Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty

Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty

Orthopaedics & Traumatology: Surgery & Research (2012) 98, 398—404 Available online at www.sciencedirect.com ORIGINAL ARTICLE Interest of an intra...

711KB Sizes 0 Downloads 28 Views

Orthopaedics & Traumatology: Surgery & Research (2012) 98, 398—404

Available online at

www.sciencedirect.com

ORIGINAL ARTICLE

Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty O. Barbier ∗, D. Ollat , G. Versier Bégin Military Teaching Hospital, 69, avenue de Paris, 94160 Paris, France Accepted: 21 February 2012

KEYWORDS Total hip arthroplasty; Offset; Limb-length discrepancy; Surgical planning

Summary Introduction: Total hip arthroplasty (THA) seeks to restore a stable, mobile and pain-free joint. This requires good implant positioning and peroperative restoration of limb-length and femoral offset. Hypothesis: A mechanical measurement device (length and offset optimization device [LOOD]) fixed to the pelvis can optimize lower-limb length and offset control during THA performed on a posterolateral approach. Patients and methods: Two prospective THA series were compared: 32 using the LOOD and 26 without. Patients with more than 5 mm preoperative limb-length discrepancy were excluded. The intraoperative target was to restore individual anatomy. Radiographic analysis was based on pre- and postoperative AP pelvic weight-bearing views in upright posture, feet aligned, with comparison to peroperative LOOD data. Results: Mean deviation from target length (i.e., pre- to postoperative length differential) was 2.31 mm (range, 0.04—10.6 mm) in patients operated on using the LOOD versus 6.96 mm (0.01—178 mm) without LOOD (P = 0.0013). Mean deviation from target offset was 3.96 (0.45—13.50) mm with LOOD versus 10.16 (0.93—28.81) without (P = 0.0199). There was no significant difference between operative and radiographic measurements of length deviation using LOOD (P = 0.4); those for offset, however, differed significantly (P = 0.02). Discussion: The LOOD guides control of limb-length and offset during THA on a posterolateral approach. Reliability seems to be better for limb-length than for offset. It is a simple and undemanding means of controlling limb-length and offset during THA. Level of evidence: III, prospective case-control study. © 2012 Elsevier Masson SAS. All rights reserved.

Introduction ∗

Corresponding author. 64, rue des Fabriques, 54000 Nancy, France. Tel.: +33 6 16 90 06 80. E-mail address: [email protected] (O. Barbier).

Controlling lower-limb length in total hip arthroplasty (THA) is difficult and is a frequent cause of patient dissatisfaction and litigation [1—8]. Restoring femoral offset is also essential to a good functional result, maintaining an

1877-0568/$ – see front matter © 2012 Elsevier Masson SAS. All rights reserved. doi:10.1016/j.otsr.2012.02.004

Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty

399

effective hip abductor lever-arm to improve implant stability and longevity and quality of gait [9—15]. At the present time, length and offset restoration depends on preoperative planning by tracing [11] and peroperative tests of soft-tissue tension and length (Charnley’s ‘‘shuck’’ test [12], the ‘‘dropkick’’ test and the ‘‘leg-to-leg’’ test); these tests, however, are affected by the surgeon’s experience, the type of anesthesia [16] and other factors which make them somewhat subjective [17]. The present study sought to assess the reliability of a measurement device (length and offset optimization device [LOOD]) implemented during THA to control lowerlimb length and femoral offset peroperatively. The working hypothesis was that the LOOD would improve control of lower-limb length and offset.

Patients and method Patients A continuous prospective single center study was performed on patients undergoing THA between July 2009 and January 2010. Inclusion criteria were: primary or secondary osteoarthritis of the hip with preoperative limb-length discrepancy (LLD) less than 5 mm on preoperative AP weight-bearing pelvic X-ray, for which primary THA without bone reconstruction was indicated. Two consecutive series were included: for the first 3 months of the study, patients were operated on without LOOD, and for the following 3 months with. Exclusion criteria were: preoperative LLD > 5 mm on AP weight-bearing pelvic X-ray, and history of conservative hip surgery for hip dysplasia. The peroperative target was to restore preoperative limb-length and offset. Fifty-eight patients were included: 32 operated on with LOOD and 26 without. Age, indication, gender, laterality and body-mass index (BMI) were collected for both groups (Tables 1 and 2).

Surgical technique The LOOD device (Amplitude, Neyron, France) (Fig. 1) comprises three parts. The fixation nail, with a triangular point to ensure stable bone anchorage and prevent rotation, is graduated every 5 mm to measure offset, and is connected, via an angular protractor guide with locking screw, to a measuring device calibrated every 5 mm which is fixed on the nail. Finally, a probe is fixed on the measuring device by a second screw (Fig. 1).

Table 1

Figure 1 LOOD device, comprising of a fixation nail with a conical apex designed to be fixed in the pelvis at the superior aspect of the acetabular area (1), connected by a tightening bolt (2) to the graduated length measurer (3). The graduated measurer is connected by a measuring bolt (4) to the probe, which is designed to be in contact with the lateral aspect of the greater trochanter (5).

All patients were operated on by a posterolateral approach, in lateral decubitus. Where the LOOD was used, after exposing and sectioning the pelvi-trochanteric muscles, the fixation nail was positioned across the medial gluteal muscle in the iliac wing following the axis of the femur, 3 to 5 cm above the greater trochanter. The LOOD was then assembled and a reference point was marked on the greater trochanter by electrocautery, with the lower limb positioned reproducibly in extension in the axis of the body, foot parallel to the ground (Fig. 2). In case of difficulty in marking, the electrocautery mark could be replaced by a short screw in the greater trochanter. The moveable part of the LOOD was then removed, without disturbing the settings, its height being determined using the graduation on the iliac fixator. The implant was a model with a modular cone reproducing standard or varus necks of various lengths (Amplitude, Neyron, France). Where the LOOD was used, it was repositioned on the iliac fixator nail at the same height as initially, and the lower limb was repositioned in extension in the reference position (Fig. 3). The peroperative target was to reproduce the initial offset and length. Stability tests were systematically made to determine neck type in patients in whom the LOOD was not used and to

Characteristics of the two study groups.

Number of patients Male/female Mean age (yrs) Lateralization right/left Body-mass index

With LOOD

Without LOOD

P

32 13/19 70.5 ± 15.5 (range, 47—87) 19/13 24.9 ± 3.8 (range, 17.7—31.1)

26 16/10 65.3 ± 11.7 (range, 36—86) 13/13 23.8 ± 3.3 (range, 18.1—29.7)

NS NS NS NS NS

LOOD: length and offset optimization device; NS: non-significant.

400 Table 2

O. Barbier et al. Etiologies of indications for THA in the two groups.

Etiologies Primary osteoarthritis Rapidly destructive osteoarthritis Minor dysplasia Aseptic osteonecrosis of the femoral head Inflammatory rheumatism Posttraumatic osteoarthritis

With LOOD (n = 32) 24 1 3 4 0 0

Without LOOD (n = 26) 17 1 3 4 0 1

P NS NS NS NS NS NS

LOOD: length and offset optimization device; NS: non-significant.

• in both groups, pre- and postoperative (1 month) radiologic analysis on AP weight-bearing pelvic view with the feet together and parallel, measuring (Fig. 4): ◦ the height between the most medial point of the lesser trochanter and the horizontal line between the teardrops, to measure limb length and change in limb length, ◦ global offset, defined as the distance between the midpoint of the pelvis (determined by the sacrum-to-pubis line) and the tip of the greater trochanter, ◦ diameter of the femoral head and of the cup. Figure 2 Peroperative view of the LOOD before arthrotomy: the pin fixation was placed through the gluteus medius in the iliac bone, in the axis of the femur, 3 cm above the greater trochanter and the reference point was marked with electrocautery on the greater trochanter with the leg in extension in the axis of the body.

7 control implant stability where the LOOD was used, the priority being to achieve a stable hip.

To avoid enlargement bias related to the radiograph scale, the scale for the preoperative radiograph measurements was calculated from the ratio of the inferosuperior diameter of the femoral head, as measured peroperatively by caliper, to the femoral head diameter measured on the radiograph. The scale for the postoperative radiographs was calculated from the ratio of the implanted cup diameter as measured on the postoperative radiograph to the real cup diameter. All differential measurements (both on X-ray and on LOOD) were recorded as absolute values.

Assessment methods Data collection comprised: • in the LOOD group, peroperative measurement by ruler of the difference in length and offset between the initial reference point and the final position of the device probe;

Figure 3 Peroperative view of the LOOD after completion of THA: the probe is on the great trochanter reference mark. The LOOD enables limb-length and offset restoration to be checked.

Figure 4 Radiological measures made on the pre- and postoperative 1-month pelvic weight-bearing X-ray views, with feet placed side by side in parallel. A. Distance between the most medial point of the lesser trochanter and the horizontal line between the teardrops, to measure limb-length change. B. Distance between the middle of the pelvis, defined by the sacrum-to-pubis line, and the tip of the greater trochanter, measuring global offset. C. Diameter of the femoral head or acetabular cup.

Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty

Statistics Statistical analysis used the StatPlusTM software package (AnalystSoft, Alexandria, VA, USA). Demographic data were compared between the two groups by non-parametric Mann-Whitney and Kruskal-Wallis tests. Categoric variables (change in length and offset) were expressed as means, ranges and standard deviations and analyzed on Fisher exact tests for inter-group comparison. The significance threshold was set at P ≤ 0.05.

Results Fifty-eight patients were included. The 2 groups were comparable on the study parameters (Table 1). Etiologies were likewise comparable, with a majority of primary osteoarthritis (Table 2). There were no peroperative complications, and no failures of proximal fixation of the LOOD. Stability tests results never led to a change in neck size or contradicted the LOOD data. There were no cases of femoral stem subsidence on 1-month radiographs. Analysis of pre- and postoperative radiographs found a mean length difference of 2.31 mm (SD, 2.64 mm; range, 0.04—10.60 mm) with LOOD and 6.96 mm (SD, 4.72 mm; range, 0.01—17.80 mm) without (P = 0.0013). Mean preand postoperative offset difference was 3.96 mm (SD, 4.79 mm; range, 0.45—13.50 mm) with LOOD and 10.16 mm (SD, 7.05 mm; range, 0.93—28.81 mm) without (P = 0.0199) (Table 3). Peroperative and radiographic measurements of change in length using the LOOD did not significantly differ: 1.70 mm (0.04—8.20 mm) versus 2.31 mm (0.04—10.60 mm) (P = 0.4). Measured change in offset, on the other hand, was significantly different between the peroperative LOOD and radiographic data (respectively: mean 5.20 (0.69—20.54) versus 3.96 (0.45—13.50) (P = 0.02)) (Table 3).

Discussion The present study sought to assess the efficacy of the LOOD device in terms of improving peroperative control of limb-length and offset during THA. Post-THA limb-length discrepancy (LLD) exceeding 10 mm, raising legal and/or neurological issues, is reported in 16—32% of cases in the literature [1—7,18—24]. Beyond this 10-mm threshold, LLD is felt by the patient [25], thereby affecting functional outcome [24—27]. Usually, patients are bothered by such LLD during the first postoperative months, with symptoms fading off over time; 15%, however, remain symptomatic [28,29] and more than half of these cases require a compensatory insole to improve satisfaction [2,4]. Restoring offset is important to improve hip abductor lever-arm and implant longevity [9—15]. The present results show that the LOOD can improve peroperative control of both limb-length and offset, and enables quantification of change in limb-length, although not in offset. The main study limitation concerns the accuracy of the radiographic measurements, which may be questioned, especially in comparison with computed tomodensitometry (CT) [30]. Kjellberg et al. [30], however, reported

401

radiographic LLD measurements to be reasonably reproductible regarding interobserver reproductibility (kappa = 0.83) as well as intra-observer reproductibility (kappa = 0.90). Differences between X-ray and CT measurement of offset and limb-length can be explained by the influence of pelvic tilt and of lower-limb rotation [31—35]. To ensure reproducible lower-limb rotation in a given patient and to avoid flexion deformity during X-ray, pelvic views were taken with the feet together and parallel and postoperative views were taken only at 1 month to avoid the postoperative functional flexion deformity, which affects pelvic tilt. Other possible solutions could have been: following Jaramaz et al. [36], to associate preoperative CT to postoperative X-ray, so as to correct pelvic positioning; or following Tannast et al. [31], to use an algorithm to correct tilt and rotation. However, both of these procedures transform the change in pelvic positioning into a measurement of pelvic abduction and inclination angles, without directly relating these to the linear measurements. Stephen et al. [37] limit the impact of pelvic tilt by measuring the difference in diameter between the two hips on each X-ray view, considering the tilt-effect between any two images to be comparable in both hips. Measurements from image to image may further be subject to enlargement bias due to change in pelvis position with respect to the plane of the film and the centering of the X-rays [31,38]. To limit this source of bias, the scale for the preoperative radiograph measurements was calculated from the ratio of the inferosuperior diameter of the femoral head, as measured peroperatively by caliper, to the diameter of the femoral head measured on the radiograph. The scale for the postoperative radiographs was calculated from the ratio of the implanted cup diameter, as measured on the postoperative radiograph, to the real cup diameter. Even so, there may be residual error due to X-ray divergence, as the rays are not perpendicular to the cup on AP pelvic views but rather centered on the midline of the pelvis. Only CT [30,39] would enable precise measurement — but at the cost of extra-irradiation. The offset measurements further lost accuracy as we did not use femoral or global offset, as described by Lecerf et al. [9], but included the distance between the teardrop line and the center of the pelvis in this measurement. In all, while still using plain X-ray, we sought to minimize measurement error in this prospective study design. One final limitation concerns the choice of surgical approach, inasmuch as the present findings cannot be extrapolated to surgery using other approaches. We chose to use an implant with a modular cone so as to give the surgeon a range of possibilities in following the LOOD measurements [9]. Peroperative testing was, however, always associated to the LOOD, to ensure implant stability. Following the LOOD data never led to instability, and it was never necessary to change the neck in order to prioritize stability over restoration of limb-length and offset, abandoning the LOOD input. The primary study objective was to show that the LOOD device improved per- and postoperative control of limblength and offset. Control of both parameters was found to be significantly better in the LOOD group. Other studies have reported on similar methods [8,18—20,23,40—49]. Most seem effective, but concern only limb-length control and not offset. McGee and Scott [50] described a simple

402

O. Barbier et al.

Table 3 Comparison of change in limb-length and offset between pre- and postoperative radiographs in the two groups (with LOOD = 32, without LOOD = 26). For the LOOD group, comparison between peroperative and radiographic measurements. Mean (range) (mm) Comparative analysis of pre- and postoperative radiographs Change in lower-limb lengtha Series operated on using LOOD Series operated on without LOOD Change in offsetb Series operated on using LOOD Series operated on without LOOD Comparison of change in limb-length and offset as measured peroperatively by LOOD and on pre- and postoperative radiographs in the LOOD group (n = 32) Change in lower-limb length Change as measured by LOOD Change as measured on X-raya Change in offset Change as measured by LOOD Change as measured on X-rayb

Standard deviation

P

2.31 (0.04—10.60) 6.96 (0.01—17.80)

2.64 4.72

0.0013

3.96 (0.45—13.50) 10.16 (0.93—28.81)

4.79 7.05

0.0199

1.70 (0.04—8.20) 2.31 (0.04—10.60)

1.63 2.64

0.4 (NS)

5.20 (0.69—20.54) 3.96 (0.45—13.50)

4.35 4.79

0.02

LOOD: length and offset optimization device; NS: non-significant. a Change as measured on X-ray = absolute difference in distance between teardrops and lesser trochanter on pre- vs. postoperative views b Change as measured on X-ray = absolute difference in distance between center of pelvis and greater trochanter on pre- vs. postoperative views

and widely used method [45,46] using a wire placed in the ilium and stretched toward the greater trochanter. Woolson et al. [23] used a caliper fixed on the iliac wing, which ensured less than 6 mm LLD in 89% of their patients. Ranawat et al. [18] used a Steinman nail positioned in the ischium at the inferior part of the posterior horn of the acetabulum, obtaining less than 6 mm LLD in 87% of cases. Using a similar device, Konyves and Bannister [51] reported a mean 9-mm LLD, and Matsuda et al. [44] 2 mm (versus 4 mm without the ancillary). Bose [20] used a metal pin in the iliac wing with a level to position the limb horizontally, achieving a mean LLD of 3.4 mm according to the caliper, versus 8.8 mm without the caliper. Jasty et al. [19], using a caliper fixed to the summit of the iliac wing, reported only 13% postoperative LLD exceeding 5 mm. Takigami et al. [43], using a caliper fixed 2 to 3 cm above the acetabulum, reported a mean postoperative LLD of 4.2 mm (range, 0—13 mm). In 2003, Shiramizu et al. [41] compared a series of patients operated on with or without use of an L-shaped caliper fixed on the anterosuperior iliac crest with the long arm parallel to the limb: mean postoperative LLD was 2.1 ± 1.5 mm using the caliper, versus 8.2 ± 3.8 mm without (P < 0.0001). Other authors [25,40,42,47,49] likewise reported benefit with similar devices for controlling limb-length but none, except for Kutz [52] and Renkawitz et al. [53], assessed restoration of offset. Renkawitz et al. [53] developed a reference system glued to the operative drape, and reported good correlation between pre- and postoperative measurements for limb-length (r = 0.92; P < 0.001) and offset (r = 0.97; P < 0.001). Kutz [52], using the original superior approach described by Murphy et al. [54] with endofemoral preparation of the femur ahead

of neck sectioning, reported significantly improved control of limb-length (P = 0.0004) but not of offset (P = 0.17). The second objective of the present study was to determine whether the LOOD could quantify change in length and offset due to surgery. Results indicated good quantification of postoperative change in the former but not the latter: there was no significant difference between peroperative LOOD and pre/postoperative X-ray measurement of length change, whereas the difference was significant for offset. Ranawat et al. [18] reported good correlation (r = 0.82) for LLD as measured by caliper and on X-ray. Likewise, Shiramizu et al. [41] reported a significant correlation (P < 0.0001) between per- and postoperative measurements in their caliper group. Using the technique described by Kurtz [52], peroperative and radiographic measurements showed good correlation for limb-length (mean difference −0.1 mm; r = 0.89) but less so for offset (mean difference −0.4 mm; r = 0.57). Thus, the LOOD measurer seems to be reliable; its accuracy depends on adherence to certain technical imperatives. Firstly, limb position during the various peroperative measurements must be reproducible. Sarin et al. [21] demonstrated that a slight difference in limb position induced considerable difference in measurement: a deviation of 5◦ in abduction/adduction led to 8 mm apparent difference in limb-length. Following the literature, we set the reference position as the hip in extension in the axis of the body. Wixson and MacDonald [38] designed a navigation technique for keeping the limb in the same position during surgery. The second imperative is stable peroperative pelvis fixation, as a change in femur positioning relative to the pelvis induces

Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty error in limb-length and offset measurement [18]. Ranawat and Rodriguez [12] showed that pelvic tilt caused an apparent limb-length discrepancy. Di Gioia et al. [55], studying pelvic movement during THA in lateral decubitus, reported 23◦ (from —15.9 to 7.0◦ ) rotation in abduction/adduction, 16◦ (from —8.1 to 7.5 mm) in flexion/extension and 40◦ (from —19.5 to 20.1◦ ) in version. There is thus a risk of error using an ancillary fixed to the pelvis, and the further the fixation screws or nails are situated from the center of rotation of the hip, the greater the error is likely to be [18]. The distance between this iliac reference point and the center of rotation of the hip therefore needs to be kept to a minimum, as is the case with the LOOD device. It is also necessary to reduce this distance in order to reduce bias in peroperative measurement: Shiramizu et al. [41] showed that, if the measuring device and the femoral reference point were not in the same axis, parallel to the femur, data collection would be biased by lack of parallax. Unlike the LOOD ancillary, most measuring devices reported in the literature use the anterosuperior iliac crest as reference, although it is remote from the center of rotation of the hip and not in the axis of the femur. Ranawat et al. [18] positioned their pin in the posterior infra-acetabular groove to be closer to the center of rotation.

Conclusion

[10]

[11]

[12]

[13]

[14]

[15]

[16]

[17] [18]

The LOOD device is a useful means of improving control of limb-length and offset during THA on a posterolateral approach.

[19]

Disclosure of interest

[20]

The authors declare that they have no conflicts of interest concerning this article.

References

[21]

[22] [23]

[1] Ranawat CS, Rodriguez JA. Functional leg-length inequality following total hip arthroplasty. J Arthroplasty 1997;12:359—64. [2] Williamson JA, Reckling FW. Limb length discrepancy and related problems following total hip joint replacement. Clin Orthop Relat Res 1978;134:135—8. [3] Turula KB, Friberg O, Lindholm TS, Tallroth K, Vankka E. Limb length inequality after total hip arthroplasty. Clin Orthop Relat Res 1986;202:163—8. [4] Edeen J, Sharkey PF, Alexander AH. Clinical significance of leg-length inequality after total hip arthroplasty. Am J Orthop 1995;24:347—51. [5] Maloney WJ, Keeney JA. Leg length discrepancy after total hip arthroplasty. J Arthroplasty 2004;19(4 Suppl. 1):108—10. [6] Austin MS, Hozack WJ, Sharkey PF, Rothman RH. Stability and leg length equality in total hip arthroplasty. J Arthroplasty 2003;18(3 Suppl. 1):88—90. [7] Plaass C, Clauss M, Ochsner PE, Ilchmann T. Influence of leg length discrepancy on clinical results after total hip arthroplasty: a prospective clinical trial. Hip Int 2011;21:441—9. [8] Desai AS, Connors L, Board TN. Functional and radiological evaluation of a simple intraoperative technique to avoid limb length discrepancy in total hip arthroplasty. Hip Int 2011;21:192—8. [9] Lecerf G, Fessy MH, Philippot R, et al. Femoral offset: anatomical concept, definition, assessment, implications for

[24]

[25]

[26]

[27]

[28]

[29]

[30]

403

preoperative templating and hip arthroplasty. Orthop Traumatol Surg Res 2009;95:210—9. Charles MN, Bourne RB, Davey JR, Greenwald AS, Morrey BF, Rorabeck CH. Soft tissue balancing of the hip: the role of femoral offset restoration. Instr Course Lect 2005;54:131—41. McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME. Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg (Br) 1995;77:865—9. Yamaguchi T, Naito M, Asayama I, Ishiko T. Total hip arthroplasty: the relationship between posterolateral reconstruction, abductor muscle strength, and femoral offset. J Orthop Surg 2004;12:164—7. Davey JR, O’Connor DO, Burke DW, Harris WH. Femoral component offset: its effect on strain in bone-cement. J Arthroplasty 1993;8:23—6. Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH. Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 2001;388: 125—34. Bachour F, Marchetti E, Bocquet D, Vasseur L, Migaud H, Girard J. Radiographic preoperative templating of extra-offset cemented THA implants: how reliable is it and how does it affect survival? Orthop Traumatol Surg Res 2010;96:760—8. Sathappan SS, Ginat D, Patel V, Walsh M, Jaffe WL, Di Cesare PE. Effect of anesthesia type on limb length discrepancy after total hip arthroplasty. J Arthroplasty 2008;23:203—9. Muller ME. Straight stem total hip replacement system. Berne, Switzerland: Proteck Ltd. Manual; 1982. Ranawat CS, Rao RR, Rodriguez JA, Bherde BS. Correction of limb-length inequality during total hip arthroplasty. J Arthroplasty 2001;16:715—20. Jasty M, Webster W, Harris W. Management of limb length inequality during total hip replacement. Clin Orthop Relat Res 1996;333:165—71. Bose WJ. Accurate limb length equalization during total hip arthroplasty. Orthopedics 2000;23:433—6. Sarin VK, Pratt WR, Bradle GW. Accurate femur repositioning is critical during intraoperative total hip arthroplasty length and offset assessment. J Arthroplasty 2005;20:887—91. Bourne RB, Rorabeck CH. Soft tissue balancing: the hip. J Arthroplasty 2002;17(4 Suppl. 1):17—22. Woolson ST, Hartford JM, Sawyer A. Results of a method of leglength equalization for patients undergoing primary total hip replacement. J Arthroplasty 1999;14:159—64. Wylde V, Whitehouse SL, Taylor AH, Pattison GT, Bannister GC, Blom AW. Prevalence and functional impact of patientperceived leg length discrepancy after hip replacement. Int Orthop 2009;33:905—9. O’Brien S, Kernohan G, Fitzpatrick C, Hill J, Beverland D. Perception of imposed leg length inequality in normal subjects. Hip Int 2010;20:505—11. Benedetti MG, Catani F, Benedetti E, Berti L, Di Gioia A, Giannini S. To what extent does leg length discrepancy impair motor activity in patients after total hip arthroplasty? Int Orthop 2010;34:1115—21. Meermans G, Malik A, Witt J, Haddad F. Preoperative radiographic assessment of limb-length discrepancy in total hip arthroplasty. Clin Orthop Relat Res 2011;469:1677—82. ParviziI J, Sharkey FP, Bissett GA, Rothman RH, Hozac WJ. Surgical treatment of limb-length discrepancy following total hip arthroplasty. J Bone Joint Surg (Am) 2003;85:2310—7. Hernigou Ph, Filippini P, Zilber S, Mathieu G, Poignard A, Chgouk A, De Moura A. Histoire naturelle de l’inégalité de longueur après PTH durant la première année. Rev Chir Orthop 2007;93:134—5. Kjellberg M, Al-Amiry B, Englund E, Sjödén GO, Sayed-Noor AS. Measurement of leg length discrepancy after total hip

404

[31]

[32]

[33]

[34] [35]

[36]

[37]

[38]

[39]

[40] [41]

[42]

[43]

O. Barbier et al. arthroplasty. The reliability of a plain radiographic method compared to CT-scanogram. Skeletal Radiol 2012;41:187—91. Tannast M, Zheng G, Anderegg C, Burckhardt K, Langlotz F, Ganz R, Siebenrock KA. Tilt and rotation correction of acetabular version on pelvic radiographs. Clin Orthop Relat Res 2005;438:182—90. Anda S, Svenningsen S, Grontvedt T, Benum P. Pelvic inclination and spatial orientation of the acetabulum: a radiographic, computed tomographic and clinical investigation. Acta Radiol 1990;31:389—94. Goergen TG, Resnick D. Evaluation of acetabular anteversion following total hip arthroplasty: necessity of proper centreing. Br J Radiol 1975;48:259—60. Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg (Br) 1993;75:228—32. Yao L, Yao J, Gold RH. Measurement of acetabular version on the axiolateral radiograph. Clin Orthop Relat Res 1995;316:106—11. Jaramaz B, DiGioia III AM, Blackwell M, Nikou C. Computer assisted measurement of cup placement in total hip replacement. Clin Orthop Relat Res 1998;354:70—81. Stephen B, Murphy MD, Timo M, Ecker MD. Evaluation of a new leg length measurement algorithm in hip arthroplasty. Clin Orthop Relat Res 2007;463:85—9. Wixson R, MacDonald M. Total hip arthroplasty through a minimal posterior approach using imageless computer-assisted hip navigation. J Arthroplasty 2005;20:51—6. Kitada M, Nakamura N, Iwana D, Kakimoto A, Nishii T, Sugano N. Evaluation of the accuracy of computed tomography-based navigation for femoral stem orientation and leg length discrepancy. J Arthroplasty 2011;26:674—9. Naito M, Ogata K, Asayama I. Intraoperative limb length measurement in total hip arthroplasty. Int Orthop 1999;23:31—3. Shiramizu K, Naito M, Shitama T, Nakamura Y, Shitama H. Lshaped caliper for limb length measurement during total hip arthroplasty. J Bone Joint Surg (Br) 2004;86:966—9. Itokazu M, Masuda K, Ohno T, et al. A simple method of intraoperative limb length measurement in total hip arthroplasty. Bull Hosp Jt Dis 1997;56:204—5. Takigami I, Itokazu M, Itoh Y, Matsumoto K, Yamamoto T, Shimizu K. Limb-length measurement in total hip arthroplasty

[44]

[45]

[46]

[47]

[48]

[49] [50]

[51]

[52] [53]

[54]

[55]

using a calipers dual pin retractor. Bull NYU Hosp Jt Dis 2008;66:107—10. Matsuda K, Nakamura S, Matsushita T. A simple method to minimize limb-length discrepancy after hip arthroplasty. Acta Orthop 2006;77:375—9. Hossain M, Sinha AK. A technique to avoid leg-length discrepancy in total hip arthroplasty. Ann R Coll Surg Engl 2007;89:314—5. Iagulli ND, Mallory TH, Berend KR, Lombardi Jr AV, Russell JH, Adams JB, Groseth KL. A simple and accurate method for determining leg length in primary total hip arthroplasty. Am J Orthop 2006;35:455—7. Pooler Archbold HA, Mohammed M, O’Brien S, Molloy D, McConway J, Beverland DE. Limb length restoration during total hip arthroplasty: use of a caliper to control femoral component insertion and accurate acetabular placement relative to the transverse acetabular ligament. Hip Int 2006;16: 33—8. Huddleston HD. An accurate method for measuring leg length and hip offset in hip arthroplasty. Orthopedics 1997;20: 331—2. Bal BS. A technique for comparison of leg lengths during total hip replacement. Am J Orthop 1996;25:61—2. McGee HMJ, Scott JHS. A simple method of obtaining equal leg length in total hip arthroplasty. Clin Orthop Relat Res 1985;194:269—70. Konyves A, Bannister GC. The importance of leg length discrepancy after total hip arthroplasty. J Bone Joint Surg (Br) 2005;87:155—7. Kurtz WB. In situ leg length measurement technique in hip arthroplasty. J Arthroplasty 2012;27:66—73. Renkawitz T, Shuster T, Grifka J, Kalteis T, Sendtner E. Leg length and offset measures with a pinless femoral reference array during THA. Clin Orthop Relat Res 2010;468:1862—8. Murphy SB, Ecker TM, Tannast M. THA performed using conventional and navigated tissue-preserving techniques. Clin Orthop Relat Res 2006;453:160—7. DiGioia AM, Jaramaz B, Blackwell M, et al. The Otto Aufranc award. Image guided navigation system to measure intraoperatively acetabular implant alignment. Clin Orthop Relat Res 1998;355:8—22.