Total Endovascular Repair of Thoracoabdominal Aortic Aneurysms With Non-Customized Stent Grafts

Total Endovascular Repair of Thoracoabdominal Aortic Aneurysms With Non-Customized Stent Grafts

ADULT CARDIAC Total Endovascular Repair of Thoracoabdominal Aortic Aneurysms With Non-Customized Stent Grafts Yongxue Zhang, MD, PhD,* Qingsheng Lu, ...

1MB Sizes 0 Downloads 30 Views

ADULT CARDIAC

Total Endovascular Repair of Thoracoabdominal Aortic Aneurysms With Non-Customized Stent Grafts Yongxue Zhang, MD, PhD,* Qingsheng Lu, MD,* Yifei Pei, MD, Mengtao Wu, MD, Suming Zhang, BS, Yi Hong, MD, and Zaiping Jing, MD Division of Vascular Surgery, Changhai Hospital; and Military Institute of Vascular Disease, the First Affiliated Hospital of the Second Military Medical University, Shanghai, China

Background. Total endovascular repair of thoracoabdominal aortic aneurysms with customized branched or fenestrated endografts could be technically challenging outside large-volume centers. This study aims to describe a new endovascular strategy for use of both noncustomized stent grafts and flow-diverting stents in treating complicated thoracoabdominal aortic aneurysms. Methods. Patients diagnosed with thoracoabdominal aortic aneurysms and deemed unfit for open surgical repair were recruited. The aim of the procedure was to cover the renovisceral segment of the aorta with flowdiverting uncovered stents, while covering the remaining aneurysm with stent grafts. Aneurysm morphologic evolution and the patency of the visceral branches were assessed at follow-up. Results. Between February 2012 and August 2013, 6 selective patients (4 men, mean age 58 years) underwent the novel joint procedure. During mean follow-up of

14 months, aneurysm shrinkage (maximum diameter decrease >5 mm) was demonstrated in 4 patients and aneurysm stabilization (maximum diameter decrease <5 mm) was observed in 2 patients. No aneurysm expansion was observed in any participants. Mean aneurysm diameter decreased from 65.0 ± 8.8 mm to 58.5 ± 12.2 mm (p [ 0.054), with a significant increase in average sac thrombus deposition volume (sac thrombosis ratio increased from 23.3% ± 7.4% to 98.0% ± 3.3%, p<0.001). The majority of side branches (23 of 24) were successfully preserved. Conclusions. Complete endovascular repair of thoracoabdominal aortic aneurysms with this novel joint procedure may be a feasible alternative in high surgical risk patients. Further validation of this technique is required to substantiate these results.

D

Driven by the clinical need for an easy-to-perform endovascular approach, Sultan and colleagues [11] recently reported their initial experience of using a new concept of multilayer uncovered stent (Cardiatis, Isnes, Belgium) to treat TAAAs. Technical success was reported at 97.3% in their series, with the 1-year survival being 86.8%. However, although there was no aneurysm rupture documented within the follow-up period, adverse changes including an increase in sac volume and aneurysm diameter were documented. Aneurysm size has been traditionally considered as an important predictor of the risk of future rupture [12], which indicates the essentiality of further validation this technique before its large-scale application. In a pilot study using multiple overlapping uncovered stents as an alternative to the commercial Cardiatis multilayer stents in the treatment of pararenal aortic aneurysms [13], we have demonstrated that aneurysm morphologic characteristics, such as the aneurysm “sac entrance (SE)” (defined as the threshold between the normal laminar flow of the native aorta and the turbulent flow entering the aneurysmal sac; Figs 1A, 1B) could impact on the thrombosis process after stenting. Short-SE aneurysms were observed to have a better clinical prognosis compared with wide-SE ones. In a following case of a wide-SE TAAA

espite significant advances in surgical technique and perioperative adjuncts, open surgical repair of thoracoabdominal aortic aneurysms (TAAAs) remains a challenging procedure associated with significant mortality and morbidity rates [1–3]. Open-endovascular hybrid procedures have also been shown to carry a similar operative risk [4–7] despite the theoretic advantages associated with reducing the total ischemic time. Through development of advanced branched or fenestrated endografts, total endovascular repair of TAAAs have become possible, offering high surgical risk patients a minimally invasive alternative [7, 8]. Though attractive, satisfactory procedural outcomes are only reproducible in experienced high-volume centers. Although standardized off-the-shelf endografts are available [9], around a third of the patients will still require custom-made “devices due to the complexity of the aneurysm anatomy [10], which prohibits its application in emergency cases.

Accepted for publication June 24, 2014. *Yongxue Zhang and Qingsheng Lu contributed equally to this work. Address correspondence to Dr Lu, Division of Vascular Surgery, Changhai Hospital, 168 Changhai Rd, 200433 Shanghai, China; e-mail: luqs@ xueguan.net.

Ó 2014 by The Society of Thoracic Surgeons Published by Elsevier

(Ann Thorac Surg 2014;98:1606–12) Ó 2014 by The Society of Thoracic Surgeons

0003-4975/$36.00 http://dx.doi.org/10.1016/j.athoracsur.2014.06.064

ZHANG ET AL JOINT ENDOVASCULAR PROCEDURE FOR TAAAS

1607

anesthesiologists according to the following standards: age greater than 80 years; American Society of Anesthesiologists classification IV or greater; New York Heart Association classification (1994) of cardiac function III/C or greater; or dysfunction of other important organ systems (ie, severe chronic obstructive pulmonary disease or renal insufficiency) [15]. Patient records were reviewed for preoperative, intraoperative, and follow-up data, including demographics (gender and age at hospitalization time), concomitant morbidities (hypertension, stroke, coronary artery disease, diabetes mellitus, chronic obstructive pulmonary disease, renal failure, and hepatitis), intervention details (length of procedural time, X-ray exposure time, blood loss, number of stents and stent grafts implanted) and postoperative events (aneurysm morphologic evolvement, sac thrombosis ratio, survival rate).

Procedural Details

Fig 1. The aneurysm sac entrance was defined as the threshold between the normal laminar flow of the native aorta and the turbulent flow of aneurysm sac (A: sac entrance for saccular aneurysms; B: sac entrance for fusiform aneurysms), whose length could exert an influence on the thrombosis process of the aneurysm sac according to our previous study [13]. The length of the aneurysm sac entrance could be artificially shortened by a stent graft covering part of it. The remaining part was named as residual sac entrance (C: residual sac entrance for saccular aneurysms; D: residual sac entrance for fusiform aneurysms). (SE ¼ sac entrance.)

involving the celiac artery (CA) and superior mesenteric artery [14], we attempted to artificially shorten the aneurysm SE by a stent graft covering part of it (Figs 1C, 1D), followed by 2 layers of uncovered stents to cover the residual SE. This case was associated with an acceptable short-term outcome, with a corresponding decrease in aneurysm diameter by up to 19 mm, and all major branches being patent. In the present study, we analyzed the results of the first clinical series of this “joint” endovascular procedure consisting of stent grafts and bare stents in the treatment of TAAAs, and evaluated the safety and efficacy of this novel technique on midterm outcomes.

Patients and Methods This study was approved by the local ethical review board. The written informed consent was obtained from each patient prior to surgery.

Study Design and Patient Cohort Data on all patients diagnosed with TAAAs and treated in our center with the joint procedure of stent grafts and uncovered stents between February 2012 and August 2013 were retrospectively analyzed. All patients enrolled into study were judged as poor candidates for open surgical repair by a multidisciplinary surgical team consisting of vascular surgeons, cardiothoracic surgeons, and

For a type Ⅴ TAAA (extended from the distal thoracic aorta to above the renal arteries), a standard joint procedure could be performed (Fig 2A: schematic for the procedure; Fig 2B: pre-surgical image; Fig 2C: poststenting image). Under spinal anesthesia, the femoral artery was exposed and heparin given intravenously (80 IU/kg). An angiogram was performed through a pigtail catheter to conform the size and location of the aneurysm. A stiff wire was then placed bridging the lesion segment, after which a tubular stent graft (Hercules-T, MicroPort, Shanghai, China) was advanced and placed covering the proximal part of the aneurysm SE, with a proximal landing zone of 15 mm or greater on the normal aorta. The distal edge of the stent graft should approach the ostia of the CA, but not cover it. Subsequently, multiple overlapping bare stents (Sinus-XL stents; OptiMed, Ettlingen, Germany) were placed beneath the previous stent graft to cover the residual SE that involved the visceral branches, with an overlapping zone of 15 mm or greater with the stent graft proximally, and a distal landing zone of 15 mm or greater on the distal normal aorta. The number of the bare stents implanted was determined by intraoperative angiogram, with the criterion that a decrease of flow velocity within the aneurysm sac was achieved on fluoroscopy. The diameter of the stents and stent grafts conformed to manufacturers’ instructions, with 10% to 20% oversizing to the aneurysm neck. Technical success was defined as successful deployment of the stents to target locations without procedure-related complications (stent migration, stent twist and kinks, etc.) [16]. In cases of long-segmental aneurysmal lesions in the thoracic aorta (type Ⅰ TAAA), 2 tubular stent grafts (Hercules-T, MicroPort, Shanghai, China) could be needed to cover the entire proximal aneurysm SE (Fig 3A). For a TAAA that mainly affected the infrarenal aorta rather than the thoracic aorta (type Ⅳ TAAA), a “reversed procedure” could be considered, in which the stent graft (Hercules-B, MicroPort) could be placed at the infrarenal aorta to cover the distal aneurysm SE, while the multiple bare stents were deployed above it to cover the proximal aneurysm residual SE (Fig 3B).

ADULT CARDIAC

Ann Thorac Surg 2014;98:1606–12

1608 ADULT CARDIAC

ZHANG ET AL JOINT ENDOVASCULAR PROCEDURE FOR TAAAS

Ann Thorac Surg 2014;98:1606–12

Fig 2. A standard joint procedure of stent grafts and bare stents. (A) The whole stenting zone (1) consisted of 2 parts: A stent graft was first placed to cover the proximal aneurysm sac entrance (2), with at least 15-mm landing zone on proximal normal aorta (5). Multiple overlapping bare metal stents were then deployed beneath the previous stent graft (3), with an overlapping zone (4) and a distal landing zone (6) of 15 mm at least, respectively. (B) Application of the standard joint procedure in a clinical case with type Ⅴ thoracoabdominal aortic aneurysm involving the visceral branches. (C) Follow-up imaging at 12 months revealed complete thrombosis of the aneurysm sac and patent side branches. (White arrow ¼ stent graft coverage zone; white arrowhead ¼ bare stents coverage zone; BMS ¼ bare metal stent; SG ¼ stent graft.)

Follow-Up and Parameter Measurement The patients were examined by computed tomography angiography (CTA) on a multislice CT scanner (Sensation

Cardiac 64; Siemens, M€ unchen, Germany) before intervention and postoperatively at 3 and 6 months initially, and then every 6 months afterward. Data in DICOM format were then

Fig 3. (A) In the treatment of type Ⅰ thoracoabdominal aortic aneurysm that involved most of the descending thoracic aorta and extended to the suprarenal abdominal aorta, 1 stent graft might be insufficient to cover the whole proximal part of the aneurysm sac entrance. Therefore, 2 stent grafts (2, 3) were deployed consecutively at the proximal sac entrance, with an overlapping zone of 15 mm (6). The distal end of the stent graft reached the origin of the celiac artery, but should not cover it; then multiple overlapping uncovered stents were placed beneath the stent grafts to cover the distal residual sac entrance involving the visceral branches (4), with an overlapping zone of 15 mm with the stent graft (7); the whole stenting zone (1) had a proximal landing zone (5) and a distal landing zone (8) of larger than 15 mm, respectively. (B) Patients with type Ⅳ TAAAs (mainly affected abdominal aorta below the diaphragm) could receive a “reversed” joint procedure. The whole stenting zone (1) consisted of 2 parts: a bifurcated stent graft (2) and multiple overlapping uncover stents (3). The bifurcated stent graft (2) was placed under the renal arteries to cover the distal sac entrance. The distal landing zone on the iliac arteries was 15 mm or more (6). For the residual sac entrance above the renal arteries, multiple overlapping uncovered stents were applied (3). The overlapping zone of the stent graft and bare stents was greater than 15 mm (4). The proximal landing zone for the bare stents was 15 mm or more (5). (BMS ¼ bare metal stent; SG ¼ stent graft.)

imported into CT-imaging process software (Aquarius Workstation, Version 3.7.0.13; TeraRecon, Foster City, CA) for three-dimensional reconstruction of the aortic contour and measurement (Figs 4A, 4B). Preoperative (Fig 4C) and postoperative aneurysm maximum diameters (Fig 4D) were measured on the cross-sectional images on the same aortic level referring to the vertebrae, as utilized in previous studies [13, 16]. The visceral branches covered by the bare stents were noted during surgery, and the patency of each was assessed during follow-up. Sac thrombosis ratio, defined as the ratio of the sac thrombus volume to the total sac volume, was calculated from the integration of the thrombus and sac areas of each slice. All measurements were made by the first author and independently verified by 2 coauthors. Clinical success was defined as either aneurysm shrinkage or stabilization, combined with patency of major visceral branches [16].

Statistical Analysis Continuous variables were presented as mean  standard deviation. Changes in aneurysm diameter and sac thrombosis ratio were analyzed using a paired t test. Statistical significance was considered when a p value less than 0.05. All analyses were performed with PASW Statistics software (IBM, Armonk, NY).

Results A total of 6 patients (4 men, mean age 58 years) with TAAAs were treated with this joint procedure during the study period. Patients’ demographics, comorbidities,

ZHANG ET AL JOINT ENDOVASCULAR PROCEDURE FOR TAAAS

1609

and surgical details are listed in Table 1. Four patients presented with type Ⅴ TAAAs (extended from the distal thoracic aorta to above the renal arteries) and received the standard joint procedure with 1 stent graft covering the aneurysm proximal SE, and 2 overlapping bare stents covering the distal SE; in a patient with type I TAAA (involved most of the descending thoracic aorta from the origin of the left subclavian artery to the suprarenal abdominal aorta), 2 stent grafts were placed to cover the aneurysm proximal SE, and 2 uncovered stents for the aneurysm distal SE; another patient presenting with type Ⅳ TAAA (mainly affected abdominal aorta below the diaphragm) received a “reversed procedure,” with a bifurcated stent graft placed at the aneurysm distal SE and bare stents at the proximal SE. Average procedural and X-ray exposure time were 88.3  19.7 and 41.7  8.2 minutes, respectively; mean contrast usage was 103.3  19.7 mL, and mean blood loss was 125  43.7 mL. Technical success was achieved in all patients. Three patients experienced a postimplantation fever, all of which resolved prior to discharge. No signs or symptoms of spinal cord ischemia were observed. Mean length of hospital stay was 9.7  2.8 days. In 3 patients with previous coronary artery disease or stroke, routine antiplatelet medications including aspirin and clopidogrel were prescribed upon hospital discharge. All patients underwent regular follow-up (mean 14 months, range 6 to 24 months). Aneurysm shrinkage (decrease in diameter  5 mm) was documented in 4 patients, aneurysm stabilization (decrease in diameter

Fig 4. Preoperative and follow-up imaging. Three-dimensional reconstruction of the aortic contour was performed with a dedicated computed tomographic imaging workstation. (A) Visceral branches within the stenting zone were noted before intervention (white arrows), and (B) the patency of each was assessed during follow-up (white arrows); the maximum aneurysm diameter (C) before and (D) after surgery was measured on the same aortic level referring to the vertebrae.

ADULT CARDIAC

Ann Thorac Surg 2014;98:1606–12

1610 ADULT CARDIAC

ZHANG ET AL JOINT ENDOVASCULAR PROCEDURE FOR TAAAS

Ann Thorac Surg 2014;98:1606–12

Table 1. Patient Demographics, Comorbidities, and Surgical Details

Gender/Age No. (Years) 1

M/38

2

M/54

3

M/58

4

M/54

5

F/83

6

F/62

a

Comorbidities

Pre/Postoperative Aneurysm Follow-Up Maximal Pre/Postoperative Procedural Details (Months) Diameter (mm) STR(%)

Hypertension, CAD

Standard procedure SG1,BMS2 Hypertension, Standard procedure Liver dysfunction SG1,BMS2 Hypertension, Standard procedure Stroke SG1,BMS2 CAD,DM Reversed procedure SG1,BMS2 Hypertension, Standard procedure Renal failure, DM SG1,BMS2 Hypertension, COPD Standard procedure SG2,BMS2

Visceral Branches Covered

Visceral Branches Occluded

6

68.2/65.1

22.6/92.2

SMA,CA,RA

None

6

59.5/54.3

36.6/100

SMA,CA,RA

None

18

60.9/54.6

19.1/100

SMA,CA,RA

None

12

55.4/50.2

23.9/100

SMA,CA,RA

RA1a

18

65.5/46.8

26.4/100

SMA,CA,RA

None

24

80.3/79.9

14.7/98.5

SMA,CA,RA

None

This renal artery was already stenotic before surgery, and it was occluded at 12-month follow-up.

BMS ¼ bare metal stent; CA ¼ celiac artery; diabetes mellitus; F ¼ female; M ¼ male; artery; STR ¼ sac thrombosis ratio.

CAD ¼ coronary artery disease; COPD ¼ chronic obstructive pulmonary disease; DM ¼ No. ¼ number; RA ¼ renal artery; SG ¼ stent graft; SMA ¼ superior mesenteric

< 5 mm) was seen in 2 patients, and no aneurysm expansion was observed. Mean aneurysm diameter decreased from 65.0  8.8 mm to 58.5  12.2 mm (p ¼ 0.054); while a significant increase in average sac thrombus deposition volume was documented (sac thrombosis ratio increased from 23.3%  7.4% to 98.0%  3.3%, p < 0.001). Complete thrombosis of the aneurysm sac was observed in 4 patients, while mild residual perfusion was seen in the other 2 patients. No proceduralrelated mortality or complications occurred during the follow-up period. The superior mesenteric artery, CA, and bilateral renal arteries (RAs) were covered by 2 layers of bare stents in all patients (24 major branches in total). During follow-up, the majority of major branches were patent, based on CT imaging (23 of 24). Only 1 RA, which had been stenotic before intervention, was occluded at 12-month follow-up, translating into a clinical success rate of 83.3% (5 of 6) in the study cohort. No clinical signs of intestinal ischemia were found in any patient.

Comment The concept of using bare metal stents to treat aneurysms was first described 2 decades ago by Geremia and colleagues [17], who suggested that an uncovered stent of low mesh porosity would alter the local blood flow pattern by decreasing the mean flow velocity within the sac, thereby creating a favorable environment for thrombus deposition. Secondary to the formation of mural thrombus, which increases the effective wall thickness, the tensile stress of the aneurysm wall could be decreased according to the Laplace Law, which illustrates that the wall tension is inversely proportional to the wall thickness in an aneurysm. Early clinical application of this theory has been mainly focusing on

intracranial aneurysms, with a favorable aneurysm occlusion rate of 85.7% to 94.4% [18–21]. After Henry and colleagues [22] first reported of a renal artery aneurysm successfully treated with the commercial Cardiatis multilayer stents in 2008, this technique has been widely applied in treating complicated peripheral aneurysms. Data from independent centers seem encouraging for both short and midterm outcomes. A multicenter study conducted by Ruffino and colleagues [23] demonstrated an overall aneurysm thrombosis rate of 93.3%, with a branch patency rate of 96.1% at 1-year follow-up. However, concerns have been raised regarding the safety and efficacy of this technique in treating aortic aneurysms [24]. Persistent sac perfusion and subsequent aneurysm rupture have been reported by Lazaris and colleagues in 2012 [25]; stent rigidity leading to aortic rupture has been noticed by Ferrero and colleagues in 2013 [26]. Sultan and colleagues [27] recently released a result of 38 patients with TAAAs treated by off-label use of multilayer stents, in which an aneurysm-related mortality rate of 71.1% was documented. These drawbacks indicate that further refinements of this technique are required before its large-scale clinical application. In a computational fluid dynamic study performed by Seshadhri and colleagues [28], the hemodynamic effects of the flow-diverting stents were analyzed, taking into account the morphologic characters of the aneurysm. The investigators found that the “dome to neck ratio (D/N ratio)” was crucial in determining the flow-modulating effect of the porous stents (Fig 5). A larger D/N ratio could lead to a more significant decrease in wall shear stress within the aneurysm sac. This theory is in accordance with our previous clinical investigation [13], in which we found that the aneurysm SE length could influence the thrombosis process within the sac.

Fig 5. Computational analysis of flow-diverting stents in treating aneurysms [27]. The dome to neck ratio (D/N ratio) was defined as the ratio between the length of the dome and the length of the neck. A larger D/N ratio would decrease the aneurysm wall shear stress more significantly. In this theory, the “neck” was equivalent to “sac entrance” described in our study.

While the dome length is fixed for a given aneurysm, the neck length could be shortened by a stent graft covering part of it. The first validation of this theory was successful, with significant aneurysm shrinkage documented after intervention [14]. One-year outcomes within our clinical series were also satisfactory; technical success was achieved in all cases, decrease in maximum diameter by over 5 mm was documented in 4 out of 6 patients (range, 5.2 to 18.6 mm), aneurysm stabilization was seen in the remaining 2 patients, and no aneurysm expansion was noticed. This result is in contrast with previously published less-favorable data regarding the flow-diverting stents in treating aortic pathologies [11, 27], and offers a preliminary experience in improving this technique by using the stent grafts as adjuncts. Branched and fenestrated endografts have been designed for total endovascular repair of complicated TAAAs in high surgical risk patients. Pooled data from different centers revealed a technical success and branch patency rate of 94.2% and 95%, respectively [29]. However, a successful branched or fenestrated procedure often requires meticulous patient selection, appropriate customized device, and experienced technical expertise with endovascular skills, which makes the encouraging figures from high-volume centers difficult to duplicate in less-experienced ones. Furthermore, the endovascular procedure can be technically challenging and timeconsuming even for experienced surgeons, especially in the presence of unusual anatomy [30]. The joint endovascular procedure described in this study could offer a potential alternative to the traditional complex and cumbersome branched and fenestrated procedures.

ZHANG ET AL JOINT ENDOVASCULAR PROCEDURE FOR TAAAS

1611

Technically, a joint procedure could be finished within 90 minutes on average and the total X-ray exposure time could be limited to 50 minutes, with low contrast dose and little blood loss. Additionally, this procedure can be conducted safely and effectively without customized stent grafts, which makes it easy to perform in less-experienced centers. In our series, occlusion of one RA was observed in a 54-year-old male. This event occurred despite administration of dual antiplatelet therapy as recommended by Euringer and colleagues [31]. Interestingly, in the 3 patients without antiplatelet therapy, RA side branches were all successfully preserved at follow-up. This suggests that ostial or proximal side-branch disease before intervention, rather than the antiplatelet therapy initiation, may be a more important factor in determining future side-branch vessel patency. In conclusion, we describe acceptable midterm outcomes in a pilot study assessing the “joint” endovascular procedures for the treatment of complex TAAAs in highrisk surgical patients. Further validation of this novel and promising technique is now required before adoption into clinical practice. This work is supported by the following public foundations: National Nature Science Foundation of China (81170291, 39800177, and 30772140, Qingsheng Lu), Chinese Scholarship Council (201306580020, Yongxue Zhang), and the Innovation Foundation for PhD Student of Second Military Medical University (20122045, Yongxue Zhang). The authors are grateful to Dr Adam J Brown, University of Cambridge for his valuable suggestions in preparing the manuscript.

References 1. Coselli JS, Bozinovski J, LeMaire SA. Open surgical repair of 2286 thoracoabdominal aortic aneurysms. Ann Thorac Surg 2007;83:S862–4. 2. Richards JM, Nimmo AF, Moores CR, Hansen PA, Murie JA, Chalmers RT. Contemporary results for open repair of suprarenal and type IV thoracoabdominal aortic aneurysms. Br J Surg 2010;97:45–9. 3. Rigberg DA, McGory ML, Zingmond DS, et al. Thirty-day mortality statistics underestimate the risk of repair of thoracoabdominal aortic aneurysms: a statewide experience. J Vasc Surg 2006;43:217–22. 4. Black SA, Wolfe JH, Clark M, Hamady M, Cheshire NJ, Jenkins MP. Complex thoracoabdominal aortic aneurysms: endovascular exclusion with visceral revascularization. J Vasc Surg 2006;43:1081–9. 5. Chiesa R, Tshomba Y, Marone EM, et al. Hybrid procedures for the treatment of thoracoabdominal aortic aneurysms and dissections. J Cardiovasc Surg (Torino) 2010;51:821–32. 6. Drinkwater SL, B€ ockler D, Eckstein H, et al. The visceral hybrid repair of thoraco-abdominal aortic aneurysms–a collaborative approach. Eur J Vasc Endovasc Surg 2009;38: 578–85. 7. Clough RE, Modarai B, Bell RE, et al. Total endovascular repair of thoracoabdominal aortic aneurysms. Eur J Vasc Endovasc Surg 2012;43:262–7. 8. Amiot S, Haulon S, Becquemin JP, et al. Fenestrated endovascular grafting: the French multicentre experience. Eur J Vasc Endovasc Surg 2010;39:537–44. 9. Chuter T, Greenberg RK. Standardized off-the-shelf components for multibranched endovascular repair of

ADULT CARDIAC

Ann Thorac Surg 2014;98:1606–12

1612 ADULT CARDIAC

10. 11.

12.

13. 14.

15.

16. 17. 18.

19.

20.

ZHANG ET AL JOINT ENDOVASCULAR PROCEDURE FOR TAAAS

thoracoabdominal aortic aneurysms. Perspect Vasc Surg Endovasc Ther 2011;23:195–201. Ferreira M, Lanziotti L, Cunha R, d’Utra G. Best surgical option for thoracoabdominal aneurysm repair - the endovascular approach. Ann Cardiothorac Surg 2012;1:345–9. Sultan S, Sultan M, Hynes N. Early mid-term results of the first 103 cases of multilayer flow modulator stent done under indication for use in the management of thoracoabdominal aortic pathology from the independent global MFM registry. J Cardiovasc Surg (Torino) 2014;55:21–32. Lo RC, Lu B, Fokkema MT, et al. Relative importance of aneurysm diameter and body size for predicting abdominal aortic aneurysm rupture in men and women. J Vasc Surg 2014;59:1209–16. Zhang YX, Lu QS, Feng JX, et al. Endovascular management of pararenal aortic aneurysms with multiple overlapping uncovered stents. J Vasc Surg 2013;58:616–23. Zhang YX, Lu QS, Jing ZP. Innovative sandwich technique in treating a symptomatic saccular suprarenal aneurysm involving the celiac trunk. Vasc Endovascular Surg 2013;47: 102–5. Lu Q, Feng J, Zhou J, et al. Endovascular repair of ascending aortic dissection: a novel treatment option for patients judged unfit for direct surgical repair. J Am Coll Cardiol 2013;61:1917–24. Chaikof EL, Blankensteijn JD, Harris PL, et al. Reporting standards for endovascular aortic aneurysm repair. J Vasc Surg 2002;35:1048–60. Geremia G, Haklin M, Brennecke L. Embolization of experimentally created aneurysms with intravascular stent devices. AJNR Am J Neuroradiol 1994;15:1223–31. Lylyk P, Miranda C, Ceratto R, et al. Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 2009;64:632–42. Szikora I, Berentei Z, Kulcsar Z, et al. Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device. AJNR Am J Neuroradiol 2010;31:1139–47. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D. The pipeline embolization device for the intracranial treatment of aneurysms trial. AJNR Am J Neuroradiol 2011;32:34–40.

Ann Thorac Surg 2014;98:1606–12

21. McAuliffe W, Wycoco V, Rice H, Phatouros C, Singh TJ, Wenderoth J. Immediate and midterm results following treatment of unruptured intracranial aneurysms with the pipeline embolization device. AJNR Am J Neuroradiol 2012;33:164–70. 22. Henry M, Polydorou A, Frid N, et al. Treatment of renal artery aneurysm with the multilayer stent. J Endovasc Ther 2008;15:231–6. 23. Ruffino MA, Rabbia C; Italian Cardiatis Registry Investigators Group. Endovascular repair of peripheral and visceral aneurysms with the Cardiatis multilayer flow modulator: one-year results from the Italian Multicenter Registry. J Endovasc Ther 2012;19:599–610. 24. Lazaris AM, Charalampopoulos A, Maheras AN, Vasdekis SN. Flow-diverting multilayer stents: a promising but questionable solution for aortic pathologies. J Endovasc Ther 2013;20:378–80. 25. Lazaris AM, Maheras AN, Vasdekis SN. A multilayer stent in the aorta may not seal the aneurysm, thereby leading to rupture. J Vasc Surg 2012;56:829–31. 26. Ferrero E, Gibello L, Ferri M, Viazzo A, Nessi F. Aortic arch rupture after multiple multilayer stent treatment of a thoracoabdominal aneurysm. J Vasc Surg 2013: http://dx.doi.org/ 10.1016/j.jvs.2013.08.006. [Epub ahead of print]. 27. Sultan S, Hynes N, Sultan M; MFM Collaborators. When not to implant the multilayer flow modulator: lessons learned from application outside the indications for use in patients with thoracoabdominal pathologies. J Endovasc Ther 2014;21: 96–112. 28. Seshadhri S, Janiga G, Beuing O, Skalej M, Th evenin D. Impact of stents and flow diverters on hemodynamics in idealized aneurysm models. J Biomech Eng 2011;133: 071005. 29. Bakoyiannis CN, Economopoulos KP, Georgopoulos S, et al. Fenestrated and branched endografts for the treatment of thoracoabdominal aortic aneurysms: a systematic review. J Endovasc Ther 2010;17:201–9. 30. Riga CV, Jenkins MP. Best surgical option for thoracoabdominal aneurysm repair - the hybrid approach. Ann Cardiothorac Surg 2012;1:339–44. 31. Euringer W, S€ udkamp M, Rylski B, Blanke P. Endovascular treatment of multiple HIV-related aneurysms using multilayer stents. Cardiovasc Intervent Radiol 2012;35:945–9.