SURVEY OF OPHTHALMOLOGY VOLUME 42 • NUMBER 5 • MARCH–APRIL 1998
DIAGNOSTIC AND SURGICAL TECHNIQUES
Intraocular Lens Complications Requiring Removal or Exchange ALAN N. CARLSON, MD,1 WILLIAM C. STEWART, MD,2,3 AND PATRICK C. TSO, MD1 1 Department of Ophthalmology, Duke University Eye Center, Durham, North Carolina, 2 Pharmaceutical Research Corporation, Charleston, South Carolina, and 3 University of South Carolina Medical School, Columbia, South Carolina, USA
Abstract. Intraocular lens (IOL)-related complications are caused primarily by mechanical trauma, inflammatory or infectious complications, or optical problems. Complications may occur at the time of surgery or be the result of an ongoing postoperative process. Mechanical and inflammatory injury may produce corneal decompensation, cystoid macular edema, hyphema, uveitis, and glaucoma, causing reduced vision and in some cases chronic pain. Optical problems may be due to a wrong power of the IOL or to postoperative decentration or dislocation of the lens. Ophthalmologists should be aware of the indications for IOL removal or exchange in those patients who have ongoing IOL-induced injury or impairment. Removal or exchange of an IOL frequently involves a complex decision-making process and is often associated with immense technical challenge. Various medical and surgical treatments may be tried to correct IOL problems before the decision is made to remove or exchange the lens. (Surv Ophthalmol 42:417–440, 1998. © 1998 by Elsevier Science Inc. All rights reserved.) Key words. bullous keratopathy • cataract surgery • complications of intraocular lenses corneal edema • cystoid macular edema • glaucoma • hyphema • malposition of intraocular lenses • removal or exchange of intraocular lenses • uveitis
Although complications from intraocular lenses (IOLs) were common in the early days of cataract surgery,8,215 improvements in surgical technique and IOL design and quality have reduced the problems. Mechanisms for IOL-related complications fall into four groups: mechanical trauma, chronic inflammation, infectious disease, and optical problems (Table 1). Mechanical trauma from an IOL may occur at the time of original surgery or it may be the result of ongoing intraocular injury. Surgical contact with the corneal endothelium causes endothelial cell loss, which may cause transient or persistent dysfunction with corneal decompensation. Anterior chamber (AC)-IOLs also may produce ongoing mechanical trauma to the cornea, iris, ciliary body, or the anterior chamber angle. Mechanical injury to the iris
•
may produce atrophy and chronic uveitis, potentially compromising aqueous outflow and leading to glaucoma. Erosion of the IOL into the ciliary body also may cause ocular discomfort. A more subtle form of injury may result from chronic inflammation produced by ongoing mechanical trauma. The IOL may erode into uveal tissue, liberating inflammatory mediators that damage other ocular structures; this is especially common with oversized or older AC-IOLs with a closed-loop design. These inflammatory products may contribute to progressive corneal endothelial cell loss, cystoid macular edema (CME), and glaucoma. A more serious form of inflammatory injury occasionally results from infectious endophthalmitis. Acute infectious endophthalmitis usually is not di417
© 1998 by Elsevier Science Inc. All rights reserved.
0039-6257/98/$19.00 PII S0039-6257(97)00140-9
418
Surv Ophthalmol 42 (5) March–April 1998 TABLE 1
Mechanisms for IOL-Related Complications Mechanical Intraoperative Postoperative Inflammatory Noninfectious Infectious Optical IOL Power IOL Malposition IOL 5 intraocular lens.
rectly attributable to IOL implantation, although epidemiologic data suggest that polypropylene loops are a risk factor for endophthalmitis.166 Menikoff et al reported that one-piece polymethylmethacrylate lenses appear less likely to be associated with infectious endophthalmitis than multi-piece design lenses with polypropylene loops.166 Successful treatment of endophthalmitis seldom requires IOL removal, although some patients with localized endophthalmitis are resistant to treatment until the IOL and the capsular bag are removed.165 Optical problems may be caused by improper power selection or abnormal IOL position. Improper IOL power may produce an unacceptable refractive error, which may require IOL exchange. Malposition may produce symptoms associated with the optic edge, IOL tilt, or other IOL structures, such as positioning holes, lens loops, or laser ridges. Failure of an IOL represents a great disappointment to the patient, who is usually aware of many successful outcomes of cataract/IOL surgery among friends. The patient often blames the surgeon and seeks a new ophthalmologist for follow-up care of the complications; thus, the original surgeon may not even be aware of the eventual need for IOL explantation. The consulting physician can help the patient to understand that the original surgeon shared a high level of optimism and that, although an IOL may currently be recognized as a poor design, it may have been acceptable or even the stateof-the-art IOL at the time of the original surgery.
Corneal Complications Corneal complications were the leading indication for IOL explantation before the development of newer IOLs and surgical techniques.8,143 The corneal endothelium consists of a monolayer of nonregenerative cells responsible for maintaining corneal clarity by its pump function. The cells that make up this layer are among the most delicate type in the human body.252 Corneal endothelial dysfunction after cataract surgery may be mild-to-moderate and transient (resulting in temporary microcystic edema and visual
CARLSON ET AL
impairment), or it may be severe and permanent. Persistent endothelial dysfunction with corneal decompensation is currently the most common indication for penetrating keratoplasty.22,24,43,51,90,99,117,132,151, 157,174,176,177,208,217,236,238,241
Surgical trauma can cause corneal decompensation,15,142,189,218,256 but the condition may be hastened by pre-existing corneal disease (e.g., Fuchs’ dystrophy). Endothelial cell loss related to the IOL may be caused by direct mechanical trauma, including excessive vaulting of an AC-IOL, lens mobility (pseudophakodonesis) of an undersized or poorly fixated IOL (tilting, rocking, or rotating [propellering]), or IOL malposition (e.g., migration through a peripheral iridectomy). Even peripheral corneal-IOL contact has the potential to cause central endothelial dysfunction. Peripheral cell loss may stimulate endothelial migration and reduce endothelial cell density centrally, causing corneal decompensation.106,107,110–112 Erosion of an IOL into uveal tissue may incite inflammation and cause the liberation of inflammatory mediators, such as prostaglandins, which may damage the corneal endothelium.8 Inflammation is often the mechanism for late corneal decompensation associated with closed-loop AC-IOL designs that erode into peripheral iris, anterior chamber angle, and ciliary body.44,135,162,163,211,236 This mechanism can explain the late corneal decompensation that may occur in patients who appear to have stable IOLs that are not directly traumatizing the corneal endothelium. Because the inflammation often is very mild, careful examination is required to reveal subtle cell and flare in the anterior chamber and fine inflammatory precipitates on the IOL surface. Inflammation associated with the semiflexible, closed-loop style AC-IOLs contributed to their removal from the USA market.9 Corneal failure may also result from continuing postoperative endothelial cell loss associated with aging, ocular disease, laser treatment, or iris or vitreous contact with the endothelium. PSEUDOPHAKIC CORNEAL EDEMA
Intraocular Lens Design We prefer pseudophakic corneal edema over the term pseudophakic bullous keratopathy to describe patients with corneal decompensation, because not all patients develop bullae. Pseudophakic corneal edema is the leading indication for penetrating keratoplasty.22,58,217,236,241,271 Kraff et al reported that 36% of IOL removals are performed because of pseudophakic corneal edema.143 In a series of 101 eyes of 98 consecutive patients, Lyle and Jin reported bullous keratopathy as the causative indication in 86.7% of patients who underwent combined penetrating keratoplasty and IOL exchange.153
419
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
Fig. 1. Pseudophakic corneal edema associated with a Stableflex-style, closed-loop design AC-IOL. Successful treatment requires IOL exchange at the time of penetrating keratoplasty.
When corneal decompensation is associated with the IOL design (e.g., the semiflexible, closed-loop design [Fig. 1]),26,44,47,118,135,162,163,211,226,236,243,255,272,274 the IOL should always be exchanged or removed when penetrating keratoplasty is performed. Such designs include the Leiske Style 10 (Surgical Corp., Goleta, CA), Azar 91Z (IOLAB Corp., Covina, CA), Optical Radiation Corp. model 11 Stableflex (Azusa, CA), the Intermedics Hessburg model 024 (Intraocular Inc., Pasadena, CA), and the Pannu AC-IOL (Allergan Medical Optics, Irvine, CA).8 Also to be avoided are AC-IOL designs featuring positioning holes or eyelets, which may erode into the angle and act as a smaller version of the closed-loop designs.8,155 Flexible, open, thin-loop lenses, such as the Dubroff-style (Intermedics Pharmacia, Monrovia, CA) AC-IOL, also may erode into the peripheral angle and create low-grade, chronic inflammation with concomitant corneal endothelial damage and CME.147 Surgical results of penetrating keratoplasty after removal or exchange of closed-loop AC-IOLs have been good. Kornmehl et al reported an average recovery of corrected visual acuity to 20/44 in 40 consecutive patients who underwent exchange of a closed-loop AC-IOL (30 Leiske, 10 Stableflex) for a Kelman Omnifit AC-21 AC-IOL (Allergan Medical Optics) at the time of penetrating keratoplasty.135 Koenig et al reported good results in 20 consecutive patients with corneal decompensation associated with a Stableflex-style AC-IOL who underwent penetrating keratoplasty, anterior vitrectomy, and IOL exchange for a Kelman Multiflex-style, open-loop AC-IOL (Cooper Vision Cilco Inc., Pomona, CA).133 Thirty-five percent (7 of 20) attained a spectacle correction of 20/40 or better, and all grafts remained clear during a follow-up period of 4–45 months.
Zaidman and Goldman reported a 90% chance of allograft clarity at the end of 12–18 months in patients undergoing penetrating keratoplasty and IOL exchange with a Kelman-style, open-loop AC-IOL or posterior chamber (PC)-IOL.285 Price et al noted that 20.6% (7 of 34) patients developed allograft failure with retained AC-IOLs compared with 5.1% (4 of 78) in patients with retained PC-IOLs at the time of keratoplasty.205 In contrast, Speaker et al reported a large number of graft failures in patients who retained closed-loop AC-IOLs at the time of keratoplasty.243 Allograft failure was noted in 60% of these patients at 1 year, 75% at 2 years, and 100% at 5 years. Smith et al,236 Insler et al,118 and Cohen et al47 reported series in which allograft outcomes were compromised by retention of closed-loop AC-IOLs. Removal of iris-supported lenses also should be considered at the time of penetrating keratoplasty, particularly if the IOL appears poorly fixated or excessively mobile. Despite initial reports suggesting a favorable surgical outcome in patients who retained these lenses, subsequent longer-term studies invariably demonstrated a higher rate of complications.12, 40,74,145,167,226,258,270 Speaker et al reported that 75% of the allograft failures in patients with retained irissupported IOLs occurred after the second postoperative year.243 Exchange did not appear to offer significant advantage over retention in their study. Rao et al found endothelial cell loss to occur at a greater rate with iris-supported IOLs than with Choyce-style AC-IOLs.207 Several additional authors noted that patients with retained iris-supported IOLs had the worst prognosis for maintenance of a clear graft.137,191,226 Reports by Meyer and Sugar,167 Kozarsky et al,137 and Waring et al274 support leaving a properly positioned, secure PC-IOL or modern flexible open-loop AC-IOL at the time of keratoplasty. Notably, Price et al reported that grafts for pseudophakic bullous keratopathy with retained posterior chamber lenses had a significantly higher failure rate from rejection and endothelial decompensation, 5.1%, compared with less than 1.0% in keratoconus, Fuchs’ dystrophy, or bullous keratopathy with secondary implants.205 Several authors have addressed the clinical and technical strategies associated with IOL exchange during penetrating keratoplasty.27,54,59,62,92,99,114,133,135, 146,199,203,205,227,241,272,281,285
Instability of the Intraocular Lens Excessive IOL mobility (pseudophakodonesis) jeopardizes allograft success. An unstable IOL should be removed at the time of keratoplasty unless it can be safely stabilized in a secure position and is considered a safe IOL design. Even modern, acceptable IOLs should be removed if they cannot be properly
420
Surv Ophthalmol 42 (5) March–April 1998
CARLSON ET AL
Fig. 2. Left: Migration of a Kelman-style II AC-IOL through a peripheral iridectomy. Right: This migration results in IOL tilt and corneal endothelial contact inferiorly. Intraocular lens exchange was performed to decrease the risk of corneal endothelial decompensation.
fixated, e.g., because of inadequate anatomic support, improper sizing, etc. Clinically, the best indication of unacceptable pseudophakodonesis is “shimmering” from mobile Purkinje-Sanson images III and IV from the anterior and posterior IOL surfaces, which can be observed during examination of subtle eye movements at the slit lamp (Holladay J: personal communication, 1996). Miller and Doane169 and Jacobs et al121 used high-speed cinematography to demonstrate pseudophakodonesis with corneal touch associated with irissupported IOLs. Rao et al identified progressive corneal endothelial cell loss over time.207 PERIPHERAL CORNEAL EDEMA
Corneal decompensation due to ongoing mechanical injury may be caused by central and/or peripheral endothelial contact with the IOL. Intermittent or constant central corneal touch may cause accelerated loss of endothelial cells.106,107,112–212 Corneal touch is more commonly associated with AC-IOLs and may be caused by improper IOL positioning, IOL migration due to improper IOL size, ocular trauma (including eye rubbing), excessive IOL vaulting, or prolonged head-down position.60 Migration of an AC-IOL through a wound dehiscence or a peripheral iridectomy (Fig. 2A) may also cause IOL-corneal touch (Fig. 2B). Patients with peripheral IOL-corneal touch may initially have no signs or symptoms of corneal decompensation. However, although the central cornea may appear clear, the peripheral contact may cause local cell loss, which can stimulate cell migration from the central cornea to the periphery. Eventually, continued endothelial cell loss will result in central corneal decompensation unless the source of damage is removed. To undergo IOL repositioning, removal, or exchange, the cornea must have adequate endothelial
reserve. Clinical signs of inadequate endothelial reserve may include a low cell density, (e.g., less than 500 cells/mm2), abnormal cell morphology (e.g., polymegathism or pleomorphism), or increased corneal thickness by pachymetry. Clinical symptoms may include decreased vision in the morning secondary to microcystic corneal epithelial edema. Microcystic corneal epithelial edema is frequently worse in the morning because of reduced evaporation and slight endothelial hypoxia from eyelid closure during sleep (Fig. 3). Patients with reduced endothelial cell density who develop microcystic corneal epithelial edema with normal intraocular pressure (IOP) are at high risk for corneal decompensation after intraocular surgery and may require penetrating keratoplasty. Patients who have IOL-corneal touch should be considered for IOL repositioning, removal, or exchange despite a clear central cornea provided that adequate endothelial cell reserve exists, as this may prevent the need for penetrating keratoplasty.106,107,160
Fig. 3. Microcystic corneal epithelial edema is best seen under high magnification using indirect retroillumination.
421
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES CLEAR CORNEA WITH INTRAOCULAR LENS-INDUCED ACCELERATED ENDOTHELIAL CELL LOSS
The decision to exchange an IOL is controversial when a patient has normal vision and no IOL-corneal touch but shows evidence of IOL pseudophakodonesis and/or chronic low-grade inflammation.28,186 Such patients may be at risk for corneal decompensation because of their IOL design, e.g., a closed-loop AC-IOL. When accelerated endothelial cell loss (i.e., cell loss at a much more rapid rate than could be attributed to aging) can be documented by serial specular cell photomicrography in a patient with a closedloop–style AC-IOL, removal or exchange of the IOL should be considered. Coli et al reported a series of 102 eyes with anterior chamber or iris-plane IOLs and signs of progressive endothelial damage and concluded that corneal decompensation may be prevented if the IOLs are removed before a critical degree of endothelial cell loss or dysfunction occurs.48 This requires serial evaluation with documentation of a reliable trend that considers the accuracy and reproducibility of each test. Evaluation should include assessment of changes in cell density, morphology, or pachymetry, as well as examination for factors that would suggest IOL-induced corneal compromise (IOL design, position, stability, proper size, evidence of erosion, or inflammatory precipitates on the IOL surface or corneal endothelium). If a patient has adequate endothelial reserve but demonstrates a clinical course that indicates likelihood of corneal decompensation, an early exchange procedure may forestall corneal transplantation. This spares the patient from potential problems, such as allograft rejection, corneal astigmatism, and prolonged corneal wound healing. Considering IOL exchange in these patients requires an assessment of the risks and benefits of observation compared with intervention. A patient with a closed-loop–style AC-IOL who continues to demonstrate good vision and a stable cell count is certainly at less risk with continued observation than a patient with documented progression of endothelial cell loss with marginal cell reserve (e.g., 500 cell/mm2 centrally [Fig. 4]). Intervention carries the risk of making the patient acutely worse, through surgically induced corneal decompensation, hemorrhage, iridodialysis, and failed or incomplete IOL removal. This is because the closed-loop AC-IOL designs may erode into the angle and ciliary body. Removal techniques must respect this erosion to preserve the integrity of the anterior segment and to avoid complications.5 The goal of IOL exchange in the presence of a clear cornea is to avoid the need for a more involved procedure later, which will be more expensive, require a longer recovery, and be associated with more
Fig. 4. Erosion of a Leiske-style, closed-loop AC-IOL into the peripheral cornea produced “corneal fixation.” This was the patient’s best eye and it maintained a stable endothelial cell count and excellent vision for over 10 years with observation. Deterioration in vision or accelerated endothelial cell loss would provide motivation for IOL exchange.
potential complications. However, patients with low endothelial cell counts (less than 500 cells/mm2) often demonstrate increased stromal thickness by pachymetry and are likely to develop acute corneal decompensation after surgical intervention. These patients are better managed with IOL exchange combined with penetrating keratoplasty after visually significant corneal decompensation occurs. Patients with pre-existing progressive corneal disorders, such as Fuchs’ corneal endothelial dystrophy, would obviously continue to demonstrate progression associated with their underlying corneal disorder despite IOL removal. It appears that all AC-IOLs accelerate endothelial cell loss to some degree. This may be an acceptable situation in older patients receiving modern versions of the Kelman-style implant (e.g., Multiflex style) that are properly sized and do not have positioning holes at the end of the haptic loop that can cause synechiae.14,155
Retinal Indications RETINAL DETACHMENT
Retinal detachment rarely requires explantation of the IOL.14 Lens removal should be considered to repair retinal detachment in patients 1) who have concomitant pathology creating other indications for IOL removal, 2) whose IOL obscures visualization of the detachment (which is more common in eyes with AC-IOLs and severe anterior proliferative vitreoretinopathy) or mechanically obstructs the surgery, 3) whose IOL may create specific intraoperative or postoperative management problems (e.g., condensation of a gas bubble on a silicone optic or
422
Surv Ophthalmol 42 (5) March–April 1998
the creation of silicone oil “droplets” on the surface of a silicone optic),17,65,76,91,144,235,253 or 4) who have significant capsular pathology (localized endophthalmitis, severe capsular contraction, or limited IOL support) that might contribute to IOL-related intraoperative or postoperative complications. CYSTOID MACULAR EDEMA
Cystoid macular edema is a well-recognized complication of intraocular surgery with or without IOL implantation.3,53,64,68,74,81,83,96–98,100,101,104,245 Potential mechanisms of pathogenesis include increased permeability of the perifoveolar capillaries, an ischemic tissue injury, a secondary response to intraocular inflammation, and direct traction on the macula after vitreous shifts.8
Incidence and Mechanisms Clinically, CME can be divided into three groups: 1) angiographic CME, in which vision is usually unaffected; 2) clinically significant CME, in which vision is decreased; and 3) chronic CME, defined as chronic or recurrent CME of more than 6 month’s duration. The presence of angiographic CME alone is of uncertain importance. In a study by Stark et al, 78% of patients with fluorescein-positive angiographic evidence of CME had 20/20 vision, and the rest had 20/25 vision.245 Clinically significant CME is associated with transient decrease in vision in up to 8% of patients after uncomplicated cataract extraction and is associated with chronic decrease in 1–2% of patients.49 A clinical study at the Wilmer Eye Institute found that clinically significant CME developed in a smaller percentage of cases (2% total incidence and 0.3% chronic) if the IOL implantation was uncomplicated.245 Visual potential probably correlates best with thickness of edema rather than size or severity of angiographic leakage.184 The peak incidence of CME after cataract surgery is approximately 4–16 weeks postoperatively, but the onset may be delayed months or even years after surgery.49,173 Spontaneous resolution occurs in 75% of cases within 6 months.49,79,82,83,104,105,173,221,284 Cystoid macular edema that causes permanent structural damage to the macula is an important cause of poor vision that persists despite IOL removal or exchange.26,153 Chronic CME may be associated with photoreceptor atrophy, lamellar hole formation, and reactive retinal pigment epithelium (RPE) changes in the fovea.80 Histopathologic examination supports the role of inflammation in the development of CME by documenting the presence of inflammation in the iris, ciliary body, vitreous, and retinal blood vessels. Breakdown of the blood-retinal barrier can be seen
CARLSON ET AL
with iris angiography and vitreous fluorophotometry.136,171 The degree and duration of breakdown of the blood-retinal barrier may correlate with the amount of intraoperative damage to the iris and the extent of the posterior synechiae.73 For this reason some authors have suggested that IOLs placed in the capsular bag may be better tolerated than those placed in the ciliary sulcus.73,172 Epidemiologic studies suggest that surgical techniques and IOL design play an important role in the development of CME. Jaffe et al found that only 3 patients in a series of 103 who underwent extracapsular cataract extraction (ECCE) with posterior chamber IOL implantation developed angiographic CME.125 The lower incidence of CME after ECCE versus intracapsular cataract extraction (ICCE) is supported by studies by Binkhorst,23 Kraff et al,139 Taylor et al,259 Stark et al,245 and Jaffe et al.124,125 Irissupported IOLs are associated with a higher incidence of CME. In a retrospective study by Taylor et al, clinical CME developed in only 2% of 630 eyes that underwent ICCE alone, versus 9.9% of 850 eyes that underwent ICCE combined with iris-supported IOL implantation.259 Rigid closed-loop AC-IOLs have also been associated with an increased incidence of CME. In a study by Maynor, CME developed in 1.9% of 207 patients who underwent ICCE and implantation of a Leiske-style IOL and in 5% of 302 patients undergoing ICCE and implantation of an Azar 91Z lens.162 Choyce found a 3% incidence of CME in a series involving 1,000 Mark IX rigid AC-IOL (Coburn, Clearwater, FL).41 Maintaining the integrity of the posterior capsule and vitreous appears to reduce the risk of CME.37 Nikica et al reported CME in 35.7% of cases of extracapsular cataract extraction complicated by posterior capsular rupture requiring anterior vitrectomy and insertion of an AC-IOL.183 This compares with a 1.5% incidence of CME in cases of uncomplicated extracapsular cataract extraction with insertion of a PC-IOL. Kraff et al randomly performed primary capsulotomies in 152 patients and left the capsule intact in 136 patients.138 Cystoid macular edema was evident, by angiography, in 5.6% of those with an intact capsule, as opposed to 21.5% of those with capsulotomy. However, overall visual acuity results were virtually identical in the two groups. Angiographic findings in patients with intact capsules did not correlate with visual acuity. Winslow and colleagues found angiographic evidence of CME in 48% of patients with vitreous prolapse after capsulotomy, whereas edema was demonstrable in only 10% of patients without vitreous loss and with intact capsules.280 Jaffe found that if capsulotomy is delayed for a year or longer after cataract extraction, CME is less likely to occur.126
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
Management Cystoid macular edema per se is only a relative indication for IOL removal or exchange, and it often occurs in the presence of other ocular inflammatory conditions, such as pseudophakic corneal edema, IOLinduced uveitis, and localized or generalized endophthalmitis. Therefore, chronic CME is infrequently the sole indication for IOL removal or exchange.158,234,239 The true incidence of CME as an associated finding in patients undergoing IOL exchange is unknown because of coexisting conditions, such as pseudophakic corneal decompensation. Isolated studies report the overall incidence of CME before IOL removal or exchange to be 3.3–5.5%.48,133,153,191,227,234 Although many authors report improvement in clinically observed CME and visual acuity after IOL removal or exchange, it is difficult to predict the response in an individual patient. Visual outcome may be limited by permanent structural changes of the fovea or from perpetuation of CME by mechanisms that are not directly attributed to the IOL. Smith reported significant improvement in visual acuity after IOL removal or exchange in some patients with chronic CME attributed to an IOL.239 In their study, 50% of the patients with CME achieved a significant improvement in vision after removal of the IOL. Lyle and Jin reported resolution of CME in 7 of 14 patients (50%) after IOL exchange; however, in their series of 101 eyes in 98 patients, the overall incidence of postoperative CME was higher than the preoperative incidence.153 In that study, the incidence of postoperative CME was significantly associated with the preoperative CME. Lyle and Jin also found that new CME was more likely to develop after penetrating keratoplasty combined with IOL exchange. They identified CME and glaucoma as the cause of a poor visual outcome in patients undergoing IOL exchange without penetrating keratoplasty. This is consistent with the findings of Busin et al, who identified glaucoma and CME as the most common cause of a final visual acuity worse than 20/60 after IOL exchange.26 In a retrospective analysis by Mamalis et al of 102 patients who underwent IOL removal or exchange, the postoperative vision improved in 39%, remained unchanged in 46%, and worsened in 15%.158 The most common cause for worsening of vision postoperatively was corneal decompensation, then glaucoma and CME, respectively. Coli et al studied 102 patients with AC-IOL– induced corneal endothelial damage who underwent IOL exchange.48 This group contained 47 patients (46%) with concomitant CME before IOL exchange. Patients with low endothelial cell counts (less than 900 cell/mm2) and either a rigid-loop– or closedloop–style AC-IOL had a higher incidence (71%) of preoperative CME. Posterior chamber-IOLs were
423
substituted for AC-IOLs in all cases, with 87 sutured to the iris, 3 sutured to the sclera, and 12 inserted into the ciliary sulcus. Among all patients with CME, visual acuity improved in 30 eyes (64%), remained unchanged in 6 eyes (13%), and worsened in 11 eyes (23%) after exchange for an iris-sutured PC-IOL and vitrectomy. In another study, penetrating keratoplasty, ACIOL removal, anterior vitrectomy, and insertion of an iris-sutured PC-IOL resulted in angiographic resolution of CME in 18 of 25 patients diagnosed with CME and pseudophakic corneal edema.203 Improvement of visual acuity to 20/40 or better was achieved in 16 patients, with visual recovery usually occurring within a year of surgery. Stark et al reported three patients who underwent IOL exchange because of pain with CME.246 Exchange of an AC-IOL for a sutured PC-IOL in the absence of capsular support relieved the pain in all the patients, and the CME improved in two of them.246 Two additional patients who underwent IOL exchange because of CME alone also demonstrated improved vision and reduced CME. Insertion of a three-piece–C-loop PC-IOL in the anterior chamber has been associated with inflammatory complications, including CME and epimacular membrane formation. Liu et al reported improvement in visual acuity and reduced CME in two patients who underwent exchange of a three-piece– C-loop PC-IOL that had been inserted into the anterior chamber.152 Many of the data addressing CME and visual outcome after IOL exchange are retrospective, and there are little data from prospective controlled interventions. Schein et al compared IOL exchange techniques in a randomized prospective study of 176 consecutive patients undergoing penetrating keratoplasty with IOL exchange.227 They found that the cumulative risk of CME with iris-fixated PC-IOLs was significantly less than that with AC-IOLs or transsclerally sutured PC-IOLs. However, the number of resolved cases and the number of newly developed cases of CME were not reported. Also, the cumulative risk of CME after penetrating keratoplasty and IOL removal alone was not determined. Three retrospective uncontrolled studies have been reported, two comparing open-loop–Kelman-style ACIOLs with iris-fixated lenses and one comparing openloop–Kelman-style AC-IOLs, iris-fixated PC-IOLs, and transscleral fixated PC-IOLs.54,99,146 None of these studies reported any differences in visual outcome or complication rates after IOL exchange among the different techniques. Harbour et al treated 24 consecutive patients with chronic pseudophakic CME with pars plana vitrectomy without IOL removal.94 Preoperatively, the pa-
424
Surv Ophthalmol 42 (5) March–April 1998
Fig. 5. Erosion into the peripheral iris is common with closed-loop AC-IOL designs, such as the Azar, producing low-grade inflammation with the potential for CME and corneal decompensation.
tients had evidence of either vitreous adhesions to anterior segment structures (23 patients) or iris capture of the IOL (1 patient), and all had responded poorly to medical therapy. In all patients, pars plana vitrectomy with removal of vitreous adhesions improved the visual acuity, with a mean improvement of 4.6 Snellen lines. The surgeon performing IOL removal or exchange for the treatment of visually significant CME should direct meticulous attention toward alleviating potential causes of chronic inflammation, such as vitreocorneal or iridocorneal adhesions, iris incarceration, IOL capture, IOL-induced iris chafing, or any vitreous traction (Figs. 5–8). This may reduce the sources of inflammation responsible for CME. Cystoid macular edema is a potential complication of any intraocular procedure, and not all pseudophakic
Fig. 7. Inflammation results from vitreous traction in addition to IOL migration through a peripheral iridectomy, producing erosion into the ciliary body.
CARLSON ET AL
Fig. 6. Vitreous entanglement with the IOL (traction at superior optic near the loop—optic junction) may produce traction with low-grade inflammation, producing CME.
patients with CME benefit from IOL removal or exchange. Patients with concomitant inflammation and clinical signs that suggest that the IOL is the cause of ongoing CME are the patients most likely to benefit from IOL removal or exchange. Surgical intervention should be reserved for cases of CME refractory to medical treatment. Cystoid macular edema may also be complicated by the development of an epimacular membrane or foveal hole formation.
Inflammatory (Noninfectious) Indications MICROHYPHEMA
Microhyphema has been noted in approximately 2–5% of anterior chamber and iris-fixated lenses after cataract extraction.66,150,283 Microhyphema associated with increased IOP and uveitis in the presence of an anterior segment lens (uveitis-glaucoma-hyphema syndrome) is discussed in the next section.
Fig. 8. Synechiae encases the loop and peripheral eyelet of a Pannu-style AC-IOL, producing CME from low-grade inflammation.
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
425
Fig. 9. Active bleeding from wound neovascularization seen by gonioscopy.
Fig. 10. Recurrent hyphema is associated with intermittent iris capture with chafing. Note the fresh clot where the IOL is in contact with the iris.
Before the development of modern surgical techniques, recurrent hyphema was more common and was associated with vascular ingrowth of the incision (Fig. 9).187,257,275 Microhyphema also has been reported after PC-IOL implantation.88,129,193 The incidence of late bleeding in the anterior segment after PC-IOL implantation has been reported to be 0.4%.283 The patients may have a history of floaters or intermittent visual decrease lasting for several hours. Some of these patients are on chronic coumadin therapy.156 Bleeding usually results from IOL trauma to the iris162 (Fig. 10) or ciliary body.194 Iris fluorescein angiography may demonstrate hyperfluorescent leakage at the bleeding site.150 On examination, the microhyphema may be observed inferiorly in the anterior chamber by slit-lamp biomicroscopy or may require gonioscopy to be visualized.156 Suspended red blood cells may be observed in the aqueous. The IOP associated with a microhyphema may be elevated. Magargal et al reported that an increased IOP occurred after a microhyphema in two of five cases.156 Treatment of microhyphema involves preventing recurrent hemorrhage by chronic topical cycloplegics,150 argon laser photocoagulation,13,182 or Nd:YAG laser photocoagulation88 to the bleeding site. Aqueous suppressants (e.g., b-adrenergic blockers, carbonic anhydrase inhibitors, apraclonidine) may be prescribed to treat an elevated IOP. Removal of the IOL has rarely been required to treat recurrent uncontrolled vitreous hemorrhage associated with chronically reduced vision.194
chanical injury to the iris and breakdown of the blood-aqueous barrier.8 Improvements in IOL design and finish have made both of these syndromes largely extinct. Intraocular lens-induced inflammation may still occur in association with excessive uveal contact resulting from IOL chafing or erosion, IOL malposition, or improper IOL sizing (particularly with AC-IOLs).269 Ocular inflammation resulting from IOL touch to the iris or the ciliary body may increase the IOP by several mechanisms. Inflammatory cells or debris may occlude the trabecular meshwork.16,119,159,223 Uveitisinduced adherence of the iris to the IOL optic may cause pupillary block.216 Glaucoma may be caused by the topical steroids used to treat the uveitis.103,149,200 In addition to inflammatory changes, outflow may be reduced by direct mechanical injury to the trabecular meshwork or peripheral anterior synechiae. The incidence of secondary glaucoma resulting from uveitis after lens implantation is 2.3% with an ACIOL, 1.7% with an iris-fixated lens, and 0.8% with a PC-IOL.283 Evidence of uveitis on clinical examination includes anterior chamber cell and flare, keratitic precipitates, and pigment debris in the inferior trabecular meshwork, as noted by gonioscopy. Gonioscopic findings in the uveitic eye may be compared with those in the uninvolved eye to determine increased angle pigment.216,251 If the increased IOP is secondary to inflammation, it should be controlled when the inflammation is controlled by topical steroids. The minimum dose of topical steroids required to control the inflammation should be used, because a steroid-induced IOP increase may occur after 3–4 weeks in sensitive individuals.39,251 Aqueous suppressants may reduce the IOP. Treatment methods that depend on altering outflow facility, such as topical miotics and argon laser trabeculo-
UVEITIS AND GLAUCOMA
Postimplantation uveitis was formerly associated with the toxic lens syndrome, which was related to IOL polishing and sterilization and the uveitis-glaucoma-hyphema syndrome, which resulted from a poorly designed and finished IOL that caused me-
426
Surv Ophthalmol 42 (5) March–April 1998
plasty, may also be tried, but might be less effective in eyes with uveitis.251 When IOP cannot be controlled by these methods, filtration surgery may be required. Occasionally, persistent uveitis that is causing secondary complications may necessitate removal of the IOL.251 INTRAOCULAR LENS-INDUCED PAIN
Intraocular lens-induced pain may be caused by local irritation or from a more generalized inflammatory response. It is not uncommon for patients with AC-IOLs to experience low-grade discomfort and tenderness with local pressure. The tenderness is often noted in the region of IOL erosion into the ciliary body and is more apparent with IOLs that have a diameter that is too large. Patients with PC-IOLs rarely have such discomfort, although Legler et al reported a patient with chronic, vague complaints of discomfort that improved after PC-IOL repositioning.148 They hypothesized that the PC-IOL loop was causing local irritation. An analysis of 6,521 explanted IOLs evaluated at the Center for IOL Research (Charleston, SC) revealed that pain was clinically present in 32.9% of patients who underwent AC-IOL removal compared with 17.7% of patients who underwent PC-IOL removal. Pain is subjective and usually not identified separately from inflammatory complications in series that address indications for IOL exchange.
Infectious Indications GENERALIZED ENDOPHTHALMITIS
Incidence Infectious endophthalmitis is an uncommon complication after modern cataract surgery with IOL insertion.36,61,67,115,123,225,226 An IOL is essentially a foreign body and, like any prosthetic device, it requires additional surgical manipulation for implantation.36 This could potentially create a higher incidence of infectious endophthalmitis in the pseudophakic eye compared with the aphakic eye; however, most studies show that this is not the case.67,247 Kattan et al reported that the risk of endophthalmitis after a standard ECCE with or without primary IOL implantation is 0.072%, but that it increases to 0.3% with a secondary IOL.131 Javitt et al reported an incidence of 0.12% in a national multicenter study.127,128 Gaping wounds are a major risk factor for infection and are present in 65% of all cases of postsurgical endophthalmitis. Stonecipher et al reported three cases after sutureless cataract surgery.254 Williams and Gills reported 27,181 consecutive cases of sutureless cataract surgery with an incidence of infection of 0.015%.278
CARLSON ET AL
Driebe et al reported one of the largest series of pseudophakic endophthalmitis.61 A bacterial source was isolated in 57 of 83 cases. Only one case required IOL removal to clear the infection. In this series, the causative organisms differed little from isolates responsible for infections in patients without IOL implantation. Risk Factors Several highly publicized clusters of endophthalmitis have been associated with contamination of solutions used for sterilizing and irrigating.77,78,85,86,120,179, 185,196,213,214,250 Clayman et al reported the potential for bacterial contamination of the phacoemulsification tip.45 Filtration of irrigating solutions along with the perioperative use of antibiotics may reduce postoperative endophthalmitis.87,89,230,240,248 The potential for contaminating the IOL intraoperatively was demonstrated by Vafidis et al,264 who reported that IOLs placed on the periocular surface before being implanted were contaminated 26% of the time. Most surgeons acknowledge the importance of avoiding IOL contact with the eyelids and lashes to reduce the risk of infectious complications. The importance of meticulous, sterile surgical technique and the appropriate preoperative treatment of pre-existing infections, such as blepharitis or conjunctivitis, is generally accepted. Menikoff et al reported that multipiece IOLs with polypropylene loops represent an independent risk factor for the development of postoperative microbial endophthalmitis.166 Additional work by Raskin et al suggested a mechanism of greater bacterial adherence to polypropylene compared with polymethylmethacrylate.209 Clinical Presentation and Signs Acute onset, generalized endophthalmitis characteristically develops within a few days after intraocu-
Fig. 11. Acute, generalized streptococcal endophthalmitis after cataract surgery.
427
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
rapidly after surgery. Occasionally, a virulent organism may create severe endophthalmitis that is delayed in onset if inoculation occurs after surgery owing to a suture tract or wound abnormality (Fig. 12).192 Management
Fig. 12. Delayed onset, generalized endophthalmitis associated with an infected filtering bleb.
lar surgery. Patients typically experience the rapid development of severe ocular pain associated with reduced vision, redness, tearing, and photophobia. Occasionally, they may present without pain.56 Clinical signs (Fig. 11) include conjunctival inflammation with chemosis; episcleral inflammation; eyelid edema with blepharospasm and secondary ptosis; corneal stromal edema secondary to endothelial dysfunction; anterior chamber inflammation, including hypopyon, cell, flare, keratic precipitates, and IOL precipitates; secondary glaucoma; uveal inflammation, including engorged iris vessels, iris nodules, posterior synechiae, and peripheral anterior synechiae; and inflammation of the posterior segment, including vitreous inflammation, inflammatory membranes, macular edema, and retinal necrosis.4,72,170 Patients who develop postoperative endophthalmitis with virulent organisms tend to have a generalized inflammatory process that is severe in nature and occurs
Successful treatment of generalized endophthalmitis requires prompt recognition and initiation of antimicrobial therapy. It usually does not require removal of the IOL, unlike infections in other parts of the body that may require removal of any associated prosthetic device to adequately clear the infection. Removal of the IOL may be required in patients who worsen while on maximum medical therapy. This is more likely to occur in patients with forms of endophthalmitis involving mycotic infections50 or localized anaerobic infections sequestered within the capsular bag. LOCALIZED ENDOPHTHALMITIS
Virtually all microbial organisms have the potential to become intraocular pathogens. Less virulent organisms may result in infectious endophthalmitis that has a delayed onset. Indolent microbial organisms may sequester within the capsular sac, creating a localized, smoldering infection.34,61,164,165,190,197,219,237, 260,261 Apple et al coined the term localized endophthalmitis to describe this condition, indicating that it should be an important consideration in the differential diagnosis of otherwise unexplainable inflammation after cataract surgery.8,10,197 Clinical Features Patients with localized endophthalmitis tend to develop inflammation that is characteristically delayed in onset, chronic, and often low grade; however, severity may be variable. Inflammation typically
Fig. 13. A: Intraoperative photo of a localized plaque within the capsular bag surrounding the polypropylene IOL loop in a patient undergoing partial capsulectomy with removal of the sequestered infection, IOL exchange, and endocapsular antibiotic inject. B: Histology (Brown and Brenn stain) showing Propionibacterium acnes densely sequestered within the capsular bag. There is an absence of inflammatory cells.
428
Surv Ophthalmol 42 (5) March–April 1998
Fig. 14. Propionibacterium acnes endophthalmitis presenting with an endocapsular plaque and chronic recurrent inflammation that is responsive to topical corticosteroids.
improves with corticosteroids, and this response may further delay the diagnosis. The patient’s symptoms of pain, reduced vision, redness, tearing, and photophobia are often less intense than in patients with acute generalized endophthalmitis. Signs of inflammation may also be less severe than those of acute generalized endophthalmitis. An endocapsular plaque (Figs. 13 and 14) may be present, representing sequestered organisms. Stefansson et al reported an endocapsular hypopyon as a clinical sign of localized bacterial endophthalmitis.249 Meisler et al were the first to recognize the importance of Propionibacterium acnes as a cause of delayedonset endophthalmitis after cataract extraction.164 They reported six patients with chronic P acnes endophthalmitis who previously would have been classified as having toxic-lens syndrome. The initial cataract surgery was routine and without complication, and each patient achieved good visual acuity before the onset of inflammation. On average, inflammation developed 4 months after surgery and improved with topical corticosteroids. The etiologic diagnosis was delayed an average of 10 additional months. These cases developed despite the use of perioperative antibiotic prophylaxis and were not specific to an individual surgeon or IOL. Although not specific to an individual IOL, reported cases of localized endophthalmitis have occurred almost exclusively in patients with IOLs with polypropylene loops. Piest et al studied the development of localized endophthalmitis in patients clinically diagnosed with toxic-lens syndrome.197 In each case, the differential diagnosis included biomaterial-related inflammation and phacoanaphylaxis. Each patient underwent removal of the IOL and lens capsule with particular care to ensure that no lens material remained in the eye. Specimens submitted for histopathologic evalu-
CARLSON ET AL
ation revealed Gram-positive organisms, suggestive of P acnes, that were sequestered in the residual lens capsule. Piest et al reported an additional case in a patient who did not have an IOL, verifying that only the presence of the capsular sac is necessary for this condition to develop. The IOL may play a role, however, in inoculating the bacterial organism.6,36,166,209, 244,260 Also unknown is the role of residual lens cortical epithelium in the development of a localized infection. Clinically, patients with localized endophthalmitis correlate well with the experimental studies in nonhuman primate eyes of Beyer et al, who reported the protective barrier effect of the posterior lens capsule in exogenous bacterial endophthalmitis.20,21 Infection spread to the vitreous cavity in only 20% of eyes that had an intact posterior capsule during the experiment. This compared with an 80% rate of subsequent vitreous infection when a primary posterior capsulotomy was performed. Localized endophthalmitis may create the clinical appearance of posterior capsular opacification (PCO).11 There are at least four cases reported of inflammation precipitated in otherwise quiet eyes by Nd:YAG laser capsulotomy.34,35,165,261 Tetz et al reported a case of P acnes endophthalmitis precipitated by Nd:YAG laser posterior capsulotomy for PCO performed 10 months after the primary surgery.261 There was no evidence of inflammation before the laser procedure, and their case was eventually treated with removal of the IOL and capsule.71 Visual acuity returned to 20/50, limited by CME. Meisler et al reported a case of P acnes endophthalmitis developing within a month after Nd:YAG laser capsulotomy in a patient with PCO without intraocular inflammation.165 The laser procedure was performed 3 months after the primary surgery, and their patient required removal of the IOL and capsule despite aggressive antimicrobial treatment. Visual acuity returned to 20/60, limited by CME. Carlson and Koch described a patient who developed P acnes endophthalmitis after Nd:YAG laser posterior capsulotomy performed 4 months after cataract surgery for PCO in an eye that was otherwise without evidence of inflammation.34 Their case demonstrated the potential for successful treatment without removal of the IOL or capsule. Their patient returned to 20/25 visual acuity despite mild CME and an epimacular membrane. Carlson et al reported another patient who developed Staphylococcus aureus endophthalmitis after Nd:YAG laser capsulotomy for PCO that developed in an otherwise quiet eye.35 The capsulotomy was performed 2 years 8 months after the original cataract surgery. The isolation of S aureus was surprising because this species is generally more pathogenic than P acnes. The organism apparently remained viable within the cap-
429
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
sule for an unusually long time without causing intraocular inflammation. This patient responded to antimicrobial and anti-inflammatory treatment without IOL removal and returned to a visual acuity of 20/25 11. In addition to the above four cases, four additional cases (three P acnes, one S epidermidis) have been reported with inflammation prior to capsulotomy that progressed and were possibly aggravated by the laser procedure.1,164,181,197 Piest et al reported a patient who underwent IOL exchange for presumed toxic lens syndrome.197 The lens capsular sac was retained and the inflammation persisted, although it improved with intense corticosteroid treatment. The patient eventually underwent Nd:YAG laser capsulotomy for PCO and developed recurrent hypopyon. The patient improved after complete removal of the capsular sac and all cortical remnants combined with anterior vitrectomy. Abrahams reported worsening of P acnes endophthalmitis after Nd:YAG laser capsulotomy for PCO in a patient who appeared to have had a localized infectious plaque on the corneal endothelium without clear documentation of PCO resulting from P acnes.1 The patient underwent sector iridectomy, removal of the retained inflammatory material, and removal of the IOL and the capsular bag along with anterior- and midvitrectomy. Histopathologic and cytologic studies of the capsular bag were unable to identify evidence of a local infection. Laboratory Diagnosis Aqueous and vitreous cultures may be negative when the infection is localized within the capsular bag. If a localized process suspicious for an infection is clearly visible within the capsular bag, laboratory evaluation should include material from the sequestrum. The material should undergo Gram stain and direct inoculation onto bacterial and fungal media. To improve sensitivity and specificity, media for anaerobic bacteria should include at least one broth (e.g., Thiol, Thioglycolate, or chopped meat enriched broth) and one agar plate (e.g., prereduced Brucella blood agar, supplemented Columbia base agar, or CDC anaerobic blood agar). The laboratory should be specifically instructed to incubate the specimen for at least 14 days before discarding it as negative and to consider organisms such as P acnes as pathogens and not laboratory contaminants. Organisms obtained in low number or under the influence of previous antibiotic therapy may take longer to grow than the standard 7-day incubation period recommended by the American Society for Microbiology. Isolating the organisms on multiple media decreases the likelihood that a contaminant or false-positive culture has occurred.
Indications for Intraocular Lens Removal Based on the literature regarding the medical and surgical management of patients with localized endophthalmitis, it is difficult to predict which patients will respond to treatment without removal of the IOL and capsular bag. Problems in interpreting the literature include the small number of well-documented cases and the variability of case definition and spectrum of case severity and treatments. Selection bias favors reporting cases that undergo IOL removal, as many of the early series were not diagnosed until the specimen was evaluated by histopathology. Isolated cases that are successfully managed without IOL removal are probably less likely to be reported than cases that involve IOL removal or exchange. Removal of the IOL and capsular bag should be considered in patients who do not respond adequately to limited capsulectomy, removal of the sequestrum, and injection of antibiotic within the capsular bag. We recommend intraoperative injection of 1 mg of vancomycin in patients suspected with P acnes, and we may repeat this postoperatively at the slit-lamp. The techniques for capsular biopsy and IOL-capsular bag removal are described elsewhere.33
Optical Indications INCORRECT INTRAOCULAR LENS POWER
Incorrect IOL power selection results in an unsatisfactory refractive error. Extreme postoperative refractive surprises have been reduced by improved understanding of the limitations of the older, empirically derived IOL calculation formulas and improved accuracy in the preoperative measurement of axial length and corneal curvature. Potential sources of error in determining the power of an IOL include 1) incorrect axial length measurement; 2) incorrect corneal power determination, particularly in patients with previous refractive surgery; 3) error associated with incorrect use or selection of IOL calculation formulas, including the reduced accuracy of empirically derived formulas with unusually long or short eyes; 4) alteration of the effective IOL power by placing the optic in a position different than expected by the preoperative calculation, such as sulcus placement instead of capsular bag placement; 5) insertion of the wrong IOL during surgery; 6) incorrect power labeling or incorrect A-constant or “surgeon factor” of the IOL; or 7) multiple sources of error that can produce an unanticipated refractive result. The majority of surprise errors are explained by remeasuring the preoperative parameters. Sources of error can be reduced by 1) measuring both eyes preoperatively for comparison; 2) repeating any measurement that is statistically unusual by
430
Surv Ophthalmol 42 (5) March–April 1998
CARLSON ET AL
Fig. 15. A: Open-loop, Kelman-style AC-IOL placed horizontally to avoid peripheral iridectomy superiorly. B: The same eye, 2 1⁄2 years later with IOL rotation and peaking of the pupil. This process may develop in association with eye rubbing and/or a size disparity between the IOL and the dimensions of the anterior chamber angle.
itself or in comparison with the other eye; 3) comparatively cross-checking keratometry with computerized corneal analysis; 4) educating the technical staff to improve consistency, minimize distraction during testing, and alert the surgeon to any difficulty encountered during measurement; 5) determining the accuracy postoperatively in preparation for IOL selection in the second eye and looking for consistent errors that might require modification of the A-constant or “surgeon factor;” and 6) anticipating tolerance of error and adjusting surgery accordingly—e.g.: targeting for 20.5 sphere in a patient who desires emmetropia, as an error in the hyperopic direction is tol-
Fig. 16. Sunrise syndrome—Asymmetric IOL loop fixation in combination with postoperative capsular contraction results in unopposed forces on the inferior IOL loop with resulting migration of the IOL superiorly.
erated less well in the nonaccommodating pseudophakic eye than an error in the myopic direction. MALPOSITION OF THE INTRAOCULAR LENS
Malposition of the IOL may be the result of the original surgical placement of the lens, or it may develop postoperatively because of external or internal forces. External forces include trauma or eye rubbing (Fig. 15). Internal forces include a size disparity between the IOL and the site of fixation, scarring, peripheral anterior synechiae, and capsular contraction. Decentration of tilting of an AC-IOL may produce optical problems from aberration or edge glare.
Fig. 17. Pea pod effect—Postoperative capsular contraction with one or more tears in the anterior capsule produces forces capable of extruding one or both loops out of the capsule, resulting in IOL decentration.
431
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES TABLE 2
Advantages of Capsulorhexis 1. Assures capsular bag fixation to improve short- and long-term IOL centration. 2. Allows use of newer IOLs with smaller optic and shorter overall length, as well as use of foldable IOL designs that tend to decenter without capsulorhexis. 3. Improves the safety of surgical techniques involving nuclear cracking or chopping. 4. Improves the efficacy of surgical techniques such as cortical-cleaving hydrodissection and facilitates cortical removal by expanding the capsular bag and decreasing aspiration of the capsular flaps. 5. Provides fixation that avoids uveal contact. 6. Improves IOL power calculations by increasing predictability of IOL location. 7. Produces less zonular stress than is created during puncture techniques. 8. Provides better support for sulcus-fixated IOLs in cases developing significant intraoperative disruption of the posterior capsule. 9. Potentially reduces posterior capsular opacification by facilitating more complete cortical removal in addition to enhanced contact inhibition from apposition between the optic and the capsule. 10. Provides an additional barrier against scleral collapse during penetrating keratoplasty when performing a triple procedure. IOL 5 intraocular lens.
Optical complications caused by wrong IOL power or malposition comprise the most common indication for removal or exchange of a modern PC-IOL.14 Although malposition of a PC-IOL may reduce the quality of vision, it is less likely than malposition of an AC-IOL to cause mechanical injury or inflammatory sequelae. The pathogenesis of malposition may be related to a variety of locations of loop fixation, to the forces of capsular contraction, or to a combined mechanism. Before the development of capsulorhexis, it was common for the surgeon to place the inferior IOL loop within the capsule while releasing the superior loop into the ciliary sulcus, thus producing
Fig. 18. Capsular contraction syndrome producing appositional fibrosis and contraction of the anterior capsular opening, evidence of capsular tension with stria formation and mild decentration of the IOL from capsular forces.
asymmetric loop fixation. Because significant decentration was anticipated, PC-IOLs had large optics (7 mm) and long total lengths (13–14 mm). Subsequent healing from capsular fusion and contraction potentially caused the inferior loop to exert forces on the optic unopposed by forces from the superior loop within the sulcus. Migration of the optic superiorly produces the “sunrise syndrome,” which may cause symptomatic edge glare or aberration (Fig. 16). A tear in the anterior capsule may allow one or both IOL loops to migrate out of the capsular bag under the forces of capsular contraction, producing “pea podding” (Fig. 17). Capsulorhexis is a major surgical advance that contributes to long-term IOL centration and reduces the likelihood of needing IOL removal or exchange (Table 2). Despite an intact capsulorhexis, dencentration may still occur, as seen in the capsular
Fig. 19. Capsular contraction with dense fibrosis of the anterior capsule and complete closure of the anterior capsular opening.
432
Surv Ophthalmol 42 (5) March–April 1998
CARLSON ET AL
Fig. 20. A: An asymmetric or eccentric capsulorhexis may allow capsular fusion peripheral to the optic with progressive adherence. B and C: This fusion produces IOL decentration from the edge of the capsulorhexis.
contraction syndrome. This has been described by several authors, including Shepherd,229 Hansen,93 Davison,55 and Masket.161 The emphasis of the original reports was the development of closure of the anterior capsular opening.18 Many of these cases develop capsular striae, a sign demonstrating that cap-
Fig. 21. Visually significant decentration of a siliconeplate design IOL requiring exchange. This design is less able than a loop design to compress within the IOL plane. Forces of capsular contraction cause the plate lens to decenter, tilt, rotate, or buckle within the capsular bag.
sular forces can produce IOL decentration (Figs. 18 and 19). This complication can be prevented by a large capsulorhexis. In a separate but related mechanism, an eccentric capsulorhexis may allow the capsulorhexis edge to play a large role in the mechanism of decentration. The capsulorhexis edge may be more peripheral to the optic in one area, with fusion developing between the anterior and posterior capsule in this region. This adherence may propagate until the capsulorhexis edge comes in contact with the optic or IOL loop, producing decentration away from the area of contact (Fig. 20).30 To reduce optically significant decentration, we recommend a large capsulorhexis that is symmetrically round and centrally located. The silicone-plate–design IOL is particularly susceptible to the forces of capsular contraction and may decenter, rotate, tilt, or buckle because of several properties specific to the plate design (Fig. 21). With forces from capsular compression, the plate lens is unable to compensate as well as a loop-design lens, which essentially adjusts for disparity between the size of the IOL and capsular bag. Plate-design lenses have a smaller arc of contact with the capsular fornix, reducing anchoring forces that normally reduce the potential for rotation and decentration. Sil-
433
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
Fig. 22. Posterior dislocation of a silicone plate IOL after Nd:YAG laser capsulotomy.
icone does not adhere well to the capsule, and compression produces folding or buckling out of the plane of the IOL rather than compression within the plane, as with a loop-design lens. An additional concern is the potential for dislocation into the vitreous after Nd:YAG laser capsulotomy (Fig. 22). Carlson et al reported 10 cases of silicone-plate IOL dislocation and attempted to identify specific risk factors for this complication.32 Eight of 10 cases occurred despite waiting more than 3 months after cataract removal before performing the laser surgery. This complication was not specific to an individual surgeon nor was there anything unusual about the IOL powers or axial lengths. Evidence of capsular tension was present in at least 6 of 10 patients who developed a capsular split during the first few laser shots. In all cases, there was no immediate dislocation of the IOL, but rather spontane-
ous dislocation occurring several days to weeks after the procedure. Most patients retained excellent vision, although 6 underwent IOL removal and 1 developed a retinal detachment resulting in vision at the level of counting fingers. Recent plate lens modifications, including enlarged holes within the plate or the presence of a loop extending from the plate, may provide additional anchoring to resist the forces associated with capsular contraction. Newer IOL designs, improved surgical techniques, and a greater understanding of IOL fixation and postoperative capsular contraction should reduce the incidence of IOL malposition. Presently, management options for IOL malposition include 1) observation, 2) pharmacologic miosis, 3) repositioning (Fig. 23), 4) repositioning combined with McCannel suture fixation (Fig. 24), 5) removal (Fig. 25), or 6) exchange. The decision to retain and reposition the IOL with or without a McCannel suture requires that the IOL be of the appropriate power and design, and be capable of adequate fixation with acceptable risk and with long-term safety and stability. In the absence of these characteristics, IOL removal or exchange should be considered. Selection of treatment is based on the patient’s symptoms, visual needs and expectations, and an assessment of which option is likely to provide the best long-term benefit with the least risk.
Patient Counseling Patients faced with removal or exchange of an IOL will have various reactions depending on their level of understanding, expectation, and emotional stability. Emotions may range from relief that their problem may be solved to anger toward their original surgeon. In a discussion of IOL removal or exchange, the content of the message should offer
Fig. 23. A: This patient was referred after a small capsulorhexis resulted in retained subincisional cortex and the original surgeon attempted sulcus fixation of the IOL. B: Malposition of the IOL was treated with enlargement of the capsulorhexis, removal of the cortex, and repositioning the IOL into the capsular bag. This procedure was the least invasive and provided the best method of achieving the surgeon’s original objective.
434
Surv Ophthalmol 42 (5) March–April 1998
CARLSON ET AL
Fig. 24. A: Visually significant PC-IOL dislocation resulting from inadequate capsular support. B: Successfully managed with a McCannel suture supporting the IOL loop to the peripheral iris. A single suture is usually adequate in patients that have residual capsule supporting at least one loop. Absence of capsular support may require an additional suture to avoid a “windshield wiper” effect.
adequate information for the patient to make an informed choice regarding additional surgery. The delivery should offer a cushion of hope around the delivery of “bad news.” The emotional environment should be maximally supportive. Helpful techniques include in-person rather than over-the-phone discussion and avoidance of interruptions or a rushed appearance during the consultation. The presence of close family members and any technical staff that the patient may be close to can provide support to the patient.206 Verbal and nonverbal feedback during counseling will help assess how the patient is comprehending and receiving the news. The staff may alert the ophthalmologist to patients’ comments about recent bad news regarding health or family stress, indicating that it may not be the most appropriate time to discuss IOL removal. Family support during the patient counseling
Fig. 25. Intraocular lens removal is the best option in a patient with an anterior chamber IOL and recurrent granulomatous uveitis unresponsive to medical management.
may be particularly helpful for patients who may have trouble remembering everything that was discussed. Most patients, when confronted with the possibility of IOL removal or exchange, are already aware of problems surrounding their previous surgery and exhibit relief and optimism when additional treatment is offered to improve their present situation or minimize future complications.
Conclusion There are few decisions in ophthalmic surgery that are as complex and problematic as the decision to remove or exchange an IOL. While some decisions appear clear-cut, e.g., removing a closed-loop AC-IOL during penetrating keratoplasty, other cases are less certain, e.g., when the patient has excellent vision that requires chronic topical corticosteroids to control inflammatory complications or when a clear cornea has IOL-induced accelerated corneal endothelial cell loss. Issues that help the surgeon determine whether or not the risk-to-benefit ratio favors IOL exchange over continued observation include the following: 1) severity, duration, and chronology of the problem; 2) response to nonsurgical treatments; 3) natural history of a specific IOL (e.g., closed-loop design AC-IOL); 4) likelihood that surgical removal would provide substantial relief or benefit; 5) ease of surgical removal and potential for aggravating or creating additional complications; 6) status of the other eye; 7) patient and family expectations and visual needs; and 8) life expectancy and overall health of the patient. The surgeon must balance these issues and determine the potential for visual rehabilitation and the likelihood that surgical intervention will improve the patient’s quality of life and ability to function. The surgeon should seek consultation if the prob-
435
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES
lem is incompletely defined or if the surgeon lacks experience with the more technically challenging IOL exchange procedures. The patient must fully understand the treatment options, including the potential risks associated with both surgical and nonsurgical treatments. Preoperative counseling allows the patient to develop reasonable expectations and make an informed choice regarding care. The frequency and spectrum of IOL-related complications is changing as a result of improved surgical techniques, instrumentation, and improved quality of IOLs. Patients continue to develop higher expectations and are less tolerant of surgical failure. Patients are also living longer and the trend has clearly developed toward the use of IOLs in younger patients. While IOL-related problems are less frequent and less severe overall, they may become more apparent in younger patients with longer follow up.
Methods of Literature Search Search terms included various combinations of the terms listed in the “Key words” section of this article. A Medline search was performed from 1985 to present through the National Library of Medicine. Articles obtained from the reference list of other articles were reviewed and included when considered appropriate. Criteria for inclusion included the present clinical value or the original importance of the article to a particular subject.
13. 14.
15. 16. 17. 18. 19.
20.
21.
22. 23. 24.
25.
References 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12.
Abrahams WI: Propionibacterium acnes endophthalmitis: An unusual manner of presentation. J Cataract Refract Surg 15:698–701, 1989 Aiello LP, Javitt JC, Canner JC: National outcomes of penetrating keratoplasty. Risk of endophthalmitis and retinal detachment. Arch Ophthalmol 111:509–513, 1993 Allen A, Jaffe NS: Symposium: intraocular lenses. Cystoid macular edema—a preliminary study. Trans Am Acad Ophthalmol Otolaryngol 81:OP133–OP134, 1976 Alpar JJ, Fechner PU: Fechner’s IOLs. New York, ThiemeStratton, 1986 Anderson WB, Cobo M, Foulks G, Mitchell C: Atlas of Ophthalmic Surgery, Vol 1, Surgery of the Anterior Segment. St Louis, Mosby Yearbook, 1991, pp 78–95 Apple DJ, Carlson AN: Clinical consultation: Postoperative endophthalmitis. J Cataract Refract Surg 8:413–419, 1992 Apple DJ, Mamalis N, Loftfield K, et al: Complications of IOLs. A historical and histopathologic report. Surv Ophthalmol 29:1–54, 1984 Apple DJ, Mamalis N, Olson RJ, Kincaid MC: IOLs: Evolution, Designs, Complications, and Pathology. Baltimore, Williams & Wilkins, 1989 Apple DJ, Olson RJ: Closed-loop anterior chamber lenses (letter). Arch Ophthalmol 105:19–20, 1987 Apple DJ, Rabb MF: Ocular Pathology; Clinical Application & Self Assessment. St Louis, Mosby Yearbook, 1991, ed 4, pp 150–152 Apple DJ, Solomon KD, Tetz MR: Posterior capsular opacification. Surv Ophthalmol 37:73–116, 1992 Arentsen JJ, Laibson PR: Surgical management of pseudophakic corneal edema: complications and visual results following penetrating keratoplasty. Ophthalmic Surg 13:371–373, 1982
26. 27.
28. 29. 30.
31. 32.
33. 34. 35. 36.
Assia EI, Blumenthal M: Recurrent hyphema associated with IOL loop displacement treated with argon laser photocoagulation. Ophthalmic Surg 24:343–345, 1993 Auffarth GU, Wesendahl TA, Brown SJ, Apple DJ: Are there acceptable anterior chamber intraocular lenses for clinical use in the 1990s? An analysis of 4104 explanted anterior chamber intraocular lenses. Ophthalmology 101: 1913–1922, 1994 Bahn CF, Sugar A: Endothelial physiology and IOL implantation. J Am Intraocular Implant Soc 7:351–364, 1981 Ballin N, Weiss DM: Pigment dispersion and IOP elevation in pseudophakia. Ann Ophthalmol 14:627–630, 1982 Bartz-Schmidt KV, Konen W, Essen P, et al: Intraocular silicone lenses and silicone oil (German). Klin Monatsbl Angenheilkd 207:162–166, 1995 Behrendt S, Wetzel W: Complete occlusion of the capsulorhexis incision by anterior capsule shrinkage (German). Ophthalmologe 91:526–528, 1994 Bergman M, Laatikainen L: IOP level in glaucomatous and nonglaucomatous eyes after complicated cataract surgery and implantation of an AC-IOL. Ophthalmic Surg 23:378– 382, 1992 Beyer TL, Vogler G, Sharma D, et al: Protective barrier effect of the posterior lens capsule in exogenous bacterial endophthalmitis—An experimental primate study. Invest Ophthalmol Vis Sci 25:108–112, 1984 Beyer TL, Vogler G, Sharmer D, et al: Protective barrier effect of the posterior lens capsule in exogenous bacterial endophthalmitis: An experimental pseudophakic primate study. J Am Intraocul Implant Soc 9:293–296, 1983 Binder PS: Secondary IOL implantation during or after corneal transplantation. Am J Ophthalmol 99:515–520, 1985 Binkhorst CD: The iridocapsular (two-loop) lens and the iris clip (four-loop) lens in pseudophakia. Trans Am Acad Ophthalmol Otolaryngol 77:589–617, 1973 Brady SE, Rapuano CJ, Arentsen JJ, et al: Clinical indications for and procedures associated with penetrating keratoplasty, 1983–1988. Am J Ophthalmol 108: 118–122, 1989 Brown RH, Lynch MG, Tearse JE, Nunn RD: NeodymiumYAG vitreous surgery for phakic and pseudophakic malignant glaucoma. Arch Ophthalmol 104:1464–1466, 1986 Busin M, Arffa RC, McDonald MB, Kaufman HE: IOL removal during penetrating keratoplasty for pseudophakic bullous keratopathy. Ophthalmology 94:505–509, 1987 Busin M, Brauweiler P, Boker T, Spitznas M: Complication of sulcus-supported intraocular lenses with iris sutures, implanted during penetrating keratoplasty after intracapsular cataract extraction. Ophthalmology 97:401–406, 1990 Cameron JD, Apple DJ, Sumsion MA, et al: Pathology of iris support intraocular lenses. Implant 5:15–24, 1987 Caplan MB, Brown RH, Love LL: Pseudophakic pigmentary glaucoma. Am J Ophthalmol 105:320–322, 1988 Carlson AN, Apple DJ, Hrdlicka ZK, et al: Dislocation of silicone plate design IOLs following Nd:YAG laser capsulotomy (abstract and poster presentation). American Academy of Ophthalmology Meeting, Atlanta, 1995 Carlson AN, Buranosky M: Clinical consultation: Antimicrobial prophylactic therapy. Ophthalmic Practice 12:288– 292, 1994 Carlson AN, Buranosky MD, Apple DJ: Evaluation and management of postoperative localized microbial endophthalmitis. American Academy of Ophthalmology Video Presentation, Atlanta, 1995. Video available through the AAO. Carlson AN, Buranosky MD, Apple DJ: IOL decentration despite intact capsulorhexis (abstract and poster presentation). Ophthalmology 101(Suppl):122, 1994 Carlson AN, Koch DD: Endophthalmitis following Nd:YAG laser posterior capsulotomy. Ophthalmic Surg 19:168–170, 1988 Carlson AN, Stewart WC, King LP, Apple DJ: Staphylococcal aureus endophthalmitis following Nd:YAG laser capsulotomy. OMIG Publication of Abstracts, 1993 Carlson AN, Tetz MR, Apple DJ: Infectious complications
436
37.
38. 39. 40. 41. 42. 43. 44. 45. 46.
47. 48. 49. 50. 51. 52. 53. 54. 55. 56. 57. 58. 59. 60. 61. 62. 63.
Surv Ophthalmol 42 (5) March–April 1998 of modern cataract surgery and IOL implantation. Infect Dis Clin N Am 3:339–355, 1989 Chambless WS: Incidence of anterior and posterior segment complications in over 3,000 cases of extracapsular cataract extractions: intact and open capsules. J Am Intraocul Implant Soc 11:146–148, 1985 Chandler PA: Glaucoma in aphakia. Trans Am Acad Ophthalmol Otolaryngol 67:483–487, 1963 Chandler PA, Grant WM: Glaucoma. Philadelphia, Lea & Febiger, 1979, pp 276–282 Charlton KH, Binder PS, Perl T: Visual prognosis in pseudophakic corneal transplants. Ophthalmic Surg 12: 411–419, 1981 Choyce DP: The first 1,000 Mark IX implants in the practice of the Regional Eye Centre, Southend General Hospital, Essex, England: 1978–1983. CLAO J 10:218–221, 1984 Cinotti DJ, Maltzman BA, Renter DJ, Cinotti AA: Neodymium:YAG laser therapy for pseudophakic block. J Cataract Refract Surg 12:174–179, 1986 Claoue C, Falcon M, Shilling J: Penetrating keratoplasty in South East London 1981–1986: epidemiological aspect and demands on medical resources. J R Soc Med 83:264–269, 1988 Clayman HM: The Surgeon’s Guide to IOL Implantation. Thorofare, Slack, 1985, pp 67–83 Clayman HM, Parel JM, Miller D: Bacterial recovery from automated cataract surgical equipment. J Cataract Refract Surg 12:158–161, 1986 Clemente P: Sekundare Vorderkammerlinsenimplantationerfahrungen bei 305 implantationen uber 10 Jahre, in Neuhann TH, Hartmann CH, Rochels R (eds): Transactions of the 6th Congress of the German IOL Implant Soc (DHG). Berlin, Springer-Verlag, 1993, pp 199–204 Cohen EJ, Brady SE, Leavitt K, et al: Pseudophakic bullous keratopathy. Am J Ophthalmol 106:264–269, 1988 Coli AF, Price FW Jr, Whitson WE: IOL exchange for anterior chamber IOL-induced corneal endothelial damage. Ophthalmology 100:384–393, 1993 Connor TB Jr, Pulido JS: Management of aphakic and pseudophakic cystoid macular edema. Curr Opin Ophthalmol 111:90–101, 1993 Cusumano A, Busin M, Spitznas M: Mycotic infection of the capsular bag in postoperative endophthalmitis. J Cataract Refract Surg 17:503–505, 1991 Damji KF, Rootman J, White VA, et al: Changing indications for penetrating keratoplasty in Vancouver. Can J Ophthalmol 25:243–248, 1990 David R, Yagev R, Schneck M, et al: The fate of eyes with anterior chamber intraocular lenses. Eur J Ophthalmol 3:42– 46, 1993 Davidorf FH: Pseudophakic cystoid macular edema. Ophthalmic Forum 1:26–28, 1983 Davis RM, Best D, Gilbert GE: Comparison of IOL fixation techniques performed during penetrating keratoplasty. Am J Ophthalmol 111:743–749, 1991 Davison JA: Capsule contraction syndrome. J Cataract Refract Surg 19:582–589, 1993 Deutsch TA, Goldberg MF: Painless endophthalmitis after cataract surgery. Ophthalmic Surg 15:837–840, 1984 Dickens CJ, Shaffer RN: The medical treatment of ciliary block glaucoma after extracapsular cataract extraction. Am J Ophthalmol 103:237, 1987 Doren GS, Stern GA, Driebe WT: Indications for and results of IOL explantation. J Cataract Refract Surg 18:79–85, 1992 Drews RC: Posterior chamber lens implantation during keratoplasty without posterior lens capsule support. Cornea 6:38–40, 1987 Drews RC: Intermittent touch syndrome. Arch Ophthalmol 100:1440–1441, 1982 Driebe WT Jr, Mandelbaum S, Forster RK, et al: Pseudophakic endophthalmitis: diagnosis and management. Ophthalmology 93:442–448, 1986 Duffey RJ, Holland EJ, Agapitos PJ, Lindstom RL: Anatomic study of transsclerally sutured IOL implantation. Am J Ophthalmol 108:300–309, 1991 Duy TP, Wollensak J: Ciliary block (malignant) glaucoma
CARLSON ET AL following posterior chamber lens implantation. Ophthalmic Surg 18:741–744, 1987 64. Easty D, Dallas N, O’Malley R: Aphakic macular edema following prosthetic lens implantation. Br J Ophthalmol 61:321–326, 1977 65. Eaton AM, Jaffe GJ, McCuen BW, Mincey GJ: Condensation on the posterior surface of silicone intraocular lenses during fluid/air exchange. Ophthalmology 102:733–736, 1995 66. Ellingson FT: The uveitis-glaucoma-hyphema syndrome associated with the Mark VIII anterior chamber lens implant. J Am Intraocular Implant Soc 4:50, 1978 67. Emery JM, McIntyre DJ: Extracapsular Cataract Surgery. St Louis, CV Mosby, 1983 68. Epstein DL: Glaucoma in aphakia, in Chandler and Grant’s Glaucoma. Philadelphia, Lea & Febiger, 1986, ed 3, pp 320–331 69. Epstein DL: Cystoid macular edema occurring 13 years after cataract extraction. Am J Ophthalmol 83:501–503, 1977 70. Epstein DL, Steinert RF, Puliafito CA: Neodymium-YAG laser therapy to the anterior hyaloid in aphakic malignant (ciliovitreal block) glaucoma. Am J Ophthalmol 98:137–143, 1984 71. Evans RB: Peripheral anterior synechiae overlying the haptics of posterior chamber lenses. Occurrence and natural history. Ophthalmology 97:415–423, 1990 72. Fechner PV: Intrakularlinsen: Grundlagen and Operationglehve. Stuttgart, Ferdinand Enke Verlag, 1980 73. Ferguson FMG, Spalton DJ: Continued breakdown of the blood aqueous barrier following cataract surgery. Br J Ophthalmol 76:453–456, 1992 74. Fine BS, Brucker AJ: Macular edema and cystoid macular edema. Am J Ophthalmol 92:466–481, 1981 75. Fine M: Keratoplasty for bullous keratopathy with IOL. J Am Intraocul Implant Soc 4:12–13, 1978 76. Franiese JE, Christ FR, Buchen SY, et al: Moisture droplet formation on the posterior surface of intraocular lenses during fluid/air exchange. J Cataract Refract Surg 21:685–689, 1995 77. Galin MA, Asano Y, El Maghraby A, et al: Studies of residual alkali on intraocular lenses sterilized with NaOH. J Am Intraocul Implant Soc 9:290–292, 1983 78. Galin MA, Turkisk L: Studies of IOL sterilization: The effect of NaOH on B subtilis spores. J Am Intraocul Implant Soc 6:18–20, 1980 79. Gass JDM (ed): Stereoscopic atlas of macular diseases. St Louis, CV Mosby, 1987, ed 2, pp 368–445 80. Gass JDM: Lamellar macular hole: A complication of cystoid macular edema after cataract extraction. Arch Ophthalmol 94:793–800, 1976 81. Gass JDM, Anderson DR, Davis ED: A clinical, fluorescein angiographic, and electron microscopic correlation of cystoid macular edema. Am J Ophthalmol 100:82–86, 1985 82. Gass JDM, Norton EWD: Follow-up study of cystoid macular edema following cataract extraction. Trans Am Acad Ophthalmol Otolaryngol 73:665–682, 1969 83. Gass JDM, Norton EWD: Cystoid macular edema and papilledema following cataract extraction: a fluorescein funduscopic and angiographic study. Arch Ophthalmol 76:646–661, 1966 84. Geisse LJ: Nd:YAG laser treatment for recurrent hyphemas in pseudophakia. Am J Ophthalmol 111:513, 1991 85. Gerding DN, Poley BJ, Hall WH, et al: Treatment of pseudomonas endophthalmitis associated with prosthetic IOL implantation. Am J Ophthalmol 88:902–908, 1979 86. Gilbert CM, Novak MA: Successful treatment of postoperative Candida endophthalmitis in an eye with an IOL implant. Am J Ophthalmol 97:593–595, 1984 87. Gills JP: Prevention of endophthalmitis by intraocular solution filtration and antibiotics. J Am Intraocul Implant Soc 11:185–186, 1994 88. Greisse LJ: Nd:YAG laser treatment for recurrent hyphemas in pseudophakia. Am J Ophthalmol 111:513, 1991 89. Gritz DC, Cevallos AV, Smolin G, Whitcher JP Jr: Antibiotic supplementation of intraocular irrigating solutions. An in vitro model of antibacterial action. Ophthalmology 103: 1204–1208, 1996 90. Haamann P, Jensen OM, Schmidt P: Changing indications
437
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES for penetrating keratoplasty. Acta Ophthalmol (Copenh) 72:443–446, 1994 91. Hainsworth DP, Chen SN, Cox TA, Jaffe GJ: Condensation on polymethylene thacrylate, acrylic polymer and silicone intraocular lenses after fluid/air exchange in rabbits. Ophthalmology 103:1410–1418, 1996 92. Hall JR, Muenzler WS: IOL replacement in pseudophakic bullous keratopathy. Trans Ophthalmol Soc UK 104:541– 545, 1985 93. Hansen SO, Crandall AS, Olsen RJ: Progressive constriction of the anterior capsular opening following intact capsulorhexis. J Cataract Refract Surg 19:77–82, 1993 94. Harbour JW, Smiddy WE, Rubsamen PE, et al: Pars plana vitrectomy for chronic pseudophakic cystoid macular edema. Am J Ophthalmol 120:302–307, 1995 95. Hardten DR, Lindstrom RL: Corneal complications of cataract extraction and IOL implantation. Curr Opin Ophthalmol 4:99–105, 1993 96. Harris WS: Cystoid macular edema in extracapsular and intracapsular aphakia and pseudophakia, in Kwito ML, Praeger DL (eds): Pseudophakia: Current Trends and Concepts. Baltimore, Williams & Wilkins, 1980, pp 367–373 97. Harris WS, Kogan I: Retrospective study of retinal complications with posterior chamber lenses. Contact Intraocular Lens Med J 7:345–347, 1981 98. Harris WS, Taylor BC, Winslow RL: Cystoid macular edema following IOL implantation. Ophthalmic Surg 8:134–137, 1977 99. Hassan TS, Soong JK, Sugar A, Meyer RF: Implantation of Kelman-style, open-loop anterior chamber lenses during keratopathy for aphakic and pseudophakic bullous keratopathy. Ophthalmology 98:875–880, 1991 100. Henkind DP: The first international cystoid macular edema symposium. Surv Ophthalmol 28 (Suppl):431–620, 1984 101. Henry MM, Henry LM, Henry LM: A possible cause of chronic cystic maculopathy. Ann Ophthalmol 9:455–457, 1977 102. Herschler J: Laser shrinkage of the ciliary processes. A treatment for malignant (ciliary block) glaucoma. Ophthalmology 87:1155–1159, 1980 103. Herschler J: Glaucoma and the IOL. Ann Ophthalmol 11:1057–1058, 1979 104. Hitchings RA: Aphakic macular oedema: a two-year followup study. Br J Ophthalmol 61:628–630, 1977 105. Hitchings RA, Chisolm IH, Bird AL: Aphakic macular edema: Incidence and pathogenesis. Invest Ophthalmol Vis Sci 14:68–72, 1975 106. Hoffer KJ: Anterior chamber lens exchange: Comment on. J Cataract Refract Surg 18:321–326, 1992 107. Hoffer KJ: Anterior chamber lens exchange: Comment on. J Cataract Refract Surg 19:536–537, 1992 108. Hoffer KJ: Rapid fibrosis of ORC Stableflex haptics (letter). J Am Intraocul Implant Soc 10:470–472, 1984 109. Hoffer KJ: Effects of extracapsular implant techniques on endothelial density. Arch Ophthalmol 100:791–792, 1982 110. Hoffer KJ: Vertical endothelial cell disparity. Am J Ophthalmol 87:344–349, 1979 111. Hoffer KJ: Corneal decompensation after corneal endothelium cell count (letter to the editor). Am J Ophthalmol 87:252–253, 1979 112. Hoffer KJ, Phillippi G: A cell membrane theory of endothelial repair and vertical cell loss after cataract surgery. J Am Intraocul Implant Soc 4:18–25, 1978 113. Holladay JT, Prager TC, Long SA, et al: Determining IOL power within the eye. J Am Intraocul Implant Soc 11:353– 365, 1985 114. Holland EJ, Daya SM, Evangelista A, et al: Penetrating keratoplasty and transscleral fixation of posterior chamber lens. Am J Ophthalmol 114:182–187, 1992 115. Hopen G, Mondino BJ, Kozy D, et al: Intraocular lenses and experimental bacterial endophthalmitis. Am J Ophthalmol 94:402–407, 1982 116. Horris RJ, Bates AK: Changing indications for keratoplasty. Eye 3:455–459, 1989 117. Hyman L, Wittpenn J, Yang C: Indications and techniques of penetrating keratoplasties, 1985–1988. Cornea 11:573– 576, 1992
118.
119. 120. 121. 122. 123. 124.
125. 126. 127.
128. 129. 130. 131. 132. 133.
134. 135.
136. 137. 138.
139. 140. 141.
Insler MS, Kook MS, Kaufman HE: Penetrating keratoplasty in pseudophakic bullous keratopathy associated with semiflexible, closed-loop anterior chamber intraocular lenses. Am J Ophthalmol 107:252–256, 1989 Insler MS, Zatzkis SM: Pigment dispersion syndrome in pseudophakic corneal transplants. Am J Ophthalmol 102:762–765, 1986 Isenberg RA, Weiss RL, Apple DJ, et al: Fungal contamination of balanced salt solution. J Am Intraocul Implant Soc 11:485–486, 1985 Jacobs PM, Cheng J, Price NC: Pseudophakodonesis and corneal endothelial contact: direct observations by highspeed cinematography. Br J Ophthalmol 67:650–654, 1983 Jaffe NS: Extracapsular cataract extraction with a posterior chamber IOL-technique and results. Trans Ophthalmol Soc N Z 34:28–30, 1982 Jaffe NS: Cataract Surgery and Its Complications. St Louis, CV Mosby, ed 5, 1990 Jaffe NS, Clayman HM, Jaffe MS: A comparison of ICCEBinkhorst IOL and ECCE-posterior chamber IOL, thirtyfour to forty months postoperatively. J Am Intraocul Implant Soc 8:128–130, 1982 Jaffe NS, Clayman HM, Jaffe MS: Cystoid macular edema after intracapsular and extracapsular cataract extraction with and without an IOL. Ophthalmology 89:25–29, 1982 Jaffe NS, Jaffe MS, Clayman HM, Light DS: The results of lens implantation in eyes with operative loss of vitreous. J Am Intraocul Implant Soc 6:243–245, 1980 Javitt JC, Street DA, Tielsch JM, et al: National outcomes of cataract extraction. Retinal detachment and endophthalmitis after outpatient cataract surgery. Cataract Patient Outcomes Research Team. Ophthalmology 101:100–105, 1994 Javitt JC, Vitale S, Canner JK, et al: National outcomes of cataract extraction. Endophthalmitis following inpatient surgery. Arch Ophthalmol 109:1085–1089, 1991 Johnson SH, Kratz RP, Olson PF: Iris transillumination defect and microhyphema syndrome. J Am Intraocular Implant Soc 10:425, 1984 Jones DB: Antibiotic supplementation of intraocular irrigating solutions. An in vitro model of antibacterial action (discussion). Ophthalmology 103:1208–1209, 1996 Kattan HM, Flynn HW Jr, Pflugfelder SC, et al: Nosocomial endophthalmitis survey. Current incidence of infection after intraocular surgery. Ophthalmology 98:227–238, 1991 Kervick GN, Shepherd WFI: Changing indications for penetrating keratoplasty. Ophthalmic Surg 21:227, 1990 Koenig SB, McDermott MI, Hyndiuk RA: Penetrating keratoplasty and IOL exchange for pseudophakic bullous keratopathy associated with closed-loop anterior chamber intraocular lenses. Am J Ophthalmol 108:43–48, 1989 Koerner FH: Anterior pars plana vitrectomy in ciliary and iris block glaucoma. Graefes Arch Clin Exp Ophthalmol 214:119–125, 1980 Kornmehl EW, Steinert RF, Odrich MG, Stevens JB: Penetrating keratoplasty for pseudophakic bullous keratopathy associated with closed-loop anterior chamber intraocular lenses. Ophthalmology 97:407–412, 1990 Kottow M, Hendrickson P: Iris angiography in cystoid macular edema after cataract extraction. Arch Ophthalmol 93: 487–493, 1975 Kozarsky AM, Stopak S, Waring GO III, et al: Results of penetrating keratoplasty for pseudophakic corneal edema with retention of IOL. Ophthalmology 91:1141–1146, 1984 Kraff MC, Sanders DR, Jampol LM, Lieberman HL: Effect of primary capsulotomy with extracapsular surgery on the incidence of pseudophakic cystoid macular edema. Am J Ophthalmol 98:166–170, 1984 Kraff MC, Sanders DR, Jampol LM, et al: Prophylaxis of pseudophakic cystoid macular edema with topical indomethacin. Ophthalmology 89:885–890, 1982 Kraff MC, Sanders DR, Lieberman HL: Monitoring for continuing endothelial cell loss with cataract extraction and IOL implantation. Ophthalmology 89:30–34, 1982 Kraff MC, Sanders DR, Lieberman HL: Specular micros-
438
142. 143. 144. 145. 146.
147. 148. 149. 150. 151. 152. 153. 154. 155. 156. 157. 158.
159. 160. 161. 162.
163. 164.
165. 166.
Surv Ophthalmol 42 (5) March–April 1998 copy in cataract and IOL patients: A report of 564 cases. Arch Ophthalmol 98:1782–1784, 1980 Kraff MC, Sanders DR, Lieberman HL: Endothelial cell loss and trauma during IOL implantation: a specular microscopic study. J Am Intraocul Implant Soc 4:107–109, 1978 Kraff MC, Sanders DR, Raanan MG: A survey of IOL explantations. J Cataract Refract Surg 12:644–650, 1986 Kusaka S, Kodana T, Ohashi Y: Condensation of silicone oil on the posterior surface of a silicone IOL during vitrectomy. Am J Ophthalmol 121:574, 1996 Kwartz J, Leatherbarrow B, Dyer P, et al: Penetrating keratoplasty for pseudophakic corneal edema. Br J Ophthalmol 79:435–438, 1995 Lass JH, DeSantis DM, Reinhart WJ, et al: Clinical and morphometric results of penetrating keratoplasty with onepiece anterior-chamber or suture-fixated posterior-chamber lenses in the absence of lens capsule. Arch Ophthalmol 108:1427–1431, 1990 Lee DA, Price FW Jr, Whitson WE: Intraocular complications associated with the Dubroff anterior chamber lens. J Cataract Refract Surg 20:421–425, 1994 Legler UFC, Apple DJ, Hund P, Kirkconnel WS: Chronic ciliary pain secondary to posterior chamber IOL loop incarceration. Am J Ophthalmol 111:513–514, 1991 Leyden WE: Glaucoma and IOL implantation, in Ritch R, Shields MB (eds): The Secondary Glaucomas. St Louis, CV Mosby, 1982, pp 367–380 Lieppman ME: Intermittent visual “white out.” A new IOL complication. Ophthalmology 89:109–112, 1982 Lindquist TD, McGlothan JS, Rotkis WM, Chandler JW: Indications for penetrating keratoplasty: 1980–1988. Cornea 10:210–216, 1991 Liu JF, Koch DD, Emery JM: Complications of implanting three-piece C-loop posterior chamber lenses in the anterior chamber. Ophthalmic Surg 19:802–807, 1988 Lyle WA, Jin JC: An analysis of IOL exchange. Ophthalmic Surg 23:453–458, 1992 Lynch MG, Brown RH, Michels RG, et al: Surgical vitrectomy for pseudophakic malignant glaucoma. Am J Ophthalmol 102:149–153, 1986 MacRae S: Positioning hole synechias with anterior chamber lenses. J Cataract Refract Surg 17:521, 1991 Magargal LE, Goldberg RE, Uram M, et al: Recurrent microhyphema in the pseudophakic eye. Ophthalmology 90:1231–1234, 1983 Mamalis N, Craig MT, Coulter VL, et al: Penetrating keratoplasty 1981–88: clinical indications and pathologic findings. J Cataract Refract Surg 14:163–167, 1990 Mamalis N, Crandall AS, Pulsipher MW, et al: IOL explantation and exchange. A review of lens style, clinical indications, clinical results and visual outcome. J Cataract Refract Surg 17:811–818, 1991 Masket S: Postoperative complications of capsulorhexis. J Cataract Refract Surg 19:721–724, 1993 Masket S: Consultation section. J Cataract Refract Surg 18:321–326, 1992 Masket S: Pseudophakic posterior iris chafing syndrome. J Cataract Refract Surg 12:252–256, 1986 Maynor RC Jr: Lens-induced complications with anterior chamber lens implants: a comparison with iris supported and posterior chamber lenses. J Am Intraocul Implant Soc 9:450–452, 1983 McDonnell PJ, Green WR, Maumenee AE, Iliff WJ: Pathology of intraocular lenses in 33 eyes examined postmortem. Ophthalmology 90:386–403, 1983 Meisler DM, Palestine AG, Vastine DW, et al: Chronic propionibacterium endophthalmitis after extracapsular cataract extraction and IOL implantation. Am J Ophthalmol 102:733–739, 1986 Meisler DM, Zakov ZN, Bruner WE, et al: Endophthalmitis associated with sequestered intraocular Propionibacterium acnes. Am J Ophthalmol 104:428–429, 1987 Menikoff JA, Speaker MG, Marmor M, et al: A case controlled study of risk factors for postoperative endophthalmitis. Ophthalmology 98:1761–1775, 1991
CARLSON ET AL 167. 168. 169. 170. 171. 172. 173. 174. 175. 176. 177. 178. 179. 180. 181. 182. 183. 184. 185. 186. 187. 188. 189. 190. 191. 192. 193.
194.
Meyer RF, Sugar A: Penetrating keratoplasty in pseudophakic bullous keratopathy. Am J Ophthalmol 90:677–681, 1980 Miami Study Group: Cystoid macular edema in aphakic and pseudophakic eyes. Am J Ophthalmol 88:45–48, 1979 Miller D, Doane MG: High-speed photographic evaluation of IOL movements. Am J Ophthalmol 97:752–759, 1984 Mills KB: Complications of IOL implantation, in Haining WM, Arnott EJ (eds): IOL Implantation. St Louis, CV Mosby, 1984, pp 550–562 Miyake K: Fluorophotometric evaluation of the blood-ocular barrier function following cataract surgery and IOL implantation. J Cataract Refract Surg 14:560–568, 1988 Miyake K, Asakura M, Kobayashi H: Effect of IOL fixation on the blood-aqueous barrier. Am J Ophthalmol 98:451– 455, 1984 Moa LK, Holland PM: “Very late onset” cystoid macular edema. Ophthalmic Surg 19:633–635, 1988 Mohamadi P, McDonnell JM, Irvine JA, et al: Changing indications for penetrating keratoplasty, 1984–1988. Am J Ophthalmol 107:550–552, 1989 Momoeda S, Hayashi H, Oshima K: Anterior pars plana vitrectomy for phakic malignant glaucoma. Jpn J Ophthalmol 27:73–79, 1983 Morris RJ, Bates AK: Changing indications for keratoplasty. Eye 3:455–459, 1989 Morrison LK, Waltman SR: Management of pseudophakic bullous keratoplasty. Ophthalmic Surg 20:205–210, 1989 Moses L: Complications of rigid anterior chamber implants. Ophthalmology 91:819–825, 1984 Mosier MA, Lusk B, Pettit TH, et al: Fungal endophthalmitis following IOL implantation. Am J Ophthalmol 83:1–81, 1977 Murphy GE: Long-term gonioscopy follow-up of eyes with posterior chamber lens implants and no iridectomy. Ophthalmic Surg 17:227–228, 1986 Neuteboom GHG, DeVreis-Knoppert WAEJ: Endophthalmitis after Nd:YAG laser capsulotomy. Doc Ophthalmol 70:175–178, 1988 Nicholson DH: Occult iris erosion: a treatable cause of recurrent hyphema in iris-supported intraocular lenses. Ophthalmology 89:113–119, 1982 Nikica G, Ljerka HP, Jelena P, et al: Cystoid macular edema in anterior chamber lens implantation following posterior capsule rupture. Doc Ophthalmol 81:309–315, 1992 Nussenblatt RB, Kaufman SC, Paletine AG, et al: Macular thickening and visual acuity: Measurement in patients with cystoid macular edema. Ophthalmology 94:1134–1139, 1987 O’Day DM: Fungal endophthalmitis caused by Paecilomyces lilacinus after IOL implantation. Am J Ophthalmol 83:130– 131, 1977 Obstbaum SA, Galin MA: Cystoid macular oedema and ocular inflammation; the corneo-retinal inflammatory syndrome. Trans Ophthalmol Soc UK 99:187–191, 1979 Odufuwa TO, Bolger J: Late hyphema after small incision cataract surgery. J Cataract Refract Surg 20:342–343, 1994 Ohrloff C, Oldend OJ, Puck A: Minimal endothelial cell loss following phacoemulsification and implantation of a posterior chamber lens. Ophthalmology 90:627–636, 1983 Olson RJ, Slappey TE: Corneal endothelial damage induced by intraocular lenses, an in vitro study. J Am Intraocul Implant Soc 5:321–323, 1979 Ormerod LD, Paton BG, Haaf J, et al: Anaerobic bacterial endophthalmitis. Ophthalmology 94:799–808, 1987 Pande M, Noble BA: The role of IOL exchange in the management of major implant-related complications. Eye 7:34– 39, 1993 Parrish R, Minckler D: Editorial: “Late endophthalmitis”— Filtering surgery time bomb? Ophthalmology 103:1167– 1168, 1996 Pavlin CJ, Harasiewicz K, Foster FS: Ultrasound biomicroscopic analysis of haptic position in late-onset, recurrent hyphema after posterior chamber lens implantation. J Cataract Refract Surg 20:182–185, 1994 Pazandak B, Johnson S, Kratz R, Faulkner GD: Recurrent intraocular hemorrhage associated with posterior chamber
439
REMOVAL OR EXCHANGE OF INTRAOCULAR LENSES lens implantation. L Am Intraocular Implant Soc 9:327– 329, 1983 195. Pearce JL: Long-term results of the Choyce anterior chamber lens implant Mark V, VII, VIII. Br J Ophthalmol 59:99– 106, 1975 196. Petit TH, Olson RJ, Foos RY, et al: Fungal endophthalmitis following IOL implantation: a surgical epidemic. Arch Ophthalmol 98:1025–1039, 1980 197. Piest KL, Kincaid MC, Tetz MR, et al: Localized endophthalmitis: A newly documented cause of toxic lens syndrome. J Cataract Refract Surg 13:498–510, 1987 198. Pignalosa B, Toni F, Ligurori G: Pigmentary dispersion syndrome subsequent IOL implantation in P.C. documenta. Ophthalmologica 73:231–243, 1990 199. Polack FM: Results of keratoplasty for aphakic and pseudophakic corneal edema with IOL implantation or exchange. Cornea 7:239–243, 1988 200. Poleski SA, Willis WE: Angle-supported intraocular lenses: A goniophotographic study. Ophthalmology 91:838–840, 1984 201. Price FW Jr: Factors contributing to corneal decompensation with the Stableflex lens. J Cataract Refract Surg 14:53– 57, 1988 202. Price FW Jr, Whitson WE: Natural history of cystoid macular edema in pseudophakic bullous keratopathy. J Cataract Refract Surg 16:163–169, 1990 203. Price FW Jr, Whitson WE: Visual results of suture-fixated posterior chamber lenses during penetrating keratoplasty. Ophthalmology 96:1234–1240, 1989 204. Price FW Jr, Whitson WE, Collins K, Johns S: Explantations of posterior chamber lenses. J Cataract Refract Surg 18:475–479, 1992 205. Price FW Jr, Whitson WE, Marks RG: Graft survival in four common groups of patients undergoing penetrating keratoplasty. Ophthalmology 98:322–328, 1991 206. Ptacek JT, Eberhardt TL: Breaking bad news—A review of the literature. JAMA 276:496–502, 1996 207. Rao GN, Stevens RE, Harris JK, Aquavella JV: Long-term changes in corneal endothelium following IOL implantation. Ophthalmology 88:386–397, 1981 208. Rapuano CJ, Cohen EJ, Brady SE, et al: Indications for and outcomes of repeat penetrating keratoplasty. Am J Ophthalmol 109:689–695, 1990 209. Raskin EM, Speaker MG, McCormick SA, et al: Influence of haptic material on the adherence of staphylococci to intraocular lenses. Arch Ophthalmol 111:250–253, 1993 210. Reed JE, Thomas JV, Lytle RA, Simmons RJ: Malignant glaucoma induced by an IOL. Ophthalmic Surg 21:177– 180, 1990 211. Reidy JJ, Apple DJ, Googe JM, et al: An analysis of semiflexible, closed-loop anterior chamber lenses. J Am Intraocul Implant Soc 11:344–352, 1985 212. Richards BW, Richards WI, Olson RJ, Richards SC: The effects of nonfixated lower lens haptics of Binkhorst lenses on corneal endothelial cell density. Ophthalmic Surg 17:286–289, 1986 213. Richburg FA: Hypopyon and ultrasonic cleaning solution (letter to the editor). J Am Intraocul Implant Soc 11:170, 1985 214. Richburg FA, Reidy JJ, Apple DJ, et al: Sterile hypopyon secondary to ultrasonic cleaning solution. J Cataract Refract Surg 12:248–251, 1986 215. Ridley H: Further observations on intraocular acrylic lenses in cataract surgery. Trans Am Acad Ophthalmol Otolaryngol 57:98–106, 1953 216. Ritch R, Shields MB, Krupin T (eds): The Glaucomas. St Louis, CV Mosby, 1989, p 1206 217. Robin JB, Gindi JJ, Koh K, et al: An update of the indications for penetrating keratoplasty: 1979 through 1983. Arch Ophthalmol 104:87–89, 1986 218. Roper-Hall MJ, Wilson RS: Reduction in endothelial cell density following cataract extraction and IOL implantation. Br J Ophthalmol 66:516–517, 1982 219. Roussel TJ, Culbertson WW, Jaffe NS: Chronic postoperative endophthalmitis associated with Propionibacterium acnes. Arch Ophthalmol 105:1199–1201, 1987
220. 221. 222. 223. 224. 225. 226. 227. 228. 229. 230. 231. 232. 233. 234. 235.
236. 237. 238. 239. 240. 241.
242. 243. 244. 245.
Rowsey JJ, Gaylor JR: IOL disasters. Peripheral anterior synechia. Ophthalmology 87:646–664, 1980 Ruiz R, Saatci OA: Visual outcome in pseudophakic eyes with clinical cystoid macular edema. Ophthalmic Surg 22:190–193, 1991 Samples JR, Bellows AR, Rosenquist RC, et al: Pupillary block with posterior chamber intraocular lenses. Arch Ophthalmol 105:335–337, 1987 Samples JR, Van Buskirk EM: Pigmentary glaucoma associated with posterior chamber intraocular lenses. Am J Ophthalmol 100:385–388, 1985 Savage JA, Thomas JJV, Belcher CD, et al: Extracapsular cataract extraction and posterior chamber IOL implantation in glaucomatous eyes. Ophthalmology 92:1506–1516, 1985 Schanzlin DJ, Goldberg DB, Brown SI: Staphylococcus epidermidis endophthalmitis following IOL implantation. Br J Ophthalmol 64:684–686, 1980 Schanzlin DJ, Robin JB, Gomez DS, et al: Results of penetrating keratoplasty for aphakic and pseudophakic bullous keratopathy. Am J Ophthalmol 98:302–312, 1984 Schein OD, Kenyon KR, Steinert RF, et al: A randomized trial of IOL fixation techniques with penetrating keratoplasty. Ophthalmology 100:1431–1443, 1993 Schulze RR, Copeland JR: Posterior chamber IOL implantation without peripheral iridectomy. A preliminary report. Ophthalmic Surg 13:567, 1982 Shepherd JR: Continuous-tear capsulotomy and insertion of a silicone bag lens. J Cataract Refract Surg 15:335–339, 1989 Shields MB: Glaucoma following ocular surgery, in Shields MB (ed): Textbook of Glaucoma. Baltimore, Williams and Wilkins, 1987, pp 334–358 Shrader CE, Belcher CD, Thomas JV, et al: Pupillary and iridovitreal block in pseudophakic eyes. Ophthalmology 91:831–837, 1984 Simel PF: Posterior chamber implants without iridectomy. J Am Intraocular Implant Soc 8:141–143, 1982 Singh G: Endophthalmitis after cataract extraction: A retrospective case study. Ann Ophthalmol 13:629–631, 1981 Sinskey RM, Amin P, Stoppel JO: Indications for results of a large series of IOL exchanges. J Cataract Refract Surg 19:68–71, 1993 Slusher MM, Seaton AD: Loss of visibility caused by moisture condensation on the posterior surface of a silicone IOL during fluid/gas exchange after posterior vitrectomy (letter). Am J Ophthalmol 118:667, 1994 Smith PW, Wong SK, Stark WJ, et al: Complications of semiflexible, closed-loop anterior chamber intraocular lenses. Arch Ophthalmol 105:52–57, 1987 Smith RE: Inflammation after cataract surgery (editorial). Am J Ophthalmol 102:788–799, 1986 Smith RE, McDonald HR, Nesburn AB, Minckler DS: Penetrating keratoplasty: Changing indications, 1947 to 1978. Arch Ophthalmol 98:1226–1229, 1980 Smith SG: IOL removal for chronic cystoid macular edema. J Cataract Refract Surg 15:442–445, 1989 Snyder HP: Intravenous chloramphenicol to prevent postoperative endophthalmitis in cataract surgery—520 consecutive cases. Ann Ophthalmol 10:1041–1042, 1978 Soong HK, Musch DC, Kowal V, et al: Implantation of posterior chamber intraocular lenses in the absence of lens capsule during penetrating keratoplasty. Arch Ophthalmol 107:660–665, 1989 Spaide RF, Yannuzzi LA, Sisco LJ: Chronic cystoid macular edema and predictors of visual acuity. Ophthalmic Surg 24:262–267, 1993 Speaker MG, Lugo M, Laibson PR, et al: Penetrating keratoplasty for pseudophakic bullous keratoplasty: Management of the IOL. Ophthalmology 95:1260–1268, 1988 Speaker MG, Milch FA, Shah MK, et al: Role of external bacterial flora in the pathogenesis of acute postoperative endophthalmitis. Ophthalmology 98:639–650, 1991 Stark WJ Jr, Maumenee AE, Fagadau W, et al: Cystoid macular edema in pseudophakia. Surv Ophthalmol 28 (Suppl): 442–451, 1984
440 246. 247. 248. 249. 250.
251. 252. 253. 254.
255. 256. 257. 258. 259. 260. 261.
262.
263. 264. 265. 266. 267. 268.
Surv Ophthalmol 42 (5) March–April 1998 Stark WJ, Gottsch JD, Goodman DF, et al: Posterior chamber IOL implantation in the absence of capsular support. Arch Ophthalmol 107:1078–1083, 1989 Stark WJ, Worthen DM, Holladay JT, et al: The FDA report on intraocular lenses. Ophthalmology 90:311–317, 1983 Starr MB, Lally JM: Antimicrobial prophylaxis for ophthalmic surgery. Surv Ophthalmol 39:485–501, 1995 Stefansson E, Cobo LM, Carlson AN, Kesler JL: Endocapsular hypopyon: A clinical sign of localized endophthalmitis. Ophthalmic Surg 21:221–222, 1990 Stern WH, Tamura E, Jacobs RA, et al: Epidemic postsurgical Candida parapsilosis endophthalmitis: Clinical findings and management of 15 consecutive cases. Ophthalmology 88:942–946, 1981 Stewart WC: Clinical Practice of Glaucoma. Thorofare, Slack, 1990, pp 149–196 Stocker FW: The Endothelium of the Cornea and Its Clinical Implications. Springfield, Charles C Thomas Stolba V, Binder S, Velikay M, Wedrich A: Intraocular silicone lenses in silicone oil: An experimental study. Graefes Arch Clin Exp Ophthalmol 234:55–57, 1996 Stonecipher KG, Parmley VC, Jensen H, Rowsey JJ: Infectious endophthalmitis following sutureless cataract surgery. Arch Ophthalmol 109:1562–1563, 1991; see comment Arch Ophthalmol 110:913, 1992 Sugar A: An analysis of corneal endothelial and graft survival pseudophakic bullous keratopathy. Trans Am Ophthalmol Soc 87:762–801, 1989 Sugar J, Mitchelson J, Kraff M: Endothelial trauma and cell loss from IOL insertion. Arch Ophthalmol 96:449–450, 1978 Swan KC: Hyphema due to wound vascularization after cataract extraction. Arch Ophthalmol 89:87–90, 1973 Taylor DM, Atlas BF, Romanchuk KG, Stern AL: pseudophakic bullous keratopathy. Ophthalmology 90:19–24, 1983 Taylor DM, Sachs SW, Stern AL: Aphakic cystoid macular edema. Longterm clinical observations. Surv Ophthalmol 28 (Suppl): 437–441, 1984 Tetz MR, Apple DJ, Hanson SO, et al: “Localized endophthalmitis”: A complication of extracapsular cataract extraction. Implant Ophthalmol 1:93–97, 1987 Tetz MR, Apple DJ, Price FW, et al: A newly described complication of neodymium-YAG laser capsulotomy: Exacerbation of an intraocular infection. Arch Ophthalmol 105: 1324–1325, 1987 Tomey KF, Senft SH, Antonios SR: Aqueous misdirection and flat chamber after posterior chamber implants with and without trabeculectomy. Arch Ophthalmol 105:770– 773, 1987 Tomey KF, Traverso CE: The glaucomas in aphakia and pseudophakia. Surv Ophthalmol 36:79–112, 1991 Vafidis GC, Marsh RJ, Stacey AR: Bacterial contamination of IOL surgery. Br J Ophthalmol 68:520–523, 1984 Vajpayee RB, Angra SK, Titiyal JS, et al: Pseudophakic pupillary-block glaucoma in children. Am J Ophthalmol 111:715–718, 1991 Van Buskirk EM: Late onset, progressive, peripheral anterior synechiae with posterior chamber intraocular lenses. Ophthalmic Surg 18:115–117, 1987 Van Buskirk EM: Pseudophakic glaucoma, in Weinstein GE (ed): Contemporary Issues in Ophthalmology. New York, Churchill Livingstone, 1986, pp 133–154 Van Buskirk EM: Pupillary block after IOL implantation. Am J Ophthalmol 95:55–59, 1983
CARLSON ET AL 269.
Van Liefferinge T, Van Oye R, Kestelyn P: Uveitis-glaucoma-hyphema syndrome: a late complication of posterior chamber lenses. Bull Soc Belge Ophthalmol 252:61–66, 1994 270. Waltman SR: Penetrating keratoplasty for pseudophakic bullous keratopathy. Arch Ophthalmol 99:415–416, 1981 271. Waring GO III: The 50-year epidemic of pseudophakic corneal edema (editorial). Arch Ophthalmol 107:657–659, 1989 272. Waring GO III, Kenyon KR, Gemmill MC: Results of anterior segment reconstruction for aphakic and pseudophakic corneal edema. Ophthalmology 95:836–841, 1988 273. Waring GO III, Stulting RD, Street D: Penetrating keratoplasty for pseudophakic corneal edema with exchange of intraocular lenses. Arch Ophthalmol 105:58–62, 1987 274. Waring GO III, Welch SN, Cavanagh HD, Wilson LA: Results of penetrating keratoplasty in 123 eyes with pseudophakic or aphakic corneal edema. Ophthalmology 90:25– 33, 1983 275. Watzke RC: Intraocular hemorrhage from vascularization of the cataract incision. Ophthalmology 87:19–23, 1980 276. Weber PA, Henry MA, Kapetansky FM, Lohman LF: Argon laser treatment of the ciliary processes in aphakic glaucoma with flat anterior chamber. Am J Ophthalmol 97:82– 85, 1984 277. Weinreb RN, Wasserstrom JP, Forman JS, et al: Pseudophakic pupillary block and angle-closure glaucoma in diabetic patients. Am J Ophthalmol 102:325–328, 1986 278. Williams DL, Gills JP: Infectious endophthalmitis following sutureless cataract surgery (letter, discussion). Arch Ophthalmol 110:913–915, 1992 279. Willis DA, Stewart RH, Kimbrough RL: Pupillary block associated with posterior chamber lenses. Ophthalmic Surg 16:108–109, 1985 280. Winslow RL, Taylor BC, Harris WS: A one-year follow-up of cystoid macular edema following IOL implantation. Ophthalmology 85:190–196, 1978 281. Wong SK, Stark WJ, Gottsch JD, et al: Use of posterior chamber lenses in pseudophakic bullous keratopathy. Arch Ophthalmol 105:856–858, 1987 282. Woodhams JT, Lester JC: Pigmentary dispersion glaucoma secondary to posterior chamber intraocular lenses. Ann Ophthalmol 16:825–855, 1984 283. Worthen DM, Boucher JA, Buxton JN, et al: Interim FDA report on intraocular lenses. Ophthalmology 87:267–271, 1980 284. Wright PL, Wilkinson CP, Balyeat HD, Reinke M: Angiographic cystoid macular edema after posterior chamber lens implantation. Arch Ophthalmol 106:740–744, 1988 285. Zaidman GW, Goldman S: A prospective study on the implantation of anterior chamber intraocular lenses during keratoplasty for pseudophakic and aphakic bullous keratopathy. Ophthalmology 97:757–762, 1990
Supported in part by an unrestricted grant from Research to Prevent Blindness, Inc., New York, New York. The authors acknowledge Dr. David Apple for helping to initiate this project, Dr. Todd D. Gwin for his contribution in the CME section, and Teresa Thompson, COT, and Tina Mitchell, COA, for their expertise in ophthalmic photography. Reprint address: Alan N. Carlson, MD, Director, Corneal and Refractive Surgery Service, Box 3802, Duke University Eye Center, Durham, NC 27710, USA.