NUCLEAR INSTRUMENTS
AND M E T H O D S 72 (1969)
254-268; ©
NORTH-HOLLAND
PUBLISHING
CO.
N E U T R O N C A P T U R E G A M M A RAY~FACILITY USING AN E L E C T R O N L I N E A R A C C E L E R A T O R * V. J. ORPHAN, C. G. HOOT, A. D. CARLSON, JOSEPH JOHN and J. R. BEYSTER
Gulf General Atomic Incorporated, San Diego, California 92112, U.S.A. Received 14 March 1969 A versatile facility for the study of resonance neutron capture gamma-ray spectra with high energy resolution is described. An electron linear accelerator is used to provide a pulsed beam of neutrons. Gamma-ray spectra are measured using a Ge(Li)NaI(TI) spectrometer. The gamma-ray pulse height and the
neutron time-of-flight information are stored using an on-line computer. The salient features of the capture facility are described and spectra resulting from the capture of neutrons in natural tungsten are displayed.
1. Introduction
parities of nuclear states and average widths of transitions from many initial states to individual final states can be obtained from an analysis of the average spectra. A versatile facility has, therefore, been constructed at Gulf General Atomic Inc., to study resonance neutron capture gamma-ray spectra. It is possible to study the gamma-ray spectra from the capture of neutrons of any set of energy ranges. These ranges, chosen at the time of analysis, may scan just one resonance or several resonances, may correspond to a small region between two resonances, or cover the entire range from thermal energy to a few hundred keV. An electron linear accelerator is used to produce a pulsed beam of neutrons. The capture sample is located at the end of a 16-metre long evacuated flight path. G a m m a - r a y spectra are measured using a coaxial Ge(Li) detector surrounded by a NaI(TI) annulus. G a m m a - r a y spectra resulting from each of many distinct capturing states are observed in the same timeof-flight experiment. This facility combines the high neutron energy resolution resulting from small burst widths and the high gamma-ray energy resolution from the use of the Ge(Li) spectrometer. This paper describes the principal features of this facility. Operating characteristics of the linear accelerator neutron source and the Ge(Li) spectrometer are summarized. The spectra resulting from the resonance capture of neutrons in natural tungsten are displayed. The performance of this system is compared with that of a fast chopper assemblyS).
An investigation of the g a m m a radiation emitted following the capture of neutrons provides information concerning the states of the resultant nucleus over a wide range of excitation energy. Radiative capture of thermal neutrons has been investigated in detail for several years. Only recently have efforts been directed toward the study of resonance neutron capture spectra. Radiative capture is usually a very complex process and detailed study of the resulting gamma-ray spectra is interesting from several points of view. It is desirable to determine the distribution of the partial radiation widths F~j. for transitions to the final state j from a series of initial states ,,, of a particular spin and parity formed by the resonance capture of neutrons. Some such measurements made in the past are in good agreement with the Porter-Thomas 1) description of the distribution of the widths associated with a single exit channel. Several other measurements 2) suggest that the partial widths have a distribution corresponding to more than one degree of freedom. The neutron-capture mechanism also requires detailed investigation. The existence of narrow resonances in the low-energy capture cross section suggests a compound nuclear process in which a complex unbound state is formed. However, the study of the intensities of high energy g a m m a rays emitted from thermal neutron capture led to the description of the capture mechanism as a simple direct 3-5) type in which the capture state consists of a neutron coupled to the target nucleus. The emitted g a m m a ray results from a single particle transition made by the captured neutron. The mechanism of the (n,'/) reaction can be investigated 6) by examining the interference effects between resonance and off-resonance amplitudes and the partial width correlations. Spectroscopic information can also be obtained by studying the average 3:-ray spectra 7) from the capture of neutrons over a wide energy range. Spins and 254
2. The neutron source The neutron source for this facility is the Gulf General Atomic Electron Linear Accelerator. In this application the Linac typically provides an electron * Supported by the U.S. Atomic Energy Commission and performed under subcontract No. 3032 with Union Carbide Corporation.
NEUTRON CAPTURE GAMMA RAY FACILITY W.
7. 7 5 0 D I A M
\ ,< \ \ \
DEPLETED URANIUM
TITANIUM WINDOW
\
i
6.00 BEAMELECTRON. . . . . . . . . . . . .L.--~ .. ~__
255
CAN
, T R-COO,,NG
\
\ \
AT ER-COOLE[ ~NSTEEL FANSTEEL CONVERTER 3NVERTER "
H------~-------
/ALUMINUM
\ \ \
3 O0
!!
\\\
6 DIAM
~ "
\ \ \ \ \ \ \
THERMOULD
Fig. 1. The electron target assembly. The water-cooled fansteel converter, the depleted uranium cylinder and the titanium entrance window are shown. All dimensions are in inch.
LINAC TARGET NEUTRON SOURCE
ECTOR ;TAL
F---3
URANIUH
W
LEAD
CONCRETE
LI2CO 3 IN LI2CO 3 IN H3BO 3 IN PARAFFIN
EPOXY
EPOXY
rULIclnTL~MC
DIRT EXPERIMENT LAYOUT
Fig. 2. Schematic drawing of the capture gamma-ray facility. Details of the evacuated flight path, the beam-defining collimators and the shielding arrangement are shown. To the right is the Ge(Li)-Nal(T1) spectrometer with its shield and the movable carriage.
256
v . J . ORPHAN et al. 10 6
l
I
I
I
I
--
I
~ - - I
l
I
I
I
I
I
[
t
BF3
t ~t~
I
I
FLUX MEASUREMENT
&E/E
,,j
=0.2
10 5
:::L
10 4 o=
E
X d LI.,,
10 5
i0 2 I0 -J
!
I
i
I
100
J
J
i
t
I
t
101
i
I
t
I
i0 2
t
t
r
K I
103
r
~
i
t
i0 4
NEUTRON ENERGY [eV)
Fig. 3. T h e relative n e u t r o n flux m e a s u r e d with a BF3 counter. T h e flux, as a function o f flight-time, is plotted against n e u t r o n energy. These m e a s u r e m e n t s were m a d e with the Linac operating at reduced power. T h e data points are averages over a r a n g e A E . T h e rapid drop below 0.8 eV is the effect of the c a d m i u m shutter in the flight path.
beam of 40 MeV to 60 MeV, the pulse width being adjustable from 3 nanosecond to 4.5/~sec. It has been operated for extended periods with repetition rates as high as 600 pps. The Linac target consists of a water-cooled fansteel (a tungsten alloy) converter surrounded by a 15.3-cm dia. right circular cylinder of depleted uranium. The electron target assembly is illustrated in fig. 1. The uranium cylinder is encapsulated in an aluminium can to provide containment of radio-active material. A thin titanium window is provided on the aluminium can for the entry of the electron beam. The electrons slowing down in the fansteel target produce bremsstrahlung which in turn produces high energy neutrons, principally through the (?,n) reaction, with a broad maximum near 1 MeV. These neutrons are slowed down to the epithermal and thermal region by a 2.54cm thick piece of polyethylene, shaped to fit around the side of the target facing the neutron flight path. Water surrounding the aluminium can provides
cooling for the uranium and contributes additional moderation. This target arrangement is chosen because it gives a favourable neutron-to-gamma-flash ratio without seriously sacrificing the neutron yield. The neutron beam traverses an evacuated flight path 16 metre long, at the end of which is located the capture sample. A plan view of this arrangement is shown in fig. 2. Shielding employed around the neutron flight path to minimise the background at the spectrometer is also displayed in this figure. Five irises are located along the flight path to prevent neutrons from scattering off the wall of the flight tube into the capture sample. Each iris consists of a 30.5-cm thickness of boric acid in epoxy and 10.2 cm of lead. The beam defining collimator is positioned at the terminus of the flight path and allows a 15.3-cm diameter capture sample to view all of the Linac target. This collimator consists of 33.0 cm of LizCO3 in epoxy and 10.2 cm of lead. A cadmium shutter is provided at 14 m fi'om the neutron source. This shutter can be closed
NEUTRON
500
I
CAPTURE
I
GAMMA
RAY
I
I
257
FACILITY
I
I
1
600
700
OOUBLE- ESCAPE PEAKS 6610 key 6624 keV
400
30C z
o
2OO
7279 keV SINGLE -ESCAPE PEAKS 712I keV
I00
4 •
O~
7135 keV
-.-;.
I
I00
-
.
-.
L
200
Z00
400 C H A N N E L NUMBER
500
Fig. 4. The high energy portion of the gamma-ray spectrum from the capture of thermal neutrons by iron. This spectrum is obtained with the Ge(Li) detector operated in the singles mode. The 14-keV doublet is partially resolved. to eliminate neutron pulse overlap at higher repetition rates. The energy o f neutrons is d e t e r m i n e d by the timeof-flight technique. The flight-time d e p e n d e n c e o f the neutron flux incident on the c a p t u r e sample is measured by placing a 2.5-cm dia., 12.7-cm long BF 3 c o u n t e r (20 'h C e n t u r y Electronics M o d e l 5EB70) at the exit point o f the c o l l i m a t e d neutron beam. A l/v detection efficiency is assumed for the BF 3 counter. The d a t a o b t a i n e d using this c o u n t e r are corrected for backg r o u n d , self-shielding effects o f the BF 3 c o u n t e r and for the loss o f counts due to analyzer dead time. If c(t) is the corrected counting rate from the BF 3 c o u n t e r as a function o f the neutron flight time, then the neutron flux go(t) is given by
go(,) = kc(t)
[e(t)]t
(1)
where k is a c o n s t a n t and E(t) is the energy corres p o n d i n g to the flight time t. Fig. 3 is a plot o f log{ c(t)[ E ( t ) ] ~ }, as a function o f log E(t) derived from the BF 3 c o u n t e r data. The linear b e h a v i o u r of the log-log plot in the range l eV to 1000 eV implies that
go(t) = / q - E ( t ) y ,
(2)
in this energy range. The e x p o n e n t ~ is found from the slope o f the curve and the c o n s t a n t K d e t e r m i n e s the absolute value o f the flux, g0(t). F o r the present arrangement o f the target and the m o d e r a t o r , the value o f = 0.585 _+ 0.003. The neutron flux as a function o f the neutron energy, go(E), is given by go(E) = go(t)dt/dE = k'[E(t)] ~- ~,
(3)
and d r o p s off a p p r o x i m a t e l y as l/E for the a b o v e value o f c~. BF 3 m e a s u r e m e n t s described a b o v e are not employed at 16 m to measure neutron flux for energies greater than 1000 eV because o f the finite time taken for the detection system to recover from the g a m m a flash. However, m e a s u r e m e n t s °) m a d e at 200 m verify that the expressions (2) and (3) for relative flux shape can be extended to a b o u t 100 keV neutron energy. The relative flux shape is converted to a b s o l u t e values by measuring the counting rate in an energy region in which all the incident neutrons are c a p t u r e d . A convenient m e t h o d is to use a " b l a c k " I°B abs o r b e r and measure the yield o f the 478 keV g a m m a
v . J . O R P H A N et al.
258 250(
1592 keY I DOUBLE- ESCAPE PEAK
+
I
h C"
I SINGLES SPECTRUM
200(: 2614 keY FULL-ENERGY PEAK 150(
100(
-
ti
2103 keY SINGLE-ESCAPE PEAK 1620 keY (2,2 PO) 1~1FULL ENERGY PEAK
50(:
.. .
d 7=
c
,] .. /~. ~,.,,w....-~.. j • :"W,: "; :
1FULL ENERGY PEAK~1 I
I
I
-
*
I
COMPTON SUPPRESSION SPECTRUM
2oo0
8
I
2614 keV FULL ENERGY PEAK 150(2
IOOC
1592 keV 504 |
DOUBLE-ESCAPE PEAKx
I
1620 keV(ZlZPo)
",, /FULL-ENERGY PEAK
2,03 keY /SINGLE ESCAPE
/PEAK f
~1 |l
JT ,~.. PAIR SPECTRUM
IOOC 1592 keY DOUBLE-ESCAPE PEAK
L~O0
~' 250
-
, 300
35O CHANNEL NUMBER
,i 450
.;,..,u
J_ 550
Fig. 5. Comparison of the response of the Ge(Li)-NaI(TI) spectrometer in the singles (top), Compton suppression (middle) and pair (bottom) mode to the 2614 keV gamma ray from ThC". The three spectra are accumulated simultaneously. Full-energy, single-escape and double-escape peaks are labelled in the singles spectrum. Reduction in the intensities of the single-escape and double-escape peaks are seen in the Compton suppression spectrum. A single line response is shown in the pair spectrum.
NEUTRON
CAPTURE
GAMMA
rays from the I°B(n,c07Li * reaction as a function of the neutron time-of-flight. From the known efficiency o f the g a m m a - r a y spectrometer for the 478 keV g a m m a ray, the area of the I°B sample and the fraction o f 478 keV g a m m a rays emitted for every neutron captured in I°B, the absolute flux can be easily calculated.
RAY
259
FACILITY
The Ge(Li) detector is mounted in a "dip-stick" cryostatt at the end of a 38.2 cm long horizontal copper cold finger inside a 5.7-cm dia. aluminium cap. The spectrometer may be simultaneously operated in the following three modes and the data stored in the CDC-1700 on-line computer: 1. Singly-operated Ge(Li) detector;
3. The Ge(Li)-NaI(TI) spectrometer The gamma-ray spectrometer consists o f a 30-cm 3 Ge(Li) detector located at the centre of a large annulus of NaI(TI) split into two optically decoupled halves. This annulus has 6.3 cm i.d., 20.3 cm o.d. and is 30.5 cm long. Three R C A 8053 photomultiplier tubes are optically coupled to each half o f the NaI(TI) annulus. /
COAXIAL GE(Lt)
(
3
2. C o m p t o n suppression spectrometer; 3. Three-crystal pair spectrometer. Each event is suitably "tagged" before storage in the computer to identify the mode in which the event is detected. t ManufacturedbyNuclear Diodes, Prairie View, Illinois.
NOI
8 IN o,d, X 12 IN SPLIT ANNULUS PM TUBES ON EACH H~,LF)
I I
TENNELEC TC-200 AMPL FIER
~1~
I C
I 1417 AMPLIFIER
TIMING SCA I INTEGRAL DISCRIMINATOR
I
I TIMING SEA J 511 keV WINDOW DISCRIMINATOR
ORTEC
260 ~ O R T E C 40:~A TIME PICKOFF ] - - I CONTROL UNIT I
C.I. 1417 AMPLIF ER
TIMING SCA J 511 keY WINDOW D SCRIMIN ATOR
!
TIMING SCA 1 INTEGRAL DISCRIMINATOR
1
1
ORTEC 414A TRIPLE FAST COINCIOENCEJ - -
LIGATEANDDELAY GENERATOR
J__ OVERLAP COINCIDENCE | ~
I
GATE AND DELAY GENERATOR
ANDDELAYl GENERATOR JGATE
BIT 13 BIT 14 I __ CDC-170O INTERFACE UNIT I ~
ICDC-WOO0N-LINECOMPUTER
TMC 217A ANALOGITO-DIGITAL CONVERTER
J
I
Fig. 6. Block diagram of the electronics for the operation of the Ge(Li)-NaI(Tl) spectrometer. The triple fast coincidence circuit picks out the pair events and the overlap coincidence circuit helps to identify the Compton suppression events.
260
v . J . ORPHAN et al. l
I
I
1
No 22 0
i0 -5
t~
~co6O o6°~
q_
~
t;0
N
o
22
C
LP Z
2" Ld t~ EL Ld <:{
NO2 4
Q~ uJ Z I
i0 -4
0
I 600
I t200
I t8 0 0 E?.
I 2400
3000
(keY)
Fig. 7. Full-energy peak efficiency of the spectrometer operated in the Compton suppression mode. The effects of the relative solid angle and the attenuation of lead and ~LiH are included.
The spectrometer assembly is mounted on a movable carriage. This allows the distance between the spectrometer and the capture sample to be varied. It also facilitates the rotation of the spectrometer assembly about the centre of the sample. An annular shield consisting of a 5.1-cm layer of lead surrounded by a 5.1-cm thick layer of LizCO3-epoxy mixture protects the NaI(TI) on all sides except the backface. A conical gamma-ray collimator is centred along the axis of the NaI(T1) annulus in the front face of the shield. Details of this arrangement are contained in fig. 2. 3.1. THE Ge(Li) DETECTOR A 30-cm 3 coaxial Ge(Li) crystal, fabricated in our laboratory and similar to a detector described by Orphan and Rasmussen1°), is used as the primary gamma-ray detector. The response of the Ge(Li) detector to low energy gamma rays is studied using several radio-active sources. Its response to higher energy g a m m a rays is investigated using the 4432 keV line from a Pu-Be source and the lines resulting from the
radiative capture of thermal neutrons in chlorine, iron, and mercury. Using optimum time constants in the pulse-shaping amplifier, the average energy resolution of the Ge(Li) detector is 3.3 keV (fwhm) for the 122 keV and 136 keV lines from 57Co and 5.3 keV (fwhm) for the 2754 keV g a m m a ray from ZgNa. During actual data taking using the Linac, it is necessary to shorten the time constants to insure fast recovery from the g a m m a flash. This worsens the over-all resolution of the detector somewhat. For gamma rays of energies above 7 MeV a resolution of 8 keV (fwhm) is typical for data obtained with the Linac as is demonstrated by the partial resolution of the 14 keV doublet at 7632 and 7646 keV in the capture of thermal neutrons by iron. This is illustrated in fig. 4. The response of the Ge(Li) detector to the 2614 keV g a m m a ray from ThC" is shown in the top portion of fig. 5. The full-energy peak, the single-escape peak and the double-escape peak are identified. The energy resolution is about 7 keV (fwhm). The ratio of the
NEUTRON
CAPTURE
GAMMA
RAY
261
FACILITY
10 -5 1
>(D
--
l
- -
~-
1
- - - - J
~
1
I0 4
Z LO u_ b_ W
tx/ 0 b--
0
CI
•
NO 24 ~ A D I O A C / I V E
THERMAL
CAPTUIRE
GAMMA
RAYS
DECAY GAMMA
ok
10
10-6
I
I
I
2
5
4
1
5 E ¥ (MeV)
I
I
I
6
7
8
Fig. 8. Efficiency o f the pair spectrometer to high energy gamma rays.
height of lhe full-energy peak to the Compton edge is about 2.9. 3.2. Ge(Li)-Nal(T1)
S P E C T R O M E T E R IN T H E C O M P T O N
S U P P R E S S I O N MODE
The operation of the spectrometer in the Compton suppression mode is illustrated by the block diagram of the electronics given in fig. 6. The signals from the three photomultiplier tubes associated with each half of the NaI(TI) annulus are amplified and shaped into a bipolar pulse and passed through a timing single channel analyzer operating as an integral discriminator. Every gamma-ray event in the Nal(TI) above a bias set at 100 keV generates a timing pulse. When a timing pulse from either half of the NaI(TI) annulus is in coincidence with a pulse from the Ge(Li) detector, the linear signal from the Ge(Li) is "tagged". All Compton scattering events occurring in the Ge(Li) detector in which the scattered gamma rays are detected by the NaI(TI) annulus are rejected from the Compton suppression spectrum, The central spectrum in fig. 5 is
measured in the Compton suppression mode. It illustrates the improved peak-to-background ratio ill comparison with the spectrum measured in the singles mode shown at the top of the same figure. Note that the average Compton background is reduced by about a factor of 5.3. At the Compton edge, however, the reduction is only about 4.0. The reduced effectiveness in the region near the Compton edge is the result of large-angle scattering events in which the scattered photon passes through little or no Nal. The relative intensity of the single- and double-escape peaks is significantly reduced in comparison to the full-energy peak. The ratio of the full energy peak to the average Compton background is about 20: 1. The ratio of the intensities of the full energy peak to the double escape peak is improved by a factor of 11.5. The full energy peak efficiency of the spectrometer in the Compton suppression mode, measured over the energy range 500 to 2754 keV using calibrated radioactive sources, is shown in fig. 7. Below 500 keV the exact shape of the efficiency curve is not determined,
262
v . J . ORPHAN et al.
This curve includes the attenuation of 1.6 mm of lead and two 2.5 cm pieces of °LiH in 1 mm thick aluminium cans placed in the gamma-ray collimator. The lead greatly reduces the intensity of the gamma flash scattered from the capture sample and the 6LiH helps shield the detectors from sample-scattered neutrons. 3.3. THREE CRYSTAL PA1R SPECTROMETER The logic used for the operation of the pair spectrometer is also shown in fig. 6. The timing single channe! analyzers used in this segment of the circuitry are used as differential discriminators. Windows are set on the 511-keV peaks that result from the absorption of annihilation radiation in the Nal crystals. A crossover pick-off circuit is used in each Nal channel to generate a timing logic pulse whenever a pulse is within the 511-keV window. The timing signal from the Ge(Li) detector is obtained from a time pick-off unit. When the two timing pulses from the NaI halves and the timing pulse from the Ge(Li) detector are in
triple coincidence, the linear signal from the Ge(Li) detector is "tagged" as a pair event prior to storage in the on-line computer. This arrangement yields spectra in which only double-escape peaks appear; full energy and single escape peaks and the Compton background are rejected. The performance of the spectrometer in the pair mode can be seen in the bottom portion of fig. 5 which shows its response to a ThC" source. A single line corresponding to the double-escape peak of the 2614 keV g a m m a ray is seen. Except for a small tail on either side of the pair peak, the background is almost completely eliminated. The three spectra contained in fig. 5 are measured simultaneously using the on-line computer. A comparison of the peak in the pair spectrum and the double-escape peak in the singles spectrum denotes a loss of efficiency in the pair spectrometer by about a factor of 3.1. The efficiency of the pair spectrometer for g a m m a rays between 1370 keV and 8500 keV is shown in
[~
Ge (Li) DETECTOR
TEST 1 TC-135 L_~ 260 / ]-NPUT "~-[PREAMPLIFIERI --[TTIME P1CKOFF]
LINAC PRE- NJECTOR TRIGGER ;
MAIN GAIN ] STAGE [ TC-200 / IAMPLIFIERI
I CONTROLI I UNIT I
_L~ -~
GATE AND I ~ DELAY I GATE I, TNF£RI GENERAT 0 RFE~-~ O~~ :~'~E:L ]
[--TWO-FOLD I O V E R L A P t-[COINCIDENCE
LINAC PRE-INJECTOR TRIGGER
AND F-GATE DELAY GENERATOR #2
SG.EMAT C
GA ,NG
TIME SEQUENCE
'NTI 2 ION
i
TC- 200 AMPLIFIER
GATE AND DELAY I [ GENERATOR
),- FLASH|GATE # I
LINAC PRE- INJECTOR TRIGGER STOP
T
E (p. sec]
[
{ START
EL DORADO TIME jCOINCIDENCEINPUT ANALYZER
PHA
COMPLETE CONVERSION INTERUPT L
_ _
t GATE AND /] BIT r5 PULSER TAG DELAY GENERATOR/
-- I CDC- 1700 ON - LINE [ COMPUTER
t
1.25 MILLION WORD MAGNET C DISK UN YJ Fig. 9. Block diagram of the electronics required for two parameter data acquisition. The electronics associated with the operation of the spectrometer are not shown. The time sequence, where the system is made inoperative during the gamma flash and a short period thereafter, and turned on for the required duration when data is accumulated, is illustrated.
NEUTRON
CAPTURE
fig. 8. The efficiency curve is determined from the known j ~) intensities of g a m m a rays from the capture of thermal neutrons in chlorine and from the two g a m m a rays from a calibrated Z4Na source. A sample of CC14, 15.2 cm dia. and 5 cm thick placed at 45" to the beam axis, is used as the source of capture g a m m a rays for these efficiency measurements. The uncertainty in the measured values of the efficiency is about + 10°;,.
4. Two-parameter data acquisition and sorting The gamma-ray energy and the corresponding neutron-time-of-flight are digitized and stored sequentially on the magnetic disc of the CDC-1700 on-line computer. Hence for every event detected in the Ge(Li) spectrometer, two computer words are written on the disc, the first word contains the pulse height information and the second word makes up the time-offlight information. In addition to the pulse height information, the first word contains two bits, one of which identifies events detected in the pair mode (bit 14) and the lother labels events in the Compton suppression mode (bit 13) of the gamma-ray spectrometer. At regular intervals, or at the end of an experimental period, data are transferred on to magnetic tapes for subsequent analysis. A block diagram of the two parameter data acquisition system is given in fig. 9. Only the electronics directly connected with data acquisition are shown in this figure. For simplicity, the electronics associated with the operation of the Ge(Li)-NaI(TI) spectrometer, described in sections 3.2 and 3.3, are omitted. For every gamma-ray event in the Ge(Li) detector two pulses are derived: a linear pulse (gamma-ray energy) and a timing pulse (neutron time-of-flight). The gamma-ray energy signal passes through a linear gate that is normally open. This gate is closed, by gate generator labelled # 1, for about 4 l~sec during the burst of electrons and a short period thereafter to prevent the gamma-flash pulses from being analyzed. A second gate generator (labelled # 2 ) is set to cover the time interval after a Linac pulse corresponding to neutron energies of interest. Those timing pulses that are in coincidence with this gate pulse are used to enable the pulse-height analyzer. In this way a one-to-one correspondence between the pulse-height and the time-of-flight variables is ensured. An El Dorado time analyzer, which has channel widths adjustable from 5 nsec to 80 nsec, is used to measure the time interval between the pre-injector pulse, used as a reference, and a timing pulse from a gamma-ray event. The pre-injector pulse occurs about 2/~sec before the
GAMMA RAY FACILITY
263
actual Linac pulse. The timing of the Linac pulse is determined by observing the time duration between the trigger pulse and the arrival of the gamma-flash at the detector. 4.1. DATA ACQUISITION CODE
The data acquisition code contains a digital gain and zero stabilization routine which corrects the incoming data for any electronic gain or zero-intercept shift. Two very stable pulser peaks, one at either end of the spectrum, are used as reference peaks. The computer code recalibrates the energy scale after a preset number of counts have been accumulated in the reference pulser peaks and adjusts the incoming pulse height data. The smallest incremental shift currently used is 0.25 channel out of the spectral range of 4096 channels. Since the test pulse is injected at the pre-amplifier, the reference peaks are sensitive to gain shifts throughout the system of electronics. The stabilizing routine can be interrogated at any time during a run to determine the amount of gain and zero shift correction that the programme is currently applying to the data. Use of the stabilization routine has permitted the successful addition of four sets of spectra accumulated over a onemonth period without any serious energy shift or loss of energy resolution. A pulse in coincidence with each reference pulse, is used to turn bit 15 on to identify these reference pulses so that the data-accumulation programme may exclude them from the gamma-ray pulse-height spectra. The data acquisition programme allows singles, pair and Compton suppression events to be accumulated simultaneously. External logic circuitry (fig. 6) is used to set bits 13 and 14 to identify Compton suppression and pair events respectively. This code classifies all events as singles data, events having bit 14 set to 1 as pair data and those having bit 13 set to 0 as Compton suppression data. In addition, different thresholds can be pre-set for separate modes of detection, the particular event being registered only if the pulse height falls above the corresponding limit. The code also generates and stores in the computer memory a time-of-flight spectrum (corresponding to all pulse-heights) and a pulse-height spectrum (for all neutron energies). Either of these spectra is available for live display during the experiment. 4.2. THE DATA SORTING PROGRAMME The data are sorted off-line using the Univac 1108 computer located at the central computing facility. The main function of the Fortran IV Programme SORTER is to generate gamma-ray pulse-height
264
v.J. ORPHAN et al.
40,000
200 I
I00 I
0
300 I
NEUTRON FLIGHT TIME (t~sec) 400 500 I I
600 i
700 I
800 I
900 I
TUNGSTEN
35,000
30,000
25,000 < 20,000
g 15,000
I0,000
1
X I
5,000 f L
0o
I00
200
..... 3~
4~
5~
600
700
i1 800
CHANNEL NUMBE~
Fig. 10. Time-of-flightspectrum using natural tungsten sample. This spectrum is measured in the singles mode of tile spectrometer. The low energy resonances are clearly resolved. spectra resulting from the capture of neutrons of energies between two specified limits. The programme can handle 10 neutron-energy limits at a time. For a given neutron energy interval, the programme generates three pulse height spectra, one spectrum corresponding to each operating mode of the spectrometer. In addition, three time-of-flight spectra are also developed, one spectrum corresponding to each mode of detection of the gamma ray. The two 16-bit words from the on-line computer, corresponding to each event, are treated as one word by the Univac 1108 central processor in which the basic computer word consists of 36 bits. The programme SORTER splits each word into the original 16-bit words by suitable masking. It then examines the timeof-flight information to determine to which of the neutron energy ranges the capture event belongs. Once this is ascertained, the programme inspects the first word which contains the pulse-height information and the spectrum label. The spectrum label determines to which of the three pulse-height spectra corresponding to this particular neutron energy range the event should be added. One count is then registered in this spectrum at the channel number deduced from the pulse height. Each magnetic disc has the capacity to store information for about 0.63 x 10 6 events. The programme, on the average, uses 3½ rain to analyze the contents of one disc and generate thirty spectra. It can handle the
contents of several discs. In addition, the programme generates plots of these spectra.
5. Capture gamma-ray spectra Capture gamma-ray data have been recorded for natural tungsten covering the range from thermal energy to 100 keV neutrons. These measurements have been made using a cylindrical disc of natural tungsten with its axis at 45 ° to the axis of the neutron beam. The spectrometer was located at 90 ° to the beam, viewing the front face of the sample. The sample was 15.2 cm in dia. and had an effective thickness of 0.0071 nuclei/barn. A time-of-flight spectrum measured for all gamma rays detected in the singles mode of the spectrometer is shown in fig. 10. The channel width for this display is 1.2psec. The low-energy resonances in the various isotopes of tungsten are well resolved. The peak at the lowest energy corresponds to the 4.16 eV resonance in 1S2W, the next one is due to the 7.65 eV resonance in l s3W. There are two resonances in the neighbourhood of 20 eV; one at 18.84 eV in ~S6W and the second at 21.2 eV in 182W. The display of these two resonances is somewhat distorted by the effects of the thick capture sample used in this measurement. Fig. 11 shows the pair spectra for neutron capture in the energy range 0.02 eV to 1.0 eV and three neutron energy ranges covering the four low energy resonances in natural tungsten. These spectra are dominated by
NEUTRON
CAPTURE
GAMMA
the effects of the resonances. As has been reported previously1 z - 14), the gamma-ray spectra vary markedly with neutron energy. This is partly due to the fact that these resonances are accounted for by several isotopes of tungsten. 5
4
tOO|f"
~
RAY
265
FACILITY
A comparison of the capture gamma-ray spectrum measured in the singles mode (upper spectrum) and in the pair mode (lower spectrum) for neutrons between 2.5 eV and 6.0 eV is given in fig. 12. It should be noted that the Compton background, which increases sharply
)'-RAY ENERGY (MeV) 5 6 r
7
8
,
~
I
/
!'
W(n,¥)
80[-
t
40
E n = 0 0 2 0 8 TO 10020 eV
..
20
~. UO
300
400 500 CHANNEL NUMBER
600
700
T- RAY ENERGY (MeV) 5 500 i 400
4
5
6
7
8
I
i
i
i
l
200
En =2.5 TO 6.0 eV RESONANCE 4,16 eV IN
[
160
18z w
300~
dz
z I
120
z z <
200 z D 12) I O0
-- 4 0
z
80L
En = 6 0
60 ~
RESONANCE 7.65 eV IN le3 w
TO 9 0 eV
2 0 ¢;"'~"
250
En z g . 0 T O
l
2 5 . 0 eV
'
RESONANCES 1884 eV IN 186W 21 2 eV IN m2W
200v 150
,OOo 200
300
400 500 CHANNEL NUMBER
600
700
F i g . 11. Capture g a m m a - r a y spectra f r o m tungsten measured with the pair spectrometer, The spectra corresponding to four n e u t r o n e n e r g y i n t e r v a l s a r e s h o w n . T h e n e u t r o n e n e r g y r a n g e and the r e s o n a n c e s in t u n g s t e n c o n t a i n e d in t h i s r a n g e a r e i d e n t i f i e d . T h e spectrum for 0 . 0 2 e V t o 1.002 e V was measured in a r u n d i f f e r e n t from the o n e in which the other three spectra were measured.
266
v . J . ORPHAN et al. 5000-
i
i
<
r
1
[----
I
-
-
"[
W {n,Y) E n =25T0 60eV SINGLES S P E C T R U M
k %
5000 F
.
G,I T 2
:
% 'Q
2000
S6
©~
\., "I;.
•
0 v oJ
..
t°t
x,o
_ ~
~o c~
'
ed O0
I000 z z
8 -.
O-q-
o
~o
En=25
-
PAIR
TO
60eV
SPECTRUM
tO
250
200
o
£~
150
~
0
J i
, oo
~o
I~
ss
. • :'. 5 0 ~. . " ' . ~ .~'~'%.'.
.
Sl
'(u
1
T
S
i
~,
.~.
[
• 200
500
4C0
500
600
CHANNEL
I
700
800
9OO
NUMBER
Fig. 12. Comparison of capture gamma-ray spectra from tungsten obtained with singly-operated Ge(Li) detector (upper) and pair spectrometer (lower). Easy identification of peaks and high peak-to-background ratio in the pair spectrum is illustrated. with decreasing gamma-ray energy in the singles spectrum, is practically eliminated in the pair spectrum. The full-energy and single-escape peaks that are present in the singles spectrum are totally absent in the pair spectrum. The peak-to-background ratio at 5164 keV is improved by a factor of 16. Use of a pair spectrometer greatly simplifies the spectrum unfolding required to obtain capture gamma-ray intensities for elements having complex capture spectra. To demonstrate the flexibility of this system the average gamma-ray spectrum formed by neutron capture in many resonances has also been generated by the
programme. Fig. 13 shows the gamma-ray spectrum resulting from the capture of neutrons from 1.5 eV to 100 keV measured with the pair spectrometer. This gamma-ray yield is averaged by an approximately lIE spectrum of incident neutron flux. Such averaged spectra may be used to obtain spectroscopic information. The spectra shown above are only a part of over 30 spectra from tungsten that have been generated and analyzed. A computer code, G A M A N L ' 5), developed at the Massachusetts Institute of Technology has been modified for the analysis of these data. The observed
NEUTRON
CAPTURE
267
GAMMA RAY FACILITY
GAMMA-RAY ENERGY (MeV) 4003
4
5
6
I
l
I W (n,Y) PAIR SPECTROMETER En = I 5 eV TO I 0 0 keV
3OO w Z Z I
Z
D
© 200
I00
0
1 20O
} 300
[ 40o
I 500
600
l 700
CHANNEL NUMBER
Fig. 13. Averagespectrum obtained from the capture o f l . 5 e V to~10~keV[neutrons|in~tungsten. Thespectrum is averaged byan approximately I/E flux of neutrons. Capture occurs in many resonances in tungsten. d a t a are first s m o o t h e d using a F o u r i e r analysis technique. The p r o g r a m m e then a u t o m a t i c a l l y locates the peaks a n d determines their centres and areas. F r o m this analysis the intensities o f the v a r i o u s transitions are obtained. 6. P e r f o r m a n c e
characteristics
The c a p t u r e g a m m a - r a y facility has been in o p e r a t i o n for a b o u t a year. The Linac has been o p e r a t e d with TABLE l Performance characteristics of the Linac capture gamma-ray facility. Narrow pulse Wide pulse Pulse width (#sec) Electron energy (,MeV) Electron current (A) Repetition rate (,pps) Flux* at 1 eV [n/(.cmZ.sec.lnE)] Flux* atl00keV [n/(cm"-.sec.lnE)] Timing resolution (nsec/m)
0.1 40.0 3.0 600 1.56x i0 a 4.14x10 a 6.25
4.5 40.0 0.6 180 4.2× 103 1.12x104 280
* The measured values of flux have an uncertainty of _+ 20%.
repetition rates up to 600 pps. Typical p e r f o r m a n c e characteristics o f the system are s u m m a r i z e d in table 1. Figures are quoted for two different modes o f o p e r a t i o n - one for n a r r o w pulses and the other for wide pulses. It is o f interest to c o m p a r e the flux and timing resolution o f the Linac a r r a n g e m e n t with those of a fast c h o p p e r facility 8) which is also used to study c a p t u r e g a m m a - r a y spectra. There is a distinct difference between the two systems. In the case of the Linac the pulse width and the pulse-repetition rate can be varied independently. Hence, high average beam intensity can be m a i n t a i n e d without sacrificing the neutron energy resolution. In the fast chopper, however, the burst width and the repetition rate are intimately c o n n e c t e d by the nature o f the c h o p p e r o p e r a t i o n . In that case high average beam intensity necessarily implies p o o r neutron energy resolution. in the wide pulse mode, with a pulse width o f 5.0 /~sec, the c h o p p e r provides a flux per unit lethargy o f 8 x 103 n e u t r o n / c m Z - s e c which is quite c o m p a r a b l e to the flux per unit lethargy o f 4 . 2 x 103 to 1.12x 104 n e u t r o n / c m z. sec o b t a i n e d with the Linac for a 4.5 ~tsec pulse width at 180 pps. The c h o p p e r with a 22-m flight
268
v . j . ORPHAN et al.
path has a timing resolution of 227 nsec/m compared to 280 nsec/m for the 16-m Linac system. However, in n a r r o w pulse operation, the Linac outperforms the chopper facility. The chopper operated with a 1.0 psec pulse width has a flux of 4.4 x 10z n e u t r o n / ( c m 2. sec. • ln E interval). F o r a 0.1/~sec pulse width, the Linac provides a flux of 1.56 x 103 to 4.14 x 103 n e u t r o n / (cm 2• sec- In E i n t e r v a l ) at 600 pps which is at least three times the flux from the chopper. The timing resolution for the Linac system is 6.25 nsec/m c o m p a r e d to 4 5 . 5 nsec/m for the chopperS). It is therefore evident that the Linac system is superior to the chopper assembly for high resolution studies in the epithermal energy range. This superiority is m a i n t a i n e d even with the new chopper flight path of 48 m. We are grateful to Y. D. Naliboff for p r o g r a m m i n g the CDC-1700 C o m p u t e r a n d to Prof. N. C. Rasmussen for supplying us with the G A M A N L code. The assistance of the Linac O p e r a t i o n s staff, the electronics personnel and the w o r k s h o p staff in the development, o p e r a t i o n and m a i n t e n a n c e of this facility is acknowledged.
References 1) c. E. Porter and R. G. Thomas, Phys. Rev. 104 (1956) 483. 2) e.g., D. L. Price, R. E. Chrien, O. A. Wasson, M. R. Bhat,
M. Beer, M. A. Lone and R. Graves, Nuclear Physics AI21 (1968) 630. a) A. M. Lane and J. E. Lynn, Nuclear Physics 17 (1960) 586. 4) C. K. Bockelman, Nuclear Physics 13 (1959) 205. a) H. Morinaga and C. Ishii, Prog. Theor. Phys. (Kyoto) 23 (1960) 161. ~) O. A. Wasson, R. E. Chrien, M. R. Bhat, M. A. Lone and M. Beer, Phys. Rev. 173 (1968) 1170. 7) L. M. Bollinger and G. E. Thomas, Phys. Rev. Letters 18 (1967) 1143; ibid. 21 (1968) 233. s) R. E. Chrien and. M. Reich, Nucl. Instr. and Meth. 53 (1967) 93. •~) W. M. kopez, M. P. Fricke, D. A. Costello and S. J. Friesenhart, Gulf General Atomic Report GA-8835 (Sept. 1968) unpublished. lo) V. J. Orphan and N. C. Rasmussen, Nucl. Instr. and Meth. 48 (1967) 282. 11) N. C. Rasmussen, Y. Hukai, T. lnouye and. V. J. Orphan, Massachusetts Institute of Technology Report M1TNE-85, to be published. va) E. R. Rae, W. R. Moyer, R. R. Fullwood and. J. L. Andrews, Phys. Rev. 155 (1967) 1301. 13) R. R. Spencer and K. T, Faler, Phys. Rev. 155 (1967) 1368. 14) W. V. Prestwich and R. E. Cote, Phys. Rev. 160 (1967) 159. 15) T. Harper, T. Inouye and N. C. Rasmussen, Massachusetts Institute of Technology Report MIT-3944-2 (August 1968).