Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes

Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes

Accepted Manuscript Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes Supriya Lath, Emma R. Knight, Divina A...

985KB Sizes 0 Downloads 9 Views

Accepted Manuscript Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes Supriya Lath, Emma R. Knight, Divina A. Navarro, Rai S. Kookana, Michael J. McLaughlin PII:

S0045-6535(19)30103-1

DOI:

https://doi.org/10.1016/j.chemosphere.2019.01.096

Reference:

CHEM 23011

To appear in:

ECSN

Received Date: 14 October 2018 Revised Date:

12 January 2019

Accepted Date: 16 January 2019

Please cite this article as: Lath, S., Knight, E.R., Navarro, D.A., Kookana, R.S., McLaughlin, M.J., Sorption of PFOA onto different laboratory materials: Filter membranes and centrifuge tubes, Chemosphere (2019), doi: https://doi.org/10.1016/j.chemosphere.2019.01.096. This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

ACCEPTED MANUSCRIPT 1

Sorption of PFOA onto different laboratory materials: filter membranes and centrifuge

2

tubes

3 Supriya Lath ([email protected]) *, a, 1

5

Emma R. Knight ([email protected]) a, b

6

Divina A. Navarro ([email protected]) a, b

7

Rai S. Kookana ([email protected]) b

8

Michael J. McLaughlin ([email protected]) a, b

M AN U

SC

RI PT

4

9 10 11

a

School of Agriculture Food and Wine, The University of Adelaide, PMB 1 Glen Osmond, SA 5064, Australia. b

13

17

EP

1

Present address: AECOM, Collins Square, Level 10, Tower Two 727 Collins Street, Melbourne, VIC 3008.

AC C

16

CSIRO Land and Water, PMB 2 Glen Osmond, SA 5064, Australia. * Corresponding author ([email protected])

14 15

TE D

12

1

ACCEPTED MANUSCRIPT Abstract

19

Measurement and reporting of concentrations of contaminants of emerging concern such as

20

per- and polyfluoroalkyl substances (PFASs), including perfluorooctanoic acid (PFOA), is an

21

integral part of most investigations. Sorption losses of PFAS analytes onto particular

22

laboratory-ware (e.g. glass containers) have been suggested in the published literature but

23

have not been investigated in detail. We examined sorption losses from aqueous PFOA

24

solutions in contact with different commonly-used materials in filter units and centrifuge

25

tubes (glass and plastics). Sorption of PFOA onto different filter membrane types ranged

26

from 21 to 79 % indicating that filtration can introduce a major source of error in PFOA

27

analysis; pre-treatment of filter membranes with phosphate or methanol solutions did not

28

improve PFOA recovery. Substantial adsorption of PFOA was also observed on tubes made

29

from polypropylene (PP), polystyrene (PS), polycarbonate (PC), and glass where losses

30

observed were between 32 to 45%, 27 to 35%, 16 to 31% and 14 to 24%, respectively.

31

Contrary to suggestions in the literature, our results indicated the greatest sorption losses for

32

PFOA occurred on PP, whereas losses on glass tubes were much lower. Variations in ionic

33

strength and pH did not greatly influence PFOA recovery. When PFOA concentrations were

34

increased the percent recovery of PFOA increased, indicating that binding sites on tube-

35

walls were saturable. This study draws attention towards analytical bias that can occur due

36

to sorption losses during routine procedures, and highlights the importance of testing the

37

suitability of chosen laboratory-ware for specific PFAS analytes of interest prior to

38

experimental use.

39

Keywords: PFOA; filter membranes; centrifuge tubes; plastics; polypropylene; glass

AC C

EP

TE D

M AN U

SC

RI PT

18

40

2

ACCEPTED MANUSCRIPT 41

1. Introduction

Perfluorooctanoic acid (PFOA) and other related per- and polyfluoroalkyl substances

43

(PFASs) have been recognised as contaminants of emerging concern due to their ubiquitous

44

and persistent nature in the environment, as well as their bioaccumulative properties. As a

45

result, these chemicals are being studied extensively with respect to their human and

46

ecological toxicity (Hekster et al., 2003, Sundström et al., 2011), their occurrence, fate and

47

transport in different environmental compartments (Ahrens et al., 2015, Hansen et al., 2002)

48

as well as management and remediation strategies (Ochoa-Herrera and Sierra-Alvarez,

49

2008, Ross et al., 2018). At various stages of such laboratory and field studies, the PFAS

50

analytes being researched come in contact with a variety of apparatus that are usually made

51

from glass, steel or plastics. The most common apparatus in any study are sampling and

52

storage containers, including a range of tubes, vials and bottles. Others examples include

53

disposable polystyrene (PS) well-plates used as exposure-vessels for toxicity assays, glass

54

aquariums for fish toxicity tests, as well as disposable filtration membranes used for the

55

separation of aqueous phases from solid or particulate matrices prior to analysis.

56

Regardless of the types of experiments conducted, one aspect that is commonly cited in the

57

methods’ section of several published studies relates to the use of sample containers made

58

of particular materials. Specifically, several studies exclude the use of equipment made from

59

glass in experimental or analytical protocols involving PFASs (Ahrens et al., 2015, Hansen et

60

al., 2002). The justification regularly provided for this is that glass adsorbs PFAS analytes.

61

The USEPA and ISO methods, which are the most widely accepted test methods for PFAS

62

analytes, also stipulate that PFAS standards, extracts and samples should not come in

63

contact with any glass containers or pipettes (ISO, 2009, Shoemaker et al., 2009). They

64

recommend that polypropylene (PP) containers be used for all sample, standard and

65

extraction preparation and storage, and suggest that other plastics may be used if they meet

66

quality control requirements (Shoemaker et al., 2009). As a result, PP has been adopted as

AC C

EP

TE D

M AN U

SC

RI PT

42

3

ACCEPTED MANUSCRIPT the material of choice in several studies (Ahrens et al., 2010, Ahrens et al., 2011, Hellsing et

68

al., 2016, Higgins et al., 2007). However, a number of studies have used materials other

69

than PP for PFAS-related studies. For example, Higgins and Luthy (2006) chose PS tubes

70

over PP or glass because their preliminary, unpublished data provided higher recoveries

71

when using PS tubes for a range of PFASs. Other studies have used polyethylene (PE),

72

polycarbonate (PC) and high-density PE containers (Johnson et al., 2007, Washington et al.,

73

2010), often without any indication as to the suitability of these materials.

74

One consensus in the published literature is the avoidance of equipment containing

75

fluoropolymer materials like polytetrafluoroethylene (PTFE) in PFAS studies, as their

76

manufacture has historically involved the use of some PFASs as a ‘polymerisation aid’

77

(Prevedouros et al., 2006). Leaching of remnant PFASs from such products can cause

78

contamination of the dissolved phase (Martin et al., 2004). However, in the case of glass and

79

plastics, there is considerable contradiction and inconsistency in the published literature

80

regarding which materials may be best suited. Losses to laboratory ware could lead to

81

considerable bias in analytical data. Despite the severe implications of such routine losses

82

due to adsorption onto glass and plastics, it has not been investigated in its own right.

83

Apart from choice of containers, filtration is another routine consideration in sample handling

84

and preparation. Filtration is often employed as a major clean-up stage for most

85

environmental and laboratory samples, during which additional losses can occur (Ahmad et

86

al., 2001, Carlson and Thompson, 2000). A study investigating active air-sampling of

87

gaseous PFOA using glass fibre (GF) filters observed that gas-phase PFOA was

88

underestimated due to its sorption onto the GF filters (Johansson et al., 2017). However, an

89

investigation using a variety of PFASs from the aqueous phase, testing four different filter

90

membranes, determined GF filters to be suitable for several PFASs, but recommended

91

specific testing to account for unpredictable effects (Chandramouli et al., 2015).

AC C

EP

TE D

M AN U

SC

RI PT

67

4

ACCEPTED MANUSCRIPT The aim of this study was to examine the sorption losses of PFOA on common glass and

93

plastic materials – specifically, centrifuge tubes and disposable syringe filter membranes –

94

during routine laboratory procedures. Sorption of PFOA from aqueous solutions onto a

95

variety of filter-membranes was tested; the influence of varying concentrations, and two pre-

96

treatments were investigated. Additionally, sorption onto a variety of glass and plastic (PP,

97

PC, PS) tubes was tested. The influence of contact time, pH, ionic strength and PFOA-

98

concentrations were examined to gain insights into the nature of PFOA-binding interactions.

100

2. Materials and Methods 2.1.

SC

99

RI PT

92

Materials

Radiolabelled 14C-PFOA with a specific activity of 2.04 GBq/mmol was purchased from

102

American Radiolabelled Chemicals Incorporation. Optiphase HiSafe 3 scintillation fluid for

103

radiochemical analysis was purchased from PerkinElmer, Australia. All other chemicals used

104

including calcium chloride dihydrate (CaCl2.2H2O), dipotassium hydrogen phosphate

105

(K2HPO4), methanol, hydrochloric acid (HCl) and sodium hydroxide (NaOH) were of

106

analytical grade. The specifications of filter-types and tube-types used in this study are listed

107

in Table 1 and Table 2 respectively.

TE D

2.2.

Filter sorption studies

EP

108

M AN U

101

The sorptive losses through retention of dissolved PFOA onto different filter membrane-

110

types was investigated in triplicate using syringe-filtration through disposable filter units. By

111

means of single-use 10 mL PP syringes, 4 mL of a 14 µg/L PFOA solution was drawn into

112

the syringes, following which filter units were attached to the end of the syringe before

113

plunging the solution through the filter membrane. The initial 2 mL volume of the filtrate was

114

discarded and the subsequent 2 mL was collected, of which 0.5 mL aliquots were used for

115

quantitative analysis. Control samples to account for any sorption occurring on the walls of

116

the syringe were prepared by taking up PFOA solution in the syringe, and dispensing the

117

solutions without attaching any filter units (details in supplementary material). To determine if

AC C

109

5

ACCEPTED MANUSCRIPT recovery of PFOA from the filters could be improved, two pre-rinsing treatments, using

119

phosphate solution (100 mM K2HPO4) and methanol, were applied to the filters. For the

120

phosphate treatment, 10 mL of a 100 mM K2HPO4 solution was plunged through the filter

121

units, followed by 10 mL of Milli-Q water, prior to PFOA filtration. Similarly, for the methanol

122

treatment, 10 mL of methanol was plunged through the filters; membrane-types that were

123

incompatible with methanol were excluded. Finally, the influence of increasing

124

concentrations of PFOA on sorptive losses on three filter membranes (PP, RC and GF) was

125

also tested using an environmentally relevant range of concentrations (0 to 529 µg/L). The

126

critical micelle concentration (CMC) of PFOA has been reported to be in the range of 1 to 1.6

127

x 107 µg/L (i.e., 25 to 38 mM), at which PFOA molecules can agglomerate to form micelles

128

and hemi-micelles; such phase separation can affect interfacial activity like sorption (Harada

129

et al., 2005, Rattanaoudom et al., 2012). The concentrations used in our studies were

130

around 104 to 106 times lower than the CMC, so the impact of micelle-formation on sorption

131

losses is expected to be negligible.

SC

Tube sorption studies

M AN U

2.3.

TE D

132

RI PT

118

To measure sorption of PFOA onto different centrifuge tube-types (PP, PS, PC, G1 and G2,

134

Table 2), tests were performed in triplicate under different conditions by varying the contact

135

times, solution pH, ionic strength as well as PFOA concentrations. When not being varied,

136

the standard test parameters included the use of 8 mL volumes of PFOA solutions of

137

concentrations of 20.5 ± 1 µg/L, prepared in a 10 mM CaCl2 background electrolyte solution

138

or Milli-Q water, and a contact time of 24 to 48 hrs. To investigate the influence of contact

139

time, tubes were subjected to shaking times of 1, 2 and 7 days. The influence of pH was

140

examined across a pH range from 4 to 9 where 0.1M HCl or 0.1M NaOH was used to adjust

141

solution pH. Background electrolyte solutions of varying concentrations (0 to 100 mM CaCl2)

142

were used to determine the effect of ionic strength on retention of PFOA. Sorptive losses as

143

a function of concentration were also examined by testing initial PFOA concentrations

144

ranging from 0 to 415 µg/L. All tubes with test PFOA solutions were placed on an end-over-

AC C

EP

133

6

ACCEPTED MANUSCRIPT 145

end shaker for the duration of the test, after which subsamples (0.5 mL) were taken for

146

quantitative analysis. 2.4.

147

Instrumental analysis

Aliquots (0.5 mL) of all samples, blanks (Milli-Q water) and stock solutions were combined

149

with 4 mL of scintillation fluid in a scintillation vial and analysed radio-chemically using a β-

150

liquid scintillation counter (Perkin Elmer Tri-Carb 3110 RT). Self-normalisation and

151

calibration (SNC) protocol was performed and the 14C counting efficiency was determined to

152

be >90%. The activity of 14C in the samples was measured as disintegrations per minute;

153

each sample was measured three times, for 2 min each. At least 3 background vials

154

(containing 0.5mL of Milli-Q water instead of the sample) were run with each batch of

155

samples to determine the background count, which was subtracted from each sample count.

156

Concentrations of PFOA were then calculated using the specific activity and the measured

157

14

158

to be 0.11 µg/L.

SC

M AN U

C activity. The limit of detection for the method calculated using the blanks was determined

2.5.

TE D

159

RI PT

148

Data and statistical analysis

All experiments were carried out at least in triplicate. The coefficient of variations for the

161

measurement of all stock solutions used in the experiments ranged from 0.35 to 4.83%.

162

Losses of PFOA through adsorption onto the filters and tubes were calculated using the

163

difference between the amounts of PFOA in the samples before and after the experimental

164

steps (i.e. filtration or shaking). Results are reported as a percentage recovery of PFOA

165

compared to the controls. A mass balance was not attempted as each additional handling,

166

extraction and transfer of every sample would increase the chances for loss of analyte,

167

introducing additional error to the method.

168

Statistical software, IBM SPSS (v. 24), was used to determine if there were significant

169

differences (p < 0.05) between treatments. After using analysis of variance test (ANOVA), a

170

Tukey’s honestly significant difference test was used to compare treatment means.

AC C

EP

160

7

ACCEPTED MANUSCRIPT 171

3. Results and Discussion

172

3.1.

173

3.1.1. Sorption losses observed on different filter membrane-types

Filter sorption studies

The sorption of dissolved PFOA onto filter membranes was determined using syringe-

175

filtration. As no sorption of PFOA analyte was identified as occurring on the syringe-surfaces

176

(Table 1, supplementary material) likely due to short residence time (< 10 seconds) in the

177

syringes, the sorption losses were calculated using the difference between the amounts of

178

PFOA in the unfiltered and the filtered solutions. Irrespective of the type of membrane

179

tested, recovery of PFOA was < 100% in all cases (Figure 1). This implies that at least a

180

certain proportion of PFOA was retained onto the filter membrane or housing material due to

181

adsorption, leading to underestimation of dissolved PFOA concentrations. As it was not

182

possible to make a distinction between the amount of PFOA retained on the membrane and

183

the housing material separately, sorption was considered for the filter unit as a whole.

184

Specifically, recovery ranged from 76.0% at best to 21.2% at worst, depending on the type of

185

membrane used (Figure 1). The highest recoveries were achieved when using PVDF

186

(76.0%), glass-fibre (74.2%), RC (74.0%) and PP (72.3%) membranes. The lowest recovery

187

due to adsorption was displayed by the NY membrane filter (21.2%). Overall, the percentage

188

recoveries after filtration were in the order PVDF ≈ GF ≈ RC ≈ PP > CA > PES ≈ PTFE > NY.

189

While PVDF showed a high recovery, it is a fluoropolymer, and like PTFE, may cause

190

contamination of analytical blanks, or over-estimation of dissolved PFAS concentrations

191

(Martin et al., 2004), so it should generally be avoided in PFAS-based analysis. We did not

192

test for PFAS contamination issues during filtration as a radiolabelled PFOA solution was

193

used. Interestingly, despite both PVDF and PTFE being fluoropolymers, PFOA-sorption on

194

PTFE was more than 2 times greater than on PVDF; this may be attributed to the differences

195

in their chemical properties. The carbon-backbone of the polymeric structure in PVDF

196

(chemical formula [–CH2–CF2–]n) is partially fluorinated, whereas that in PTFE (chemical

AC C

EP

TE D

M AN U

SC

RI PT

174

8

ACCEPTED MANUSCRIPT formula [–CF2–CF2–]n) is fully fluorinated, making PTFE more hydrophobic than PVDF. It is

198

likely that the PFOA tail, being hydrophobic, is able to undergo greater hydrophobic

199

interactions with PTFE, leading to greater PFOA sorption.

200

When testing a range of filter media with PFAS-spiked water, Chandramouli et al. (2015)

201

reported that the poorest recoveries were observed with PTFE (2 to 24%), followed by NY

202

(62 to 80%) filters, while GF filters displayed the best recovery (> 85%) overall. For a range

203

of perfluorinated carboxylate and sulphonate compounds, Labadie and Chevreuil (2011)

204

also reported better recoveries using GF (70 to 98%), as opposed to NY (40 to 98%);

205

however, specifically in the case of PFOA, GF and NY were reported to perform equally.

206

Like in the case of GF membrane, the recoveries obtained from RC as well as PP

207

membranes in our study were greater compared to other filters, however accounts of the use

208

of these membranes in the published literature are rare. Chain-length of the PFASs were an

209

important factor controlling sorption losses reported in these studies (Chandramouli et al.,

210

2015, Labadie and Chevreuil, 2011), where greater sorption occurred as chain-length

211

increased. This has been reported in studies related to sorption in environmental media such

212

as soil and sediments, as well as adsorbents such as activated carbon (Du et al., 2014).

213

While the literature focussing on PFAS-investigations in water, soil and sediments commonly

214

use GF filters (Ahrens et al., 2015, Kwadijk et al., 2010, Lein et al., 2008), the literature

215

dealing with air-sampling of PFAS is exploring ways to reduce the sorption of PFAS onto

216

GF-filters (Arp and Goss, 2008, Johansson et al., 2017). This contrast may be attributed to

217

the differences in sampling and filtration equipment set-up for different environmental media.

218

Moreover, it appears that membrane-types, apart from GF and quartz-fibre filters, have not

219

been tested for PFAS sorption.

220

While GF filters have been reported to show the best recoveries for aqueous phases

221

(Chandramouli et al., 2015, Labadie and Chevreuil, 2011), it is important to note that glass

222

fibres used in membranes usually have variable structural integrity. As a result, they are

AC C

EP

TE D

M AN U

SC

RI PT

197

9

ACCEPTED MANUSCRIPT often ascribed with a particle retention rating, covering a range of sizes, and in our case it

224

was 0.8 to 8 µm rather than a specific pore-size. The GF membranes used in this study were

225

stated to be of 20 µm pore size, but the manufacturer cannot guarantee an actual pore size.

226

An interesting observation was that the inclusion of a GF membrane as a pre-filter in the

227

case of PES and CA membranes, caused recovery to improve by up to 18% and 11%,

228

respectively, when compared to the PES and CA membranes where no pre-filter was

229

included. Usually, pre-filters are effective at improving recovery in samples with high

230

particulate load. The solutions used in this study were devoid of any particulate matter, thus

231

it is unclear why an improvement was observed.

232

One approach to reduce sorption of PFOA onto the filters could be through pre-treatment of

233

the filters to pre-saturate or block the available binding sites for PFOA on the membrane – a

234

strategy that has been demonstrated for other organic compounds (Bin et al., 2000, Hupfeld

235

et al., 2009). However, on pre-treating the membranes with either a phosphate solution (100

236

mM), or methanol (99% purity), no significant improvements in PFOA-recovery were

237

observed (Figure S1 and Figure S2; supplementary material). In the case of the methanol

238

treatment, in fact, a decrease in recovery of 7 to 28%, was observed for some membranes

239

(RC, PP and PES; Figure S2). The exact mechanisms controlling the binding of PFOA to the

240

different filter membranes are not known. In the study on active air-sampling of PFOA using

241

GF filters, pre-treatment with a siliconizing reagent led to an appreciable reduction in

242

sorption of gaseous PFOA onto the GF filters (Johansson et al., 2017), particularly when

243

atmospheric concentrations of PFOA were high. While no specific mechanism was identified,

244

it was attributed to deactivation of surface active sites on the GF filter. Given that no benefits

245

in terms of reduction in sorption losses were observed in our study, combined with the added

246

inconvenience, time and expense associated with the process, pre-treatment of disposable

247

filter-units is not recommended for improving recovery from solution. However, pre-rinsing

248

remains a suitable strategy to reduce contamination of the dissolved phase from certain

249

filters (e.g. PTFE) (Labadie and Chevreuil, 2011).

AC C

EP

TE D

M AN U

SC

RI PT

223

10

ACCEPTED MANUSCRIPT 250

3.1.2. Effect of PFOA concentration on PFOA recovery

To test the influence of increasing PFOA concentrations on recovery, three of the filters

252

exhibiting the lowest sorptive losses – PP, RC and GF – were used; PVDF was excluded on

253

account of being a fluoropolymer. In terms of percentage of PFOA recovered (Figure 2),

254

there was no evidence for consistent effect of PFOA concentration on recovery. A similar

255

outcome was reported for PFOA by Chandramouli et al. (2015) when tested on NY filters;

256

PFOA concentrations used ranged from 0.02 to 1 µg/L, which were 2 to 3 orders of

257

magnitude lower than the concentrations tested in our study. We used higher concentrations

258

with the expectation that sorption might be saturable, but this was not observed.

259

As all filter-membranes displayed some losses of PFOA through adsorption, we recommend

260

that it is better to avoid filtration, wherever possible. Other studies have also advised against

261

the use of filtration in sample-preparation steps for the same reasons (Schultz et al., 2006,

262

Voogt and Sáez, 2006). For instance, Schultz et al. (2006) elected centrifugation as their

263

only viable sample clean-up step when analysing PFAS-contaminated municipal wastewater

264

samples due to such losses. If, however, filtration is considered necessary in any

265

procedures, we advise that specific testing of analyte-sorption onto filter media is undertaken

266

to account for potential underestimation of dissolved PFAS concentrations due to such

267

losses.

268

3.2.

269

3.2.1. Effect of contact time on recovery of PFOA

AC C

EP

TE D

M AN U

SC

RI PT

251

Tube sorption studies

270

Recovery of PFOA decreased in the order G1 ≈ G2 > PC > PS > PP (Figure 3), with greater

271

recoveries from glass tubes, compared to the plastic tubes. Specifically, PFOA-recovery

272

observed from glass (G1 and G2) tubes ranged from 77 to 86%, whereas that in PP tubes

273

ranged from 55 to 68%, which is contrary to what is widely implicit in the published literature.

274

Amongst the plastics tested, the recoveries from PC (69 to 78%) and PS tubes (65 to 73%) 11

ACCEPTED MANUSCRIPT were greater than from PP tubes. No significant changes were observed over time (1, 2 and

276

7 days) in the case of G1, G2 and PC tubes (Figure 3).

277

In the case of PP and PS tubes, however, some small time effects were observed. Losses at

278

day 7 were similar to losses observed at day 1. In the case of PP1, PP2 and PS tubes, in

279

fact, losses at day 2 were greater (by 8 to 13%) than those at days 1 and 7 (Figure 3).

280

Theoretically, at a fixed analyte concentration, recovery was expected to decrease with time

281

and reach equilibrium at a stage where no more PFOA was being sorbed onto the tube

282

walls. As this trend was not observed in our data, it is possible that PFOA sorption onto the

283

glass and plastic surfaces occurred in a much shorter time frame (within hours) than was

284

investigated in this study. In a way, this idea is supported by our observations during

285

syringe-filtration, where despite using polypropylene syringes, no sorption of PFOA was

286

observed on the syringe-surfaces (Table 1, supplementary material). The PFOA solutions

287

were in contact with the syringes for no more than 5 – 10 seconds, which may have been too

288

short a time for any PFOA analyte to be retained on the surface via sorption. Further studies

289

using shorter time-frames (e.g. < 6hrs) would be required to corroborate this.

SC

M AN U

TE D

290

RI PT

275

3.2.2. Effect of solution chemistry on recovery of PFOA

The influence of solution chemistry on sorption of PFOA in different tubes was determined

292

by measuring PFOA-recovery under varying pH (Figure 4a) and ionic strength (Figure 4b)

293

conditions. Both tests corroborated our previous observations that greater recoveries were

294

obtained from the glass tubes, compared to the plastic tubes (G1 ≈ G2 > PC > PS > PP).

295

Specifically, PFOA-recovery observed from glass (G1 and G2) tubes ranged from 93 to

296

103%, whereas that in PP tubes ranged from 74 to 81%. Recoveries from PC and PS tubes

297

were 85 to 89% and 81 to 86%, respectively.

298

On increasing the pH from 4 to 8, the PFOA-recovery from the glass tubes (G1 and G2)

299

remained largely unaffected. In the PC, PS and PP, a slight increase in recovery, by 4.1%,

AC C

EP

291

12

ACCEPTED MANUSCRIPT 4.7% and 6.5%, respectively, was apparent. Glass tubes, G1 and G2 displayed the least

301

fluctuations and PP the most. Despite some of these pH effects being statistically significant

302

(p < 0.05), on the whole, the sorption losses due to variations in pH (i.e., 1.4 to 6.5%) were

303

minor when compared to the underlying losses due to the inherent nature of the different

304

materials being tested (i.e., 5 to 25%).

305

By using PFOA solutions prepared in background CaCl2 electrolyte solutions of varying

306

concentrations, effects of ionic strength on PFOA recovery were examined. No significant

307

effects of increasing CaCl2 concentrations were observed on recovery of PFOA in PP, PS

308

and PC tubes (p > 0.05). Compared to the control PFOA solution prepared in Milli-Q water,

309

recovery of PFOA from G1 tubes decreased by 5.3% and 6.2%, respectively, in the

310

presence of 25 mM and 100 mM CaCl2 (p = 0.043 and 0.015), but no effects were observed

311

for other ionic strength treatments (1 mM, 10 mM and 50 mM CaCl2). In the case of G2

312

tubes, compared to the control, recovery decreased by 3.5 to 6.4% (p < 0.025) for all ionic

313

strength treatments, except for the 50 mM CaCl2 treatment (p = 0.132).

314

Ionic strength usually affects the electrostatic nature of a surface as well the solubility of the

315

compound, thus controlling interactions occurring at that surface. Glass is known to have a

316

negative surface charge above pH 2.6 (Gu and Li, 2000), whereas a variety of plastics

317

(including PP) have been reported to carry a negative surface charge above pH 3.5 to 4

318

(Lameiras et al., 2008, Leininger et al., 1964). In the experimental conditions used in this

319

study (pH 5.6 ± 0.2), all tubes may be expected to carry a slight negative charge. Divalent

320

cations like Ca2+ have been found to act as a bridge between the negatively charged

321

functional head groups of PFASs and the negatively charged surfaces of a variety of

322

adsorbents (Higgins and Luthy, 2006). It is thus possible that as ionic strength increased, the

323

Ca2+-induced bridging effect caused more PFOA to be retained on the negative surfaces of

324

the glass tubes, thereby slightly decreasing recovery. However, as in the case of solution

325

pH, these variations were relatively minor compared to the considerably greater underlying

326

losses resulting due to the physiochemical nature of the materials themselves.

AC C

EP

TE D

M AN U

SC

RI PT

300

13

ACCEPTED MANUSCRIPT Glass, which is composed of several polar inorganic components like silica (silicon dioxide),

328

boron trioxide and sodium oxide is known to be hydrophilic in nature (DeRosa et al., 2003).

329

On the other hand, plastic polymers like PP and PS are known to be hydrophobic materials

330

as they are both solid hydrocarbons composed entirely of carbon and hydrogen (Kovalchuk

331

et al., 2014). In comparison, PC may be expected to have a lower hydrophobicity than PP

332

and PS (given the presence of polar oxygen functional groups in its polymeric chemical

333

structure), but greater hydrophobicity than glass. While sorption mechanisms need

334

investigation, greater sorption of PFOA onto the plastics compared to glass may be related

335

to potential hydrophobic interactions of PFOA with the hydrophobic components of the

336

polymeric plastics. Further investigation is required at a molecular level to identify the

337

mechanisms controlling the interactions between PFOA and the different materials it may be

338

in contact with.

M AN U

SC

RI PT

327

339

3.2.3. Effect of PFOA concentration on recovery of PFOA

TE D

340

On increasing the concentrations of PFOA in the test solutions, the percentage of PFOA

342

recovered from the solutions within each tube-type increased (Figure 5). For instance, as the

343

spiked concentration of PFOA increased from 12 to 415 µg/L, the recovery of PFOA from

344

PP, PC and G1 tubes improved from 53.7% to 85.5%, 66.7 to 95.1% and 75.3 to 106.3%,

345

respectively. Essentially, the higher the concentration of PFOA in the test solutions, the

346

lower the proportional loss of PFOA onto the container walls, irrespective of the tube-type.

347

This suggests that all plastic and glass tubes tested herein contained a limited number of

348

binding sites on their surface, and can only interact with a finite amount of PFOA. Therefore,

349

as the concentration of PFOA increases, the binding sites on the container walls become

350

increasingly saturated. Similar concentration-dependent results for sorption losses have

351

been reported in the case of other organic chemicals such as pesticides (Sharom and

352

Solomon, 1981) and polycyclic aromatic hydrocarbons (Chlebowski et al., 2016), as well as

353

inorganic substances such as silver nanoparticles (Malysheva et al., 2016).

AC C

EP

341

14

ACCEPTED MANUSCRIPT Poor recovery, particularly at low concentrations, can present serious implications. One

355

current topic of interest in PFAS-research is the determination of the toxicity profiles of these

356

chemicals. When conducting ecotoxicological testing, if significant sorption losses occur, the

357

test organisms will be exposed to reduced concentrations of the analytes, resulting in

358

inaccurate toxicity thresholds (Sekine et al., 2015). Similarly, when testing drinking water

359

quality, and comparing to guideline values to determine safety, erroneous risk assessments

360

may be made due to such losses. The lowest PFOA concentration tested in this study (12

361

µg/L) was around 170 times greater than the current USEPA drinking water health advisory

362

limit for PFOA (0.07 µg/L) (USEPA, 2016). Consequently, greater levels of losses may be

363

expected to occur in trace monitoring and health-related studies, causing substantial

364

inaccuracies.

SC

M AN U

365

RI PT

354

4. Conclusions and Implications

To our knowledge, this is the first study to systematically quantify and report on sorptive

367

losses of dissolved PFOA observed on tubes made from different materials (PP, PC, PS and

368

glass), despite the serious implications of this issue. References to such losses have been

369

made in the published literature, but remain largely unsubstantiated. Moreover, contrary to

370

what is implicit in standard protocols (e.g. USEPA and ISO methods) and the published

371

literature, our data emphasise that greater sorption losses of PFOA occurred on PP

372

containers than on glass containers. Irrespective of solution chemistry (pH and ionic

373

strength) or concentration of PFOA tested, sorption of PFOA onto the tube-walls increased

374

in the order glass < PC < PS < PP. Proportional losses decreased when PFOA

375

concentrations of test solutions increased. Due to this concentration-effect, losses can be

376

especially exaggerated when dealing with PFAS solutions of low concentrations, for

377

instance, when reporting on the quality of drinking water samples. Filtration of aqueous

378

PFOA solutions also introduced a major source of error, leading to an underestimation of

AC C

EP

TE D

366

15

ACCEPTED MANUSCRIPT dissolved concentrations. Our recommendation is therefore to avoid filtration of PFAS

380

solutions where possible, unless it is inevitable.

381

It is possible that even when using tubes or filters made of the same materials, sorption

382

losses may vary widely under different conditions, or with different batches and suppliers –

383

hence it is difficult to ascertain what the ‘best’ material for PFOA or other PFASs may be.

384

However, if adsorptive losses are monitored using appropriate controls and measured with

385

suitable precision, it would be possible to minimise errors associated with such losses.

386

Although it is not possible to directly extrapolate from the current dataset for PFOA to apply

387

to other types of PFASs, it is reasonable to suggest that the specific trends may differ

388

depending on the type of PFAS, due to differences in their chain lengths and functional

389

properties. For instance, compared to PFOA, greater losses may occur in the case of

390

perfluorooctanesulfonic acid (PFOS), which binds more strongly than PFOA to most

391

surfaces. Factors such as the co-occurrence of other surfactants and chemicals, as well as

392

the matrix being tested (e.g. freshwater, sea water, soil, sediments, biological fluids and

393

tissues) can also impact sorption dynamics. Consequently, our study highlights the need to

394

account for losses associated with common laboratory ware for each analyte separately as

395

part of sampling and experimental protocols, and carefully consider the choice of suitable

396

materials for use in PFAS-work.

397

5. Acknowledgements

AC C

EP

TE D

M AN U

SC

RI PT

379

398

Financial support from Australian Research Council Discovery Grant DP150101760 and

399

Ziltek Pty. Ltd is gratefully acknowledged. The authors also thank Gavin Stevenson from

400

National Measurement Institute (NMI, Australia) for technical insights on the analytical

401

procedures used by NMI, including laboratory ware and filters.

402

6. Declarations of Interests: None.

16

ACCEPTED MANUSCRIPT 403 404

7. References Ahmad R, Kookana SR, Alston MA (2001). Syringe filtration as a source of error in pesticide

406

residue analysis in environmental samples. Bulletin of Environmental Contamination

407

and Toxicology, 66, 313-318.

RI PT

405

Ahrens L, Norström K, Viktor T, Cousins AP, Josefsson S (2015). Stockholm Arlanda Airport

409

as a source of per- and polyfluoroalkyl substances to water, sediment and fish.

410

Chemosphere, 129, 33-38.

411

SC

408

Ahrens L, Taniyasu S, Yeung LWY, Yamashita N, Lam PKS, Ebinghaus R (2010). Distribution of polyfluoroalkyl compounds in water, suspended particulate matter and

413

sediment from Tokyo Bay, Japan. Chemosphere, 79, 266-272.

414

M AN U

412

Ahrens L, Yeung LWY, Taniyasu S, Lam PKS, Yamashita N (2011). Partitioning of

415

perfluorooctanoate (PFOA), perfluorooctane sulfonate (PFOS) and perfluorooctane

416

sulfonamide (PFOSA) between water and sediment. Chemosphere, 85, 731-737. Arp HPH, Goss K-U (2008). Irreversible sorption of trace concentrations of

TE D

417 418

perfluorocarboxylic acids to fiber filters used for air sampling. Atmospheric

419

Environment, 42, 6869-6872.

422 423 424 425 426

EP

421

Bin T, McCrosky L, Kulshreshtha AK, Hem SL (2000). Adsorption of esters of phydroxybenzoic acid by filter membranes: Mechanism and effect of formulation and

AC C

420

processing parameters. Pharmaceutical Development and Technology, 5, 95-104.

Carlson M, Thompson RD (2000). Analyte loss due to membrane filter adsorption as determined by high-performance liquid chromatography. Journal of Chromatographic

Science, 38, 77-83. Chandramouli B, Benskin JP, Hamilton MC, Cosgrove JR (2015). Sorption of per- and

427

polyfluoroalkyl substances (PFASs) on filter media: implications for phase partitioning

428

studies. Environmental Toxicology and Chemistry, 34, 30-6.

17

ACCEPTED MANUSCRIPT 429

Chlebowski AC, Tanguay RL, Simonich SLM (2016). Quantitation and prediction of sorptive

430

losses during toxicity testing of polycyclic aromatic hydrocarbon (PAH) and nitrated

431

PAH (NPAH) using polystyrene 96-well plates. Neurotoxicology and Teratology, 57,

432

30-38.

434 435

DeRosa RL, Schader PA, Shelby JE (2003). Hydrophilic nature of silicate glass surfaces as

RI PT

433

a function of exposure condition. Journal of Non-Crystalline Solids, 331, 32-40.

Du Z, Deng S, Bei Y, Huang Q, Wang B, Huang J, Yu G (2014). Adsorption behavior and mechanism of perfluorinated compounds on various adsorbents—A review. Journal

437

of Hazardous Materials, 274, 443-454.

439 440

Gu Y, Li D (2000). The ζ-potential of glass surface in contact with aqueous solutions. Journal of Colloid and Interface Science, 226, 328-339.

M AN U

438

SC

436

Hansen KJ, Johnson HO, Eldridge JS, Butenhoff JL, Dick LA (2002). Quantitative

441

characterization of trace levels of PFOS and PFOA in the Tennessee River.

442

Environmental Science & Technology, 36, 1681-1685.

Harada K, Xu F, Ono K, Iijima T, Koizumi A (2005). Effects of PFOS and PFOA on L-type

TE D

443 444

Ca2+ currents in guinea-pig ventricular myocytes. Biochemical and Biophysical

445

Research Communications, 329, 487-94. Hekster F, Laane RPM, de Voogt P 2003. Environmental and toxicity effects of

EP

446

perfluoroalkylated substances. Reviews of Environmental Contamination and

448

Toxicology. Springer New York.

AC C

447

449

Hellsing MS, Josefsson S, Hughes AV, Ahrens L (2016). Sorption of perfluoroalkyl

450

substances to two types of minerals. Chemosphere, 159, 385-391.

451 452 453

Higgins CP, Luthy RG (2006). Sorption of perfluorinated surfactants on sediments. Environmental Science & Technology, 40, 7251-7256. Higgins CP, McLeod PB, MacManus-Spencer LA, Luthy RG (2007). Bioaccumulation of

454

perfluorochemicals in sediments by the aquatic oligochaete Lumbriculus variegatus.

455

Environmental Science & Technology, 41, 4600-4606.

18

ACCEPTED MANUSCRIPT 456

Hupfeld S, Ausbacher D, Brandl M (2009). Asymmetric flow field-flow fractionation of

457

liposomes: 2. Concentration detection and adsorptive loss phenomena. Journal of

458

Separation Science, 32, 3555-3561.

459

ISO (2009). Water quality - Determination of perfluorooctanesulfonate (PFOS) and perfluorooctanoate (PFOA) - Method for unfiltered samples using solid phase

461

extraction and liquid chromatography/mass spectroscopy. Geneva, Switzerland.

462

RI PT

460

Johansson JH, Berger U, Cousins IT (2017). Can the use of deactivated glass fibre filters eliminate sorption artefacts associated with active air sampling of perfluorooctanoic

464

acid? Environmental Pollution, 224, 779-786.

465

SC

463

Johnson RL, Anschutz AJ, Smolen JM, Simcik MF, Penn RL (2007). The adsorption of perfluorooctane sulfonate onto sand, clay, and iron oxide surfaces. Journal of

467

Chemical & Engineering Data, 52, 1165-1170.

M AN U

466

Kovalchuk NM, Trybala A, Starov V, Matar O, Ivanova N (2014). Fluoro- vs hydrocarbon

469

surfactants: Why do they differ in wetting performance? Advances in Colloid and

470

Interface Science, 210, 65-71.

471

TE D

468

Kwadijk CJAF, Korytár P, Koelmans AA (2010). Distribution of perfluorinated compounds in aquatic systems in the Netherlands. Environmental Science & Technology, 44, 3746-

473

3751.

EP

472

Labadie P, Chevreuil M (2011). Biogeochemical dynamics of perfluorinated alkyl acids and

475

sulfonates in the River Seine (Paris, France) under contrasting hydrological

476 477 478 479

AC C

474

conditions. Environmental Pollution, 159, 3634-3639.

Lameiras FS, Souza ALd, Melo VARd, Nunes EHM, Braga ID (2008). Measurement of the zeta potential of planar surfaces with a rotating disk. Materials Research, 11, 217219.

480

Lein NPH, Fujii S, Tanaka S, Nozoe M, Tanaka H (2008). Contamination of perfluorooctane

481

sulfonate (PFOS) and perfluorooctanoate (PFOA) in surface water of the Yodo River

482

basin (Japan). Desalination, 226, 338-347.

19

ACCEPTED MANUSCRIPT 483

Leininger RI, Mirkovitch V, Beck RE, Andrus PG, Kolff WJ (1964). The zeta potentials of

484

some selected solids in respect to plasma and plasma fractions. ASAIO Journal, 10,

485

237-243.

486

Malysheva A, Ivask A, Hager C, Brunetti G, Marzouk ER, Lombi E, Voelcker NH (2016). Sorption of silver nanoparticles to laboratory plastic during (eco)toxicological testing.

488

Nanotoxicology, 10, 385-390.

489

RI PT

487

Martin JW, Kannan K, Berger U, Voogt PD, Field J, Franklin J, Giesy JP, Harner T, Muir DCG, Scott B, Kaiser M, Järnberg U, Jones KC, Mabury SA, Schroeder H, Simcik M,

491

Sottani C, Bavel BV, Kärrman A, Lindström G, Leeuwen SV (2004). Analytical

492

challenges hamper perfluoroalkyl research. Environmental Science & Technology,

493

38, 248A-255A.

495 496

M AN U

494

SC

490

Ochoa-Herrera V, Sierra-Alvarez R (2008). Removal of perfluorinated surfactants by sorption onto granular activated carbon, zeolite and sludge. Chemosphere, 72, 1588-1593. Prevedouros K, Cousins IT, Buck RC, Korzeniowski SH (2006). Sources, fate and transport of perfluorocarboxylates. Environmental Science & Technology, 40, 32-44.

498

Rattanaoudom R, Visvanathan C, Boontanon SK (2012). Removal of concentrated PFOS

TE D

497

and PFOA in synthetic industrial wastewater by powder activated carbon and

500

hydrotalcite. Journal of Water Sustainability, 2, 245-258.

EP

499

Ross I, McDonough J, Miles J, Storch P, Thelakkat Kochunarayanan P, Kalve E, Hurst J, S.

502

Dasgupta S, Burdick J (2018). A review of emerging technologies for remediation of

503

AC C

501

PFASs. Remediation Journal, 28, 101-126.

504

Schultz MM, Barofsky DF, Field JA (2006). Quantitative determination of fluorinated alkyl

505

substances by large-volume-injection liquid chromatography tandem mass

506

spectrometry - Characterization of municipal wastewaters. Environmental Science &

507

Technology, 40, 289-295.

508

Sekine R, Khurana K, Vasilev K, Lombi E, Donner E (2015). Quantifying the adsorption of

509

ionic silver and functionalized nanoparticles during ecotoxicity testing: Test container

510

effects and recommendations. Nanotoxicology, 9, 1005-1012. 20

ACCEPTED MANUSCRIPT 511

Sharom MS, Solomon KR (1981). Adsorption and desorption of permethrin and other

512

pesticides on glass and plastic materials used in bioassay procedures. Canadian

513

Journal of Fisheries and Aquatic Sciences, 38, 199-204.

514

Shoemaker JA, Grimmett PE, Boutin BK (2009). Determination of selected perfluorinated alkyl acids in drinking water by solid-phase extraction and liquid

516

chromatography/tandem mass spectrometry (LC/MS/MS). Tandem Mass

517

Spectrometry (LC/MS/MS), in, US Environmental Protection Agency, Washington,

518

DC.

Sundström M, Ehresman DJ, Bignert A, Butenhoff JL, Olsen GW, Chang S-C, Bergman A

SC

519

RI PT

515

(2011). A temporal trend study (1972–2008) of perfluorooctanesulfonate,

521

perfluorohexanesulfonate, and perfluorooctanoate in pooled human milk samples

522

from Stockholm, Sweden. Environment International, 37, 178-183.

524 525 526

USEPA (2016). Drinking water health advisory for perfluorooctanoic acid (PFOA). EPA Document Number: 822-R-16-005. Washington, D C.

Voogt Pd, Sáez M (2006). Analytical chemistry of perfluoroalkylated substances. TrAC

TE D

523

M AN U

520

Trends in Analytical Chemistry, 25, 326-342. Washington JW, Yoo H, Ellington JJ, Jenkins TM, Libelo EL (2010). Concentrations,

528

distribution, and persistence of perfluoroalkylates in sludge-applied soils near

529

Decatur, Alabama, USA. Environmental Science & Technology, 44, 8390-8396.

AC C

530

EP

527

21

ACCEPTED MANUSCRIPT Figure 1. Percentage recovery of 14C-PFOA filtrate solution (13.58 ± 0.33 µg/L) from different syringe filter membranes. Error bars represent the standard deviation (n = 4). Different letters denote statistically significant differences (p < 0.05) between groups.

RI PT

Figure 2. Percentage recovery of dissolved PFOA after filtering through three different filters types (PP, RC, GF) when using a range of 14C-PFOA concentrations (0 – 528.90 µg/L). Error

SC

bars represent standard deviation (n = 3).

M AN U

Figure 3. Average concentration of 14C-PFOA remaining in solution after 1, 2 and 7 days when in contact with different centrifuge-tube materials. Error bars represent the standard deviation (n = 3). Concentration of 14C-PFOA used was 21.32 ± 0.07 µg/L. Different letters

TE D

denote significant differences between days on individual centrifuge tube types.

Figure 4. Effect of solution chemistry, pH (a) and ionic strength (b), on the average percentage recovery of PFOA when in contact with different centrifuge tubes. Test

EP

conditions ranged from pH 3 to pH 9 and ionic strength ranged from 0 to 100 mM CaCl2. Concentrations of 14C-PFOA used were 20.79 ± 0.07 µg/L and 19.18 ± 0.10 µg/L, for pH and

AC C

ionic strength tests, respectively. Error bars represent the standard deviation (n = 3). The lines across the data points represent linear regressions.

Figure 5. Average percentage recovery of 14C-PFOA in different centrifuge-tube types when in contact with increasing concentrations of 14C-PFOA (0 – 415.19 µg/L) in solution. Error bars represent standard deviation (n = 3). The lines across the data points represent moving averages.

ACCEPTED MANUSCRIPT

Table 1. Specifications of different filter-types tested for sorption of PFOA. Membrane

Housing

Pore size

Diameter

Source

code

type

material

(µm)

(mm)

(product number)

PP

Polypropylene

Polypropylene

0.45

30

MicroAnalytix (30AP045AN)

GF

Glass fibre

Polyvinyl chloride

20

25

Millipore (SLAP02550)

PVDF

Polyvinylidene fluoride

Polypropylene

0.45

PES

Polyethersulphone

Polypropylene

PES+GFa

PES with GF pre-filter

PTFE

SC

RI PT

Filter

Millipore Millex (SLHV033NK)

0.45

35

MicroAnalytix (MS SF35PS045)

Polypropylene

GF 1, PES 0.45

35

MicroAnalytix (MS SF35GPS045)

Polytetrafluoroethylene

Polypropylene

0.45

25

Sartorius (17576-K)

RC

Regenerated cellulose

Polypropylene

0.45

25

Sartorius (17765-K)

CA

Cellulose acetate

Acrylic resin MBSb

0.45

28

Sartorius (16555-K)

CA+GFa

CA with GF pre-filter

Acrylic resin MBS

GF 0.7µm2, CA 0.45

28

Sartorius (17829-K)

NY

Nylon

Polypropylene

0.45

25

ProSciTech (WS1-04525N)

AC C

EP

TE D

M AN U

33

a

These filter units have two filter membranes in one housing unit, where the GF pre-filter membrane preceded the PES or CA membranes.

b

MBS - methacrylate butadiene styrene

ACCEPTED MANUSCRIPT Table 2. Specifications of different tubes tested for sorption of PFOA. Tube code

Material

Lid material

PP1

Polypropylene

Polyethylene

Capacity

Source

(mL)

(product number)

10

LabServ (LBSCT1202)

Polypropylene

Polyethylene

10

LabServ

RI PT

PP2

(LBSSP1201)

PS1

Polystyrene

Polyethylene

10

Rowe Scientific (S10316UU)

PS2

Polystyrene

Polyethylene

10

ThermoFisher

PC

Polycarbonate

Polypropylene

SC

(LBSCT1002)

10

ThermoFisher

G1

Glass

Screw cap, Teflon liner

Rubber

EP

TE D

Glass

AC C

G2

M AN U

(NAL 3118-0010)

12

10

Kimble Kimax (45066A-16100) BD Vacutainer (366430)

ACCEPTED MANUSCRIPT

Recovery of PFOA (%)

100 80

ab

a

a

ab

60 40

c

d e

b

e f

0

AC C

EP

TE D

M AN U

SC

Filter membrane type

RI PT

20

ACCEPTED MANUSCRIPT 100

PFOA recovery (%)

95 90 85 80

RI PT

PP RC

75

GF

70 100 200 300 400 500 PFOA concentration (μg/L)

600

AC C

EP

TE D

M AN U

SC

0

ACCEPTED MANUSCRIPT

Day 1 Day 2 Day 7

ab a

A

B

b

A

PP1

PP2

60

RI PT

80

40 20 0

G2

M AN U

PS PC G1 Tube material

TE D EP AC C

SC

Recovery of PFOA (%)

100

ACCEPTED MANUSCRIPT

100 90 80 PP1 PS2 PC G1 G2

70 60 50 75 150 225 300 375 Solution (PFOA ng/mL)

450

AC C

EP

TE D

M AN U

SC

0

RI PT

Recovery of PFOA (%)

110

105

105

100

100

PP1

95

95

PS2

90

90

PC

85

85

80

80

75

75

G1 G2

70 3

4

5 6 7 pH (In solution)

8

-10 10

30 50 70 CaCl2 (mM)

(b)

90 110

EP

TE D

M AN U

SC

(a)

RI PT

70

AC C

PFOA recovery (%)

ACCEPTED MANUSCRIPT

ACCEPTED MANUSCRIPT Highlights:

EP

TE D

M AN U

SC

RI PT

Sorption of PFOA onto lab-ware can cause analytical bias during routine procedures. Filtration of PFOA solutions can lead to underestimation of dissolved concentrations. Sorption losses of PFOA onto polypropylene tubes was greater than on glass tubes. Sorption losses decreased when concentrations of PFOA in test solutions increased.