Muscle oxygenation measurement in humans by noninvasive optical spectroscopy and Locally Weighted Regression

Muscle oxygenation measurement in humans by noninvasive optical spectroscopy and Locally Weighted Regression

Accepted Manuscript Title: Muscle Oxygenation Measurement in Humans by Noninvasive Optical Spectroscopy and Locally Weighted Regression Author: Lorile...

736KB Sizes 0 Downloads 197 Views

Accepted Manuscript Title: Muscle Oxygenation Measurement in Humans by Noninvasive Optical Spectroscopy and Locally Weighted Regression Author: Lorilee S.L. Arakaki Kenneth A. Schenkman Wayne A. Ciesielski Jeremy M. Shaver PII: DOI: Reference:

S0003-2670(13)00652-1 http://dx.doi.org/doi:10.1016/j.aca.2013.05.003 ACA 232574

To appear in:

Analytica Chimica Acta

Received date: Revised date: Accepted date:

7-12-2012 29-4-2013 3-5-2013

Please cite this article as: L.S.L. Arakaki, K.A. Schenkman, W.A. Ciesielski, J.M. Shaver, Muscle Oxygenation Measurement in Humans by Noninvasive Optical Spectroscopy and Locally Weighted Regression, Analytica Chimica Acta (2013), http://dx.doi.org/10.1016/j.aca.2013.05.003 This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

1 2

THIS IS THE HIGHLIGHTED COPY. ALL REVISIONS TO THE ORIGINAL MANUSCRIPT ARE UNDERLINED AND ARE IN RED.

3 Muscle Oxygenation Measurement in Humans by Noninvasive Optical Spectroscopy and

5

Locally Weighted Regression

6

Lorilee S. L. Arakaki,a,* Kenneth A. Schenkman,a,b Wayne A. Ciesielski,a and Jeremy M. Shaverc

cr

ip t

4

7

us

8 9

11

a

12

University of Washington

13

Box 356320

14

Seattle, WA 98195

15

USA

an

10

te

d

M

Department of Pediatrics

Ac ce p

16 17

b

18

University of Washington

19

Box 356540

20

Seattle, WA 98195

21

USA

Department of Anesthesiology

22 23

c

24

3905 West Eaglerock Dr.

25

Wenatchee, WA 98801

26

USA

Eigenvector Research, Inc.

Page 1 of 25

*Corresponding Author

28

Department of Pediatrics

29

University of Washington

30

Box 356320

31

Seattle, WA 98195

32

USA

33

Tel: 1-206-543-3373

34

Email: [email protected]

us

cr

ip t

27

an

35 36

M

37 Email addresses for co-authors:

39

Kenneth A. Schenkman: [email protected]

40

Wayne A. Ciesielski: [email protected]

41

Jeremy M. Shaver: [email protected]

te

Ac ce p

42

d

38

Page 2 of 25

42 Highlights

44

We propose a new method for measuring muscle oxygenation (Mox) from optical spectra.

45

Optical reflectance spectra were measured from the hand in the visible-NIR region (540-800 nm).

46

Locally Weighted Regression (LWR) with Partial Least Squares (PLS) was implemented.

47

Muscle oxygenation using LWR-PLS was more accurate than standard PLS.

48

Multivariate Curve Resolution can estimate Mox under controlled experimental conditions.

cr

us

49

Abstract

We have developed a method to make real-time, continuous, noninvasive measurements of

M

52

an

50 51

ip t

43

muscle oxygenation (Mox) from the surface of the skin. A key development was measurement in both

54

the visible and near infrared (NIR) regions. Measurement of both oxygenated and deoxygenated

55

myoglobin and hemoglobin resulted in a more accurate measurement of Mox than could be achieved

56

with measurement of only the deoxygenated components, as in traditional near-infrared spectroscopy

57

(NIRS). Using the second derivative with respect to wavelength reduced the effects of scattering on the

58

spectra and also made oxygenated and deoxygenated forms more distinguishable from each other.

59

Selecting spectral bands where oxygenated and deoxygenated forms absorb filtered out noise and

60

spectral features unrelated to Mox. NIR and visible bands were scaled relative to each other in order to

61

correct for errors introduced by normalization. Multivariate Curve Resolution (MCR) was used to

62

estimate Mox from spectra within each data set collected from healthy subjects. A Locally Weighted

63

Regression (LWR) model was built from calibration set spectra and associated Mox values from 20

64

subjects using 2562 spectra. LWR and Partial Least Squares (PLS) allow accurate measurement of Mox

65

despite variations in skin pigment or fat layer thickness in different subjects. The method estimated Mox

66

in five healthy subjects with an RMSE of 5.4%.

Ac ce p

te

d

53

Page 3 of 25

67 KEYWORDS

69

Near infrared spectroscopy, Visible spectroscopy, Locally weighted regression, Muscle oxygenation,

70

Tissue oxygenation, Hemoglobin

ip t

68

71 ABBREVIATIONS

73

Muscle oxygenation (Mox), Near infrared (NIR), Near-infrared spectroscopy (NIRS), Locally Weighted

74

Regression (LWR), Multivariate Curve Resolution (MCR), Partial Least Squares (PLS), total absorbance

75

summed across all wavelengths (Aobs), effective path length (B), total absorptivities summed across all

76

wavelengths of the oxygenated and deoxygenated species, respectively (Eo and Ed), concentrations of

77

oxygenated and deoxygenated species, respectively (Co and Cd), root mean squared error of cross-

78

validation (RMSECV), principal components (PCs), root mean square error (RMSE)

us

an

M

d te

80

Ac ce p

79

cr

72

Page 4 of 25

80 81 82

1. INTRODUCTION Optical spectroscopy is increasingly being used in the clinical setting to noninvasively monitor patients. The nonionizing radiation used and the potential for inexpensive, portable devices make

84

optical spectroscopy attractive for continuous monitoring in challenging clinical settings like

85

ambulances, emergency departments, intensive care units, and operating rooms.

cr

Hemoglobin is one of the most commonly measured biological chromophores. It is a protein in

us

86

ip t

83

blood that has four oxygen binding sites. Pulse oximetry is the standard of care in hospitals and uses two

88

wavelengths, one in the visible region and one in the infrared, to measure the oxygen saturation of

89

hemoglobin in arterial blood. While pulse oximetry is valuable in the assessment of oxygen availability in

90

the arterial blood, it does not yield information about whether enough oxygen is actually getting into

91

tissues to meet metabolic demands.

M

an

87

Because oxygen is essential for life, an enormous amount of medical treatment centers around

93

normalizing or stabilizing oxygen levels in the body. Despite significant advances in medical monitoring,

94

there currently is no device that continuously and quantitatively measures oxygen levels at the cellular

95

level, where oxygen is metabolized. Recently, there has been interest in optical monitoring of tissue

96

oxygenation in brain [1, 2] and muscle [2, 3]. Near-infrared spectroscopy (NIRS) devices typically

97

measure 2-6 wavelengths in the NIR region. Measurement of only a few wavelengths limits the ability to

98

sensitively measure tissue hemoglobin saturation. In a study of trauma patients, a commercially

99

available NIRS device that measures six wavelengths could identify severe hemorrhagic shock, but could

100

not distinguish mild and moderate shock from healthy controls [4]. Another NIRS device that measures

101

two NIR wavelengths has found increasing popularity as a trend monitor in operating rooms and

102

intensive care units, but its clinical value has been questioned [5, 6].

Ac ce p

te

d

92

Page 5 of 25

2 103

Our laboratory has measured myoglobin saturation [7-10] and hemoglobin saturation [11, 12] in vivo using Partial Least Squares (PLS) calibrated with spectra of solutions acquired in vitro. Myoglobin is

105

a protein found within muscle cells, and has one oxygen binding site. It functions as an intracellular store

106

and carrier for oxygen [13]. Our approach using PLS with full spectra in the visible and NIR regions

107

allowed accurate and separate measurements of myoglobin and hemoglobin saturation despite the fact

108

that the two proteins have very similar and highly overlapping absorbance spectra. PLS calibration on in

109

vitro spectra is not practical for clinical monitoring because PLS estimates are typically only valid for

110

subject spectra containing physiological interferences very similar to those in the calibration set [14]. It

111

is impractical and perhaps impossible to make appropriate in vitro calibration sets when attempting to

112

make single, instantaneous muscle oxygenation (Mox) estimates on subjects randomly selected from a

113

heterogeneous population.

M

an

us

cr

ip t

104

The goal of the present study was to develop a spectral analysis method to yield accurate, real-

115

time measurements of Mox that could potentially be used in a clinical monitor. Mox is a measure of the

116

percentage of oxygenated myoglobin and hemoglobin out of the total myoglobin and hemoglobin in the

117

optical path. Individual saturations of myoglobin and hemoglobin were not determined. Mox has

118

contributions from venous, capillary, and arterial blood and myoglobin.

Ac ce p

te

d

114

119

Full spectra were acquired between 540 and 800 nm, allowing measurement of both

120

oxygenated and deoxygenated forms of myoglobin and hemoglobin. This is in contrast to NIRS devices

121

that measure only deoxygenated hemoglobin in the NIR. Preprocessing included second derivatives,

122

variable selection, scaling between visible and NIR regions to balance information content between the

123

two regions, and total signal normalization. Multivariate Curve Resolution (MCR) was used to estimate

124

Mox in twenty healthy subjects. Those Mox estimates were then used in the Locally Weighted

125

Regression (LWR) training set. LWR provides a means to utilize the stability and simplicity of a linear

126

regression model in the presence of non-linear, difficult-to-model interferences [15]. In the case of

Page 6 of 25

3 transcutaneous Mox estimation, the various physiological variations from subject to subject (e.g.,

128

variations in skin tone and intervening fat layer thickness) are the nonlinear effects to be minimized. For

129

each measurement of a subject for which Mox is to be estimated, LWR automatically selects the most

130

similar subset of the training set and builds a "local" PLS model to estimate Mox. This provides a PLS

131

model that is more valid than one based on in vitro spectra because it is built from human subjects with

132

similar physical characteristics, notably skin tone and body mass index. The method was validated on

133

five additional healthy subjects.

us

cr

ip t

127

135

2. MATERIALS AND METHODS

136

2.1 EXPERIMENTAL

2.1.1. Optical spectroscopy. The optical system is composed of an imaging spectrometer (JY

M

137

an

134

Horiba iHR320), a thermoelectrically-cooled CCD camera (JY Horiba Synapse, 256x1024 format), a

139

custom-designed LED light source (Innovations in Optics, Inc.), and a custom-designed fiber optic probe

140

(Fiberoptic Systems, Inc.). The 300-groove grating in the spectrometer was centered at 670 nm and the

141

wavelength resolution was approximately 0.25 nm/pixel.

te

Ac ce p

142

d

138

The fiber optic probe is bifurcated, with one branch connected to the light source and the other

143

branch coupled to the spectrometer entrance slit. The common end of the probe was placed on the

144

back of the subject’s hand over the thenar muscle group and was held in place with an elastic bandage.

145

Three probes were used, each with a constant separation between source and detection fibers of 13, 14,

146

or 15 mm. The largest probe that yielded spectra with adequate signal-to-noise ratios was used in each

147

individual subject. Hand spectra were referenced to a spectrum collected from 1% Intralipid to calculate

148

an approximate absorbance response.

149 150

2.1.2. Human subject protocol. With approval from the University of Washington Institutional Review Board, healthy controls between the ages of 18 and 80 were recruited. After consent was

Page 7 of 25

4 obtained, subjects were interviewed to determine whether they had cardiovascular or pulmonary

152

disease, diabetes, or a history of deep vein thrombosis. Subjects were excluded from the study if they

153

had one or more of these conditions. Baseline blood pressures and pulse rates were recorded, in

154

addition to height and weight. Twenty-five subjects were studied, with body mass indices of 24.2 ± 3.0.

155

ip t

151

Spectra (540-800 nm) were acquired at a rate of either 0.33 or 0.5 Hz during a 3-min period of rest while each subject breathed room air. An oxygen mask was then placed over the face for 3 min, and

157

the subject breathed 100% O2. A blood pressure cuff was inflated on the upper arm to a pressure that

158

was 60 mmHg above the subject’s systolic blood pressure or to 200 mmHg, whichever was less. The cuff

159

was inflated for 15 min, and the oxygen mask was removed approximately two minutes into the period

160

of arm ischemia. A recovery period of 5 min followed to ensure normalization of blood perfusion to the

161

arm. Spectral acquisition was continuous throughout the entire protocol.

us

an

M

162

cr

156

Consecutive spectra were averaged to improve signal-to-noise ratios. When the acquisition rate was 0.33 Hz, three spectra were averaged together, producing a new rate of 0.11 Hz. When the

164

acquisition rate was 0.5 Hz, five spectra were averaged together for a new rate of 0.1 Hz.

te

d

163

Spectra collected from the hand at rest while breathing 100% O2 represented the most

166

oxygenated condition that we could practically produce in the subjects (Fig.1). We assumed that these

167

spectra represented Mox values of 100% but we have not verified with an independent measure that

168

myoglobin and hemoglobin in the hand were fully saturated with oxygen. Spectra collected at the end of

169

the 15-min period of arm ischemia were assumed to represent Mox values of 0% (Fig. 1). In healthy

170

subjects, it has been observed that 15 min of arm ischemia is sufficient to achieve complete

171

deoxygenation of skeletal muscle and blood in the hand because myoglobin and hemoglobin saturations

172

plateau to a minimum value [12]. Spectra acquired during the period of arm ischemia represent

173

intermediate Mox values between 0 and 100%.

174

2.2. ANALYSIS

Ac ce p

165

Page 8 of 25

5 175 176

All analyses were performed using PLS_Toolbox Version 6.2 (Eigenvector Research, Inc.) running in Matlab R2009b (Mathworks, Inc.). To help remove baseline and background signals due to scattering phenomena, second

178

derivatives with respect to wavelength were taken using the Savitzky-Golay algorithm with a 65-point

179

window and a second order polynomial [16]. The visible wavelength region was restricted to 562.2 –

180

601.7 nm and the NIR region was restricted to 744.7 – 802.2 nm. These wavelength regions were

181

selected because they encompass the absorbances of the oxygenated and deoxygenated species,

182

respectively. The two regions were scaled relative to each other to correct for differences in absorptivity

183

of the species in the visible and infrared regions of the spectra. The magnitude of this adjustment was

184

determined using the method described in section 2.2.2. The two discontinuous wavelength regions

185

were then concatenated. Finally, to correct for changes in illumination and path length, as well as other

186

total-signal influences, each spectrum was normalized to its total summed absorbance.

cr

us

an

M

2.2.1. Estimation of Mox Reference Values by MCR. As no accurate primary method exists to

d

187

ip t

177

determine the actual Mox in human subjects at any given point in time, MCR [17, 18] was used to

189

estimate Mox during ischemia. Separate two-factor MCR models were calculated for each subject to

190

recover estimates of the oxygenated and deoxygenated component spectra and corresponding

191

contributions of each species at each point in time.

Ac ce p

te

188

192

Each model used a closure constraint to assure that the summation of the two recovered

193

contribution profiles would always be 100%. Assuming that the spectra collected just prior to cuff

194

inflation during breathing of pure oxygen are close to 100% Mox, the five spectra at this point in every

195

run were constrained to give a 100%:0% ratio of oxygenated to deoxygenated species. Likewise, the last

196

five spectra in each run prior to cuff release were constrained to give a 0%:100% ratio. The average

197

spectra from each of these time periods were used as the initial guesses for the pure-component

198

spectra in the MCR. Because the closure constraint requires that the two contribution profiles sum to

Page 9 of 25

6 199

100%, the recovered oxygenated species contribution profile is equivalent to the estimated Mox for the

200

subject. An example Mox profile is shown in Fig. 2.

201

It is important to note that this procedure to recover the Mox profiles is only an approximation of the actual Mox, which may be influenced by optical variations (such as skin-probe boundary changes).

203

However, it is also assumed that when all the subjects' results are combined during the LWR modeling

204

later, these effects will be diminished by an averaging of the pool of all subjects.

cr

ip t

202

2.2.2. Correction of Absorptivity Effects. Because there is always an ambiguity in the total

206

volume of muscle interrogated, the total observed absorbance is also ambiguous. The absorbance

207

observed for the deoxygenated species, for example, may increase due to decreased Mox, or due to an

208

increase in total interrogated volume. To correct for these effects, each spectrum is normalized to the

209

total absolute value of the observed signal (the sum of the absolute value) after taking the second

210

derivative and before MCR. This has two side effects: First, it means relative absorbances of the

211

oxygenated and deoxygenated species can be used to calculate Mox. Second, it introduces an

212

unintended nonlinearity between calculated Mox and true Mox due to the difference in absorptivity

213

between the oxygenated and deoxygenated species, as described below. Note that this effect is not a

214

nonlinearity between observed absorbances at different wavelengths (which would lead to a higher

215

multivariate rank of the set of spectra), but one simply between the calculated and true Mox.

217 218

an

M

d

te

Ac ce p

216

us

205

The total absorbance summed across all wavelengths, Aobs, can be written as a summation of the absorbances of the oxygenated and deoxygenated species: Aobs = EoBCo + EdBCd

(1)

219

where B is the effective path length, Eo and Ed are the total absorptivities summed across all

220

wavelengths of the oxygenated and deoxygenated species, respectively, and Co and Cd are the

221

concentrations of oxygenated and deoxygenated species, respectively. If the two absorptivity profiles

222

had the same integrated areas, Aobs would only be influenced by changes in path length or total

Page 10 of 25

7 223

myoglobin + hemoglobin concentration (Co + Cd) and dividing each spectrum by Aobs would correct for

224

changes in those properties. However, because Eo and Ed are not equal, the total summed absorbance,

225

Aobs, will vary with Mox even in the absence of changes in path length or Co + Cd. Fig. 3 illustrates the problem that would occur with normalization if there were a 4:1 ratio

ip t

226

between Eo and Ed. The dotted lines represent "true" oxygenated and deoxygenated species

228

concentrations (in relative %) as Mox decreases linearly from left to right. The solid lines show the

229

concentrations that would be calculated from normalized spectra. At 50% true Mox, the observed Mox

230

would be 80%, reflecting the 4:1 contributions to Aobs from Eo and Ed when Co = Cd. At 100% Mox, the

231

summed absorbance is dominated by the oxygenated species absorbance (Aobs = EoBC), while at 0% Mox,

232

the summed absorbance is dominated by the deoxygenated species (Aobs = EdBC). Aobs at 100% Mox

233

would be four times greater than Aobs at 0% Mox. Thus, as Mox varies, Aobs is not a constant but is

234

changing as a proportion of the absorptivity of the oxygenated or deoxygenated species. The effect of

235

the non-constant Aobs is that observed Mox would be consistently overestimated in this example.

us

an

M

d

One approach to correct for this effect is to scale the visible region of the spectrum (dominated

te

236

cr

227

by the oxygenated species) relative to the NIR region (dominated by the deoxygenated species) by a

238

factor corresponding to the ratio between Eo and Ed. Estimating an appropriate scaling factor was

239

achieved by taking advantage of the fact that when there are no major path length effects (such as in

240

the highly-controlled ischemia trials), normalization has little effect except to introduce the nonlinearity.

241

Once the scaling factor is correct, normalization should have almost no effect on Mox values recovered

242

using MCR. By adjusting the NIR:Visible scaling factor and then examining Mox profiles recovered with

243

and without normalization, we identified a factor at which the profiles were not significantly changed by

244

normalization. This was considered the "optimal scaling factor" for the given subject. The 50% Mox point

245

was used as a reference to indicate how similar the profiles were with and without normalization.

Ac ce p

237

Page 11 of 25

8 246

Figs. 4A and 4B show an example of the change in Mox profile resulting from normalization

247

before a scaling factor was applied. In Figs. 4C and 4D, the 50% Mox points of unnormalized and

248

normalized data sets were identical when a scaling factor of 5.3 was used. This process was repeated for all 20 training subjects and the median optimal scaling factor of

ip t

249

4.7 was selected for all subsequent spectral adjustments. Each training subject's spectra were corrected

251

by the median optimal scaling factor and the Mox profiles were estimated using the MCR procedure

252

described earlier. Finding a single factor to use for all subjects provides for the most general correction

253

of absorptivity and possible optical effects without over-correcting for a single subject. It also provides a

254

means to analyze subjects for which the normalized and non-normalized results were too severely

255

different (probably due to actual path length or optical effects) and enables use of the models for

256

application to future clinical subjects for which we will have only a single point, and not an entire

257

ischemic profile.

us

an

M

2.2.3. Locally Weighted Regression Modeling. Given the estimated Mox values and the

d

258

cr

250

individual spectra for all subjects as a whole, an LWR model was created that used a two principal

260

component PCA model to select training spectra. For any given Mox estimate to be made, 200 training

261

spectra were selected and a one latent variable PLS model was calculated and used for predictions. An

262

attractive feature of such an empirical, inverse least squares PLS model is that having relatively few

263

constraints allows flexibility in correcting for interferences. In contrast, a classical least squares method

264

like MCR will have a greater number of constraints that will likely degrade prediction accuracy.

Ac ce p

265

te

259

The training spectra provided to the LWR model were pre-processed similarly to that described

266

for the MCR estimation procedure (2nd derivative, 4.7 NIR:Vis scaling, area normalization) but the

267

spectra and estimated Mox values were also mean-centered to allow for an intercept in the linear

268

regression. The number of principal components, number of training spectra selected, and number of

269

PLS latent variables to be used were all selected using cross-validation in which the entire set of spectra

Page 12 of 25

9 from a given subject were left out of the training set, and the remaining 19 subjects were used to

271

estimate the left-out subject's Mox. The error in these Mox estimations was summarized in the root

272

mean squared error of cross-validation (RMSECV) which is the square root of the average error of

273

prediction for all samples when they were left out of the training set [14]. This was repeated for various

274

settings of the LWR algorithm with general agreement that the selected LWR parameters gave stable

275

and reliable results.

cr

2.2.4. Partial Least Squares Modeling. For comparison to the LWR models, an identical

us

276

ip t

270

modeling effort was made using standard (non-locally-weighted) PLS modeling. For these models, the

278

identical calibration data were analyzed with a standard PLS algorithm using cross-validation to

279

determine the number of PLS latent variables. The best smallest cross-validation error was observed at 2

280

latent variables.

M

281

an

277

2.2.5. Quality of Spectral Fit Parameters. In both the PLS and LWR models, it is important to assess how well the model fits a given spectrum when making a Mox prediction for that spectrum. For

283

factor-based models such as PLS and LWR, this can be done using the Hotelling’s T2 and Q residuals [14].

284

Generically, these models decompose a data matrix, X (in which rows are spectra), into scores, T, and

285

loadings, P, and spectral residuals, R, according to the equation:

Ac ce p

te

d

282

286

X = TPT + R

287

Hotelling’s T2 is calculated as a weighted sum of the scores and describes whether the features

(2)

288

described by the loadings of the model are present in similar scale to what was observed in the

289

calibration data. The Q residuals are calculated as the sum of the squared residuals in each row of the R

290

matrix [14]. When making a prediction for a new sample, a new Hotelling’s T2 value and Q residual value

291

are calculated and compared to the values obtained for the calibration data. If a new value is

292

substantially different from the calibration data, it is inferred that the new spectrum contains some

Page 13 of 25

10 293

feature(s) which were not observed in the calibration data and, thus, the predictions for Mox should not

294

be used.

295

297

3. RESULTS AND DISCUSSION

ip t

296

3.1. Cross-Validation. While building the models, cross-validation was used as described earlier. For the LWR model, cross-validation indicated very little change as a function of either the number of

299

principal components (PCs) in the model or the number of samples selected for each local model.

300

Models were examined with 1 to 6 PCs, 100, 150, or 200 selected samples, and 1 to 3 latent variables in

301

the local PLS model. The RMSECV ranged from 6.5% at 2 PCs and 200 samples selected for each local

302

model up to 8.1% at 6 PCs and 100 samples selected. In comparison, the best standard PLS model (not

303

LWR-based model) provided an RMSECV of 9.3% obtained with 1 latent variable. All LWR and PLS

304

models with more than 1 latent variable gave RMSECVs above 9.3%. In all cases, the subjects in the

305

calibration set displayed Hotelling’s T2 and Q residuals that indicated there were no unusual subjects or

306

samples in the calibration set.

us

an

M

d

te

3.2. Validation Subject Results. In order to validate the model, Mox measurements were made

Ac ce p

307

cr

298

308

from five healthy subjects that were not included in the model. These subjects underwent the same

309

protocol as the 20 subjects included in the model. Fig. 5 shows that the LWR model produced accurate

310

Mox measurements compared to MCR estimates of Mox obtained from each subject’s data set. The

311

best-fit line had a slope of 0.99 and a y-intercept of -2.2, very similar to the line of unity. The root mean

312

square error (RMSE) for all 654 spectra was 5.4%.

313

For comparison, the results for the standard PLS model applied to the same five test subjects are

314

shown in Fig. 6. Because the PLS model is required to take into account all subjects equally, the model

315

must be a “consensus,” resulting in a per-subject bias. Three of the five subjects show notable

Page 14 of 25

11 316

systematic offset from the MCR estimates. The RMSE for the test subjects using the PLS model was

317

9.6%.

318

Finally, an additional LWR model was built with the wavelength range limited to only the NIR portion of the spectrum (744.7 – 802.2 nm). In this model, the predictions showed little to no trend with

320

Mox and produced an RMSE of 29.3% (Fig. 7). Such a poor result is expected considering that removing

321

the visible portion of the spectrum leaves no ability to correct for path length and absorbance effects. It

322

demonstrates the need for obtaining both the visible and NIR portions together for the Mox evaluation.

323

Interestingly, these poor results are probably overly optimistic considering that reference Mox values

324

were obtained from MCR using the full wavelength range. If the MCR step was similarly limited in

325

wavelength window, no sensible reference values could even be obtained for the calibration samples.

cr

us

an

3.3. Application to Trauma Patients. Mox was measured in a preliminary study of trauma

M

326

ip t

319

patients admitted to Harborview Medical Center in Seattle, WA. Spectra were acquired for 15 min as

328

soon as possible upon admission to the emergency department. The LWR model was applied to the

329

spectra from each patient. Fig. 8 shows the Mox time course in one patient. Mox has a good correlation

330

with blood pressure; it provides a continuous and real-time indication of tissue perfusion. The sharp

331

increases in Mox at approximately t = 2 and 8 min correlate with intravenous saline infusions given for

332

low blood pressure that temporarily improved tissue perfusion. In contrast, arterial blood saturation

333

measured by pulse oximetry was not sensitive to changes in tissue perfusion.

335 336

te

Ac ce p

334

d

327

4. CONCLUSIONS

The results for both the validation subjects and a trauma patient indicate that the use of joined

337

visible and NIR spectra can provide Mox estimates better than what has been previously reported for

338

single- or double-wavelength NIR methods. In addition, these results show that the LWR models are

339

capable of making reasonable estimates of Mox, even when presented with the limited range of

Page 15 of 25

12 340

physiological variability provided in this small calibration set. The accuracy of these predictions should

341

only improve when additional calibration subjects, representing a wider variety of skin tones and body

342

mass indices, are added to the calibration set. Further improvements in model performance would also be expected by using a modified

ip t

343

method for training sample selection. In this work, the sample-to-sample distance in the PCA score

345

space was used. This PCA model captured the variance associated with both the subject-to-subject

346

variability (which is most critical to correct) and the varying Mox caused by ischemia in individual

347

subjects. Thus, samples were selected both on the similarity of the subject as well as the similarity of the

348

extent of ischemia to test spectra. The predictions made from the ordinary PLS model show that the

349

calibration and test subjects were generally in error by a constant slope and/or offset across all Mox

350

values. From this, we would anticipate that a modified sample selection method which focuses more on

351

subject-to-subject variability (for example, discarding all spectra for individuals which are dissimilar and

352

keeping all spectra for individuals which are similar) would tend to create models that capture the

353

specific slope and offset scaling relevant for the tested individual. Such models may provide better

354

predictions than our current model.

us

an

M

d

te

Ac ce p

355

cr

344

This work also demonstrates that the use of MCR with simple equality and closure constraints

356

provides a reasonable inference of Mox in the absence of any true primary reference. This use of a

357

carefully controlled experimental protocol and inferential estimates of Mox can be used to easily

358

produce additional calibration data for the models.

359 360 361

ACKNOWLEDGMENTS This work was funded in part by grants from the Coulter Foundation, the Life Sciences Discovery

362

Fund and NIH 1R21EB009099. None of these granting agencies played any role in the study design or in

363

the collection, analysis or interpretation of the data.

Page 16 of 25

13 364 REFERENCES

366 367 368 369 370 371 372 373 374 375 376 377 378 379 380 381 382 383 384 385 386 387 388 389 390 391

[1] M. Smith, Philos Transact A Math Phys Eng Sci 369 (2011) 4452-4469. [2] J.M. Murkin, M. Arango, Br J Anaesth 103 Suppl 1 (2009) i3-13. [3] T.W. Scheeren, P. Schober, L.A. Schwarte, J Clin Monit Comput 26 (2012) 279-287. [4] B.A. Crookes, S.M. Cohn, S. Bloch, J. Amortegui, R. Manning, P. Li, M.S. Proctor, A. Hallal, L.H. Blackbourne, R. Benjamin, D. Soffer, F. Habib, C.I. Schulman, R. Duncan, K.G. Proctor, J Trauma 58 (2005) 806-813; discussion 813-806. [5] J.C. Hirsch, J.R. Charpie, R.G. Ohye, J.G. Gurney, Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu 13 (2010) 51-54. [6] D. Highton, C. Elwell, M. Smith, Curr Opin Anaesthesiol 23 (2010) 576-581. [7] G.A. Cohen, L.C. Permut, L.S. Arakaki, W.A. Ciesielski, D.M. McMullan, A.R. Parrish, K.A. Schenkman, ASAIO J 57 (2011) 314-317. [8] K.A. Schenkman, D.R. Marble, D.H. Burns, E.O. Feigl, Appl Spectrosc 53 (1999) 332-338. [9] L.S.L. Arakaki, W.A. Ciesielski, B.D. Thackray, E.O. Feigl, K.A. Schenkman, Appl Spectrosc 64 (2010) 973-979. [10] L.S.L. Arakaki, D.H. Burns, M.J. Kushmerick, Appl Spectrosc 61 (2007) 978-985. [11] D.J. Marcinek, W.A. Ciesielski, K.E. Conley, K.A. Schenkman, Am J Physiol Heart Circ Physiol 285 (2003) H1900-H1908. [12] C.E. Amara, D.J. Marcinek, E.G. Shankland, K.A. Schenkman, L.S. Arakaki, K.E. Conley, Methods 46 (2008) 312-318. [13] J.B. Wittenberg, B.A. Wittenberg, J Exp Biol 206 (2003) 2011-2020. [14] H. Martens, T. Naes, Multivariate Calibration, John Wiley & Sons, Chichester, 1989. [15] T. Naes, T. Isaksson, B. Kowalski, Anal Chem 62 (1990) 664-673. [16] A. Savitzky, M.J.E. Golay, Anal Chem 36 (1964) 1627. [17] M.H. Van Benthem, M.R. Keenan, D.M. Haaland, J Chemom 16 (2002) 613-622. [18] P.J. Gemperline, E. Cash, Anal Chem 75 (2003) 4236-4243.

cr

us

an

M

d

te

Ac ce p

392

ip t

365

Page 17 of 25

14

us

cr

ip t

392

an

393

Figure 1. Spectra acquired from a healthy human subject at extremes of muscle oxygenation. The solid

395

spectrum was acquired after 3 min of 100% O2 breathing. The dashed spectrum was acquired after 15

396

min of arm ischemia. Oxymyoglobin and oxyhemoglobin absorb near 580 nm in the visible region;

397

deoxymyoglobin and deoxyhemoglobin absorb near 760 nm in the near infrared region. The visible and

398

NIR bands that were used to estimate Mox are indicated by the sets of vertical dashed lines. In the

399

wavelength region between 610 and 730 nm, the detector was saturated. These spectral values do not

400

represent measurements of tissue absorbance and are not used in the spectral analysis.

d

te

Ac ce p

401

M

394

Page 18 of 25

15 401

an

us

cr

ip t

402

403

Figure 2. Muscle oxygenation (Mox) in a healthy human subject estimated by MCR. The subject

405

breathed air for 3 min, then 100% O2 for 3 min, and then underwent a 15-min period of arm ischemia.

M

404

Ac ce p

te

d

406

Page 19 of 25

16 406

an

us

cr

ip t

407

408

Figure 3. Normalization bias effect. Since the integrated areas of absorptivity profiles of oxygenated and

410

deoxygenated species are not equal, normalization introduces an unintended nonlinearity to recovered

411

relative oxygenated and deoxygenated concentrations. Dotted lines represent true relative

412

concentrations and by definition, true Mox is equal to “True Oxygenated.” True Mox decreases from left

413

to right, as during ischemia. If the ratio of the integrated areas of oxygenated to deoxygenated

414

absorptivity profiles were 4:1, the observed relative concentrations would be those shown by the solid

415

lines. The observed Mox, equal to “Observed Oxygenated,” would be consistently overestimated.

d

te

Ac ce p

416

M

409

Page 20 of 25

17 416 417 418

ip t

419 420

cr

421

Ac ce p

te

d

M

an

us

422 423 424 425 426

427 428 429 430 431 432 433 434 435 436 437 438 439 440

Page 21 of 25

18 Figure 4. Scaling correction for absorptivity effects. A) MCR results without (left) and with (right)

442

normalization show that the 50% Mox point is affected by normalization. B) Concatenated second-

443

derivative spectra. The visible region comprises 562-601 nm and the NIR region comprises 744-802 nm.

444

C) Up-weighting the NIR region where deoxygenated species have predominant intensity corrects for

445

the normalization effect. This example shows results when the NIR region was up-scaled by a factor of

446

5.3. The optimal scaling factor was determined by adjusting the scaling until the 50% Mox points for

447

unnormalized (left) and normalized (right) cases were equal. D) Concatenated second-derivative spectra

448

with the scaling, without and with normalization.

us

cr

ip t

441

Ac ce p

te

d

M

an

449

Page 22 of 25

19

us

cr

ip t

449

an

450

Figure 5. Accuracy of muscle oxygenation (Mox) measurements by LWR in five healthy human subjects.

452

Mox was measured accurately across its range from 0% to 100%.

M

451

Ac ce p

455

te

454

d

453

456 457 458 459 460 461 462 463

464

Figure 6. Accuracy of muscle oxygenation (Mox) measurements by PLS in five healthy human subjects.

465

Most subjects show a systematic bias at all Mox levels.

Page 23 of 25

us

cr

ip t

20

466

Figure 7. Muscle oxygenation (Mox) measurements by LWR in five healthy human subjects when only

468

the NIR portion of the spectra is used. Accuracy is very poor. The advantage of including the visible

469

region in the analysis is apparent when these results are compared to those in Figure 5.

M

an

467

470

d

471

Ac ce p

te

472 473 474 475 476 477 478 479

480

Figure 8. Muscle oxygenation (Mox) is indicative of tissue perfusion. Mox was measured in a trauma

481

patient who received saline infusions at t = 2 and 8 min (indicated by arrows), causing increases in blood

482

pressure (BP) and corresponding increases in Mox. Arterial saturation measured by pulse oximetry was

483

not sensitive to changes in tissue perfusion.