Liberman et al
Thoracic
Prospective, multicenter, international phase 2 trial evaluating ultrasonic energy for pulmonary artery branch sealing in video-assisted thoracoscopic surgery lobectomy Moishe Liberman, MD, PhD,a Eric Goudie, MD,a Christopher Morse, MD,b Wael Hanna, MD, MBA,c Nathaniel Evans, MD,d Kazuhiro Yasufuku, MD, PhD,e and John Sampalis, PhD,f the VATS PA Study Working Group* ABSTRACT Objectives: The study objectives were to evaluate the immediate, short-, and medium-term efficacy and safety of pulmonary artery branch sealing using an ultrasonic vessel-sealing device in minimally invasive anatomic lung resection.
Results: A total of 150 patients with a minimum of 1 pulmonary artery branch sealed with an ultrasonic vessel-sealing device were prospectively enrolled in the trial. Resections included 139 lobectomies and 11 segmentectomies. A total of 424 pulmonary artery branches were divided: 239 with the ultrasonic vesselsealing device, 181 with endostaplers, and 4 with endoscopic clips. The mean and median pulmonary artery diameters were 4.7 mm/5.0 mm, 10.3 mm/ 10.0 mm, and 6.5 mm/6.5 mm for each method, respectively. Three of the pulmonary artery branches divided with the ultrasonic vessel-sealing device (1.3%) and 4 pulmonary artery branches divided with endostaplers (2.2%) bled intraoperatively. Among the patients with seal failures, 1 patient required conversion to thoracotomy. There was no postoperative bleeding from divided pulmonary artery branches with either sealing method. There was no mortality at 30 days. Conclusions: Pulmonary artery branch sealing with ultrasonic energy during video-assisted thoracoscopic surgery lobectomy is safe for vessels 7 mm or less. The use of an ultrasonic device is a reasonable sealing method for pulmonary artery branches 7 mm or less. (J Thorac Cardiovasc Surg 2019;-:1-11)
From the aUniversity of Montreal, Montreal, Quebec City, Canada; bMassachusetts General Hospital, Boston, Mass; cMcMaster University, Hamilton, Ontario, Canada; dJefferson University, Philadelphia, Pa; eUniversity of Toronto, Toronto, Ontario, Canada; and fMcGill University, Montreal, Quebec City, Canada. Funding was received through an Investigator Initiated Study Request for Proposals Grant from Ethicon Endosurgery, Johnson and Johnson, Cincinnati, Ohio. Clinical trial registry number: clinicaltrials.gov NCT02719717. Date and number of Institutional Review Board approval: CRCHUM IRB #15.302, April 11, 2016. Read at the 99th Annual Meeting of The American Association for Thoracic Surgery, Toronto, Ontario, Canada, May 4-7, 2019.
Lingular PA branch ultrasonic sealing in VATS lobectomy. Central Message PA branch sealing with ultrasonic energy during VATS lobectomy is safe for vessels of 7 mm or less. The use of an ultrasonic device is a reasonable sealing method for PA branches 7 mm or less.
THOR
Methods: This study consists of a prospective, phase 2, multicenter, international clinical trial (clinicaltrials.gov: NCT02719717) that enrolled patients planned for video-assisted thoracoscopic surgery/robotic anatomic lung resection in 7 centers (United States, Canada, United Kingdom). Pulmonary artery branches of 7 mm or less were sealed and divided with an ultrasonic energy vessel-sealing device. The remainder of the lobectomy was performed according to surgeon preference. Intraoperative, in-hospital, and 30-day postoperative bleeding and complications were prospectively recorded.
Perspective We performed a prospective, phase 2, multicenter, international clinical trial that evaluated ultrasonic PA branch sealing for vessels 7 mm or less. There was no difference in bleeding between ultrasonic-sealed vessels and vessels sealed with endostaplers. PA branch sealing with ultrasonic energy during VATS lobectomy is safe for vessels 7 mm or less.
See Commentary on page XXX.
* A list of complete VATS PA Study Working Group is shown in the Acknowledgments. Received for publication May 6, 2019; revisions received Aug 31, 2019; accepted for publication Sept 3, 2019. Address for reprints: Moishe Liberman, MD, PhD, Centre Hospitalier de l’Universite de Montreal, Centre de Recherche du CHUM-Room R04.402-1, 900 Rue SaintDenis, Montreal, Quebec, Canada H2X 0A9 (E-mail: moishe.liberman@ umontreal.ca). 0022-5223/$36.00 Copyright Ó 2019 by The American Association for Thoracic Surgery https://doi.org/10.1016/j.jtcvs.2019.09.061
The Journal of Thoracic and Cardiovascular Surgery c Volume -, Number -
1
Thoracic
Liberman et al
MATERIALS AND METHODS Abbreviations and Acronyms PA ¼ pulmonary artery RATS ¼ robotic-assisted thoracic surgery SD ¼ standard deviation VATS ¼ video-assisted thoracoscopic surgery
Scanning this QR code will take you to the article title page to access supplementary information. To view the AATS Annual Meeting Webcast, see the URL next to the webcast thumbnail.
THOR
Although video-assisted thoracoscopic surgery (VATS) lobectomy has been proven to be effective and safe in experienced hands, it is not devoid of risk. Intraoperative surgical complications can be catastrophic at times. The most important intraoperative complication is pulmonary vascular injury necessitating urgent conversion to open thoracotomy and even death.1 Published conversion rates range from 2% to 20%.2-4 In a Canadian study published in 2011 analyzing the causes of VATS conversion to thoracotomy in 237 cases, the conversion rate from VATS to open was 13.5%.1 Pulmonary artery (PA) injury alone constituted 37.5% of all conversions. Currently, despite being a safe and effective technique in experienced hands, anatomic pulmonary resections amenable to minimally invasive resections are still being performed by thoracotomy.5 Data from the National Surgical Quality Improvement Program database show that between 2007 and 2015 the incidence of thoracoscopy for lobectomies/segmentectomies increased from 11.6% to 60.6%. However, we do not know what proportion of those cases are true VATS approaches versus a ‘‘hybrid’’ approach consisting of a thoracotomy with the aid of a thoracoscope. The technical difficulty and danger of VATS lobectomy are directly related to PA branch manipulation, stapling, and division. PA manipulation is the main hesitation of many thoracic surgeons regarding the adoption of VATS lobectomy. If we can decrease the manipulation and dissection required on the PA branches, we can make these procedures safer and therefore more prevalent for anatomic pulmonary resections. Energy sealing devices have the potential to achieve adequate sealing of smaller PA branches with the advantage of having a smaller footprint than endostaplers. The objectives of this trial were to evaluate the immediate, short-, and medium-term efficacy and safety of PA branch sealing using an ultrasonic vessel-sealing device in minimally invasive anatomic lung resection. 2
This international phase 2 clinical trial evaluated the safety of an ultrasonic energy device for sealing and dividing pulmonary arteries 7 mm or less during minimally invasive anatomic lung resection (clinicaltrials. gov: NCT02719717) (Figure 1). There were 7 recruitment centers: 4 centers in Canada, 2 centers in the United States, and 1 center in the United Kingdom. The Institutional Review Board at the main study site (Centre de Recherche du Centre Hospitalier de l’Universite de Montreal – CRCHUM) and at the participating study sites approved this trial (CRCHUM IRB #15.302). Written informed consent was obtained for all participating patients.
Patient Population Patients who planned to undergo an anatomic lung resection (segmentectomy or lobectomy) by VATS or robotic-assisted thoracic surgery (RATS) were approached for enrollment in this clinical trial. Eligibility criteria were as follows: (1) ability to consent; (2) age more than 18 years; (3) nonhilar tumors; (4) preoperative imaging (chest computed tomography and positron emission tomography–computed tomography), nonsuggestive of hilar or mediastinal nodal disease; (5) invasive mediastinal staging requirement was based on current American College of Chest Physicians lung cancer staging criteria and was performed by any of the following tests, in appropriate patients, alone or in combination based on study site preference in accordance with American College of Chest Physicians guidelines: mediastinoscopy, mediastinotomy, VATS, endobronchial ultrasound, and endoscopic ultrasound. Exclusion criteria were as follows: (1) previous ipsilateral thoracic surgical procedure or trauma; (2) history of mediastinal or pulmonary irradiation; (3) anticoagulation with inability to stop anticoagulants before surgery; (4) systemic vascular disease or vasculitis; (5) uncorrectable coagulopathy; and (6) use of systemic steroids or immunosuppressive medication. Pulmonary hypertension was not an exclusion criteria because patients with pulmonary hypertension were shown to have higher bursting in previous studies.6
Surgical Technique The Harmonic ACEþ7 Shears with Advanced Hemostasis (ACEþ7; Ethicon, Cincinnati, Ohio) was the ultrasonic energy vessel-sealing device evaluated in this trial. Surgeons were instructed to perform the planned anatomic lung resection following their standard technique except for PA division. For the trial, all PA vessels that required division were dissected and measured in vivo using a sterile ruler. PA branches greater than 7 mm in diameter were divided using a vascular endostapler. Division and sealing of the lobar and segmental PA branches 7 mm or less in diameter were performed with the ACEþ7 (Video 1). PA vessels with diameters 1 to 5 mm were sealed on the Min setting with generator set at level ‘‘3’’, and vessels between 5 and 7 mm were sealed using the Advanced Hemostasis setting. The Harmonic ACEþ7 was used in both VATS and RATS approaches with the same technique for PA sealing. To seal and divide a PA branch with the ultrasonic energy device, surgeons were instructed to verify that the PA would lie flat inside the 2 blades of the instrument without any folding of the vessel wall before instrument activation. Only 1 activation of the instrument was used until the vessel was completely divided. Surgeons were also instructed to avoid putting tension on the vessel being sealed. Finally, surgeons were encouraged to leave a long stump of the PA being divided for safety purposes. No clips or sutures were added for reinforcement on PA stumps divided. To ensure uniformity of the PA sealing technique with the device, all participating surgeons had to follow a training module outlining all of the steps and technical aspect of the technique. The module also contained several videos with examples of proper sealing technique. The study credentialed surgeon investigator remained responsible to decide the suitability of PA branches for sealing with the device. This was assessed and
The Journal of Thoracic and Cardiovascular Surgery c - 2019
Liberman et al
Thoracic
150 Minimally Invasive Anatomical Lung Resections
PA ≤7 mm
239 PA branches divided with ultrasonic energy vessel sealing device
3 seal failures (1.3%)
PA >7 mm
181 PA branches divided with vascular endostapler
4 seal failures (2.2%)
424 pulmonary artery branches divided
Surgeon appreciation of the use of the ACE+7 for pulmonary artery branch sealing
Level of difficulty of procedure 80
40
Percentage
30 20
60 40 20
10 0
THOR
tra l S ffi om cu e lt w to ha us t Ve e ry di ffi cu lt us to e
eu
di
N
rd ha h uc
ry to eas us y e S ea om sy ew to ha us t e
er
r de ar H M
m
ila
rl di eve ffi l cu of lty
er si Ea Si
M
uc
h
ea
si
er
0
Ve
Percentage
50
Compared to your standard technique, did the use of the Harmonic ACE+7 make this procedure:
For PA branch sealing, the Harmonic ACE+7 was:
FIGURE 1. Top: In 150 minimally invasive lung resections, 424 PA branches were divided. PA branches 7 mm or less were divided with an ultrasonic energy vessel-sealing device and PA branches larger than 7 mm in diameter were divided with endostaplers. Seal failure rate was 1.3% for the ultrasonic energy vessel-sealing device and 2.2% for endostaplers (P ¼ .47). Bottom: Most surgeons reported that the use of the ultrasonic energy vessel-sealing device made the lung resection easier or much easier. The same survey revealed that most surgeons thought that the ultrasonic energy vessel-sealing device was very easy to use. PA, Pulmonary artery.
decided upon intraoperatively according to anatomy, vascular dissection, and patient-specific factors. Surgeons were not required to seal PA branches less than or equal to 7 mm with the device if they had a reason to believe it was not safe. Of the 15 surgeons participating in the trial, 3 had previous experience with the device for PA branch sealing during minimally invasive lung resections. All other surgeons learned the technique specifically for the trial. All procedures were filmed and recorded during PA manipulation and division for analysis in case of a seal failure and for monitoring purposes. Patients were included in the study analysis only if they underwent the study intervention: PA branch sealing using ACEþ7 of at least 1 PA branch during a minimally invasive anatomic lung resection. Conversion to thoracotomy before sealing a PA branch with the device also excluded patients from study analysis. Standard postoperative care was provided to all patients according to surgeon practice and preference. No additional monitoring or tests were necessary for patients in the postoperative period.
the initial operation that required direct intervention, such as compression, clip placement, or suture. Delayed seal failure was defined as bleeding from a PA branch sealed with the device in the first 30 postoperative days. It was assumed that postoperative bleeding from a PA branch would cause hemodynamic instability necessitating prompt investigations and reintervention in a patient, thus allowing study investigators to identify all primary outcomes. Secondary outcomes were overall postoperative complications according to the Clavien-Dindo Classification and 30-day mortality.7 The safety outcomes in this trial were any seal failure related to the study device, as described earlier, and any complication directly related to the use of the study device, such as a thermal injury. Efficacy in the trial was defined as the ability of the ACEþ7 to seal PA branches 7 mm or less in a practical manner for the surgeons. This was assessed with the ability of surgeons to perform the procedure as described in the protocol. Postoperative data were recorded prospectively during patient hospital stay. After hospital discharge, patients were followed in the outpatient postoperative clinic or with a phone call.
Study Outcomes The primary outcome of this trial was intraoperative and postoperative seal failure on a sealed PA branch with the ultrasonic energy device. Immediate seal failure was defined as bleeding from a divided PA branch during
Surgeon Questionnaire Immediately after each operation, surgeons were asked to fill out a short questionnaire to evaluate their impression of the use of the energy sealing
The Journal of Thoracic and Cardiovascular Surgery c Volume -, Number -
3
Thoracic
Liberman et al
device during the specific operation. The questionnaire was developed by the surgeons who had previous experience with the energy-sealing device. The questions were as follows: 1. Compared with your standard technique, did the use of the Harmonic ACEþ7 make this procedure (a) much easier, (b) easier, (c) similar level of difficulty, (d) harder, or (e) much harder. 2. For PA branch sealing, the Harmonic ACEþ7 was (a) very easy to use, (b) somewhat easy to use, (c) neutral, (d) somewhat difficult to use, or (e) very difficult to use. 3. In your opinion, did the Harmonic ACEþ7 Shears avert a conversion to thoracotomy in this case? Yes/No.
THOR
Enrollment
VIDEO 1. Left upper lobectomy with all PAs divided with an ultrasonic energy sealing device. Video available at: https://www.jtcvs.org/article/ S0022-5223(19)32065-3/fulltext.
Statistics This trial is evaluating precision and does not aim at testing a hypothesis. The sample size has been calculated on the basis of a 2-sided 99% confidence interval. We assumed a rate of seal failure of 2% at 30 days. This
363 assessed for eligibility
Allocation
270 allocated to intervention 150 received allocated intervention 120 did not receive allocated intervention: - All PA branch diameter >7 mm (n = 77) - Conversion to thoracotomy unrelated to ACE (N = 26) - Lobectomy not performed because of intraoperative findings (n = 7) - ACE not used for technical reasons (n = 10)
Follow-up
0 lost to follow-up
Analysis
93 excluded - Refused to participate (n = 40) - Not meeting inclusion criteria (n = 28) - Other reasons (n = 25)
150 included in analysis
FIGURE 2. Prospective, multicenter, international phase 2 trial evaluating ultrasonic energy for PA branch sealing in VATS lobectomy consort diagram. PA, Pulmonary artery.
4
The Journal of Thoracic and Cardiovascular Surgery c - 2019
Liberman et al
Thoracic
TABLE 1. Lobes resected and associated number of pulmonary artery vessels divided
Lobe
No.
Mean no. of PA branches divided
Median no. of PA branches divided
Total no. of PA branches divided with the Harmonic ACEþ7 (Ethicon, Cincinnati, Ohio)
Total no. of PA branches divided with staplers
Total no. of PA branches divided with clips
RUL
56
2
1
72
63
1
LUL
43
4
2
96
74
3
RLL
21
2
1
33
22
0
RML
11
2
1
16
4
0
LUL segmentectomy
9
2
1
9
9
0
LLL
7
2
1
8
7
0
RLL segmentectomy
2
2
2
4
0
0
1
3
1
RML and RLL Total
150
1
2
0
239
181
4
estimate is based on our previous experience with the technique and on published data.8-11 On the basis of this assumption, the absolute minimum number of patients was 85. However, allowing for subgroup analyses, we aimed at recruiting 150 patients. nQuery (Statistical Solutions Ltd, Cork, Ireland) was used for sample size calculation, and IBM SPSS Statistics (version 25; New York, NY) was used for descriptive statistics.
RESULTS Between July 2016 and November 2018, 270 patients were approached and consented for study enrollment (Figure 2). These patients were recruited across 7 sites and were planned for surgery with 1 of 15 board certified general thoracic surgeons with experience in minimally invasive lobectomy/segmentectomy. A total of 150 patients were included and represent the cohort for study analysis. In these patients, a minimum of 1 PA branch was sealed with the ultrasonic energy sealing device (ACEþ7) during VATS or RATS anatomic lung resection. Mean age was 66 years (range, 33-82 years). TABLE 2. Stage and type of resected lung tumors Stage 0 - in situ
No. of cases (%) 1 (0.7)
IA
83 (55)
IB
23 (15)
IIA
4 (2.6)
IIB
15 (10)
IIIA
4 (2.6)
IIIB
2 (1.3)
Other cancers Benign lesions Total
10 (6.7) 8 (5.3) 150
There were 93 female and 57 male patients. Mean body mass index was 26.6 kg/m2 (standard deviation [SD], 5.1; range, 16-42). A total of 46 patients (31%) were receiving aspirin therapy at the time of operation. Ten patients (6.7%) were receiving continuous anticoagulant therapy before surgery. Anticoagulants were stopped before surgery and resumed afterward according to site-specific standard clinical practice. A total of 139 lobectomies and 11 anatomic segmentectomies were performed in the trial (Table 1). A total of 128 resections were performed using a VATS approach, and 22 were performed using a RATS approach. Tumor stage and pathological findings are detailed in Tables 2 and 3. A total of 424 PA branches were divided: 239 with the ultrasonic vessel-sealing device, 181 with endostaplers, and 4 with endoscopic clips (Table 4). The mean and median PA diameters were 4.7 mm/5.0 mm, 10.3 mm/ 10.0 mm, and 6.5 mm/6.5 mm for each method, respectively. Twenty-nine branches 7 mm or less in diameter were not divided with the ACEþ7; 27 of those were divided with endostaplers and 2 with clips. Seal failure occurred in 7 cases: 3 after sealing with the ACEþ7 device and 4 after division with an endostapler (P ¼ .47). All seal failures occurred intraoperatively, and no patient experienced a bleed from a divided PA stump postoperatively. The first seal failure related to the use of the energy sealing device was on a 5-mm posterior ascending PA branch during a right upper lobectomy (Video 2). The branch had sealed initially; however, the corner of the stump started bleeding at the end of the operation because of suction trauma. A clip was applied on the PA stump to control the bleeding. Blood loss was 300 mL, and no blood transfusion was necessary. The second bleed related to the ACEþ7 was on a 5-mm anterior PA branch during a right upper lobectomy. The bleeding was difficult
The Journal of Thoracic and Cardiovascular Surgery c Volume -, Number -
5
THOR
PA, Pulmonary artery; RUL, right upper lobe; LUL, left upper lobe; RLL, right lower lobe; RML, right middle lobe; LLL, left lower lobe.
Thoracic
Liberman et al
TABLE 3. Histology of resected lung tumors Histology
No. of cases (%)
Lung adenocarcinoma
101 (67)
Squamous cell lung cancer
16 (11)
Lung adenocarcinoma and squamous cell lung cancer
1 (0.7)
Carcinoid lung tumor
13 (8.7)
Lung adenocarcinoma and small cell lung cancer
1 (0.7)
Colorectal adenocarcinoma
7 (4.7)
Urothelial carcinoma
1 (0.7)
Testicular teratoma
1 (0.7)
B-cell lymphoma
1 (0.7)
Benign lung lesion
8 (5.3)
Total
150
THOR
to control during thoracoscopy; therefore, conversion to thoracotomy was necessary to allow for suture of the PA stump that controlled the bleeding. Blood loss was 550 mL, and no blood transfusion was necessary. The third bleed with ultrasonic sealing was on a 6-mm anterior basal artery during a left lower lobectomy. The bleeding was coming from the corner of the PA stump immediately after the energy device was applied and necessitated a single clip placement. No conversion to thoracotomy was necessary, blood loss was 50 mL, and no transfusion was required. Bleeding associated with endostaplers occurred on PA branches of 7, 7, 8, and 12 mm. In all cases, the seal failure was partial and visible pulsatile bleeding was seen on a portion of the staple line. All necessitated clip placement to control the bleeding, and no conversion to thoracotomy or blood transfusion was necessary. There was no specific cause for seal failures associated with the endostaplers. In our experience, the mismatch between staple size and vessel wall thickness can explain the incomplete seal of a PA branch with endostaplers. Two other patients required conversion to thoracotomy after the PA branches had been sealed with the energy sealing device. One conversion was necessary for a bronchoplasty due to left main bronchus injury during left upper lobectomy. The second conversion was during a right lower
VIDEO 2. Right upper lobectomy; bleeding from a posterior ascending PA branch after ultrasonic energy sealing and division. Video available at: https://www.jtcvs.org/article/S0022-5223(19)32065-3/fulltext.
lobectomy. An angioplasty was necessary because of tumor invasion into the main PA. Neither of these conversions were related to the use of the energy device. Mean operative time was 126 minutes (SD, 79 minutes). Mean blood loss was 109 mL (SD, 119 mL). Median chest tube removal day was postoperative day 3 (range, 121 days). Mean and median length of stay were 4.1 days (SD, 3.1 days) and 3.8 days (range, 1-30 days), respectively. Table 5 describes the postoperative complications in the trial. One patient required reoperation because of bleeding from a bronchial artery. At the time of reoperation, all the PA stumps were carefully examined, and it was confirmed that no bleeding was coming from any PA branch. All patients were followed up until postoperative day 30. There was no mortality at 30 days. A total of 148 questionnaires were answered by 15 surgeons with a response rate of 98.7%. Figure 3 demonstrates surgeon perception of the utility of the ultrasonic energy sealing device on the procedure compared with the surgeon’s standard approach. In 74.3% of cases, surgeons declared that the device made the surgical procedure much easier (n ¼ 72) or easier (n ¼ 38) compared with their standard practice. Figure 4 demonstrates surgeon perception regarding the ease of use of the energy sealing device during the procedure. In 91.2% of cases, surgeons declared
TABLE 5. Postoperative complications Complications
TABLE 4. Divided pulmonary artery characteristics PA division method (no. of PAs) Ultrasonic energy (239) Endostapler (181) Clip (4) PA, Pulmonary artery.
6
Mean diameter (mm)
Median diameter (mm)
Range (mm)
4.7
5.0
1-7
10.3
10.0
3-24
6.5
6.5
3-10
No. of cases (%)
Prolonged air leak
23 (15.3)
Atrial fibrillation
10 (6.7)
Pneumonia
9 (6.0)
Pneumothorax
4 (2.7)
Delirium
4 (2.7)
Urinary tract infection
2 (1.3)
Pulmonary embolism
1 (0.7)
Hemothorax
1 (0.7)
The Journal of Thoracic and Cardiovascular Surgery c - 2019
Liberman et al
Thoracic
TABLE 6. Recommended technical and safety points for the use of the ultrasonic energy sealing device on pulmonary arteries 7 mm or less
Level of difficulty of procedure
Technical and safety points for the use of the ACEþ7 on PA branches
40 30 20 10 er
r
rd
de
ha
ar uc
h
H M
ila
rl di eve ffi l cu of lty
r ie Ea s Si m
M
uc
h
ea
si
er
0
Compared to your standard technique, did the use of the Harmonic ACE+7 make this procedure: FIGURE 3. Surgeon perception of the utility of the ultrasonic energy sealing device during the procedure compared with their standard approach. After each case using the energy sealing device, surgeons were asked to answer the following question: Compared with your standard technique, did the use of the Harmonic ACEþ7 (Ethicon, Cincinnati, Ohio) make this procedure (a) much easier, (b) easier, (c) similar level of difficulty, (d) harder, or (e) much harder.
that the device was very easy to use (n ¼ 112) or somewhat easy to use (n ¼ 23). Finally, in 18% of cases, surgeons subjectively thought that the use of the energy sealing device during minimally invasive anatomic lung resection prevented conversion to thoracotomy.
Surgeon appreciation of the use of the ACE+7 for pulmonary artery branch sealing
Percentage
80 60 40 20
tra l di S ffi om cu e lt w to ha us t Ve e ry di ffi cu lt us to e
N eu
Ve ry to eas us y e So ea m sy ew to ha us t e
0
For PA branch sealing, the Harmonic ACE+7 was: FIGURE 4. Surgeon perception regarding the ease of use of the energy sealing device during the procedure. After each case using the energy sealing device, surgeons were asked to answer the following question: For PA branch sealing, the Harmonic ACEþ7 was (a) very easy to use, (b) somewhat easy to use, (c) neutral, (d) somewhat difficult to use, or (e) very difficult to use. PA, Pulmonary artery.
Use on arteries 7 mm. Apply on well-dissected arteries. Use proper instrument function: Min for arteries 5 mm and advanced hemostasis for arteries >5 mm. Only activate the sealing device once and allow sufficient time for complete sealing and cutting. Place the artery flat inside the instrument in the middle of the blades. Avoid touching surrounding tissues with the instrument tip during or after activation to avoid thermal trauma. Leave a stump long enough for clip placement if a seal failure occurs. Avoid putting tension on the PA while sealing. PA, Pulmonary artery.
DISCUSSION A minimally invasive approach is the standard of care for anatomic lung resections for stage I and II lung cancer.12 The incidence of VATS and RATS lobectomy has increased consistently over the last decade.5 Several factors may have contributed to this increase, including the availability of instruments adapted to VATS/RATS to ease dissection, manipulation, and division of hilar structures.13,14 This phase 2 trial aimed specifically at evaluating the safety of an ultrasonic energy sealing device for sealing PA branches during thoracoscopic anatomic lung resections. The current trial evaluated a device that has already been commercialized; however, there is a lack of safety data for its use on pulmonary vasculature because safety studies on energy devices are typically performed on systemic arteries. Therefore, this trial is the result of a comprehensive step-bystep approach carried out by our group to demonstrate the efficacy and safety of energy sealing for PA vessels 7 mm or less. Ex vivo studies evaluated burst pressures of PA branches sealed with different energy vessel-sealing devices and with standard endostaplers.6,15 Results showed that PAs sealed with ultrasonic energy can sustain higher pressures than PAs sealed with advanced bipolar energy devices or endostaplers. In an animal survival study, 10 lobectomies using the device were completed in 10 dogs without any bleeding complications at 30 days.9 Two phase 1 trials evaluated the safety of ultrasonic energy for PA sealing in open and VATS lobectomy.10,11 These 2 trials allowed us to obtain safety data to support the current phase 2 multicenter trial. There were 3 seal failures with the ACEþ7. In the first case, the PA stump started bleeding after a suction rubbed the stump. This complication can be avoided by not touching the freshly sealed PA stumps. The second bleed associated with the ACEþ7 was during a right upper lobectomy on a 5-mm branch. No specific cause was identified to
The Journal of Thoracic and Cardiovascular Surgery c Volume -, Number -
7
THOR
Percentage
50
Thoracic
THOR
explain the seal failure. The surgeon hypothesized that tension on the PA during activation could have caused premature cutting of the branch. The third case in which bleeding occurred was on a 6-mm branch during a left lower lobectomy. Retrospectively, we believe this seal failure can be explained by the positioning of the PA inside the instrument. The PA branch was wedged into the crotch of the instrument jaws, and sealing at this location is not as effective because the vessel may be folded. The 2 blades do not have full contact in the crotch, and surgeons should avoid placing the PA inside the ‘‘V’’ at the instrument. The 3 seal failures related to the energy sealing device all occurred intraoperatively and were managed with minimal complications. There was no difference in failure rates between energy sealed vessels and endostaplers in this trial. There were no postoperative seal failures in either group. We believe that technical details are important to ensure the safe use of this instrument (Table 6). PA dissection is necessary to safely apply the blades of the ACEþ7 around the PA branch. This avoids having other tissues (lung, lymph nodes) between the vessel wall and the instrument, which can compromise the quality of the seal. Also, sufficient dissection allows keeping a safe distance between the tip of the instrument and the surrounding tissues. At the end of activation, the tip can reach temperatures up to 200 C.16 In circumstances where only limited clearance of the main PA (<2 mm) can be obtained, some other form of vascular division should be considered. In the current trial, no thermal injury caused any clinically significant complication. Proper apposition of the PA branch inside the blades of the instrument is important. The vessel needs to lie flat because folds in the vessel wall can compromise the quality of the seal. Also, proper sealing is best achieved in the middle of the blades. Positioning the artery in the crotch of the instrument should be avoided. We recommend leaving at least a few millimeters of stump to allow easier vascular control in case of sealing failure. The ACEþ7 has 3 buttons: maximum (Max), minimum (Min), and advanced hemostasis. We do not recommend using the Max function on PA branches because it spends less time sealing the tissue and cuts the tissue faster than other functions. However, some authors suggest the Max function is sufficient for PA sealing.8 The Min function is the appropriate function to seal and divide named blood vessels of 5 mm or less. This function spends more time sealing tissues before dividing them. Finally, the advanced hemostasis function spends more time sealing vessel walls before dividing the tissue. Advanced hemostasis is appropriate for vessels between 5 and 7 mm. On the basis of our experience, we advise avoiding tension or torsion on the PA branch while the instrument is activated. This avoids premature division of the PA before a complete seal is achieved. Because of the device profile and small tip size, the surgeon’s stress level when getting around small PA branches 8
Liberman et al
may be decreased. In a Japanese study, surgeon stress was evaluated with a visual analog scale and was found to be significantly less when using an energy device compared with an endostapler for PA divison.13 To our knowledge, the current trial is the largest prospective human trial evaluating an ultrasonic energy sealing device for PA sealing. In 2016, White and colleagues8 performed a retrospective study looking at the use of the Harmonic ACE for PA and vein sealing. They divided 97 pulmonary arteries and 21 veins in 82 lobectomies. There was 1 seal failure. Likewise, for a seal failure in this trial, their explanation for the seal failure seemed to be related to the positioning of the PA branch within the jaws of the instrument. Other publications report on the use of advanced bipolar energy sealing devices.17-21 In our opinion, there are not enough safety data available for these devices compared with ultrasonic devices for PA sealing. In 1 ex vivo study evaluating the burst pressures of pulmonary arteries sealed with different devices, some PA stumps did not seal at all with advanced bipolar devices.6 We advise caution on applying the results of this trial to other energy devices that are commercially available. Sufficient safety data should be obtained for a specific device before its routine use can be recommended for the pulmonary vasculature. CONCLUSIONS PA branch sealing with an ultrasonic energy device during VATS lobectomy is safe for vessels 7 mm or less. There was no difference in bleeding between ultrasonic-sealed vessels and vessels sealed with endostaplers. The use of an ultrasonic device is a reasonable sealing method for small PA branches. Webcast You can watch a Webcast of this AATS meeting presentation by going to: https://aats.blob.core.windows.net/media/ 19%20AM/Saturday_May4/203BD/203BD/S14%20-%20 Looking%20to%20the%20future/S14_5_webcast_11105 9219.mp4.
Conflict of Interest Statement Authors have nothing to disclose with regard to commercial support. VATS PA Study Working Group members: R Malthaner (University of Western Ontario, London, Ontario, Canada); J Dunning
The Journal of Thoracic and Cardiovascular Surgery c - 2019
Liberman et al
Thoracic
(James Cook University Hospital, Middlesbrough, United Kingdom); E Lafontaine (University of Montreal, Montreal, Quebec, Canada); P Ferraro (University of Montreal, Montreal, Quebec, Canada); C Wright (Massachusetts General Hospital, Boston, Mass); H Auchincloss (Massachusetts General Hospital, Boston, Mass); M Lanuti (Massachusetts General Hospital, Boston, Mass); J Martin (University of Montreal, Montreal, Quebec, Canada); H Ott (Massachusetts General Hospital, Boston, Mass); and H Gaissert (Massachusetts General Hospital, Boston, Mass).
19. Schuchert MJ, Abbas G, Landreneau JP, Luketich JD, Landreneau RJ. Use of energy-based coagulative fusion technology and lung sealants during anatomic lung resection. J Thorac Cardiovasc Surg. 2012;144: S48-51. 20. Schuchert MJ, Abbas G, Pettiford BL, Luketich JD, Landreneau RJ. Preliminary results of anatomic lung resection using energy-based tissue and vessel coagulative fusion technology. J Thorac Cardiovasc Surg. 2010; 140:1168-73. 21. Okada M, Miyata Y, Takamochi K, Tsutani Y, Oh S, Suzuki K. Prospective feasibility study of sealing pulmonary vessels with energy in lung surgery. J Thorac Cardiovasc Surg. 2019;157:388-95.
References
Key Words: pulmonary artery, vessel sealing, ultrasonic energy, VATS lobectomy
Discussion Dr Scott J. Swanson (Boston, Mass). You and your co-investigators should be congratulated for successfully designing and carrying out a prospective phase 2 trial looking at new technology, and you looked at ultrasonic energy to divide PA branches during a lobectomy and a segmentectomy. You showed what we and others have seen, that you can successfully divide 7-mm PA branches using the ultrasonic shears. We did it in animals too and saw the same thing: minimum morbidity and no mortality. It’s a big advance. You did observe several failures, which were immediately remedied at the time and had no significant consequence. I have 3 questions. You mentioned in your article it’s hard to know exactly how many of these branches that could be divided were divided by each surgeon, but do you have any idea if, say, 6 or 7 or smaller, if 3 were done or 2 were done, and how those decisions were made? Do you have any insight on that? Dr Moishe Liberman (Montreal, Quebec, Canada). We have all those data. Every vessel had to be measured, even if you were going to staple it or put a clip on it, so we know which vessels were sealed with energy and which were not. And I would say in the trial I think, I could be wrong, but I think except for the 9 or 10 vessels that were 7 mm and less, they were all sealed with the ultrasonic device. The reasons reported by the surgeons who had not sealed those smaller vessels, some of them were due to anatomy or I guess the vessel probably was very, very short, and some were at the beginning of the trial when surgeons were a little less comfortable and were in their learning curve, sort of getting used to seeing that this actually works. I think at the end of the trial, almost every size
The Journal of Thoracic and Cardiovascular Surgery c Volume -, Number -
9
THOR
1. Gazala S, Hunt I, Valji A, Stewart K, Bedard ER. A method of assessing reasons for conversion during video-assisted thoracoscopic lobectomy. Interact Cardiovasc Thorac Surg. 2011;12:962-4. 2. McKenna RJ Jr, Houck W, Fuller CB. Video-assisted thoracic surgery lobectomy: experience with 1,100 cases. Ann Thorac Surg. 2006;81:421-5; discussion 5-6. 3. Gharagozloo F, Tempesta B, Margolis M, Alexander EP. Video-assisted thoracic surgery lobectomy for stage I lung cancer. Ann Thorac Surg. 2003;76:1009-14; discussion 14-5. 4. Nomori H, Horio H, Naruke T, Suemasu K. What is the advantage of a thoracoscopic lobectomy over a limited thoracotomy procedure for lung cancer surgery? Ann Thorac Surg. 2001;72:879-84. 5. Mehta H, Osasona A, Shan Y, Goodwin JS, Okereke IC. Trends and outcomes of thoracoscopic lobectomy or segmentectomy: a national surgical quality improvement project analysis. Semin Thorac Cardiovasc Surg. 2018;30:350-9. 6. Liberman M, Khereba M, Goudie E, Kazakov J, Thiffault V, Lafontaine E, et al. Pilot study of pulmonary arterial branch sealing using energy devices in an ex vivo model. J Thorac Cardiovasc Surg. 2014;148:3219-23. 7. Dindo D, Demartines N, Clavien P-A. Classification of surgical complications. Ann Surg. 2004;240:205-13. 8. White A, Kucukak S, Lee DN, Swanson SJ. Energy-based ligation of pulmonary vessels: a six-year experience with ultrasonic shears in video-assisted thoracoscopic lobectomy and segmentectomy. Ann Thorac Surg. 2016;101:1334-7. 9. Goudie E, Khereba M, Tahiri M, Hegde P, Thiffault V, Hadjeres R, et al. Pulmonary artery sealing with an ultrasonic energy device in video-assisted thoracoscopic surgery lobectomy: an animal survival study. Ann Thorac Surg. 2016; 102:1088-94. 10. Goudie E, Oliveira RL, Thiffault V, Jouquan A, Lafontaine E, Ferraro P, et al. Phase 1 trial evaluating safety of pulmonary artery sealing with ultrasonic energy in VATS lobectomy. Ann Thorac Surg. 2018;105:214-20. 11. Goudie E, Thiffault V, Jouquan A, Lafontaine E, Ferraro P, Liberman M. Pulmonary artery sealing with ultrasonic energy in open lobectomy: a phase I clinical trial. J Thorac Cardiovasc Surg. 2017;153:1600-7. 12. Howington JA, Blum MG, Chang AC, Balekian AA, Murthy SC. Treatment of stage I and II non-small cell lung cancer: diagnosis and management of lung cancer, 3rd ed: American College of Chest Physicians evidence-based clinical practice guidelines. Chest. 2013;143:e278S-313S. 13. Toishi M, Yoshida K, Agatsuma H, Sakaizawa T, Eguchi T, Saito G, et al. Usefulness of vessel-sealing devices for ¼7 mm diameter vessels: a randomized controlled trial for human thoracoscopic lobectomy in primary lung cancer. Interact Cardiovasc Thorac Surg. 2014;19:448-55. 14. Rocco G, Internullo E, Cassivi SD, Van Raemdonck D, Ferguson MK. The variability of practice in minimally invasive thoracic surgery for pulmonary resections. Thorac Surg Clin. 2008;18:235-47. 15. Liberman M, Khereba M, Nasir B, Goudie E, Danino A, Giot JP, et al. Pulmonary artery sealing using the HARMONIC ACEþ shears for video-assisted thoracoscopic surgery lobectomy. Ann Thorac Surg. 2015;100:898-904. 16. Kim FJ, Chammas MF Jr, Gewehr E, Morihisa M, Caldas F, Hayacibara E, et al. Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT). Surg Endosc. 2008;22: 1464-9. 17. Yamada T, Sowa T, Bando T, Date H. Experimental study in pulmonary artery sealing with a vessel-sealing device. Asian Cardiovasc Thorac Ann. 2016;24: 562-7. 18. Lesser TG, Wolfram F, Boltze C. Sealing of pulmonary arteries with LigaSure: in vivo and ex vivo examinations. J Thorac Cardiovasc Surg. 2013; 145:1525-8.
Thoracic
THOR
appropriate vessel is being sealed with energy. So it’s a small percentage. Dr Swanson. We have seen, shown, and observed that you can use it on veins, no problem, same size, 7 mm. Any reason not to look at that or did you just want to keep it as a pure study? Dr Liberman. Good point, and I agree with that 100%. We also use it in anatomic segments, on veins and middle lobe veins. In reality, when you are doing a lobectomy, it is rare to find a very small vein that drains an actual lobe. The trial was really designed for the PA. It would have been confusing, for the sample size calculation, to have added veins to it, especially because we had a small percentage of segments. But I think there is no hesitation to use it on veins. The worry with veins is not the same as with arteries, because systemic veins and pulmonary veins are pretty much the same. This has already been well shown in the belly to work on veins, so I’m not worried about it. But it’s a good question. Dr Swanson. I think the economics of these things are important, and the more you use it and the less you use other devices the better. The other thing is, I think the 7-mm size is based on the size of the probe, it’s about an 11- to 12-mm probe, and when you flatten out the vessel, you want no wrinkling, as you have mentioned in your article, and that’s where 7 mm comes in. So my guess is when probes get bigger, we will be able to do bigger vessels. The last question is really around the technique, and I have a bit of difference in your sort of interpretation about what setting to use, and I think it showed up well in the videos. I think the longer you have to pull the vessel, the more likely you can sometimes tear or be under tension, and tension is really the problem here. With the fast speed, you need less time and you get less tension, and in our animal model and in our practice the fast speed worked. And 1 of the vessels that didn’t work out, you mentioned maybe because it was under some tension. So I wonder if you have data as to why you think the fast setting is worse than the slow setting, because we found the opposite, and I think you would avoid some of that tension. But it was a great study. I really appreciate it. Dr Liberman. The answer is, I don’t know. I would say that in our animal studies we found that the slow setting for larger vessels was better, and that was, again, we were just at 5 mm, we weren’t going to 7 mm at that time. When you have a very small 2-mm vessel, yes, I think the fast is great. When you are at 6 to 7 mm, we were using advanced hemostasis; when you are at 5, I’m not sure the fast setting is enough. We tried to keep it standard for the trial. We haven’t had problems. We have done more than 500 cases now without a problem using these settings. And I’m not saying the other way is wrong, it’s just that’s the way we do it, and we wanted everyone in the trial to do it the same way. 10
Liberman et al
Dr Mark F. Berry (Stanford, Calif). Two quick comments. One, congratulations on taking a tool, introducing it to clinical practice, and then actually systematically evaluating it, which is really awesome to see. Two, I’m sure you’ll never need to have a chemical stress test, because I feel like just watching that was a stress test in itself. But my question is, beyond what Dr Swanson asked you about when you were talking, can you give some technical tips for when you start to use this in practice? One, I can imagine you have to control the tension while you are burning the vessel so it doesn’t rip too soon before it’s sealed, and if you can just give some insight on how to gauge that technical aspect. The other question is, how important is it to avoid those stumps after the vessel is sealed, such as manipulating the stumps or pushing on the stumps, or causing any tension on the cauterized area? Dr Liberman. I would say that the reason we leave a long stump isn’t necessarily because we think it seals better, but we like to have that measure of security as you saw on the video to apply a clip or a stitch if there is a problem. Obviously these devices are very hot, and we have done studies showing that some of these devices heat up to about 300 C. So putting that near the main PA is obviously good to avoid, if you can, not because we have proven that it makes a difference, but I think that everyone in this room would be worried about putting a match on a main PA. In the trial and in clinical practice, we try to dissect the vessel out so there is no fat, tissue, or lung or lymph node behind or around it. So you are not just heating like you would be doing a right hemicolectomy, but you are actually dissecting the vessel the way you would do it with scissors in a regular open lobectomy. We found that if you have tissue that’s not artery in the device, you have a higher chance of failure. One of the patients who failed in the study had the vessel shoved into the crotch of the jaw, and it was a bit folded. We have videos on all of the patients. It failed, and it actually failed at that crotch. Once we have it dissected and we have a stump, then we try and relax all tension on the lobe, so you are not pulling hard on the vessel, and I think that’s what most of us would do with the stapler as well. Unidentified Speaker. That was just my next question. At OHSU we have been using it, but we only use it for about 5-mm vessels or smaller. So it’s good to see that it can be used for bigger vessels. And that was going to be my question, because we try to aim it so it’s just in the wide part of the Harmonic and not necessarily in the crotch, and I was just going to ask that question. Because some of the videos showed that the vessels were in the crotch and it still worked, but my question is, do they fail more when they are in the crotch versus not?
The Journal of Thoracic and Cardiovascular Surgery c - 2019
Liberman et al
Thoracic
Dr Liberman. Again, when we see a longer stump, we’re talking about 5 mm, maybe 4 mm. When you use an endostapler, you are usually leaving 5 or 4 mm on each side anyway for the artery. I’m not saying in terms of the stump. If you leave a 5-mm stump on a venous branch, I don’t know that that is actually a problem. That’s a good question, can that cause an embolism? We haven’t seen it, but I’m not saying it’s not possible. It’s a good question and a concern. Again, we are really only using it on segments and not for lobectomies. Dr Thomas A. D’Amico (Durham, NC). It’s reported in the literature for lobectomies.
THOR
Dr Liberman. A good question. It is exactly as we said, don’t put it in the crotch, it should really lie flat, because the problem with the crotch is that the tension is not evenly distributed and the vessel often folds, even a minor bit that you don’t see, and we have seen that when we have done ex vivo studies looking under the electron microscope at what happens to those vessels. Dr Gunda Leschber (Berlin, Germany). I would like to return to the length of the stump. On a PA, I don’t think that’s so much of a problem because it is relatively long, although I was at least trained to have a very short stump only. But I would caution if you do this on a vein, because don’t you think it could be where an embolism could develop and then you end up having systemic emboli?
The Journal of Thoracic and Cardiovascular Surgery c Volume -, Number -
11
Thoracic
000
Prospective, multicenter, international phase 2 trial evaluating ultrasonic energy for pulmonary artery branch sealing in video-assisted thoracoscopic surgery lobectomy Moishe Liberman, MD, PhD, Eric Goudie, MD, Christopher Morse, MD, Wael Hanna, MD, MBA, Nathaniel Evans, MD, Kazuhiro Yasufuku, MD, PhD, and John Sampalis, PhD, the VATS PA Study Working Group, Montreal, Quebec City, Hamilton and Toronto, Ontario, Canada; Boston, Mass; and Philadelphia, Pa PA branch sealing with ultrasonic energy during VATS lobectomy is safe for vessels of 7 mm or less. The use of an ultrasonic device is a reasonable sealing method for PA branches 7 mm or less.
THOR The Journal of Thoracic and Cardiovascular Surgery c - 2019
Liberman et al