Journal of Chromatography A, 1356 (2014) 211–220
Contents lists available at ScienceDirect
Journal of Chromatography A journal homepage: www.elsevier.com/locate/chroma
Hydrophilic interaction chromatography versus reversed phase liquid chromatography coupled to mass spectrometry: Effect of electrospray ionization source geometry on sensitivity夽 Aurélie Periat a , Isabelle Kohler a , Aurélie Bugey b , Stefan Bieri b , Franc¸ois Versace c , Christian Staub a,c , Davy Guillarme a,∗ a
School of Pharmaceutical Sciences, University of Geneva, University of Lausanne, 20, Boulevard d’Yvoy, 1211 Geneva 4, Switzerland Official Food Control Authority and Veterinary Affairs of Geneva, 22, Quai Ernest-Ansermet, 1211 Geneva 4, Switzerland c University Center of Legal Medicine, Forensic Toxicology and Chemistry Unit, 1, Rue Michel Servet, 1211 Geneva 4, Switzerland b
a r t i c l e
i n f o
Article history: Received 5 March 2014 Received in revised form 14 May 2014 Accepted 16 June 2014 Available online 26 June 2014 Keywords: Hydrophilic interaction chromatography HILIC–MS Sensitivity Electrospray ionization Source geometry ESI
a b s t r a c t In this study, the influence of electrospray ionization (ESI) source design on the overall sensitivity achieved in hydrophilic interaction chromatography (HILIC) and reversed phase liquid chromatography (RPLC), was investigated. State-of-the-art triple quadrupole mass analyzers from AB Sciex, Agilent Technologies and Waters equipped with brand specific source geometries were tested with various mobile phase pH on 53 pharmaceutical compounds. The design of the ESI source showed to strongly influence the gain in sensitivity that can be achieved in HILIC compared to RPLC mode. The 6460 Triple Quadrupole LC/MS system from Agilent Technologies was particularly affected by mobile phase settings. Indeed, compared to RPLC conditions, 92% of the compounds had an increased signal-to-noise ratio at a flow rate of 300 L/min in HILIC mode at pH 6, while this percentage dropped to only 7% at 1000 L/min and pH 3. In contrast, the influence of flow rate and mobile phase pH on the gain in sensitivity between RPLC and HILIC was found very limited with the API 5000TM LC/MS/MS system from AB Sciex, as only 15 to 36% of the tested compounds showed an enhanced sensitivity in HILIC mode. With the Xevo TQ-S instrument from Waters, superior sensitivity in HILIC was noticed for 85% of the compounds with optimal conditions (i.e., pH 3 and 1000 L/min), whereas at sub-optimal conditions (i.e. pH 6 and 300 L/min), it represented less than 50%. The gain in sensitivity observed in HILIC was found less significant with the recent LC–MS platforms used in this study than for old-generation instruments. Indeed, the improved ESI sources equipping the recent mass analyzers allow for enhanced evaporation efficiency, mainly for RPLC mobile phases containing high proportion of water and this even at high flow rates © 2014 Elsevier B.V. All rights reserved.
1. Introduction Hydrophilic interaction chromatography (HILIC) appears today as an attractive alternative to reversed phase liquid chromatography (RPLC) for analyzing polar and/or ionizable compounds [1–4]. This chromatographic mode is characterized by a polar stationary phase and aqueous-polar organic mobile phase containing an important proportion (>60%) of organic modifier, usually acetonitrile. This highly volatile organic mobile phase
夽 Presented at the 13th International Symposium on Hyphenated Techniques in Chromatography and Separation Technology, Bruges, Belgium, 29–31 January 2014. ∗ Corresponding author. Tel.: +41 22 379 34 63; fax: +41 22 379 68 08. E-mail address:
[email protected] (D. Guillarme). http://dx.doi.org/10.1016/j.chroma.2014.06.066 0021-9673/© 2014 Elsevier B.V. All rights reserved.
provides low backpressure due to its weak viscosity. In addition, it enhances droplet formation and desolvation efficiency, leading to a significant sensitivity improvement in electrospray ionization (ESI) [5,6]. This is also true for other detectors based on nebulization/evaporation processes (e.g. corona charged aerosol and evaporative light scattering detectors) [7]. Indeed, a gain in sensitivity for a large range of compounds has been reported several times in HILIC-ESI/MS compared to RPLC–ESI/MS [7–14]. A recent systematic study based on signal-to-noise ratios (S/N) comparison highlighted a ca. 4-fold median gain in sensitivity in HILIC vs. RPLC using a dataset of 56 basic drugs covering a wide range of physico-chemical properties [13]. Nevertheless, large differences were observed within the compounds set, with some of them even reaching a >100-fold sensitivity improvement in HILIC. Moreover, for ca. 95% of the compounds, highest S/N ratios were observed in HILIC mode.
212
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
Not only are the chromatographic conditions, e.g., mobile phase composition and pH, influencing the ionization efficiency, but also the ESI source geometry. Indeed, the multiple steps involved in ESI process, including droplet formation, desolvation, creation of gasphase ions and ion sampling, may be significantly affected by the design of the source [15]. Since the first ESI-MS experiments performed by Dole et al. in 1968, numerous developments have been carried out to improve the ionization efficiency, including modifications of the source geometry, the sprayer position relative to MS entrance, the sampling cone diameter, the number of transfer capillaries, as well as addition of heated gas [15–17]. Today, several source geometries are commercially available, i.e., (i) off-axis geometry, where the spray is at 30–45◦ to the x-axis between sampling capillary and the first quadrupole (e.g. Ion Spray interface), (ii) orthogonal geometry, where the spray is positioned at 90◦ relative to the x-axis (e.g., Jet Stream ESI source from Agilent Technologies or Turbo VTM source from AB Sciex), and (iii) Z-Spray geometry, which presents a double orthogonal sampling (e.g., Z-Spray from Waters). Orthogonal and Z-Spray configurations are usually providing better sensitivities than other source designs due to the prevention of clogging the MS orifice by non-volatile materials [15,17,19]. Until now, only a limited number of comparative studies have highlighted the effect of ESI source geometry on matrix effect or on signal suppression caused by ion-pairing reagent [16,18–21]. In the present work, the influence of ESI source design on the overall sensitivity observed between HILIC vs. RPLC was assessed using three state-of-the-art triple quadrupole (QqQ) mass spectrometers, i.e., Agilent 6460 Triple Quadrupole instrument from Agilent Technologies, API 5000TM System from AB Sciex, and Xevo TQ-S from Waters. Signal intensities, background noises and S/N ratios measured in HILIC and RPLC conditions were systematically compared using a training set of 53 basic drugs covering a broad range of lipophilicity and ionization properties. Besides, the influence of mobile phase pH and flow rate was also investigated. 2. Experimental 2.1. Chemicals and reagents Ultrapure water was supplied by a Milli-Q Advantage A10 purification system from Millipore (Bedford, MA, USA). Acetonitrile (ACN), methanol (MeOH), formic acid and acetic acid were of ULC–MS grade and purchased from Biosolve (Valkenswaald, Netherlands). The ammonium hydroxide (28%, m/v) solution was obtained from Sigma–Aldrich (Buchs, Switzerland). The 10 mM ammonium formate buffer at pH 3 was prepared with an adequate volume of formic acid and further adjusted to pH 3.0 with an ammonium hydroxide solution (28%, m/v). The pH values were measured with a SevenMulti pH meter (Mettler-Toledo, Schwerzenbach, Switzerland). 2.2. Compound dataset The training set of 53 basic drugs, covering a broad spectrum of pKa (i.e., between 6 and 11, except for a few compounds in the 2–6 pKa range) and log P values (i.e., between −1.2 and 5.6), was composed of the following compounds: 6-monoacetylmorphine, acebutolol, adenosine, alprazolam, alprenolol, amphetamine, antipyrine, bisoprolol, buprenorphine, bupropion, buspirone, clonidine, cocaethylene, cocaine, codeine, dextromethorphan, dibucaine, diltiazem, doxepin, fentanyl, flurazepam, heroin, hydroxyzine, imipramine, ketamine, N-methyl-1-(1,3-benzodioxol-5-yl)-2-butanamine (MBDB), 3,4methylenedioxyethamphetamine (MDEA), methadone, methamphetamine, methylephedrine, naloxone, naltrexone, norcocaine,
norephedrine, nortriptyline, noscapine, papaverine, perphenazine, pethidine, pindolol, prilocaine, propranolol, pseudoephedrine, pyrilamine, reserpine, salbutamol, sulpiride, terfenadine, tetracaine, thebaine, tolazoline, tramadol and triprolidine, which were either purchased in powder form from Sigma–Aldrich (Steinheim, Germany) or in methanolic solution (1 mg/mL) from Lipomed (Arlesheim, Switzerland). Stock solutions of compounds in powder form were prepared at 1 mg/mL in pure MeOH. Two mixtures of compounds were prepared. The first mixture was composed of 26 compounds including adenosine, alprenolol, bupropion, heroin, norephedrine, perphenazine, propranolol and terfenadine each at 1000 ng/mL; alprazolam, amphetamine and thebaine each at 500 ng/mL; naloxone at 400 ng/mL; dextromethorphan, methylephedrine and pseudoephedrine each at 200 ng/mL; dibucaine, doxepin and fentanyl each at 100 ng/mL; buspirone, MBDB, norcocaine, pethidine, prilocaine, sulpiride and tramadol each at 50 ng/mL and diltiazem at 30 ng/mL. The second mixture consisted of 27 compounds including nortriptyline at 2000 ng/mL, buprenorphine at 1500 ng/mL, codeine and reserpine each at 1000 ng/mL, clonidine, naltrexone and triprolidine each at 500 ng/mL; acetylmorphine, antipyrine and tolazoline each at 400 ng/mL; MDEA, methamphetamine, pindolol and pyrilamine each at 200 ng/mL; cocaethylene, flurazepam, hydroxyzine, ketamine, methadone, noscapine and salbutamol each at 100 ng/mL; cocaine, papaverine and tetracaine each at 50 ng/mL; imipramine and acebutolol each at 25 ng/mL; and bisoprolol at 5 ng/mL. Both samples were then diluted 10-fold for the analyses with Agilent and Sciex instruments, and 100-fold for the Waters detector. At least 95% ACN or water was used for dilution in HILIC and RPLC conditions, respectively, and 2 L of sample volume were injected. 2.3. Instrumentation 2.3.1. Mass spectrometry Various QqQ mass analyzers were used for this study, namely (i) Agilent 6460 Triple Quadrupole System (Agilent Technologies, Santa Clara, CA, USA), (ii) AB Sciex API 5000TM System (AB Sciex, Concord, Canada) and (iii) Waters Xevo TQ-S (Waters Corporation, Milford, MA, USA). All experiments were carried out in positive ionization mode and Selected Reaction Monitoring (SRM) acquisition. Source parameters were systematically investigated in HILIC and RPLC mode using a mixture of 4 model compounds representative of the training set, i.e., adenosine, clonidine, doxepin and norcocaine at 600 L/min and pH 3; optimal conditions are summarized in Table 1. Nitrogen was used as the nebulizing gas. Dwell times and inter-channel delays were both set at 5 ms, allowing for a sufficient number of data points across the narrow chromatographic peaks obtained with columns packed with sub-2 m particles [17]. Cone voltages and collision energies were determined for each compound and are presented in Table S1 (Supplementary material). 2.3.2. Liquid chromatography An Agilent 1290 Infinity LC System (Agilent Technologies, Waldbronn, Germany) was used for the experiments with the Agilent 6460 Triple Quadrupole. Data acquisition, instrument control and data handling were performed with MassHunter software. A Thermo Scientific Dionex UltiMate 3000 RSLC System (Dionex, CA, USA) was used for experiments with AB Sciex API 5000TM . Data were acquired and processed with Analyst software. A Waters Acquity UPLCTM I-Class System from Waters (Milford, MA, USA) was used in combination with Waters Xevo TQ-S. Data acquisition, instrument control and data handling were performed with MassLynx Software. A Zorbax HILIC Plus Rapid Resolution HD column (50 mm × 2.1 mm i.d., 1.8 m) from Agilent Technologies was
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
213
Table 1 Settings of source parameters. TurboVTM Source (AB Sciex API 5000TM )
RPLC
HILIC
Jet Stream source (Agilent 6460 Triple Quadrupole)
RPLC
HILIC
Z-spray Source (Waters Xevo TQ-S)
RPLC
HILIC
Spray voltage (V)
3000
5000
Capillary voltage (V)
3500
3000
Nebulizer gas pressure (psi)
40
30
Nebulizer gas pressure (psi)
50
40
3000 50 7
2000 50 7
Heater gas pressure (psi)
50
40
Sheath gas flow rate (L/min)
11
11
Capillary voltage (V) Cone voltage (V) Nebulizer gas pressure (bar) Desovation gas flow rate (L/h) Desovation gas temperature (◦ C) Cone gas flow rate (L/h)
1000
1000
600
600
150
150
◦
750
650
Sheath gas temperature ( C)
400
400
Curtain gas pressure (psi)
20
20
Drying gas flow rate (L/min) Drying gas temperature (◦ C)
10 300
10 300
Heater gas temperature ( C)
◦
used for HILIC experiments while an Acquity UPLC BEH C18 (50 mm × 2.1 mm i.d., 1.7 m) from Waters was employed for all RPLC experiments.
2.4. Systematic evaluation of sensitivity The 53 model compounds were analyzed in HILIC and RPLC modes using generic stationary phases, i.e., C18 and unmodified bare silica, respectively. Two mobile phase pH, i.e., pH 3 and 6, were investigated in both modes. Experiments were conducted at three different mobile phase flow rates, namely 300, 600 and 1000 L/min. For HILIC experiments at 600 L/min, the gradient profile was the following: 95% ACN for 1 min, 95–65% ACN in 2.2 min, and a final isocratic step at 65% ACN for 0.5 min. The same gradient profile was used for experiments at 300 and 1000 L/min but the times were geometrically scaled leading to the following gradient: 95% ACN for 2 min, 95–65% ACN in 4.4 min, and 65% ACN for 1 min for experiments at 300 L/min; 95% ACN for 0.6 min, 95–65% ACN in 1.32 min, and 65% ACN for 0.3 min for experiments at for 1000 L/min. For RPLC experiments at 600 L/min, the gradient profile consisted of 2% ACN for 1 min, 2–70% ACN in 5 min, and a final isocratic step at 70% ACN for 0.5 min. The same gradient profile was employed for the two other flow rates, leading to the following gradients: 2% ACN for 2 min, 2–70% ACN in 10 min, and 70% ACN for 1 min for experiments at 300 L/min; 2% ACN for 0.6 min, 2–70% ACN in 3 min, and 70% ACN for 0.3 min for experiments at 1000 L/min. Both mixtures of compounds were injected in triplicate under each experimental condition. In order to ensure the validity of the methodology, a system suitability experiment (SSE) consisted in an injection of a mixture of 4 analytes (i.e., adenosine, clonidine, doxepin and norcocaine) was performed every 18 analysis with a systematic evaluation of chromatographic performances (e.g., retention times variability, peak width, etc.) and MS signal intensity. Prior to each SSE, a blank injection was performed to avoid any carry-over effect. The S/N values of individual compounds were reported for each condition and automatically calculated by the data acquisition software, considering the same methodology for noise determination, i.e., the selected background noise zones for flow rates of 300, 600 and 1000 L/min were between 0.5 and 2.5 min; 0.5 and 1.5 min and 0.6 and 1.0 min, respectively. For the compounds which elute in this region, the noise area was between 3.0 and 5.0 min; 1.5 and 2.5 min; and 1.0 and 1.5 min for 300, 600 and 1000 L/min respectively. This systematic procedure was strictly similar to the one implemented in a previous study described elsewhere which dealt with a previous-generation MS instrument [13]. The results obtained in the present study can thus also be compared to the previous one, increasing the number of tested QqQ instruments to four.
3. Results and discussion 3.1. Description of ESI sources This study compares the sensitivity achieved in RPLC and HILIC modes on three QqQ instruments representing the latest generation of MS instruments. These systems present obvious differences, mostly in their ionization source geometry, as illustrated in Fig. 1, which may significantly impact both RPLC–MS and HILIC–MS overall sensitivity. The ionization process of all these interfaces is pneumatically assisted by a nitrogen gas coaxially blowing around the spray capillary, to help droplet formation and desolvation [22]. Besides this nitrogen stream, additional features have been developed by the different manufacturers to support the droplet desolvation. The Turbo VTM source (Fig. 1A) has been developed by AB Sciex. It is compatible with flow rates of up to 3 mL/min and able to vaporize from 100% aqueous to 100% organic mobile phases. In this source geometry, a nebulizer gas (up to 90 psi) and a heater gas that originates from the turbo heaters (≤90 psi and temperature of up to 750 ◦ C) are added for a more efficient nebulization. The combination of the IonSpray effluent and the heated drying gas from the turbo sprayer are then projected at a 90◦ -angle to the MS sampling orifice [23]. Recent MS instruments from Agilent Technologies are equipped with the so-called Jet Stream thermal gradient focusing technology, illustrated in Fig. 1B, where a heated nitrogen sheath gas (up to 400 ◦ C and 12 L/min) is surrounding the nebulizer. In this configuration, according to the manufacturer, a 5–10-fold gain in sensitivity may be achieved for mobile phase flow rates between 0.25 and 2 mL/min which is attributed to a spray confinement and a more efficient desolvation process [24]. In the Z-spray source (Fig. 1C) proposed by Waters (Milford, MA, USA), the probe is positioned perpendicularly to the sampling cone, while the second extraction cone is also perpendicular to the ion beam, thus improving the mobile phase elimination. In addition, the source block is heated at 150 ◦ C and the heated nebulizing gas flow and temperature can be as high as 1200 L/h and 650 ◦ C, respectively. Under such extreme conditions, the negative impact of elevated mobile phase flow rates on sensitivity seems to be moderate.
3.2. Impact on sensitivity in RPLC and HILIC The results obtained for the evaluation of sensitivity in HILIC and RPLC modes are reported in Figs. 2–4 for Agilent, AB Sciex and Waters instruments, respectively, based on similar experiments. The whole series of experiments on a given device were systematically performed in a timeframe as narrow as possible to lower the time-dependant variability. The sensitivity was estimated at three different flow rates, i.e., 300, 600 and 1000 L/min, adapted to the column dimensions, and with two mobile phase pHs, i.e., pH 3 and
214
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220 analyte soluon
analyte soluon
nebulizing gas
heater
heater
B. curtain gas
A.
super-heated sheath gas heated drying gas
curtain gas
nozzle voltage heated nitrogen
spray
exhaust
C.
nebulizing gas
resisve sampling capillary
vent heated desolvaon gas
baffle plate
analyte soluon
nebulizing gas sampling cone
extracon cone
isolaon valve
ion block cone gas
Fig. 1. Graphical representation of the investigated source geometries. (A) TurboVTM source of AB Sciex API 5000TM LC/MS System; (B) Jet Stream source of Agilent 6460 Triple Quadrupole LC/MS system; and (C) Z-Spray source of Waters Xevo TQ-S. Adapted from [15] with permission.
6. In order to have comparable S/N ratios between the LC–MS systems, analytes were diluted 10-fold for use with Agilent and AB Sciex instruments, and 100-fold for Waters instrument which was more sensitive. The percentages reported in Figs. 2–4 correspond to the proportion of compounds among the set of 53 basic drugs presenting a specific behaviour. The group labeled in dark gray highlights the compounds giving a lower sensitivity in HILIC compared to RPLC (<0.8-fold). The category labeled in intermediate gray corresponds to the compounds showing a negligible variation of sensitivity (0.8 ≤ x < 1.2). The last class underlines the analytes presenting a significant sensitivity enhancement in HILIC of 1.2–5-fold, in light gray, and higher than 5-fold, in white, respectively. Therefore, these results emphasize the modifications of sensitivity observed between HILIC and RPLC conditions which were mostly depending on the MS devices and, thus of the source geometry and ion transmission efficiency.
than HILIC mode, as already discussed in a previous study [13]. Therefore, the optimal combination of pH and flow rate allowing for the highest sensitivity gain using HILIC was pH 6 and 300 L/min. On the contrary, at relatively high flow rates (1000 L/min) and acidic pH conditions (pH 3), the HILIC mode was considered irrelevant to improve the sensitivity due to a significant noise increase, as discussed later in Section 3.3. The ionization behaviour of this source seems difficult to predict and the ionization efficiency is highly dependent on the mobile phase composition and its flow rate, as well as sheath gas parameters, i.e., flow rate and temperature. Indeed, the efficiency of the spray plume confinement as well as desolvation seems to be closely linked to all the experimental conditions. It is worth mentioning that this source may also have a different influence on matrix effects compared to other sources, as demonstrated in a comparative study where much higher ion suppression was observed with this design [16].
3.2.1. Agilent Jet Stream source As illustrated in Fig. 2, a major impact of mobile phase pH on the gain in sensitivity in HILIC compared to RPLC was noticed on the Agilent instrument. Indeed, the HILIC mode was found much more beneficial to improve sensitivity at pH 6 than pH 3, when compared to RPLC in the same conditions. For instance, at a flow rate of 300 L/min and at pH 6, 92% of the compounds presented higher S/N ratios in HILIC compared to RPLC, while this proportion dropped to 74% at pH 3. This difference of behaviour linked to the pH was even more pronounced at a higher flow rate, as shown at 1000 L/min where 7% of the compounds showed a better sensitivity in HILIC at pH 3, compared to 59% at pH 6. This important dissimilarity was mainly explained by the significant decrease in sensitivity observed in RPLC at pH 6 (data not shown). Indeed, in comparison to conditions at pH 6, the overall ionization efficiency was enhanced at pH 3, due to the basic properties of all the compounds of the training set. Furthermore, the influence of mobile phase pH was frequently considered more pronounced in RPLC
3.2.2. AB Sciex Turbo VTM source As shown in Fig. 3, the influence of flow rate and mobile phase pH on sensitivity between RPLC and HILIC was found negligible with AB Sciex API 5000TM LC/MS/MS System. Indeed, 51–68% of the compounds provided a higher sensitivity in RPLC, while only 15–36% of the compounds showed an enhanced sensitivity in HILIC. Surprisingly, RPLC mode was most of the time leading to a higher sensitivity compared to HILIC despite expected better desolvation efficiency with the latter. This may be explained by the geometry of the source which has been specially designed to maintain high ionization efficiencies, even under extreme conditions of flow rate or with highly aqueous mobile phases, as described in Section 3.1. Moreover, the selected experimental conditions, i.e., 750 ◦ C for heater gas temperature and 50 psi for nitrogen gas pressure, respectively, have probably contributed to minimize the impact of the mobile phase composition and flow rate, leading to comparable desolvation efficiency between HILIC and RPLC. Therefore, the
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
215
Fig. 2. Gain in sensitivity, expressed by S/N achieved in HILIC vs. RPLC conditions at different flow rates (i.e., 300, 600 and 1000 L/min) and pH (i.e., 3 and 6) on the Agilent 6460 Triple Quadrupole LC/MS/MS system. The percentages of the 53 tested compounds presenting different S/N ratios between RPLC and HILIC were highlighted in the pie charts in black, i.e., <0.8-fold higher sensitivity in HILIC; dark gray, i.e., 0.8 ≤ x < 1.2-fold higher sensitivity in HILIC; light gray, i.e., 1.2 ≤ x < 5-fold higher sensitivity in HILIC; and black, i.e., ≥5-fold higher sensitivity in HILIC.
HILIC mode did not relevantly improve the overall sensitivity with this source geometry. 3.2.3. Modern Z-Spray source – Waters Xevo TQ-S Fig. 4 highlights the modifications of S/N ratios in HILIC and RPLC observed with different mobile phase flow rates and pH, which followed a similar trend than the results obtained with the Agilent source but to a lesser extent. In this particular case, the HILIC mode significantly enhanced the ionization efficiency at high flow rate and low pH. Indeed, for 85% of the compounds, the S/N ratios were typically improved in HILIC at 1000 L/min and pH 3, while only 47% were enhanced at 300 L/min and pH 6. It is worth mentioning that this trend is exactly the opposite than the one observed
with the Agilent instrument. This behaviour may be explained by the negligible effect of the mobile phase pH on the sensitivity in RPLC conditions with the Waters source, which enables to maintain a constant noise level independently of the column flow rate as discussed in Section 3.3. The higher gain in sensitivity under HILIC conditions at high flow rates can be attributed to the fact that the effluent may be less efficiently evaporated at high flow rates and especially with a high proportion of water. This leads to higher desolvation efficiency in HILIC, while this effect is likely rather pronounced at low flow rates, whatever the mobile phase composition. Compared to the AB Sciex Turbo VTM source, the ZSpray source geometry seems thus more adapted to HILIC than RPLC conditions.
216
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
Fig. 3. Gain in sensitivity, expressed by S/N achieved in HILIC vs. RPLC conditions at different flow rates (i.e., 300, 600 and 1000 L/min) and pH (i.e., 3 and 6) on the AB Sciex API 5000TM LC/MS/MS System. See Fig. 2 for pie charts labeling.
3.2.4. Previous Z-Spray source – Waters Acquity TQD The conclusions drawn in this study were relatively different from the ones observed with a Waters Acquity TQD instrument of a previous generation [13]. Indeed, with the latter instrument, the capability of HILIC to improve sensitivity was much more pronounced, as shown by the proportion of compounds showing a higher sensitivity in HILIC, i.e., between 77 and 94%, depending on the mobile phase conditions. Moreover, the absolute gain in sensitivity achieved in HILIC was noticeable, with ca. 10% of the tested compounds presenting a >30-fold gain of sensitivity. It is worth mentioning that such a drastic increase in sensitivity was hardly observed with Agilent and recent Waters instruments, (and never with the AB Sciex instrument). This may likely be explained by the recent developments and the significant improvements in ESI source designs. Since new ESI geometries have been optimized to be readily compatible with highly aqueous mobile phase and high
flow rates, the desolvation efficiency is almost no more influenced by mobile phase conditions and composition. As illustrated herein with Waters instruments, the expected gain in sensitivity in HILIC mode may therefore be particularly relevant for old-generation interfaces, where the ionization efficiency remains significantly lower than for the most recent MS systems. 3.3. Evaluation of background noise With Waters TQD instrument (old-generation MS device), it has been reported that the background noise in HILIC and RPLC was statistically equivalent, whatever the pH and flow rate conditions [13]. The noise was considered purely electronic-dependent, due to the high selectivity of the SRM acquisition mode. Table 2 reports the noise values measured in the present study using the procedure detailed in Section 2.4, for all experimental
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
217
Fig. 4. Gain in sensitivity, expressed by S/N achieved in HILIC vs. RPLC conditions at different flow rates (i.e., 300, 600 and 1000 L/min) and pH (i.e., 3 and 6) on the Waters Xevo TQ-S system. See Fig. 2 for pie charts labeling.
at 1000 L/min for ca. 55% and 72% of the compounds at pH 3 and 6, respectively. On the Waters Xevo TQ-S instrument, the observed noise was 1.2–5-fold higher in HILIC for 51–67% of the compounds depending on – experimental conditions, and larger than 5-fold for ca. 20–37% of them. A very similar trend was observed with the
conditions. As shown here, the reported values could be significantly different between both chromatographic modes, which was not true in our previous study [13]. Amongst these results, the most relevant was a 5-fold increase in noise level under HILIC conditions with the Agilent instrument
Table 2 Proportion of compounds showing a noiseHILIC /noiseRPLC ratio < 0.8 (i.e., higher noise in RPLC); between 0.8 and 1.2 (i.e., similar noise in RPLC and HILIC); between 1.2 and 5 (i.e., higher noise in HILIC); and above 5 (i.e., higher noise in HILIC).
<0.8 (%) 0.8–1.2 (%) 1.2–5 (%) >5 (%)
Agilent 6460 Triple Quadrupole LC/MS
AB Sciex API 5000TM LC/MC/MC
Waters Xevo TQ-S
pH 3
pH 3
pH 3
pH 6
pH 6
pH 6
300
600
1000
300
600
1000
300
600
1000
300
600
1000
300
600
1000
300
600
1000
53 26 15 6
2 9 74 15
0 8 38 55
32 23 40 6
21 23 45 11
2 0 26 72
25 25 47 4
13 21 57 9
13 38 42 8
6 4 64 26
9 9 60 21
11 13 66 9
14 6 58 23
8 2 58 33
2 8 67 23
4 8 51 37
4 2 58 36
10 12 59 20
218
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
Table 3 Proportion of compounds showing a heightHILIC /heightRPLC ratio < 0.8 (i.e., higher noise in RPLC); between 0.8 and 1.2 (i.e., similar noise in RPLC and HILIC); between 1.2 and 5 (i.e., higher noise in HILIC); and above 5 (i.e., higher noise in HILIC). AB Sciex API 5000TM LC/MC/MC
Waters Xevo TQ-S
pH 3
pH 3
pH 3
pH 6
pH 6
pH 6
300
600
1000
300
600
1000
300
600
1000
300
600
1000
300
600
1000
300
600
1000
18.9 9.4 69.8 1.9
15.1 5.7 77.4 1.9
22.6 5.7 67.9 3.8
3.8 1.9 9.4 84.9
1.9 1.9 11.3 84.9
0.0 0.0 13.2 86.8
45 30 23 2
34 36 28 2
47 30 21 2
11 11 64 13
15 21 55 9
21 25 49 6
9 4 60 26
0 0 25 76
0 0 8 93
9 6 53 32
8 2 57 34
4 2 30 64
obtained with an old-generation QqQ instrument [13]. The median gain in sensitivity was selected for the graphical representation instead of the average, since few compounds presented an extreme behaviour. As previously discussed, the old-generation waters TQD presented a significant gain in sensitivity under HILIC conditions, much higher than the one achieved with recent MS systems evaluated in the present study. Similarly, the Agilent instrument follows the same trend than old-generation Waters device since the sensitivity gain obtained in HILIC mode decreased along with an increased flow rate. Nevertheless, the absolute median gains in sensitivity were very different, varying from 5.2- to 3.3-fold better S/N ratios in HILIC for the old-generation Waters TQD (purple) and from 2.1- to 0.7-fold for the Agilent system (blue) from 300 to 1000 L/min, respectively. An opposite behaviour was observed with the recent Waters Xevo TQ-S (green), where a higher sensitivity gain in HILIC at high flow rate was observed, ranging from 1.0- to 3.0-fold at 300 and 1000 L/min, respectively. The opposite trend observed with the two Waters devices may be explained by major modifications on the ESI source, sampling cone and ion
A.
6.0 5.2 5.0
Median gain in HILIC
AB Sciex instrument where a 1.2–5-fold increase in background noise was measured in HILIC for 42–66% of the compounds, while the proportion of compounds providing a 5-fold superior noise in HILIC was less important, in comparison with the Waters Xevo TQ-S system. As already mentioned, the background noise was basically similar between RPLC and HILIC modes on an old-generation instrument, whereas a significant difference was reported in the present study with recent state-of-the-art mass analyzers, i.e., a higher noise observed under HILIC conditions. Thus, this greater background noise contributes to reduce the expected positive impact of HILIC conditions on S/N ratios, even though enhanced signal heights are mostly observed. Table 3 illustrates the differences in reported peak heights between RPLC and HILIC, showing significantly higher peak intensities in HILIC for a large number of compounds. Indeed, for more than three quarters of the compounds, an increase in signal height was measured in HILIC conditions compared to RPLC with both Agilent and Waters instruments. For these instruments, the amplified noise level in HILIC hindered a significant gain of overall sensitivity. A different situation was observed with the AB Sciex instrument, where peak signals were enhanced for only ca. 24% and ca. 56% of test compounds in HILIC at pH 3 and 6, respectively. According to Tables 2 and 3, the poor S/N ratios obtained in HILIC with the AB Sciex instrument compared to Agilent and Waters devices is a consequence of a higher background noise and similar or lower peak heights under HILIC conditions with these latter systems. Several reasons may explain the differences in the measured background noise levels between the two chromatographic modes. With the most recent state-of-the-art MS instruments, the technological developments allow for reaching ever-lower analyte detection limits thanks to improvements in source geometry, ion sampling, ion optics, and vacuum capability. Therefore, an underlying issue is the strong need for noise reduction by mainly suppressing extraneous peaks caused by solvent background. In RPLC, it is well known that the background noise generally increases along with the highly organic portion of the gradient [25]. This noise increase may be attributed to low solvent purity, potential clusters between the target analytes and organic solvent, contamination with phtalates (plasticizers) or metal ion content, which can lead to the formation of mass adducts with the analytes. Moreover, a high acetonitrile-containing mobile phase may form clusters of chloride ions with ammonium [26]. Besides the contamination from the mobile phase itself, column bleeding may be more pronounced at a higher percentage of organic modifier, leading to additional contaminants entering the MS and thus increasing the background noise. As HILIC involves the use of high proportion of acetonitrile in the mobile phase, it certainly explains the higher background observed with this chromatographic mode using highly sensitive modern QqQ devices.
4.3
3.3 2.9
3.0 2.1 2.0
1.5
1.4 0.8
0.7
Fig. 5 reports the median change in sensitivity achieved in this study with different mobile phase conditions, including the results
1.4
1.1
1.0
0.7 0.8
0.6
0.0 all condions
B.
300μL/min
600 μL/min
1000 μL/min
6.0 4.9
5.0 4.3 4.0
3.7
3.6
3.0 2.0 1.0
1.9 1.5
1.4 0.7
1.3
1.2 0.7
0.7
0.0 all condions
3.4. Overall HILIC and RPLC sensitivity
4.1
4.0
1.0
Median gain in HILIC
<0.8 (%) 0.8–1.2 (%) 1.2–5 (%) >5 (%)
Agilent 6460 Triple Quadrupole LC/MS
pH 3
pH 6
Fig. 5. Influence of mobile phase flow rate (A.) and pH (B.) on the median sensitivity gain in HILIC vs. RPLC for Agilent 6460 Triple Quadrupole LC/MS system, in blue; AB Sciex API 5000TM LC/MC/MC System, in red; Waters Xevo TQ-S, in green; and Waters TQD, in purple.
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
transmission on the Waters Xevo TQ-S, leading to better ionization and ion transmission efficiencies. Finally, with the AB Sciex API 5000TM system (red), HILIC did not promote MS sensitivity. Indeed, whatever the mobile phase flow rate, no significant difference were observed in the median gain in sensitivity under HILIC conditions, rather providing a systematic lower sensitivity than those obtained under RPLC conditions. The influence of mobile phase pH on the sensitivity gain under HILIC conditions is shown in Fig. 5B. In agreement with previous results, the pH showed a negligible effect on sensitivity gain with the AB Sciex instrument (0.7-fold at both pH values). With Waters TQD and Xevo TQ-S, the sensitivity improvement was reported to be barely greater at pH 3 than pH 6, with a 1.9- vs. 1.2-fold and 3.7 vs. 3.6-fold enhancement at pH 3 and 6, respectively. On the other hand, the Agilent instrument was strongly influenced by the mobile phase pH, as illustrated by the 1.3- up to 4.9-fold median gain in sensitivity between pH 3 and 6, respectively. This gain in sensitivity in HILIC at pH 6 was largely due to the strong decrease of S/N ratios in RPLC conditions owing to poorer ionization efficiency at pH 6 for basic compounds than at pH 3. When considering the whole set of test compounds, the lowest sensitivity gains in HILIC were generally observed with salbutamol, pindolol, acetylmorphine and pseudoephedrine, while the highest improvements were mainly obtained for clonidine, adenosine, alprazolam, reserpine, bupropion and naloxone. Results for these highlighted compounds are in agreement with previous findings and confirm that the better desolvation in HILIC conditions is not the unique explanation for improved sensitivity, as previously discussed [13].
4. Conclusion This study highlights the crucial role played by the design of the ionization source on sensitivities obtained in HILIC-MS and RPLC–MS. Since modern ESI sources are able to efficiently desolvate highly aqueous mobile phases, even at high flow rates up to 1000 L/min, the differences in sensitivity observed between RPLC and HILIC are much more limited nowadays. Old-generation devices seem to be more prone to significant differences in sensitivity between HILIC and RPLC. For instance, a systematic larger gain in sensitivity in HILIC was reported for the Waters TQD instrument compared to the recent Xevo TQ-S, although both generations are equipped with a Z-Spray ionization source. In HILIC, the most significant gain was obviously observed at a high flow rate (1000 L/min), since the mobile phase was more efficiently desolvated than in RPLC. On AB Sciex API 5000TM system, HILIC mode was found to be of limited interest to improve S/N ratios. Depending on the conditions, a higher sensitivity was observed for 51–68% of the compounds in RPLC, while only 15–36% of the compounds showed a sensitivity improvement in HILIC. It is worth mentioning that the AB Sciex Turbo VTM source has been specially designed to maintain high sensitivity even under extreme conditions of flow rate as well as with highly aqueous mobile phases. Therefore, the impact of the mobile phase composition can be considered as negligible, offering similar desolvation efficiency in HILIC and RPLC. Finally, the Agilent 6460 Triple Quadrupole LC–MS system showed a strong influence of the mobile phase conditions and compositions on the overall measured sensitivity. Indeed, at a flow rate of 300 L/min and pH 6, 92% of compounds presented better S/N ratios in HILIC vs. RPLC, while this number dropped to only 7% at 1000 L/min and pH 3, which was merely explained by a significant increase of the background noise under HILIC conditions. Because a limited number of instruments were tested, the conclusions drawn in this study cannot be generalized to any entire brands of MS devices.
219
Acknowledgments The authors wish to thank Waters, and particularly Martin Rühl, Joel Fricker, Stéphane Canarelli and Marco Rentsch for the kind opportunity to use the Waters Acquity I-class and Xevo TQ-S instruments in their demo-lab in Baden-Daettwil, Switzerland.
Appendix A. Supplementary data Supplementary data associated with this article can be found, in the online version, at http://dx.doi.org/10.1016/j.chroma. 2014.06.066.
References [1] D.V. McCalley, Is hydrophilic interaction chromatography with silica columns a viable alternative to reversed-phase liquid chromatography for the analysis of ionisable compounds? J. Chromatogr. A 1171 (2007) 46–55. [2] J. Ruta, J. Boccard, D. Cabooter, S. Rudaz, G. Desmet, J.L. Veuthey, D. Guillarme, Method development for pharmaceutics: Some solutions for tuning selectivity in reversed phase and hydrophilic interaction liquid chromatography, J. Pharm. Biomed. Anal. 63 (2012) 95–105. [3] D. Guillarme, J. Ruta, S. Rudaz, J.L. Veuthey, New trends in fast and high-resolution liquid chromatography: a critical comparison of existing approaches, Anal. Bioanal. Chem. 397 (2010) 1069–1082. [4] A. Kumar, J.C. Heaton, D.V. McCalley, Practical investigation of the factors that affect the selectivity in hydrophilic interaction chromatography, J. Chromatogr. A 1276 (2013) 33–46. [5] H.P. Nguyen, K.A. Schug, The advantages of ESI-MS detection in conjunction with HILIC mode separations: fundamentals and applications, J. Sep. Sci. 31 (2008) 1465–1480. [6] W. Naidong, Bioanalytical liquid chromatography tandem mass spectrometry methods on underivatized silica columns with aqueous/organic mobile phases, J. Chromatogr. B 796 (2003) 209–224. [7] C.R. Mitchell, Y. Bao, N.J. Benz, S. Zhang, Comparison of the sensitivity of evaporative universal detectors and LC/MS in the HILIC and the reversed-phase HPLC modes, J. Chromatogr. B 877 (2009) 4133–4139. [8] C. Apostolou, C. Kousoulos, Y. Dotsikas, Y.L. Loukas, Comparison of hydrophilic interaction and reversed-phase liquid chromatography coupled with tandem mass spectrometric detection for the determination of three pharmaceuticals in human plasma, Biomed. Chromatogr. 22 (2008) 1393–1402. [9] W. Naidong, W. Shou, Y.L. Chen, X. Jiang, Novel liquid chromatographic-tandem mass spectrometric methods using silica columns and aqueous-organic mobile phases for quantitative analysis of polar ionic analytes in biological fluids, J. Chromatogr. B 754 (2001) 387–399. [10] M. Michi, L. Ferrari, Validation of a high-sensitivity assay for zatebradiWaters Acquityne in dried blood spots of human blood at pg/ml concentrations using HILIC-MS/MS, Bioanalysis 2 (2010) 1863–1871. [11] E.S. Grumbach, D.M. Wagrowski-Diehl, E.R. Mazzeo, B. Alden, P.C. Iraneta, Hydrophilic interaction chromatography using silica columns for the retention of polar analytes and enhanced ESI-MS sensitivity, LCGC North Am. 22 (2004) 1010–1023. [12] R. ChiritaTampu, C. West, L. Fougere, C. Elfakir, Advantages of HILIC mobile phases for LC-ESI-MS-MS analysis of neurotransmitters, LC–GC Eur. 26 (2013) 128–140. [13] A. Periat, J. Boccard, J.-L. Veuthey, S. Rudaz, D. Guillarme, Systematic comparison of sensitivity between hydrophilic interaction liquid chromatography and reversed phase liquid chromatography coupled with mass spectrometry, J. Chromatogr. A 1312 (2013) 49–57. [14] N. Gray, J. Heaton, A. Musenga, D.A. Cowan, R.S. Plumb, N.W. Smith, Comparison of reversed-phase and hydrophilic interaction liquid chromatography for the quantification of ephedrines using medium-resolution accurate mass spectrometry, J. Chromatogr. A 1289 (2013) 37–46. [15] I. Manisali, D.D.Y. Chen, B.B. Schneider, Electrospray ionization source geometry for mass spectrometry: past, present, and future, Trends Anal. Chem. 25 (2006) 243–256. [16] H. Stahnke, S. Kittlaus, G. Kempe, C. Hemmerling, L. Alder, The influence of electrospray ion source design on matrix effects, J. Mass Spectrom. 47 (2012) 875–884. [17] M. Rodriguez-Aller, R. Gurny, J.L. Veuthey, D. Guillarme, Coupling ultra highpressure liquid chromatography with mass spectrometry: constraints and possible applications, J. Chromatogr. A 1292 (2013) 2–18. [18] M. de Person, P. Chaimbault, C. Elfakir, Characterization of low-molecular weight peptides in champagne wine by liquid chromatography/tandem mass spectrometry, J. Mass Spectrom. 43 (2008) 204–215. [19] M. Holcapek, K. Volna, P. Jandera, L. Kolarova, K. Lemr, M. Exner, A. Cirkva, Effects of ion-pairing reagents on the electrospray signal suppression of sulphonated dyes and intermediates, J. Mass Spectrom. 39 (2004) 43–50. [20] H. Mei, Y. Hsieh, C. Nardo, X. Xu, S. Wang, K. Ng, W.A. Korfmacher, Investigation of matrix effects in bioanalytical high-performance liquid
220
A. Periat et al. / J. Chromatogr. A 1356 (2014) 211–220
chromatography/tandem mass spectrometric assays: application to drug discovery, Rapid Commun. Mass Spectrom. 17 (2003) 97–103. [21] C. Ghosh, C.P. Shinde, B.S. Chakraborty, Influence of ionization source design on matrix effects during LC-ESI-MS/MS analysis, J. Chromatogr. B 893–894 (2012) 193–200. [22] N.B. Cech, C.G. Enke, Practical implications of some recent studies in electrospray ionization fundamentals, Mass Spectrom. Rev. 20 (2001) 362–387. [23] http://www.absciex.com/Documents/Downloads/Literature/turbo-v-ionsource-operator-guide-eng.pdf, consulted in January 2014.
[24] A. Mordehai, J. Fjeldsted, Agilent Technical Note 5990–3494, 2009. [25] http://fscimage.fishersci.com/cmsassets/downloads/segment/Scientific/pdf/ Literature/LCMS OptimizeMobilePhase.pdf?&storeId=10652, consulted in January 2014. [26] A. Ekdahl, M.C. Johansson, M. Ahnoff, Tracing and separating plasma components causing matrix effects in hydrophilic interaction chromatographyelectrospray ionization mass spectrometry, J. Chromatogr. B 923–924 (2013) 83–91.