Posterolateral Trajectories Favor a Longer Motor Domain in Subthalamic Nucleus Deep Brain Stimulation for Parkinson Disease

Posterolateral Trajectories Favor a Longer Motor Domain in Subthalamic Nucleus Deep Brain Stimulation for Parkinson Disease

Accepted Manuscript Posterolateral trajectories favor a longer motor domain in STN DBS for Parkinson’s disease Idit Tamir, Odeya Marmor-Levin, Renana ...

3MB Sizes 2 Downloads 130 Views

Accepted Manuscript Posterolateral trajectories favor a longer motor domain in STN DBS for Parkinson’s disease Idit Tamir, Odeya Marmor-Levin, Renana Eitan, Hagai Bergman, Zvi Israel PII:

S1878-8750(17)31086-0

DOI:

10.1016/j.wneu.2017.06.178

Reference:

WNEU 6049

To appear in:

World Neurosurgery

Received Date: 21 March 2017 Revised Date:

26 June 2017

Accepted Date: 29 June 2017

Please cite this article as: Tamir I, Marmor-Levin O, Eitan R, Bergman H, Israel Z, Posterolateral trajectories favor a longer motor domain in STN DBS for Parkinson’s disease, World Neurosurgery (2017), doi: 10.1016/j.wneu.2017.06.178. This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

ACCEPTED MANUSCRIPT Tamir Title: Posterolateral trajectories favor a longer motor domain in STN DBS for

2

Parkinson’s disease

3

Authors: Idit Tamir1,2,4, Odeya Marmor-Levin2, Renana Eitan3, Hagai Bergman2,3

4

and Zvi Israel4

RI PT

1

5 Affiliations:

7

1 – Department of Neurological Surgery, University of California San Francisco, San

8

Francisco, CA, USA. Postal address: 505 Paranassus St. San Francisco, California, 94143

9

(present address)

M AN U

SC

6

10 11

2 – Department of Medical Neurobiology, Hadassah Hebrew University Medical Center,

12

Jerusalem, Israel. POB 12271, ZIP 9112102.

TE D

13

3 – Edmond and Lily Safra Center for Brain Research, The Hebrew University, Ein Krem

15

Campus, Jerusalem, Israel. POB 12271, ZIP 9112102.

AC C

16

EP

14

17

4 – Department of Neurosurgery, Center for Functional and Restorative Neurosurgery,

18

Hadassah-Hebrew University Medical Center, Jerusalem, Israel, 91120.

19 20

1

ACCEPTED MANUSCRIPT Tamir Corresponding author:

22

Dr. Idit Tamir, MD, PhD. Department of Neurosurgery, Center for Functional and

23

Restorative Neurosurgery, Hadassah-Hebrew University Medical Center, Jerusalem,

24

Israel, 91120. Email address: [email protected].

RI PT

21

25

SC

26 Odeya Marmor-Levin, M.sc. [email protected]

28

Renana Eitan, MD. [email protected]

29

Hagai Bergman, MD, PhD. [email protected]

30

Zvi Israel, MD. [email protected]

M AN U

27

EP

33

AC C

32

TE D

31

2

ACCEPTED MANUSCRIPT Tamir 34

Funding

35

This research was supported by the following grants:

37

1. Adelis Foundation and Israel-US Binational Science Foundation (BSF) to ZI, RE and HB.

RI PT

36

2. The Israel Science Foundation (ISF).

39

3. The German Israel Science Foundation (GIF).

40

4. The Canadian friends of the Hebrew university.

41

5. The Rosetrees and Vorst Foundations.

42

6. The Simone and Bernard Guttman chair in research – to HB

AC C

EP

TE D

M AN U

SC

38

3

ACCEPTED MANUSCRIPT Tamir Abstract

44

Objective: The clinical outcome of Parkinson's disease (PD) patients that undergo

45

Subthalamic Nucleus Deep Brain Stimulation (STN DBS) is, in part, determined by the

46

length of the electrode trajectory through the motor STN domain, the Dorso-Lateral

47

Oscillatory Region (DLOR). Trajectory length has been found to correlate with the

48

stimulation-related improvement in patients’ motor function (estimated by part III of the

49

United Parkinson’s Disease Rating Score (UPDRS)). Therefore, it seems that ideally

50

trajectories should have maximal DLOR length.

51

Methods: We retrospectively studied the influence of various anatomical aspects of PD

52

patients’ brains and the geometry of trajectories planned on the length of the DLOR and

53

STN recorded during DBS surgery. We examined 212 trajectories and 424 electrode

54

MER tracks in 115 patients operated in our center between the years 2010 and 2015.

55

Results: We found a strong correlation between the length of the recorded DLOR and

56

STN. Trajectories that were more lateral and/or posterior in orientation had a longer STN

57

and DLOR pass, although the DLOR/STN fraction length (%DLOR) remained constant.

58

The STN target was more lateral when the third ventricle was wider, and the latter

59

correlated with older age and male gender.

60

Conclusions: Trajectory angles correlate with the recorded STN and DLOR lengths, and

61

should be altered toward a more posterolateral angle in older patients and atrophied

62

brains in order to compensate for the changes in STN location and geometry. These fine

63

adjustments should yield a longer motor domain pass, thereby improving the patient's

64

predicted outcome.

65

Keywords: STN; Beta oscillations; DBS; Parkinson’s disease; LFP; Trajectory

AC C

EP

TE D

M AN U

SC

RI PT

43

4

ACCEPTED MANUSCRIPT Tamir 66

Introduction

67

It is generally accepted that precise lead location in STN DBS will determine 1

68

motor outcome in PD patients

69

(motor) score.

70

most centers target the dorsal motor part of the STN, some aim to different targets,

71

including extra-STN areas, such as the Zona Incerta. 4-6 Therefore, it should come as little

72

surprise that several studies have failed to find an association between changes in

73

UPDRS scores and target coordinates.

74

between UPDRS scores and beta band oscillatory activity in the STN DLOR.

75

Therefore, the DLOR area, representing the motor sub-division of the STN, is believed to

76

be the preferred target for DBS implantation. 10 11

3

While

7

SC

RI PT

The optimal STN target is a subject of unresolved controversy.

Other studies have shown a strong association 8 9

M AN U

2

as validated by improvement in the UPDRS part III

By this reasoning, directing implantation of the lead contacts to the longest

78

recordable DLOR should be a central operative objective. Anatomical MRIs used for pre-

79

operative target and trajectory planning as well as Diffuse Tensor Imaging (DTI)

80

protocols are not yet routinely capable of differentiating between sub-territories of the

81

STN.

82

DLOR target thus avoiding suboptimal clinical results.

83

Factors affecting STN targeting are thought to include variable target selection between

84

different neurosurgeons,

85

(MER),

86

and awake MER-guided vs. asleep intraoperative Magnetic Resonance Imaging (iMRI) -

87

guided DBS implantation. 21-23 Some papers discuss trajectory planning in the context of

88

adverse effects,

Therefore, MER is still the only method to accurately detect and validate the

AC C

EP

12 13

TE D

77

14-16

3

effects of anesthesia on STN Microelectrode Recording

the type of imaging used for pre-op planning of target and trajectory,

24 25 26

17-20

due to proximity to blood vessels, eloquent brain areas and the

5

ACCEPTED MANUSCRIPT Tamir 89

ventricles, and others discuss trajectory planning for DBS targets other than STN, such as

90

Globus Pallidum internus (GPi)

91

investigated the effects of patient-specific characteristics on STN geometry as revealed

92

by intra-operative STN physiological recordings.

and Vim

28

. However, none of these studies have

RI PT

93

27

The STN, first characterized by Jules in 1865, is a lens shaped diencephalic structure. It contains around 430,000 neurons, comprising a volume of 100-130 mm3.

12

95

The STN lies between the Zona Incerta postero-dorsally, the Internal Capsule laterally

96

and the Substantia Nigra antero-ventrally. Its long axis extends from anterolateral (in its

97

dorsal portion) to posteromedial (in its ventral portion). Despite its preserved anatomy

98

across species, some variation in its postural angle is evident within and between patients.

99

Age has a particularly important effect on the anatomy of the STN in both healthy

100

human subjects as well as in PD patients. The volume of the STN as well as its neuronal

101

count decreases with age, with small variations between healthy human subjects.

102

distance of the STN from the mid-sagittal plane increases with age, becoming more

103

lateralized. The latter finding is observed in PD patients as well as in healthy controls. 30

29

The

TE D

M AN U

SC

94

In addition to age-related STN changes, other sub-cortical structures also tend to

105

change morphometrically with age. These include decreased thalamic height and

106

increased third ventricle width. In contrast, the AC-PC (Anterior Commissure – Posterior

107

Commissure) length seems to be stable within various age groups. 31 All these anatomical

108

alterations inevitably affect the way we target the STN and plan out trajectories.

109

Intuitively, they should also have impact on the recorded STN length, and may have an

110

effect on DLOR length as well. Brain atlases (Schaltenbrand-Warren and Schaltenbrand

AC C

EP

104

6

ACCEPTED MANUSCRIPT Tamir 111

Bailey) that were based on a limited, non-diverse patient population, and are still being

112

used for indirect targeting by some centers, should be viewed as inherently inaccurate. Driven by the finding that STN beta activity is directly correlated with the clinical

114

outcome of PD patients undergoing STN DBS, 8 and by a recent case report that indicated

115

a dramatic effect of trajectory angles on the side effects profile of the STN DBS,

116

investigated the possible contribution of patient-specific demographic and anatomic

117

factors on intra-operative STN physiology. We studied the contribution of three groups of

118

factors: 1) Demographic: patients’ age, gender, and PD duration. 2) Anatomic: third

119

ventricle length (AC-PC length), third ventricle width (both maximal width and width at

120

mid-commissural point; MCP), and laterality (right vs. left hemisphere, and first vs.

121

second operated side). 3) Trajectory-related: antero-posterior (ring) and medio-lateral

122

(arc) trajectory angles used to target the STN, and microelectrode location (relative to the

123

Microdrive’s BenGun central track).

RI PT

113

we

AC C

EP

TE D

M AN U

SC

32

7

ACCEPTED MANUSCRIPT Tamir Methods

125

Data Collection

126

This study was authorized and supervised by the Institutional Review Board (IRB) of

127

Hadassah Medical Center (reference code: 0168-10-HMO). The charts and operative

128

reports of all PD patients who had undergone STN DBS at the Hadassah Hebrew

129

university medical center in the years 2010-2015 were retrospectively reviewed. All

130

patients signed informed consent and release forms for participating in studies that

131

include analysis of data related to their clinical records and MERs.

SC

RI PT

124

Overall, 115 patient datasets were included in this study. Most patients were

133

implanted bilaterally, usually both sides in the same session or in some cases as a staged

134

procedure. “Trajectory” was defined as a single pass to the target, having a specific set of

135

trajectory angles (arc and ring). Routinely, we use two micro-electrodes in the BenGun,

136

thus all trajectories had two MER datasets: one located centrally in the BenGun, and the

137

other being either anterior (ventral) or posterior (dorsal) to the central microelectrode.

138

Most patients had one pass to each brain target (right or left). However, a few patients

139

required a second pass, due to unsatisfactory microelectrode recordings or clinical

140

response to macro-stimulation in the first one. In these cases, a new set of trajectory

141

angles might be used. Overall, 212 trajectories and 424 MER tracks were included in our

142

data analysis.

144 145 146

TE D

EP

AC C

143

M AN U

132

DBS surgery and MER

Surgery and MER techniques used are similar to those previously reported.

33

Briefly, stereotactic localization was performed based on fusion of pre-operative

8

ACCEPTED MANUSCRIPT Tamir Magnetic Resonance Imaging (MRI; either 1.5 or 3T axial and coronal slices using

148

2*0.4*0.4 mm isotropic T1 and T2 weighted MR sequences) and stereotactic Computed

149

Tomography (CT; performed on the morning of surgery, after fixation of a CRW

150

(Cosman Robert Wells) stereotactic frame (Radionics, Burlington, MA, USA)), using

151

Framelink 5 software (Medtronic, Minneapolis, USA).

RI PT

147

STN target coordinates were chosen based on direct visualization on the T2 MRI

153

(targeting for the posterior dorsolateral STN). Trajectories were planned to avoid

154

penetrating sulci, ventricles, blood vessels (as seen in T1 MRI with contrast media) and

155

also eloquent brain areas. In some cases, trajectory angles were slightly modified during

156

surgery, due to cortical blood vessels. Both the pre-operative targeting coordinates and

157

MER- defined STN borders used in the analysis were those originally used for planning

158

and treatment, and thus were blind to the goals of this study.

M AN U

SC

152

MER was used in all patients to identify the STN DLOR for implantation of the

160

DBS lead. MER was performed while the patients were awake (no sedation was given

161

prior to recordings), and off dopaminergic medications (for at least 12 hours). The

162

planned trajectory was used to insert two microelectrodes: one at the center and the

163

second two mm apart, either anterior (ventral) or posterior (dorsal) to the central

164

electrode. Overall, we had 212 central, 157 anterior and 55 posterior MER tracks. In all

165

cases, recordings started at 10 mm above the planned target. This starting point fitted our

166

targeting method in a way that we always encountered the STN at least a few millimeters

167

below our starting point. Recorded data was acquired using MicroGuide and Neuro-

168

omega systems (AlphaOmega Engineering, Nazareth, Israel). Neurophysiological activity

169

was recorded via polyamide coated tungsten microelectrodes (Alpha Omega) with

AC C

EP

TE D

159

9

ACCEPTED MANUSCRIPT Tamir impedances ranging between 0.3-0.9 megaOhms (measured at 1 kHz at the beginning of

171

each trajectory). The signal was amplified, band-passed from 250 to 6000 Hz (using a

172

hardware four-pole Butterworth filter), and sampled at 48 kHz by a 12- or 16-bit Analog

173

to Digital Converter.

RI PT

170

174

Spontaneous and evoked STN multi-unit MER activity was recorded. While in the

175

STN, the two electrodes were simultaneously advanced in small discrete steps of ∼0.1

176

mm.

177

as a sharp increase and decrease in the background activity, respectively. The STN

178

boundaries and the dorsolateral oscillatory region were further confirmed using a method

179

for automatic detection of STN and DLOR borders.

180

used to detect DLOR and STN borders are found in previous papers of our group. 33

SC

The STN entry and exit points were discerned visually by the neurophysiologist

M AN U

34

181 MER data analysis

Further details of the methods

TE D

182

8 33 35

For each MER track, we calculated the STN length, the DLOR length and the

184

fraction of the DLOR length out of STN length (%DLOR). Due to the insertion of two

185

parallel microelectrodes in the microdrive for every trajectory, each trajectory included

186

two distinct MER datasets, one from each microelectrode. Inasmuch as only the central

187

microelectrode was truly directed toward the preoperatively defined target (and to a 2

188

mm shifted target as the other non-central ones), we used only the central MER track for

189

all statistical analysis. The non-central MER track was used only when a comparison was

190

made between the central and the non-central MER tracks. In some trajectories, the

191

central microelectrode failed to record STN activity. These trajectories were omitted for

192

STN analysis (either averaging or correlations). Similarly, some central MER tracks

AC C

EP

183

10

ACCEPTED MANUSCRIPT Tamir 193

failed to record oscillatory activity despite good STN recordings. These trajectories were

194

omitted from the DLOR analysis. These “failed” STN and DLOR MER tracks were

195

analyzed separately in regards to possible differences in trajectory angles.

RI PT

196 Demographic and Anatomic Patients Data

198

Data extracted from the subjects’ medical records included: age at time of implantation,

199

gender, Parkinson’s disease duration, and laterality of implantation (right vs. left

200

hemisphere implantation).

SC

197

The AC-PC length, the AC-PC coordinates of the lead target (xyz), the two

202

trajectory angles (angle 1=arc, i.e., lateral to the mid sagittal plane; angle 2=ring, i.e.,

203

posterior to the anterior portion of the “axial” AC-PC plane), the length of the recorded

204

STN (by MER), and the order of the implanted leads (the side operated and implanted

205

first) were all extracted from the operative report. No correction was made for intra-

206

operative fine tuning of the trajectory angles due to cortical vessels. However, these

207

changes were very small (less than one degree) and unusual as most superficial vessels

208

can be detected and avoided in the pre-operative planning session.

209

The third ventricle length was defined as the AC-PC distance, as both AC and PC were

210

marked at the anterior and posterior borders of the third ventricle, respectively. In order

211

to unify the third ventricle length measurements, a single neurosurgeon (ZI) defined the

212

AC and PC location for all patients, extracting the AC-PC length.

213

Finally, the pre-op MRI (axial T2 and T1+Gadolinium protocol) was used in order to

214

measure third ventricular width. This was measured by a single neurosurgeon (IT), in two

215

different ways: as the maximal third ventricular width, and as the width at MCP. The two

AC C

EP

TE D

M AN U

201

11

ACCEPTED MANUSCRIPT Tamir 216

measurement methods were compared in relation to the other parameters to further

217

improve the reliability of these measurements.

218 Statistical Analysis

RI PT

219

A statistical software program SigmaStat (Ver. 4, Systat software Inc.; San Jose,

221

CA, USA) was used to determine potential interaction between age, gender, laterality,

222

third ventricular width and length, AC-PC coordinates, STN and DLOR length, and

223

trajectory angles.

SC

220

Each of the measured parameters mentioned above passed normality test

225

(Kolmogorov-Smirnov test). Then, multiple linear regression analysis was performed

226

with each of the trajectory angles as a function of age, gender, laterality, disease duration,

227

third ventricle width and length, STN and DLOR length and relative DLOR length (as a

228

% of STN length).

TE D

M AN U

224

As there were no obvious nonlinear relationships seen, a linear regression analysis

230

was chosen for ease of interpretation. The suitability of a linear fit was further determined

231

by the Constant Variance test and the Durbin Watson Statistic. Pearson’s r correlation

232

coefficient was used to determine the strength of correlation between the individual

233

factors (age, target coordinates, third ventricle width and length, disease duration, and the

234

two trajectory angles). Statistical significance was defined as p < 0.05.

AC C

EP

229

12

ACCEPTED MANUSCRIPT Tamir 235

Results

236

Patients’ clinical and anatomical characteristics Overall, 115 patients (74 males and 41 females) fitted the study inclusion criteria.

238

Table 1 summarizes the demographic and anatomic characteristics of these patients. The

239

average values are aligned with the numbers reported previously in the DBS and

240

anatomical literature.

241

(Table 2), despite a slightly younger age at surgery.

Interestingly, men had larger third ventricles than women

242

………..Table 1 near here …………..

M AN U

243 244 245

SC

31 36-40

RI PT

237

STN Target and trajectory coordinates

The AC-PC coordinates of the STN target are summarized in Table 2. Overall,

247

these values are in agreement with the Schaltenbrand-Wahren atlas STN coordinates as

248

well as other previous studies in the DBS literature.

249

lateralized STN than women (Table 2), in line with the finding of wider third ventricle in

250

men. Other AC-PC related coordinates were similar in men and women.

253 254

Interestingly, men had a more

EP

252

39 41

………..Table 2+3 near here …………..

AC C

251

TE D

246

The two trajectory angles are summarized in Table 3 (arc=angle 1, i.e., lateral to

255

the mid sagittal plane; ring=angle 2, i.e., posterior to the anterior portion of the “axial”

256

AC-PC plane). These angles were planned in a way to allow a maximal yet safe intra-

257

nuclear trajectory. Overall, they fit a traditional frontal bi-oblique approach to STN.

13

ACCEPTED MANUSCRIPT Tamir Figure 1 demonstrates an example of typical bilateral lead trajectories to the STN in a PD

259

patient. The trajectory avoids penetration of the ventricular system, crossing cortical and

260

subcortical blood vessels, eloquent brain areas, and the caudate nucleus. It sometimes

261

crosses the lateral border of the Thalamus.

RI PT

258

The 3389 Medtronic lead is implanted so that the second contact (contact 1,

263

contact 0 is the most ventral) will cover the center and lower half of the DLOR,

264

according to the MER data of the same patient. The volumetric segmentation of the STN

265

on top of T2 MRI was done using a prediction framework of subcortical structures based

266

on PD patients’ 7 Tesla MRI data base.

268

………..Figure 1 near here …………..

269 STN and DLOR recordings

TE D

270

M AN U

267

SC

262

The recorded STN lengths, DLOR length and %DLOR (fraction of DLOR length

272

out of STN length in percentage) for the central and non-central electrodes are

273

summarized in Table 4. Interestingly, the central electrode recorded on average

274

significantly longer STN and DLOR lengths (Table 4), while the %DLOR was not

275

significantly different between the two. In 35% of trajectories the recorded STN length

276

was longer in the non-central electrode, while in 37% of the trajectories the recorded

277

DLOR length was longer in the non-central electrode.

279

AC C

278

EP

271

………..Table 4 near here …………..

280

14

ACCEPTED MANUSCRIPT Tamir 281

Association between recorded DLOR length and STN length We found a strong and significantly positive correlation between the STN

283

recorded length and the DLOR recorded length (Fig. 2A). Omitting the MER tracks in

284

which no DLOR was identified did not significantly change this correlation (Fig. 2B).

RI PT

282

A significant positive correlation was found between the DLOR (as a percentile

286

fraction of the STN (%DLOR; see Methods) and the recorded STN length (Fig. 2C).

287

However, when omitting the MER epochs that failed to detect DLOR activity, no

288

correlation was found between STN length and %DLOR (Fig. 2D). To summarize, when

289

DLOR was detected in the MER, its length increased linearly with the recorded STN

290

length and the DLOR fraction remained unchanged.

M AN U

SC

285

Next, we addressed the clinical relevance of longer STN and DLOR trajectories in

292

DBS surgeries. We studied the correlation between the patients’ response to dopamine

293

(pre-operative off – pre operative on medication UPDRS) and the recorded STN and

294

DLOR lengths. We found a significant positive correlation between the right hemibody

295

UPDRS scores and the left brain recorded STN lengths (Figure 2E) and DLOR lengths

296

(Figure 2F). However, we were not able to detect similar significant correlations on the

297

opposite side.

EP

Previous studies of our group showed highly significant positive correlation

AC C

298

TE D

291

299

between the post-operative improvement in UPDRS (on vs. off medication and

300

stimulation) and the DLOR length (r=0.67, p<0.001).

301

longer STN trajectories will result in longer DLORs and better clinical outcome.

302 303

15

8

Accordingly, we suggest that

ACCEPTED MANUSCRIPT Tamir 304

………..Fig. 2 near here …………..

305 306

The effect of Trajectory angles on MER Due to the three dimensional (3D) complexity of the STN anatomy and trajectory

308

parameters, we chose to look at each trajectory angle separately. All correlation data and

309

significance are summarized in Table 5. We found a positive correlation between STN as

310

well as DLOR lengths and the arc angle, meaning that STN and DLOR lengths increase

311

as the trajectory becomes more laterally oriented. In contrast, the %DLOR was not

312

correlated with arc angle. STN length, but not DLOR length or %DLOR, was also

313

correlated with the ring angle. Together, these results mean that posterolateral trajectories

314

tend to have a longer recorded STN, and more laterally oriented trajectories (regardless of

315

the anteroposterior angle) also have a longer DLOR. However, the fraction of DLOR out

316

of STN was independent of trajectory angles.

TE D

M AN U

SC

RI PT

307

Trajectories in which the central microelectrode failed to record any STN activity

318

were slightly more anterior than trajectories that yielded STN recordings (Fig. 3A). No

319

difference was found in regards to laterality between successful STN trajectories and

320

failed ones (Fig. 3A). Interestingly, trajectories in which the central microelectrode failed

321

to identify DLOR were significantly more medial (Fig. 3A) and slightly more anterior

322

(Fig. 3A) than trajectories that succeeded in recording DLOR activity. Taken together,

323

these data suggest higher failure rates for STN and/or DLOR detection in anteromedial

324

trajectories. Still, the reported trajectory angles are within the standard trans-frontal

325

approach angles to the STN.

AC C

EP

317

326

16

ACCEPTED MANUSCRIPT Tamir 327

………..Table 5 near here …………..

328 329

According to previous studies, showing an enlarged lateral and third ventricle in 42 43

atrophied brains

we suspected lead trajectories in patients with brain atrophy to be

331

more laterally oriented. Surprisingly, there was no association between laterality of

332

trajectories and third ventricular width, or laterality of trajectories and laterality of STN

333

target (Table 5). On the other hand, we did observe an association between third

334

ventricular length and anteriorly displaced trajectories angles (Table 5).

SC

RI PT

330

We then combined the two trajectory angles by multiplication to create a vector

336

directed to the STN. This choice was based on the working assumption that postero-

337

lateral trajectories are more likely to parallel the long axis of the STN, and therefore more

338

likely to result in a longer pass of the electrode through the STN. In addition, the beta

339

oscillatory region of the STN is thought to be located on the dorso-lateral (postero-

340

superior) aspect of the STN.

TE D

M AN U

335

We found a positive correlation between the STN length and the trajectory vector

342

(Fig. 3B). However, when we omitted the MER tracks that failed to detect the DLOR, the

343

correlation between STN length and trajectory vector did not reach statistical significance

344

(Fig. 3C). Therefore, we suspected a positive correlation between STN length in failed

345

DLOR trajectories and trajectory vectors. Indeed, we found a significant correlation

346

between the two (Fig. 3D).

AC C

347

EP

341

Similarly, DLOR length was positively correlated with the trajectory vector when

348

failed DLOR trajectories were included (Fig. 3E), but not correlated when failed DLOR

349

trajectories were omitted from the correlation (Fig. 3F). Also DLOR fraction was

17

ACCEPTED MANUSCRIPT Tamir 350

positively correlated with trajectory vector when failed DLOR trajectories were included

351

(Fig. 3G), but not when they were omitted (Fig. 3H).

353

………..Fig. 3 near here …………..

354 355

RI PT

352

The effect of patient- and disease- specific characteristics on MER

Correlations of other variables that might contribute to the diversity of the STN

357

and DLOR recorded length between patients are presented in Fig. 4. These included

358

patient-specific clinical parameters, anatomy-related parameters, and 3D planning-related

359

parameters.

M AN U

SC

356

Regarding patient-specific parameters, the age of the patients at the time of

361

surgery was not correlated with either STN length (Fig. 4A) or DLOR length (Fig. 4B).

362

However, when omitting failed DLOR MER tracks, there was a trend towards negative

363

correlation between patients’ age and DLOR length (Fig. 4C). Similarly, disease duration

364

was not significantly correlated with either STN length (Fig. 4D) or DLOR length (Fig.

365

4E). Omitting the failed DLOR MER tracks did not change the lack of correlation

366

between DLOR length and disease duration (Fig. 4F).

EP

STN length and DLOR length were not significantly different between right and

AC C

367

TE D

360

368

left hemisphere recordings (Fig. 4G). Also the %DLOR was not significantly different

369

between right and left sides (43±27% vs. 43±25%, p = 0.33, respectively; Data not

370

shown). In addition, the STN and DLOR lengths and the %DLOR were not significantly

371

different between the side that was operated first (n=147 trajectories) or the second

372

operated side (n = 65 trajectories; includes only patients that were operated bilaterally on

18

ACCEPTED MANUSCRIPT Tamir the same day; STN length: 5.4 ± 1.6 mm vs. 5.24 ± 1.7 mm, respectively, p = 0.26;

374

DLOR length: 2.96 ± 1.2 mm vs. 2.88 ± 1.2 mm, respectively, p = 0.33; %DLOR: 52.9 ±

375

17.8% vs. 53.8 ± 19.4%, respectively, p = 0.38). However, the fraction of failed STN and

376

DLOR MER tracks (when considering only the central MER tracks) was slightly higher

377

on the second operated side than on the first one (failed STN: 9.23% vs. 7.48%; failed

378

DLOR: 13.4% vs. 11.0%, respectively).

RI PT

373

Finally, no differences were found in STN length and DLOR length between men

380

and women (Fig. 4H). Also %DLOR was not different between men and women (47.7%

381

vs. 45.7%, p = 0.6, respectively; Data not shown).

M AN U

382 383

………..Fig. 4 near here …………..

384

The effect of anatomy- specific characteristics on MER

TE D

385

SC

379

We next studied the contribution of ventricular size variability to the recorded

387

STN and DLOR lengths. According to the literature, the variability in human ventricular

388

system between patients is related to differences in brain volume, size, age and brain

389

atrophy. 31 42 43 Third ventricle width is a commonly used quantification measure, among

390

others, such as frontal horn index, Evans’ index, sulci width, and sub-arachnoid space

391

size.

392

ventricle width and length as our measures to explore atrophy-related variability in STN

393

targeting.

AC C

EP

386

31 42 43

Due to the proximity of the STN to the third ventricle, we chose third

19

ACCEPTED MANUSCRIPT Tamir 394

We found no correlation between either STN, or DLOR lengths and third

395

ventricle length (Fig. 5A and Fig. 5B, respectively). Omitting failed DLOR MER tracks

396

from the correlation analysis did not change these results (Fig. 5C). Third ventricle width measurements were not correlated with STN or DLOR

398

length, whichever one of the two measuring methods was used (Width at MCP: Fig. 5D;

399

maximal width: Fig. 5E, Maximal width vs. STN length and Maximal width vs. DLOR

400

length: Table 5). Omitting the failed DLOR MER tracks from the correlation in fig. 5E

401

did not alter the results (Fig. 5F).

SC

RI PT

397

Taken together, we conclude that in patients with atrophied brains an

403

“unintentional” lateral correction of the trajectory angle was used in the pre-op planning

404

in order to get a maximal intra-nuclear trajectory. Therefore, no marked differences in

405

STN or DLOR lengths were found in these patients.

M AN U

402

In order to specify which electrode position (either the central microelectrode

407

position or the non-central (anterior/posterior) electrode) better targets the STN, we

408

compared the length of STN and DLOR recorded in the central electrode (black bars)

409

with the ones recorded at the non-central electrodes (white bars). We found significantly

410

longer STN and DLOR trajectories at the central electrode compare to the non-central

411

one (Figure 5G). In addition, the fraction of the recorded DLOR out of STN was higher

412

in the central electrode compared to the non-central one (43 ± 26% vs. 33 ± 32%, p =

413

0.03).

EP

AC C

414

TE D

406

To further analyze this data, we divided the non-central electrode group into two

415

subgroups: anterior electrode position (used in our center until November 2014) and

416

posterior electrode position (subsequent to Nov 2014; see Methods). We found no

20

ACCEPTED MANUSCRIPT Tamir difference between the average length of recorded STN in these two groups (4.4 ± 2.2

418

mm vs. 4.9 ± 1.8 mm, respectively; n=142 and n=54, respectively; p = 0.13). Robust

419

changes in DLOR length (anterior electrodes: 1.89±1.5 mm; posterior electrodes:

420

2.72±1.6 mm; p = 0.002) and %DLOR (anterior electrodes: 40.34 ± 32%; posterior

421

electrodes: 52.73 ± 26%; p = 0.01) were however detected between these two groups, to

422

favor the posterior trajectories.

RI PT

417

SC

423 424

………..Fig. 5 near here …………..

426

M AN U

425

Anatomy considerations for targeting purposes

Brain atrophy is a prominent consideration in DBS surgery. Not only may the

428

cortical entry point be deeper to the skull burr-hole and the dural opening, also the brain

429

target might have distorted relations to its surroundings. To clarify this, we studied the

430

relationship between the AC-PC coordinates of the STN target (xyz) and the various

431

parameters mentioned earlier.

TE D

427

The x parameter (which represents the laterality of the target) was positively

433

correlated with the third ventricle width (Fig. 5H), and to a lesser extent also with the

434

third ventricle length (Fig. 5I). This means a more laterally positioned STN as the third

435

ventricle enlarges. Despite these results, we did not find any correlation between x and

436

patient age or PD duration (Table 5).

AC C

437

EP

432

The y parameter (represents the antero-posterior dimension) was slightly

438

correlated with age, but not correlated with disease duration, third ventricle length or

439

third ventricle width (Table 5).

21

ACCEPTED MANUSCRIPT Tamir 440

The z parameter (representing the rostro-caudal dimension) was significantly

441

correlated with disease duration and third ventricle width, but not with age or third

442

ventricle length (Table 5). We next studied the possible contribution of the targeting coordinates to the

444

variability of the STN and DLOR lengths. However, none of these multiple comparisons

445

yielded a significant effect of targeting on the MERs (Table 5).

RI PT

443

In order to identify a possible contribution of selected STN target coordinates to

447

trajectory failures, we compared AC-PC coordinates between MER tracks that

448

successfully recorded STN activity and those that failed to detect STN activity. However,

449

no significant differences were found between AC-PC target coordinates of the two sub-

450

groups (x: 11.96 ± 0.7 vs. 11.96 ± 0.6 mm, respectively, p = 0.5; Y: -3.19 ± 1.0 vs. 3.43 ±

451

0.5 mm, respectively, p = 0.18; Z: -4.54 ± 1.5 vs. -4.62 ± 0.7 mm, respectively, p = 0.4).

M AN U

SC

446

A similar comparison was conducted between successful DLOR identifying MER

453

tracks and DLOR failure sets. No differences were found in AC-PC coordinates between

454

the groups (x: 11.96 ± 0.7 vs. 11.82 ± 0.8 mm, respectively, p = 0.19; Y: -3.22 ± 1.0 vs. -

455

3.19 ± 1.0 mm, respectively, p = 0.47; Z: -4.59 ± 1.4 vs. 4.32±2.2 mm, respectively, p =

456

0.2).

AC C

EP

TE D

452

22

ACCEPTED MANUSCRIPT Tamir 457

Discussion This study shows for the first time that recorded STN DLOR (STN motor

459

domain) length is primarily determined by the STN length. The chosen trajectory angles

460

affect STN and DLOR lengths, favoring postero-lateral trajectories that tend to have

461

longer STNs and DLORs, while antero-medial ones tend to “fail” more often in detecting

462

STN and/or oscillatory (motor) activities. Nevertheless, atrophy-related changes in STN

463

geometry do not seem to alter the chosen trajectories. In patients with larger ventricles,

464

the STN target itself seems to shift laterally such that trajectory angles are not altered,

465

and the resulting STN and DLOR recorded lengths remain without significant change.

466

This scenario of enlarged ventricles seems to be more frequent in men than in women,

467

despite younger age and similar disease duration at surgery.

M AN U

SC

RI PT

458

468

DLOR-to-STN ratio is highest in the center of the STN and stable across different

470

trajectories

TE D

469

This study further provides compelling evidence that the beta oscillatory region of

472

the STN is located at the dorsal postero-lateral part of the nucleus. We show for the first

473

time that DLOR-to-STN proportion is stable across different trajectories. Therefore,

474

aiming for the longest STN trajectory should also result in a longer DLOR.

AC C

475

EP

471

Another interesting finding is the larger fraction of DLOR in the centrally-located

476

electrode in comparison with the peripherally (either anterior or posterior) located ones.

477

Since our target is planned to enter the center of the DLOR, these results suggest that the

478

oscillatory region is longer in the center of the postero-dorsolateral region of the STN

479

compared to the periphery of the nucleus. This is in agreement with a stronger beta signal

23

ACCEPTED MANUSCRIPT Tamir 480

(either power or coherence extracted from LFP signal or single unit analysis) in the

481

central trajectories, decreasing in strength from dorsal to ventral (along the recording

482

electrode path in the STN).

483

dorsolateral motor beta oscillatory activity toward the center of the dorsolateral STN

484

compared to its periphery. Therefore, in order to achieve maximal stimulation of the STN

485

motor area and inhibition of beta hyper-synchronized activity, the lead should be directed

486

to the center of the dorsolateral STN.

RI PT

Together, these data might suggest a higher density of

SC

44-46

This notion might be particularly important for future applications of multiple -

488

contact current steering electrodes and closed-loop DBS systems that rely on dorsolateral

489

STN beta power - derived algorithms for stimulus adaptation. 47 48 Such systems require a

490

reliable, strong and stable signal and a high signal to noise ratio. For this reason, precise

491

targeting of the DLOR is essential.

493

TE D

492

M AN U

487

STN and DLOR recorded lengths are larger in posterolateral trajectories Postero-lateral trajectories record longer STN and DLOR activity. Moreover,

495

these trajectories tend less to “fail”, in terms of detecting any STN and/or beta oscillatory

496

activity. This notion is true, despite significant variability of STN target coordinates

497

across patients. Most probably, postero-lateral trajectories follow the long axis of the

498

STN in the 3D space, therefore yielding longer recordings. However, DLOR geometry

499

might be more complex than that.

AC C

500

EP

494

Medial trajectories have been shown in the past to be less favorable due to worse 25

501

cognitive outcome, possibly due to penetration of the caudate nucleus.

502

these findings may also be attributed to a presumed longer pass through the ventromedial

24

Nevertheless,

ACCEPTED MANUSCRIPT Tamir ‘limbic/cognitive’ part of the STN when medial trajectories are used. Taken together with

504

our results, it might be beneficial in terms of targeting the oscillatory region, to aim for a

505

more postero-lateral trajectory. However, this should not, in any way, be considered as a

506

compensation for accidentally targeting the ventro-medial STN.

507 508

RI PT

503

Atrophy-related changes in STN geometry alter DBS target but not trajectory

An intuitive assumption that we had prior to this study was the tendency toward

510

more lateralized trajectory angles in older patients. We assumed older patients to have

511

larger ventricles, therefore forcing a more lateralized trajectory angle to the STN.

512

Surprisingly, we found that the trajectory angles were independent of ventricle size.

513

However, the STN target itself did show significant lateralization with increasing

514

ventricular size. This finding supports similar previous reports in the literature. 30 36 39 41

M AN U

SC

509

We could not find any association between patients’ age and ventricular size or

516

target lateralization. This is in contrast to recent studies showing that the STN target

517

becomes more lateral with age in PD patients, either by using pre-operative planning or

518

post-operative results to define actual STN target location.

519

factors, such as disease severity or age at diagnosis, may explain brain atrophy changes in

520

this patient population.

Other contributing

EP

30 49

AC C

521

TE D

515

Overall, our data led us to conclude that the lateralization shift of the STN target

522

compensates for the need of more lateralized trajectory angles. Due to the lateral shift of

523

the STN target and the more lateralized borders of the third ventricles, the distance

524

between these two intuitively should remain constant. This could be used as an extra

525

safety measure of target coordinates, but requires further study.

25

ACCEPTED MANUSCRIPT Tamir The laterality of the STN target in our study was also correlated with third

527

ventricle length, although to a lesser extent than width, meaning a more lateralized STN

528

target with longer third ventricles. The third ventricle length in general is a measure of

529

brain length and overall brain size 37 31 and is less likely to vary across patients, although

530

some studies have found such variability with age.

RI PT

526

We found men to have larger third ventricles than women (both length and

532

width), despite a slightly younger age at surgery. This finding is in accordance with

533

previous anatomical studies of normal brains revealing larger ventricles in men, measured

534

both directly and relative to brain size. 43 50

536 537

Study Importance and Limitations

M AN U

535

SC

531

These results emphasize the importance of using microelectrode spike and Local Field Potential (LFP) recordings to guide electrode implantation in DBS surgery for PD.

539

Nevertheless, this study has several limitations. Analysis of the electrophysiology yields

540

the “functional” borders of the STN, while the anatomical STN borders as per their

541

appearance on imaging were not directly studied. Also the 3D relations between the

542

electrode trajectory and the STN anatomical borders were not thoroughly studied.

EP

This study did not attempt to correlate patient outcomes (pre/ post- surgery and

AC C

543

TE D

538

544

on/off state - UPDRS scores), beta power and coherence with trajectory angles. In

545

addition, the results of this study have clinical implications only if the STN targeting

546

method used is directed to the DLOR, and if direct stimulation of the DLOR has clinical

547

benefit.

26

ACCEPTED MANUSCRIPT Tamir In addition, we cannot exclude the possibility that different parameters have

549

opposing effects. For example, brain atrophy may increase beta–band oscillatory activity

550

as part of disease progression, but also decrease STN volume and length. These

551

contradictory effects might obscure some of the correlations studied. Future studies

552

should try to overcome these obstacles.

RI PT

548

Finally, it is possible that minor differences in the recorded STN length are

554

negligible when inducing large electrical fields, especially with the current electrodes

555

available. Larger prospective studies, including post-operative UPDRS clinical data and

556

different lead types should be conducted in order to address these important questions.

M AN U

SC

553

AC C

EP

TE D

557

27

ACCEPTED MANUSCRIPT Tamir Acknowledgements

559

The authors thank Dr. Noam Harel (Center for Magnetic Resonance Research, University

560

of Minnesota Medical School, Minneapolis, MN, USA) for his kindness to provide us

561

with figure 1.

AC C

EP

TE D

M AN U

SC

RI PT

558

28

ACCEPTED MANUSCRIPT Tamir 562

References

563

1. Israel Z, Bergman H. Location, location, location: Validating the position of deep brain

564

stimulation electrodes. Movement Disorders. 2016; 31:259-259. 2. Welter ML, Schupbach M, Czernecki V, Karachi C, Fernandez-Vidal S, Golmard JL et

566

al. Optimal target localization for subthalamic stimulation in patients with

567

Parkinson disease. Neurology. 2014; 82:1352-1361.

RI PT

565

3. Hamel W, Koppen JA, Alesch F, Antonini A, Barcia JA, Bergman H et al. Targeting

569

of the subthalamic nucleus for deep brain stimulation: a survey among Parkinson's

570

disease specialists. World Neurosurg. 2016.

M AN U

SC

568

4. Burrows AM, Ravin PD, Novak P, Peters ML, Dessureau B, Swearer J et al. Limbic

572

and motor function comparison of deep brain stimulation of the zona incerta and

573

subthalamic nucleus. Neurosurgery. 2012; 70:125-130; discussion 130-121.

574

5. Plaha P, Ben-Shlomo Y, Patel NK, Gill SS. Stimulation of the caudal zona incerta is

575

superior to stimulation of the subthalamic nucleus in improving contralateral

576

parkinsonism. Brain. 2006; 129:1732-1747.

TE D

571

6. Blomstedt P, Fytagoridis A, Astrom M, Linder J, Forsgren L, Hariz MI. Unilateral

578

caudal zona incerta deep brain stimulation for Parkinsonian tremor. Parkinsonism

579

Relat Disord. 2012; 18:1062-1066.

AC C

EP

577

580

7. Nestor KA, Jones JD, Butson CR, Morishita T, Jacobson CEt, Peace DA et al.

581

Coordinate-based lead location does not predict Parkinson's disease deep brain

582

stimulation outcome. PLoS One. 2014; 9:e93524.

29

ACCEPTED MANUSCRIPT Tamir 583

8. Zaidel A, Spivak A, Grieb B, Bergman H, Israel Z. Subthalamic span of beta

584

oscillations predicts deep brain stimulation efficacy for patients with Parkinson's

585

disease. Brain. 2010; 133:2007-2021. 9. Beudel M, Oswal A, Jha A, Foltynie T, Zrinzo L, Hariz M et al. Oscillatory Beta

587

Power Correlates With Akinesia-Rigidity in the Parkinsonian Subthalamic

588

Nucleus. Mov Disord. 2016.

RI PT

586

10. Pozzi NG, Arnulfo G, Canessa A, Steigerwald F, Nickl R, Homola GA et al.

590

Distinctive neuronal firing patterns in subterritories of the subthalamic nucleus.

591

Clin Neurophysiol. 2016; 127:3387-3393.

M AN U

SC

589

592

11. Accolla EA, Dukart J, Helms G, Weiskopf N, Kherif F, Lutti A et al. Brain tissue

593

properties differentiate between motor and limbic basal ganglia circuits. Hum

594

Brain Mapp. 2014; 35:5083-5092.

12. Zwirner J, Mobius D, Bechmann I, Arendt T, Hoffmann KT, Jager C et al.

596

Subthalamic nucleus volumes are highly consistent but decrease age-dependently-

597

a combined magnetic resonance imaging and stereology approach in humans.

598

Hum Brain Mapp. 2017; 38:909-922.

EP

TE D

595

13. Horn A, Neumann WJ, Degen K, Schneider GH, Kuhn AA. Toward an

600

electrophysiological "sweet spot" for deep brain stimulation in the subthalamic

601

AC C

599

nucleus. Hum Brain Mapp. 2017.

602

14. Raz A, Eimerl D, Zaidel A, Bergman H, Israel Z. Propofol decreases neuronal

603

population spiking activity in the subthalamic nucleus of Parkinsonian patients.

604

Anesth Analg. 2010; 111:1285-1289.

30

ACCEPTED MANUSCRIPT Tamir 605

15. Krishna V, Elias G, Sammartino F, Basha D, King NK, Fasano A et al. The effect of

606

dexmedetomidine on the firing properties of STN neurons in Parkinson's disease.

607

Eur J Neurosci. 2015; 42:2070-2077. 16. Lee WW, Ehm G, Yang HJ, Song IH, Lim YH, Kim MR et al. Bilateral Deep Brain

609

Stimulation of the Subthalamic Nucleus under Sedation with Propofol and

610

Fentanyl. PLoS One. 2016; 11:e0152619.

RI PT

608

17. Chen SY, Tsai ST, Hung HY, Lin SH, Pan YH, Lin SZ. Targeting the subthalamic

612

nucleus for deep brain stimulation--a comparative study between magnetic

613

resonance images alone and fusion with computed tomographic images. World

614

Neurosurg. 2011; 75:132-137; discussion 122-134, 129-131.

M AN U

SC

611

18. Kocabicak E, Aygun D, Ozaydin I, Jahanshahi A, Tan S, Onar M et al. Does probe's

616

eye subthalamic nucleus length on T2W MRI correspond with microelectrode

617

recording in patients with deep brain stimulation for advanced Parkinson's

618

disease? Turk Neurosurg. 2013; 23:658-665.

TE D

615

19. Al-Helli O, Thomas DL, Massey L, Foltynie T, Limousin P, Holton JL et al. Deep

620

brain stimulation of the subthalamic nucleus: histological verification and 9.4-T

621

MRI correlation. Acta Neurochir (Wien). 2015; 157:2143-2147.

EP

619

20. Southwell DG, Narvid JA, Martin AJ, Qasim SE, Starr PA, Larson PS. Comparison

623

of Deep Brain Stimulation Lead Targeting Accuracy and Procedure Duration

624 625

AC C

622

between 1.5- and 3-Tesla Interventional Magnetic Resonance Imaging Systems: An Initial 12-Month Experience. Stereotact Funct Neurosurg. 2016; 94:102-107.

31

ACCEPTED MANUSCRIPT Tamir 626

21. Benazzouz A, Breit S, Koudsie A, Pollak P, Krack P, Benabid AL. Intraoperative

627

microrecordings of the subthalamic nucleus in Parkinson's disease. Mov Disord.

628

2002; 17 Suppl 3:S145-149. 22. Sterio D, Zonenshayn M, Mogilner AY, Rezai AR, Kiprovski K, Kelly PJ et al.

630

Neurophysiological refinement of subthalamic nucleus targeting. Neurosurgery.

631

2002; 50:58-67; discussion 67-59.

RI PT

629

23. Ostrem JL, Ziman N, Galifianakis NB, Starr PA, Luciano MS, Katz M et al. Clinical

633

outcomes using ClearPoint interventional MRI for deep brain stimulation lead

634

placement in Parkinson's disease. J Neurosurg. 2016; 124:908-916.

M AN U

SC

632

635

24. Pourfar MH, Mogilner AY. Lead Angle Matters: Side Effects of Deep Brain

636

Stimulation Improved With Adjustment of Lead Angle. Neuromodulation. 2016;

637

19:877-881.

25. Witt K, Granert O, Daniels C, Volkmann J, Falk D, van Eimeren T et al. Relation of

639

lead trajectory and electrode position to neuropsychological outcomes of

640

subthalamic neurostimulation in Parkinson's disease: results from a randomized

641

trial. Brain. 2013; 136:2109-2119.

EP

TE D

638

26. Le Goff F, Derrey S, Lefaucheur R, Borden A, Fetter D, Jan M et al. Decline in

643

verbal fluency after subthalamic nucleus deep brain stimulation in Parkinson's

644 645

AC C

642

disease: a microlesion effect of the electrode trajectory? J Parkinsons Dis. 2015;

5:95-104.

646

27. Sharim J, Yazdi D, Baohan A, Behnke E, Pouratian N. Modeling Laterality of the

647

Globus Pallidus Internus in Patients With Parkinson's Disease. Neuromodulation.

648

2016.

32

ACCEPTED MANUSCRIPT Tamir 649

28. Moscovich M, Morishita T, Foote KD, Favilla CG, Chen ZP, Okun MS. Effect of

650

lead trajectory on the response of essential head tremor to deep brain stimulation.

651

Parkinsonism Relat Disord. 2013; 19:789-794. 29. Zwirner J, Mobius D, Bechmann I, Arendt T, Hoffmann KT, Jager C et al.

653

Subthalamic nucleus volumes are highly consistent but decrease age-dependently-

654

a combined magnetic resonance imaging and stereology approach in humans.

655

Hum Brain Mapp. 2016.

SC

RI PT

652

30. Pereira JL, B AS, Sharim J, Yazdi D, DeSalles AA, Pouratian N. Lateralization of the

657

Subthalamic Nucleus with Age in Parkinson's Disease. Basal Ganglia. 2016;

658

6:83-88.

M AN U

656

31. Wang XY, Zhao L, Yu T, Qiao L, Ni DY, Zhang GJ et al. Assessment of Age-

660

Related Morphometric Changes of Subcortical Structures in Healthy People

661

Using Ultra-High Field 7 Tesla Magnetic Resonance Imaging. Front Aging

662

Neurosci. 2016; 8:224.

664

32. Pourfar MH, Mogilner AY. Lead Angle Matters: Side Effects of Deep Brain Stimulation Improved With Adjustment of Lead Angle. Neuromodulation. 2016.

EP

663

TE D

659

33. Valsky D, Marmor-Levin O, Deffains M, Eitan R, Blackwell KT, Bergman H et al.

666

Stop! border ahead: Automatic detection of subthalamic exit during deep brain

667

AC C

665

stimulation surgery. Mov Disord. 2016.

668

34. Shamir RR, Zaidel A, Joskowicz L, Bergman H, Israel Z. Microelectrode Recording

669

Duration and Spatial Density Constraints for Automatic Targeting of the

670

Subthalamic Nucleus. Stereotactic and Functional Neurosurgery. 2012; 90:325-

671

334.

33

ACCEPTED MANUSCRIPT Tamir 672

35. Zaidel A, Spivak A, Shpigelman L, Bergman H, Israel Z. Delimiting subterritories of

673

the human subthalamic nucleus by means of microelectrode recordings and a

674

Hidden Markov Model. Mov Disord. 2009; 24:1785-1793. 36. Zhu XL, Hamel W, Schrader B, Weinert D, Hedderich J, Herzog J et al. Magnetic

676

resonance imaging-based morphometry and landmark correlation of basal ganglia

677

nuclei. Acta Neurochir (Wien). 2002; 144:959-969; discussion 968-959.

679

37. den Dunnen WF, Staal MJ. Anatomical alterations of the subthalamic nucleus in

SC

678

RI PT

675

relation to age: a postmortem study. Mov Disord. 2005; 20:893-898. 38. Fiandaca MS, Salegio EA, Yin D, Richardson RM, Valles FE, Larson PS et al.

681

Human/nonhuman primate AC-PC ratio--considerations for translational brain

682

measurements. J Neurosci Methods. 2011; 196:124-130.

M AN U

680

39. Houshmand L, Cummings KS, Chou KL, Patil PG. Evaluating indirect subthalamic

684

nucleus targeting with validated 3-tesla magnetic resonance imaging. Stereotact

685

Funct Neurosurg. 2014; 92:337-345.

TE D

683

40. Hrabovsky D, Balaz M, Rab M, Feitova V, Hummelova Z, Novak Z et al. Factors

687

responsible for early postoperative mental alterations after bilateral implantation

688

of subthalamic electrodes. Br J Neurosurg. 2016; 1-5.

EP

686

41. Daniluk S, K GD, Ellias SA, Novak P, Nazzaro JM. Assessment of the variability in

690

the anatomical position and size of the subthalamic nucleus among patients with

691 692 693 694

AC C

689

advanced Parkinson's disease using magnetic resonance imaging. Acta Neurochir

(Wien). 2010; 152:201-210; discussion 210.

42. Celik HH, Gurbuz F, Erilmaz M, Sancak B. CT measurement of the normal brain ventricular system in 100 adults. Kaibogaku Zasshi. 1995; 70:107-115.

34

ACCEPTED MANUSCRIPT Tamir 695 696

43. LeMay M. Radiologic changes of the aging brain and skull. AJR Am J Roentgenol. 1984; 143:383-389. 44. Alavi M, Dostrovsky JO, Hodaie M, Lozano AM, Hutchison WD. Spatial extent of

698

beta oscillatory activity in and between the subthalamic nucleus and substantia

699

nigra pars reticulata of Parkinson's disease patients. Exp Neurol. 2013; 245:60-71.

700

45. Lourens MA, Meijer HG, Contarino MF, van den Munckhof P, Schuurman PR, van

701

Gils SA et al. Functional neuronal activity and connectivity within the

702

subthalamic nucleus in Parkinson's disease. Clin Neurophysiol. 2013; 124:967-

703

981.

M AN U

SC

RI PT

697

704

46. Feng H, Zhuang P, Hallett M, Zhang Y, Li J, Li Y. Characteristics of subthalamic

705

oscillatory activity in parkinsonian akinetic-rigid type and mixed type. Int J

706

Neurosci. 2016; 126:819-828.

47. West T, Farmer S, Berthouze L, Jha A, Beudel M, Foltynie T et al. The Parkinsonian

TE D

707

Subthalamic

Network:

Measures

of

Power,

Linear,

and

Non-linear

709

Synchronization and their Relationship to L-DOPA Treatment and OFF State

710

Motor Severity. Front Hum Neurosci. 2016; 10:517.

EP

708

48. Bour LJ, Lourens MA, Verhagen R, de Bie RM, van den Munckhof P, Schuurman PR

712

et al. Directional Recording of Subthalamic Spectral Power Densities in

713 714

AC C

711

Parkinson's Disease and the Effect of Steering Deep Brain Stimulation. Brain Stimul. 2015; 8:730-741.

715

49. Keuken MC, Bazin PL, Schafer A, Neumann J, Turner R, Forstmann BU. Ultra-high

716

7T MRI of structural age-related changes of the subthalamic nucleus. J Neurosci.

717

2013; 33:4896-4900.

35

ACCEPTED MANUSCRIPT Tamir 718

50. Gyldensted C, Kosteljanetz M. Measurements of the normal ventricular system with

719

computer tomography of the brain. A preliminary study on 44 adults.

720

Neuroradiology. 1976; 10:205-213.

RI PT

721

AC C

EP

TE D

M AN U

SC

722

36

ACCEPTED MANUSCRIPT Tamir Figure Legends

724

Figure 1. 3D relationships between DBS lead trajectory and contacts to STN, and

725

surrounding structures. The STN (in green) was volumetrically segmented using a

726

prediction framework of subcortical structures based on PD patients’ 7 Tesla MRI

727

database. The electrode contacts (dark gray) were reconstructed from the post-operative

728

CT fused to the 3 Tesla pre-op MRI. The second contact (counting from the lead tip) was

729

aimed to the center of the DLOR. Therefore, the first contact covers the ventral STN, the

730

third contact was on the dorsolateral border of the STN, and the fourth contact was

731

outside of the STN, at the ZI (Zona Incerta). The intra-nuclear path was planned to be as

732

long as possible in the dorsolateral area of the STN, ideally > 6 mm inside the STN. Most

733

patients were set on monopolar programming configuration, at the second contact.

734

Bottom left (3D cube) – H=horizontal (axial) view, P=posterior (coronal) view. The

735

picture is in courtesy of Noam Harel (see Acknowledgements).

AC C

EP

TE D

M AN U

SC

RI PT

723

37

ACCEPTED MANUSCRIPT Tamir Figure 2. Longer intra-nuclear trajectories have longer recorded DLOR, but similar

737

DLOR fraction. A. A plot of the relationship between the recorded DLOR length and the

738

recoded STN length in 195 trajectories in 115 patients. This graph precludes trajectories

739

in which no STN activity was found. Linear regression analysis showed a significant

740

correlation between the two parameters (Pearson’s r = 0.67; p< 0.0001). B. A plot of the

741

recorded DLOR vs. the recorded STN lengths. This graph excludes trajectories that failed

742

to detect both STN activity and DLOR activity (additional 23 trajectories). The

743

correlation was still significantly positive (r = 0.61; p < 0.0001). C. Linear regression

744

analysis of the percentile fraction of recorded DLOR from recorded STN length vs. the

745

STN length, in the same patient group as in A. Pearson’s r = 0.22, p < 0.0001. D.

746

Correlation of %DLOR and STN length, using the same patient group as in B. Omitting

747

the failed DLOR trajectories resulted a non-significant correlation (r = 0.02; p = 0.8). E.

748

Significant positive correlation of the patients’ right hemibody improvement of UPDRS

749

scores (pre-operative off – pre operative on medication UPDRS part III) and the left brain

750

recorded STN lengths (r=0.17, p=0.05). F. Significant positive correlation of the patients’

751

right hemibody improvement of UPDRS scores (pre-operative off – pre operative on

752

medication UPDRS part III) and the left brain recorded DLOR lengths (r=0.17, p=0.05).

SC

M AN U

TE D

EP

AC C

753

RI PT

736

38

ACCEPTED MANUSCRIPT Tamir Figure 3. Trajectory angles influences likelihood of long STN and DLOR

755

recordings. A. Comparison of trajectory angles between successful and failed STN

756

recording sets and between successful and failed DLOR sets. Failed STN recording sets

757

had slightly more anterior trajectories (57.38 ± 3.5 degrees vs. 59.76 ± 5.2 degrees, p =

758

0.08) but were not different in their laterality (arc: 20.64 ± 2.7 vs. 20.75 ± 2.8,

759

respectively; p = 0.89). Failed DLOR sets had slightly more antero-medial trajectories in

760

comparison to successful DLOR sets (arc: 19.61 ± 2.8 degrees vs. 20.95 ± 2.8 degrees, p

761

= 0.01; ring: 58.16 ± 4.7 vs. 59.92 ± 5.1 degrees, respectively; p = 0.07). B. The X axis

762

was truncated at 600 degrees^2 in all graphs. Positive correlation of the recorded STN

763

length (in mm) with the trajectory vector (measured as the multiplication of the two

764

trajectory angles (degrees^2); r = 0.22; p = 0.002). C. Same correlation as in B but

765

including only the MER sets that succeeded in finding the DLOR. This correlation was

766

not statistically significant (r = 0.03; p = 0.7). D. Positive correlation between STN

767

length and trajectory vector in MER sets that failed to detect the DLOR (r = 0.46; p =

768

0.05). E. Same correlation as in D but excluding failed DLOR MER sets did not find

769

significant correlation between DLOR length and trajectory vector (r = 0.05; p = 0.5). F.

770

Correlation graph of the recorded DLOR length with the trajectory vector. A significant

771

positive correlation was found between the two parameters (r = 0.21; p = 0.003). G.

772

Correlation graph of %DLOR vs. the trajectory vector showing a significant positive

773

correlation between the two parameters (r = 0.18, p = 0.01). H. Same correlation graph as

774

in G but excluding failed DLOR MER sets showed no significant correlation between

775

%DLOR and trajectory vector (r = 0.02; p = 0.8).

AC C

EP

TE D

M AN U

SC

RI PT

754

39

ACCEPTED MANUSCRIPT Tamir Figure 4. STN and DLOR lengths are not associated with patients’ age, PD

777

duration, gender or lead laterality. A. Linear regression analysis found no significant

778

correlation between the recorded STN length and the patient’s age (r = 0.05; p = 0.57). B.

779

No correlation was found between the recorded DLOR length and the patient’s age (r =

780

0.02; p = 0.8). C. Omitting the failed DLOR trajectories from the correlation presented in

781

B revealed a trend of negative correlation between patients’ age and DLOR length (r =

782

0.14; p = 0.08). D. No correlation was evident between STN length and the patients’

783

disease duration (r = 0.04; p = 0.6). E. No correlation was evident between DLOR length

784

and disease duration (r = 0.09; p = 0.25). F. Omitting the failed DLOR recording sets did

785

not change the lack of correlation between DLOR length and disease duration (r = 0.01; p

786

= 0.8). G. A bar graph presenting the differences in lengths of STN and DLOR as

787

recorded from the right hemisphere (black bars) and the left hemisphere (white bars).

788

Results are presented as mean ± SD. No significant difference was found (STN: 4.76 ±

789

2.0 vs. 5.06 ± 2.3, respectively. p = 0.3; DLOR: 2.33 ± 1.6 vs. 2.4 ± 1.5, respectively. p =

790

0.7). H. Similar bar graph presenting the differences in STN and DLOR lengths between

791

males (black bars) and females (white bars). No significant differences were found (STN:

792

5.42 ± 1.6 vs. 5.23 ± 1.9, respectively. p = 0.46; DLOR: 2.63 ± 1.4 vs. 2.45 ± 1.5,

793

respectively. p = 0.4).

AC C

EP

TE D

M AN U

SC

RI PT

776

40

ACCEPTED MANUSCRIPT Tamir Figure 5. Recorded STN and DLOR are longer in central electrode position, and are

795

not influenced by third ventricle anatomy. A. scatter graph showing no correlation

796

between STN length and third ventricle length (r = 0.11; p = 0.13). B. No correlation

797

between DLOR length and third ventricle length (r = 0.11; p = 0.15). C. Omitting the

798

failed DLOR MER sets from the correlation in B did not alter the resulted analysis (r =

799

0.1; p = 0.2). D. No correlation was found between STN length and third ventricle width,

800

measured at the MCP (r = 0.05; p = 0.5). E. No correlation was evident between DLOR

801

length and third ventricle width (r = 0.02, p = 0.7). F. Omitting the failed DLOR MER

802

sets from the correlation in E did not alter the resulted analysis (r = 0.03; p = 0.7). G. A

803

bar graph comparing the differences in length of recorded STN and DLOR between the

804

central electrode (black bars) and the non-central one (anterior or posterior; white bars).

805

Results are presented as mean ± SD. Significant differences were found in STN length

806

(5.06 ± 2.2 vs. 3.79 ± 2.5, respectively; p < 0.0001) as well as in DLOR length (2.4 ± 1.6

807

vs. 1.7 ± 1.7, respectively; p = 0.0002) between the two electrodes. H. The x coordinate

808

of the STN target was positively correlated with the third ventricle width (r = 0.69, p <

809

0.0001). I. The x coordinate of the target was also correlated with the third ventricle

810

length (r = 0.29, p = 0.0002).

AC C

EP

TE D

M AN U

SC

RI PT

794

41

ACCEPTED MANUSCRIPT

Table 1. Summarized Patients Demographics and Anatomical Characteristics.

(n = 115 patients) 62.04 ± 8.6 years

Disease Duration

9.43 ± 5.9 years

Third ventricle width (at MCP)

5.02 ± 1.8 mm

AC C

EP

TE D

Third ventricular length (AC-PC line)

M AN U

Maximal third ventricular width

SC

Age of Patients

RI PT

Mean ± SD

7.15 ± 2.0 mm

24.99 ± 1.4 mm

ACCEPTED MANUSCRIPT

Table 2. Averaged anatomical differences between men and women Men

Women

(n = 74)

(n = 41)

p value

RI PT

Parameter

25.3 ± 1.4 mm

24.5 ± 1.4 mm

3rd ventricle width

5.4 ± 1.6 mm

4.5 ± 2.0 mm

0.02*

60.77 ± 9.0 years

63.87 ± 7.5 years

0.08

11.6 ± 0.8 mm

0.0008†

STN |x| coordinate

12.1 ± 0.7 mm

M AN U

Age at surgery

SC

3rd ventricle length

0.006†

TE D

The numbers represent mean ± standard deviation

* denotes statistical significance (p < 0.05) when comparing central to non-central electrodes

AC C

EP

† denotes statistical significance (p < 0.01) when comparing central to non-central electrodes

ACCEPTED MANUSCRIPT

Table 3. Averaged target coordinates and trajectory angles Mean ± SD

RI PT

(n = 212 trajectories)

11.95 ± 0.74 mm

Y of Target

-3.2 ± 0.99 mm

Z of Target

-4.54 ± 1.51 mm

M AN U

SC

|X| of Target

Angle 1 (Arc)

20.82 ± 2.81 degrees 59.7 ± 5.09 degrees

AC C

EP

TE D

Angle 2 (Ring)

ACCEPTED MANUSCRIPT

Table 4. Averaged STN and DLOR lengths from MER

5.35±1.6 mm † (195)

DLOR length (not including failed DLOR)

2.94±1.2 mm * (172)

% DLOR (DLOR length*100/STN length)

53.17±18.2 % (172)

4.57±2.2 mm (185)

2.68±1.3 mm (148)

M AN U

SC

STN length (not including failed STN)

Non-central electrode (n)

RI PT

Central electrode (n)

55.06±24.9 % (148)

The numbers represent mean ± standard deviation

* denotes statistical significance (p < 0.05) when comparing central to non-central electrodes

AC C

EP

TE D

† denotes statistical significance (p < 0.01) when comparing central to non-central electrodes

ACCEPTED MANUSCRIPT

Table 5. Correlation results between various patients parameters Pearson’s r

p value

STN length

Arc angle

0.16

0.02*

DLOR length

Arc angle

0.19

0.01*

%DLOR

Arc angle

STN length

Ring angle

DLOR length

SC

RI PT

Independent Parameter

Dependent parameter

0.5

0.16

0.02*

Ring angle

0.04

0.6

%DLOR

Ring angle

0.08

0.3

Arc angle

3rd ventricle width

0.1

0.22

STNx

0.02

0.75

3rd ventricle length

0.18

0.02*

0.001

0.9

Max 3rd ventricle width

0.003

0.9

Patients age

0.11

0.16

STNx

PD duration

0.04

0.6

STNy

Patients age

0.15

0.05

TE D

Arc angle

EP

Ring angle

Max 3rd ventricle width

AC C

STN length DLOR length STNx

M AN U

0.04

PD duration

0.08

0.3

STNy

3rd ventricle length

0.04

0.6

STNy

3rd ventricle width

0.06

0.4

STNz

PD duration

0.3

0.0001†

STNz

3rd ventricle width

0.16

0.04*

STNz

Patients age

0.03

0.64

STNz

3rd ventricle length

0.07

0.3

STN length

STNz

0.06

0.42

DLOR length

STNz

0.04

0.53

STNx

0.1

0.19

STNx

0.01

0.89

STNy

0.06

0.45

STNy

0.06

0.43

Right UPDRS on-off

Left STN length

0.17

0.05*

Left UPDRS on-off

Right STN length

0.02

0.8

STN length

DLOR length

M AN U

TE D

AC C

DLOR length

EP

STN length

RI PT

STNy

SC

ACCEPTED MANUSCRIPT

ACCEPTED MANUSCRIPT

Left DLOR length

0.17

0.05*

Left UPDRS on-off

Right DLOR length

0.06

0.5

AC C

EP

TE D

M AN U

SC

RI PT

Right UPDRS on-off

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

AC C

EP

TE D

M AN U

SC

RI PT

ACCEPTED MANUSCRIPT

ACCEPTED MANUSCRIPT

Highlights Recorded STN DLOR length is primarily determined by the STN length.



Postero-lateral trajectories have longer STNs and DLORs.



DLOR-to-STN proportion is stable across different trajectories.



Atrophy-related changes in STN geometry do not alter the chosen trajectories.



In patients with larger ventricles, the STN target shifts laterally.



Central electrodes has larger fraction of STN DLOR than peripheral ones.

AC C

EP

TE D

M AN U

SC

RI PT



ACCEPTED MANUSCRIPT

Abbreviation List %DLOR = the fraction of the DLOR length out of STN length 3D = Three Dimentional

CRW = Cosman Robert Wells CT = Computed Tomography DBS = Deep Brain Stimulation

GPi = Globus Pallidum internus IC = Internal capsule

M AN U

DLOR = Dorso Lateral Oscillatory Region

iMRI = intraoperative Magnetic Resonance Imaging IRB = Institutional Review Board

TE D

LFP = Local Field Potential MCP = Mid Commissural Point

MER = Micro Electrode Recording

EP

MRI = Magnetic Resonance Imaging PC = Posterior Commissure

AC C

PD = Parkinson’s Disease

PSA = Posterior Subthalamic Area SN = Substrantia Nigra

STN = Sub Thalamic Nucleus UPDRS = United Parkinson’s Disease Rating Scale ZI = Zona Incerta

SC

RI PT

AC = Anterior Commissure

ACCEPTED MANUSCRIPT

Disclaimer

AC C

EP

TE D

M AN U

SC

RI PT

Conflict of interest: none.