Talonto, Vol.24,pp.611-623. Pergamon Press, 1977. Printed in Great Britain.
THE SPECIFICITY CHARACTERISTIC OF ANALYTICAL METHODS-II ESTIMATION AND EXPER~ENTAL VERIFICATION OF THE NON-SPECIFICITY COEFFICIENTS LEON PSZONICKI and ANNA EUKSZO-BIE&KOWSKA
Institute of Nuclear Research, Department of Analytical Chemistry, Warsaw (Received 13 September 1976. Revised 1 February 1977. Accepted 22 April 1977) Su~ry-A procedure for estimation of non-s~~city coefficients by use of small amounts of standard samples is described. This procedure enables also the simultaneous verification if the specificity of a method with respect to the investigated interferent may be characterized by means of these coefficients. Four various analytical methods were tested and a good agreement of experimentally found values of interferent effects with the values calculated by means of non-specificity coefficients have been stated.
NOTATION USED
ESTIMATION OF NON-SPECIFICITY COEFFICIENTS
a
-intercept of a straight line graph on the ordinate b slope of the straight line --concentration of interferent Cb --concentration of analyte Cd --relative concentration of interferent, relae, tion of cs to c, KP and KA -non-specificity coefficients R --total analytical signal itnalytical signal of analyte Rd -“parallel shift” component of interfering Rb signal -standard deviation for results indicated by sx subscript mean “lack of fit” square S,. mean “within group” square SW -number of standards in series Y t -number of parallel determinations of a standard sample
The non-specificity coefficients must be calculated from the experimental data obtained by analysis of standard samples. The standards must be prepared separately for every interferent tested The most precise way to estimate the coefficients is to prepare a set of calibration curves with varied interferent concentration,
followed
values
every
for
were defined
two
kinds
such
However,
because
if every curve
method
is tested
interferent, samples
cussed
Here
coefficient
of their
of anaiyt&l
we present
of these coefhcients
methods
a procedure
with solution
was dis-
standards
for esti~~on
very strongly
from the analyte signal;
polarogra-
trations
and would usually chosen
difficult
and thus provide
ideas. In addition,
affect the an~~cal
before measurement
effects are particularly tatively,
to
prepare
be separated
the analyte
used for preparation
to
interferent
improve
to characterize
quanti-
the results obtained
requires
the
for which
to make q = 5
and the total
are easily pres-
617
of
simple, and it may not be used
choice
of
the
samples
procedure,
interferent allows
however.
concen-
simplification
According
I,’ the analytical
to the signal
of
analytical
signal in the presence
(1) of one
is R = kdcd -I- k,c,
ented graphically.
even
the preparation
Rd = k,,c,,
interferent
at
confidence
is
such large
a good test of the proposed
of
of interferer-it.
be lower than 3, testing
one
(4) and (5) Part
of the analytical because
standard
is very time-consuming
in the standard
equations
q2
practice.
of the testing investigated
they were specially
of
and
is relatively
Appropriate
phy and coulometry. results
concentrations
for even
samples,
in analytical
verification
as spectrophotometry,
All the ~terferents
and
q
curve in absence
or 6. Such a procedure
application
and its experimental
for such methods
on q standards
against
the value of q cannot
9 standards,
of
is based
plus the standards calibration
the
range
is un~no~c,
limits of the results it is recommended
and the possibility
to the characterization
a method
it is necessary
the normal
least
of non-specificity
whether
interest.
of the method
I,’
of the coefficient shows
over the concentration
Because
In Part
which
values are constant
the
~TRODU~ON
by calculation
curve,
+ kacdG.
(2)
618
LEONPSZONICKIand ANNALUKSZO-BIE&KOWSKA
The ratio of R to Rd is k + 2% kd
k A, k,
R,=l+
(3)
where c, is the relative concentration of interferent (c, = %/Cd). Taking into account the definitions of K, and KA we obtain R, = 1 + Kpc, + KAc,,.
(4)
When a series of standards is prepared so that c, is constant, then R, is a function of cb only. This function R&J, described by equation (4), enables us to calculate the K-values and to test their constancy by means of 4 standards. When R&J is linear, then K, and KA are constant hnd they are described by the parameters of its regression line: a-l KP = -
(5)
c, KA = b
(6)
The standard deviations and confidence limits of the
K-coefficients may be calculated from the standard deviations of a and b: %P = s&,
and
sKA= s,.
(7)
In the calculation of the regression parameters and verification of the linearity it is important that the variances of all measurements should be homogeneous. In chemical analysis the variance of the measurements usually increases with increasing concentration and this requirement is not fulfilled. It is therefore necessary to use weighting factors in all the statistical calculations and hence the dispersion of the measurements at various concentrations ought to be estimated. The variances of the R,-values can be calculated from the variances of corresponding values of R,, and R, taking into account the law of propagation of errors. The variations in parallel determinations of Rd and R are not correlated, so the following simplified formula may be used:
Calculation scheme 1
1. Calculate the values of R,, . . . Rrq as the ratios of values of R (Table 1, column 8) to corresponding values of Rd (column 3). 2. Calculate the variances si, from the corresponding variances si and s&. 3. Calculate the parameters a and b of regression line RJc,,) by the least-squares method, using the weighting factors, and then calculate the variances s,’ and s:.
R2
2 2 AL++2 SR, Rd’
R;
The general schedule of measurements shown in Table 1 is sufficient for complete calibration of an analytical method. The variances of Rd and R as well as of Rb must be estimated experimentally on the basis of t parallel determinations on standard samples. The minimum value of t may be taken as 3, but if it does not lead to too much extra work this number ought to be higher, e.g., 5 or 6. In practice, an interferent rarely affects the dispersion of results significantly and the magnitude of the variance is dependent only on the magnitude of the analytical signal. Because small errors in estimation of si or s& values do not have a significant influence on the values of the calculated coefficients the variances of R and Rb may be taken as equal to the variance of Rd obtained for a standard that emits a corresponding magnitude of signal. The variance of R need be determined experimentally only for the range of concentration where R significantly exceeds Rd, but such cases happen rather rarely in practical analysis. When t parallel determinations are done, then the values given in Table 1 represent the mean values and their variances. When the variances for R or Rb are taken as the same as for Rd, then the values of R and Rb can be based on single determinations and the variances of single determinations must also then be used. For all interferents for which Rr(c,,) is linear, the values of the measurements tabulated in columns 6 and 7, 11 and 12 etc. are redundant and can be omitted. The calculations are carried out according to the following scheme.
Rd
Table 1. Schematic table of measurements Analyte Standard No.
$, 3
4
Interferent A c,, = R Cb Rb (Cb= C,,.Cd) (Cd= 0) %b (c, = const.) 5
6
7
8
Interferent B cr, = $
Cb
Rb
s&
R
s;
9
10
11
12
13
14
1
2
1 2
cd, Cd*
cO,‘cd,
C,,‘Cdl
‘k’cdz
cr,‘cdz
4
‘d,
cr.., ’ cd.
&a’
etc.
cd,
Note. The values of cd,, cdl . . cd,,and c, are dependent only on the concentration ranges of interest. The data given in columns 6 and 7, 11 and 12 etc. are necessary only when R, = Rr(cb) is not linear.
Specificity characteristics of analytical methods-11 4. Using the weighting factors, calculate the “within-group” sum of squares Of results [i.e., C~i - yi)‘] and the “lack-of-fit” sum of squares, which is the difference between the residual sum of squares about the regression and the “wi~n-group” sum of squares. 5. Calculate the mean “lack-of-fit” square (St) and mean “within-group” square (S,) and apply the F-test to verify the hypothesis of linearity of R,(c,,):~ F = S,_/S,.
6. If F > Ftab the function R&J is linear. Calculate the values of the K, and K:, coefficients and their standard deviations. An example of calculation according to scheme 1 of K-values for spectrophotometric determination of uranium with respect to thorium as interferent is given in Appendix I. When an analytical method does not fulfll all the simplifying assumptions made in deducing the K-coefficients, then the function Rr(cb) is not linear and at least one of the K-coefficients is not constant. If the interferent signal R, is responsible for the deviation from the linearity and this effect is not large, the specificity of the method may be characterized by means of K-coefficients expressed as linear functions of c,. In such a case the &-value must be calculated from the functions Rd(cd) and R$.z,,), the second of which describes the component of the interfering signal that is responsible for the parallel‘shift of the calibration curve and is inde~ndent of the concentration of anaiyte (see Part I).’ For this purpose it is necessary to prepare an additional series of standards (see columns 6, 11 etc. in Table 1) that contain the corresponding concentrations of interferent (cb = c,-cd)but no analyte. The total number of standards necessary for testing the method with respect to such an interferent is 2q. The K-coefficients must then be calculated according to scheme 2. Calculation scheme 2
Steps l-5 are in scheme 1. 6. When F < Ftab calculate the parameters of the regression lines for R&cd) and R&T,,), using the weighting factors, verify the linearity by means of the F-test and check whether RJO) = 0 and R;(O) = 0 by means of the t-test. Rd(0) and R;(O) may be accepted as equal to zero when the corresponding values of lai/s, are less than tfab. Note. When Rd(cd) or R&J is linear and does not pass through the origin then the analytical procedure is incorrect, e.g., it is necessary to use a blank correo tion. 7~. When Rd(cd) and R;I(c,,) are linear and pass through the origin Kp must be constant. Calculate its value as the relation of the slope of R&J to that 0f R&i). 7b. When R&J is not linear, calculate the local values of k, and then the local values of Kp as the ratio of k, to kd, find the parameters of the regression line of K&c,,) and verify its linearity. If it is
619
linear then KP = a + bc,. Its confidence limit may be estimated only locally for individual c,-values: KP f Cf+bQ.*f0.95@-2). Note. If R&J is not linear or K&J is not linear and not constant, the specificity of the method relative to the interferent tested cannot be characterized by means of the K-coefficients. 8. Using the constant value of Kp or local values of Kp, calculate the local values of KA and their variances. 9. When Kr # const., test whether KA = const. by means of the t-test. Note. K, may be constant despite R&J not being linear, when only the Rb-signal is responsible for this deviation from linearity. 10. If KA Z const. calculate the parameters of the regression line K&z,) and verify its linearity. When it is linear KA may be expressed as KA = a + bcb and its confidence limit may be estimated only locally (see 76). Note, If KA Z const. and KA(c,,) is not linear, the specificity of the method relative to the interferent tested cannot be characterized by means of the K-coefficients. An example of calculation (according to scheme 2) of the K-coefficients for coulometric determination of uranium with molybdenum as interferent is given in Appendix II. Sometimes the coefficient KA ought to be expressed as a linear function of the analyte concentration; then all the operations given in step 10 must be carried out for K&J. The expression of the K-coefficients as linear functions is necessary when the influence of the interferent on the analytical signal is approximated better by a second-order function than a linear function and it usually concerns the coefficient KA that describes the inter-effect between the analyte and the interferent. The application of this procedure widens the possibilities of the approach. It is particularly useful when the K-coefficients are used for mathematical correction of the analytical results obtained in the presence of certain interferents, some of which do not fultil the ~mplif~g assump~ons well. The K-coefficients can also be expressed by means of functions other than linear, but this complicates the problem and finds application only for special cases. The procedure described allows the complete testing of an analytical method for all interferents of interest by use of relatively few standards, and also whether the simplifying assumptions are satisfied. However, it does not permit verification of the assumption that the activity of individual interferents is independent of their mutual proportions in the sample. One way of verifying this is to compare the result for a standard mixture containing known concentrations of the analyte and all interferents of interest with the result (corrected by means of the K-coeficients) for a sample containing the same concentration of pure analyte. The complete statistical evaluation of the data used
LEON PSZONKKI
620
and
ANNA LUKSZO-BI~~KOWSKA
for calibration of an analytical method according to schemes 1 and 2 is rather arduous and requires the use of a computer. However, particularly when the results are read from a graphical calibration curve, a much simpler procedure can often be used, that gives sufficiently correct results. The estimation of the parameters of all the test-functions as well as the verification of their linearity may be done graphically and the K-coefficients these parameters.
calculated.
simply on the basis of
EXPERIMENTAL
trodes. Equipment for controlled-potential coulometry based on the Tacussel PRT-500 Lc potentiostat with IG-38 integrator. Minicomputer. Methods
used
1. Spectrophotometric
uranium(W) determination with Arsenazo-III according to Savvin (without extraction).3 2. Spectrophotometric manganese determination with formaldoxime according to Marczenko.4 3. Polarographic determination of uranium(V1) in 2N sulphuric acid medium according to Kolthoff and Harris.’ 4. Controlled-potential coulometric determination of uranium(W) in 5N sulphuric acid medium according to Farrar et a1.6
Apparatus
Standard samples
Unicam SP 500 spectrophotometer. Radelkis OH-102 polarograph with dropping mercury and calomel elec-
The standard samples used for testing the methods above were prepared according to Table 1.
Table 2. K, and KA values for spectrophotometric determination of uranium(W) with Arsenazo III (uranium concentration range %201ugl 10 ml)
Interferent
Interf. concn. range, w/l0 ml
Th Ce Gd
&20 620 Cl5
0.95 1.62 1.64
-1.16 x 1O-2 -7.30 x 10-J -7.15 x 10-a
Table 3. KP and KA values for spectrophotometric determination of manganese with formaldoxime (manganese concentration range O-25 pg/ 1Oml)
Interferent
Interf. concn. range w/10 ml
Ni Fe co MO
O-20 O-40 O-300 0-1000
1.30 0.43 0.053 0
0 0 -3.00 x 1o-4 3.70 x 1o-5
Table 4. KP and KA values for polarographic determination uranium(V1) (uranium concentration range O-1000 pg/lO ml)
Interferent
Interf. concn. range w/l0 ml
KP
V cu MO
O-400 @200 O-250
0 2.336
0.357
of
(J;l)-,’ 5.58 x 1O-4 -5.02 x 1O-4 8.01 x lo-’
Table 5. K1’ and KA values for coulometric determination of uranium(W) (uranium concentration range &4 mg/sample) Interf. concn. range Interferent
mglsample
V MO Fe cu Ce
W.5 0.5 O-l.25 k2.0 W3.6
KA
KP
2.242 5.366 2.010 3.607 0.669
@g/sample)-
0.506 3.069-5.433 cb 0 0 0
621
Specificity characteristics of analytical methods-II
,5. UtlOThtSGdtSCe
Mnt2ONi
d
1.C
,’
UtlOGd
ut5ce Ut5Th U
Fig. 1. Comparison of experimentally determined (points) and calculated (dashed lines) interference effects in the spectrophotometric determination or uranium 1-Uranium; 2+individual interferents; 5-mixture of interferents. The numbers by the symbols of interferents indicate the concentrations in pgg/lO ml.
pg
Mn/lOml
Fig. 2. Comparison of experimentally determined (points) and calculated (dashed line) interference effects in spectrophotometric determination of manganese. l-Manganese; 2-5-individual interferents; 6-mixture of interferents. The numbers by the symbols of interferents indicate the concentrations in ng/lO ml.
RESULTS AND DISCUSSION The
ficients
possibility of applying the non-specificity coeffor determining the practical specificity
characteristics of the four methods mentioned above was investigated. Each method was tested with respect to the strongest interferents. The KP and KA values obtained experimentally are shown in Tables 2-5. The full data and calculations for an example of each computation scheme are given in Appendices I and II. The comparison of the experimental results with the results corrected by means of the K-coefficients is presented in Figs. 14. The points plotted represent the experimental values of the analytical signal for the analyte in presence of one interferent or a mixture of some interferents. The continuous line represents the normal calibration curve in the absence of interferents and the dashed lines represent the theoretically calculated values, i.e., the calibration curve values corrected by means of the K-coefficients for a defined concentration of one interferent or a mixture of interferents. It is seen from Tables 2-5 that for 15 interferents tested in the four methods the K-coefficients are constant and that only in one case, molybdenum in the coulometric determination of uranium, the coefficient KA must be described by a linear function. The agreement of the calculated and experimental values is excellent for the determinations in presence of single interferents in all the methods investigated. This fact confirms that the K-coefficients may be used successfully as a characteristic of analytical methods and that
the proposed calibration procedure based on a small number of standards gives sulkiently precise results. The worst situation is for the multi-interferent mix-
c+‘4.
Ut250
,3.
ut4oov
/5.
Utloocui .IOOV t 50Mo
,I.
u
,, /’
7’ ,
p’
I
I
I’ ,
,’
Id
’
I’ d’ , /
I’ I’
/, 200
I 400 pg
I’
MO
,o
/
I 600
I 800
U /IO
ml
/
/*.,2.
ut2oOcu
I 1000
Fig. 3. Comparison of experimentally determined (points) and calculated (dashed lines) interference effects in polarographic determination of uranium. l-Uranium; 2&individual interferents; 5-mixture of interferents. Numbers by the symbols of interferents indicate the concentrations in pg/lO ml.
LEONPSZONICKIand ANNALUKSZO-B&KOWSKA
622
ture. In this case the deviations of the calculated results from the experimental are significantly larger than the reproducibility of the methods. However, taking into consideration the very strong interfering signals and the possibility of accumulation of errors made in the estimation of K-values for single interferents, the additivity of interfering signals can be accepted as established. On the basis of these experiments we conclude that the non-specificity coefficients characterize quantitatively the specificity of the analytical procedures investigated.
,6.6. ,’
,’
Ut2.OCu
/’ 1 u+o.5v+o.I
/’
t
>-
REFERENCES 1 L. Pszonicki, Talanta, 1971, 24, 613. 2. 0. W. Davies and P. L. Goldsmith (eds.), Statistical Methods in Research and Production, 4th Ed., pp.
I
199-201. Oliver and Boyd, Edinburgh, 1972. 3. S. B. Savvin, Talanta, 1961, 8, 673. 4. Z. Marczenko, Spectrophotometric Determination of Elements, p. 342. Horwood, Chichester, 1976. 5. I. M. Kolthoff and W. E. Harris, J. Am. Chem. Sot.,
i
1946, 68, 1175.
L
6. L. Farrar, P. Thomason and M. T. Kelly, Anal. Chem., 1958, 30, 1511.
mg U /sample
I. S. Brandt, Statistical and Computational Methods in Data Analysis, North-Holland,
Amsterdam, 1976.
8. P. R. Bevington, Data Reduction and Error Analysis for the Physical Sciences, McGraw-Hill, New York, 1969. APPENDIX I Calculation of K, and KA values for spectrophotometric determination of uranium(V1) in presence of thorium. Note. All statistical calculations presented in Appendices I and II were carried out according to the formulae given by Brandt’ and Bevington.’ Number of standards in series; q = 5. Number of parallel determinations of each standard:
t= 5.
Fig. 4. Comparison of experimentally determined (points) and calculated (dashed lines) interference effects in coulometric determination of uranium. l-Uranium; 2-bindividual interferents; 7-mixture of interferents. Numbers by the symbols of interferents indicate the concentrations in mg per sample.
Relative concentration of interferent: c, = 1. cd and cb in &lOml. R, and R in absorbance units. All R, and R values are the mean values of five measurements and the variances are the variances of these means.
Experimental data Standard No. 1 2 3
4 5
Cd
Rd
s;, x lo6
2 5
0.043 0.107
0.2 0.8
10 15 20
0.214 0.319 0.428
1.8 5.0 5.0
R
s; x lo6
2 5
0.085 0.198
0.8 0.8
10 15 20
0.388 0.568 0.742
3.2 12.8 28.8
cb
Calculated values of the total relative signal R, Standard No:
Ch
R,
$, x 104
1 2 3 4 5
2 5 10 15 20
1.977 1.850 1.813 1.781 1.734
8.6 3.2 2.0 2.8 2.4
Regression parameters of R&J: b = -9.68 x 10m3; S, = 1.621 x lo-*; SW = 8.951 x 1O-4; F = 18.10 > F,.,,(,,,, = 10.13; R, (cb) is linear. SK, = 0.03; SK, = 2.3 x 10-3; to.ss(3j = 3.18; Kp = 0.92 f 0.09 and K, = -0.01 f 0.007 (pg/loml)-1. a = 1.92;
623
Specificity characteristics of analytical methods-II APPENDIX
Relative concentration of interferent: c. = 0.1. cd and c,, in mg/sample. Rdr Rb and R in coulombs. All Rd, Rb and R values are the mean values of five measurements and the variances are the variances of these means.
II
Calculation of KP and K,, values for coulometric determination of uranium(V1) in presence of molybdenum Number of standards in series: q = 5. Number of parallel determinations of each standard:
t= 5. Experimental data Standard No. 1 2 3 4 5
0.2 0.5 1.0 2.0 4.0
R,
& x lo6
Cb
Rb
s&, x lo6
R
s; x lo5
0.142 0.409 0.812 1.621 3.254
1.8 1.8 5.0 7.2 20.0
0.02 0.05 0.10 0.20 0.40
0.086 0.216‘ 0.435 0.868 1.742
1.8 1.8 7.2 9.8 12.8
0.275 0.685 1.455 3.133 6.157
0.32 0.98 2.0 4.6 18.0
.
Calculated values of total relative signal R, Standard No. I
2 3 4 5
0.02 0.05 0.10 0.20 0.40
1.586 1.675 1.792 1.933 1.892
29.5 8.9 5.5 2.8 2.4
Function R&j: a = 1.75; b = 0.445; S, = 1.84 x lo-‘; SW = 8.51 Function R&J: s, = 0.005; b = 0.811; a = 1.41 x 10-J; F = 1500 $ F0.95(1,3)= 10.13. R,(c,) is linear and passes through the origin.
x
10-s; F = 2.16 < Fe.sstf,s) = 10.13; Rt(cb) is not linear. t = 0.28 < t0.s5(aj = 3.18;
SL = 2.10;
S, = 1.4 x 10-d;
Function R&J: t = 1.44< t0,95(3j = 3.18; S, = 0.793; SW = 1.46 x lo-“; a = -9.90 x 10-4; b = 4.36; s, = 6.9 x 10-4; F = 5.4 x 10’ r> F0.95c1,3, = 10.13; R’&)b) is linear and passes through the origin; sK. = 0.047; Kr = 5.37 + 0.15. Local values of K, Standard No. 1
2 3 4 5
cb
0.02 0.05 0.10 0.20 0.40
KA
2.45 2.16 2.55 1.98 0.89
&A
6.4 x 3.2 x 5.2 x 7.2 x 1.6 x
10-s lo-& 1O-5 1O-6 10-e
Function K&J: U = 3.07; b = 5.46; S, = 1.17; S, = 1.19 x lo-+; F = 9800 % FO,,so,,, = 10.13; K&q,) is linear. KA = 3.07 - 5.46 q, (mg/sample)-‘.