C. B u r g e s s a n d K.D. M i e l e n z ( E d i t o r s ) , Advances
in Standards
and Methodology
in
Spectrophotometry
1 9 8 7 Elsevier S c i e n c e P u b l i s h e r s Β. V . , A m s t e r d a m — P r i n t e d in T h e N e t h e r l a n d s
PHYSICAL STANDARDS IN ABSORPTION AND REFLECTION SPECTROPHOTOMETRY
J F VERRILL Division of Quantum Metrology, National Physical Laboratory Teddington, Middx, TW11 OLW, UK
ABSTRACT
In recent years there has been a significant growth in the availability of different types of reference materials and transfer standards for the calibration of spectrophotometers and colorimeters. The quantities that need to be checked are the photometric linearity of the absorbance/ transmittance/reflectance scale, the accuracy of the wavelength scale and the stray radiation performance, together with the accuracy of several possible colour specification scales used on instruments dedicated to colour measurement. The principal areas of interest are regularly transmitting standards for use by analytical chemists and diffusely reflecting standards for use in colour measurement, solar reflectance and other areas. Transfer standards for specular reflectance are also available but the whole area of diffuse transmittance spectrophotometry has been less well researched, possibly due to the lack of a well defined need.
INTRODUCTION When optical reflected
and
radiation in
part
falls
on any material
absorbed.
In
some
transmitted. Spectrophotometry can be divided and
transmittance. Transmission
and
object
cases
it
it
may
will also
in
part
be
in
be
part
into two main areas, reflectance
reflection
can
be
of two kinds,
regular
(specular) and diffuse. Many samples have both regular and diffuse properties. The
measurement
of
spectral
reflectance
and
transmittance
spectrophotometry, a misleading word since it is not region
of
the
spectrum
as
the
syllable
"phot"
confined would
is to
imply.
known the
as
visible
The
term
"spectrometry" is preferable, though not endorsed by the CIE. As transmittance and reflectance are both dimensionless ratios there is no need
for a fundamental physical standard
such as is needed
for time or
mass.
However, the accuracy of any measurement will depend greatly on the design of the instrument and possibly on the quality of the sample itself. Measurements should be traceable to a reference instrument in which systematic uncertainties
112 have
all been
through made
carefully
which
the
traceable
The major
to
of
since
Where
or
medium for
because
they
not,
in
as
prepares
the
This raises
for chemistry
However,
low
high are
the
standards
of
radiometric
in
but can
i f the be
stray
done
wavelength with
Calibration
transmittance decoupled
requires
is
from
calibration
known
independent wavelength
are needed
to
of
radiation
characterisation
of
monochromator
or
require
lines
methods however,
laser are not
a
single
laser for
range
are
is
be
but
by m e a n s
lines be
scale
firstly
values
skill
of
are
whoever
This
full
working
a
difficult
used
as
a guide
of c u t - o f f
to
is o f t e n
of
range
the
be with
of
subject
most
stray
the
Double users,
radiation
filters. Several
the
filters holder. whose
errors
to
degrees
instrument.
and
radiation
or
filters
ratio
that
of
sample
differing
source. For
little
lamps
the
enables
requires the
into
ideally
radiation calibration
Calibration
discharge
which
calibration
stray
instrument
there
filters
for
scale,
the
after
from
should
evaluation.
desirable.
of
inserted
Several
the
full
liquids.
preferable.
liquids,
and
instrument.
monchromator
to a
practicable
be o b t a i n e d
full wavelength
a
to
be
and
necessary
on
spectrophotometry
scales
wavelength.
cover
are
calibration
the w a v e l e n g t h
better
can
errors.
evaluation
the
is i n a d e q u a t e
a
ratio
attenuation
well
expertise
in
of
the
spectral
of
standards
may
measurement
to
that
radiometric
stray
and
get
peaks
be
laboratory.
scale
to
can
chemistry
measurements
preferable
the
standards
industry
laboratories.
solid their
because on
transfer
and
are analytical
liquids
quantities
performance
than
absorption of the
given
of
are
dependent
c a n be m a d e
radiation
scale
narrow
solids secondly
for
science
the n a t i o n a l
needed
and
are needed,
other
arises
in
whether
work
any
ratio
of
are
particular
properties. Adjustments
in
the m a j o r i t y
of liquids,
physical the
the q u e s t i o n
stable
three
thus
used
of spectrophotometry
accuracy
case
are
instruments
use
accuracies
more
The need
instruments
chemists do
the s o l u t i o n s
There
of
reference
areas
colorimetry.
evaluated.
scales
the from
be of The
complete a
double
monochromators however,
these
performance
of these
covering
can, the
113
(α)
(b) (Ο
transmittance or reflectance
wavelength
Fig.1
Ideal spectral profiles: a) neutral density b) absorption peak c) cut off.
Figure 1 indicates schematically are required, the neutral density
the
three
types
of spectral
filter, the narrow
absorption
cut-off filter. For instruments dedicated
to colour
measurement
added
standards
should
requirement
highly
saturated
colours.
Physical
that
physical
colours
and,
standards
transfer
if numbers should
be
with
peak there
cover
permit, a range stable
profile
of
time,
a
low
that
and
the
is
the
range
of
saturation
durable,
of
low
temperature coefficient, and readily available. A summary of currently available reference materials and transfer
standards
for testing the performance of spectrophotometers and colorimeters has recently been prepared by the CIE Committee 2 - 1 3 (ref 1 ) . This paper concentrates on the question of how well these materials
meet
current
requirements
and
indicates
areas where further development is needed.
REFERENCE MATERIALS AND TRANSFER STANDARDS CURRENTLY AVAILABLE A brief summary of available materials is listed will
be
found
in
the
report
of
CIE
Committee
as
follows. Full
2-13.
Note
that
details
chemicals
requiring preparation are not listed here.
Wavelength 1.
Spectral emission lines from the following elements:
deuterium, cadmium,
caesium, helium, neon, argon, krypton, mercury, potassium, zinc and rubidium 2.
Absorption filters of didymium and holmium glasses.
114 Regular
transmittance
3.
Neutral
4.
Metal
Regular
First
6.
Second
7.
First
No
fused
glass
filters
silica neutral
density
filters
reflectance
5.
Diffuse
density
on
surface
aluminium
surface
mirror.
aluminium
surface gold
mirrors
with or without
wedge.
mirror.
transmittance reference
materials
for
spectral
diffuse
transmittance
are
as
yet
available.
Diffuse 8.
reflectance
Barium
sulphate
9.
Halon
10.
Russian Opal, Ever
11.
White
12.
Vitrolite
13.
Black
ceramic
glazed
14.
Black
15.
Ever
16.
Ceramic
17.
Enamel
Stray 18.
ceramic
porcelain
tile
enamel
Black
Colour Colour
Standards Standards
light Cut-off
The opals
list and
filters.
excludes
WAVELENGTH
Clarke
not
standards.
there are many
yet
It
commercially
also
excludes
available
printed
such
as
or painted
coloured
papers
and
types.
STANDARDS
The wavelength known
materials
fluorescent
c a r d s of w h i c h
well
White
tile
scales
spectral
(ref. 2 ) .
of
lines. An
reference A
list
uncertainty
of of
spectrophotometers those
most
.01 nm
is
are
calibrated
frequently
used
adequate
for
is
against given
almost
by all
115 requirements in analytical chemistry and colorimetry. Many spectral lines known to higher
accuracies
than
used with any spectrophotometer
this. In
principle, spectral
but in practice there may
lines
be major
are
could
be
problems.
Most commercial instruments are not designed to permit arbitrary sources to be focussed on the entrance slit of the monochromator.
Fig. 2
The
Wavelength standards; a) holmium glass b) didymium glass. percentage transmittance as a function of wavelength in nm. big
advantage
inserted directly
of glasses
with
absorption
peaks
is
into the sample holder. The most widely
that used
they
can
materials
be for
wavelength absorption peaks are holmium and didymium oxides in a glass matrix (ref. 3 ) . The wavelength of the absorption independent
of
temperature
but
the
peaks
is, for practical
transmittance
significant changes with temperature. Transmittance
values
at
the
purposes, peaks
show
curves for the ultraviolet
116 and
visible
data for
regions
stations routine
spectral
are able
work,
it
emission
absorption
are
peaks
shown
in
to l o c a t e is
a much
lines.
to be u s e d
is
TRANSMITTANCE
STANDARDS
3 illustrates
some
at
compartment in
the
sample
happens the
(a) and
when
sample
compartment detector a so
that
Fig.3
sample
is
increased
the
means
is d e f l e c t e d
at t h e
the
rays
Departures
significant
errors
reflectance
with
(ref. 4 ) .
within
from will
angle
normal
this
can
be
is
materials
with
errors For
at
that
can
simplicity beam
the
arise
with
narrower
only
will
have
detector.
two a
Now
within
what
the
refractive
length
within
cross
error
section
is n o t
arises
of
the
uniform
if the
are
consider
path
the
rays
area
Because
sensitivity
the
finite
beam.
that
to
adequate
achieved
optical
arise. A similar
c h a n g e of b e a m c r o s s of sample sample wedge p a r a s i t i c b e a m s from
Generally
the
While
than for
of the
area
the
If the d e t e c t o r
Systematic errors arising spectrophotometer :
b) c)
in
nm.
coupled
standards.
section
unity
which
need
systematic
finite
placed
•+ 0.2
a
spectrophotometer.
than
will
to
spectrophotometers
uncertainty
clearly
the c r o s s
is
greater
error
of
different
is
the beam
a)
sample.
a
is c h a n g e d .
systematic
a
(b) but
and a
of
peaks
as w a v e l e n g t h
Fig.
shown
the
2. M o d e r n
higher
There
REGULAR
sample
figure
index
the beam
across
sample
of
sample at
its
has a
the area
wedge
detector.
in t h e
sample
section
compartment
at d e t e c t o r
on
of a
insertion
interreflections. the
sample
incidence
arise. These of incidence
can or
compartment must
be
limited
be e i t h e r
increased
are
as
path
a
not to
a
result
length
normal few of
to
the
degrees
or
variation
of
within
the
sample
117 All materials reflect a percentage of the incident radiation reflected from the sample may be reflected the
spectrophotometer
to
Likewise radiation may
pass
through
be reflected
reflection at the sample. The
the
from
parasitic
sample
the
radiation.
and
detector
beams
give
Some
of
the
back from a component of reach and
an
the
detector.
be returned
error
in
the
after
measured
value of transmittance. Interreflection errors can be avoided by careful design with
components
suitably
angled
so
that
parasitic
beams
do
not
reach
the
detector. For most instruments interreflection errors are negligibly small for non-metallic
samples
but
they
often
become
significant
for
metallic
samples
where the reflectance is higher. If
there
is
a
significant
component
of
diffuse
transmittance
then
the
instrumental reading will be dependent on the solid angle of collection. Total transmittance of samples with a significant should
be measured
with
the
sample
at
component of diffuse
the
entrance
port
of
transmittance
an
integrating
sphere. There are, of course, many sample
compartment
dependent
on
both
other
but
the
the
quality
sources
preceding of
of
summary
the
error
that
indicates
sample
and
lie
outside
the
that
errors
are
the
quality
of
the
of
two
spectrophotometer. Currently
available
physical
regular
transmittance
standards
types, neutral density glass filters and metal film on silica The
spectral
transmittance
curves
of
four
filters
of
each
are
(fused type
quartz).
of
nominal
transmittance 92%, 56%, 32% and 10% are shown in fig. 4. The advantages of the glass filters are that they are very stable and the surface reflectance
is low
and similar to that of cuvettes. However, they absorb strongly below 400 nm and so cannot be used
in the ultraviolet. Metal film filters consisting of a thin
layer of a nickel-chromium alloy on a silica overcome
this
problem
and
can
be
used
(quartz plate) were developed
down
to
200 nm.
They
are
also
to
more
neutral than glass filters in the near infrared. But, the higher reflectance of the metal film does cause problems in some spectrophotometers (refs. 5 , 6 ) . What is needed
is a material which is approximately
neutral down to 200 nm
with a
reflectance similar to that of silica. At the present time there are no obvious candidates.
118
200
400
300
500
600
700
800nnr
Wavelength
neutral density m e t a l
f i l m on s i l i c a f i l t e r s
§ 50%
400
Fig. 4
500 Wavelength
600
Transmittance curves of four neutral density glass filters and four metal film on silica filters.
REGULAR REFLECTANCE STANDARDS Regular
reflectance
spectrophotometer need
built
is
usually
for regular
measured
with
a
transmittance
for a series of neutral mirrors of
special
attachment
measurements.
different
There
transmittances
to is
because
a no
the
linearity of the radiometric ratio can be checked with the same filters as are used for transmittance. However, with regular reflectance attachments the path of the beam may be very different for the reference ( 1 0 0 % ) Take,
for
example
the
VW
type
of
reflectance
and sample readings.
accessory,
fig. 5.
reference reading the beam follows the V path and for the sample beam
follows
the
W
path.
The
method
gives
the
square
of
For
the
reading
the
the
spectral
119 reflectance as the beam is incident twice on the sample. If the mirrors are not perfectly aligned then the beam will not fall on the same patch of the detector for the
reference
uniform
over
spectrally
and
sample
readings.
its area
then an
calibrated
regular
reflectance. It is
important
error
If
will
reflectance
that
the
the
detector
result. Thus standards
reference
sensitivity the
with
standard
need a
and
is
not
arises
for
high the
neutral
sample
are
mounted in the same plane. Therefore a front surface mirror should be used as the standard where front surfaces are to be measured. Aluminium and gold are both used for reflectance standards. Aluminium is neutral region whereas
gold
is not, but gold
is neutral
in the
in
infrared
the
films
visible
and
has
a
higher reflectance in that region than aluminium. Back surface mirrors are also available and are more stable
because
the metal
substrate. However they should only be used
in
surface the
same
is protected plane
as
surface as many reflectance attachments give readings that are a
by
the
the
sample
function
of
sample position. Back surface mirrors with a wedge are also available and have the
advantage
that
the
front
and
rear
surface
reflected
beams
are
not
coincident. They cannot, of course, be used with reflectance attachments where the front surface is used for location.
Fig. 5
VW regular reflectance attachment. Misalignment of the sample causes a displacement of the beam at the detector.
120 DIFFUSE REFLECTANCE STANDARDS
Fig. 6
Variation of radiance factor with angle for a glossy and a matt Russian opal.
Diffuse reflectance standards are required primarily for colorimetry but in recent
years
other
areas
such
as
integrated
solar
reflectance
important. A major consideration is whether such standards
have
should
become
be matt
or
glossy. The big advantage of glossy standards is that they are much easier to keep
clean
than
disadvantages.
matt
standards.
Firstly
they
are
However,
a
less
glossy
good
standards
approximation
to
diffuser than a matt standard (refs. 7 , 8 ) . This is illustrated gives the variation Russian
opal.
A
of
luminance
perfect
factor with angle
diffuser
would
have
a
have a
Lambertian
in fig. 6 which
for a glossy
luminance
several
and
factor
a matt
of
unity
independent of angle. Secondly the specular component may not be collected with the same efficiency as the diffused light in the integrating sphere giving rise to a systematic from
error
the detector
collected with
but
(ref. 9 ) . In
fig. 7 the diffuse
the
component
a higher
specular
efficiency.
is
If a gloss
not trap
radiation and
will
is used
is
screened
therefore
to exclude
be the
specular component it may well not be perfectly efficient. Thirdly the radiance factors for ρ and s polarized light are much more different for glossy
samples
than for matt samples as shown in fig. 8 (refs. 7 , 8 ) . This point is of importance
in
instruments with
a
0°/45°
(or 45°/0°) measuring
head.
great
If
the
state of polarization of the incident radiation is uncertain then the radiance factor for the glossy sample
of figure 8 could
be anywhere
between 0.93
and
1.01 but for the matt sample it will lie between 0.98 and 0.99. In spite of all these disadvantages reflectance
have
those
opted
laboratories
predominantly
issuing for
transfer
glossy
rather
standards than
matt
because of the much greater durability and ease of keeping clean.
of
diffuse
materials
121
Detector
Fig. 7
Integrating sphere with a single screen. The diffuse component of reflection is screened from the detector. The regular component which falls on the opposite side of the sphere is unscreened.
Fig. 8
Differences in variation of radiance factor with angle of ρ and s polarized light for glossy and matt Russian opals.
The
most
widely
(ref. 1 0 ) and pressed
used
matt
PTFE powder
reflectance (halon)
standards
are
(ref. 1 1 ) . Barium
either as a pressing or with a binder as a
barium
sulphate
sulphate
is used
paint. A number
of
manufacturers
supply painted barium sulphate reference standards recessed back into a metal plate. Recessing prevents scuffing of the surface when
placed
against
a
port
but it introduces a major new problem because the standard will not be in the same plane as the sample and thus the efficiency of collection by will be different
for the reference and sample
the
sphere
(ref. 9 ) . Halon is widely used
122 in North America but less so in Europe
possibly
because
historically
opal has been more readily available in Europe. Halon has a higher than barium sulphate at around
2000
Russian
reflectance
nm and is therefore to be preferred as an
integrating sphere coating for use in the infrared. Black
tiles have
a total
reflectance
typically
of
4 to 5%,
materials generally have a refractive index of around
Because
solid
1.5 or greater a smooth
surface will always give a glossy reflectance of about 4%. Abrading the surface does not reduce this. It merely converts the glossy reflectance reflectance. Reflectances below H% can be achieved but where a very low reflectance
is
required
with
a trap
a
in
into a diffuse
structured
the
form
surface
of a
glass
wedge is preferred (ref. 9) fig. 9.
Fig. 9
Black glass wedge gloss trap.
Fig. 1 depicts
two
other
types
of spectral
profile
required
standards. Unfortunately diffusely reflecting materials
with
for
sharp
transfer
absorption
peaks are not available at the present time although there is a definite need for them as wavelength standards. Spectral profiles with a single steep slope are
available
and
the
mid
point
calibration. The difficulty
here
dependent so one must always standard
of
the
is that
be certain
slope
can
the mid that
the
be
point
used
for
wavelength
value
is
temperature
surface
temperature
of
the
is the same as that when calibrated, if a wavelength error is to be
unambiguously distinguished
from a thermochromic shift of the spectral
Generally,
below
the
reflectance
the
evaluating stray radiation performance.
steep
slope
is
too
high
for
slope. use
in
123 DIFFUSE TRANSMITTANCE STANDARDS This is a neglected area. Indeed, the author has been unable to identify any calibrated
reference
materials
or
transfer
standards
for
spectral
diffuse
transmittance. In measuring diffuse transmittance a reference reading is taken with the sample removed so that the incident radiation falls on one small patch of the sphere wall opposite the entrance port. A second reading is then taken with the sample at the entrance port and the ratio of the two readings taken to give the diffuse transmittance. The difficulty
is that
rays entering the sphere are detected with equal evidence
to
uncertainty
justify of
Calibrated
this
less
opal
assumption
than
about
diffusers
are
2%
it
is
where
frequently
this assumes
efficiency.
not
possible
absolute used
Unless to
values
as
ascribe
are
transfer
that
all
there
is an
required.
standards
in
densitometry. Systematic uncertainties are always large because of uncertainty in the absolute value of transmittance of the standard, lack of a well defined geometry
of
collection,
non
Lambertian
diffusion
and
multiple
reflections
between the sample and the detector.
STRAY LIGHT STANDARDS A number of chemical standards for stray radiation measurement are available but these lie outside the scope of this paper. The most widely used glass cut off filters are those produced many
of
these
fluoresce.
fluorescence, the
cut
by Schott. It should, however, be noted
Unless
it
is
off filter must
known
that
be placed
there
between
is the
no
that
significant
source and
the
monochromator rather than between the monochromator and the detector.
CONCLUSIONS Although
a wide
range
of physical
standards
for spectrophotometry
is
now
available, there are still several areas where there are no suitable
standards
or
need
where
improvements
are
needed.
transmittance standards of low down
to
peaks,
200 and
nm,
transmitting
diffuse
reflectance wavelength
reflectance
better cut off properties.
In
standards
particular with
a wavelength
standards with
there
with
narrow
is
a
range
narrower
absorption
for
extending absorption peaks
and
124
REFERENCES
1
CIE
(in
preparation).
Survey
of
reference
materials
for
testing
the
performance of spectrophotometers and colorimeters. Report of CIE Committee TC2-13. 2
Commission Internationale de l'Eclairage, Paris.
Clarke F J J absorption
(l°/8l).
Reduction
spectrometry.
UV
of
the
uncertainties
Spectrometry
Group
of
standards No 9 .
Bulletin,
in
Part 2 ,
8I-9O.
3
Dodd
4
Mielenz K D ( 1 9 7 2 ) ,
C X
and
West
Τ W
(I96I).
Spectral
transmittance
properties
of
rare
earth glasses. J. Opt. Soc. Am, Vol 5 1 , 9 1 5 - 9 1 6 .
J.
5
Res.
NBS.
Mielenz
Vol
Κ D
Physical parameters in high accuracy
76A,
and
spectrophotometry.
455-^67·
(1973)·
Mavrodineanu R
Reflection
correction
for
high-accuracy transmittance measurements on filter glasses. J. Res. NBS. Vol 77A
699-703.
6
Verrill J F ( 1 9 8 3 ) .
7
Clarke F J J,
A re-evaluation of metal film on silica neutral density
filters. UV Spectrometry Group Bulletin, No 1 1 , 3 0 - 3 8 . Garforth
F A
and
Parry D J
(1977).
Goniophotometric
and
polarization properties of the common white reflection standards. NPL Report MOM 2 6 . 8
Clarke F J J, polarization
9
Garforth F A properties
of
and
Parry D J
white
reflection
Research and Technology, Vol 1 5 ,
133-149.
Clarke F J J and Compton J Anne
(I986).
(1983).
Goniophotometric
standard
materials.
Correction methods
for
and
Lighting
integrating
sphere measurement of hemispherical reflectance. Col. Res. Appl. Vol 11 (in press). 1 0 CIE ( 1 9 7 9 ) . material
A review of publications on properties and reflection values of
standards.
CIE
publication
46,
Commission
Internationale
de
l'Eclairage, Paris. 1 1 Weidner R
and
Hsia J J
(I98I).
Reflection
properties
polytetrafluorethylene powder. J. Opt. Soc. Am. Vol 7 1 . 8 5 6 - 8 6 I .
of
pressed