Journal of the American College of Cardiology © 2011 by the American College of Cardiology Foundation Published by Elsevier Inc.
Vol. 58, No. 21, 2011 ISSN 0735-1097/$36.00 doi:10.1016/j.jacc.2011.09.009
FOCUS ISSUE: STRUCTURAL HEART DISEASE
Prosthetic Valve Failure State-of-the-Art Paper
Transcatheter Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves Ronen Gurvitch, MBBS,*† Anson Cheung, MD,* Jian Ye, MD,* David A. Wood, MD,* Alexander B. Willson, MBBS,* Stefan Toggweiler, MD,* Ronald Binder, MD,* John G. Webb, MD* Vancouver, British Columbia, Canada; and Melbourne, Victoria, Australia When bioprosthetic cardiac valves fail, reoperative valve replacement carries a higher risk of morbidity and mortality compared with initial valve replacement. Transcatheter heart valve implantation may be a viable alternative to surgical aortic valve replacement for high-risk patients with native aortic stenosis, and valve-in-valve (Vin-V) implantation has been successfully performed for failed surgical bioprostheses in the aortic, mitral, pulmonic, and tricuspid positions. Despite some core similarities to transcatheter therapy of native valve disease, V-in-V therapy poses unique clinical and anatomic challenges. In this paper, we review the challenges, selection criteria, techniques, and outcomes of V-in-V implantation. (J Am Coll Cardiol 2011;58:2196–209) © 2011 by the American College of Cardiology Foundation
Despite the shorter durability of bioprosthetic cardiac valves compared with mechanical prostheses, the former are often used to reduce thromboembolic risk and to avoid anticoagulation and the associated increased risk of bleeding (1,2). When bioprosthetic valves degenerate, reoperative valve replacement is the current standard of care (3–5). However, patients requiring redo valvular surgery are generally at an increased risk of adverse perioperative events. Factors that have been shown to portend a higher reoperative risk include advanced age, renal failure, pulmonary disease, cognitive impairment, reduced ejection fraction, higher New York Heart Association functional class, need for concomitant bypass surgery, need for more than 1 reoperation, and reoperation for failed mitral valves (higher risk than aortic) (3,5–12). Operative mortality for redo valve surgery may be higher than 20% in such high-risk patients, whereas in low-risk patients, it may be ⬍4% (8,12). Transcatheter aortic valve replacement for native aortic stenosis has evolved as a viable alternative to open heart surgery in patients with high or prohibitive surgical risk (13–20). As transcatheter aortic valve replacement evolved and clinical outcomes improved, transcatheter treatment of
From the *Department of Cardiology and Cardiothoracic Surgery, St. Paul’s Hospital, University of British Columbia, Vancouver, British Columbia, Canada; and the †Department of Cardiology, The Royal Melbourne Hospital, Melbourne, Victoria, Australia. Drs. Cheung, Ye, and Webb are consultants to Edwards Lifesciences Inc. Dr. Wood is a consultant to St. Jude Medical and Edwards Lifesciences. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose. Manuscript received September 12, 2011; accepted September 13, 2011.
degenerated bioprosthetic valves became possible using valve-in-valve (V-in-V) implantation, in which the new transcatheter valve is inserted inside the degenerated bioprosthesis. V-in-V implantation has been successfully performed in degenerated aortic, mitral, pulmonic, and tricuspid bioprostheses as well as in pulmonary conduits. A variety of percutaneous and minimally invasive surgical techniques have been used. This article reviews the challenges, selection criteria, techniques, and what is known of the outcomes of this evolving therapy. Structural elements of bioprosthetic valves. Contemporary surgical bioprostheses generally incorporate leaflets derived from porcine valve leaflets or bovine pericardium (heterograft/xenograft). These are generally preserved in glutaraldehyde to attenuate calcific degeneration by crosslinking collagen fibers, reducing antigenicity, and preventing remodeling of extracellular matrix (21,22). Human tissue (homograft) is less commonly used. For the less common pulmonic position, valved conduits of various types are often used. Bioprosthetic valves may be stented or stentless. Stented valves are usually constructed of whole porcine aortic valves or bovine pericardium, suspended from a support structure such as a stent or frame (Fig. 1). The support structures of current valves are composed of various alloys or plastics, designed to absorb forces on the leaflets during valve function and maintain structural integrity. The support structure is attached to a basal ring, which is covered by a fabric sewing cuff. The basal ring may be circular or scallop shaped. Newer valve designs often incorporate low-profile
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
or supra-aortic design features intended to optimize hemodynamic functioning (e.g., Mitroflow or Soprano valve, Sorin, Canada). Stentless valves are sutured to the root in the position of a native valve and eliminate the support stent/frame to improve hemodynamic performance and durability (23,24). These may be of autograft, heterograft, or homograft origin. Although stentless valves tend to have better laminar flow and lower post-operative gradients, clinical outcomes have not been superior to stented valves (25,26). Some stentless valves have been associated with exuberant calcification of the aortic root, posing a particular challenge for reoperation. Stentless valves pose unique challenges for V-in-V therapy given the lack of a frame to anchor the new transcatheter heart valve (THV), and the absence of radiopaque markers to aid with positioning. Causes of bioprosthetic valve failure. Failure of bioprosthetic valves may occur due leaflet failure or nonleaflet failure or both. Leaflet failures are the result of leaflet degeneration (wear and tear or calcification) or leaflet destruction (due to endocarditis). Nonleaflet failures are typically the result of pannus, thrombus, or paravalvular leaks. The outcomes of these failures may be stenosis, regurgitation, or a combination of both. In general, failure due to calcification, pannus formation, or thrombus results in valvular stenosis, and failure due to leaflet destruction (endocarditis) or paravalvular leak results in regurgitation. In failure due to wear and tear, which is frequently associated with calcification and abnormal leaflet coaptation, mixed stenosis/regurgitation is common. With current bio-
Figure 1
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
2197
prosthetic valves, freedom from Abbreviations failure at 10 years is between 70% and Acronyms and 90% and at 15 years between THV ⴝ transcatheter heart 40% and 70% (22,27–31). Comvalve mon risk factors for bioprosthetic V-in-V ⴝ valve-in-valve failure include mitral valve position, younger age at implantation, renal failure, and hyperparathyroidism (22,29 –31). Bioprosthetic valve calcification is markedly accelerated in younger patients, with failure rates as high as 40% at 4 years in patients younger than 30 years old (29). Not surprisingly, the lifetime risk of needing reoperation decreases with the increasing age of the patient (29,32). When bioprosthetic valves degenerate, calcific deposits are often located within the leaflet tissue, with particular predisposition to areas of high leaflet stress (e.g., commissural and attachment points). Glutaraldehyde fixation treatments creating stable crosslinks between collagen aim to render the valve material inert, although residual glutaraldehyde may also serve as calcium binding sites (22). Various anticalcification treatments aim to reduce residual glutaraldehyde or phospholipids (e.g., Medtronic AOS [Medtronic Inc., Minneapolis, Minnesota] or Edwards ThermaFix [Edwards Lifesciences, Irvine, California]). Frequently, leaflet calcification and subsequent tears also result in some degree of valvular regurgitation. Although calcification is responsible for the majority of bioprosthetic valve failures, progressive collagen damage related to a number of immune and atherosclerotic processes may also be contributory (33–35). Pannus formation results from growth of host
Schematic Diagram of Stented Bioprosthetic Valves
Valve leaflet (A), stent frame (B), and external sewing ring (C). The internal, outer, and external diameters all represent different dimensions of surgical bioprostheses.
2198
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
tissue and is to some extent part of the normal healing reaction to prosthesis implantation. More than 60% of explanted valves show evidence of mild pannus at the tissue-valve interface. In those with severe pannus formation, leaflet dysfunction can occur largely due to leaflet stiffness and restricted mobility. An important distinction must be made between pannus and thrombus, with qualitative and quantitative ultrasound intensity on transesophageal echocardiography helping differentiate them (36). In patients with bioprosthetic valve regurgitation, clinicians must clarify whether it is transvalvular or paravalvular regurgitation. Although the former may be successfully treated using V-in-V therapy, the latter is not suitable for such techniques. Patients with bioprosthetic paravalvular
Figure 2
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
regurgitation who are deemed unsuitable for reoperation may, however, benefit from other percutaneous techniques to occlude the paravalvular leak (37). Technical Considerations Transcatheter systems for V-in-V implantation. To date, 3 devices have been described in the setting V-in-V implantation. 1) The Edwards SAPIEN valve (Edwards Lifesciences) is balloon expandable, composed of a metal stent frame with bovine pericardial leaflets crimped onto a balloon catheter (15,16) (Fig. 2). Initial models used stainless steel frames, whereas recent versions use a lower profile cobalt chromium frame (38). The current SAPIEN XT
The 3 Types of Widely Available Transcatheter Systems
Edwards SAPIEN valve (A) and Edwards Novaflex delivery system (with Edwards SAPIEN XT valve) (B), Medtronic CoreValve (C), Medtronic CoreValve delivery system (D), Medtronic Melody valve (E), and Medtronic Melody delivery system (F).
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
valve is manufactured with expanded external diameters of 20, 23, 26, and 29 mm, allowing treatment of failed bioprostheses in all positions. 2) The CoreValve system (Medtronic Inc.) is composed of a self-expandable nitinol multilevel frame with porcine pericardial leaflets (Fig. 2). This valve is available in external diameters of 26 and 29 mm, with additional sizes anticipated. The device is deployed by retraction of a sheath. The valve can only be mounted in 1 direction within the restraining sheath; hence, the current device can only be delivered in a retrograde orientation via a transfemoral or subclavian/axillary/ transaortic approach. The long frame limits its utility for V-in-V implantation for failed aortic bioprostheses. 3) The Melody transcatheter valve (Medtronic Inc.) is composed of a bovine jugular venous valve attached to a platinum iridium stent scaffold delivered using a balloon-in-balloon system to facilitate positioning during expansion (Fig. 2). The valve is designed for use in the pulmonary circulation, mostly to treat dysfunctional right ventricular outflow tract conduits or other pulmonary bioprostheses in patients with congenital heart disease. Manufacturer sizing and labeling of surgical bioprosthetic heart valves. The methodologies for labeling valve sizes are not standardized and vary by the different manufacturers.
Internal diameters vary markedly for a given labeled size (39,40). The labeling of valve sizes may refer to internal or outer diameters of the valve (in the case of stented valves, Fig. 1) or the external diameter for most stentless valves. In most cases, the labeled size of a stented valve refers to the stent outer diameter (apart from the Soprano valve in which the label reflects the internal diameter). Different bioprostheses with the same label size frequently have different internal diameters and external sewing ring diameters. The external sewing ring diameter generally reflects the original native annulus diameter as measured by the surgeon at the time of the original valve implantation. The exact internal dimensions of a surgical bioprosthesis are those that are most relevant for V-in-V therapy, and in the case of most stented valves, these diameters are significantly smaller than the labeled valve size. Most stented bioprostheses with labeled diameters of 19 to 21 mm may have internal diameters that are too small to achieve acceptable hemodynamics with V-in-V therapy. Given the heterogeneity of valve types and sizing nomenclature, operators contemplating V-in-V therapy must familiarize themselves with the structural elements and dimensions of the specific bioprosthesis that they are treating. Obtaining a detailed operative report is critical. Tables 1, 2, and 3
per Manufacturer Valve Dimensions for Selected Information 18-Selected to 23-mm Stented Surgical Bioprostheses, Valve Product Dimensions for 18to 23-mm Stented Surgical Bioprostheses, Table 1 per Manufacturer Product Information Valve Label Size
Valve Type/Model (Manufacturer)
Sewing Ring External Diameter, mm
Stent Outer Diameter, mm
Stent Internal Diameter, mm
18
Soprano (Sorin Biomedica)
26
21
17.8
19
Magna (Edwards Lifesciences)
24
19
18
Perimount (Edwards Lifesciences)
26
19
18
Mosaic (Medtronic)
25
19
17.5
Hancock Ultra (Medtronic)
24
19
17.5
Hancock II (Medtronic)
N/A
N/A
N/A
Mitroflow (Sorin Biomedica)
21
18.6
15.4
20 21
22 23
Trifecta (St. Jude Medical)
24
19
N/a
Epic/Biocor (St. Jude Medical)
N/A
N/A
N/A
Epic Supra/Biocor Supra (St. Jude Medical)
N/A
N/A
N/A
Soprano (Sorin Biomedica)
28
23
19.8
Magna (Edwards Lifesciences)
26
21
20
Perimount (Edwards Lifesciences)
29
21
20
Mosaic/Hancock II (Medtronic)
27
21
18.5
Hancock/Hancock Ultra (Medtronic)
26
21
18.5
Mitroflow (Sorin Biomedica)
23
20.7
17.3
Trifecta (St. Jude Medical)
26
21
N/A
Epic/Biocor (St. Jude Medical)
N/A
21
19
Epic Supra/Biocor Supra (St. Jude Medical)
N/A
21
21
Soprano (Sorin Biomedica)
30
25
21.7
Magna (Edwards Lifesciences)
28
23
22
Perimount (Edwards Lifesciences)
31
23
22
Mosaic/Hancock II (Medtronic)
30
23
20.5
Hancock/Hancock Ultra (Medtronic)
28
23
22
Mitroflow (Sorin Biomedica)
26
22.7
19
Trifecta (St. Jude Medical)
28
23
N/A
Epic/Biocor (St. Jude Medical)
N/A
23
21
Epic Supra/Biocor Supra (St. Jude Medical)
N/A
23
23
N/A ⫽ not available.
2199
2200
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
provide the relevant dimensions for a range of commonly implanted surgical bioprostheses. In general, a V-in-V implant is chosen with a nominal external diameter matching or exceeding the reported internal diameter of the failed surgical valve to allow for secure fixation and sealing. It is important to consider that excess V-in-V oversizing may be associated with significant underexpansion the newly implanted valve, resulting in compromised hemodynamics and durability. The acceptable boundaries of underexpansion remain undefined in this context, although clinical experience suggests that a range of 10% to 15% may be acceptable. The mechanism of failure of the initial valve deserves consideration. Regurgitant valves with torn leaflets may have relatively larger internal diameters, whereas valves with prominent pannus or calcification may have reduced internal dimensions. Sizing consideration needs to take into account the reported internal diameter, bulkiness of the degenerated valve, nature of the valve failure, and location of calcification or pannus. Pre-procedural screening with both computed tomography and transesophageal echocardiography may help quantify discrepancies between in vitro and in vivo bioprosthetic internal diameters; provide information regarding valve bulkiness, calcification, and pannus; and
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
define the internal diameters when the specific valve model or size is unknown. Access route. Depending on the valve location and the delivery system, a number of approaches may be considered. Aortic V-in-V implantation may be performed using transapical access, which may facilitate crossing a stenotic valve and coaxial positioning, or various transarterial approaches that may avoid the need for a thoracotomy and a general anesthetic. Mitral V-in-V implantation appears ideally suited, at least anatomically, for the transapical approach, although the more circuitous transvenoustransseptal approach has been successfully used and avoids the need for a thoracotomy and possibly the need for general anesthesia (41– 46). Pulmonary conduits are ideally suited for percutaneous femoral, subclavian, or internal jugular venous access depending on the characteristics of the particular delivery system (47– 49). Tricuspid V-in-V implantation has been achieved using percutaneous transvenous access from the superior vena cava and the femoral vein, as well as using an open transatrial approach with direct right atrial puncture (44,48,50,51). The role of pre-dilation. Balloon pre-dilation may be performed to facilitate crossing or positioning during THV implantation. However, degenerated bioprostheses are often
per Manufacturer Valve Dimensions for Selected Information 24-Selected to 29-mm Stented Surgical Bioprostheses, Valve Product Dimensions for 24to 29-mm Stented Surgical Bioprostheses, Table 2 per Manufacturer Product Information Valve Label Size
Valve Type/Model (Manufacturer)
Sewing Ring External Diameter, mm
Stent Outer Diameter, mm
Stent Internal Diameter, mm
24
Soprano (Sorin Biomedica)
32
27
23.7
25
Magna (Edwards Lifesciences)
28
23
22
Perimount (Edwards Lifesciences)
31
23
22
Mosaic/Hancock II (Medtronic)
30
23
20.5
Mosaic Ultra/Hancock I Ultra (Medtronic)
30
25
22.5
Mitroflow (Sorin Biomedica)
29
25.1
21
Trifecta (St. Jude Medical)
31
25
N/A
Epic/Biocor (St. Jude Medical)
N/A
25
23
Epic Supra (St. Jude Medical)
N/A
25
25
26
Soprano (Sorin Biomedica)
35
29
25.6
27
Magna (Edwards Lifesciences)
32
27
26
Perimount (Edwards Lifesciences)
35
27
26
Mosaic/Hancock II (Medtronic)
36
27
24
Mosaic Ultra/Hancock II Ultra (Medtronic)
32
27
24
Mitroflow (Sorin Biomedica)
31
27.3
22.9
Trifecta (St. Jude Medical)
33
27
N/A
Epic/Biocor (St. Jude Medical)
N/A
27
27
Epic Supra (St. Jude Medical)
N/A
27
27
28
Soprano (Sorin Biomedica)
38
31
27.6
29
Magna (Edwards Lifesciences)
34
29
28
Perimount (Edwards Lifesciences)
37
29
28
Mosaic/Hancock II (Medtronic)
39
29
26
Mosaic Ultra/Hancock II Ultra (Medtronic)
34
29
26
Mitroflow (Sorin Biomedica)
33
29.5
24.7
Trifecta (St. Jude Medical)
35
29
N/A
Epic/Biocor (St. Jude Medical)
N/A
29
27
Epic Supra (St. Jude Medical)
N/A
N/A
N/A
N/A ⫽ not available.
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
2201
per Manufacturer Valve Dimensions for Selected Information Stentless Valves, Valve Product Dimensions for Selected Stentless Valves, Table 3 per Manufacturer Product Information Valve Label Size 19
21
23
25
Bioprosthesis
Manufacturer
Outer Diameter, mm
Internal Diameter, mm
Freestyle
Medtronic
19
16
Prima Plus
Edwards Lifesciences
19
16
3F Therapeutics
ATS Medical
19
17
Toronto SPV
St. Jude Medical
N/A
N/A
Pericarbon Freedom
Sorin Biomedica
19
17
Freestyle
Medtronic
21
18
Prima Plus
Edwards Lifesciences
21
18
3F Therapeutics
ATS Medical
21
19
Toronto SPV
St. Jude Medical
21
18
Pericarbon freedom
Sorin Biomedica
21
19
Freestyle
Medtronic
23
20
Prima Plus
Edwards Lifesciences
23
20
3F Therapeutics
ATS Medical
23
21
Toronto SPV
St. Jude Medical
23
20
Pericarbon Freedom
Sorin Biomedica
23
21
Freestyle
Medtronic
25
21
Prima Plus
Edwards Lifesciences
25
21
3F Therapeutics
ATS Medical
25
23
Toronto SPV
St. Jude Medical
25
21
Pericarbon Freedom
Sorin Biomedica
25
23
N/A ⫽ not available.
bulky and friable. Debris embolization is of particular concern for left-sided bioprostheses, and leaflet tears may result in acute severe regurgitation, which may be problematic if there is any delay in implanting the THV. Predilation may be necessary when a retrograde technique is used for a severely stenosed and calcified valve (e.g., transarterial crossing of an aortic bioprosthesis). Predilation is generally not necessary with antegrade crossing of a stenotic bioprosthesis as the funneled inflow and leaflets facilitate THV crossing. Similarly, predilation is not generally necessary with severely regurgitant bioprostheses, regardless of the access mode or the type of THV chosen. Device selection for V-in-V implantation. No doubt many of the valves currently undergoing evaluation or under development will be used for future V-in-V implantations. To date only 3 types of THVs have been used in this setting. The Edwards SAPIEN/XT and the Medtronic CoreValve devices have been widely used in the aortic position. The Medtronic Melody valve has also been used, but the durability of this venous valve in the systemic circulation is doubtful.
AORTIC.
MITRAL. Only the SAPIEN/XT valves have been used in the mitral position to our knowledge (Fig. 3). The design of the current CoreValve system is not suited to mitral implantation and the Melody device would not be anticipated to be sufficiently durable in the systemic circulation. PULMONARY. The Melody and SAPIEN/XT (Fig. 4) valves have been used in this position. Pulmonary valve disease is mostly part of more complex congenital structural disease,
with pulmonary conduits of relatively low pressure and mostly tubular, often in patients who have undergone multiple previous procedures and hence ideally suited to V-in-V therapy. The Edwards SAPIEN/XT (Fig. 5) and Melody valves have been used in this position. Positioning principles and deployment. For secure deployment without embolization, the THV is ideally positioned to overlap the surgical valve annular sewing ring (Fig. 6). To achieve this, the exact type of bioprosthesis and its characteristics must be known, including whether it is stented or stentless, dimensions, and radiologic appearance (Fig. 7). Some valves may have radiopaque markers or rings at the basal portion close to the anatomic sewing ring (Fig. 8), whereas in others, the markers may be close to the tip of the stent posts or there may be no radiopaque markers at all. Accurate placement requires coaxial positioning within the valve, which in turn requires determining the appropriate C-arm angulation representing perpendicularity to the valve plane (52). Fluoroscopically visible components of surgical valves can be used effectively to achieve perpendicularity. If the radiographic marker is a basal ring, rotating the C arm until the ring is side on (appearing as a straight “line”) suggests perpendicularity to the valve plane and an appropriate implant view (Fig. 9). The typical aortic valve is directed slightly to the right and posterior, such that left anterior oblique-cranial views are often helpful. The typical mitral valve is directed to the left and anterior, such that right anterior oblique views are often helpful. Further TRICUSPID.
2202
Figure 3
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
Mitral Valve-in-Valve
(A) Fluoroscopic image of a regurgitant Carpentier Edwards valve in the mitral position, with the stent frame radiopaque (white arrow). The valve is perpendicular to the image intensifier. (B) Same valve is shown “down the barrel.” This would be a poor angiographic projection in which to position a transcatheter valve. (C) Perpendicular projection after implantation of an Edwards SAPIEN valve. (D) “Down the barrel” projection after implantation of an Edwards SAPIEN valve (black arrow) inside the old valve (white arrow).
manipulation of the delivery system or guidewire may then be required to achieve coaxial alignment. For radiolucent valves (e.g., stentless) angiography with a pigtail catheter in one of the valve cusps and transesophageal echocardiographic guidance may be helpful. When treating failed pulmonic homografts/conduits, pre-stenting with a bare metal stent may provide additional radial strength and reduce the risk of stent fracture with the Melody valve. Similarly, pre-stenting may provide a longer “landing zone” with a greater margin of safety during positioning with the relatively short Edwards SAPIEN/XT valve (less necessary when treating failed stented bioprostheses that may provide an adequate support structure). Pre-stenting with a covered stent may also reduce the risk associated with conduit rupture when this is a concern. Pacing. The Edwards valves have generally been deployed with rapid ventricular pacing to aid accurate positioning of the relatively short valve. This may not be required in the lower pressure venous system; although, if cardiac motion is problematic, we have used either rapid pacing or induced ventricular fibrillation. The Medtronic Melody valve is generally deployed without rapid pacing in the more commonly treated pulmonary and tricuspid
positions (50), but with pacing in the aortic position. Rapid pacing is generally not used when using the CoreValve device, although this may be of value during difficult deployment, particularly when treating a severely regurgitant bioprosthesis.
Results Early pre-clinical studies. Boudjemline et al. (53) evaluated the concept of mitral V-in-V therapy in a sheep model. A bovine jugular valve was mounted onto a stent and implanted off-pump using a transatrial approach. Walther et al. (54) evaluated the use of a more contemporary transcatheter system in both the aortic and mitral positions in pigs. Edwards THVs were implanted into Carpentier Edwards xenografts transapically in the beating heart model. Azadani et al. (55) evaluated the hemodynamic performance of 23-mm THVs within degenerated surgical bioprostheses. They demonstrated in an in vitro model that incomplete stent expansion resulted in leaflet distortion and central regurgitation when implanted in 19- and 21-mm bioprostheses. In a subsequent report, the same group used a
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
Figure 4
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
2203
Pulmonic Valve-in-Valve
(A) A Carpentier Edwards Perimount 27-mm valve (a left pulmonary artery stent is also seen) in a 21-year-old man with repaired tetralogy of Fallot. The valve was severely regurgitant with moderate to severe stenosis. (B) An Edwards SAPIEN 26-mm valve is being positioned. (C) Post-deployment pulmonary angiogram demonstrating no residual pulmonary regurgitation. (D) Final image post-implantation.
custom-designed supravalvular THV (56), achieving more favorable hemodynamics. Valve hemodynamics post–V-in-V. THV expansion during V-in-V therapy is constrained by the internal dimen-
Figure 5
sions of the bioprosthetic valve, especially by the relatively undilatable sewing ring (Fig. 8F). Although this may reduce the chance of annular rupture, heart block, and coronary occlusion compared with therapy with native aortic valve
Tricuspid Valve-in-Valve
(A) A stenosed and regurgitant tricuspid Mitroflow Synergy 27-mm valve is seen perpendicular to the image intensifier (a bileaflet mechanical valve in the mitral position is also seen above it). (B) A sheath is introduced via the right atrium through a mini right thoracotomy. An Edwards SAPIEN 26-mm valve is being positioned. (C) Final image after implantation.
2204
Figure 6
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
In Vitro Demonstration of a Transcatheter Valve (Edwards SAPIEN) Implanted Within a Carpentier Edwards Valve
(A) Incorrect positioning: the transcatheter valve is implanted too high within the outflow tract of the surgical valve. This may result in splaying of the surgical valve posts and transcatheter valve embolization. (B) Correct valve positioning: the transcatheter valve (arrow) is implanted such that it overlaps the surgical sewing ring, allowing better anchoring and a more secure position. Reprinted, with permission, from Webb et al. (46).
stenosis, valve underexpansion may contribute to hemodynamically significant residual gradients after V-in-V implantation. Table 4 provides a summary of the hemodynamic outcomes associated with V-in-V implantation (57– 66). Valve underexpansion may be expected to affect transvalvular gradients, effective orifice areas, and leaflet and stent durability. Post-procedural transaortic mean gradients are in the range of 10 to 22 mm Hg in most series (range: 4 to 30 mm Hg). In comparison, mean gradients after transcatheter therapy for native aortic stenosis are generally around 10 mm Hg (67,68). Although elevated gradients may be acceptable in some patients who cannot undergo open-heart surgery, they may be inadequate in others with longer expected survival. There are currently insufficient long-term outcome data to understand the clinical sequelae of such residual gradients. From our own experience, we have yet to see patients
Figure 7
Fluoroscopic Appearance of Various Stented Surgical Bioprostheses
Carpentier Edwards (CE) Perimount Standard (A), CE Porcine SupraAnnular (B), CE pericardial (the radiopaque base ring has 3 “holes”) (C), Hancock II (D), CE Porcine (E), Mosaic (radiopaque rings at the top of each stent post only) (F), Mitroflow (G), Epic (very faint base ring only) (H).
fail to improve or return for further intervention because of elevated gradients. For example, 2 patients from our cohort had transaortic gradients of 24 and 27 mm Hg, and at a follow-up of 610 and 540 days, respectively; both remain improved and relatively asymptomatic. However, it is likely that some patients may fail to obtain adequate symptomatic improvement if sufficiently low transvalvular gradients are not obtained. V-in-V designs that minimize residual gradients will become increasingly important as younger and lower risk patients are treated in the future.
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
Figure 8
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
2205
Fluoroscopic Positioning for V-in-V Implantation for Bioprosthetic Aortic Valve Failure
Importance of knowing the radiological appearance of the surgical valve treated. (A) Carpentier Edwards pericardial valve (the wire frame within the valve posts is visible, although the rigid sewing ring below this required for V-in-V fixation is radiolucent in this model). (B) Positioning of the Edwards SAPIEN just below the lowest radiopaque portion (arrow indicates lowest portion of the SAPIEN valve). (C) Sorin Mitroflow valve. (D) Positioning the Edwards SAPIEN just below the lowest radiopaque portion. (E) Medtronic Mosaic valve: the radiopaque markers are near the top of the surgical stent posts (black arrows), hence the valve is positioned completely below these markers (white arrows). (F) Deployed Edwards SAPIEN valve. The “waist” at the lower part of the implanted valve demonstrates the narrowest location of the surgical valve (white arrow). The SAPIEN valve remained slightly underexpanded, and the residual mean gradient was 30 mm Hg.
Regurgitation post–V-in-V. Although paravalvular leaks are common after transcatheter aortic valve replacement for native aortic stenosis, regurgitation appears to be absent or mild in most published V-in-V reports thus far. The
Figure 9
circular sewing ring of surgical bioprostheses appears to facilitate intervalvular sealing. We suggest reporting 3 sources of regurgitation in the context of V-in-V therapy (Fig. 10):
Perpendicular Alignment for Implantation
Flouroscopic images of a Carpentier Edwards pericardial valve in the mitral position. (A) In the posterior-anterior projection, the valve is not perpendicular to the image intensifier, leading to foreshortening. (B) In the 25o right anterior oblique projection, the valve is now perpendicular/orthogonal to the image intensifier, allowing more accurate positioning of a transcatheter valve.
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
2206
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
Clinical Hemodynamic Outcomes ofOutcomes V-in-V Implantation Table 4and Clinical and Hemodynamic of V-in-V Implantation Valve Position First Author (Ref. #)
No. of Valves Reported
THV Implanted
Residual Transvalvular Gradient, mm Hg
30-Day Survival
Aortic Khawaja et al. (42)
4
CoreValve
PG 30–39
Maroto et al. (57)
2
Edwards SAPIEN
PG 13 and 17
100% Not reported
14
Edwards SAPIEN
MG 3–23 (mean MG 13.1 ⫾ 6.4)
86%
Seiffert et al. (59)
4
Edwards SAPIEN
MG 12–30 (mean MG 19.0 ⫾ 12.4)
75%
Kempfer et al. (60)
11
Edwards SAPIEN
MG 11 ⫾ 4
100%
CoreValve
Mean MG 16.4 ⫾ 3.6
100%
Edwards SAPIEN
Mean MG 20.2 ⫾ 6.7
100% 100%
Pasic et al. (58)
Gotzmann et al. (61) Webb et al. (46)
5 10
Mitral Cheung et al. (43)
11
Edwards SAPIEN
Mean MG 7 (range 5–8)
Seiffert et al. (59)
1
Edwards SAPIEN
MG 3
0%
De Weger et al. (62)
1
Edwards SAPIEN (into a failed mitral annuloplasty ring)
MG 4
100%
Cerillo et al. (44)
3
Edwards SAPIEN
MG 4–9
75%
Tricuspid Roberts et al. (48)
Melody
MG 3.9
Webb et al. (46)
15 1
Edwards SAPIEN
MG 4
93%
Cerillo et al. (44)
1
Edwards SAPIEN
MG 9
100%
Van Garsee et al. (63)
1
Edwards SAPIEN
MG 2
Not reported
100%
Pulmonic Zahn et al. (64)
34
Melody
Mean MG 17.3 ⫾ 7.3
Khambadkone et al. (65)
59
Early version of the Melody valve
Mean MG 19.5 ⫾ 15.3
100%
Edwards SAPIEN
MG 14.9 ⫾ 6.9
100%
Melody
Median peak 12
99%
Boone et al. (66) McElhinney et al. (47)
7 124
100%
MG ⫽ mean gradient; PG ⫽ peak gradient; THV ⫽ transcatheter heart valve.
1. Paravalvular: arising between the native valve and the original failed bioprosthetic valve. THV implantation is unlikely to ameliorate this, and paravalvular leak closure with an occlusive plug may be an option. 2. Intervalvular: arising from the area between the old “outer” bioprosthetic valve and the new “inner” transcatheter valve. This may occur from incomplete THV expansion (undersizing, severe calcification of bioprosthetic valve leaflets). This has not been a major issue in published series thus far. 3. Transvalvular: arising through the newly implanted THV. This may occur from incomplete leaflet coaptation due to an oversized and deformed THV or a THV leaflet “stuck” in an open position. This has not been a major issue in published series thus far. Coronary concerns. Most stented surgical bioprostheses have leaflet tissue mounted internal to the stent frame (Fig. 11). To some extent, this ensures that degenerated bioprosthetic tissue will not be displaced external to the valve posts. When bioprosthetic tissue is mounted external to the valve frame or there is no stent frame, there is no such assurance. For example, acute coronary ostial occlusion by displaced externally mounted pericardial tissue has been reported after V-in-V implantation within the Mitroflow valve (Sorin Group, Burnaby, British Columbia, Canada) (69), although many successful and uncomplicated such implantations have also been performed. Coronary obstruction has similarly been observed with stentless prostheses. It is reasonable to think that post–V-in-V implantation coronary compromise might also occur with other failed bioprostheses that have externally mounted leaflets, are nonstented, implanted in a
supra-annular position, or have unusually bulky diseased leaflets, particularly in patients with low-lying coronary ostia and shallow sinuses. When considering aortic V-in-V im-
Figure 10
Possible Sources of Regurgitation After V-in-V Implantation
“Down the barrel/end on” image of an Edwards SAPIEN valve (white arrow) implanted within a Medtronic Mosaic valve (black arrows indicate the radiopaque markers). The large surrounding circle represents surrounding native tissue. Regurgitation may arise from 3 areas depicted schematically by the 3 black circles: 1, paravalvular, arising between the native tissue and the original failed bioprosthetic valve; 2, intervalvular, arising from the area between the old “outer” surgical valve and the new “inner” transcatheter valve; and 3, transvalvular, arising from within the transcatheter valve.
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209
Figure 11
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
2207
Selected Stented Surgical Bioprosthetic Valves
Carpentier Edwards Perimount (A), Medtronic Mosaic (B), St. Jude Medical EPIC (C), and Sorin Medical Mitroflow (D) valves. In A to C, the leaflets are inside the stent frame. In D, the leaflets are outside of the stent frame (the stent frame is not visible).
plants, the relationship of the failed valve to the coronary ostia must be carefully evaluated. Conclusions Initial results with V-in-V therapy are very encouraging. However, in the absence of rigorous evaluation and longterm follow-up, V-in-V therapy is probably best considered only for patients who present with a prohibitive reoperative risk. Therapy of small (e.g., ⱕ21-mm aortic valves) should be approached with caution as significant residual gradients may remain with currently available valves. Operators should be encouraged to share their experience, whether favorable or unfavorable. Future technologic advances may continue to improve both hemodynamic and clinical outcomes. Reprint requests and correspondence: Dr. John G. Webb, St. Paul’s Hospital, 1081 Burrard Street, Vancouver, British Columbia V6Z 1Y6, Canada. E-mail:
[email protected].
REFERENCES
1. Gummert JF, Funkat A, Beckmann A, et al. Cardiac surgery in Germany during 2008. A report on behalf of the German Society for Thoracic and Cardiovascular Surgery. Thorac Cardiovasc Surg 2009; 57:315–23. 2. Ruel M, Chan V, Bedard P, et al. Very long-term survival implications of heart valve replacement with tissue versus mechanical prostheses in adults ⬍60 years of age. Circulation 2007;116:I294 –300.
3. Maganti M, Rao V, Armstrong S, Feindel CM, Scully HE, David TE. Redo valvular surgery in elderly patients. Ann Thorac Surg 2009;87: 521–5. 4. Balsam LB, Grossi EA, Greenhouse DG, et al. Reoperative valve surgery in the elderly: predictors of risk and long-term survival. Ann Thorac Surg 2010;90:1195–200, discussion 1201. 5. Jones JM, O’kane H, Gladstone DJ, et al. Repeat heart valve surgery: risk factors for operative mortality. J Thorac Cardiovasc Surg 2001; 122:913– 8. 6. Akins CW, Buckley MJ, Daggett WM, et al. Risk of reoperative valve replacement for failed mitral and aortic bioprostheses. Ann Thorac Surg 1998;65:1545–51, discussion 1551–2. 7. Albeyoglu SC, Filizcan U, Sargin M, et al. Determinants of hospital mortality after repeat mitral valve surgery for rheumatic mitral valve disease. Thorac Cardiovasc Surg 2006;54:244 –9. 8. Leontyev S, Borger MA, Davierwala P, et al. Redo aortic valve surgery: early and late outcomes. Ann Thorac Surg 2011;91:1120 – 6. 9. Beurtheret S, Gariboldi V, Feier H, et al. Short-term results of repeat valve replacement: a predictive factor analysis. J Heart Valve Dis 2010;19:326 –32. 10. Lytle BW, Cosgrove DM, Taylor PC, et al. Reoperations for valve surgery: perioperative mortality and determinants of risk for 1,000 patients, 1958-1984. Ann Thorac Surg 1986;42:632– 43. 11. Piehler JM, Blackstone EH, Bailey KR, et al. Reoperation on prosthetic heart values. Patient-specific estimates of in-hospital events. J Thorac Cardiovasc Surg 1995;109:30 – 48. 12. Kirsch M, Nakashima K, Kubota S, Houel R, Hillion ML, Loisance D. The risk of reoperative heart valve procedures in Octogenarian patients. J Heart Valve Dis 2004;13:991– 6, discussion 996. 13. Cribier A, Eltchaninoff H, Bash A, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation 2002;106:3006 – 8. 14. Webb JG, Altwegg L, Masson JB, Al Bugami S, Al Ali A, Boone RA. A new transcatheter aortic valve and percutaneous valve delivery system. J Am Coll Cardiol 2009;53:1855– 8.
2208
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
15. Webb JG, Chandavimol M, Thompson CR, et al. Percutaneous aortic valve implantation retrograde from the femoral artery. Circulation 2006;113:842–50. 16. Lichtenstein SV, Cheung A, Ye J, et al. Transapical transcatheter aortic valve implantation in humans: initial clinical experience. Circulation 2006;114:591– 6. 17. Walther T, Dewey T, Borger MA, et al. Transapical aortic valve implantation: step by step. Ann Thorac Surg 2009;87:276 – 83. 18. Webb JG. Percutaneous aortic valve replacement will become a common treatment for aortic valve disease. J Am Coll Cardiol Intv 2008;1:122– 6. 19. Piazza N, Grube E, Gerckens U, et al. Procedural and 30-day outcomes following transcatheter aortic valve implantation using the third generation (18 Fr) corevalve revalving system: results from the multicentre, expanded evaluation registry 1-year following CE mark approval. EuroIntervention 2008;4:242–9. 20. Tamburino C, Capodanno D, Mule M, et al. Procedural success and 30-day clinical outcomes after percutaneous aortic valve replacement using current third-generation self-expanding CoreValve prosthesis. J Invasive Cardiol 2009;21:93– 8. 21. Hilbert SL, Ferrans VJ. Porcine aortic valve bioprostheses: morphologic and functional considerations. J Long Term Eff Med Implants 1992;2:99 –112. 22. Schoen FJ, Levy RJ. Calcification of tissue heart valve substitutes: progress toward understanding and prevention. Ann Thorac Surg 2005;79:1072– 80. 23. Kallikourdis A, Jacob S. Is a stentless aortic valve superior to conventional bioprosthetic valves for aortic valve replacement? Interact Cardiovasc Thorac Surg 2007;6:665–72. 24. Dunning J, Graham RJ, Thambyrajah J, Stewart MJ, Kendall SW, Hunter S. Stentless vs. stented aortic valve bioprostheses: a prospective randomized controlled trial. Eur Heart J 2007;28:2369 –74. 25. Perez de Arenaza D, Lees B, Flather M, et al. Randomized comparison of stentless versus stented valves for aortic stenosis: effects on left ventricular mass. Circulation 2005;112:2696 –702. 26. Chambers JB, Rimington HM, Hodson F, Rajani R, Blauth CI. The subcoronary Toronto stentless versus supra-annular Perimount stented replacement aortic valve: early clinical and hemodynamic results of a randomized comparison in 160 patients. J Thorac Cardiovasc Surg 2006;131:878 –2. 27. Vesey JM, Otto CM. Complications of prosthetic heart valves. Curr Cardiol Rep 2004;6:106 –11. 28. Jamieson WR, Munro AI, Miyagishima RT, Allen P, Burr LH, Tyers GF. Carpentier-Edwards standard porcine bioprosthesis: clinical performance to seventeen years. Ann Thorac Surg 1995;60:999 –1006, discussion 1007. 29. Jamieson WR, Rosado LJ, Munro AI, et al. Carpentier-Edwards standard porcine bioprosthesis: primary tissue failure (structural valve deterioration) by age groups. Ann Thorac Surg 1988;46:155– 62. 30. Ruel M, Kulik A, Rubens FD, et al. Late incidence and determinants of reoperation in patients with prosthetic heart valves. Eur J Cardiothorac Surg 2004;25:364 –70. 31. Jamieson WR, Miyagishima RT, Burr LH, Lichtenstein SV, Fradet GJ, Janusz MT. Carpentier-Edwards porcine bioprostheses: clinical performance assessed by actual analysis. J Heart Valve Dis 2000;9: 530 –5. 32. van Geldorp MW, Eric Jamieson WR, Kappetein AP, et al. Patient outcome after aortic valve replacement with a mechanical or biological prosthesis: weighing lifetime anticoagulant-related event risk against reoperation risk. J Thorac Cardiovasc Surg 2009;137:881– 6, 886e1-5. 33. Sacks MS, Schoen FJ. Collagen fiber disruption occurs independent of calcification in clinically explanted bioprosthetic heart valves. J Biomed Mater Res 2002;62:359 –71. 34. Vesely I, Barber JE, Ratliff NB. Tissue damage and calcification may be independent mechanisms of bioprosthetic heart valve failure. J Heart Valve Dis 2001;10:471–7. 35. Mohammadi S, Baillot R, Voisine P, Mathieu P, Dagenais F. Structural deterioration of the Freestyle aortic valve: mode of presentation and mechanisms. J Thorac Cardiovasc Surg 2006;132:401– 6. 36. Barbetseas J, Nagueh SF, Pitsavos C, Toutouzas PK, Quinones MA, Zoghbi WA. Differentiating thrombus from pannus formation in obstructed mechanical prosthetic valves: an evaluation of clinical, transthoracic and transesophageal echocardiographic parameters. J Am Coll Cardiol 1998;32:1410 –7.
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209 37. Nietlispach F, Johnson M, Moss RR, et al. Transcatheter closure of paravalvular defects using a purpose-specific occluder. J Am Coll Cardiol Intv 2010;3:759 – 65. 38. Nietlispach F, Wijesinghe N, Wood D, Carere RG, Webb JG. Current balloon-expandable transcatheter heart valve and delivery systems. Catheter Cardiovasc Interv 2010;75:295–300. 39. Christakis GT, Buth KJ, Goldman BS, et al. Inaccurate and misleading valve sizing: a proposed standard for valve size nomenclature. Ann Thorac Surg 1998;66:1198 –203. 40. Shibata T, Inoue K, Ikuta T, Bito Y, Yoshioka Y, Mizoguchi H. Which valve and which size should we use in the valve-on-valve technique for re-do mitral valve surgery? Interact Cardiovasc Thorac Surg 2009;8:206 –10. 41. Sharp AS, Michev I, Colombo A. First trans-axillary implantation of Edwards Sapien valve to treat an incompetent aortic bioprosthesis. Catheter Cardiovasc Interv 2010;75:507–10. 42. Khawaja MZ, Haworth P, Ghuran A, et al. Transcatheter aortic valve implantation for stenosed and regurgitant aortic valve bioprostheses CoreValve for failed bioprosthetic aortic valve replacements. J Am Coll Cardiol 2010;55:97–101. 43. Cheung AW, Gurvitch R, Ye J, et al. Transcatheter transapical mitral valve-in-valve implantations for a failed bioprosthesis: a case series. J Thorac Cardiovasc Surg 2011;141:711–5. 44. Cerillo AG, Chiaramonti F, Murzi M, et al. Transcatheter valve in valve implantation for failed mitral and tricuspid bioprosthesis. Catheter Cardiovasc Interv 2011 Apr 28 [E-pub ahead of print]; doi: 10.1002/ccd.23112. 45. Montorfano M, Latib A, Chieffo A, et al. Successful percutaneous anterograde transcatheter valve-in-valve implantation in the mitral position. J Am Coll Cardiol Intv 2011. In press. 46. Webb JG, Wood DA, Ye J, et al. Transcatheter valve-in-valve implantation for failed bioprosthetic heart valves. Circulation 2010; 121:1848 –57. 47. McElhinney DB, Hellenbrand WE, Zahn EM, et al. Short- and medium-term outcomes after transcatheter pulmonary valve placement in the expanded multicenter US melody valve trial. Circulation 2010;122:507–16. 48. Roberts PA, Boudjemline Y, Cheatham JP, et al. Percutaneous tricuspid valve replacement in congenital and acquired heart disease. J Am Coll Cardiol 2011;58:117–22. 49. Huber CH, Hurni M, Tsang V, von Segesser LK. Valved stents for transapical pulmonary valve replacement. J Thorac Cardiovasc Surg 2009;137:914 – 8. 50. Roberts P, Spina R, Vallely M, Wilson M, Bailey B, Celermajer DS. Percutaneous tricuspid valve replacement for a stenosed bioprosthesis. Circ Cardiovasc Interv 2010;3:e14 –5. 51. Kenny D, Hijazi ZM, Walsh KP. Transcatheter tricuspid valve replacement with the Edwards SAPIEN valve. Catheter Cardiovasc Interv 2011;78:267–70. 52. Gurvitch R, Wood DA, Leipsic J, et al. Multislice computed tomography for prediction of optimal angiographic deployment projections during transcatheter aortic valve implantation. J Am Coll Cardiol Intv 2010;3:1157– 65. 53. Boudjemline Y, Pineau E, Borenstein N, Behr L, Bonhoeffer P. New insights in minimally invasive valve replacement: description of a cooperative approach for the off-pump replacement of mitral valves. Eur Heart J 2005;26:2013–7. 54. Walther T, Falk V, Dewey T, et al. Valve-in-a-valve concept for transcatheter minimally invasive repeat xenograft implantation. J Am Coll Cardiol. 2007;50:56 – 60. 55. Azadani AN, Jaussaud N, Matthews PB, Ge L, Chuter TA, Tseng EE. Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses. Interact Cardiovasc Thorac Surg 2010;11: 70 –7. 56. Azadani AN, Jaussaud N, Matthews PB, et al. Valve-in-valve implantation using a novel supravalvular transcatheter aortic valve: proof of concept. Ann Thorac Surg 2009;88:1864 –9. 57. Maroto LC, Rodriguez JE, Cobiella J, Marcos P. Transapical offpump aortic valve-in-a-valve implantation in two elderly patients with a degenerated porcine bioprosthesis. Eur J Cardiothorac Surg 2010; 37:738 – 40. 58. Pasic M, Unbehaun A, Dreysse S, et al. Transapical aortic valve implantation after previous aortic valve replacement: clinical proof of the ”valve-in-valve” concept. J Thorac Cardiovasc Surg 2011;142:270–7.
JACC Vol. 58, No. 21, 2011 November 15, 2011:2196–209 59. Seiffert M, Franzen O, Conradi L, et al. Series of transcatheter valve-in-valve implantations in high-risk patients with degenerated bioprostheses in aortic and mitral position. Catheter Cardiovasc Interv 2010;76:608 –15. 60. Kempfert J, Van Linden A, Linke A, et al. Transapical off-pump valve-in-valve implantation in patients with degenerated aortic xenografts. Ann Thorac Surg 2010;89:1934 – 41. 61. Gotzmann M, Mugge A, Bojara W. Transcatheter aortic valve implantation for treatment of patients with degenerated aortic bioprostheses— valve-in-valve technique. Catheter Cardiovasc Interv 2010;76:1000 – 6. 62. de Weger A, Tavilla G, Ng AC, et al. Successful transapical transcatheter valve implantation within a dysfunctional mitral bioprosthesis. J Am Coll Cardiol Img 2010;3:222–3. 63. Van Garsse LA, Ter Bekke RM, van Ommen VG. Percutaneous transcatheter valve-in-valve implantation in stenosed tricuspid valve bioprosthesis. Circulation 2011;123:e219 –21. 64. Zahn EM, Hellenbrand WE, Lock JE, McElhinney DB. Implantation of the melody transcatheter pulmonary valve in patients with a dysfunctional right ventricular outflow tract conduit early results from the U.S. clinical trial. J Am Coll Cardiol 2009;54:1722–9.
Gurvitch et al. Valve-in-Valve Therapy for Failed Surgical Valves
2209
65. Khambadkone S, Coats L, Taylor A, et al. Percutaneous pulmonary valve implantation in humans: results in 59 consecutive patients. Circulation 2005;112:1189 –97. 66. Boone RH, Webb JG, Horlick E, et al. Transcatheter pulmonary valve implantation using the Edwards SAPIEN transcatheter heart valve. Catheter Cardiovasc Interv 2010;75:286 –94. 67. Grube E, Buellesfeld L, Mueller R, et al. Progress and current status of percutaneous aortic valve replacement: results of three device generations of the CoreValve Revalving system. Circ Cardiovasc Interv 2008;1:167–75. 68. Webb JG, Altwegg L, Boone RH, et al. Transcatheter aortic valve implantation: impact on clinical and valve-related outcomes. Circulation 2009;119:3009 –16. 69. Gurvitch R, Cheung A, Bedogni F, Webb JG. Coronary obstruction following transcatheter aortic valve-in-valve implantation for failed surgical bioprostheses. Catheter Cardiovasc Interv 2011;77:439 – 44. Key Words: failed surgical bioprosthesis y transcatheter aortic valve implantation y valve-in-valve.