199
Thermochimica Acta, 94 (1985) 199-204 Elsevier Science Publishers B.V., Amsterdam - Printed in The Netherlands
RESULTS
OF R ROUND ROBIN EXPERIMENT
DIFFERENTIRL
SCRNNING
‘Sektion
fur
D-7900
Ulm
and
Kalorimetrie
K.-H.
der
BREUER;’
Universitst,
Postfach
4066
(FRGI
‘Mineralogisch-Petrographisches D-6900
OF
CRLORIMETERS”
HflHNE?, , W. EYSEL2
G.W.H.
ON THE CRLIBRRTION
Institut
Heidelberg
der
Universitat,
Postfach,
(FRG)
FlBSTRRCT Three substances (indium, rubidiumnitrate and Lithiumsulfate) with five samptes each were run on four scanninq calorimeters of different type to measure the heat of five well-defined transitions in the temoerature ranoe 150 - 580 -C. In order to avoid errors due to differences in composition and weighing, identical samples were The results of the two power measured on the different apparatus. compensated scanning calorimeters fit very wetl, but the heats measured on calorimeters of heat-flux type are systematically to smalt in cases of higher transition temperatures and large specific transition enthalpies. Explanation is possible by the theory of heat-flux calorimeters predicting such effects.
INTRODUCTION calibration
of
quires
The
standard
reference
(resp.
heat-flux)
and
orimeter.
parameters
sition
heat,
known.
Moreover,
nufacturers periments of
‘interest, In
the
this
four
not
testing
(RbNO, different
the and of
0040-6031/85/$03.30
results
of
LizSO,
at
the
6th
on tran-
shoutd
be of
trustworthy.
Thus and
substances
at1
the
are
welt the
maex-
suitabi-
of
high
experiments. testing by
two
inorganic
a miniature
are
not
cal-
factor specific
calorimeters
heat
respective
statements
reported.
calibration-factor is
the
etc.
the
very of
robin
of
re-
and
calibration
sample
reference
round
it
the
(USC)
temperature
temperature,
as
calibration standard
the
the
experience
taken
calorimeters
Certainly, presented
be
the as
of
of to
for
knowledge
transition
conductivity
materials paper
size,
calorimeters
both
dependence
according
should
determination
orimeter. *Poster
as
including
substances on
the
heat
scanning
substances a sufficient
Especially
sample
Lity
differential
possible
robin
Indium
was
of to
draw
Utm Calorimetry
0 1985 Elsevier Science Publishers B.V.
nonmetattic
round the
experiment added
respective
definitive
Conference,
for cal-
concluMarch
1985
200
sions
since
theless
the
the
cular
as
population
results
a base
of
of for
this
the
test
first
enlarged
was
attempt
test
rather are
limited.
Never-
valuable,
in
parti-
series.
EXPERIMENTDL Preceeding got
from
experiments metal-samples
with
literature)
tion
heat
was
of
fluctuate c,‘C were
Five cision for
salts
(at
choosen
for
the about
balance
into
sample
in
to
To normalize
the
samples
Indium fifteen
comparable samples
high were
were
to
be
institute.
The
in
the
ELMER DSC-2
Hohne,
ELMER DSC-2C
Wruck
C
DUPONT TFI 1090/910
D
HERFIEUS TF1 500
E
same
F1,B:
power
compensated
there
are
the the
age
samples
carried
out
these
return run
whether the
to
samples were
of
Indium of heat
the of
institutes calorimetry.
5 Klmin. of
1711
The the
calorime-
respective
used: of
Ulm and
Hannove
and
Braunschweig
Cammenga
holder
assembly
C,D,E:
heat
of
up
scientists other
scanning
flux
calorimeter)
calorimeters apploying
to
r
Heidelberg
Eysel
types or
with
contact
us
in
them. in
use,
May
order
to
experiments. of
once
within
one
pre-
suitable
melting
Salje
Type
hunt
a high
University
sample
more
encourages in
were
other
much
succeed
paper
participate Flfter
of
method
and
Reichelt with
we didn’t
they
scanning
were
Breuer
C but
at
the
Investigators
as
solidone
calibration
scientific
rate
calorimeters
PERKIN
this
in
individual
PERKIN
be
on
samples
the
we defined
precise
B
but
with
J/mot.
R
Indeed
defined
Rl-pans)
Five
a heating
with
well Li;,SO,
weighed
Pont
compare
measured
applying
transi-
apparatus
experiment. were (Du
results
Type
(Type
robin
and
3280
following
three
results (compared
the
different
<‘C) and
investigation.
control
were
calibrated
with
285
containers
experience
run
and
the
calorimeters
with
RbNO,
of
whereas
IO mg mass
the
samples
scanning
round
under
order
coincidence
(Ref.1,2,3),
225
of
equipment.
ters
Thus,
166,
calorimeters
the
determined
each
added
The
good
samples
all
were
different
more.
transitions
580
by rather
inorganic
sligthly
solid
showed,that
the
more
samples on
the
remained kept half
from
unchanged. in
a year,
the
calorimeter dry
During
atmosphere. the
round Fl in
control
robin order
experiment to
transport Rll run
test and
measurements after
one
storwere
year.
201
TRBLE 1 Results
of
the
round
robin
experiment
on
transition
heat
determination Calorimeter
Heat
of
transition
and
standard
deviation
(kJlmo1)
of
III
Li ,SO,
1.301+.015
24.8t
Type
Indium
RbNO,
I
II
R (l.runI
3.2802.003
3.83t
.Ol
3.16
B
3.28Of.006
3.850:.004
3.192t.008
1.31
f.02
24.82
C
3.28
+.0’S
3.86+
.04
3.19
t.04
1.29
t.O1
23.9
*, .3
II
3.28
5.06
3.82
+ . 13
3.09
f.10
1.21
t.04
22.8i
.5
E
3.28
2.03
3.72t
.03
3.02
2.03
1.235-+.006
22.121.3
R (2.rut-t)
3.28Ot.004
3.‘10
+-.04
1.28
24.52
3.81t.02
+-.02
t.01
.2 .3
.4
RESULTS The
experimenters
onset
and
of
expected well The
in
with results
this
whereas
the
results
in
obtained
temperature
transitions the
in
transition
of
the
question.
peaks
calibration,
from
factor
I%
fit
rather
however,
is
the
indium
(defining
samples
the
true
was
not
used
melting
of
the
respective
calorimeter
were
to
get
comparable
data.
1 includes
factor
the
round of
two
robin
the
power
case
specific
heats
of
observed
heats
from
heat
higher
transition. the
samples
compensated of
of
experiment,
five
i.e. as
the
well.
flux
type
transition
fit
rather
of
yielded
2 includes the
two
and
be
more
temperature
power
seepI
the the
well,
diverging
and
the
as
mul-
average
Fls can
calorimeters
Table average
Table
to
heat
this
calorimeters the
five of
the
!A11 results of
the
and
paper.
uncertainty of
the
heat
Temperature
heat
JlmoL).
results
of
temperatures
a normaLization
tiplied
random
the
calorimeters.
transition
caLcuLate
the
maximum
onset
all
considered
3280
the
the
on
reported
Larger
deviation compensated
of
the
calo-
rimeters.
TFlBLE 2 Difference the
average
between measured
the
heats on
power
measured compensated
on
heat
flux
calorimeters.
calorimeters
and
202
DISCUSSION F)s can lity’
be
of
seen
the
systematic
from
four
the
errors
in
the
factor
of
of
sample
holder
factor
versus
temperature
rather
unknown
the
Rnother (at
calorimeters
for
‘BC)
Here
was
the
a clear
than
that
of
curves
(Ref.51
the
Ref.4,5).
lead
to
straight-line
area
for
of
results
heat
third
the
peak
of
step
power
the
‘qua-
seem
to
calibration
calibration
unity,
a fact
transition calorimeters in
compensated
up
to
of
deviation
of
RbNO,
of
heat
baseline
(Fig.11
(for
calorimeters determining
three
applied
This
the
the
percent.
in
this is
The
case
peak
area
method
of
yields
an
considerably
lar-
calorimeters.
557
Plots of RbNOJ transition at 285 UC measured a) by a heat flux calorimeter (Du-Pont TFI 1090/910) b) by a power compensated calorimeter (Perkin-Elmer the same sample. The left curve shows a larger baseline
T/ K
Fig.1:
of
than ____
the right straight-line
one
be
replacement the
from
change
methods
small.
the the
altered
differ
the
correction
too flux
to
for
varying
baseline
the
of
E
has
1
calorimeter.
Larger
Different
systematically
ger
found
curves
showed
see
of
of
calorimeter
obviously
type
Table there
dependence
For
function
this
type
theory
E.
in
Moreover,
temperature
assembly
flux
deviations
differs.
D and
phenomenon
285
standard
calorimeters
caused baseline
by
the
heat flux method. . . . . . ‘true’ baseline
DSC-2)
shift
203 Furthermore
the
type
involves
the
calibration
C,D to
of
factor
method
errors
heat
of
transition
transition
heat
of
l_i,SO,
and E and be
caused
about
that by
eight
measured
this
times
calorimeters
parameters
to
heat
flux
a dependence
the
of
as mass,
heat
etc.
The very
deviation
measured
by means
of
The than
of
leading
by power
fact.
larger
of
(Ref.41
on sample
specific
vity, the
measuring
sources
compensated
transition that
of
calorimeters
calorimeters
heat the
the
conductibetween
of
other
this
seems
material
is
transitions.
SUMMFlRY The
round
cautions
robin
experiment
because
its
has
limited
ments
and
drawn
which
are
also
- 5)
and
other
literature
indicate
a slight
(Ref .I RI1 wer
materials.
results
compensated
need its
test
of
full
task
for
international of
of
permanent results
flux the
comparable the
differential the
the
the
calibration
factor
of
the
in
pretensions
ing
to
thermal
a less For
heat
analysis so
is
it
system
of
available.
time
calibraand cle-
care
and
and maintain
DSC.
On the
(oxidation)
the
heat
and
experimental
conducting
to
and
of
obtain
compensated by aging
transition
one
of
the
material,
a certain
extent
and
recorwnended
in
on
indi-
results
heat
DSC
can
and
accuracy
are
relation
to
close
electronic
reason
scanning far.
flux
its
yet
quality
to
depends of
covering
by great
possible
po-
the
on tempera-
question.
repeatability sensitive
that
quantity
in
advantage
compensated system
the
transition
Nevertheless
price.
and
(Ref.41
the
The major
probe
the
Only
the
instruments
development
on expense
a power
theoretical
combination
of
of
is
not
their
moment
may alter
hand
that
ture
it
those
be
authors
of
flux
are be
depends
factor
the
calibration
standards
experimenter.
temperature
other
cate,
to
of
heat
heat
should
Fit the
control
calibration
can
reproducibility the
for
Such
calorimeters
calibration
pre-
instru-
conclusions
experiences
better
future
respective
great
data.
standards
near
with
participants,
some
Moreover,
range.
the
evaluated of
on previous
recommendation.
heat
verness
hand
of
temperature
!A major tion
based
set
be
Nevertheless
instruments.
an excellent
to
number
be
measurement
is
easy
to
with
very
low
calorimeter
is
include costs,
available
case
slightly the
reduced.
DTA method
equipment a DSC module whereas only
the
the as
lead-
at in
a lower a
power
a complete
204
FICKNOWLEDGEMENT The B.
authors
Wruck
Hannover
of and
Physikalische Braunschweig The
investigations
wish the
to
Institut
H.
K.
und for
express fur
Caronenga
their
and
Theoretische careful were
gratitude
to
Kristallographie J.
Reichelt
Chemie
measurements supported
by
E.
der of
the
der
Technischen
and
very
the
Deutsche
Salje
and
Universitat Institut
fur
Universitat
helpful
cooperation.
Forschungs-
gemeinschaft.
REFERENCES The Calorimetric Calibration of K.-H. Breuer and W. Eysel, Differential Scanning Calorimetry Cells, Thermochimica Rcta 57 (1982) 317-329. W. Eysel and K.-H. Breuer, Differential Scanning Calorimetry: Simultaneus Temoerature and Calorimetric Calibration, in J. F. Johnson and P. S. Gill (Ed.): Analytical Calorimetry; Plenum Publishing Corporation, 1984, pp. 67-80. G. W. H. Hohne, K.-H. Breuer and W. Eysel, Differential Scanning Calorimetry: Comparison of Power Compensated and Heat Flux Instruments, Thermochimica Flcta 69 (1983) 145-151. G. W. H. Hohne, Problems with the Calibration of Differential Temoerature Scannino Calorimeters, Thermochimica Flcta 69 (1983) 175-197. W. Hemminger and G. Htihne: Calorimetry, Fundamentals and Practice, UCH Verlags-Gesellschaft Weinheim, Deerfield Beach (Florida), Easel, 1984.