IMPROVEMENT OF MECHANICAL PROPERTIES OF ACRYLIC BONE CEMENT BY FIBER REINFORCEMENT* SUBR,ATA S-\HA+ and SUBRATA Biomechanics
PAL:
Department c7i Orthopaedic Surgery. L.S.U. 33931. Shrr\eport. LA 71130, U.S.A.
Laboratory,
Medical
Center.
P. 0.
Box
Abstract-Acrylic bone cement is signiticantl) weaker and less still’than compact bone. Bone cement is also weaker in tension than in compression. This hmits its use in orthopaedics to areas where tensile stresses are minimum. We have attempted to improve the mechanical properties of PMMA by reinforcing it with metal wires. and graphite and aramid fibers. Normal. carbon fiber reinforced and aramid fiber reinforced bone cement specimens were tested in compression. Addition of a small percentage I l-2 “” by weight for carbon and up to 6”” for aramid) of these fibers improved the mechanical properties signiticantly. Due to the improved mechanical properties of tiber reiniorced bone cement, its clinical use may reduce the incidence of cement fracture and thus loosening of the prosthesis
bone, carbon
INTRODUCTIOS
or other fiber reinforcement
may be used
in those cases. Self curing joint
acrylic
replacement
gical fractures the
though
cement
is not
bone cement of the joint
without
replacement
proved
as compact
bone even ~1 trl..
improvement
in the
achieved Trent, The
Like weaker
other
brittle
in tension
Therefore,
than
could
been used clinically (Scoville
vestigated
concrete.
incorporating
et al.,
1967;
Dunn,
cement to match
(Table
1).
compact
wire
re-
reinforcement
significantly
1977).
and showed increased
annular
in total joint space
We
that
the failure
replacement
between
the
(Wright
and
was to compare bone
the
cement
fiber reinforced
if it is possible
with
PMMA.
to reinforce
bone
and elastic properties
of
bone.
LITERATURE
in-
Mechanical
(Saha
points anical
stress
of self-curing
properties
properties physical
properties
This
is not of
bone
reported surprising cement
pressure.
also affects due
are
investi-
atfected
by
many
like: (1) propor-
(Haas et al.. 1975) and time and speed (Lee er
of handiing
The strain
the strength to
this table
by different
variables,
(2) mixing
al.. 1977); (3) method
on the mech-
However,
since the mechanical
and monomer
their compositions:
cement
view
scatter in the data on specific
and environmental
tions of polymer
without
information
of bone cement.
a considerable
mechanical gators.
Receired April 1981: in recisedjhz February 1981. *Presented in part at the 7th and 8th Annual Meetings of the Society of Biomaterials, held at Troy, New York. on 28-31 March, 1981, and at Birmingham, Alabama. on 27 April-l May, 1983; and at the 29th Annual Meeting of the Orthooaedic Research Society. held at Anaheim. Califcrnia, on 7-i0 March, 1983. _ tAuthor to whom reprint requests should be addressed. ZCurrent address: Department of Mechanical Engineering. Jadavpur University. Calcutta 700032, India.
bone cement
from various
(Saha and Pal. 19S4). Tabie 2 shows a summary
indicates
and
REVIEH
by many authors
of some of the published
\vire
due to the
prosthesis
properties
have been studied
in all these modes. As use of metal wire is
not practicable
study
of normal
the strength
of the cervical
and shear (Saha and Warman,
of PMMA
in
has also been
metal wires have
and Saha, 1977). bending
1978,1979)
1978a) properties
reinforcement
capacity
the effect of metal wire on the tensile (Saha
and Kraay,
and
also be im-
improvement
reinforcement
and aramid
is
metal
fiber
of PMMA
of this
properties
cement
the load carrying
in the stabilization
er al., 1976; Taitsman
narrow
bone
in compression
improve
of bone cement. PMMA
of PMMA
materials,
as in reinforced
inforcement
spine
objective
We also investigated most
et al., 1981).
could
by graphite
propertics
those of carbon
future.
(Robinson
that the compressive cement
fiber
strength
life and the fracture
1979~ b). Similar
by aramid
or carbon
tensile
1979: Saha cr (11.. 19Slb).
mechanical
bone cement in the
of
significantly
the mechanical
but also open up greater
for new applications
of bone
(Saha and Warman.
of bone cement. This will not only failures.
cement
we have shown
shear properties
the cement is
1975). Moreover,
a need for
Previously,
sometimes
it is used to replace bone (Crobvninshield
possibilities
the
of bone
1974). This suggests strength
that graphite
increases
toughness
bone
in tension
reduce bone cement
have shown
et ul., 1975, 1976; Ray e’f al.. 1974).
and leads to failure
mechanical
authors
reinforcement
PI al., 1978), the fatigue
For example,
25 Y, as strong
fiber
(Litchman
(Weber and Charnley, only
Other
patholowith bone
use of
complications. fractures
used in artificial
1970). Likewise.
and bony defects are repaired
cement (Harrington However,
is widely
cement
(Charnley,
and molding
with or
rate used during
and
its viscoelastic
stiffness nature
testing
properties (Saha
of
el al..
19Sla). From Table 1, one can get an idea of the ranges rather than the exact value of each parameter 467
reported.
SUBRATA SAHA and SUBRATA PAL
465
Table
I. Mechanical
CTS (MN m-‘) Industrial
PMMA
and prosthesis
properties of human bone. PMMA
68.9
2.1
30.8 ‘7.6 ‘5
2.0
uss
LCS
E* IGN m-l)
(MN
material
m-:)
(MN
Reference
m-?)
103
(injection molded) Orthopaedic PMMA (hand mixed)
Wet femoral cortical bone
123
66.3 77 17.2
166
Wet lumbar vertebrae (adult av.)
3.7
0.3-l
4.6
Wet cervical
3.1
0.55
10.1
vertebrae
29.7 41
Litchman et al. (19781 Saha et 01. (1976.19791 Lee et al. (1977)
84
Yamada
(1970)
Yamada
(1970)
Yamada
(1970)
2.8
(adult av.)
Stainless steel
1724
UTS = u, = ultimate tensile strength; strength; E = modulus of elasticity.
199.96 UC’S = ultimate
Again. these ranges do not reflect actual differences in the bone cement, but rather dilrerences which depend on the testing methods, speed of loading and some of the other variables mentioned before. Different investigators have studied the effect of various inclusions, e.g. blood (Lee et al., 1977; Helm, 1977);additives like antibiotic mixtures (Lautenschlager or al., 1976); effect of irradiation (Greenwald et al., 1977; Murray et al., 1974) and the effect of in km environment (Rostoker et al., t979; Weinstein n al., 1976). Table 2 depicts some of the variables, the details of which have been reviewed in a separate paper (Saha and Pal, 1984). 11 may be mentioned here that bone cement is the weakest link in any orthopaedic prosthetic joint. In order to limit the wide difference in mechanical strengths of bone, bone cement and prosthesis, many authors utilized the idea of fiber reinforcement of bone cement with various biocompatible and chopped fibers, e.g. glass (Litchman et al., 1977), carbon (Knoell er al., 1975; Pilliar et al., 1976; Saha and Warman, 1978b, 1979a,b), stainless steel (Fishbane and Pond, 1977), and aramid (Wright and Trent, 1979; Saha et al., 1981b) fibers. Fiber concentration, i.e. percentage by weight or volume, was chosen somewhat arbitrarily or it was mostly guided by the ease of mixing and handling during rapid polymerization of the cement. Most of the authors kept it limited to 1 or 2 w/o of the cement. Out ofthe various fibers mentioned, randomly oriented carbon fibers (or graphite) were used by the majority of the researchers. Table 3 summarizes the mechanica properties of various fiber reinforced PMMA as reported by different authors. Here also those discrepancies in mixing and handling parameters, rate of straining, and method of testing resulted in wide variation in the results of their tests. Similarly. Table 4 shows the test results of carbon fiber reinforced PMMA. The carbon fibers or graphite fibers have been used by most of the investigators because of its better biocompatibility and high strength and stiffness qualities. Most previous investi-
compressive
strength;
USS = ultimate
shear
gators used hand mixing in preparing fiber reinforced bone cement specimens. Uniform dispersion of fibers is difficult to obtain by such mixing. Therefore, some specimens might have contained bundles of fibers, thus diminishing the ultimate strength (Table 3). MATERIALS AND METHODS
We used chopped graphite fibers (6mm long, 8/1 diameter. a, = 43 GN m-‘. E = 3380GN m-’ Hercules A-S Type) hand mixed with PMMA powder, to prepare the carbon fiber reinforced PMMA samples. Similarly. chopped aramid fibers (Kevlar-29, E. I. DuPont Nemours & Co. 12-13mm long, density 1.44gcm-‘, 6, = 2758 MNm-‘) were used for preparation of the aramid fiber reinforced samples. Surgical grade PMMA powder and the chopped fibers were well hand mixed prior lo the addition of liquid monomer. The cylindrical compression specimens were 18 mm in diameter and 33 mm long and were cast in a cylindrical Teflon@mold as shown in Fig. 1, and pressed by hand to ensure compactness. The specimens were kept at room temperature for several days and then stored in water for 48 hr before testing them in compression using a servohydraulic Instron testing machine at a crosshead speed of 4.5 mm min - ‘. Compression tests were performed on control specimens of cement along and on three groups of test specimens of cement reinforced with 1% by weight of carbon fibers and 2 % and 4 % by weight of aramid, respectively. The number of specimens tested in each group is shown in Table 5. Before the compression test, the end faces of the cylindrical samples were checked for parallelism and were greased to reduce the friction between the sample and the plantens of the Instron. RESULT’S
Typical compressive stress-strain curves of normal, carbon fiber reinforced (Hercules AS type chopped fibers) and aramid fiber reinforced bone cement
Improvement
of mechanical
properties
469
of acrylic bone cement
4 f ij
x
Fumich er al. (1979)
Sih er 01. (1980)
Rostoker YI al. (1979)
Kusy (1978)
Holm (1977)
Lautenschlager and Marshall (1976)
_-
Reference
1.21
21.1
46.2 + 3.5 46.2k4.1 48.2 f 2.8 33.8 f 3.5
42.1 f 5.5
29.18+
Tensile
Ultimate
98.5 95.3
80.66 + 5.10
Compressive
60.5 4-l. 1
48.3 48.8 50.9 40.6
58.2 53.0 62.6
+ + + +
3.8 4.9 5.7 .60
Flexural
stress (MNm-*)
Table
Shear
2. (continued)
2324
2756 k 261 X + 2549 f 2411+ 2549 f
2480 2428 2265
(MN
m-l)
138 (Tension) 69 (Tension) 275 (Tension) 69 (Tension) 275 (Tension)
Elastic modulus ____-.___-
5bcatmin’ 120beatmin-’
Surgical Simplex
_____
The bone cement were implamed and tested slier 1 day, 6, 12, and 24 monlhs
CMW Palaces R hktCOS RG Sulphix 6 Surgical Simplex
CM W ccmcnt Simplex Palaces R
Strain rate = O.OZ/mi.
Commcnl
___-
Surgical Simplex
Aramid Kevlar-29 El Dupoint
Aramid Kelvar-29 EI Dupoint
Glass fiber
Stainless steel AlSl-316
Type
*All length and diameters are in mm. tThese are fractures toughness in Mnm-“’ 0, = maximum stress in tension; ran = maximum v/o = percentage by volume.
Saha and Pal (1981)
rf al.
PMMA (KerrSybron Corp)
Simplex
Litchman (1978)
Wright (1979)
Surgical grade PMMA
Fisbane and Pond (1977)
er ul.
type
Author
Cement
-
13
stress in compression;
I=
r = maximum
2. w/o 4. w/o
PMMA
79.39 + 16.74 85.19+6.95
25:; increase
id
1.88t 2.3 1 2.85
Other
of elasticity;I =
(4
Strength (MPa) Compressive Shear
of reinforced
stress in shear; E = modulus
36. I 38.20 42.8
1. w/o 4 w/o 7 w/o
I=
13
41.7 54.4
(0,)
4.5, v/o 8.6 v/o
7, used
Tensile
properties
I = 12.5 I/d = 5000 bu =2Gnm-’
short fiber I=O.S to I* area = 31.75 x 10T4 mm*
Reinforcing material Prc&ties
Table 3. Mechanical
of Elasticity MPa _
length: tl = diameler;
Comment
percentage
~~_~...
-_
by weight;
- ~.
(r was measured a; 4” ,, strain
32 y,, increase in 6, 74 9” increase in fracture toughness
Ram Speed, 2.5 mm min _ ’
W/O =
______~___
l880+ 180(c) I556 _+240(c)
324 1 (T) 3482(T)
Modulus
P
Cement
Osteobond
CMW
CM W-Porous
Simplex P RG.
Surgical simplex (RO)
Author
Knoell rr al. (1975)
Pilliar et ul. (1976)
Pilliar cf crl. (1976)
Saha cr ul. (1979)
Litchman er ul. (1978)
Graphite ThorneI-300 Union (Carbide)
Chopped Carbon Hercules As-type
Carbon
Carbon
Graphite GY-70
Type
I. and 6.6, w/o
0.53, v/o 1.05, v/o 1.57, v/o
I = 12.5 d = 5.0
I and 2 v/o
2 v/o
1,2,3 and 10 by w/o
y0 used
i=6mm d = 8Itm
/=6mm d = lfkl5prn E = 380 to 460GPa
/=6mm d=7flm
/=6mm chopped
Fiber Property
34.1 37.8 42.2
38
Tensile (e,)
74 72
1.7
35 47
Strength (MPa) Compressive Shear (%) (r)
Table 4. Mechanical properties of carbon fiber reinforced PMMA
47
Flexure (er)
4008(T) 5730(T)
3241(T)
Static: 3700 5800 Cyclic: I:’ = 4000(7’) E = 375oc)
5560(T)
4600(c)
Modulus of Elasticity MPa
Crosshead speed = 2.5 mm min _ ‘.
a, t = 5 x lo-‘*s-’ at E = 2 x 10-zS~’
607; increase in u,, 100‘,, increase in E
Comment
Fig. 1. Normal (left) carbon fiber reinforced (right) bone cement specimens with their Teflon@ molds. The arrow indicates recessed aluminium bottom plate for facing the molds.
473
Fig. 5. Electron photomicrograph of fracture surface of a CFR-PMMA
474
sample failed in compression
Table 5 Fiber percentage Sample group
Type of fiber
Sum&r
thy weighti
rpximens
I
x
(normal bone cement) 7
I (’II
Graphite
;
Aramid
1 /b - ‘3
4
Aramid
-l “,,
COMPRESSIOR
TEST
s 6
6
spccimsw
are sho\
in Fig. 2. As portrayed
figure, the stress-strain 1oc) -
of tested
2 ‘t, ARAMID
relationships
in this
were linear up to
a fairly
high stress level and there were no sharp yield
points.
However.
with increasing strain, there was a
decrease in the load carrying capacity of normal
and
carbon fiber reinforced bone cement beyond the initial peaks. but this was not true reinforced
PMMA
for the aramid
different
failure
samples.
For instance. while carbon
mechanisms
of these three groups
specimens developed longitudinal aramid
fiber
(Fig. 2). This may be due to the
fibers increased
of
fiber reinforced
cracks, samples with
uniformly
in diameter
with
Increasing compression. The means and standard ive properties
0
,.
specimens
7
05
are compared
deviations and carbon
of thecompressfiber
reinforced
in Figs 3 and 4. Since there
was no sharp failure point. the ultimatestrength
0 10
stress)
STRAIN
of
aramid
calculatedat Fig. 2. Typical stress-strain behavior of normal, carbon fiber (1 w/o) and aramid tibcr (2 w/o) reinforced
of aramid
PMMA
4”,strain,asat
failed
reinforced
COMPRESSION PMMA
limit for I ‘I,, by
TEST
0
2% ARAMID
I
4% ARAMID
0 IJ 004
T
90 80
T
704
T
PROP
Fig. 3. Comparison
LIMIT
ULT
were
theCF-
in most of the cases.
0
1% CARBON
(proof
specimens
rhis level ofstrain,
As shown in Fig. 3, the proportional
PMMA.
NORMAL
fiber
STRENGTH
of (means and 1 S.D.) proporuonal limit and ultimatestrength PMMA and 2”, and 4”” AFR-PbfMA.
of normal.
I I’,,CFR-
SUBRATA
476
SAHA
and SLBRATA PAL
possible with normal bone cement. In order to examine the mechanism of strength improvement. the fracture surfaces of CFR-PMMA were examined in a scanning U 4% ARAMID 1800 electron microscope. 1600 Figure 5 shows the electron photomicrograph of a fracture surface of a specimen failed in compression. It 1400 reveals that even after failure of the cement, some of F1 E 1200 the fibers are intact and still can carry some load. The 2 failure mechanism was the interracial shear of the fiber I 1000 and cement. This shows that the energy absorption 800 capacity of CFR-PM MA could be increased due to the 600 fiber pull outs. A high rise in temperature during the setting of bone 400 cement may cause tissue necrosis (Jefferiss et al., 1975) 200 and thus reduction in peak temperature is highly desirable in the clinical use of bone cement. We moniMOO. ELASTICITY tored the temperature rise in similar sized cylindrical IN COMPRESSION specimens for normal, carbon fiber reinforced and aramid fiber reinforced specimens and found that the Fig. 4. Comparison of modulus of elasticity of I r, CFRPMMA and Z”,, and 4’” AFR-PMMA. peak temperature of bone cement could be significantly reduced by the use of fiber reinforcements. This may be a highly beneficial side effect of using fiber weight of carbon fiber reinforced samples is comreinforcement in the orthopaedic use of bone cement. parable to 21:” by weight of aramid fiber reinforceThe improved mechanical properties of PMMA ment. However, the workability of 29, aramid fiber reinforced with ultra high strength graphite fiber, reinforced PMMA was much better than similar 1To Thornel-300 (Union Carbide, Inc.), or aramid fiber, carbon reinforced cement. When normal bone cement Kevlar-29, are compared with those of human compact (radiopaque) specimens were tested in compression, bone in Fig. 6. About 2 ‘:I,,Thornel-300 fibers improved they failed at an ultimate strength of 66.3 + 6.6 MPa. the tensile strength of PMMA to about 50% and Thus, 1 Oncarbon and 2 Y,,aramid fiber reinforcement modulus of elasticity to about 40”, that of compact improved the ultimate strength by 20.5 “” and 19.5 ‘:,, human bone. These data on AFR-PMMA were obrespectively. Four percent by weight of aramid fibers tained from Wright and Trent (1979) and on CFRimproved the ultimate strength by 28.7”“. The moPMMA from Litchman er al. (1978). dulus of elasticity obtained by 2 I:,, aramid fibers was Our result suggests that although significant imI1 Y0greater than that obtained by 1 O0carbon fiber provement in the mechanical properties of bone reinforcement (Fig. 4). cement can be achieved by fiber reinforcement, it may not be possible to match the properties of compact bone by such fiber reinforcement. Further improveDISCUSSIOS AKD CONCLUSIOS Cl 1% CARBON
zoo0 -
12%
ARAMID
The results of mechanical testing of graphite and aramid fiber reinforced specimens show that incorporation of such fibers produces a significant increase in the compressive strength of bone cement. As shown in Table 1, we have also shown previously that the shear strength of PMMA could be significantly improved by graphite fiber reinforcement (Saha and Warman, 1978a, b, 1979a, b). The modulus of elasticity of fiber reinforced PMMA also increases with increased fiber content when the volume percent is kept small. Other investigators have shown before that the tensile and fatigue strength, and fracture toughness of bone cement can be improved by graphite and aramid fiber reinforcements (Litchman rr al., 1978; Pilliar et al., 1976, and Wright and Trent, 1979). Recently we have also demonstrated that creep deformation of carbon fiber reinforced bone cement was significantly less than that of normal bone cement (Saha and Pal, 1982). The improved mechanical properties of graphite fiber reinforced PMMA would allow the clinical use of this material in more diverse applications than is presently
TENSILE
TEST
150I-
140
120
16000
v/o 0 THORNEL-300.1.05 I THORNEL-300, 2.08 v,c 0 ARAMIDI w/o 0 ARAMID- 4 w/o 0 Human Bone
NE z‘ L 100 E
14000
12000,
80
8000
SO
6000
40
4000
20
2000
6 E
E
9 10000 E
ULT. STRENGTH
z 0 f u) 5 =
MODULI ELASTI
Fig. 6. Comparison of ultimate tensile strengths and mod&i of elasticity of CFR-PMMA, AFR-PMMA and human compact bone.
Improrement of mschamcal properties of acrylic bone cement ment in the mechanical properties of bonecement may be achieved by the use of different types of fibers and by a more uniform mixing of the fiber and cement. A machine-mixed carbon fiber reinforced bone cement (Zimmer, Inc.) in which the fiber cross-sections are dog bone shaped for better bonding, is presently being used for limited clinical trial. We are in the process of testing this CFR-PMMA mechanically and our preliminary result indicates somewhat higher strength values compared to those reported in this paper.
Mechanical properties of bone cements conrammg large doses of antibiotic powders. J. bwmrd. .!far. Rrs. 10, 929-938.
Lautenschlager. E. P. and Marshall, G. W. (19761 bfschanical strength of acrylic bone cements impregnated Hith antlbioti&. J. biomid. Mat. Rrs. 10. 837184% Lee. A. J. C.. Line. R. S. M. and Vaneala. S. S. 119771 The _ mechanical properties of bone cements. J. .ift~l. Etlyny Tech. 2. 137-140. Lee A. J. C., Ling, R. S. M. and Wrighton, J. D. (19’3) Some properties Of polymethlmethdcrylatc with reference to Its use in orthopaedic surgery. C/in. Or~hop. 95. 2s I -287. Lida. M.. Furuya, K., Kawachi. S., Masuhara, E. and Tarumi. J. (1974) New improved bone cement (MMA-TBB). Cfin. Orrhop.
REFERENCES
Astleford. W. J.. Asher, M. D.. Lindholm. U. S. and Rockwood, C. A. (1975) Some physical and mechanical factors affecting ihe shear strength of methylmethacrylate. Cim. Or&p. 108, 145-148. Charnley, J. (1970) Acrylic Cement in Orrhopuedic Suryer)‘. Williams and Wilkins, Baltimore, MD. Crowninshield, R. D.. Pope, M. H. and Hoaglund, F. T. (1974) A comparison of the mechanical properties of bone and polymcthylmethacrylatc. J. Botre JI Surq. S6A. 865. DeWijn. J. R., Sloff. T. J. J. H. and Driessens, F. C. M. (1975) Characterization of bone cement. Acta orthop. stand.. 46, 38-51. Dumbleton, J. H. and Black, J. (1975) An Infroduciion fo C. Thomas, Orrhopuedic ,%.IuferiuIs p. 185. Charles Springfield, IL. Dunn, E. J. (1977) The role of methylmethacrylate in the stabiliz;ltion and replacement of tumors of the cervical spine. Spine 2, 15-24. Eftekhar, N. S. and Thurrton. C. W. (1975) Effect of irradiation on acrylic cement with special reference 10 fixation of pathological fractues. J. Biomrchonirs 8.53-56. Fishbane. B. M. and Pond, R. B. (1977) Stainless steel fiber reinforcement of polymethylmethacrylate. C/in. Orlhop. Ret. Rrs. 128. 194199. Fumich. R. M. and Gibbons, D. F. (1979) Rate of mixing and rhe strength of methylmethacrylate bone cements. Or~hop. RzL’. 8, 414. Greenwald, A. S., Wilde, A. H. and Matejczyk, M. B. (1977) Chnical applications and properties of acrylic bone cement. Orrhop. Diy. 5, 2, 16. Haas. S. S., Brauer. G. M. and Dickson, G. (1975) A characterization of polymethylmethacrylate bone cwment. J. Bonr JI Sury. 57A, 380. Harrington, K. D. (1975) The use of methylmethalcrylate as an adjunct in the internal fixation of unstable comminuted intertrochanteric fractures in osteoporotic patients. J. Bone Jf Sury. 57A, 744-750. Harrington. K. I)., Sim, F. H.. Enis, J. E., Johnston, J. O., Dick, H. M. and Gristina. A. G. (1976) Methylmethacrylate as an adjunct in internal fixation of pathological fractures. J. Bonr Jr. Sury. 58A, 1047-1055. Holm. N. J. (1977) The modulus of elasticity and Rexural strength of some acrylic bone cements. Acru orrhop. stand. 48, 436-442.
Jefferiss, C. D., Lee, A. J. C. and Ling, R. M. S. (1975) Thermal aspects of self-curing polymethylmethacrylate. J. Bone Jr S&y. 57B, 5ll-518.Kellv. A. and Davies. G. J. 119651. The DrinciDles of fiber reinforcement of mitals. %irral/. Rrc. lb, I. ’ Knoell. A., Maxwell, H. and Bechtol, C. (1975) Graphie fiber reinforced bone cement. Ann. biomrd. Enyng. 3, 225-229. Kusy, R. P. (1978) Characterization of self-cuing acrylic bone cements. J. biomrd. Mut. Rrs. 12, 271-305. Lautenschlager, E. P., Jacobs, J. J. and Marshall, G. W. (1976)
177
Rrl.
Rrs.
100.
279-286.
Litchman, H. M., Richman. M. H., Warman. M. and Mitchell. J. (1978) Improvement of the mechanical properties of polymethylmethacrylate by graphite tiber reinforcement. Transuct ions o/’ fhr Orfhopurdic R twurch SOWI?. Vol. 2. p. 86. Litchman. H. M.. Richman. M. H.. Warman. M. and Mitchell, J. (1977) Improvement of mechanical properties of PMMA by glass tiber reinforcement. Unpublished work. Morrison, J. (1968) Analysis of force actions transmitted by knee joint. J. biomed. Enyy. 3. 164-170. Murray, J. A., Bruels, M. C. and Lindberg, R. D. (1974) Irradiation of polymethylmethacrylate. J. Born,, Jr. Sury. 56A, 31 I-312. Pal, S. and Saha, S. (1982) Stress relaxation and creep behavior of normal and carbon fiber bone cement. Biomutrriuls
3. 93-96.
Pilliar, R. M. and Blackwell, R. (1976) Carbon fiberreinforced bone cement in orthopaedic surgery. J. biomrd. Mul.
Rcs.
10, 893-906.
Pilliar, R. M., Bratina, W. J. and Blackwell. R. (1976) Fatigue studies of carbon fibre reinforced polymethylmrthacrylate for surgical implant applications. Presented dt ASTM Filamentary Composites. Denver, Symposium on November 1976. Ray, A. K., Romine. J. S. and Pankovich, A. Xl. (1974) Stabilization of Pathogenic Fractures with Acr>hc Cement. C/in.
Orlhop.
202,
182-185.
Robinson, R. P.. Wright, T. M. and Burnstein, A. H. (1981) Mechanical properties of poly(methyl methacrylate) bone cements. J. biomrd. hful. Rrs. 15, 203-208. Rostoker, W., Lereim, P. and Galante, J. 0. (1979) EtTect ofan in-cico environment of the strength of bone cement. J. biomrd. Mar. Rrs., 13, 365-370. Saha, S. and Kraay. M. J. (1978) Flrxural properties of acrylic bone cement reinforced with stainless steel wires. Proceedings olfhr 3 Isf Annual Confirmcr on Enyinrrring in Medic& and Eioloyy, 19, 55. Saha. S. and Kraay, M. J. (1979) Improved strength characteristics of polymethylmethacrylate beam specimens reinforced with metal wires. J. biomrd. ,\ful. Res. 13.443457. Saha, S., Pal, S. and Albright, J. A. (198la) Time dependent mechanical properties of normal and carbon fiber reinforced bone cement. 7run.sucrron.s o/’ Ihe 27111 Annuul Conjrrencr oJ thr Or~hopu~‘dic Rtwurch Socrrlj,. Vol. 6. p. 296, ORS. Chicago. Saha. S., Pal. S. and Albright, J. A. (1981bl Improved properties of aramid fiber reinforced polymethglmethacrylate.Transactions of 7th Annual Biomarrrials, Vol. 4, p. 21.
Mrrring
of Ihe SocierJjor
Saha, S. and Pal, S. (1982) Effect of strain rate on mechanical properties of normal and carbon fiber reinforced bone cement. Transaction of the 7th Annual Meeting of the Socirfy jar Biomareriats. p. 1. Saha. S. and Pal. S. (1984) Mechanical properties of acrylic bone cement: a review. J. biomrd. .ifur. Rrs. 18, -135-462. Saha, S., Taitsman, J. P., Johnson, T. R.. and Albright, J. A. (1976) Metal reinforced bone cement I: tensile behavior. Proccedinys 01 rhr 4rh Nr,v Enylund Biornyinwriny
SIJBRATASAHA and SUBRUA PAL
478
Confirrnce (Edited by Saha, S.) pp. 105-108. Pergamon Press, Oxford. Saha. S. and Warman. M. L. (1978a) Shear strength of metal wire reinforced acrylic bone cement. Proceedings ofrhe 6th New England Bioenyinreriny Conference. pp. 306-308. Pergamon Press, Oxford. Saha S. and Warman, M. L. (1978b) Graphite fiber reinforced polymethylmethacrylate as a biomaterial. Proceedings of fhe
61h
Nrw
England
Biocnginrering
Confrrence
pp. 31 l-3 12, Pergamon Press, Oxford. Saha. S. and Warman, M. L. (1979a) Improved mechanical properties of graphite fiber reinforced polymethylmethacrylate. Transactions of the 1 Ilh International Biomaferiak Symposium and 5th Annual Meeting o/ the Society of Biomuferiak, Vol. 3, p. 86. Saha, S. and Warman, M. L. (1979b) Compressive and shear properties of graphite fiber reinforced bone cement. Orthop. Transaclions
3. 169.
Scoville, W. B., Palmer, A. H., Samra. K. and Chong,G. (1967) The use of acrylic plastic for vertebral replacement or fixation in metastatic diseaseof the spine. J. Nnlrosurg. 27, 274-279. Sih, G. C. and Berman, A. T. (1980) Fracture Toughness
Concept Applied to Methly Methacrylate. J. biomed. Mat. Rex 14, 311-324. Stubbs, B. E., Matthews, L. S. and Sonstegard, D. A. (1975) Experimental fixation of fractures of the femur with methylmethacrylate. J. Bone Jt Surg. 57A, 317-321. Taitsman, J. P. and Saha, S. (1977) Tensile strength of wireeinforced bone cement and twisted stainless-steel wire. J. Bone JI Surg. 59A, 419-425. Weber. F. A. and Charnley, J. (1975) A radiological study of fractures of acrylic cement in relation to the stem of a femoral head prosthesis. J. Bone Jr Surg. 57B, 297301. Weinstein, A. M., Bingham, D. N., Sauer, B. W. and Lunceford. E. M. (1976) The effect of high pressure insertion and antibio inclusion upon the mechanical properties of polymethlymethacrylate. C/in. Orrhop. 121, 67-73. Wilde, A. H. and Greenwald, A. S. (1975) Shear strength of self-curin acrylic cement. C/in. Orrhop. Rel. Res. 106. 126. Wright, T. M. and Trent. P. S. (1979) Mechanical properties ofaramid fibre reinforced acrylic bone cement. J. Mar. Sci. 14, 503-505. Yamada. H. (1970) Strength oj’Eio/ogical Mawials. Williams & Wilkins, Baltimore.