks are 2.4 keV FWHM and 1.8 keV FWHM, respectively. The inset

4 Pulser 662 keV Gamma Peak

Peak Width, FWHM (keV)

3.5 3 2.5 2 1.5 1 0.5 0

0

2

335

4 6 8 Strip Number

10

336

Figure 4. The width of the 662 keV peak from 137Cs and an electronic pulser for each of

337

the ten collecting strips.

338 339 340 341 342 343 344

345 346

Figure 5. A gamma-ray spectrum from a single, central strip (strip 6) recorded in the

347

presence of a

348

kcps. The inset shows the 32 keV and 36 keV X-ray peaks associated with

349

the 59 keV and 69 keV peaks that result from fluorescence X-rays emitted from Tungsten

350

shielding surrounding the 137Cs source.

351 352 353 354 355 356 357 358 359

137

Cs source and acquired at an incident single-strip count rate of 178.5 137m

Ba and

360 361

Figure 6. The energy resolution of the 662 keV photopeak as a function of the incident

362

count rate for two central strips and the two edge strips.

363 364 365 366 367 368 369 370

371 372

Figure 7. The event throughput as a function of the incident count rate for two central

373

strips and the two edge strips.

374 375 376 377 378 379 380 381

382 383

Figure 8. The incident count rate (top), event throughput (middle), and energy resolution

384

at 662 keV (bottom) of all collecting strips at the highest total count rate studied.