Accepted Manuscript Laser-induced breakdown spectroscopy for the quantitative analysis of metals in sediments using natural zeolite matrix
E.S. Austria, E.M. Fuentes, G.M. Nuesca, R.B. Lamorena PII: DOI: Reference:
S0584-8547(16)30155-0 doi: 10.1016/j.sab.2017.07.001 SAB 5274
To appear in:
Spectrochimica Acta Part B: Atomic Spectroscopy
Received date: Revised date: Accepted date:
15 August 2016 25 April 2017 5 July 2017
Please cite this article as: E.S. Austria, E.M. Fuentes, G.M. Nuesca, R.B. Lamorena , Laser-induced breakdown spectroscopy for the quantitative analysis of metals in sediments using natural zeolite matrix, Spectrochimica Acta Part B: Atomic Spectroscopy (2017), doi: 10.1016/j.sab.2017.07.001
This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
ACCEPTED MANUSCRIPT Laser-induced breakdown spectroscopy for the quantitative analysis of metals in sediments using natural zeolite matrix E. S. Austria Jr.1,2 , E. M. Fuentes1 , G. M. Nuesca1 and R. B. Lamorena1,*
T
Institute of Chemistry, College of Science, University of the Philippines, Diliman, Quezon City,
IP
1
2
CR
Philippines
Materials Science and Engineering Program, College of Science, University of the Philippines,
US
Diliman, Quezon City, Philippines
AC
CE
PT
ED
M
AN
*corresponding author
ACCEPTED MANUSCRIPT Abstract The dependence of laser-induced breakdown spectroscopy (LIBS) to the matrix of the sample remains an important consideration in performing quantitative analysis. In this study, a new matrix was introduced in the preparation of solid powder calibration curves. Heat-treated natural zeolite
IP
T
and KBr were mixed separately into high purity metal powders to generate calibration curves using
CR
a univariate approach. A LIBS technique was used in the detection and quantitative analysis of Cr, Cu and Pb in river sediment samples. The relative percent difference (RPD) was calculated to
US
describe the variability measurements made using ICP/OES and LIBS as well as to evaluate the accuracy of the method. Calculated limits of detection in the matrices prepared were comparable
AN
with literature values and ranged from 0.41 to 6.1 ppm. The resulting metal concentrations indicate
M
that the natural zeolite matrix were closer to the reference values compared to the KBr matrix. By employing principal component analysis (PCA), heat treated zeolite was shown as a potential
ED
diluent or binder for generating calibration curves and could provide matrix-matched standards in
PT
identification of target metals from unknown sediment samples. The natural zeolite appeared to closely simulate the ablation behavior and property of the samples, and it is found to be a potential
AC
CE
suitable matrix for the quantitative LIBS analysis of sediments.
Keywords: laser-induced breakdown spectroscopy; matrix-match standard; principal component analysis; heat-treated natural zeolite; sediment samples
ACCEPTED MANUSCRIPT 1. Introduction Laser-induced breakdown spectroscopy (LIBS) makes it possible to perform rapid qualitative and quantitative analysis of trace, major, and minor elements in sediment and soil samples. In this technique, laser pulses ablate and ionize the sample surface. Atomic and ionic
IP
T
emissions from the plasma-material interaction provide fingerprint lines in the spectra that can be
CR
used for elemental analysis [1-4]. Compositional analysis using LIBS has been carried out in a wide range of samples such as rocks, sediments, sludge, aerosols, liquids, alloys and biological
US
samples [5-11]. LIBS analysis in sediments offers several advantages in comparison with classical methods of analysis like atomic absorption spectroscopy (AAS) and inductively coupled
AN
plasma/optical emission spectrometry (ICP/OES). The advantages of LIBS include a fast
M
turnaround time, the in situ analysis of multiple elements in virtually any phase, the potential in obtaining depth-profile composition , and minimal sample amount, treatment, and preparation.
ED
However, LIBS also suffers from several drawbacks due to the nature of analysis. The ablation
PT
mechanism is heavily dependent on instrumental parameters such as laser energy, spot size and spectrometer delay. Repeatability of the analysis becomes challenging because of the laser energy
CE
fluctuations that happen from shot-to-shot [12]. Moreover, since the laser firing is localized, the
AC
spatial distribution of the target analyte in the sample is influenced by the homogeneity of the sample itself. The biggest disadvantage of quantitative LIBS analysis is the dependence of the calibration standards with the sample matrix, also known as the matrix effect [13]. The matrix plays a significant role in the plasma formation process of LIBS and methods to reduce the uncertainties of each pulse measurement have been a focus of various research works [14-16]. Sample-related matrix effects may significantly affect the accuracy of LIBS quantitative analysis. Emission lines and matrix effects occur simultaneously, leading to LIBS signals that are either
ACCEPTED MANUSCRIPT under or over quantified. There is a need for alternative or potentially better matrices that can be used for the detection of metals in sediment samples, especially the ones that provide lower detection limits. LIBS analyses with univariate or multivariate calibration models are commonly coupled to
IP
T
perform quantitative and classification analyses on many environmental samples, such as sediment
CR
and soil samples [17-18]. Commonly used multivariate techniques include principal component analysis (PCA) and partial least-squares discriminant analysis (PLS-DA). Nonetheless, both
US
univariate and multivariate techniques utilize calibration samples and/or standards which must be closely matched with the matrix of the target sample. The use of internal standard and/or
AN
background normalization is commonly done in order to correlate the calibration curves with those
M
from the sample for quantitative analysis, especially for those samples with no suitable standard matrices [19]. Both methods are done to offset the matrix effects of samples and standards. Internal
ED
standard utilizes the intensity of the major element emission line that is independent from the
PT
concentration. In the normalization with background procedure, the intensity of the analyte is normalized with the continuum radiation that is produced from the radiative recombination of ions
CE
with electrons to create atoms. The target analyte emission becomes normalized because the
AC
continuum radiation is directly related to the plasma density and the plasma luminosity [20]. There were several studies reported regarding the analysis of metals in solids utilizing normalization with different calibration matrices to improve the sensitivity and accuracy of the LIBS technique [21-24]. However, the matrices of the binders used in these studies were mostly boric acid (H3BO3) and potassium bromide (KBr) which do not closely emulate the same plasma-material interaction of their respective samples.
ACCEPTED MANUSCRIPT The aim of this study is to evaluate an alternative, inexpensive and stable matrix modifier to improve detection limits in sediment samples. In this work, we performed the detection and quantification of three contaminant metals (Cr, Cu and Pb) in river sediment samples. The prepared standard diluent used was a naturally-occurring zeolite; thus, it potentially matched the matrix of
T
the samples. For this study, the zeolite material acted as both a diluent and the sample binder. The
IP
effects of varying matrices will be highlighted by comparing the natural zeolite with an established
CR
KBr binder matrix.
US
2. Materials and methods
AN
2.1. Sample collection and preparation
Three representative mining areas were used as sediment collection sites in this study.
M
These mining areas run along the Acupan Creek located at Benguet, Philippines. The Acupan
ED
Creek flows through the mining and residential areas of Acupan town , where the livelihood is
PT
primarily based on gold mining operations near the river. The origin, depth, elevation and relevant
Table 1
CE
properties are listed in Table 1.
Code
AC
Location of the sampling sites along the Acupan Creek. Description
Elevation (m)
Depth (m)
Texture
818
0.2
Silt
Midstream section located at the
community, main tributary of the creek S1 system, flows directly from the upstream
ACCEPTED MANUSCRIPT Midstream section located at the S2
community, secondary tributary of the
821
0.2
Silt
828
0.2
Silt
creek system Midstream section located at the S3
IP
T
community, confluence of S1 and S2
CR
Sediment samples collected in this study were sent to a service laboratory for reference
US
measurements. The reference concentrations of Cr (0.01g/mL), Cu (0.02 g/mL) and Pb (0.02 g/mL) were obtained through ICP/OES analysis using Agilent 720 Series ICP-OES (Agilent
AN
Technologies, Santa Clara, CA, USA). ediment samples were initially digested in 20 mL-50 mL of HNO3, HCl and HF mixture, and were subsequently filtered. All solutions were prepared using
M
deionized water. Table 2 shows the metal concentrations of the samples obtained via ICP/OES.
ED
Table 2
PT
Metal concentrations (mg kg-1) in sediment samples analyzed via ICP/OES*
Samples
ICP/OES values (ppm) Cu
Pb
17
147
216
S2
44
200
640
S3
28
165
293
AC
S1
CE
Cr
* Complete elemental composition of the sediment samples are found in Supporting Data
For the preparation of sediment samples for LIBS analysis, the samples were initially oven-dried at 50°C for 2 days. The oven-dried samples were sieved into a stainless steel mesh (0.707 mm, US
ACCEPTED MANUSCRIPT Sieve No. 24). The resulting samples were ground into finer particles using a mortar and pestle. The sediments were further crushed and ground using an electrical ball mill. The homogenized samples were pelletized by subjecting the samples into a 4-ton load for 10 seconds in a uniaxial hydraulic press.
IP
T
Calibration curve preparation
CR
Two binders, as constituents of the calibrating materials, namely, potassium bromide (KBr), and natural zeolite (NZ) mined from Pangasinan, Philippines, were examined in order to
US
choose the most suitable matrix for the calibration standards. NZ powders were calcined and
AN
referred to as heat-treated NZ in the study. The calcination involved the heating of the NZ powder for 2 hours at 100°C and for 6 hours at 700°C in order to remove aluminum, volatile organic
M
materials and residual metal content. High purity metal standard reference material (SRM)
ED
powders (99.99% purity, Sigma-Aldrich) were used as metal source. The target metals for the study were Cr, Cu and Pb. Calibration standards were prepared by mixing the standards with the
PT
matrices in terms of mass. The calibration standard powders were mixed initially with either KBr
CE
or NZ using mortar and pestle before the resultant mixture was homogenized in an electrical ball mill. The homogenized calibration standards with the binder were then pelletized using a 4-ton
AC
load for 10 seconds in a uniaxial hydraulic press, similar to that used for sediment samples. 2.2. Optimization of instrument parameters Four instrument parameters were optimized in terms of intensity prior to analysis. Laser energy, laser spot size, spectrometer delay and type of firing were optimized. The main atomic emission lines used for elemental identification are 520.6 nm, 324.7 nm and 405.8 nm for Cr, Cu, and Pb, respectively.
ACCEPTED MANUSCRIPT A commercial LIBS set-up (J200, Applied Spectra, Fremont, CA, USA) was used in this study. Axiom software (Applied Spectra, Fremont, CA, USA) was used in operating the LIBS setup. The laser was a Q-switched Nd:YAG laser operated at the fundamental wavelength of 1064 nm with repetition rate of 20 Hz and pulse width of 6 ns. The laser was fired into the pelletized
T
calibration standards and samples at atmospheric pressure. Analyses were conducted under
IP
optimized parameters, namely: 0.5 µs (gate delay), 45 mJ (laser energy), 75 µm (spot size), and
CR
raster firing. A 5x5 grid with a total area of approximately 65 mm2 was ablated through raster patterns with the speed of 10 mm/s totaling to 219 shots. Light coming from the plasma was
US
collected and recorded into a high resolution 6-channel CCD spectrometer through a high
AN
transmission optical fiber collection system. The built-in spectrometer had a spectral range of 190 nm to 1040 nm, and full spectrum analysis had been performed within this range to get complete
M
information on the elements of interest. The plasma temperature was obtained using Boltzmann
ED
plot method, while the electron density was obtained using the Saha-Boltzmann equation. Table 3 shows the Fe emission lines used for the estimation of plasma parameters. Fe I and II were used
PT
due to the high percent abundance of Fe in the sediment samples. The plasma temperature reached
CE
10808.2 K at a gate delay time of 500 ns. On the other hand, the estimated electron density at 500
Table 3
AC
ns was 4.20634 x1027/cm3.
Spectroscopic parameters used for Tplasma determination Upper Transition energy Wavelength Element probability level (nm) (1x107 s-1) (1x104 cm-1) Fe I
246.26
4.06
5.85
296.69
3.37
2.72
Fe I
344.06
2.91
1.71
Fe I
344.10
2.95
1.24
Fe I
347.55
2.95
0.98
Fe I
349.06
2.91
0.61
Fe I
358.12
3.48
10.20
Fe I
360.89
3.59
8.13
Fe I
361.88
3.56
7.22
Fe I
367.99
2.72
0.14
Fe I
374.56
2.74
1.15
Fe I
374.83
2.76
0.92
Fe I
374.95
3.40
Fe I
375.82
3.43
6.34
Fe I
376.38
3.45
5.44
Fe I
381.58
Fe I
382.04
Fe I Fe I
ED
PT
7.63
11.20
3.31
6.67
385.64
2.63
0.46
385.99
2.59
0.97
387.86
2.65
0.62
Fe I
388.63
2.61
0.53
Fe I
404.58
3.67
8.62
Fe I
406.36
3.72
6.65
Fe I
432.58
3.61
5.16
Fe I
438.36
3.48
5.00
AC
CE
3.82
Fe I
T
Fe I
IP
34.40
CR
4.08
US
249.06
AN
Fe I
M
ACCEPTED MANUSCRIPT
ACCEPTED MANUSCRIPT Fe II
249.12
4.10
29.10
Fe II
259.94
3.85
23.50
T
2.3. PCA Analysis
IP
The Aurora software (Applied Spectra, Fremont, CA, USA) was used to determine the spectral
CR
line information for the target elements. The preliminary evaluation was done with Principal Component Analysis (PCA) in order to get an understanding on existing trends and to recognize
US
property variations in the chemical data for each sample. PCA was also employed to validate the
AN
emission lines used and to qualitatively distinguish the uniformity of the binder with the sediment samples. A total of 12,288 spectral data points were analyzed and were used as input variables for
M
the PCA method. No pre-processing technique for each LIBS spectrum was employed before
ED
performing PCA analysis. The constructed PCA models were used only for visualization purposes
PT
and no optimization were done for determining the number of principal components.
CE
3. Results and Discussion
AC
3.1. Calibration curve
The performance of the two types of binders employed in this study has been primarily evaluated by linearity and limit of detection. In order to examine the capability of LIBS for singleelement quantitative analysis in sediments, a calibration set was prepared for each target metal wherein known amounts of metal targets were added with KBr or natural zeolite matrix.. Both univariate and multivariate techniques need calibration standards that are capable of simulating
ACCEPTED MANUSCRIPT the laser-material interaction of the sample. However, exact replication of the behavior and particle interaction between the laser and sample material is not possible. Thus, finding a suitable matrix with nearly the same ablation threshold is necessary for the quantitative analysis of the target metals in the sediments. The first matrix used was potassium bromide (KBr). Along with boric
T
acid (H3BO3), the KBr binder has been used in most of the LIBS-based analyses because of its
IP
good binding ability and the uncomplicated matrix that it forms with calibration samples [25-29].
CR
The normalized intensity of the target metals, Cr (520.6 nm), Cu (324.7 nm) and Pb (405.8 nm), were plotted versus the known metal concentration. The metal concentrations used in all
US
calibration sets were 0, 10, 25, 50, 100 and 200 ppm. Figure 1 (a-c) shows the calibration curves
AN
for Cr, Cu and Pb in the KBr matrix.
M
The three calibration curves prepared using KBr (Figure 1) have satisfactory values of coefficient of variation (R2 = 0.900 - 0.973). This indicates that the LIBS signal intensities of the
ED
calibration standards were linearly correlated at low concentrations. Even though these R2 values
PT
are not as high compared to those obtained from standards used in ICP-OES analysis, it should be noted that the matrix phase used for LIBS analysis was solid which induces inherent problems
CE
with homogeneity. Nonetheless, the linear correlations of these calibration curves were
AC
comparable to previous LIBS literature [7, 25, 30-31]. The second matrix used was a natural zeolite obtained from a mining quarry in the province of Pangasinan, Philippines. The composition of the natural zeolite after calcination showed negligible concentrations of the target metals as shown in Table 4. The total metal concentrations for the three metals were below the method limit of detection (MDL). Table 4 Metal concentration of natural zeolite.
ACCEPTED MANUSCRIPT Metal
Concentration (ppm)
Cr
<5.0*
Cu
<1.0*
Pb
<5.0*
IP
T
* metal concentration values fall below the method detection limit
CR
The NZ binder was heat treated to be free from metal impurities but may have caused misleading
US
interpretation of the results. Pre-treatment step allowed the natural zeolite to be a suitable binder in the preparation of the calibration curves. Figure 2 shows the calibration curves for Cr, Cu and
AN
Pb using the natural zeolite. The standard deviation values are found to be high and this can be
M
associated to the innate heterogenous composition nature of the zeolite in contrast to the relative
ED
homogenous nature of KBr.
Similar with the R2 values for the KBr matrix, the calibration curves using the NZ heat-
PT
treated binder have generated high R2 values (R2 = 0.96 - 0.99). These values were comparable
CE
with literature values cited previously. Considering that multiple minerals/elements constitute the natural zeolite, the linear correlation was still comparable with the much simpler KBr matrix. This
AC
positive correlation may come from two reasons, (1) homogeneity in the calibration sample preparation, and (2) output data normalization. Since the calibration samples were solid pellets, the preparation was largely affected by the homogeneity of the powder mixture. The mixing process was initially done by a mortar and pestle set-up. These powders were further mixed in an electrical ball mill that was set to 100 rpm. This two-step mixing process ensured uniformity of the powder samples for ablation. Another factor that helped to improve homogeneity is by increasing the sample area during the ablation process. By averaging the spectral intensities
ACCEPTED MANUSCRIPT coming from a much larger firing area, better representative data were collected. The second reason for the comparable linearity between the natural zeolite matrix and KBr was the normalization of the spectra with the background. In quantitative LIBS analysis, adding an internal standard to the calibration standards is a common practice in order to correlate the calibration curves with the
T
sample. This is particularly necessary for those samples with no matching matrices [32]. The
IP
intensity of the target metal will then be normalized with the intensity of the internal standard
CR
placed. Another common approach of data processing technique is normalization with the background, with or without the presence of an internal standard. The intensity of the target metal
US
is normalized with the continuum radiation that is produced when ions interact with electrons
AN
through radiative recombination forming atoms [33]. In this study, normalization with the background was done in order to offset the effects of the varying matrices between the samples
ED
3.2. Limit of detection
M
and the calibration standards.
PT
The sensitivity of the procedure was evaluated using the prepared calibration curves.
CE
Equation 1 was used for the calculation of the limit of detection (LOD) of the calibration curves. The σ is the standard deviation of the blank (0 ppm) and s is the slope of the line of the calibration
AC
curve.
𝐿𝑂𝐷 (𝑝𝑝𝑚) =
3𝜎 𝑠
Equation 1. Limit of detection. For the calculation of the LOD, 25 trials using the variability of signals from the blank sample were done for each metal in both matrices. Each trial consisted of 219 shots equivalent to 219 spectra. The peak areas of intensities of the target analyte in all spectra were averaged and
ACCEPTED MANUSCRIPT were normalized with the continuum background intensity. Table 5 summarizes the results obtained from the analysis of limit of detection for each metal considered in a either KBR or NZ matrix. Table 5
LOD in KBr (ppm)
Cr
1.43 ± 0.23
Cu
0.41 ± 0.01
Pb
2.47 ± 0.02
LOD in NZ (ppm) 6.10 ± 0.02 4.26 ± 0.01 2.65 ± 0.01
AN
US
CR
Metal analyte
IP
T
Limit of detection of each metal in KBr and natural zeolite (NZ) matrix.
M
The values presented in Table 5 were comparable with the LOD values using LIBS
ED
reported in previous literature [7, 30, 34-36]. Comparison between the two matrices shows that use of KBr provided a lower detection limit compared to the natural zeolite. This trend was
PT
expected since the KBr presented a less complicated matrix compared to the natural zeolite. Aside
CE
from the aluminosilicate framework, elements such as calcium, magnesium and iron were also present in the natural zeolite. The heat-treatment was performed to eliminate organic and more
AC
volatile metal components from the zeolite. This pre-treatment step did not produce significant changes in the framework of the zeolite as seen in XRD data (Supporting Data). 3.3. LIBS vs. ICP/OES analysis Three midstream sediment samples (S1, S2 and S3) were used as real samples for the study. ICP/OES was used in validating the results of the LIBS analysis. The representative LIBS spectra of the three samples, along with the integration peaks used per element, were illustrated in Figure
ACCEPTED MANUSCRIPT 3. The results from the ICP/OES analysis were averaged and used as the reference values in this study. Tables 6-8 summarizes the reported composition of the S1 sediment sample using LIBS and ICP/OES analysis as well as the relative percent difference (RPD) between these two analytical methods. The RPD was calculated to describe the variability between ICP/OES and LIBS. RPD
IP
T
was not determined for values less than the laboratory reporting level or method detection limit.
CR
Table 6 Comparison of LIBS and ICP/OES values for S1.
Matrix
Analyte
LIBS value (ppm)
US
ICP/OES value
RPD (%) #
(ppm) NA
Cu
121 ± 8.18
147
19.4
Pb
<2.47*
216
NA
Cr
16.9 ± 0.63
17
0.590
Cu
125 ± 16.1
147
16.2
216
125
Pb
AN
17
M
<1.43*
ED
NZ
Cr
PT
KBr
938 ± 27.0
CE
* metal concentrations values fall below the method detection limit
AC
# % Relative percent difference was calculated by taking the absolute value of the difference over the mean times 100 NA: not applicable
Several values were not determined (indicated as NA) for S1 using the calibration curves generated for Cr and Pb measurements in the KBr matrix as shown in Table 6. These results may be caused by the difference in material ablation threshold of the KBr and the S1 sediment sample. Moreover, it should be noted that the percent difference observed for Cu was 18.6% and considered
ACCEPTED MANUSCRIPT to be a relatively low value. Results of LIBS analysis for the natural zeolite were comparable with the ICP/OES analysis and can be calculated using the generated calibration curves. Interpolation of concentration for Cr and Cu yielded results close to the reference value. In comparison to KBr, the zeolite provided a matrix with comparable laser threshold value with sediment. However, the
T
extrapolated concentration of Pb was far from the ICP/OES value. The calculated error in
IP
extrapolation could be due to the low linear correlation of signal-to-noise ratio with metal
CR
concentration at higher concentration. Similar trend were obtained for S2 (Table 7) and S3 (Table
Table 7
ICP/OES value (ppm)
RPD (%)#
<1.43*
44
NA
Cu
166 ± 30.9
200
18.6
Pb
ED
AN
Comparison of LIBS and ICP/OES values for S2.
<2.47*
640
NA
45.1 ± 1.82
44.0
2.47
191 ± 19.8
200
4.61
630 ± 19.2
640
1.57
Analyte
LIBS value (ppm)
KBr
Cr
PT
M
Matrix
Cr Cu
AC
Pb
CE
NZ
US
8) sediment samples.
* metal concentrations values fall below the method detection limit # % Relative percent difference was calculated by taking the absolute value of the difference over the mean times 100 NA: not applicable
Table 8
ACCEPTED MANUSCRIPT Comparison of LIBS and ICP/OES values for S3. Analyte
LIBS value (ppm)
ICP/OES value (ppm)
RPD (%) #
KBr
Cr
<1.43*
28
NA
Cu
120 ± 9.16
165
31.6
Pb
<2.47*
293
NA
Cr
28.2 ± 0.99
28
Cu
141 ± 20.0
165
15.7
Pb
693 ± 115
293
134
IP
CR
0.712
US
NZ
T
Matrix
* metal concentrations values fall below the method detection limit
AN
# % Relative percent difference was calculated by taking the absolute value of the difference over the mean times 100
M
NA: not applicable
ED
Measurements done on samples prepared in NZ matrix were all determined and can be
PT
compared with ICP/OES results. Calculated concentration values of Cr and Pb in KBr were lower than the corresponding calculated value for the S2 and S3 samples prepared in NZ matrix. Results
CE
for Cu concentration using LIBS remained satisfactory with respect to ICP/OES values for the
AC
KBr matrix. On the other hand, both Cr and Cu have low deviations from the ICP/OES values in NZ matrix. Extrapolated LIBS result of S2 sample for Pb in zeolite was close with the reference value. However, the S3 sample produced erroneous result which was far from the reference value for Pb in the zeolite matrix. 3.4. Principal component analysis
ACCEPTED MANUSCRIPT In order to evaluate and visualize the difference in material property of the matrices and the samples, principal component analysis (PCA) was utilized. The full LIBS spectra of the three samples (S1, S2 and S3), along with the two matrices (KBr and NZ), were transformed into several principal components (PCs) through PCA. The first 5 PCs explained 96.04% of the total variation
T
in the data. The loading plot of the first five components (Figure 4) validates the presence of the
IP
chosen emission lines (Cu 324.7, Pb 405.8 and Cr 520.6) in the five PCA samples. However, the
CR
loading values for the three elements were relatively low (from -0.1 to 0.1) and can be directly
US
related to the total amount of the respective elements in the sample & matrices. Moreover, the degree of variation between the samples and binders were further analyzed
AN
using the same PCA run. Figure 5 shows the orthographic view of the PCA plot of the samples
M
and matrices. The first two components explained 94.17% (PC 1: 92.59%; PC 2: 1.58%) of the variations of the spectral information presented. Each circle corresponds to the linear combination
ED
of the LIBS spectra of the first two principal components. The preparation of the calibration curves
PT
involved the use of binders that act as the diluent of the high purity metal powders. A homogeneous matrix that produces signals similar with the sample is needed in order to obtain results that are
CE
comparable with classical analytical methods. Interpretation of the PC score plot suggests that
AC
discrimination between the two matrices and the samples can be characterized in terms of the lasermaterial interaction. A pattern can be recognized, as seen in the PCA plot (Fig. 5), wherein for the natural zeolite the pattern is clustered closer to the sediment samples compared to the KBr matrix. It can also be observed from the plot that that the sediment samples and natural zeolite are identified in proximity to PC 2, while the KBr matrix approached PC 1.The PCA plot was generated from the laser intensity signals of each of the matrices and the samples. The plot showed the natural distribution of samples indicating the similar chemical composition among the
ACCEPTED MANUSCRIPT sediment samples and natural zeolite. The KBr matrix was clearly separated from the zeolite matrix and sediment samples. This signifies that the natural zeolite simulated closer the ablation threshold and emission intensity of the sediment samples compared to KBr. This would indicate that laser and matrix would have a better interaction and both the zeolite and sediment samples
IP
T
would be much better recognized as homogeneous or uniform.
CR
4. Conclusion
The utilization of PCA showed that heat-treated zeolite is a potential diluent or binder for
US
generating calibration curves and provide matrix-matched standards in the identification of target
AN
metals from unknown sediment samples. The score plot of PCs was used to reveal the similarity of the natural zeolite and sediment samples based on their respective LIBS spectra. The results
M
showed that the detection and quantification of Cr, Cu and Pb in sediments via LIBS can be
ED
comparable to an established analytical technique like ICP/OES by using an appropriate matrix, such as heat-treated natural zeolite as reported in this study. Single-element calibration curves
PT
with good linear correlation for each metal target were generated. However, extrapolation of data
CE
from the curve did not produce satisfactory results which signified the non-linearity of concentration and signal-to-noise ratio at higher concentrations. Limit of detections for each curve
AC
produced comparable results with previous literature. The accuracy of the LIBS method was evaluated by relating the values to ICP/OES analysis. The similarity between the material properties of the natural zeolite matrix and the sediment samples produced comparable statistical values in comparison with the KBr matrix and other LIBS literature.
ACCEPTED MANUSCRIPT Acknowledgments The research was supported and funded by the Department of Science & Technology - Philippine Council for Industry, Energy and Emerging Technology Research and Development (DOSTPCIEERD). The study was under Project 1 of the Chemical Sensors for Mine Site Monitoring
AC
CE
PT
ED
M
AN
US
CR
IP
T
Program implemented at the Institute of Chemistry, University of the Philippines – Diliman.
ACCEPTED MANUSCRIPT References [1] F.J. Fortes, J. Moros, P. Lucena, L.M. Cabalin, J.J. Laserna, Laser-induced breakdown spectrosocopy, Analytical. Chemistry, 85 (2013) 640-669. [2] R.S. Harmon, R.E. Russo, R.R. Hark, Applications of Laser-induced breakdown spectroscopy for
IP
T
geochemical and environmental analysis: A comprehensive review , Spectrochimica Acta Part B, 87 (2013)
CR
11-26.
[3] F. Anabitarte, A. Cobo, J.M. Lopez-Higuera, Laser-induced breakdown spectroscopy: Fundamentals, and
challenges,
ISRN
Spectroscopy,
Article
ID:
285240
(2012)
12
pages.
US
applications
AN
doi:10.5402/2012/285240.
[4] D.A. Cremers, and L.J. Radziemski, Handbook of Laser-induced breakdown spectroscopy, John Wiley
M
and Sons Ltd, West Sussex, England, 2006.
ED
[5] J.-B. Sirven, B. Salle, P. Mauchien, J.-L. Lacour, S. Maurice, G. Manhes, Feasibility study of rock identification at the surface of Mars by remote laser-induced breakdown spectroscopy and three
PT
chemometric methods, Journal of Analytical Atomic Spectrometry, 22 (2007) 1471-1480.
CE
[6] M.Z. Martin, S.D. Wullschleger, C.T. Garten, A.V. Palumbo, Laser-induced breakdown spectroscopy
2077.
AC
for the environmental determination of total carbon and nitrogen in soils, Applied Optics, 42 (2003) 2072-
[7] F. Capitelli, F. Colao, M. Provenzano, R. Fantoni, G. Brunetti, N. Senesi, Determination of heavy metals in soils by laser induced breakdown spectroscopy, Geoderma, 106 (2002) 45-62. [8] W.B. Lee, J. Wu, Y.I. Lee, J. Sneddon, Recent applications of laser‐induced breakdown spectrometry: a review of material approaches, Applied Spectroscopy Reviews, 39 (2004) 27-97.
ACCEPTED MANUSCRIPT [9] M. Sabsabi, P. Cielo, Quantitative analysis of aluminum alloys by laser-induced breakdown spectroscopy and plasma characterization, Applied Spectroscopy, 49 (1995) 499-507. [10] A. Kumar, F.-Y. Yueh, J.P. Singh, S. Burgess, Characterization of malignant tissue cells by laserinduced breakdown spectroscopy, Applied Optics, 43 (2004) 5399-5403.
IP
T
[11] Q. Sun, M. Tran, B.W. Smith, J.D. Winefordner, Zinc analysis in human skin by laser induced-
CR
breakdown spectroscopy, Talanta, 52 (2000) 293-300.
[12] Y.L. Yang, J.X. Li, F.W. Meng, C.H. Ma, Quantitative analysis of LIBS reduces the matrix effect
US
methods, Applied Mechanics and Materials, Trans Tech Publ, 2013, 579-582.
AN
[13] D.W. Hahn, N. Omenetto, Laser-induced breakdown spectroscopy (LIBS), part II: review of instrumental and methodological approaches to material analysis and applications to different fields,
M
Applied spectroscopy, 66 (2012) 347-419.
ED
[14] K-R. Kim, G. Kim, J-Y. Kim, K. Park, K-W. Kim. Kriging interpolation method for laser-induced breakdown spectroscopy analysis of Zinc in various soils, Journal of Analytical Atomic Spectrometry, 29
PT
(2014) 76-84.
CE
[15] S-J. Choi, K-J. Lee, J.J. Yoh. Quantitative laser-induced breakdown spectroscopy of standard reference
AC
materials of various categories, Applied Physics B, 3 (2013) 379-388. [16] N. Gilon, J. El-Haddad, J. Stankova, W. Lei, Q. Ma, V. Motto-Ros, J. Yu, A matrix effect and accuracy evaluation for the determination of elements in milk powder LIBS and laser ablation/ICP-OES spectroscopy, Analytical and Bioanalytical Chemistry, 401 (2011) 2681-2689. [17] M.Z. Martin, N. Lobbe, N. Andre, R. Harris, M. Ebinger, S.D. Wullschleger, A.A. Vass. High reolution applications of laser-induced breakdown spectroscopy for environmental and forensic applications. Spectrochimica Acta B, 62 (2007) 1426-1432.
ACCEPTED MANUSCRIPT [18] S.M. Zaytsev, A.M. Popov, E.V. Chernykh, R.D. Voronina, N.B. Zorov, T.A. Labutin, Comparison of single-and multivariate calibration for determination of Si, Mn, Cr and Ni in high-alloyed stainless steels by laser-induced breakdown spectrometry, Journal of Analytical Atomic Spectrometry, 29 (2014) 14171424.
T
[19] D. Diaz, D.W. Hahn, A. Molina, Evaluation of laser-induced breakdown spectroscopy (LIBS) as a
IP
measurement technique for evaluation of total elemental concentration in soils, Applied Spectroscopy, 66
CR
(2012) 99-106.
US
[20] G. Senesi, M. Dell’Aglio, R. Gaudiuso, A. De Giacomo, C. Zaccone, O. De Pascale, T. Miano, M. Capitelli, Heavy metal concentrations in soils as determined by laser-induced breakdown spectroscopy
AN
(LIBS), with special emphasis on chromium, Environmental Research, 109 (2009) 413-420. [21] S. Gornushkin, I. Gornushkin, J. Anzano, B. Smith, J. Winefordner, Effective normalization technique
M
for correction of matrix effects in laser-induced breakdown spectroscopy detection of magnesium in
ED
powdered samples, Applied Spectroscopy, 56 (2002) 433-436.
PT
[22] F. Colao, R. Fantoni, V. Lazic, L. Caneve, A. Giardini, V. Spizzichino, LIBS as a diagnostic tool during the laser cleaning of copper based alloys: experimental results, Journal of Analytical Atomic
CE
Spectrometry, 19 (2004) 502-504.
AC
[23] N.B. Zorov, A.A. Gorbatenko, T.A. Labutin, A.M. Popov, A review of normalization techniques in analytical atomic spectrometry with laser sampling: From single to multivariate correction, Spectrochimica Acta Part B: Atomic Spectroscopy, 65 (2010) 642-657. [24] X. Li, Z. Wang, S.-L. Lui, Y. Fu, Z. Li, J. Liu, W. Ni, A partial least squares based spectrum normalization method for uncertainty reduction for laser-induced breakdown spectroscopy measurements, Spectrochimica Acta Part B: Atomic Spectroscopy, 88 (2013) 180-185.
ACCEPTED MANUSCRIPT [25] D. Santos, L.C. Nunes, L.C. Trevizan, Q. Godoi, F.O. Leme, J.W.B. Braga, F.J. Krug, Evaluation of laser induced breakdown spectroscopy for cadmium determination in soils, Spectrochimica Acta Part B: Atomic Spectroscopy, 64 (2009) 1073-1078. [26] A. Sarkar, D. Alamelu, S.K. Aggarwal, Laser-induced breakdown spectroscopy for determination of
IP
T
uranium in thorium–uranium mixed oxide fuel materials, Talanta, 78 (2009) 800-804. [27] N. Carmona, M. Oujja, S. Gaspard, M. García-Heras, M. Villegas, M. Castillejo, Lead determination
CR
in glasses by laser-induced breakdown spectroscopy, Spectrochimica Acta Part B: Atomic Spectroscopy,
US
62 (2007) 94-100.
[28] T. Ctvrtnickova, M.-P. Mateo, A. Yanez, G. Nicolas, Characterization of coal fly ash components by
AN
laser-induced breakdown spectroscopy, Spectrochimica Acta Part B: Atomic Spectroscopy, 64 (2009)
M
1093-1097.
[29] M. Gondal, T. Hussain, Z. Yamani, M. Baig, The role of various binding materials for trace
ED
elemental analysis of powder samples using laser-induced breakdown spectroscopy, Talanta, 72 (2007)
PT
642-649.
CE
[30] F. Hilbk-Kortenbruck, R. Noll, P. Wintjens, H. Falk, C. Becker, Analysis of heavy metals in soils using laser-induced breakdown spectrometry combined with laser-induced fluorescence, Spectrochimica
AC
Acta Part B: Atomic Spectroscopy, 56 (2001) 933-945. [31] D.D. Pace, C. D'Angelo, D. Bertuccelli, G. Bertuccelli, Analysis of heavy metals in liquids using Laser Induced Breakdown Spectroscopy by liquid-to-solid matrix conversion, Spectrochimica Acta Part B: Atomic Spectroscopy, 61 (2006) 929-933. [32] R. Gaudiuso, M. Dell’Aglio, O.D. Pascale, G.S. Senesi, A.D. Giacomo, Laser induced breakdown spectroscopy for elemental analysis in environmental, cultural heritage and space applications: a review of methods and results, Sensors, 10 (2010) 7434-7468.
ACCEPTED MANUSCRIPT [33] G.S. Senesi, M. Dell'Aglio, A. De Giacomo, O. De Pascale, Z.A. Chami, T.M. Miano, C. Zaccone, Elemental Composition Analysis of Plants and Composts Used for Soil Remediation by Laser‐Induced Breakdown Spectroscopy, CLEAN–Soil, Air, Water, 42 (2014) 791-798. [34] W.T.Y. Mohamed, Improved LIBS limit of detection of Be, Mg, Si, Mn, Fe and Cu in aluminum
T
alloy samples using a portable Echelle spectrometer with ICCD camera, Optics & Laser Technology, 40
IP
(2008) 30-38.
CR
[35] M.A. Ismail, H. Imam, A. Elhassan, W.T. Youniss, M.A. Harith, LIBS limit of detection and plasma
US
parameters of some elements in two different metallic matrices, Journal of Analytical Atomic Spectrometry, 19 (2004) 489-494.
AN
[36] R. Knopp, F. Scherbaum, J. Kim, Laser induced breakdown spectroscopy (LIBS) as an analytical tool for the detection of metal ions in aqueous solutions, Fresenius' journal of analytical chemistry, 355
AC
CE
PT
ED
M
(1996) 16-20.
ACCEPTED MANUSCRIPT
CR
IP
T
List of Figures
AC
CE
PT
ED
M
AN
US
Figure 1. Calibration curves of (a) Cr, (b) Cu and (c) Pb prepared with KBr binder.
Figure 2. Calibration curves of (a) Cr, (b) Cu and (c) Pb prepared with natural zeolite binder.
PT
ED
M
AN
US
CR
IP
T
ACCEPTED MANUSCRIPT
Figure 3. Overlapped LIBS spectra of S1, S2 and S3. Integration peaks (red boxed region)
AC
CE
highlights the respective emission lines used (Cu 324.7, Pb 405.8 and 520.6).
CE
PT
ED
M
AN
US
CR
IP
T
ACCEPTED MANUSCRIPT
Figure 4. Loading plot of the first five principal components from the PCA of the LIBS spectra of
AC
the three sediment samples and two matrices. Emission lines used for Cr (520.6 nm), Cu (324.7 nm) and Pb (405.8 nm) were identified.
AC
CE
PT
ED
M
AN
US
CR
IP
T
ACCEPTED MANUSCRIPT
Figure 5. Scatter plot of two principal components (PC1 and PC2) for KBr, heat-treated natural zeolite, and sediment samples S1, S2 and S3.
ACCEPTED MANUSCRIPT GRAPHICAL ABSTRACT
AC
CE
PT
ED
M
AN
US
CR
IP
T
Laser-induced breakdown spectroscopy for the quantitative analysis of metals in sediments using natural zeolite matrix
ACCEPTED MANUSCRIPT
HIGHLIGHTS Laser-induced breakdown spectroscopy for the quantitative analysis of metals in sediments using natural zeolite matrix Heat-treated natural zeolite is a suitable matrix for the quantitative LIBS analyses of sediments
Use of an alternative and inexpensive binder that act as diluent of high purity metal powders in preparation for the calibration curves
Detection and quantification of Cu, Cr, Pb in river sediment samples using a potentially matrix-matched standard
AC
CE
PT
ED
M
AN
US
CR
IP
T