Electrophysiology
Inappropriate therapy from atrial fibrillation sinus tachycardia in automated implantable cardioverter defibrillators Kumar Nanthakumar, MD,” Miney Paquette, MSc,= Bruce Gunderson, MSC,~ Jim Gilkerson,
MSc,= David Newman, PhD,e Mary Greene,
Toronto,
St Paul,
Ontario,
Canada;
Background [ST])
Inappropriate
in patients
algorithms
with
from
all episodes
of inappropriate Therapy
electrograms,
history,
sensitivity
and
turer
receiver
and
Results available.
specificity
episodes
of ST from
facturers
at about
caused
and
for VT. We 19 patients. An onset
62
value
(atrial
Sunnyvale,
fibrillation
challenge.
We
therapy
delivered
because
tachycardia
and
(29%)
received
(VT),
inception.
tested
Calq
[AF] the
and
sinus
tachycardia
performance
of AF and
ST, or AF from By using
of onset
was
inappropriate for AF,
19 for
of VT from
in the clinically
of 80%
study percentage
therapy
83 episodes
Specificity,
and
of stability
an onset
algorithm
for
30
values
review
to 60
were
of stored
R-R intervals
calculated
for
intracardiac before
each
therapy,
manufac-
generated.
received
analyzed
Texas;
tachyarrhythmias
before
values
curves
1 patients
Angleton,
is a major
by ventricular R-R data
of stability
monitored, 2
MD,” D. Curtis Deno, MD, PhD,b Lisa Malden, MSC,~ Denis Heng, BSC,~ and Paul Dorian, MDa
by ST.
as caused
characteristic
patients,
40%.
of inappropriate
a family
Of the 2 17 patients therapy
for episodes
information,
operating
Of the 40
appropriate
for
supraventricular defibrillators
classified
clinical
from
Minn;
cardioverter therapy
was
and
therapy
implanted
3 manufacturers
Methods
rithm
Minneapolis
and
relevant
associated
with
therapy,
18 patients,
94
sensitivity
range
91%
sensitivity
and
1 patient
ST, and
episodes
had
for
noise;
complete
95%
was
R-R information
15 (38%)
of AF from
of 295%, and
40
comparable
specificity
also
1 patients,
2
received and
across
for the specific
56 manualgo-
tested.
Conclusions
Inappropriate
stability
used
algorithm
algorithms accurately
therapy
to differentiate
differentiates
is a common AF from
ST from
problem
VT was
VT. (Am
less than
Heart
Submitted August 16, 1999; accepted December 3, I999 Reprmt requests: Paul Dorion, MD, Division of Cardiology, St Mlchoel’s Hospital, 30 Bond St, Q-705 1, Toronto, Ontario, Canada, MSB I W8. E-mall.
[email protected] co Copyright 0 2000 by Mosby, Inc. 0002.8703/2000/$ I2 00 + 0 4/l/104757 doi: IO. 1067/mh~.2000 104757
ideal,
though
cardiac comparable
defibrillators. across
The
performance
manufacturers.
of the The
onset
J 2000;139:797-803.)
It is estimated that more than 50,000 cardiac defibrillators (ICD) were implanted worldwide in 1997.’ Although ICDs are very accurate and effective in the diagnosis and treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), inappropriate therapy from supraventricular tachyarrhythmias is a major challenge and is reported to affect 20% to 30% of patients.z-9 Atria1 fibrillation and sinus tachycardia are the most common reasons for inappropriate delivery of therapy.” It is estimated that approximately 12% of all patients with an ICD will receive inappropriate therapy because of atria1 fibrillation,x~” and in those with a prior From 0St Mlchoel’s Hospifol, Toronto; bfnfermedrcs Inc. Angleton; cSl Jude/Venlritex Inc. Svnnyvole; dMedkonic Inc. Mmneopohs; and eGwdonf Inc. St Paul.
in implantable
history of atria1 fibrillation, twice as many will receive inappropriate therapy.‘O The consequences of inappropriate therapy include pain and its psychological sequelae (particularly anxiety), induction of ventricular arrhythmias, decreased battery longevity, and rarely, fatal outcomes. l l-l2 The primary detection criterion in all current ICDs is heart rate (R-R intervals). To decrease the frequency of inappropriate therapy, several detection criteria beyond R-R interval alone may be available in single-chambered tiered therapy (ICDs with antitachycardia pacing and low energy cardioversion capability) devices. These algorithms are based on R-R interval pattern or QRS morphologic criteria. Those relying on R-R intervals include sudden onset, stability, and sustained rate duration. The most commonly used algorithms rely on the expectation that atria1 fibrillation is more irregular than VT, with rate stability programmed to withhold therapy for rhythms that exceed the rate cutoff for tachycardia detection but exhibit R-R variability exceeding a pre-
Amerox
798
Nonthokumor
et al
specified value. Similarly, rhythms with a “Sudden onset” are thought to be likely caused by VT and those with more gradual onset by sinus tacliycardia5.‘3,1$ therapy can be withheld if onset is more gradual than the programmed value. These detection algorithms are, however, not programmable in the VF zone. However, clinicians have been hesitant to enable these functions because of the risk of withholding therapy for VT.‘5-17 Studies to date on the perfomiance of onset and stability algorithms have primarily been based on atrlal fibrillation induced at electrophysiology study and sinus tachycardia and atria1 fibrillation from Holter monitors.1j.14 There are few studies that prospectively evaluated the performance of single algorithms in a small number of inappropriate therapies.21*z3 The irregularity noted at the onset of spontaneous episodes of VT and the irregularity of polymorphic VT may result in overlap of stability values with atrial fibrillation with rapid ventricular response.5,17 Sinus tachycardia that leads into VT may have an overlap of onset values with VT. We therefore tested the hypothetical performance of stability algorithms from 3 different ICD manufacturers for episodes of inappropriately delivered therapy from atrlal fibrillation and performance of a particular onset algorithm for episodes of inappropriate therapy caused by sinus tachycardia and assessed their accuracy in detecting spontaneous VT ln the same group of patients.
Methods Patient population All patients who had ICDs with data-logging capability (II = 217) implanted between 1990 and 1998 at St Michael’s Hospital were included in the study. AI1 had single-chamber, third-generation tiered therapy devices with intracardiac electrocardiographic storage from 1 of the 4 major manufacturers (Guidant Inc, St Paul, Minn; Intermedics Inc. AngIeton. Texas; St Jude/Ventritex Inc, Sunnyvale, Calif; and Medtronic Inc, Minneapolis, Minn). AII tachycardia episodes leading to device therapy had been prospectively classified as appropriate or inappropriate by the 2 attending electrophysiologists and personnel in the ICD clinic. Inappropriate therapy was defined as therapy for other than VT or VF from a review of stored electrograms, R-R interval data, history, and clinical data for each episode. Ail patients who received any inappropriate therapy were included for analysis, whether or not they had received appropriate therapy for VT or VF episodes. Arrhythmia detection Arrhythmia detection was based on the evaluation of consecutive R-R intervals. The detection criterion is fulfilled in some devices if x/y (for example 8/ 12 or 12/ 16) R-R intervals (range 8 to 16) are shorter than the programmed value and in other devices if a certain number of consecutive beats have cycle length shorter than the programmed value. After this criterion has been met, some devices recont3-m the presence of arrhy-thmia while preparing for therapy, whereas others start therapy as soon as criteria have been met. None of the patients had additional discriminators programmed such as
Heart Journal May 2000
stability or onset. However, all patients had programmed sustained rate durdtion (when available). Sustained rate durdtion is a function of the ICD that triggers therapy after a certain period of progrdmn%IbIe time has elapsed during a tachycardia event independently of the classification of that rhythm. Study design We reviewed data logs of all our ICD patients and collected cIinicaI information and demographic data. Patients who had received inappropriate therapy were identified. Inappropriate therapies were classified as caused by atrial fibrillation, sinus tachycardia. or noise. The R-R intervals for all episodes leading to inappropriate therapy for atrial fibrillation or sinus tachycardia and R-R intervals of all episodes of true VT in these patients were retrieved from the data logs. Depending on the manufacturer, between 26 to 40 R-R intervals preceding device declaration of VT and 8 to 20 additional intervals before delivery of therapy were recorded. The data were stored in a Microsoft Excel database (Microsoft Inc. Redmond, Wash) and analyzed by a customized program designed to calculate a stability and onset value for each of the different algorithms. Stability criterion The performance of stability algorithms from 3 different manufacturers (Guidant, Intermedics, and St. Jude/Ventritex) in distinguishing between ventricular and supraventricular arrhythmias was tested with the stored R-R data from the episodes of detected tachycardia. The Guidant stability algorithm This algorithm measures differences between consecutive R-R intervals after the rate criterion has been satisfied. The stability of the rhythm is analyzed by the variance in cycle length. The weighted average variance (the recent R-R intervals contribute more than the remote R-R Intervals) is calculated after every newly detected R-R interval. This current weighted average difference in consecutive R-R intervals is compared with the programmed stability parameter. If the value is equal to or less than the programmed value, the rhythm is thought to be stable and therapy is delivered. However, if the calculated vziability value is greater than the programmed value, the rhythm is deemed unstable and therdpy is withheld. The Intermedics stability algorithm In this algorithm a ventricular rate criterion operates in parallel, although not independently because a single long R-R interval out of the tachycardia zone resets a stability counter. Each new R-R interval contributes to a running assessment of the longest R-R (R-Rmax) and shortest R-R (R-Rmin) vaIues over the beats since a reset. Evidence for stability is accumulated with each successive R-R interval that falls in the tachycardia zone and that does not result in the R-Rmax/R-Rmin difference exceeding a programmed stability parameter. When the stabilit) evidence counter exceeds the programmed value for a specific number of consecutive intervals, then the rhythm is judged both stable and fast enough for VT therapy to be delivered. The Ventritex stability algorithm The measured interval stability value is the difference between the second longest and second shortest intervals in
Amer,con Heart Journal Volume 139, Number 5
Figure
Nonthokumar
et al 799
1 Dlrtributlon
of Stability
Values
20
q VT slab n AF stab
16 16 f
'4
5 0 s
10
j
8
z'
6
12
4 2 0
J 25-30-35-40-45-60-5s-60.c70-75-60-&5. 30354045505560a70758-38590
Stability
Distribution
of stability
of 5 ms; y-axis, significant
overlap
number
values
for VT and
of episodes
in stability
values
atrial
of VT (open between
fibrillation bars) atrial
for and
fibrillation
a single
atrial
manufacturer.
fibrillation and
(solid
x-axis, barsj.
Stability
Histogram
values
in bins
demonstrates
VT.
a recent group of intervals defined by the programmable interval stability window. If the measured difference is greater than or equal to the programmed stability delta, the discriminator indicates SVT and therapy is withheld.
manufacturer and an ROC curve was generated. Sensitivity refers to the proportion of VT episodes correctly classified as VT, and specificity refers to the proportion of sinus tachycardia episodes correctly classified as sinus tachycardia.
The onset criterion We tested the performance of the onset criterion from one manufacturer (Medtronic). The onset criterion is a feature that inhibits the ICD from detecting episodes of sinus tachycardia when the ventricular rate undergoes a gradual increase and enters the VT zone. If an R-R interval is in the VT zone but the rhythm does not satlsIy the onset criterion, that cycle is marked as a normal sinus beat and the VT counter resets. After a sensed event the ICD compares the most recent 4 intervals to the 4 that preceded them. The criterion is fulfilled if the avenge of the most recent 4 intervals is less than the average of the 4 intervals that preceded them by a programmable percentage (sudden onset delta percentage).
Results
Data analysis The sensitivity and specificity for the detection of VT, as opposed to atria1 fibrillation, were retrospectively calculated for a wide range of different (programmable) stability values. Sensitivity refers to the proportion of VT episodes correctly classified as VT, and specificity refers to the proportion of atria1 fibrillation episodes correctly classified as atrial fibrillation. Rrceiver operating characteristic curves (ROC) were generated for stability values starting at 5 ms and every 5 ms increment thereafter. This analysis was performed separately for the 3 different algorithms from the 3 different manufacturers. Sensitivity and specificity in the detection of VT as opposed to sinus tachycardia were retrospectively calculated for a family of programmable percentage onset values for the single
Patient
characteristics
Of 2 17 patients, 62 (29%) received inappropriate therapy, of whom 40 patients had complete R-R information available for analysis. Of these 40 patients, 15 (38%) also received therapy for appropriate reasons for VT. Their mean age was 63 f 11 years, and 75% were male. The mean ejection fraction was 33% f 15%, and patients received a mean of 4.1 f 2.4 inappropriate therapies. In this population of 40 patients, 21 patients received therapy for atria1 fibrillation, 19 for sinus tachycardia, and 1 for noise (1 patient received therapy for both atrial fibrillation and for sinus tachycardia). We analyzed a total of 183 episodes of VT from 18 patients (15 patients who had also received inappropriate therapy and 3 patients who had only VT without other inappropriate therapy), 94 episodes of atrial fibrillation from 2 1 patients, and 56 episodes of sinus tachycardia from 19 patients. Stability We compared calculated stability values for the episodes of VT with stability values for atrial fibrillation. The mean stability value for all VT episodes was 17.9 f 15.0 ms, and the mean stability value for atria1 fibrillation was 29.7 f 19.6 ms (P c ,001) for 1 manufacturer (Guidant Inc). Figure 1 describes the distribution of sta-
800
Nonthakumor
American Heart jx~rnal May 2000
et 01
.
Figure 2
Elf:
Stability
1.0 0.6 E 5 0.6 E ee 0.4 cc
_ $Esurderlhewrve Man#l=0.70 Man#2=0.76 Man#3=0.73
0.2
0.0 0.0
0.2
0.4
0.6
0.8
1.0
l-speciicity Three
curves
depict
manufacturers.
ROC
y-axis,
clinically
relevant
ers have
similar
Guidant;
Man
of stability
Sensitivity;
range
of sensitivity
performance #2,
algorithm x-axis,
Intermedics;
(90%),
regarding
Man
for
In
all 3 manufoctur-
specificity. #3,
3 different
1 - specificity. Man
#I,
Venitrex.
bility values for VT and atrial fibrillation for this manufacturer. Table I describes the sensitivity and specificity values for the different stability values for this particular manufacturer. The stability value of 40 ms had a sensitivity of 88% and a specificity of 31%. If we had programmed stability at 40 ms we would have failed to treat 10 episodes of VT (12%) and would have inappropriately treated 65 episodes (69%) of atrial fibrillation in this select population for one particular manufacturer. The stability value 60 ms would have been associated with 98% sensitivity; however, 92% of episodes of atria1 fibrillation would have been treated inappropriately at this value. We then performed the same series of experiments for stability algorithms from 2 other manufacturers. Figure 2 depicts the ROC of the stability algorithm for the 3 different manufacturers with the same analysis. The specificity, in the clinically relevant sensitivity range of 95% and above, was comparable across manufacturers at about 40%. However, the absolute values of stability that apply for the particular algorithm for the different manufacturers were different. The stability values associated with 95% sensitivity were 50, 45, and 65 ms for Guidant, Intermedics, and Venitrex, respectively. Unstable VT at rhythm onset To assess if the VT would stabilize after its onset, we reanalyzed all episodes of VT with stability values greater than 40 ms. We recalculated stability values with R-R intervals recorded after detection, while the device was charging and reconfirming. Ten episodes of VT had stability values greater than 40 ms at onset of the rhythm, but when stability was recalculated includ-
1. Sensitivity
(ms)
and
specificity
Sensitivity 1.2% 4.8% 6.0% 9.5% 15.5% 36.9% 51.2% 67.9% 75.0% 83.3% 86.9% 88.1% 90.5% 91.7% 95.2% 97.6% 97.6% 97.6% 98.8% 98.8% 98.8% 98.8% 98.8% 98.8%
values
for different
stability
Specificity 98.9% 98.9% 96.8% 93.6% 93.6% 80.9% 70.2% 60.6% 55.3% 43.6% 35.1% 30.9% 23.4% 18.1% 13.8% 8.5% 4.3% 2.1% 2.1% 0.0% 0.0% 0.0% 0.0% 0.0%
ing the R-R intervals after detection until delivery of therapy, 8 of 10 episodes had stability values less than 40 ms (Guidant Inc). Onset Fii-six episodes of inappropriate therapy from ST in 19 patients, compared with 83 spontaneous episodes of VT in 18 patients, were analyzed. Table II describes the sensitivity and specificity values for different programmable onset values for a particular manufacturer. An onset of 75% would have been associated with a sensitivity of 91% and a specificity of 95%, and an onset of 85% would have been associated with a sensitivity of 99% and a specificity of 70%. Figure 3 describes a family of onset percentages and associated sensitivity and specificity values.
Discussion This study demonstrates that inappropriate therapy is a common problem in patients with ICDs. The performance of the stability algorithms used to differentiate AF from VT was less than ideal, though comparable across manufacturers. The study also demonstrates that the onset algorithm accurately differentiates sinus tachycardia from VT. Stability In patients with ICDs prone to atria1 fibrillation, the stability algorithm has poor specificity at acceptable
Amer,con
kteort Journal
Volume 139, Number 5
Table
II. Sensitivity percentage values
Onset percentage value
Nanthakumor
and
specificity
values
for different
onset
Figure
3
1.0
Sensitivity
\
Specificity
onSel=&i% *
0.6
30-35 35-40 40-45 45-50 50.55 55-60 60-65 65-70 70-75 75-80 80-85 85.90 90-95 95-100
1.2% 7.1% 21.4% 34.5% 53.6% 70.2% 73.8% 8 1 .O% 83.3% 90.5% 92.9% 98.8% 100.0% 100.0%
100.0% 100.0% 100.0% 98.2% 98.2% 98.2% 98.2% 96.4% 96.4% 94.6% 89.3% 69.6% 58.0% 0.0%
Srmcitidty=M%
jr onwt=75% sensis~91%. SpeciIicily=95%
-
t 0.2 .
of a stability
algorithm
from
other
Neuzner et als monitored 27 patients with ICDs with Ventak PR devices (CPI, St Paul, Minn) over a period of 11 months. In this series, there were 264 episodes detected; 201 were VT, 13 were supraventricular tachycardia, and 52 were atrial arrhythmias. Despite signi cant differences in the value of mean stability between VT and SVT, there was a broad zone of overlap. A programmed stability value of 35 ms resulted in a specificity of 46.2%. This finding is similar to our analysis, in which the specificity for 35 ms in our population was 35%. Swerdlow et ali3 demonstrated that a programmed stability of 40 ms for a particular algorithm (Medtronic) would result in decreased detection of induced, paroxysmal, and chronic atria1 fibrillation by 95%. However, these were only achieved by increasing the detection
.
.
,
,
,
.
.
.
.
/
.
.
.
,
.
.
/
.
.
I
0.0
Depicts
ROC
approaching
levels of sensitivity for VT at the initiation of the arrhythmia episode. This is especially true of fast atrial fibrillation from the regularity of the R-R intervals. Most clinicians do not use stability for programming ICDs for the fear of underdetecting VT; this study suggests that this fear is well-founded. However, VT that would be initially underdetected because of the irregularity at the beginning of the episode may be detected appropriately with progression of the episode if VT becomes regular. Our data suggest that there is no universal cutoff that leaves adequate sensitivity for VT yet adequate specificity for atria1 Iibrillation at onset of arrhythmia episode. In addition, the actual stability values that give ideal performance are different for different manufacturers, and hence the programming cardiologist needs to be aware of the specific algorithm for each manufacturer.
.
,
.
,
*
, ’
*
0.0 Y
Sensitivity;
Performance studies
et al 801
0.2
of onset x-axis,
algorithm
1 - specificity.
ideal
0.6
o.&.spedficityo.6
for a single Curve
manufacturer.
describes
1.0
y-axis,
an algorithm
performance.
interval to 16 R-R intervals. In that same report, a programmed value of 50 ms and 8 R-R intervals would have resulted in 60% of the atrial fibrillations being detected inappropriately as VT. Barold et alzo prospectively evaluated the stability criteria in 17 patients with Medtronic 721X ICD. They induced atrial fibrillation at 3 and 6 months, and of the 33 inductions 4 resulted in inappropriate therapy. Brugada et al*3 prospectively monitored 82 patients with CPI-Guidant devices; in this series, stability programmed to 40 ms resulted in underdetection of 19 of 497 episodes of VT from stability being greater than 40 ms. The comparison of these studies to our data is limited by the differences in study design, differences in the rate detection algorithm, and the algorithm for stability used in the investigated devices. It is our belief that data from small investigational studies do not reflect the actual rate of inappropriate therapy in larger noninvestigational populations. Recently Rosenqvist et al*l published a prospective multicenter study of 778 patients with either Medtronic model 7219 C (392 patients) or model 7219 D (386 patients) ICDs with the primary intent of monitoring for adverse events. The average duration of follow-up was 4.0 * 4.6 months. Of 294 patients who received ICD therapy during the follow-up period, inappropriate detection and subsequent therapy delivery were observed 119 times in 102 patients. Stabilization
of unstable
VT after onset
of VT
Unstable VT in the ICD population may be caused by nonsustained VT with occasional capture beats or ventricular premature beats followed by a compensatory
American
802
Nanthakumar
Heart journal May 2000
et al
pause at the origin of VT, resulting in initial irregularity of R-R intervals.17 Several studies have reported variability in the cycle length of induced VT, particularly at the beginning of the episode.ls.19 Irregular VT may also be caused by polymorphic VT that has varying R-R intervals as a consequence of antiarrhythmic drug therapy.12 In our population, stability values decreased as the rhythm progressed in 8 of 10 episodes of VT. This finding suggests that the unstable VT seen in our population was caused in part by instability at the onset and not polymorphic VT. This finding is consistent with the series of Brugada et al’3 in which all 19 episodes of VT with stability greater than 40 ms dropped to below 40 ms within 2 * 1 seconds. Onset criterion Our study demonstrates that the enabling onset algorithm could result in considerable reduction in inappropriate therapy from sinus tachycardia. We did not evaluate the performance of the onset criterion in atria1 fibrillation because the onset of this arrhythmia is expected to be as sudden as VT. Performance of onset algorithm from other studies In a previous study of devices manufactured by Guidant Inc, onset programmed at 9% led to undertreatment of 27 (5.4%) of 497 episodes with 17 episodes of sinus tachycardia treated inappropriately. These 27 episodes of VT were ultimately treated after 72 f 34 seconds as the sustained rate duration monitor elapsed.
diagnostic data from dualchamber systems or QRS morphologic criteria may be needed to provide meaningful discrimination between atria1 fibrillation and VT in patients prone to atrial fibrillation. Our study does imply that onset should be programmed to the ICD for patients who are at risk to receive inappropriate therapy from sinus tachycardia. Limitations The main limitation of this cal and did not reflect actual However, device algorithms R-R events from patients for
We thank Guidant Inc, Intermedics, St Jude/Ventriter Inc, and Medtronic for their cooperation in participating in this project and Connie Yang for her assistance in the preparation of the manuscript.
References 1. Ruskin JN, O’Callaghon dioverter
2. Anderson
MH,
detection
3. O’Nunain
1:2204-l
of significant
and
cordioverter
implica-
late com-
defibrillotors.
3. cardioverter
symptoms.
by stored
tachycardio
defibrillators:
T, et al. Limitations
automatic
BG, et al. Implantable
absence
ventricular
cardioverter
M, Trouton
1995;9
l-707.
PACE 1997;20:2975-83.
of third generation
Circulation
car-
1997;22:64
of basic
in implantable
programming.
defibrillator
Rhythm
electrogram
therapy
diagnosis
analysis.
and
Circulation
in the
manage-
1993;
87: i 897.906. 5. Neuzner
J. Clinical
experience
capable
of biphasic
results
of CI prospective
investigator
6. Wood data
MA,
pulse
study.
PACE 1994;17:
implantable
cardioverter
Multicentre
Investigator
European
storage:
Ventok
RI, et al. lessons
clinical
trial using
defibrillator. Group.
data
P2
1243.55.
BS, Domiano
in o multicenter
defibrillator
and enhanced
m¢er
Stambler
logging
with o new cordioverter
waveform
group.
The Guardian
J Am Coil
learned
from
o late generation ATP42
Cordial
10
1994;24:
1692.9.
7. Brachmann follow
J, Sterns
implantable
a.
Grimm
defibrillotors.
BF, Marchilinski
unnecessary,
implantable
9. Fogoros
with nonthorocotomy
cardioverter
W, Flares
documented
PACE 1994;17:499-505.
shocks
PACE 1992;15:
implontoble
of shocks cordioverter
and resuscitative incidence
following
shock
study:
[abstract]
J Am Coil Cordial
11. Johnson
NJ, Morchilinski
tochycardia
therapy
146573.
of defibrillators
fibrillation
increase
the
defibrillotors
1:27BA.
FE. Arrhythmias
due to diagnostic
with
and pottern
1989;12:
by implanted
1993;2
patients
of automatic
evaluation
does otriol
shocks
incidence
PACE
Comprehensive
of inappropriate
in 241
1667.73.
implantation
defibrillotors.
investigators.
and chronic
generation
FE. Electrocardiographically
spontaneous
defibrillators.
efficacy third
RN, Elson JJ, Bonnet CA. Actuoriol
10. CEDERS
12.
LD, Hilbel T, et al. Acute
up OF patients
of occurrence
Clinical implications In patients prone to atrial fibrillation, and especially those with an expected rapid ventricular response, the specificity of diagnostic algorithms needs to be improved. Our results imply that the stability algorithm has limitations and may not be adequate for patients prone to atrial fibrillation if VT be appropriately detected 98% to 100% of the time. Other discrimination strategies such as atrial
et al. Performance
S, Roelke
plications
status of implontable
Curr Probl Cardiol
algorithms
tions for device
4. Hook
PA. Current
defibrillotors.
ment aided
Dual-chamber devices and tachycardia discrimination Discriminating VT from ST or atria1 fibrillation may be difficult in singlechamber ICDs with existing tachycardia discriminators. 14 Dual-chamber ICDs have the advantage of monitoring atrial rate and its relation to the ventricular rate. The hypothesis that tachycardia discriminators in dual-chamber ICDs may be superior to singlechamber ICDs was tested in a recent trial.24 In this trial patients with dual-chamber devices were randomly assigned either to a strategy of enhanced detection with dual-chamber algorithms or to a rate based only on detection strategy. There was an absolute reduction of inappropriate therapy of 15% between the 2 strategies at 6 months. However, no trials have directly compared singlechamber devices with discrimination algorithms activated to dual-chamber devices.
analysis is that it is theoretidevice behavior in patients. are best tested on recorded safety reasons.
induced nonspecificity.
Cardiol
1991;18:141B-25.
tuderitz
B, Jung W, Deister A, et al. Potient acceptance
by device
onti-
J Am Coil
of implontoble
Amer,can Heart Journal Volume 139, Number 5
cordioverter
Nanthakumor
defibrillator
in ventricular
tochyarrhythmias.
PACE 1993;
ventricular
16: 18 15-2 1. 13. Swerdlow
409-l
CD, Chen
tochycordia therapy
from
PS, Kass RM, et al. Discrimination
sinus tochycordio
cordioverter
of ventricular
and atria1 fibrillation
defibrillator.
J Am Coil
Cardiol
20.
in a tiered
CD, Ahern
Bardy
GH,
1994;23:
16.
GH, Troutman
tiered
therapy
17. Framer cordio 18. Geibeil onset pacing. 19. Volosin
specificity
J Am Coil Cordial
B, Johnson
defibrillotors.
Bardy
1992;85:
PS, et al. Underdetection to enhance
defibrillator.
Hoffer
cordioverter
T, then
by algorithms
opy cordioverter 15.
Borold
HS, Newby
ond
ventricular
1998;2
lar tachycordio
1994;24:4
G, et al. lmplontoble
Circulation
1993;87:
antiarrhythmia
device.
with a
Circulation
cycle
length [obstroct].
A, Zehender of sustained
M, Brugoda
KJ, Beouregord
tochy-
P. Changes
in cycle
for antitachycardio
Jewel
CD, Ahern
defibrillators.
PACE
events
with tronsve-
a prospective
ICD investigators.
Circulation
multicen1998;98:
T. Effects
of antiarrhythmic
tachycardio:
o new drug
drugs device
on interval interaction
PACE 1994;17:11-865.
Brugodo implantable
length
ot the
J. Is inappropriate
therapy
cordioverter
defibrillotors?
24.
Dorion
therapy changes
P, Newman
of a standardized
15:588-92. R. Spontaneous
in implontable
evoluotion
suproventricular
a resolved
issue with current
Am J Cardiol
1999;83:
8, et al. A randomized
clinical
40D-44D.
importance
LA, Fobiszewski
between
T, Block M, et al. Adverse
of ventricular
[abstract].
PACE 1987;10:.593.
tachycordios:
Am Heart J 1988;l
of ventricular
1991;17:
G, et 01. Prospective
to discriminate
cardioverterdefibrillators:
European
Swerdlow stability
23. R, et al. Oscillation
J Am Coil Cardiol
663.70. 22.
1689.98. M, Kus T, Nadeou
M, Beyer
ter study.
transvenous
experience
KH, Tomossoni
tachycardio
nous implantable
16-24.
length.
1: 1347-55.
2 1. Rosenqvist
ther-
1 152-68.
G, Poole JE, et al. Clinical
multiprogromoble
of ventricuin CJtiered
cycle
4.
of new and old criteria
1342.55. 14. Swerdlow
tachycordio
et al
in
[obstrac$
D, Thiboult protocol
using a dual chamber Circulation
for the prevention implantable
1999;100:1786.
trial
of inappropriate
cordioverter
defibrillator
803