How to perform noncontact mapping

How to perform noncontact mapping

HANDS ON How to perform noncontact mapping Larry A. Chinitz, MD, Jesse S. Sethi, MD From the New York University Medical Center and the Leon H. Charn...

264KB Sizes 1 Downloads 155 Views

HANDS ON

How to perform noncontact mapping Larry A. Chinitz, MD, Jesse S. Sethi, MD From the New York University Medical Center and the Leon H. Charney Division of Cardiology, New York University School of Medicine, New York, New York.

Introduction

Signal acquisition

The anatomic and electrophysiologic complexity of arrhythmias subject to evaluation and catheter-based therapy has increased over the past several years. Use of advanced mapping systems, capable of three-dimensional rendering of cardiac chambers and superimposition of electrical information, are not designed to replace conventional mapping techniques but to be used as an adjunctive tool in the analysis and treatment of complex arrhythmias. EnSite 3000 (Endocardial Solutions, Minneapolis, MN) was the first component of the EnSite mapping system capable of advanced electroanatomic evaluation through novel catheter design and capabilities. The unique features of this system allow it to acquire data simultaneously from an entire cardiac chamber, often in a single cardiac cycle. This can be most useful in difficult arrhythmias where mapping is hampered by hemodynamic instability, sporadic ectopic beats, or multiple rhythm morphologies. The safety and utility of the EnSite system have been evaluated in animal studies and validated against conventional mapping in humans.1,2 The system has facilitated successful ablation of both focal and reentrant arrhythmias.3,4 It has been particularly useful in patients with multiple foci and/or unique anatomy such as prior cardiac surgery involving atriotomy scars or repair of congenital cardiac malformations (Figures 1 and 2). More recently, the array has been deployed in the left atrium to help elucidate the origins of focally initiated atrial fibrillation, ascertain conduction block across linear lesions delivered in the posterior left atrium and mitral isthmus, and provide guidance in the treatment of left atrial flutter (Figure 3). The promise of single-beat global mapping, on its own, may be considered an overstatement. In order to achieve the objective of the noncontact mapping technology, it is important to have a clear understanding of the mechanisms of signal acquisition, as well as its constraints and limitations.

The EnSite Array uses a catheter-based, noncontact multielectrode array (MEA), a reference patch electrode, amplifiers, filters, and a computer workstation. The system can simultaneously reconstruct more than 3,000 electrograms, translating them into a three-dimensional graphical representation of the cardiac chamber, either in an isochronal or isopotential mode. The MEA is mounted on a 9Fr catheter and consists of a 7.5-mL ellipsoid balloon surrounded by 64 electrically insulated wires. Each wire has a break in its insulation, allowing it to function as a unipolar electrode. The unipolar signals are recorded using a ring electrode as reference, located 16 cm proximal to the MEA. The array can be positioned in any cardiac chamber and sample electrophysiologic signals from the endocardial surface, feeding them into its amplifier. Data generated by the potential field are sampled at 1.2 kHz and filtered between 0.1 and 300 Hz. However, the cavitary potentials detected by the array are of relatively low amplitude and frequency and therefore are not easily interpretable in their raw form.5 The EnSite workstation analyzes and mathematically transforms the endocardial signals. The system uses an inverse solution to Laplace’s equation to ascertain how a signal detected by the array would appear on the endocardial surface. Ultimately, through its signal processing, the EnSite system allows for visualization of very small potentials (voltage sensitivity 10 ␮V) even at significant distances from the MEA. This is achieved through a redundancy in which a summation from all 64 electrodes is used in the reconstruction of potentials from a single site. Using these techniques, 3,360 virtual electrograms are produced and updated 1,200 times per second. In complex substrate-related arrhythmias (i.e., arrhythmogenic right ventricular dysplasia), use of activation mapping alone is insufficient for rhythm analysis or directing lesion delivery. Recently, Dynamic Substrate Mapping has been introduced, which allows the creation of voltage maps from a single cardiac cycle. This feature provides the capability of identifying low-voltage areas, as well as fixed and functional block, on the virtual endocardium through noncontact methodology (Figure 1). When combined with the activation sequence, substrate mapping provides essential information for guiding ablation, even when the arrhythmia

KEYWORDS Noncontact Mapping; Endocardial Solutions; EnSite; Unipolar electrograms; Activation mapping; Substrate mapping; Voltage Mapping (Heart Rhythm 2006;3:120 –123) Address reprint requests and correspondence: Dr. Larry A. Chinitz, Leon H. Charney Division of Cardiology, New York University Medical Center, 403 East 34th Street, New York, New York 10016. E-mail address: [email protected].

1547-5271/$ -see front matter © 2006 Heart Rhythm Society. All rights reserved.

doi:10.1016/j.hrthm.2005.10.030

Chinitz and Sethi

Noncontact Mapping

Figure 1 Focal right atrial tachycardia associated with an atriotomy scar. Bipolar voltage map (A) compared with Dynamic Substrate Mapping (B) showing the focal site of origin (C) and breakout (D) adjacent to the superior aspect of the scar. EnSite was able to guide successful lesion delivery to the border zone of the scar (E). TS ⫽ time step.

is nonsustained. Finally, the most recent version of the EnSite software provides the capability of point-by-point contact mapping, allowing the creation of activation and voltage maps by acquiring serial contact electrograms and displaying them on the virtual endocardium. This has utility in adding detail, familiarity, and validation of the information obtained by the noncontact method.

Geometry creation and catheter localization The noncontact mapping catheter is equipped with a balloon at the tip, over which the MEA is mounted. The catheter is

Figure 2 Ischemic ventricular tachycardia (VT) activation map shown at high-pass (HP) filtering of 1 Hz (A) and 8 Hz (B) with Dynamic Substrate Mapping (DSM) defining scar within the light blue border. Filtering at 1 Hz shows the low-frequency repolarization wave (time step [TS] ⫽ ⫺50 ms) and clearly defines presystolic activation at TS ⫽ 0 (time step with 0 ms at the onset of QRS). Raising the filter to 8 Hz eliminates the repolarization wave but shifts the visualization of presystolic activity away from the critical low-frequency, slow conduction zone.

121

Figure 3 Postablation left atrial flutter traversing around and between the left superior and inferior veins as well as across the isthmus between the left inferior pulmonary vein and the mitral annulus. Radiofrequency termination was achieved at the carina of the left-sided veins. HP ⫽ high pass.

advanced through a special 9Fr sheath into the chamber of interest and deployed by injection of a 30/70 mixture of contrast media and saline. The distal catheter is anchored within the cardiac chamber using the pigtail configuration at its tip or with a guidewire placed through the end of the catheter. The position of the array in the chamber must be secured to avoid significant movement that will invalidate the electrical and anatomic information. The array must be positioned as close as possible to the area of interest within the chamber because the accuracy of the map is sensitive to the distance between the center of the balloon and the endocardium being mapped. Moreover, for the endocardial surface in question to be effectively mapped, it must be in direct line of sight to the MEA through the blood pool. The system provides a constant display of the distance between the mapping catheter and the center of the balloon, providing feedback on the relative accuracy of the map. For the purpose of providing a global chamber map, we prefer to initially deploy the MEA in a very central location within the chamber of interest. However, without a modified MEA position, focal activity originating within the atrial appendage, mitral or tricuspid valve, or distal superior vena cava may not be adequately visualized. Thus, the MEA position may need to be altered after the area of interest is ascertained. Due to the high profile of the balloon, care must be taken to avoid obstruction of vessels or valves and to achieve appropriate anticoagulation with frequent monitoring of activated clotting times. The system can locate any conventional catheter with relation to the MEA using an impedance-based technology. Through this technique, a low-current, 5.68-kHz “locator” signal is emitted from the roving catheter electrode (up to four electrodes may be located). The signals are passed between the catheter electrodes and the ring electrodes proximal and distal to the array. The system calculates the signal angles and locates the position of each electrode in space, updating its position 400 times per second. The

122 Table 1

Heart Rhythm, Vol 3, No 1, January 2006 High pass filter settings by signal type*

Signal Type Spectral Components* High Pass Filter Typical Range Advantage Disadvantage

Repolarization atrial/ventricular

Slow Conduction scar border

Normal Myocardium atrial or ventricular

Fascicular His-Purkinje

.05– 4 Hz

1– 4 Hz

4 –16 Hz

8 –32 Hz

⬎4 Hz 8 –12 Hz eliminates far field

1 Hz 1– 4 Hz visualize slow conduction

8 Hz 4 –16 Hz minimal repolarization

eliminates slow conduction

potential repolarization artifact

eliminates slow conduction

16 Hz 8 –32 Hz eliminates far field, myocardial cond. eliminates myocardial conduction

*Estimated for unipolar virtual electrograms roving catheter then can be rapidly moved around the surface of any cardiac chamber, acquiring thousands of points during catheter movement. In order to create a detailed geometry, attempts must be made to make contact with as much endocardium as possible. This requires maneuvering on all sides of the array, which can be challenging and may require decreasing the profile of the balloon by withdrawing a few milliliters of fluid. This results in the rapid formation of a relatively accurate three-dimensional geometric model of the cardiac chamber. The geometry created by catheter movement is used to superimpose the 3,360 virtual electrograms on what becomes the reconstructed endocardial surface. This virtual endocardium may now be created with superimposed isopotential or isochronal maps with a color range representing voltage or timing of endocardial activation. Once the virtual endocardium is created, innumerable activation sequences can be acquired, analyzed, and stored.

Signal filtering and analysis In addition to the comprehensive color map displayed on the virtual endocardium, unipolar electrograms may be individually selected from the reconstructed surface and reviewed as waveforms on the system display. These selected unipolar waveforms are used to augment information obtained from the three-dimensional map by demonstrating the slope of depolarization, the presence of double potentials or fractionation, and the differentiation of far-field signals from more relevant endocardial activation. Identification of true local activation and its differentiation from far-field or repolarization waves is essential to successful utilization of the EnSite system. This problem may be emphasized with unipolar, noncontact mapping because signals detected by the array initially are amplified to derive the endocardial map. Thus, low-frequency noise, repolarization waves, or far-field signals in the measured cavity will be magnified into larger signals on the reconstructed chamber. In the unipolar electrograms, signals associated with high conduction velocity (i.e., His-Purkinje system) possess a greater slope (⫺dV/dT), and thus are characterized by higher-frequency spectral components (⬎32 Hz). Electrograms re-

corded in normally conducting atrial or ventricular myocardium possess spectral components in the mid range of 4 to 16 Hz, whereas electrograms in regions of slow conduction are composed of lower-frequency spectral components of 1 to 4 Hz. Spectral components associated with repolarization waves lie in the low end of the spectrum at 1 to 4 Hz (Table 1). Thus, the high-pass filter must be adjusted between 1 and 32 Hz, helping to modulate the extent to which low-frequency signals are visible on the three-dimensional display. However, filtering out large repolarization waves by the high-pass filter may attenuate critical information contained in relevant signals of lower frequency and amplitude. This is especially important in differentiating the origin of tachycardias from their endocardial breakout sites or tracking diastolic depolarization during reentrant arrhythmias (Figures 2 and 4). Although automated filter settings are provided, manual manipulation of high-pass filters are readily achieved, allowing for differentiation of near-field and farfield signals, essential to identifying clinically relevant information. An additional tool, vital to the analysis of EnSite images, is the dynamic range of color used on the three-dimensional map representing voltage amplitude. The gain on the threedimensional color display allows the user to concentrate on

Figure 4 Unipolar virtual electrograms filtered at 1 Hz (A), 8 Hz (B), and 32 Hz (C) with a fixed color voltage scale of 0.1 mV, each at the same instant in time. The high-frequency fascicular potential is best seen by filtering out low-frequency repolarization waves and far-field signals by progressively raising the high-pass (HP) filter. TS ⫽ time step.

Chinitz and Sethi

Noncontact Mapping

123 the scale would be opened up to include an array of colors representing a spectrum of voltages. This ability to vary the voltages visible on the color map, along with high-pass filtering, permits the operator to concentrate on particular signals in an area of interest. When these tools are combined with analysis of the unipolar and bipolar electrograms, one is empowered with invaluable information needed to understand the mechanisms of complex arrhythmias and select sites for lesion delivery.

References

Figure 5 Narrow-band (A: 0.1 mV) vs broad-band (B: 2 mV) color voltage scale used to differentiate fascicular from myocardial activation. Narrowing the scale as in panel A highlights the fascicular potential but saturates myocardial activation. Broadening the scale as in panel B displays the myocardial breakout but partially obscures the low-amplitude fascicular activation. HP ⫽ high pass; TS ⫽ time step.

a narrow or wide range of potentials. By diminishing the color scale, as might be required to see a fascicular potential or diastolic depolarization during reentry, larger-amplitude signals will be eliminated ( Figure 5). To visualize the broad spectrum of potentials present during a tachycardia cycle,

1. Thiagalingam A, Wallace EM, Campbell CR, Boyd AC, Eipper VE, Byth K, Ross DL , Kovoor P. Value of noncontact mapping for identifying left ventricular scar in an ovine model. Circulation 2004;110: 3175–3180. 2. Schilling R, Peters N, Davies DW. Simultaneous endocardial mapping in the human left ventricle using a noncontact catheter: comparison of contact and reconstructed electrograms during sinus rhythm. Circulation 1998;98:887– 898. 3. Schilling R, Peters N, Davies DW. Feasibility of a noncontact catheter for endocardial mapping of human ventricular tachycardia. Circulation 1999;99:2543–2552. 4. Fung JW, Chan HC, Chan JY, Sanderson JE. Ablation of nonsustained or hemodynamically unstable ventricular arrhythmia originating from the right ventricular outflow tract guided by noncontact mapping. Pacing Clin Electrophysiol 2003;26:1699 –1705. 5. Khoury D, Taccardi B, Lux R, Ershler P, Rudy Y. Reconstruction of endocardial potentials and activation sequences from intracavitary probe measurements: localization of pacing sites and effects of myocardial structure. Circulation 1995;91:845– 863.