JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY
VOL. 75, NO. 4, 2020
ª 2020 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION PUBLISHED BY ELSEVIER
ORIGINAL INVESTIGATIONS
Short-Term Hemodynamic and Electrophysiological Effects of Cardiac Resynchronization by Left Ventricular Septal Pacing Floor C.W.M. Salden, MD,a Justin G.L.M. Luermans, MD, PHD,b,c Sjoerd W. Westra, MD,b,c Bob Weijs, MD, PHD,b Elien B. Engels, MSC, PHD,a,d Luuk I.B. Heckman, MD,a Léon J.M. Lamerichs, RN,b Michel H.G. Janssen, BSC,b Kristof J.H. Clerx, RN,b Richard Cornelussen, PHD,a,e Subham Ghosh, PHD,f Frits W. Prinzen, PHD,a Kevin Vernooy, MD, PHDb,c
ABSTRACT BACKGROUND Cardiac resynchronization therapy (CRT) is usually performed by biventricular (BiV) pacing. Previously, feasibility of transvenous implantation of a lead at the left ventricular (LV) endocardial side of the interventricular septum, referred to as LV septal (LVs) pacing, was demonstrated. OBJECTIVES The authors sought to compare the acute electrophysiological and hemodynamic effects of LVs with BiV and His bundle (HB) pacing in CRT patients. METHODS Temporary LVs pacing (transaortic approach) alone or in combination with right ventricular (RV) (LVsþRV), BiV, and HB pacing was performed in 27 patients undergoing CRT implantation. Electrophysiological changes were assessed using electrocardiography (QRS duration), vectorcardiography (QRS area), and multielectrode body surface mapping (standard deviation of activation times [SDAT]). Hemodynamic changes were assessed as the first derivative of LV pressure (LVdP/dtmax). RESULTS As compared with baseline, LVs pacing resulted in a larger reduction in QRS area (to 73 22 mVs) and SDAT (to 26 7 ms) than BiV (to 93 26 mVs and 31 7 ms; both p < 0.05) and LVsþRV pacing (to 108 37 mVs; p < 0.05; and 29 8 ms; p ¼ 0.05). The increase in LVdP/dtmax was similar during LVs and BiV pacing (17 10% vs. 17 9%, respectively) and larger than during LVsþRV pacing (11 9%; p < 0.05). There were no significant differences between basal, mid-, or apical LVs levels in LVdP/dtmax and SDAT. In a subgroup of 16 patients, changes in QRS area, SDAT, and LVdP/dtmax were comparable between LVs and HB pacing. CONCLUSIONS LVs pacing provides short-term hemodynamic improvement and electrical resynchronization that is at least as good as during BiV and possibly HB pacing. These results indicate that LVs pacing may serve as a valuable alternative for CRT. (J Am Coll Cardiol 2020;75:347–59) © 2020 by the American College of Cardiology Foundation.
Listen to this manuscript’s audio summary by Editor-in-Chief Dr. Valentin Fuster on JACC.org.
From the aDepartment of Physiology, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Maastricht, the Netherlands; bDepartment of Cardiology, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center (MUMC+), Maastricht, the Netherlands; cDepartment of Cardiology, Radboud University Medical Centre (Radboudumc), Nijmegen, the Netherlands; e
d
Department of Medicine, University of Western Ontario, London, Ontario, Canada;
Medtronic PLC, Bakken Research Center, Maastricht, the Netherlands; and fMedtronic PLC, Cardiac Rhythm and Heart Failure
(CRHF), Mounds View, Minnesota. A part of this study was financially supported by Medtronic (Minneapolis, Minnesota). Dr. Luermans has a consultancy agreement with Medtronic; and has received an educational grant from Biotronik. Dr. Engels has received a post-doctoral research fellowship from the Cardiac Arrhythmia Network of Canada (CANet). Dr. Cornelussen is an employee of Medtronic. Dr. Ghosh is an employee of and stock holder in Medtronic. Dr. Prinzen has received research grants
ISSN 0735-1097/$36.00
https://doi.org/10.1016/j.jacc.2019.11.040
348
Salden et al.
JACC VOL. 75, NO. 4, 2020
LV Septal Pacing for CRT
FEBRUARY 4, 2020:347–59
D
ABBREVIATIONS AND ACRONYMS AV = atrioventricular
uring the last 2 decades, cardiac
with endocardial CRT [7,8]). Another advantage of
resynchronization therapy (CRT)
LVs pacing may be that lead positioning is less critical
has developed as an important
than
treatment option for symptomatic patients
BiV = conventional biventricular
CRT = cardiac resynchronization therapy
ECG = electrocardiogram EP = electrophysiological HB = His bundle LBB = left bundle branch LBBB = left bundle branch block
LV = left ventricular
during
His
bundle
(HB)
pacing,
another
approach that is gaining interest in CRT (9–13).
with left ventricular (LV) systolic dysfunc-
To further explore the potential of LVs pacing in
tion and dyssynchronous ventricular activa-
CRT, the present study aimed to answer the following
tion, especially due to left bundle branch
questions: how do the electrophysiological and he-
(LBB) block (LBBB). Randomized clinical tri-
modynamic effects of LVs in CRT candidates compare
als showed that CRT creates a more coordi-
with those of BiV and HB pacing; and how important
nated and efficient ventricular contraction,
is the location within the LVs? These questions were
improves LV systolic function and functional
investigated in a prospective dual-center interven-
capacity, reverses structural remodeling, and
tional cohort study by measuring short-term electro-
decreases heart failure hospitalization and
physiological and hemodynamic effects in heart
mortality (1,2).
failure patients with an indication for CRT. SEE PAGE 360
LVdP/dtmax = first derivative
METHODS
of left ventricular pressure
CRT is usually applied by stimulation of
LVs = left ventricular septal/
both the right ventricle (RV) and the LV
STUDY
lateral wall (biventricular [BiV] pacing). Our
patients, fulfilling the criteria for CRT implantation
group has investigated a single-lead alter-
according to the current guidelines (14), were pro-
septum
native approach for ventricular resynchro-
spectively enrolled. Patients were included in the
SDAT = standard deviation of
nization: pacing the LV endocardial side of
Maastricht University Medical Centerþ (n ¼ 23) and
activation times
the interventricular septum (LV septal [LVs]
the Radboud University Medical Center (n ¼ 4) from
VCG = vectorcardiogram
pacing). LVs pacing is based on the idea that
November 2017 to April 2019. These patients were
in the normal heart, the earliest exit of electrical im-
enrolled on the basis of the inclusion criteria of the
pulses out of the Purkinje system occurs at the LVs
presence of sinus rhythm, New York Heart Associa-
and that LVs pacing would provide a near physio-
tion functional class II to III, optimal heart failure
logical activation. This idea has been supported by
medication, LV ejection fraction <35%, and QRS
short- and long-term animal studies (3,4). Impor-
duration $130 ms in the presence of LBBB or QRS
tantly, long-term LVs pacing maintained regional
duration >150 ms in the absence of typical LBBB.
cardiac mechanics, contractility, relaxation, and effi-
Patients were excluded when they had moderate-to-
ciency near native levels, whereas RV apical or RV
severe aortic stenosis, frequent premature ven-
septal (RVs) pacing diminished these variables (4). In
tricular complexes ($2 complexes on a standard
relation to CRT, Rademakers et al. (5) showed that in
electrocardiogram [ECG]), significant peripheral vascular
canine LBBB hearts, short-term electrical and hemo-
disease, or an age below 18 years.
septum
RV = right ventricular RVs = right ventricular septal/
POPULATION. Twenty-seven
consecutive
dynamic benefits of LVs pacing were comparable to
This study was performed according to the princi-
those during BiV pacing. Moreover, in a small group
ples of the Declaration of Helsinki, and the study
of patients with an indication for CRT, short-term
protocol was approved by the ethics committee of the
hemodynamic benefits of LVs pacing and BiV pacing
Maastricht University Medical Centerþ (registration
were also comparable.
number NL58809.068.16/METC 161117). All patients
The successful permanent implantation in dog
gave written informed consent before investigation,
hearts was followed by a small study in 10 patients
and the study was monitored by the Clinical Trial
with sinus node disease in which a modified version
Center Maastricht.
of the Medtronic 3830 lead (Medtronic, Dublin,
STUDY PROTOCOL. All participants underwent CRT
Ireland) was implanted permanently using a trans-
implantation according to routine clinical practice.
venous
interventricular
The atrial lead was positioned in the right atrial
septum (6). This approach was feasible and avoids
appendage, the RV lead in the RV apical septum,
access of the LV lead into the LV cavity (as is the case
and a quadripolar LV lead in a suitable vein on the
route
and
crossing
the
from Medtronic, Abbott, Microport CRM, Biosense Webster, MDS, and Biotronik; and has served as an advisor to Oracle Health. Dr. Vernooy has received research grants from Medtronic, Abbott, and Biotronik; and has a consultancy agreement with Medtronic and Abbott. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose. Manuscript received July 24, 2019; revised manuscript received October 31, 2019, accepted November 8, 2019.
JACC VOL. 75, NO. 4, 2020
Salden et al.
FEBRUARY 4, 2020:347–59
LV Septal Pacing for CRT
F I G U R E 1 Fluoroscopic Images of the Implantation Procedure
LVs mid level
LVs mid level
LVs basal level
LVs apical level
Right anterior oblique (RAO) (upper left) and left anterior oblique (LAO) (upper right) view of the heart with right atrial (RA), right ventricular (RV), and left ventricular (LV) leads and the pressure wire. Upper panels show position of the electrophysiological catheter during mid-level LV septal (LVs) pacing, whereas the lower panels show the catheter position during LVs basal and apical level pacing.
posterolateral LV wall. Thereafter, a 0.014-inch
by a small atrial and large ventricular electrogram
PressureWire X (Abbott, St. Paul, Minnesota) was
during electrical mapping from the catheter tip was
introduced via a femoral arterial puncture using a
targeted. HB capture was noted according to previ-
4-F multipurpose catheter, into the LV cavity
ously defined criteria with correction of the bundle
(Figure 1). Subsequently, a 6-F steerable electro-
branch block (stimulus to QRS end # His to QRS end
physiological (EP) catheter was advanced via the
and paced QRS narrowing) (12).
same femoral arterial puncture into the LV for
The pacing protocol was performed at >10 beats/min above the intrinsic sinus rate to exclude spontaneous
temporary pacing. Hemodynamic and electrophysiological measure-
changes in heart rate. To ensure full ventricular
during
capture, ventricular pacing was performed during
different pacing configurations. Bipolar BiV was per-
atrioventricular (AV) sequential pacing (DDD mode)
formed at 2 levels of the LV lead, using the 2 distal
with a paced AV delay 60 ms shorter than the
(1 and 2) and 2 proximal (3 and 4) electrodes of the
intrinsic PR interval or a paced AV delay of 150 ms in
quadripolar LV lead. To perform LVs pacing, the EP
case of prolonged PR interval (>210 ms). The inter-
catheter was temporarily positioned at different LVs
ventricular pacing delay during biventricular pacing
levels (basal, mid-, and apical). The position at the
was set to 0 ms. Baseline measurements were per-
ments
were
simultaneously
performed
fluoroscopic
images
formed during atrial pacing (AAI mode) at the same
(Figure 1) and the 12-lead ECG. The LVs was paced
pacing rate, both before and after each mode of
alone as well as in combination with the RV lead
ventricular pacing.
LVs
was
determined
from
(LVsþRV). Subsequently, the EP catheter was posi-
Once the pacing protocol was completed, the EP
tioned on the RV interventricular septum via the right
catheter and pressure wire were withdrawn, and the
femoral vein to perform mid-RVs pacing and on the
permanent leads were connected to the CRT device.
HB for HB pacing. A stable HB potential accompanied
The implantation procedure was completed, which
349
350
Salden et al.
JACC VOL. 75, NO. 4, 2020
LV Septal Pacing for CRT
FEBRUARY 4, 2020:347–59
F I G U R E 2 Example of the VCG Construction
X
QRSx = 88 µVs
Y
QRSy = –10 µVs
Z
Tz = –123 µVs
Ty = 9 µVs
Tx = –64 µVs QRSarea =
QRSz = 167 µVs
QRS2area,x + QRS2area,y + QRS2area,z = 190 µ Vs QRS area
T-wave area
Example of a 3-dimensional vectorcardiogram (VCG) constructed from a standard 12-lead electrocardiogram.
automatically terminated the study participation of
principle directions (X, Y, and Z) and represents the
the patient.
extent
M EA SU R EM EN TS. E l e c t r o c a r d i o g r a p h y . Standard
ventricular activation (Figure 2) (16,17).
of
unopposed
electrical
forces
during
12-lead ECGs were recorded throughout the procedure
The customized VCG software was also used to
using an EP recording system (Bard, Boston Scientific,
measure the QRS duration manually from the 12-lead
Massachusetts, or Claris, Abbott). After the procedure,
ECGs. QRS duration was defined as the maximum
3-dimensional
were
time between the beginning of the QRS complex and
synthesized from the recorded 12-lead ECGs using
the end of the QRS complex in all 12 leads. During
the Kors matrix (15). Ectopic ventricular beats
ventricular pacing, the beginning of the QRS complex
were excluded from the analysis. VCG parameters,
was defined in 2 ways: as the first visible deviation
including QRS area, were calculated using customized
after the pacing artifact and as time of the pacing
software programmed in MATLAB R2010b software
artifact.
(MathWorks, Natick, Massachusetts). QRS area is the
M u l t i e l e c t r o d e b o d y s u r f a c e m a p p i n g . A single-
combined area under the QRS complex of the 3
use disposable ECG belt (Heartscape Technologies,
vectorcardiograms
(VCGs)
JACC VOL. 75, NO. 4, 2020
Salden et al.
FEBRUARY 4, 2020:347–59
LV Septal Pacing for CRT
derived from the pressure wire and analyzed using
T A B L E 1 Baseline Characteristics (N ¼ 27)
Age at implantation, yrs
67 9
Men
20 (74) 26 4
2
Body mass index, kg/m
NYHA functional class II/III
22 (81)/5 (19)
ICM LVEF, % QRS duration, ms
PhysioMon software, version 2.02 (Abbott). LVdP/ dtmax is the maximum rate of LV pressure rise and a marker of LV contractility. During each hemodynamic measurement, LVdP/
12 (44)
dtmax was measured for each heartbeat for 30 s
26 5
during baseline (AAI pacing), 60 s during the ven-
151 14
tricular pacing protocol, and again 30 s during base-
QRS morphology
line. To account for spontaneous changes in baseline
LBBB
24 (89)
IVCD
3 (11)
LVdP/dtmax, each ventricular pacing response was compared with the corresponding baseline as relative
Medical history
percentage change in LVdP/dtmax. The last 10 beats
Myocardial infarction/PCI
12 (44)
Hypertension
7 (26)
Prior AF
6 (22)
for analysis, so that for each pacing mode, a “CRT-on”
Diabetes mellitus
4 (15)
and “CRT-off” effect could be determined (20). The
CABG
4 (15)
results of these 2 steps were averaged. Ectopic ven-
Ischemic stroke
3 (11)
tricular beats and the 2 subsequent heartbeats were
COPD
3 (11)
before and first 10 beats after every switch were used
excluded from the analysis.
Medication 27 (100)
STATISTICAL ANALYSIS. Continuous data are pre-
ACE inhibitors/ARB
23 (85)
sented as mean SD, and discrete variables as counts
Diuretics
22 (82)
and percentages. Data showed a normal distribution
Aldosterone antagonist
2 (7)
according to the Kolmogorov–Smirnoff and Shapiro–
Angiotensin receptor-neprilysin inhibitor
2 (7)
Wilk tests. For ventricular sites with multiple
Beta-adrenergic blocking agents
measurements (BiV with distal and proximal LV
Values are mean SD or n (%). ACE ¼ angiotensin-converting enzyme; AF ¼ atrial fibrillation; ARB ¼ angiotensin II receptor blocker; CABG ¼ coronary artery bypass grafting; COPD ¼ chronic obstructive pulmonary disease; ICM ¼ ischemic cardiomyopathy; IVCD ¼ intraventricular conduction delay; LBBB ¼ left bundle branch block; LVEF ¼ left ventricular ejection fraction; NYHA ¼ New York Heart Association; PCI ¼ percutaneous coronary intervention.
electrodes, LVsþRV pacing at different LVs levels, and LVs pacing alone at different levels), the 2 best results were averaged. Differences between the pacing settings were statistically evaluated using repeated-measures
analysis
of
variance
with
Bonferroni post hoc analysis, using SPSS software, Verathon, Seattle, Washington) (with modifications
version 25.0 (SPSS, Chicago, Illinois). A p value <0.05
in recording software by Medtronic) with 17 anterior
was considered statistically significant.
and 36 posterior unipolar ECG electrodes was
RESULTS
applied to the patient before the implantation procedure. with
A
multichannel
customized
amplifier
software
and
27 patients represent a typical CRT population. The patient cohort was predominantly male (74%), nearly
were
times
one-half had an ischemic etiology, and LV ejection
and color-coded isochronal maps per setting with
fraction was 26 5%. Cardiac magnetic resonance
custom-made software written in MATLAB. The
imaging or echocardiography showed septal ischemia
activation time was defined as the time to the
in 4 patients (15%) and akinesia or hypokinesia of the
steepest negative slope, and the standard deviation
septum in 4 other patients (15%). Most patients (89%)
of activation times (SDAT) from all electrodes was
were scheduled for a de novo device implantation,
calculated.
and 11% for a CRT upgrade from an implantable car-
SDAT
is
a
into
measure
for
PATIENT CHARACTERISTICS. Table 1 shows that the
acquisition. The data of each unipolar electrogram offline
used
monitor data
transformed
were
activation
of
the
spatial
synchronicity of ventricular activation, and $10%
dioverter-defibrillator.
SDAT reduction by CRT from intrinsic rhythm has
Twenty-six patients successfully completed the
previously been shown to be a predictor of LV
ECG analysis and body surface mapping measure-
reverse remodeling (18,19).
ments, whereas hemodynamic measurements were
HEMODYNAMIC
MEASUREMENTS. The
short-term
successfully performed in all 27 patients. HB pacing
hemodynamic effect of pacing the different sites
with correction of the bundle branch block was per-
was evaluated by invasive measurement of the
formed in 16 patients, with, in most cases, nonselec-
first derivative of LV pressure (LVdP/dtmax) as
tive HB capture (88%). In 1 patient, RV apical pacing
351
Salden et al.
JACC VOL. 75, NO. 4, 2020
LV Septal Pacing for CRT
FEBRUARY 4, 2020:347–59
F I G U R E 3 Examples of EP Effects
A
Baseline
BiV
His bundle
LV septal + RV
LV septal
137 ms / 96 PVs
141 ms / 72 PVs
QRS duration / QRS area = 147 ms / 124 PVs
135 ms / 70 PVs
B
85 ms / 79 PVs
Anterior Right
Posterior Left
Right
110
Baseline SDAT = 45 ms
88 BiV SDAT = 34 ms
66 His bundle SDAT = 27 ms
44 LV septal + RV SDAT = 34 ms
Activation time (ms)
352
22
LV septal 0
SDAT = 17 ms
Example of ECGs during baseline, conventional biventricular (BiV), His bundle, LV septal in combination with RV, and LV septal pacing alone with corresponding QRS duration and QRS area (A) and examples of isochronal maps with corresponding standard deviation of activation times (SDAT) (B). EP ¼ electrophysiological; other abbreviations as in Figures 1 and 2.
(VVI mode) was used as baseline because of the
ELECTROPHYSIOLOGICAL EFFECTS. An example of
presence of atrial fibrillation during the procedure
ECG recordings and isochronal maps based on body
(after inclusion). No periprocedural complications
surface maps is shown in Figure 3. Baseline QRS
were observed.
duration was 151 14 ms, with 24 patients (89%)
JACC VOL. 75, NO. 4, 2020
Salden et al.
FEBRUARY 4, 2020:347–59
LV Septal Pacing for CRT
F I G U R E 4 Electrophysiological Effects
Subgroup (n = 16)
QRS Duration (ms)
200
150
*
*
*
* *
100
50
0 200
†
§
QRS Area (µ µVs)
‡ 150 *
*
100
*
*
50
0 50
†
SDAT (ms)
40
*
*
*
*
*
LVs
HB
LVs
30 20 10 0
BL
BiV
LVs+RV
QRS duration as time between QRS beginning and end (closed bars) and as time from pacing stimulus to QRS end (open bars) (upper panel), QRS area (middle panel), and SDAT (lower panel) (n ¼ 26) during baseline (BL), conventional BiV, LVs in combination with RV, and LVs pacing alone and in a subgroup (n ¼ 16) during His bundle (HB) and LVs pacing. Results are presented as mean SD. *p < 0.05 versus BL; †p < 0.05 BiV versus LVs; ‡p < 0.05 LVsþRV versus LVs; §p < 0.05 HB versus LVs. Abbreviations as in Figures 1 and 3.
showing a LBBB morphology of the QRS complex ac-
comparable for 3 CRT conditions (BiV 136 15 ms,
cording to the European Society of Cardiology defi-
LVsþRV 134 21 ms, and LVs pacing 135 23 ms) (all
nition. QRS duration, measured as the time between
p < 0.05 compared with baseline). HB pacing (n ¼ 16)
earliest
resulted in a QRS duration of 110 17 ms (p < 0.05
deviation
and
end
QRS
complex,
was
353
Salden et al.
JACC VOL. 75, NO. 4, 2020
LV Septal Pacing for CRT
FEBRUARY 4, 2020:347–59
F I G U R E 5 Hemodynamic Effects
Subgroup (n = 16)
†
30
*
*
*
LVs
HB
LVs
* Increase in LVdP/dtmax (%) Compared to Baseline (AAI)
354
*
20
10
0
BiV
LVs+RV
Relative change in first derivative of left ventricular pressure (LVdP/dtmax) compared with baseline (n ¼ 27) during BiV, LVs in combination with RV, and LVs pacing alone and in a subgroup (n ¼ 16) during HB and LVs pacing. Results are presented as mean SD. *p < 0.05 versus BL; †p < 0.05 LVsþRV versus LVs. Abbreviations as in Figures 1, 3, and 4.
compared with baseline). The time from pacing
significantly reduced the increase in LVdP/dtmax as
stimulus to QRS end was comparable for all 4 CRT
compared with LVs pacing alone to 11 9% from
conditions (Figure 4).
baseline. In the subgroup analysis (n ¼ 16), with HB
Baseline QRS area was 116 44 m Vs. BiV pacing
pacing, the increase in LVdP/dtmax was not signifi-
significantly reduced QRS area compared to baseline
cantly different between LVs and HB pacing (20 8%
to 93 26 m Vs, whereas LVs further reduced QRS area
vs. 18 10%, respectively).
to 73 22 m Vs (p < 0.05 compared with baseline and
LVs PACING SITE. No significant differences were
BiV pacing). LVsþRV pacing did not significantly
found in changes in LVdP/dtmax and SDAT between
reduce the QRS area. The reduction in QRS area by
the basal, mid-, and apical LVs levels. QRS area during
LVs was similar to that by HB pacing (72 17 mVs vs. to 74 36 mVs, respectively) (Figure 4). For the entire cohort, baseline SDAT was 39 6 ms. BiV pacing reduced SDAT significantly to 31 7 ms, whereas LVs pacing further decreased SDAT to 26
basal and mid-LVs pacing was significantly smaller than during apical LVs pacing (Figure 6). Of note, LVs pacing, taken into account all performed LVs pacing measurements (basal, mid-, and apical septum) in this study (n ¼ 80) that resulted in at least 10%
7 ms (p < 0.05 compared with baseline and BiV pac-
reduction in SDAT from baseline had a significantly
ing). HB pacing resulted in a significant SDAT reduc-
larger hemodynamic benefit than LVs pacing, which
tion compared with baseline to 27 7 ms, which was comparable to the reduction by LVs pacing in this subgroup (27 6 ms) (Figure 4).
reduced SDAT by <10% (LVdP/dtmax 20 9% vs. 9 8%, respectively; p < 0.05). The same was found for LVs pacing with $10% versus <10% reduction in QRS
ACUTE HEMODYNAMIC EFFECTS. Baseline LVdP/dtmax
area (LVdP/dtmax 19 8% vs. 9 10%, respec-
was 823 225 mm Hg/s. BiV and LVs pacing resulted
tively; p < 0.05).
in significant increases in LVdP/dtmax, relative to
Changes in QRS area, SDAT, and LVdP/dtmax by
baseline (17 9% vs. 17 10%, respectively)
mid-LVs pacing were significantly larger than those
(Figure 5). Interestingly, combined LVsþRV pacing
by mid-RVs pacing (71 23 m Vs vs. 121 43 m Vs,
JACC VOL. 75, NO. 4, 2020
Salden et al.
FEBRUARY 4, 2020:347–59
LV Septal Pacing for CRT
29 10 ms vs. 40 7 ms, and 1 9% vs. 16 10%, respectively).
F I G U R E 6 Results per LVs Pacing Site
BiV pacing using LV proximal (3 and 4) or distal (1
†
and 2) pacing electrodes resulted in similar SDAT (31
200
7 ms vs. 32 8 ms, respectively) and LVdP/dtmax (17
†
with the proximal electrodes resulted in a significantly larger reduction in QRS area from baseline (from 115 45 mVs to 86 28 m Vs) than pacing with the distal electrodes (from 119 48 m Vs to 103 29 m Vs).
DISCUSSION
QRS Area (μVs)
8% vs. 15 10%, respectively). However, BiV pacing
150
100
hemodynamic improvement that is at least as good as ond, basal, mid-, and apical LVs pacing positions
40
resulted in similar results, indicating that within the
pacing creates better electrical resynchronization than BiV pacing, as evidenced by larger reductions in from previous animal studies, showing that pacing cial endocardial fibers, not necessarily Purkinje fibers,
which takes 30 to 70 ms in LBBB (23,24). Other factors may be that the LVs is the location where electrical impulses break out of the rapid conduction system (25) and that activation occurs from endocardium to epicardium. The latter is expected to generate more simultaneous LV activation. Previous studies indicate that SDAT is a more accurate predictor of CRT response than QRS duration (18,19). Interestingly, SDAT was similar during
20
n = 23
n = 25
n = 21
*
*
*
n = 23
n = 26
n = 22
Basal
Mid
Apical
30 Increase in LVdP/dtmax (%) Compared to Baseline (AAI)
slow conduction across the interventricular septum,
30
0
the LV at any endocardial position captures superfi-
tion (4,21,22). Moreover, pacing the LVs bypasses
*
10
QRS area and SDAT. Potential explanations come
providing a physiological way of LV resynchroniza-
*
n = 21
* SDAT (ms)
important finding of the present study is that LVs
n = 25
50
during BiV and HB pacing (Central Illustration). Sec-
ELECTROPHYSIOLOGICAL EFFECTS OF LVs PACING. An
n = 22 0
results in short-term electrical resynchronization and
LVs, the position of the pacing electrode is not critical.
*
50
The present proof-of-principle study demonstrates that in patients with an indication for CRT, LVs pacing
*
20
10
0
HB and LVs pacing in the present study, indicating comparable synchronicity of ventricular activation
QRS area (upper panel), SDAT (middle panel), and relative increase in LVdP/
by HB and LVs pacing. A potential explanation for
dtmax (lower panel) during pacing at 3 levels of the LVs. Results are pre-
this finding comes from previous animal experi-
sented as mean SD. *p < 0.05 versus BL; †p < 0.05 between LVs levels.
ments, which showed a sinus-rhythm–like activation
Abbreviations as in Figures 1, 3, 4, and 5.
pattern when pacing into septal tissue below the point at which HB penetrates the central fibrous body (26). Conceptually, LVs pacing would not be effective if
these patients is quite proximal in the His-Purkinje system, as recently described by Upadhyay et al. (27).
most of the conduction delay in CRT patients were
Furthermore, the present study corroborates data
peripheral. The observation that HB pacing was able
from previous studies that there is a considerable
to reduce QRS duration to an average of 110 ms in our
difference in electrophysiological effects between
study indicates that most of the conduction delay in
RVs and LVs pacing (4,28). Between RVs pacing and
355
356
Salden et al.
JACC VOL. 75, NO. 4, 2020
LV Septal Pacing for CRT
FEBRUARY 4, 2020:347–59
C E NT R AL IL L U STR AT IO N Left Ventricular Septal Pacing for Cardiac Resynchronization Therapy
Salden, F.C.W.M. et al. J Am Coll Cardiol. 2020;75(4):347–59.
The present study compared acute electrophysiological and hemodynamic effects of left ventricular (LV) septal, biventricular, and His bundle pacing in patients receiving cardiac resynchronization therapy (upper left). Electrophysiological changes were assessed using multi-electrode body-surface mapping (upper right) as standard deviation of activation times. Hemodynamic changes were assessed as LVdP/dtmax. The results indicate that LV septal pacing provides acute hemodynamic improvement (lower left) and electrical resynchronization (lower right), that is at least as good as during biventricular and possibly His bundle pacing.
LVs endocardium activation is an approximately
conventional CRT or because of inability to access the
40-ms delay, and it takes another 50 ms to conduct
coronary sinus. The present study, however, was
the impulse around the LV endocardium. During LVs
prospective, and none of the patients had undergone
pacing, the transseptal conduction occurs (in the
prior attempts for CRT implantation, thereby better
rightward direction) simultaneous with the circum-
reflecting regular clinical practice.
ferential LV endocardial conduction, thus reducing
An unexpected and interesting finding in this
total LV activation time significantly (4).
study is that combined LVsþRV pacing provided a
HEMODYNAMIC
PACING. The
significantly smaller hemodynamic response than LVs
beneficial electrophysiological consequences of LVs
pacing alone. The explanation for this finding is not
pacing appear to translate to beneficial hemodynamic
clear and requires further investigation.
EFFECTS
OF
LVs
effects, especially because LVs pacing with $10%
The similarity of hemodynamic effects of HB
reduction in SDAT or QRS area was associated with a
and LVs pacing is in agreement with the results in
significantly larger hemodynamic response. The
a previous study in patients with sinus node disease.
similarity of hemodynamic effects of LVs and BiV
In these patients, LVs pacing did not induce a sig-
pacing is in line with previous animal studies as well
nificant reduction in LVdP/dtmax as compared with
as a preliminary patient study (5). In this small study,
intrinsic sinus rhythm, despite that LVs pacing
patients (n ¼ 12) underwent investigation of alterna-
increased QRS duration in these patients with a
tive pacing locations because of nonresponse to
narrow QRS complex (6).
JACC VOL. 75, NO. 4, 2020
Salden et al.
FEBRUARY 4, 2020:347–59
LV Septal Pacing for CRT
The finding that HB and BiV pacing provided
the finding that the reduction in SDAT and the in-
similar increases in LVdP/dtmax compares also with
crease in LVdP/dtmax were virtually independent of
previous studies showing no significant differences in
the LVs pacing location. However, Hou et al. (28)
echocardiographic response between HB and BiV
indicate that LBB capture may provide superior
pacing (9,29). By contrast, the recent study by Arnold
benefit compared with myocardial capture, albeit in a
et al. (11) showed that HB pacing provides a larger
population of patients with bradyarrhythmias and
increase in systolic arterial blood pressure, albeit in a
LVEF
pre-selected patient cohort. CLINICAL IMPLICATIONS. Although the present study
compare the electrophysiological and hemodynamic
>55%.
Additional
research
is
needed
to
effects of LV septal pacing with LBB area pacing.
used temporary LVs pacing, several practical benefits
This wide range in LVs pacing location possibility
of single-lead LVs pacing may be envisioned for future
would make LVs lead implantation relatively easy,
clinical application. With a similar hemodynamic
because lead positioning is less critical than during
benefit of LVs, BiV, and HB pacing, LVs may have
HB and LBBB area pacing (12). Further disadvantages
several advantages once a market-released pacing lead
of HB pacing are various sensing issues, high pacing
is available for this purpose. A custom, investigational
thresholds,
lead has been successfully employed in animal studies
implantation leading to more lead revisions and/or
(4,30) and in 10 patients with sinus node dysfunction,
faster battery drainage (12,13). In the experience with
using a transvenous approach by positioning the lead
long-term LVs pacing in patients with sinus node
deep into the interventricular septum (6). Although
disease, LVs pacing was performed from the tip of the
and
further
rising
thresholds
post-
these studies were performed with a modified version
lead with capture thresholds of 0.5 0.2 V. No lead-
of the Medtronic 3830 lead (extended helix), Huang
related complications were observed, and leads
et al. (31) were able to perform deep septal (LBB area)
remained electrically and mechanically stable during
pacing, using the same approach, by implantation of
the 6-month follow-up (6) (and are still so after 3
the standard Medtronic 3830 lead.
years) (Vernooy, unpublished data, April 2019).
As compared with BiV pacing, LVs pacing has the advantages that it avoids potential problems with lead placement in a coronary venous tributary and phrenic nerve stimulation (32). Moreover, the essentially epicardial stimulation, generated by pacing in a coronary vein, creates a less physiological, epicardialto-endocardial, activation sequence, which has been considered as a cause of some proarrhythmic effects of CRT (33). More physiological activation is also obtained using endocardial CRT. However, endocardial CRT is still in its childhood, associated with a lack of proper tools and a relatively high percentage of adverse events (7,8). Recently, pioneering studies showed the feasibility of pacing in or near the LBB (34,35). LBB area pacing is
STUDY LIMITATIONS. The present proof-of-principle
study investigated the short-term response of temporary pacing at the LVs. For long-term LVs pacing, the aforementioned transvenous transventricular septal approach can be used. This may lead to slightly different activation patterns because the penetrating lead may not entirely reach the endocardium. HB pacing was also performed using an EP catheter, which resulted in nonselective HB capture in most cases. Therefore, QRS durations and SDAT may be larger compared with selective HB capture. Finally, there are controversial data as to whether the shortterm increase in LVdP/dtmax translates to a longterm
benefit
of
CRT
(36,37),
especially
when
comparing to short-term changes in stroke work (38).
supposed to capture the rapid conduction system. This LBB area pacing is performed by a similar
CONCLUSIONS
transventricular septal approach as in long-term LVs pacing, although across the most basal part of the
LVs pacing results in short-term hemodynamic
septum (31,34,35). In our LVs pacing approach, pacing
improvement and electrical resynchronization that is
locations were not chosen on the basis of a “His-
at least as good as BiV and possibly HB pacing. These
Purkinje signal” on the local electrogram. However,
results suggest that LVs pacing with a single ven-
the QRS morphology that we present in Figure 3A has
tricular lead may serve as a novel, valuable, and
some similarities to the LBB area–paced QRS mor-
feasible alternative way to apply CRT.
phologies described by Huang et al. (31). These similarities may be explained by accidental capture of the
ADDRESS FOR CORRESPONDENCE: Dr. Floor C.W.M.
His-Purkinje system during our LVs approach. Alter-
Salden, Maastricht University, Department of Physi-
natively, pacing the rapid conduction along non-
ology,
Purkinje endocardial fibers may achieve the same
Netherlands. E-mail:
[email protected].
benefit as pacing the LBB. The latter is supported by
Twitter: @FloorSalden.
P.O.
Box
616,
6200
MD
Maastricht,
the
357
358
Salden et al.
JACC VOL. 75, NO. 4, 2020
LV Septal Pacing for CRT
FEBRUARY 4, 2020:347–59
PERSPECTIVES COMPETENCY IN PATIENT CARE AND
TRANSLATIONAL OUTLOOK: Long-term clinical
PROCEDURAL SKILLS: Pacing the cardiac conduction
studies are needed to assess the utility of a left ventric-
system along Purkinje fibers and non-Purkinje left
ular septal pacing approach to cardiac resynchronization
ventricular endocardial fibers may provide an alternative
therapy.
to biventricular, His bundle, and left bundle branch stimulation for cardiac resynchronization therapy.
REFERENCES 1. Cleland JGF, Daubert J-C, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005; 352:1539–49. 2. McAlister FA, Ezekowitz JA, Wiebe N, et al. Systematic review: cardiac resynchronization in patients with symptomatic heart failure. Ann Intern Med 2004;141:381. 3. Peschar M, De Swart H, Michels KJ, et al. Left ventricular septal and apex pacing for optimal pump function in canine hearts. J Am Coll Cardiol 2003;41:1218–26.
patients with heart failure and left bundle branch block. J Am Coll Cardiol 2018;72:3112–22.
cardiac resynchronization therapy. PACE 2011;34: 217–25.
12. Vijayaraman P, Dandamudi G, Zanon F, et al. Permanent His bundle pacing: recommendations from a Multicenter His Bundle Pacing Collaborative Working Group for standardization of definitions, implant measurements, and follow-up. Heart Rhythm 2018;15:460–8.
21. van Deursen C, van Geldorp IE, Rademakers LM, et al. Left ventricular endocardial pacing improves resynchronization therapy in canine left bundle-branch hearts. Circ Arrhythmia Electrophysiol 2009;2:580–7.
13. Zanon F, Ellenbogen KA, Dandamudi G, et al. Permanent His-bundle pacing: a systematic literature review and meta-analysis. Europace 2018;
Castellanos A, Morales A, Bassett A. The role of canine superficial ventricular muscle fibers in endocardial impulse distribution. Circ Res 1978;42: 27–35.
20:1819–26.
4. Mills RW, Cornelussen RN, Mulligan LJ, et al. Left ventricular septal and left ventricular apical pacing chronically maintain cardiac contractile coordination, pump function and efficiency. Circ Arrhythm Electrophysiol 2009;2:571–9.
14. Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013 ESC guidelines on cardiac pacing and
5. Rademakers LM, van Hunnik A, Kuiper M, et al. A possible role for pacing the left ventricular septum in cardiac resynchronization therapy. J Am
Reconstruction of the Frank vectorcardiogram from standard electrocardiographic leads: diagnostic comparison of different methods. Eur Heart J 1990;11:1083–92.
Coll Cardiol EP 2016;2:413–22. 6. Mafi-Rad M, Luermans JGLM, Blaauw Y, et al. Feasibility and acute hemodynamic effect of left ventricular septal pacing by transvenous approach through the interventricular septum. Circ Arrhythmia Electrophysiol 2016;9:e003344. 7. Morgan JM, Biffi M, Geller L, et al. ALternate Site Cardiac ResYNChronization (ALSYNC): a prospective and multicentre study of left ventricular endocardial pacing for cardiac resynchronization therapy. Eur Heart J 2016;37:2118–27. 8. Gamble JHP, Herring N, Ginks MR, Rajappan K, Bashir Y, Betts TR. Endocardial left ventricular pacing across the interventricular septum for cardiac resynchronization therapy: clinical results of a pilot study. Heart Rhythm 2018;15:1017–22. 9. Upadhyay GA, Vijayaraman P, Nayak HM, et al. On-treatment comparison between corrective His bundle pacing and biventricular pacing for cardiac resynchronization: a secondary analysis of HisSYNC. Heart Rhythm 2019;16:1797–807. 10. Narula OS. Longitudinal dissociation in the His bundle. Bundle branch block due to asynchronous conduction within the His bundle in man. Circulation 1977;56:996–1006. 11. Arnold A, Shun-Shin M, Keene D, et al. His resynchronization versus biventricular pacing in
cardiac resynchronization therapy. Eur Heart J 2013;34:2281–329. 15. Kors J, van Herpen G, Sittig A, van Bemmel J.
16. Engels EB, Mafi-Rad M, van Stipdonk AMW, Vernooy K, Prinzen FW. Why QRS duration should be replaced by better measures of electrical activation to improve patient selection for cardiac resynchronization therapy. J Cardiovasc Transl Res 2016;9:257–65. 17. van Stipdonk AMW, ter Horst IAH, Kloosterman M, et al. QRS area is a strong determinant of outcome in cardiac resynchronization therapy. Circ Arrhythmia Electrophysiol 2018;11: e006497. 18. Johnson W, Vatterott PJ, Peterson MA, et al.
22. Myerburg
R,
Gelband
H,
Nilsson
K,
23. Strik M, van Deursen CJM, van Middendorp LB, et al. Transseptal conduction as an important determinant for cardiac resynchronization therapy, as revealed by extensive electrical mapping in the dyssynchronous canine heart. Circ Arrhythm Electrophysiol 2013;6:682–9. 24. Prinzen FW, Auricchio A. Is echocardiographic assessment of dyssynchrony useful to select candidates for cardiac resynchronization therapy? Echocardiography is not useful before cardiac resynchronization therapy if QRS duration is available. Circ Cardiovasc Imaging 2008;1:70–8. 25. Durrer D, van Dam RT, Freud GE, Janse MJ, Meijler FL, Arzbaecher RC. Total excitation of the isolated human heart. Circulation 1970;41: 899–912. 26. Laske TG, Skadsberg ND, Hill AJ, Klein GJ, Iaizzo PA. Excitation of the intrinsic conduction system through His and interventricular septal pacing. Pacing Clin Electrophysiol 2006;29: 397–405.
Body surface mapping using an ECG belt to characterize electrical heterogeneity for different left ventricular pacing sites during cardiac resynchronization: relationship with acute hemodynamic improvement. Heart Rhythm 2017;14:385–91.
27. Upadhyay GA, Cherian T, Shatz DY, et al. Intracardiac delineation of septal conduction in
19. Gage RM, Curtin AE, Burns KV, Ghosh S, Gillberg JM, Bank AJ. Changes in electrical dyssynchrony by body surface mapping predict left ventricular remodeling in patients with cardiac
28. Hou cardiac branch septum.
resynchronization therapy. Heart Rhythm 2017;14: 392–9.
left bundle branch block patterns: mechanistic evidence of left intra-Hisian block circumvented by His pacing. Circulation 2019;139:1876–88. X, Qian Z, Wang Y, et al. Feasibility and synchrony of permanent left bundle pacing through the interventricular Europace 2019;21:1694–702.
20. Whinnett ZI, Nott G, Davies JER, et al. Maxi-
29. Lustgarten DL, Crespo EM, ArkhipovaJenkins I, et al. His-bundle pacing versus biventricular pacing in cardiac resynchronization ther-
mizing efficiency of alternation algorithms for hemodynamic optimization of the AV delay of
apy patients: a crossover design comparison. Heart Rhythm 2015;12:1548–57.
JACC VOL. 75, NO. 4, 2020
Salden et al.
FEBRUARY 4, 2020:347–59
LV Septal Pacing for CRT
30. Grosfeld MJW, Res JCJ, Vos DHS, de Boer TJM, Bos HJ. Testing a new mechanism for left interventricular septal pacing: the transseptal route. Europace 2002;4:439–44.
34. Vijayaraman P, Subzposh FA, Naperkowski A, et al. Prospective evaluation of feasibility, electrophysiologic and echocardiographic characteristics of left bundle branch area pacing. Heart Rhythm 2019;16:1774–82.
31. Huang W, Su L, Wu S, Xu L, Xiao F. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the
35. Li X, Li H, Ma W, et al. Permanent left bundle branch area pacing for atrioventricular block:
conduction block. Can J Cardiol 2017;33. 1736.e1– 3.
feasibility, safety and acute effect. Heart Rhythm 2019;16:1766–73.
32. Sieniewicz BJ, Gould J, Porter B, et al. Understanding non-response to cardiac resynchronisation therapy: common problems and potential solutions. Heart Fail Rev 2019;24:41–54.
36. de Roest G, Knaapen P, Gotte M, et al. Stroke work or systolic dP/dtmax to evaluate acute response to cardiac resynchronization therapy: are they interchangeable? Eur J Heart Fail 2009;11: 706–8.
33. Fish JM, Brugada J, Antzelevitch C. Potential proarrhythmic effects of biventricular pacing. J Am Coll Cardiol 2005;46:2340–7.
37. Duckett SG, Ginks M, Shetty AK, et al. Invasive acute hemodynamic response to guide
left ventricular lead implantation predicts chronic remodeling in patients undergoing cardiac resynchronization therapy. J Am Coll Cardiol 2011; 58:1128–36. 38. Zweerink A, Salden OAE, van Everdingen WM, et al. Hemodynamic optimization in cardiac resynchronization therapy: should we aim for dP/ dtmax or stroke work? J Am Coll Cardiol EP 2019; 5:1013–25.
KEY WORDS body surface mapping, cardiac resynchronization therapy, heart failure, hemodynamics, His bundle pacing, ventricular septum
359