Accepted Manuscript Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes Supriya Lath, Emma R. Knight, Divina A. Navarro, Rai S. Kookana, Michael J. McLaughlin PII:
S0045-6535(19)30103-1
DOI:
https://doi.org/10.1016/j.chemosphere.2019.01.096
Reference:
CHEM 23011
To appear in:
ECSN
Received Date: 14 October 2018 Revised Date:
12 January 2019
Accepted Date: 16 January 2019
Please cite this article as: Lath, S., Knight, E.R., Navarro, D.A., Kookana, R.S., McLaughlin, M.J., Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes, Chemosphere (2019), doi: https://doi.org/10.1016/j.chemosphere.2019.01.096. This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
ACCEPTED MANUSCRIPT 1
Sorption of PFOA onto different laboratory materials: filter membranes and centrifuge
2
tubes
3 Supriya Lath (
[email protected]) *, a, 1
5
Emma R. Knight (
[email protected]) a, b
6
Divina A. Navarro (
[email protected]) a, b
7
Rai S. Kookana (
[email protected]) b
8
Michael J. McLaughlin (
[email protected]) a, b
M AN U
SC
RI PT
4
9 10 11
a
School of Agriculture Food and Wine, The University of Adelaide, PMB 1 Glen Osmond, SA 5064, Australia. b
13
17
EP
1
Present address: AECOM, Collins Square, Level 10, Tower Two 727 Collins Street, Melbourne, VIC 3008.
AC C
16
CSIRO Land and Water, PMB 2 Glen Osmond, SA 5064, Australia. * Corresponding author (
[email protected])
14 15
TE D
12
1
ACCEPTED MANUSCRIPT Abstract
19
Measurement and reporting of concentrations of contaminants of emerging concern such as
20
per- and polyfluoroalkyl substances (PFASs), including perfluorooctanoic acid (PFOA), is an
21
integral part of most investigations. Sorption losses of PFAS analytes onto particular
22
laboratory-ware (e.g. glass containers) have been suggested in the published literature but
23
have not been investigated in detail. We examined sorption losses from aqueous PFOA
24
solutions in contact with different commonly-used materials in filter units and centrifuge
25
tubes (glass and plastics). Sorption of PFOA onto different filter membrane types ranged
26
from 21 to 79 % indicating that filtration can introduce a major source of error in PFOA
27
analysis; pre-treatment of filter membranes with phosphate or methanol solutions did not
28
improve PFOA recovery. Substantial adsorption of PFOA was also observed on tubes made
29
from polypropylene (PP), polystyrene (PS), polycarbonate (PC), and glass where losses
30
observed were between 32 to 45%, 27 to 35%, 16 to 31% and 14 to 24%, respectively.
31
Contrary to suggestions in the literature, our results indicated the greatest sorption losses for
32
PFOA occurred on PP, whereas losses on glass tubes were much lower. Variations in ionic
33
strength and pH did not greatly influence PFOA recovery. When PFOA concentrations were
34
increased the percent recovery of PFOA increased, indicating that binding sites on tube-
35
walls were saturable. This study draws attention towards analytical bias that can occur due
36
to sorption losses during routine procedures, and highlights the importance of testing the
37
suitability of chosen laboratory-ware for specific PFAS analytes of interest prior to
38
experimental use.
39
Keywords: PFOA; filter membranes; centrifuge tubes; plastics; polypropylene; glass
AC C
EP
TE D
M AN U
SC
RI PT
18
40
2
ACCEPTED MANUSCRIPT 41
1. Introduction
Perfluorooctanoic acid (PFOA) and other related per- and polyfluoroalkyl substances
43
(PFASs) have been recognised as contaminants of emerging concern due to their ubiquitous
44
and persistent nature in the environment, as well as their bioaccumulative properties. As a
45
result, these chemicals are being studied extensively with respect to their human and
46
ecological toxicity (Hekster et al., 2003, Sundström et al., 2011), their occurrence, fate and
47
transport in different environmental compartments (Ahrens et al., 2015, Hansen et al., 2002)
48
as well as management and remediation strategies (Ochoa-Herrera and Sierra-Alvarez,
49
2008, Ross et al., 2018). At various stages of such laboratory and field studies, the PFAS
50
analytes being researched come in contact with a variety of apparatus that are usually made
51
from glass, steel or plastics. The most common apparatus in any study are sampling and
52
storage containers, including a range of tubes, vials and bottles. Others examples include
53
disposable polystyrene (PS) well-plates used as exposure-vessels for toxicity assays, glass
54
aquariums for fish toxicity tests, as well as disposable filtration membranes used for the
55
separation of aqueous phases from solid or particulate matrices prior to analysis.
56
Regardless of the types of experiments conducted, one aspect that is commonly cited in the
57
methods’ section of several published studies relates to the use of sample containers made
58
of particular materials. Specifically, several studies exclude the use of equipment made from
59
glass in experimental or analytical protocols involving PFASs (Ahrens et al., 2015, Hansen et
60
al., 2002). The justification regularly provided for this is that glass adsorbs PFAS analytes.
61
The USEPA and ISO methods, which are the most widely accepted test methods for PFAS
62
analytes, also stipulate that PFAS standards, extracts and samples should not come in
63
contact with any glass containers or pipettes (ISO, 2009, Shoemaker et al., 2009). They
64
recommend that polypropylene (PP) containers be used for all sample, standard and
65
extraction preparation and storage, and suggest that other plastics may be used if they meet
66
quality control requirements (Shoemaker et al., 2009). As a result, PP has been adopted as
AC C
EP
TE D
M AN U
SC
RI PT
42
3
ACCEPTED MANUSCRIPT the material of choice in several studies (Ahrens et al., 2010, Ahrens et al., 2011, Hellsing et
68
al., 2016, Higgins et al., 2007). However, a number of studies have used materials other
69
than PP for PFAS-related studies. For example, Higgins and Luthy (2006) chose PS tubes
70
over PP or glass because their preliminary, unpublished data provided higher recoveries
71
when using PS tubes for a range of PFASs. Other studies have used polyethylene (PE),
72
polycarbonate (PC) and high-density PE containers (Johnson et al., 2007, Washington et al.,
73
2010), often without any indication as to the suitability of these materials.
74
One consensus in the published literature is the avoidance of equipment containing
75
fluoropolymer materials like polytetrafluoroethylene (PTFE) in PFAS studies, as their
76
manufacture has historically involved the use of some PFASs as a ‘polymerisation aid’
77
(Prevedouros et al., 2006). Leaching of remnant PFASs from such products can cause
78
contamination of the dissolved phase (Martin et al., 2004). However, in the case of glass and
79
plastics, there is considerable contradiction and inconsistency in the published literature
80
regarding which materials may be best suited. Losses to laboratory ware could lead to
81
considerable bias in analytical data. Despite the severe implications of such routine losses
82
due to adsorption onto glass and plastics, it has not been investigated in its own right.
83
Apart from choice of containers, filtration is another routine consideration in sample handling
84
and preparation. Filtration is often employed as a major clean-up stage for most
85
environmental and laboratory samples, during which additional losses can occur (Ahmad et
86
al., 2001, Carlson and Thompson, 2000). A study investigating active air-sampling of
87
gaseous PFOA using glass fibre (GF) filters observed that gas-phase PFOA was
88
underestimated due to its sorption onto the GF filters (Johansson et al., 2017). However, an
89
investigation using a variety of PFASs from the aqueous phase, testing four different filter
90
membranes, determined GF filters to be suitable for several PFASs, but recommended
91
specific testing to account for unpredictable effects (Chandramouli et al., 2015).
AC C
EP
TE D
M AN U
SC
RI PT
67
4
ACCEPTED MANUSCRIPT The aim of this study was to examine the sorption losses of PFOA on common glass and
93
plastic materials – specifically, centrifuge tubes and disposable syringe filter membranes –
94
during routine laboratory procedures. Sorption of PFOA from aqueous solutions onto a
95
variety of filter-membranes was tested; the influence of varying concentrations, and two pre-
96
treatments were investigated. Additionally, sorption onto a variety of glass and plastic (PP,
97
PC, PS) tubes was tested. The influence of contact time, pH, ionic strength and PFOA-
98
concentrations were examined to gain insights into the nature of PFOA-binding interactions.
100
2. Materials and Methods 2.1.
SC
99
RI PT
92
Materials
Radiolabelled 14C-PFOA with a specific activity of 2.04 GBq/mmol was purchased from
102
American Radiolabelled Chemicals Incorporation. Optiphase HiSafe 3 scintillation fluid for
103
radiochemical analysis was purchased from PerkinElmer, Australia. All other chemicals used
104
including calcium chloride dihydrate (CaCl2.2H2O), dipotassium hydrogen phosphate
105
(K2HPO4), methanol, hydrochloric acid (HCl) and sodium hydroxide (NaOH) were of
106
analytical grade. The specifications of filter-types and tube-types used in this study are listed
107
in Table 1 and Table 2 respectively.
TE D
2.2.
Filter sorption studies
EP
108
M AN U
101
The sorptive losses through retention of dissolved PFOA onto different filter membrane-
110
types was investigated in triplicate using syringe-filtration through disposable filter units. By
111
means of single-use 10 mL PP syringes, 4 mL of a 14 µg/L PFOA solution was drawn into
112
the syringes, following which filter units were attached to the end of the syringe before
113
plunging the solution through the filter membrane. The initial 2 mL volume of the filtrate was
114
discarded and the subsequent 2 mL was collected, of which 0.5 mL aliquots were used for
115
quantitative analysis. Control samples to account for any sorption occurring on the walls of
116
the syringe were prepared by taking up PFOA solution in the syringe, and dispensing the
117
solutions without attaching any filter units (details in supplementary material). To determine if
AC C
109
5
ACCEPTED MANUSCRIPT recovery of PFOA from the filters could be improved, two pre-rinsing treatments, using
119
phosphate solution (100 mM K2HPO4) and methanol, were applied to the filters. For the
120
phosphate treatment, 10 mL of a 100 mM K2HPO4 solution was plunged through the filter
121
units, followed by 10 mL of Milli-Q water, prior to PFOA filtration. Similarly, for the methanol
122
treatment, 10 mL of methanol was plunged through the filters; membrane-types that were
123
incompatible with methanol were excluded. Finally, the influence of increasing
124
concentrations of PFOA on sorptive losses on three filter membranes (PP, RC and GF) was
125
also tested using an environmentally relevant range of concentrations (0 to 529 µg/L). The
126
critical micelle concentration (CMC) of PFOA has been reported to be in the range of 1 to 1.6
127
x 107 µg/L (i.e., 25 to 38 mM), at which PFOA molecules can agglomerate to form micelles
128
and hemi-micelles; such phase separation can affect interfacial activity like sorption (Harada
129
et al., 2005, Rattanaoudom et al., 2012). The concentrations used in our studies were
130
around 104 to 106 times lower than the CMC, so the impact of micelle-formation on sorption
131
losses is expected to be negligible.
SC
Tube sorption studies
M AN U
2.3.
TE D
132
RI PT
118
To measure sorption of PFOA onto different centrifuge tube-types (PP, PS, PC, G1 and G2,
134
Table 2), tests were performed in triplicate under different conditions by varying the contact
135
times, solution pH, ionic strength as well as PFOA concentrations. When not being varied,
136
the standard test parameters included the use of 8 mL volumes of PFOA solutions of
137
concentrations of 20.5 ± 1 µg/L, prepared in a 10 mM CaCl2 background electrolyte solution
138
or Milli-Q water, and a contact time of 24 to 48 hrs. To investigate the influence of contact
139
time, tubes were subjected to shaking times of 1, 2 and 7 days. The influence of pH was
140
examined across a pH range from 4 to 9 where 0.1M HCl or 0.1M NaOH was used to adjust
141
solution pH. Background electrolyte solutions of varying concentrations (0 to 100 mM CaCl2)
142
were used to determine the effect of ionic strength on retention of PFOA. Sorptive losses as
143
a function of concentration were also examined by testing initial PFOA concentrations
144
ranging from 0 to 415 µg/L. All tubes with test PFOA solutions were placed on an end-over-
AC C
EP
133
6
ACCEPTED MANUSCRIPT 145
end shaker for the duration of the test, after which subsamples (0.5 mL) were taken for
146
quantitative analysis. 2.4.
147
Instrumental analysis
Aliquots (0.5 mL) of all samples, blanks (Milli-Q water) and stock solutions were combined
149
with 4 mL of scintillation fluid in a scintillation vial and analysed radio-chemically using a β-
150
liquid scintillation counter (Perkin Elmer Tri-Carb 3110 RT). Self-normalisation and
151
calibration (SNC) protocol was performed and the 14C counting efficiency was determined to
152
be >90%. The activity of 14C in the samples was measured as disintegrations per minute;
153
each sample was measured three times, for 2 min each. At least 3 background vials
154
(containing 0.5mL of Milli-Q water instead of the sample) were run with each batch of
155
samples to determine the background count, which was subtracted from each sample count.
156
Concentrations of PFOA were then calculated using the specific activity and the measured
157
14
158
to be 0.11 µg/L.
SC
M AN U
C activity. The limit of detection for the method calculated using the blanks was determined
2.5.
TE D
159
RI PT
148
Data and statistical analysis
All experiments were carried out at least in triplicate. The coefficient of variations for the
161
measurement of all stock solutions used in the experiments ranged from 0.35 to 4.83%.
162
Losses of PFOA through adsorption onto the filters and tubes were calculated using the
163
difference between the amounts of PFOA in the samples before and after the experimental
164
steps (i.e. filtration or shaking). Results are reported as a percentage recovery of PFOA
165
compared to the controls. A mass balance was not attempted as each additional handling,
166
extraction and transfer of every sample would increase the chances for loss of analyte,
167
introducing additional error to the method.
168
Statistical software, IBM SPSS (v. 24), was used to determine if there were significant
169
differences (p < 0.05) between treatments. After using analysis of variance test (ANOVA), a
170
Tukey’s honestly significant difference test was used to compare treatment means.
AC C
EP
160
7
ACCEPTED MANUSCRIPT 171
3. Results and Discussion
172
3.1.
173
3.1.1. Sorption losses observed on different filter membrane-types
Filter sorption studies
The sorption of dissolved PFOA onto filter membranes was determined using syringe-
175
filtration. As no sorption of PFOA analyte was identified as occurring on the syringe-surfaces
176
(Table 1, supplementary material) likely due to short residence time (< 10 seconds) in the
177
syringes, the sorption losses were calculated using the difference between the amounts of
178
PFOA in the unfiltered and the filtered solutions. Irrespective of the type of membrane
179
tested, recovery of PFOA was < 100% in all cases (Figure 1). This implies that at least a
180
certain proportion of PFOA was retained onto the filter membrane or housing material due to
181
adsorption, leading to underestimation of dissolved PFOA concentrations. As it was not
182
possible to make a distinction between the amount of PFOA retained on the membrane and
183
the housing material separately, sorption was considered for the filter unit as a whole.
184
Specifically, recovery ranged from 76.0% at best to 21.2% at worst, depending on the type of
185
membrane used (Figure 1). The highest recoveries were achieved when using PVDF
186
(76.0%), glass-fibre (74.2%), RC (74.0%) and PP (72.3%) membranes. The lowest recovery
187
due to adsorption was displayed by the NY membrane filter (21.2%). Overall, the percentage
188
recoveries after filtration were in the order PVDF ≈ GF ≈ RC ≈ PP > CA > PES ≈ PTFE > NY.
189
While PVDF showed a high recovery, it is a fluoropolymer, and like PTFE, may cause
190
contamination of analytical blanks, or over-estimation of dissolved PFAS concentrations
191
(Martin et al., 2004), so it should generally be avoided in PFAS-based analysis. We did not
192
test for PFAS contamination issues during filtration as a radiolabelled PFOA solution was
193
used. Interestingly, despite both PVDF and PTFE being fluoropolymers, PFOA-sorption on
194
PTFE was more than 2 times greater than on PVDF; this may be attributed to the differences
195
in their chemical properties. The carbon-backbone of the polymeric structure in PVDF
196
(chemical formula [–CH2–CF2–]n) is partially fluorinated, whereas that in PTFE (chemical
AC C
EP
TE D
M AN U
SC
RI PT
174
8
ACCEPTED MANUSCRIPT formula [–CF2–CF2–]n) is fully fluorinated, making PTFE more hydrophobic than PVDF. It is
198
likely that the PFOA tail, being hydrophobic, is able to undergo greater hydrophobic
199
interactions with PTFE, leading to greater PFOA sorption.
200
When testing a range of filter media with PFAS-spiked water, Chandramouli et al. (2015)
201
reported that the poorest recoveries were observed with PTFE (2 to 24%), followed by NY
202
(62 to 80%) filters, while GF filters displayed the best recovery (> 85%) overall. For a range
203
of perfluorinated carboxylate and sulphonate compounds, Labadie and Chevreuil (2011)
204
also reported better recoveries using GF (70 to 98%), as opposed to NY (40 to 98%);
205
however, specifically in the case of PFOA, GF and NY were reported to perform equally.
206
Like in the case of GF membrane, the recoveries obtained from RC as well as PP
207
membranes in our study were greater compared to other filters, however accounts of the use
208
of these membranes in the published literature are rare. Chain-length of the PFASs were an
209
important factor controlling sorption losses reported in these studies (Chandramouli et al.,
210
2015, Labadie and Chevreuil, 2011), where greater sorption occurred as chain-length
211
increased. This has been reported in studies related to sorption in environmental media such
212
as soil and sediments, as well as adsorbents such as activated carbon (Du et al., 2014).
213
While the literature focussing on PFAS-investigations in water, soil and sediments commonly
214
use GF filters (Ahrens et al., 2015, Kwadijk et al., 2010, Lein et al., 2008), the literature
215
dealing with air-sampling of PFAS is exploring ways to reduce the sorption of PFAS onto
216
GF-filters (Arp and Goss, 2008, Johansson et al., 2017). This contrast may be attributed to
217
the differences in sampling and filtration equipment set-up for different environmental media.
218
Moreover, it appears that membrane-types, apart from GF and quartz-fibre filters, have not
219
been tested for PFAS sorption.
220
While GF filters have been reported to show the best recoveries for aqueous phases
221
(Chandramouli et al., 2015, Labadie and Chevreuil, 2011), it is important to note that glass
222
fibres used in membranes usually have variable structural integrity. As a result, they are
AC C
EP
TE D
M AN U
SC
RI PT
197
9
ACCEPTED MANUSCRIPT often ascribed with a particle retention rating, covering a range of sizes, and in our case it
224
was 0.8 to 8 µm rather than a specific pore-size. The GF membranes used in this study were
225
stated to be of 20 µm pore size, but the manufacturer cannot guarantee an actual pore size.
226
An interesting observation was that the inclusion of a GF membrane as a pre-filter in the
227
case of PES and CA membranes, caused recovery to improve by up to 18% and 11%,
228
respectively, when compared to the PES and CA membranes where no pre-filter was
229
included. Usually, pre-filters are effective at improving recovery in samples with high
230
particulate load. The solutions used in this study were devoid of any particulate matter, thus
231
it is unclear why an improvement was observed.
232
One approach to reduce sorption of PFOA onto the filters could be through pre-treatment of
233
the filters to pre-saturate or block the available binding sites for PFOA on the membrane – a
234
strategy that has been demonstrated for other organic compounds (Bin et al., 2000, Hupfeld
235
et al., 2009). However, on pre-treating the membranes with either a phosphate solution (100
236
mM), or methanol (99% purity), no significant improvements in PFOA-recovery were
237
observed (Figure S1 and Figure S2; supplementary material). In the case of the methanol
238
treatment, in fact, a decrease in recovery of 7 to 28%, was observed for some membranes
239
(RC, PP and PES; Figure S2). The exact mechanisms controlling the binding of PFOA to the
240
different filter membranes are not known. In the study on active air-sampling of PFOA using
241
GF filters, pre-treatment with a siliconizing reagent led to an appreciable reduction in
242
sorption of gaseous PFOA onto the GF filters (Johansson et al., 2017), particularly when
243
atmospheric concentrations of PFOA were high. While no specific mechanism was identified,
244
it was attributed to deactivation of surface active sites on the GF filter. Given that no benefits
245
in terms of reduction in sorption losses were observed in our study, combined with the added
246
inconvenience, time and expense associated with the process, pre-treatment of disposable
247
filter-units is not recommended for improving recovery from solution. However, pre-rinsing
248
remains a suitable strategy to reduce contamination of the dissolved phase from certain
249
filters (e.g. PTFE) (Labadie and Chevreuil, 2011).
AC C
EP
TE D
M AN U
SC
RI PT
223
10
ACCEPTED MANUSCRIPT 250
3.1.2. Effect of PFOA concentration on PFOA recovery
To test the influence of increasing PFOA concentrations on recovery, three of the filters
252
exhibiting the lowest sorptive losses – PP, RC and GF – were used; PVDF was excluded on
253
account of being a fluoropolymer. In terms of percentage of PFOA recovered (Figure 2),
254
there was no evidence for consistent effect of PFOA concentration on recovery. A similar
255
outcome was reported for PFOA by Chandramouli et al. (2015) when tested on NY filters;
256
PFOA concentrations used ranged from 0.02 to 1 µg/L, which were 2 to 3 orders of
257
magnitude lower than the concentrations tested in our study. We used higher concentrations
258
with the expectation that sorption might be saturable, but this was not observed.
259
As all filter-membranes displayed some losses of PFOA through adsorption, we recommend
260
that it is better to avoid filtration, wherever possible. Other studies have also advised against
261
the use of filtration in sample-preparation steps for the same reasons (Schultz et al., 2006,
262
Voogt and Sáez, 2006). For instance, Schultz et al. (2006) elected centrifugation as their
263
only viable sample clean-up step when analysing PFAS-contaminated municipal wastewater
264
samples due to such losses. If, however, filtration is considered necessary in any
265
procedures, we advise that specific testing of analyte-sorption onto filter media is undertaken
266
to account for potential underestimation of dissolved PFAS concentrations due to such
267
losses.
268
3.2.
269
3.2.1. Effect of contact time on recovery of PFOA
AC C
EP
TE D
M AN U
SC
RI PT
251
Tube sorption studies
270
Recovery of PFOA decreased in the order G1 ≈ G2 > PC > PS > PP (Figure 3), with greater
271
recoveries from glass tubes, compared to the plastic tubes. Specifically, PFOA-recovery
272
observed from glass (G1 and G2) tubes ranged from 77 to 86%, whereas that in PP tubes
273
ranged from 55 to 68%, which is contrary to what is widely implicit in the published literature.
274
Amongst the plastics tested, the recoveries from PC (69 to 78%) and PS tubes (65 to 73%) 11
ACCEPTED MANUSCRIPT were greater than from PP tubes. No significant changes were observed over time (1, 2 and
276
7 days) in the case of G1, G2 and PC tubes (Figure 3).
277
In the case of PP and PS tubes, however, some small time effects were observed. Losses at
278
day 7 were similar to losses observed at day 1. In the case of PP1, PP2 and PS tubes, in
279
fact, losses at day 2 were greater (by 8 to 13%) than those at days 1 and 7 (Figure 3).
280
Theoretically, at a fixed analyte concentration, recovery was expected to decrease with time
281
and reach equilibrium at a stage where no more PFOA was being sorbed onto the tube
282
walls. As this trend was not observed in our data, it is possible that PFOA sorption onto the
283
glass and plastic surfaces occurred in a much shorter time frame (within hours) than was
284
investigated in this study. In a way, this idea is supported by our observations during
285
syringe-filtration, where despite using polypropylene syringes, no sorption of PFOA was
286
observed on the syringe-surfaces (Table 1, supplementary material). The PFOA solutions
287
were in contact with the syringes for no more than 5 – 10 seconds, which may have been too
288
short a time for any PFOA analyte to be retained on the surface via sorption. Further studies
289
using shorter time-frames (e.g. < 6hrs) would be required to corroborate this.
SC
M AN U
TE D
290
RI PT
275
3.2.2. Effect of solution chemistry on recovery of PFOA
The influence of solution chemistry on sorption of PFOA in different tubes was determined
292
by measuring PFOA-recovery under varying pH (Figure 4a) and ionic strength (Figure 4b)
293
conditions. Both tests corroborated our previous observations that greater recoveries were
294
obtained from the glass tubes, compared to the plastic tubes (G1 ≈ G2 > PC > PS > PP).
295
Specifically, PFOA-recovery observed from glass (G1 and G2) tubes ranged from 93 to
296
103%, whereas that in PP tubes ranged from 74 to 81%. Recoveries from PC and PS tubes
297
were 85 to 89% and 81 to 86%, respectively.
298
On increasing the pH from 4 to 8, the PFOA-recovery from the glass tubes (G1 and G2)
299
remained largely unaffected. In the PC, PS and PP, a slight increase in recovery, by 4.1%,
AC C
EP
291
12
ACCEPTED MANUSCRIPT 4.7% and 6.5%, respectively, was apparent. Glass tubes, G1 and G2 displayed the least
301
fluctuations and PP the most. Despite some of these pH effects being statistically significant
302
(p < 0.05), on the whole, the sorption losses due to variations in pH (i.e., 1.4 to 6.5%) were
303
minor when compared to the underlying losses due to the inherent nature of the different
304
materials being tested (i.e., 5 to 25%).
305
By using PFOA solutions prepared in background CaCl2 electrolyte solutions of varying
306
concentrations, effects of ionic strength on PFOA recovery were examined. No significant
307
effects of increasing CaCl2 concentrations were observed on recovery of PFOA in PP, PS
308
and PC tubes (p > 0.05). Compared to the control PFOA solution prepared in Milli-Q water,
309
recovery of PFOA from G1 tubes decreased by 5.3% and 6.2%, respectively, in the
310
presence of 25 mM and 100 mM CaCl2 (p = 0.043 and 0.015), but no effects were observed
311
for other ionic strength treatments (1 mM, 10 mM and 50 mM CaCl2). In the case of G2
312
tubes, compared to the control, recovery decreased by 3.5 to 6.4% (p < 0.025) for all ionic
313
strength treatments, except for the 50 mM CaCl2 treatment (p = 0.132).
314
Ionic strength usually affects the electrostatic nature of a surface as well the solubility of the
315
compound, thus controlling interactions occurring at that surface. Glass is known to have a
316
negative surface charge above pH 2.6 (Gu and Li, 2000), whereas a variety of plastics
317
(including PP) have been reported to carry a negative surface charge above pH 3.5 to 4
318
(Lameiras et al., 2008, Leininger et al., 1964). In the experimental conditions used in this
319
study (pH 5.6 ± 0.2), all tubes may be expected to carry a slight negative charge. Divalent
320
cations like Ca2+ have been found to act as a bridge between the negatively charged
321
functional head groups of PFASs and the negatively charged surfaces of a variety of
322
adsorbents (Higgins and Luthy, 2006). It is thus possible that as ionic strength increased, the
323
Ca2+-induced bridging effect caused more PFOA to be retained on the negative surfaces of
324
the glass tubes, thereby slightly decreasing recovery. However, as in the case of solution
325
pH, these variations were relatively minor compared to the considerably greater underlying
326
losses resulting due to the physiochemical nature of the materials themselves.
AC C
EP
TE D
M AN U
SC
RI PT
300
13
ACCEPTED MANUSCRIPT Glass, which is composed of several polar inorganic components like silica (silicon dioxide),
328
boron trioxide and sodium oxide is known to be hydrophilic in nature (DeRosa et al., 2003).
329
On the other hand, plastic polymers like PP and PS are known to be hydrophobic materials
330
as they are both solid hydrocarbons composed entirely of carbon and hydrogen (Kovalchuk
331
et al., 2014). In comparison, PC may be expected to have a lower hydrophobicity than PP
332
and PS (given the presence of polar oxygen functional groups in its polymeric chemical
333
structure), but greater hydrophobicity than glass. While sorption mechanisms need
334
investigation, greater sorption of PFOA onto the plastics compared to glass may be related
335
to potential hydrophobic interactions of PFOA with the hydrophobic components of the
336
polymeric plastics. Further investigation is required at a molecular level to identify the
337
mechanisms controlling the interactions between PFOA and the different materials it may be
338
in contact with.
M AN U
SC
RI PT
327
339
3.2.3. Effect of PFOA concentration on recovery of PFOA
TE D
340
On increasing the concentrations of PFOA in the test solutions, the percentage of PFOA
342
recovered from the solutions within each tube-type increased (Figure 5). For instance, as the
343
spiked concentration of PFOA increased from 12 to 415 µg/L, the recovery of PFOA from
344
PP, PC and G1 tubes improved from 53.7% to 85.5%, 66.7 to 95.1% and 75.3 to 106.3%,
345
respectively. Essentially, the higher the concentration of PFOA in the test solutions, the
346
lower the proportional loss of PFOA onto the container walls, irrespective of the tube-type.
347
This suggests that all plastic and glass tubes tested herein contained a limited number of
348
binding sites on their surface, and can only interact with a finite amount of PFOA. Therefore,
349
as the concentration of PFOA increases, the binding sites on the container walls become
350
increasingly saturated. Similar concentration-dependent results for sorption losses have
351
been reported in the case of other organic chemicals such as pesticides (Sharom and
352
Solomon, 1981) and polycyclic aromatic hydrocarbons (Chlebowski et al., 2016), as well as
353
inorganic substances such as silver nanoparticles (Malysheva et al., 2016).
AC C
EP
341
14
ACCEPTED MANUSCRIPT Poor recovery, particularly at low concentrations, can present serious implications. One
355
current topic of interest in PFAS-research is the determination of the toxicity profiles of these
356
chemicals. When conducting ecotoxicological testing, if significant sorption losses occur, the
357
test organisms will be exposed to reduced concentrations of the analytes, resulting in
358
inaccurate toxicity thresholds (Sekine et al., 2015). Similarly, when testing drinking water
359
quality, and comparing to guideline values to determine safety, erroneous risk assessments
360
may be made due to such losses. The lowest PFOA concentration tested in this study (12
361
µg/L) was around 170 times greater than the current USEPA drinking water health advisory
362
limit for PFOA (0.07 µg/L) (USEPA, 2016). Consequently, greater levels of losses may be
363
expected to occur in trace monitoring and health-related studies, causing substantial
364
inaccuracies.
SC
M AN U
365
RI PT
354
4. Conclusions and Implications
To our knowledge, this is the first study to systematically quantify and report on sorptive
367
losses of dissolved PFOA observed on tubes made from different materials (PP, PC, PS and
368
glass), despite the serious implications of this issue. References to such losses have been
369
made in the published literature, but remain largely unsubstantiated. Moreover, contrary to
370
what is implicit in standard protocols (e.g. USEPA and ISO methods) and the published
371
literature, our data emphasise that greater sorption losses of PFOA occurred on PP
372
containers than on glass containers. Irrespective of solution chemistry (pH and ionic
373
strength) or concentration of PFOA tested, sorption of PFOA onto the tube-walls increased
374
in the order glass < PC < PS < PP. Proportional losses decreased when PFOA
375
concentrations of test solutions increased. Due to this concentration-effect, losses can be
376
especially exaggerated when dealing with PFAS solutions of low concentrations, for
377
instance, when reporting on the quality of drinking water samples. Filtration of aqueous
378
PFOA solutions also introduced a major source of error, leading to an underestimation of
AC C
EP
TE D
366
15
ACCEPTED MANUSCRIPT dissolved concentrations. Our recommendation is therefore to avoid filtration of PFAS
380
solutions where possible, unless it is inevitable.
381
It is possible that even when using tubes or filters made of the same materials, sorption
382
losses may vary widely under different conditions, or with different batches and suppliers –
383
hence it is difficult to ascertain what the ‘best’ material for PFOA or other PFASs may be.
384
However, if adsorptive losses are monitored using appropriate controls and measured with
385
suitable precision, it would be possible to minimise errors associated with such losses.
386
Although it is not possible to directly extrapolate from the current dataset for PFOA to apply
387
to other types of PFASs, it is reasonable to suggest that the specific trends may differ
388
depending on the type of PFAS, due to differences in their chain lengths and functional
389
properties. For instance, compared to PFOA, greater losses may occur in the case of
390
perfluorooctanesulfonic acid (PFOS), which binds more strongly than PFOA to most
391
surfaces. Factors such as the co-occurrence of other surfactants and chemicals, as well as
392
the matrix being tested (e.g. freshwater, sea water, soil, sediments, biological fluids and
393
tissues) can also impact sorption dynamics. Consequently, our study highlights the need to
394
account for losses associated with common laboratory ware for each analyte separately as
395
part of sampling and experimental protocols, and carefully consider the choice of suitable
396
materials for use in PFAS-work.
397
5. Acknowledgements
AC C
EP
TE D
M AN U
SC
RI PT
379
398
Financial support from Australian Research Council Discovery Grant DP150101760 and
399
Ziltek Pty. Ltd is gratefully acknowledged. The authors also thank Gavin Stevenson from
400
National Measurement Institute (NMI, Australia) for technical insights on the analytical
401
procedures used by NMI, including laboratory ware and filters.
402
6. Declarations of Interests: None.
16
ACCEPTED MANUSCRIPT 403 404
7. References Ahmad R, Kookana SR, Alston MA (2001). Syringe filtration as a source of error in pesticide
406
residue analysis in environmental samples. Bulletin of Environmental Contamination
407
and Toxicology, 66, 313-318.
RI PT
405
Ahrens L, Norström K, Viktor T, Cousins AP, Josefsson S (2015). Stockholm Arlanda Airport
409
as a source of per- and polyfluoroalkyl substances to water, sediment and fish.
410
Chemosphere, 129, 33-38.
411
SC
408
Ahrens L, Taniyasu S, Yeung LWY, Yamashita N, Lam PKS, Ebinghaus R (2010). Distribution of polyfluoroalkyl compounds in water, suspended particulate matter and
413
sediment from Tokyo Bay, Japan. Chemosphere, 79, 266-272.
414
M AN U
412
Ahrens L, Yeung LWY, Taniyasu S, Lam PKS, Yamashita N (2011). Partitioning of
415
perfluorooctanoate (PFOA), perfluorooctane sulfonate (PFOS) and perfluorooctane
416
sulfonamide (PFOSA) between water and sediment. Chemosphere, 85, 731-737. Arp HPH, Goss K-U (2008). Irreversible sorption of trace concentrations of
TE D
417 418
perfluorocarboxylic acids to fiber filters used for air sampling. Atmospheric
419
Environment, 42, 6869-6872.
422 423 424 425 426
EP
421
Bin T, McCrosky L, Kulshreshtha AK, Hem SL (2000). Adsorption of esters of phydroxybenzoic acid by filter membranes: Mechanism and effect of formulation and
AC C
420
processing parameters. Pharmaceutical Development and Technology, 5, 95-104.
Carlson M, Thompson RD (2000). Analyte loss due to membrane filter adsorption as determined by high-performance liquid chromatography. Journal of Chromatographic
Science, 38, 77-83. Chandramouli B, Benskin JP, Hamilton MC, Cosgrove JR (2015). Sorption of per- and
427
polyfluoroalkyl substances (PFASs) on filter media: implications for phase partitioning
428
studies. Environmental Toxicology and Chemistry, 34, 30-6.
17
ACCEPTED MANUSCRIPT 429
Chlebowski AC, Tanguay RL, Simonich SLM (2016). Quantitation and prediction of sorptive
430
losses during toxicity testing of polycyclic aromatic hydrocarbon (PAH) and nitrated
431
PAH (NPAH) using polystyrene 96-well plates. Neurotoxicology and Teratology, 57,
432
30-38.
434 435
DeRosa RL, Schader PA, Shelby JE (2003). Hydrophilic nature of silicate glass surfaces as
RI PT
433
a function of exposure condition. Journal of Non-Crystalline Solids, 331, 32-40.
Du Z, Deng S, Bei Y, Huang Q, Wang B, Huang J, Yu G (2014). Adsorption behavior and mechanism of perfluorinated compounds on various adsorbents—A review. Journal
437
of Hazardous Materials, 274, 443-454.
439 440
Gu Y, Li D (2000). The ζ-potential of glass surface in contact with aqueous solutions. Journal of Colloid and Interface Science, 226, 328-339.
M AN U
438
SC
436
Hansen KJ, Johnson HO, Eldridge JS, Butenhoff JL, Dick LA (2002). Quantitative
441
characterization of trace levels of PFOS and PFOA in the Tennessee River.
442
Environmental Science & Technology, 36, 1681-1685.
Harada K, Xu F, Ono K, Iijima T, Koizumi A (2005). Effects of PFOS and PFOA on L-type
TE D
443 444
Ca2+ currents in guinea-pig ventricular myocytes. Biochemical and Biophysical
445
Research Communications, 329, 487-94. Hekster F, Laane RPM, de Voogt P 2003. Environmental and toxicity effects of
EP
446
perfluoroalkylated substances. Reviews of Environmental Contamination and
448
Toxicology. Springer New York.
AC C
447
449
Hellsing MS, Josefsson S, Hughes AV, Ahrens L (2016). Sorption of perfluoroalkyl
450
substances to two types of minerals. Chemosphere, 159, 385-391.
451 452 453
Higgins CP, Luthy RG (2006). Sorption of perfluorinated surfactants on sediments. Environmental Science & Technology, 40, 7251-7256. Higgins CP, McLeod PB, MacManus-Spencer LA, Luthy RG (2007). Bioaccumulation of
454
perfluorochemicals in sediments by the aquatic oligochaete Lumbriculus variegatus.
455
Environmental Science & Technology, 41, 4600-4606.
18
ACCEPTED MANUSCRIPT 456
Hupfeld S, Ausbacher D, Brandl M (2009). Asymmetric flow field-flow fractionation of
457
liposomes: 2. Concentration detection and adsorptive loss phenomena. Journal of
458
Separation Science, 32, 3555-3561.
459
ISO (2009). Water quality - Determination of perfluorooctanesulfonate (PFOS) and perfluorooctanoate (PFOA) - Method for unfiltered samples using solid phase
461
extraction and liquid chromatography/mass spectroscopy. Geneva, Switzerland.
462
RI PT
460
Johansson JH, Berger U, Cousins IT (2017). Can the use of deactivated glass fibre filters eliminate sorption artefacts associated with active air sampling of perfluorooctanoic
464
acid? Environmental Pollution, 224, 779-786.
465
SC
463
Johnson RL, Anschutz AJ, Smolen JM, Simcik MF, Penn RL (2007). The adsorption of perfluorooctane sulfonate onto sand, clay, and iron oxide surfaces. Journal of
467
Chemical & Engineering Data, 52, 1165-1170.
M AN U
466
Kovalchuk NM, Trybala A, Starov V, Matar O, Ivanova N (2014). Fluoro- vs hydrocarbon
469
surfactants: Why do they differ in wetting performance? Advances in Colloid and
470
Interface Science, 210, 65-71.
471
TE D
468
Kwadijk CJAF, Korytár P, Koelmans AA (2010). Distribution of perfluorinated compounds in aquatic systems in the Netherlands. Environmental Science & Technology, 44, 3746-
473
3751.
EP
472
Labadie P, Chevreuil M (2011). Biogeochemical dynamics of perfluorinated alkyl acids and
475
sulfonates in the River Seine (Paris, France) under contrasting hydrological
476 477 478 479
AC C
474
conditions. Environmental Pollution, 159, 3634-3639.
Lameiras FS, Souza ALd, Melo VARd, Nunes EHM, Braga ID (2008). Measurement of the zeta potential of planar surfaces with a rotating disk. Materials Research, 11, 217219.
480
Lein NPH, Fujii S, Tanaka S, Nozoe M, Tanaka H (2008). Contamination of perfluorooctane
481
sulfonate (PFOS) and perfluorooctanoate (PFOA) in surface water of the Yodo River
482
basin (Japan). Desalination, 226, 338-347.
19
ACCEPTED MANUSCRIPT 483
Leininger RI, Mirkovitch V, Beck RE, Andrus PG, Kolff WJ (1964). The zeta potentials of
484
some selected solids in respect to plasma and plasma fractions. ASAIO Journal, 10,
485
237-243.
486
Malysheva A, Ivask A, Hager C, Brunetti G, Marzouk ER, Lombi E, Voelcker NH (2016). Sorption of silver nanoparticles to laboratory plastic during (eco)toxicological testing.
488
Nanotoxicology, 10, 385-390.
489
RI PT
487
Martin JW, Kannan K, Berger U, Voogt PD, Field J, Franklin J, Giesy JP, Harner T, Muir DCG, Scott B, Kaiser M, Järnberg U, Jones KC, Mabury SA, Schroeder H, Simcik M,
491
Sottani C, Bavel BV, Kärrman A, Lindström G, Leeuwen SV (2004). Analytical
492
challenges hamper perfluoroalkyl research. Environmental Science & Technology,
493
38, 248A-255A.
495 496
M AN U
494
SC
490
Ochoa-Herrera V, Sierra-Alvarez R (2008). Removal of perfluorinated surfactants by sorption onto granular activated carbon, zeolite and sludge. Chemosphere, 72, 1588-1593. Prevedouros K, Cousins IT, Buck RC, Korzeniowski SH (2006). Sources, fate and transport of perfluorocarboxylates. Environmental Science & Technology, 40, 32-44.
498
Rattanaoudom R, Visvanathan C, Boontanon SK (2012). Removal of concentrated PFOS
TE D
497
and PFOA in synthetic industrial wastewater by powder activated carbon and
500
hydrotalcite. Journal of Water Sustainability, 2, 245-258.
EP
499
Ross I, McDonough J, Miles J, Storch P, Thelakkat Kochunarayanan P, Kalve E, Hurst J, S.
502
Dasgupta S, Burdick J (2018). A review of emerging technologies for remediation of
503
AC C
501
PFASs. Remediation Journal, 28, 101-126.
504
Schultz MM, Barofsky DF, Field JA (2006). Quantitative determination of fluorinated alkyl
505
substances by large-volume-injection liquid chromatography tandem mass
506
spectrometry - Characterization of municipal wastewaters. Environmental Science &
507
Technology, 40, 289-295.
508
Sekine R, Khurana K, Vasilev K, Lombi E, Donner E (2015). Quantifying the adsorption of
509
ionic silver and functionalized nanoparticles during ecotoxicity testing: Test container
510
effects and recommendations. Nanotoxicology, 9, 1005-1012. 20
ACCEPTED MANUSCRIPT 511
Sharom MS, Solomon KR (1981). Adsorption and desorption of permethrin and other
512
pesticides on glass and plastic materials used in bioassay procedures. Canadian
513
Journal of Fisheries and Aquatic Sciences, 38, 199-204.
514
Shoemaker JA, Grimmett PE, Boutin BK (2009). Determination of selected perfluorinated alkyl acids in drinking water by solid-phase extraction and liquid
516
chromatography/tandem mass spectrometry (LC/MS/MS). Tandem Mass
517
Spectrometry (LC/MS/MS), in, US Environmental Protection Agency, Washington,
518
DC.
Sundström M, Ehresman DJ, Bignert A, Butenhoff JL, Olsen GW, Chang S-C, Bergman A
SC
519
RI PT
515
(2011). A temporal trend study (1972–2008) of perfluorooctanesulfonate,
521
perfluorohexanesulfonate, and perfluorooctanoate in pooled human milk samples
522
from Stockholm, Sweden. Environment International, 37, 178-183.
524 525 526
USEPA (2016). Drinking water health advisory for perfluorooctanoic acid (PFOA). EPA Document Number: 822-R-16-005. Washington, D C.
Voogt Pd, Sáez M (2006). Analytical chemistry of perfluoroalkylated substances. TrAC
TE D
523
M AN U
520
Trends in Analytical Chemistry, 25, 326-342. Washington JW, Yoo H, Ellington JJ, Jenkins TM, Libelo EL (2010). Concentrations,
528
distribution, and persistence of perfluoroalkylates in sludge-applied soils near
529
Decatur, Alabama, USA. Environmental Science & Technology, 44, 8390-8396.
AC C
530
EP
527
21
ACCEPTED MANUSCRIPT Figure 1. Percentage recovery of 14C-PFOA filtrate solution (13.58 ± 0.33 µg/L) from different syringe filter membranes. Error bars represent the standard deviation (n = 4). Different letters denote statistically significant differences (p < 0.05) between groups.
RI PT
Figure 2. Percentage recovery of dissolved PFOA after filtering through three different filters types (PP, RC, GF) when using a range of 14C-PFOA concentrations (0 – 528.90 µg/L). Error
SC
bars represent standard deviation (n = 3).
M AN U
Figure 3. Average concentration of 14C-PFOA remaining in solution after 1, 2 and 7 days when in contact with different centrifuge-tube materials. Error bars represent the standard deviation (n = 3). Concentration of 14C-PFOA used was 21.32 ± 0.07 µg/L. Different letters
TE D
denote significant differences between days on individual centrifuge tube types.
Figure 4. Effect of solution chemistry, pH (a) and ionic strength (b), on the average percentage recovery of PFOA when in contact with different centrifuge tubes. Test
EP
conditions ranged from pH 3 to pH 9 and ionic strength ranged from 0 to 100 mM CaCl2. Concentrations of 14C-PFOA used were 20.79 ± 0.07 µg/L and 19.18 ± 0.10 µg/L, for pH and
AC C
ionic strength tests, respectively. Error bars represent the standard deviation (n = 3). The lines across the data points represent linear regressions.
Figure 5. Average percentage recovery of 14C-PFOA in different centrifuge-tube types when in contact with increasing concentrations of 14C-PFOA (0 – 415.19 µg/L) in solution. Error bars represent standard deviation (n = 3). The lines across the data points represent moving averages.
ACCEPTED MANUSCRIPT
Table 1. Specifications of different filter-types tested for sorption of PFOA. Membrane
Housing
Pore size
Diameter
Source
code
type
material
(µm)
(mm)
(product number)
PP
Polypropylene
Polypropylene
0.45
30
MicroAnalytix (30AP045AN)
GF
Glass fibre
Polyvinyl chloride
20
25
Millipore (SLAP02550)
PVDF
Polyvinylidene fluoride
Polypropylene
0.45
PES
Polyethersulphone
Polypropylene
PES+GFa
PES with GF pre-filter
PTFE
SC
RI PT
Filter
Millipore Millex (SLHV033NK)
0.45
35
MicroAnalytix (MS SF35PS045)
Polypropylene
GF 1, PES 0.45
35
MicroAnalytix (MS SF35GPS045)
Polytetrafluoroethylene
Polypropylene
0.45
25
Sartorius (17576-K)
RC
Regenerated cellulose
Polypropylene
0.45
25
Sartorius (17765-K)
CA
Cellulose acetate
Acrylic resin MBSb
0.45
28
Sartorius (16555-K)
CA+GFa
CA with GF pre-filter
Acrylic resin MBS
GF 0.7µm2, CA 0.45
28
Sartorius (17829-K)
NY
Nylon
Polypropylene
0.45
25
ProSciTech (WS1-04525N)
AC C
EP
TE D
M AN U
33
a
These filter units have two filter membranes in one housing unit, where the GF pre-filter membrane preceded the PES or CA membranes.
b
MBS - methacrylate butadiene styrene
ACCEPTED MANUSCRIPT Table 2. Specifications of different tubes tested for sorption of PFOA. Tube code
Material
Lid material
PP1
Polypropylene
Polyethylene
Capacity
Source
(mL)
(product number)
10
LabServ (LBSCT1202)
Polypropylene
Polyethylene
10
LabServ
RI PT
PP2
(LBSSP1201)
PS1
Polystyrene
Polyethylene
10
Rowe Scientific (S10316UU)
PS2
Polystyrene
Polyethylene
10
ThermoFisher
PC
Polycarbonate
Polypropylene
SC
(LBSCT1002)
10
ThermoFisher
G1
Glass
Screw cap, Teflon liner
Rubber
EP
TE D
Glass
AC C
G2
M AN U
(NAL 3118-0010)
12
10
Kimble Kimax (45066A-16100) BD Vacutainer (366430)
ACCEPTED MANUSCRIPT
Recovery of PFOA (%)
100 80
ab
a
a
ab
60 40
c
d e
b
e f
0
AC C
EP
TE D
M AN U
SC
Filter membrane type
RI PT
20
ACCEPTED MANUSCRIPT 100
PFOA recovery (%)
95 90 85 80
RI PT
PP RC
75
GF
70 100 200 300 400 500 PFOA concentration (μg/L)
600
AC C
EP
TE D
M AN U
SC
0
ACCEPTED MANUSCRIPT
Day 1 Day 2 Day 7
ab a
A
B
b
A
PP1
PP2
60
RI PT
80
40 20 0
G2
M AN U
PS PC G1 Tube material
TE D EP AC C
SC
Recovery of PFOA (%)
100
ACCEPTED MANUSCRIPT
100 90 80 PP1 PS2 PC G1 G2
70 60 50 75 150 225 300 375 Solution (PFOA ng/mL)
450
AC C
EP
TE D
M AN U
SC
0
RI PT
Recovery of PFOA (%)
110
105
105
100
100
PP1
95
95
PS2
90
90
PC
85
85
80
80
75
75
G1 G2
70 3
4
5 6 7 pH (In solution)
8
-10 10
30 50 70 CaCl2 (mM)
(b)
90 110
EP
TE D
M AN U
SC
(a)
RI PT
70
AC C
PFOA recovery (%)
ACCEPTED MANUSCRIPT
ACCEPTED MANUSCRIPT Highlights:
EP
TE D
M AN U
SC
RI PT
Sorption of PFOA onto lab-ware can cause analytical bias during routine procedures. Filtration of PFOA solutions can lead to underestimation of dissolved concentrations. Sorption losses of PFOA onto polypropylene tubes was greater than on glass tubes. Sorption losses decreased when concentrations of PFOA in test solutions increased.
AC C
• • • •